<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-1740</title>
    <subtitle>WSI is missing Frame Type</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/03/27</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para>
                    <para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
           <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para>
                   <para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
           </tr>
           <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>2017/07/09</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1740</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: WSI is missing Frame Type</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6 2018a</para>
                </td>
            </tr>
            <tr>
               <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                   <para>WSI has no functional group to specify FrameType and hence there is uncertainty over
                       conditions that depend on it, e.g., Frame Content data time and duration attributes
                       that are conditioned on originality.</para>
                    <para>Add a modality-specific Frame Type functional group macro to be consistent with all the other enhanced multi-frame family IODs.</para>
               </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
              
<section label="A.32.8" status="3" xml:id="sect_A.32.8">
        <title>VL Whole Slide Microscopy Image IOD</title>
        <section label="A.32.8.4" status="4" xml:id="sect_A.32.8.4">
          <title>VL Whole Slide Microscopy Image Functional Group Macros</title>
          <para xml:id="para_c6dd0dd8-a562-406c-a95f-7d307b307f1d">
            <xref linkend="table_A.32.8-2" xrefstyle="select: label"/> specifies the use of the Functional Group Macros used in the Section C.7.6.16<!--<xref linkend="sect_C.7.6.16" xrefstyle="select: title"/>--> for the VL Whole Slide Microscopy Image IOD.</para>
          <table frame="box" label="A.32.8-2" rules="all" xml:id="table_A.32.8-2">
            <caption>VL Whole Slide Microscopy Image Functional Group Macros</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_661654ac-dde2-4b5b-b103-22db80f7df54">Functional Group Macro</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f0df0b22-b551-45da-b912-36013f2dec5a">Section</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1d9c0115-1d06-4a82-9f8b-b435811646e1">Usage</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_eb389343-b64d-4efe-a32f-71e091518fb8">Pixel Measures</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ea1313a9-1968-4f04-8cf9-5f7e50434db1">
                    C.7.6.16.2.1<!--<xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b6971130-c9fd-4a1d-b939-8d43782d685c">M - Shall be used as a Shared Functional Group.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fa7cff4f-5a24-4b9d-9643-63f2878c0b06">Frame Content</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_927096fd-ffe5-4351-a2e2-0dbf757b07dd">
                    <xref linkend="sect_C.7.6.16.2.2" xrefstyle="select: labelnumber"/>
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fa450bb3-cf94-4482-9690-a8a0953834a5">M - Shall not be used as a Shared Functional Group.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_06fee48c-e13a-4a73-a2b0-3f7095a05e60">Referenced Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ac09cd80-0a48-4a55-b984-01ea10623663">
                    C.7.6.16.2.5<!--<xref linkend="sect_C.7.6.16.2.5" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_370b6b45-d0dc-410f-b49e-a39bbe0e4425">U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9248e796-3b6a-43c3-9d69-586501a1e427">Derivation Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b906be5b-b845-44e6-9724-a5a4f83f1afe">
                    C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e1a55ba2-e0d7-4c80-a9c9-5388750afd25">C - Required if the image or frame has been derived from another SOP Instance.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cf24da64-e959-4b71-aec1-87735508871e">Real World Value Mapping</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_697ef855-597d-464c-aa18-9bb3e0f2e5e7">
                    C.7.6.16.2.11<!--<xref linkend="sect_C.7.6.16.2.11" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d4f78b1f-ad51-4f90-9a8f-22a269acc18a">U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e9bebd70-7c3c-4d05-af0c-5393b389c5ea">Plane Position (Slide)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f0b762cb-8165-40f8-be2d-165a8c78dde2">
                    C.8.12.6.1<!--<xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b8936e04-89ca-4496-a36d-f7bfa6c6c7b5">M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a0eaee81-b5a4-4a0d-b33b-c95cf2b648c9">Optical Path Identification</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_72be2f57-a60c-4710-b61d-24465bc4a311">
                    C.8.12.6.2<!--<xref linkend="sect_C.8.12.6.2" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6ec616ba-a4ef-4b20-82c4-ca8bee1a3f30">M</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f50e43da-7a34-46de-932c-cc33b8fc980e">Specimen Reference</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bff977d3-c4c7-4996-a47c-288de4cef85d">
                    C.8.12.6.3<!--<xref linkend="sect_C.8.12.6.3" xrefstyle="select: labelnumber"/>-->
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_18241086-a4a4-49df-ba81-884fc96f6144">U</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para><emphasis role="bold"><emphasis role="underline">Whole Slide Microscopy Image Frame Type</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>
                    <xref linkend="sect_C.8.12.9" xrefstyle="select: labelnumber"/>
                  </para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para><emphasis role="bold"><emphasis role="underline">M - Shall be used as a Shared Functional Group.</emphasis></emphasis></para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      </section>
      
      
      
      <section label="C.8.12.9" status="4" xml:id="sect_C.8.12.9">
        <title><emphasis role="bold"><emphasis role="underline">Whole Slide Microscopy Image Frame Type Macro</emphasis></emphasis></title>
        <para>
        <emphasis role="bold"><emphasis role="underline"><xref linkend="table_C.8.12.9-1" xrefstyle="select: label"/> specifies the attributes of the Whole Slide Microscopy Image Frame Type Functional Group Macro.</emphasis></emphasis></para>
        <table frame="box" label="C.8.12.9-1" rules="all" xml:id="table_C.8.12.9-1">
          <caption><emphasis role="bold"><emphasis role="underline">Whole Slide Microscopy Image Frame Type Macro Attributes</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Attribute Name</emphasis></emphasis></para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Tag</emphasis></emphasis></para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Type</emphasis></emphasis></para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Attribute Description</emphasis></emphasis></para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Whole Slide Microscopy Image Frame Type Sequence</emphasis></emphasis></para>
                </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">(0040,0710)</emphasis></emphasis></para>
                </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Identifies the characteristics of this Whole Slide Microscopy Image frame.</emphasis></emphasis></para>
                <para><emphasis role="bold"><emphasis role="underline">Only a single Item shall be included in this Sequence.</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;Frame Type</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">(0008,9007)</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Type of Frame. A multi-valued attribute analogous to Image Type (0008,0008).</emphasis></emphasis></para>
                <para><emphasis role="bold"><emphasis role="underline">Enumerated Values and Defined Terms are the same as those for the four values of Image Type (0008,0008). See <xref linkend="sect_C.8.12.4.1" xrefstyle="select: label"/>.</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">For reference, unchanged:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="C.7.6.16.2.2" status="6" xml:id="sect_C.7.6.16.2.2">
              <title>Frame Content Macro</title>
              <para xml:id="para_2797cdd6-abe6-4e70-a1f8-c1b965189d4c">
<xref linkend="table_C.7.6.16-3" xrefstyle="select: label"/> specifies the attributes of the Frame Content Functional Group Macro.</para>
              <para xml:id="para_c05df9f4-9a80-4060-95e7-99b39556fd21">This Functional Group Macro may only be part of the Per-frame Functional Groups Sequence (5200,9230) attribute.</para>
              <table frame="box" label="C.7.6.16-3" rules="all" xml:id="table_C.7.6.16-3">
                <caption>Frame Content Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_524ebe21-f60a-44d2-9648-e16379da4dce">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_7fd19267-d36c-4cf5-a83c-2c238303ce38">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_f38a031a-bf90-45fb-86b0-4d9633671d73">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0467d156-3cf3-43ae-a201-93f7efa5c0d2">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_2cbd10d3-95f9-469c-b795-4aae3e62f9a3">Frame Content Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_11f48380-d331-45fe-9341-78ddd8617369">(0020,9111)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3e3dafc7-4788-43af-a588-eff012eb150d">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_47f02e7a-77c5-4bf3-9dd4-6176e50f2782">Identifies general characteristics of this frame.</para>
                      <para xml:id="para_8c19b021-9574-49de-9cea-6c45d8e229c5">Only a single Item shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a24dce14-f92c-4ab9-b15d-88a91fff6f08">&gt;Frame Acquisition Number</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_1430cd99-6dfd-4950-b2b2-ba93a5fa6a34">(0020,9156)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_a98dbfa7-d0aa-4f20-b848-168fec91a874">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3ba6b5df-b9f2-4670-9987-687ecfd20f96">A number identifying the single continuous gathering of data over a period of time that resulted in this frame.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_aced64ba-6bc1-4605-bdc7-0ce3db227b65">&gt;Frame Reference DateTime</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_991c9dc8-b3f0-4308-b8f5-cd04d893ed41">(0018,9151)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_44e259c4-5874-45df-98d9-d51d12eee3b5">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b4be9e58-cb0b-41d8-ab9b-c062befbf53c">The point in time that is most representative of when data was acquired for this frame. See Section C.7.6.16.2.2.1<!--<xref linkend="sect_C.7.6.16.2.2.1" xrefstyle="select: label"/>--> and Section C.7.6.16.2.2.2<!--<xref linkend="sect_C.7.6.16.2.2.2" xrefstyle="select: label"/>--> for further explanation.</para>
                      <note>
                        <para xml:id="para_0595ede0-6d6b-4144-979a-4d2c52ae8262">The synchronization of this time with an external clock is specified in the synchronization Module in Acquisition Time synchronized (0018,1800).</para>
                      </note>
                      <para xml:id="para_a33f713c-0957-4d73-abd2-4b97e12e5bdc">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and the SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" or "1.2.840.10008.5.1.4.1.1.4.4" or "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted). May be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_1f06faaa-748c-4ac6-bfe8-1db57aac87a6">&gt;Frame Acquisition DateTime</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_7c34ae9a-49c5-4a9e-ad76-33e2cdbbaaa0">(0018,9074)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_7462965d-82e4-490f-a0fd-353de51ee00a">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_54046302-5d66-4799-8c80-ebcf361902a7">The date and time that the acquisition of data that resulted in this frame started. See Section C.7.6.16.2.2.1<!--<xref linkend="sect_C.7.6.16.2.2.1" xrefstyle="select: label"/>--> for further explanation.</para>
                      <para xml:id="para_2865964e-056f-42ca-acf9-0db9c54872e1">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and the SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" or "1.2.840.10008.5.1.4.1.1.4.4" or "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted). May be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_4e0948b3-a4e3-4de7-8757-a9a56557051c">&gt;Frame Acquisition Duration</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_df641104-eee2-4cbe-8ab0-0fdb97459ba8">(0018,9220)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_d8afe1e8-80ce-45ab-8efe-dfaa337aaf8e">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_25eec740-e3e5-4c3d-9869-475178a0f475">The actual amount of time [in milliseconds] that was used to acquire data for this frame. See Section C.7.6.16.2.2.1<!--<xref linkend="sect_C.7.6.16.2.2.1" xrefstyle="select: label"/>--> and Section C.7.6.16.2.2.3<!--<xref linkend="sect_C.7.6.16.2.2.3" xrefstyle="select: label"/>--> for further explanation.</para>
                      <para xml:id="para_5717e579-92ca-4a89-bc96-403ded30d10a">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and the SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" or "1.2.840.10008.5.1.4.1.1.4.4" or "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted). May be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_16007e1d-d00a-4188-aef8-2aadea90fd8c">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0beb9dd5-c385-4cc8-91f1-4fbfd1463019">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0f587ad1-5890-4842-a83c-4d9e0fb1674b">...</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_5754c2c7-bd57-4a58-b841-7024d2e00748">...</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>

      <section label="C.8.12.4" status="4" xml:id="sect_C.8.12.4">
        <title>Whole Slide Microscopy Image Module</title>
        <para xml:id="para_fe13bf3f-8823-4b0a-9c2b-cc3f63af5238">
          <xref linkend="table_C.8.12.4-1" xrefstyle="select: label"/> specifies the Attributes that describe the Whole Slide Microscopy Image Module.</para>
        <table frame="box" label="C.8.12.4-1" rules="all" xml:id="table_C.8.12.4-1">
          <caption>Whole Slide Microscopy Image Module Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_19b0b4dc-4e92-4c95-9988-804684315cb7">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_96a09a1d-8879-43be-97e5-121291459369">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3e3024bf-3c64-498a-ba60-ec15eea8abd4">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_18f4a093-c520-4639-ae1b-71059a30ef26">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a877f73-4d9d-41b5-86d4-a8c3bbfedf52">Image Type</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_379f1037-05af-47e4-a7cc-263ccdb59552">(0008,0008)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_5e62e3f9-7cd2-48f1-898c-3cc78973e9f9">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8987cdf6-559b-4308-8bb6-886172153890">Image identification characteristics.</para>
                <para xml:id="para_6af97694-9c87-41bf-9a64-b1a1ab24272e">See <xref linkend="sect_C.8.12.4.1.1" xrefstyle="select: label"/> for specialization.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3bf068f2-90e2-4038-9795-a5e533a7457d">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_6fd29a76-4932-4bf1-bd94-1bc9e5e03526"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_0f85d412-d926-40c7-a05a-cd74c9d8391c"/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c0bca29-c4c8-4fba-bb7a-8d244fda6667"> </para>
              </td>
            </tr>
          </tbody>
        </table>
        <section label="C.8.12.4.1" status="5" xml:id="sect_C.8.12.4.1">
          <title>Whole Slide Microscopy Image Attribute Descriptions</title>
          <section label="C.8.12.4.1.1" status="6" xml:id="sect_C.8.12.4.1.1">
            <title>Image Type</title>
            <para xml:id="para_570df186-e80e-495a-8c3f-a885b9326afc">Image Type (0008,0008) is specified to be Type 1 with the following constraints:</para>
            <para xml:id="para_5d7c3916-3108-4847-b5d9-a9efd8c2719f">Value 1 shall have a value of ORIGINAL or DERIVED</para>
            <para xml:id="para_64085fb2-c119-461d-b5ea-597a1b2f0c02">Value 2 shall have a value of PRIMARY</para>
            <para xml:id="para_7a14bdbb-e4da-42ef-a9f0-42bf9c4a72d6">Value 3 (Image Flavor) shall have the Defined Terms in <xref linkend="table_C.8.12.4-2" xrefstyle="select: label"/>.</para>
            <table frame="box" label="C.8.12.4-2" rules="all" xml:id="table_C.8.12.4-2">
              <caption>Whole Slide Microscopy Image Flavors</caption>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f4e0b9db-57d3-40c8-be18-c56a76e704e9">LOCALIZER</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9914348f-ddb2-4d12-9137-fbec1afa0557">Collected for the purpose of planning or navigating other images.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7490619d-a47e-420d-bbab-aa7af0caad42">VOLUME</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9b847f99-2c22-472f-b17d-8f60eb58e765">Set of frames that define a regularly sampled volume</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b553e150-7ca4-4380-933d-3ff577682c88">LABEL</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_32434d33-8596-4260-86d4-c66a6bf6994e">Purpose of image is to capture the slide label; any non-label area captured is incidental to that purpose.</para>
                  </td>
                </tr>
              </tbody>
            </table>
            <para xml:id="para_fb8f0316-22e1-4eb4-bc32-cba178b99ea1">Value 4 (Derived Pixel) shall have the Defined Terms specified in <xref linkend="table_C.8.12.4-3" xrefstyle="select: label"/>.</para>
            <table frame="box" label="C.8.12.4-3" rules="all" xml:id="table_C.8.12.4-3">
              <caption>Whole Slide Microscopy Image Derived Pixels</caption>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_21cbe34a-c110-42e8-8b67-a67c8ab89150">NONE</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4b1732c6-26fc-4014-aa1a-10101b0ac589">No derivation of pixels (original)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9b1cc671-a848-44d5-87a6-6562644deb2a">RESAMPLED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9bb22eb0-ea22-48a0-8a00-0350e6287192">Pixels were derived by down sampling a higher resolution image</para>
                  </td>
                </tr>
              </tbody>
            </table>
            <para xml:id="para_3a8971ef-39d9-413a-83a7-81a386515947">No additional values shall be present.</para>
          </section>
        </section>
      </section>
      
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="6" status="1" xml:id="chapter_6">
        <title>Registry of DICOM Data Elements</title>
        
        <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
          <caption>Registry of DICOM Data Elements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                  <emphasis role="bold">Tag</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                  <emphasis role="bold">Keyword</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                  <emphasis role="bold">VR</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                  <emphasis role="bold">VM</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">(0040,0710)</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Whole Slide Microscopy Image Frame Type Sequence</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Whole​Slide​Microscopy​Image​Frame​Type​Sequence</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">SQ</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
    </chapter>
</book>