    <title>CP-1705</title>
    <subtitle>Add more texture measures</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/11/22</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para>
                    <para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>QIICR</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/03/27</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1705</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add more texture measures</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.6, 3.16 2017d</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>More texture measures exist than are currently defined in the DICOM standard.</para>
                    <para>A comprehensive and definitive summary is found in
                        Zwanenburg A et al, Image biomarker standardisation initiative version 1.4.
                        arXiv:161207003. 2017 Jul 20; Available from: <link xl:href="http://arxiv.org/abs/1612.07003"/>.
                        The existing definitions are updated to use this as the normative reference, missing concepts are added, and potential ambiguities of historical synonyms are described.</para>
                    <para>Additional radiomic features, not all of which are "texture" features, will be added in future CPs.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="2" status="1" xml:id="chapter_2">
            <title>Normative References</title>
            <bibliography>
                <title/>
                <bibliodiv label="2.1" status="2" xml:id="sect_2.1">
                    <title>General</title>
                    <biblioentry xml:id="biblio_IBSI_Features_v4">
                        <abbrev><emphasis role="bold"><emphasis role="underline">IBSI Features v4</emphasis></emphasis></abbrev>
                        <biblioset relation="journal">
                            <title><emphasis role="bold"><emphasis role="underline">arXiv</emphasis></emphasis></title>
                        </biblioset>
                        <biblioset relation="article">
                            <authorgroup>
                                <author>
                                    <personname><emphasis role="bold"><emphasis role="underline">Zwanenburg A</emphasis></emphasis></personname>
                                </author>
                                <author>
                                    <personname><emphasis role="bold"><emphasis role="underline">Leger S</emphasis></emphasis></personname>
                                </author>
                                <author>
                                    <personname><emphasis role="bold"><emphasis role="underline">Vallières M</emphasis></emphasis></personname>
                                </author>
                                <author>
                                    <personname><emphasis role="bold"><emphasis role="underline">Löck S</emphasis></emphasis></personname>
                                </author>
                            </authorgroup>
                            <date><emphasis role="bold"><emphasis role="underline">20 Jul 2017</emphasis></emphasis></date>
                            <volumenum><emphasis role="bold"><emphasis role="underline">arXiv:1612.07003v4</emphasis></emphasis></volumenum>
                            <title><emphasis role="bold"><emphasis role="underline">Image biomarker standardisation initiative - feature definitions</emphasis></emphasis></title>
                        </biblioset>
                        <bibliosource>
                            <link xl:href="http://arxiv.org/abs/1612.07003v4"/>
                        </bibliosource>
                    </biblioentry>
                </bibliodiv>
            </bibliography>
        </section>
        <section label="TID 1500" status="3" xml:id="sect_TID_1500">
            <title>Measurement Report</title>
            <table frame="box" label="TID 1500" rules="all" xml:id="table_TID_1500">
                <caption>Measurement Report</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7f96ef72-3739-4709-a6f5-d3c609e4ec05"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eb663436-d709-454e-ac93-98682212c9f0">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6142a93b-0f6d-44a4-8494-fc3299ba7c21">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_81a6fa58-162c-4816-971a-d0a224744ee1">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_97622db7-889d-46d0-a1c0-605bfed0f4c7">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_feb517dd-7464-42f5-9f1d-496753de6be4">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_220e6ae8-5f33-4f58-b426-f41cd8fd2bf4">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af3785d-8f23-4d07-809d-5639784468c9">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_80f6672b-b0af-4725-adc3-e4e8c7e83695">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_96d6cfbb-9f8d-4203-92f0-dc6f05c2d0a8">7</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6ef2d231-13a5-4959-a821-9506f815bff7">&gt;&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3f15deab-dd23-4d60-83de-dd6a73b90078">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7ce9f59e-0b11-4f3a-aa05-a88e1aa9c1ae">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_77bd714c-3790-436e-8fde-280daf9850b5">DTID 1410<!--<xref linkend="sect_TID_1410" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a9103603-6261-4637-b46b-fa22134536fe">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_15f776e2-a5af-4ee4-8902-8d861d3932b7">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d25704ea-5734-4683-8a36-210409afacf8"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a87b6e4f-bced-4eb7-85ca-72f7b53d423c">$Measurement = BCID 7469<!--<xref linkend="sect_CID_7469" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para><emphasis role="bold"><emphasis role="underline">$Measurement = B<xref linkend="sect_CID_7468" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis></para>
                            <para xml:id="para_378db2d7-b6b1-4bd5-bd55-3ecc1936deb9">$Units = BCID 7181<!--<xref linkend="sect_CID_7181" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para xml:id="para_99508cbb-2bee-4f5b-8816-1d3495eccb3c">$Derivation = BCID 7464<!--<xref linkend="sect_CID_7464" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para xml:id="para_d653b285-fe19-47d0-9b09-d0cec3ff9359">$Method = BCID 6147<!--<xref linkend="sect_CID_6147" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_237b4895-3ec5-42f6-83d7-ca4b9914ce10">8</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_34b0264b-6367-41ef-9ab3-ac6f73e5a926">&gt;&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_edd197ad-42e1-4181-b741-9f29a3ca83f8">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_38ae7415-a014-4b38-bba5-90c84f6d89c9">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8276a87d-183b-4ec0-9de3-c5b2ec064c57">DTID 1411<!--<xref linkend="sect_TID_1411" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e654f7db-b58f-4a0c-be26-9679fe2714ed">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d0a9eb71-ba77-4b3c-92b0-20807066566c">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_637e45d6-6eee-4f3f-9ce5-fb5d384ae031"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a2b0eb06-c798-4882-abec-a2632ac19b44">$Measurement = BCID 7469<!--<xref linkend="sect_CID_7469" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para><emphasis role="bold"><emphasis role="underline">$Measurement = B<xref linkend="sect_CID_7468" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis></para>
                            <para xml:id="para_d952336b-31c3-4a64-a013-7ff63702a47c">$Units = BCID 7181<!--<xref linkend="sect_CID_7181" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para xml:id="para_9f14bfcc-5741-4404-9156-21acc26e58d2">$Derivation = BCID 7464<!--<xref linkend="sect_CID_7464" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para xml:id="para_7ad66743-b635-4689-8f59-e97ffa3bde7d">$Method = BCID 6147<!--<xref linkend="sect_CID_6147" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e00ee47b-e07a-4063-9295-f00b8f902998">9</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_75af0046-0485-49d9-8863-6ce205f822c6">&gt;&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_aa985cec-610b-4445-b7f1-2ebbab06936b">CONTAINS</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_51feb8fa-c5e0-46a7-afcb-b82dc0f31423">INCLUDE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_afa5de9d-884c-4de6-8af3-0f259624c15e">DTID 1501<!--<xref linkend="sect_TID_1501" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0cc25ff-0ec1-44cd-a6c4-4ba948d2a1ae">1-n</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2c07259d-38bc-4ccd-94bd-f6802ad60bf1">U</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_87d9dd51-a193-4405-b81d-59c4610661d1"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_17c5d93d-3fbc-4a4b-9dc6-71a477409861">$Measurement = BCID 7469<!--<xref linkend="sect_CID_7469" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para><emphasis role="bold"><emphasis role="underline">$Measurement = B<xref linkend="sect_CID_7468" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis></para>
                            <para xml:id="para_56c17377-8203-49db-abf1-3e609e9ce931">$Units = BCID 7181<!--<xref linkend="sect_CID_7181" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para xml:id="para_33b0e4c5-3954-4eb8-89c2-efc8fe79d202">$Derivation = BCID 7464<!--<xref linkend="sect_CID_7464" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                            <para xml:id="para_2b135bc8-4839-49ce-9cf7-18949e1ba320">$Method = BCID 6147<!--<xref linkend="sect_CID_6147" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <para xml:id="para_868f2065-cd1a-426f-a9fe-19f831d32718">
                <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa575819-a7dc-4043-ad84-1d6ce95c8b7b">Rows 7, 8, 9</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5a0d0919-bb63-4e2f-a517-75d44bd1afe1">The baseline context groups defined allow for generic intensity<emphasis role="bold"><emphasis role="strikethrough"> and</emphasis><emphasis role="underline">,</emphasis></emphasis> size <emphasis role="bold"><emphasis role="underline">and texture </emphasis></emphasis>measurements, regardless of the geometry of the ROI (e.g., linear distance can be measured on volumes, or volume can be estimated from a linear distance), and being baseline, do not constrain the invoker from using other appropriate concepts specific to the application.</para>
                        </td>
                    </tr>
                </tbody>
            </informaltable>
        </section>
        
        <section label="CID 7467" status="3" xml:id="sect_CID_7467">
            <title>Gray Level Co-occurrence Matrix Measurements</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_d68c18bc-202f-4ee9-8465-18610a4dcc7b">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_04565925-5d06-410e-bb5c-8bb21c20cde8">
                            <emphasis role="bold"><emphasis role="strikethrough">20141110</emphasis><emphasis role="underline">20171122</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7467" rules="all" xml:id="table_CID_7467">
                <caption>Gray Level Co-occurrence Matrix Measurements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c7a6700c-50a8-4203-bc2b-2d071b998436">
                                <emphasis role="bold">Coding Scheme Designator</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e25b50e3-fe79-4f6a-91fb-f1761819ada8">
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eb595bda-d969-4d1f-8506-824e4a812125">
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8df55bcd-9484-4a21-899e-fb8e40ef8d9e">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126060" xrefstyle="select: nopage">126060</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a5484a73-c304-4355-ba50-30589d57a58a"><emphasis role="bold"><emphasis role="underline">Joint </emphasis></emphasis>Entropy of GLCM</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_812f778f-076c-4bd2-a7f7-b4befeda0937">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126061" xrefstyle="select: nopage">126061</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_714a5f55-ae6e-441a-9cbd-93a3c0c862ba"><emphasis role="bold"><emphasis role="strikethrough">Energy</emphasis><emphasis role="underline">Root Angular Second Moment</emphasis></emphasis> of GLCM</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0f42e901-3e6a-47ad-a8f9-b96021fa6de3">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126062" xrefstyle="select: nopage">126062</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5279fa7a-a897-4181-af57-cc93376427fc"><emphasis role="bold"><emphasis role="strikethrough">Homogeneity</emphasis><emphasis role="underline">Inverse Difference Moment</emphasis></emphasis> of GLCM</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0b486b02-3edf-400a-9502-41f444adecd9">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126063" xrefstyle="select: nopage">126063</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3e1cdf17-82e2-493d-be54-5ef0e96e39d9">Contrast of GLCM</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2bc3513c-0623-43df-b121-2f984746da6c">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126064" xrefstyle="select: nopage">126064</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_86c5bfe6-180d-4f02-b982-3678a7fd295c">Dissimilarity of GLCM</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ad2f3d0-65b8-47cd-bc91-0c0af91bd977">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126065" xrefstyle="select: nopage">126065</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4566551f-31e3-4fe2-8fc6-00b09b51c6b7"><emphasis role="bold"><emphasis role="strikethrough">ASM</emphasis><emphasis role="underline">Angular Second Moment</emphasis></emphasis> of GLCM</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fe9d87c3-a39f-45e8-b2df-ca69cb4d82ad">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126066" xrefstyle="select: nopage">126066</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cbb09e54-e125-4afc-90ba-a33a0cb39acf">Correlation of GLCM</para>
                        </td>
                    </tr>
                    
                    
                    
                    
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128781" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128781</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Joint Maximum of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128782" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128782</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Joint Average of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128783" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128783</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Joint Variance of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128784" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128784</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Difference Average of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128785" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128785</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Difference Variance of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128786" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128786</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Difference Entropy of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128787" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128787</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sum Average of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128788" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128788</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sum Variance of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128789" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128789</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sum Entropy of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128790" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128790</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Difference of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128791" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128791</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Difference Normalized of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128792" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128792</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Difference Moment Normalized of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128793" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128793</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Variance of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128794" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128794</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Autocorrelation of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128795" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128795</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Cluster Tendency of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128796" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128796</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Cluster Shade of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128797" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128797</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Cluster Prominence of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128798" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128798</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">First Measure of Information Correlation of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128799" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128799</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Second Measure of Information Correlation of GLCM</emphasis></emphasis></para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        
        
        
        <section label="CID 7475" status="3" xml:id="sect_CID_7475">
            <title><emphasis role="bold"><emphasis role="underline">Gray Level Run Length Based Features</emphasis></emphasis></title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold"><emphasis role="underline">Type:</emphasis></emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold"><emphasis role="underline">Extensible</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold"><emphasis role="underline">Version:</emphasis></emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold"><emphasis role="underline">20171122</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7475" rules="all" xml:id="table_CID_7475">
                <caption><emphasis role="bold"><emphasis role="underline">Gray Level Run Length Based Features</emphasis></emphasis></caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline">Coding Scheme Designator</emphasis></emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline">Code Value</emphasis></emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline">Code Meaning</emphasis></emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128801" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128801</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Short Runs Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128802" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128802</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Long Runs Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128803" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128803</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Low Gray Level Run Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128804" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128804</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">High Gray Level Run Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128805" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128805</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Short Run Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128806" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128806</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Short Run High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128807" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128807</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Long Run Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128808" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128808</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Long Run High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128809" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128809</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity in Runs</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128810" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128810</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity in Runs Normalized</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128811" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128811</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Length Nonuniformity</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128812" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128812</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Length Nonuniformity Normalized</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128813" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128813</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Percentage</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128814" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128814</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Variance in Runs</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128815" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128815</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Length Variance</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128816" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128816</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Entropy</emphasis></emphasis></para>
                        </td>
                    </tr>
                    
                    
                </tbody>
            </table>
        </section>
        <section label="CID 7476" status="3" xml:id="sect_CID_7476">
            <title><emphasis role="bold"><emphasis role="underline">Gray Level Size Zone Based Features</emphasis></emphasis></title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold"><emphasis role="underline">Type:</emphasis></emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold"><emphasis role="underline">Extensible</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold"><emphasis role="underline">Version:</emphasis></emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold"><emphasis role="underline">20171122</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7476" rules="all" xml:id="table_CID_7476">
                <caption><emphasis role="bold"><emphasis role="underline">Gray Level Size Zone Based Features</emphasis></emphasis></caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline">Coding Scheme Designator</emphasis></emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline">Code Value</emphasis></emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline">Code Meaning</emphasis></emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128821" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128821</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Small Zone Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128822" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128822</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Large Zone Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128823" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128823</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Low Gray Level Zone Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128824" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128824</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">High Gray Level Zone Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128825" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128825</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Small Zone Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128826" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128826</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Small Zone High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128827" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128827</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Large Zone Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128828" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128828</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Large Zone High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128829" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128829</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity of Zone Counts</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128830" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128830</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity of Zone Counts Normalized</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128831" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128831</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Nonuniformity</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128832" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128832</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Nonuniformity Normalized</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128833" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128833</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Percentage</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128834" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128834</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Variance in Zones</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128835" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128835</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Variance</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_128836" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128836</emphasis></emphasis></link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Entropy</emphasis></emphasis></para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        <section label="CID 7468" status="3" xml:id="sect_CID_7468">
            <title>Texture Measurements</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_6ecad3a2-9e1a-49ee-9df2-d086eeec7470">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_a11032d7-bdbe-43e9-a8d9-60db8f619884">
                            <emphasis role="bold"><emphasis role="strikethrough">20141110</emphasis><emphasis role="underline">20171122</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7468" rules="all" xml:id="table_CID_7468">
                <caption>Texture Measurements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3019298e-4248-470d-ae84-ce392ee6073b">
                                <emphasis role="bold">Coding Scheme Designator</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b490b4df-bcaf-4d8d-96f8-61ce462c1fee">
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d10076be-629d-4793-ad01-1a6b89cf4268">
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_10627633-7d75-47e0-a3de-4c112e86a100">
                                <emphasis role="italic">Include <xref linkend="sect_CID_7467" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline"><emphasis role="italic">Include <xref linkend="sect_CID_7475" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </emphasis></emphasis></emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline"><emphasis role="italic">Include <xref linkend="sect_CID_7476" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </emphasis></emphasis></emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_319676bd-7c7f-4972-aa10-505bb7bd8302">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_126050" xrefstyle="select: nopage">126050</link>
                            </para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_48152c1f-e876-4f03-ad8f-e9ce952adcb9">Fractal Dimension</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">
                                <emphasis role="bold">Definition</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">
                                <emphasis role="bold">Notes</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126050">126050</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_82a8b177-a776-4ebb-a3be-2ca98d7a774e">Fractal Dimension</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1faf3b63-fafc-4a9a-ad88-8d4c4e526824">A statistical index of complexity comparing how detail in a fractal pattern changes with the scale at which it is measured; a ratio of the change in detail to the change in scale.</para>
                        </td>
                        <td align="left" colspan="1"/>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126060">126060</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_36c29347-422c-44db-a4c2-801f855615f4"><emphasis role="bold"><emphasis role="underline">Joint </emphasis></emphasis>Entropy of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_79ddffa9-a3f3-40e6-b7ab-68c32cab6250">The zero order entropy of a Gray Level Co-occurrence Matrix (GLCM). A measure of disorder.<emphasis role="bold"><emphasis role="underline"> Abbreviated ENT.</emphasis></emphasis></para>
                            <para xml:id="para_2ff5f3fa-1f9e-4793-9ebd-4a77136fc456">See <emphasis role="bold"><emphasis role="underline">F<subscript>cm.joint.entr</subscript> in </emphasis></emphasis><emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1"/>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126061">126061</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_c631328c-5e78-4395-9607-040cf184f180"><emphasis role="bold"><emphasis role="strikethrough">Energy</emphasis><emphasis role="underline">Root Angular Second Moment</emphasis></emphasis> of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_8de41f82-bd2e-4e42-9ce0-1a7e2535d64d">The <emphasis role="bold"><emphasis role="strikethrough">energy (uniformity) (</emphasis></emphasis>square root of the Angular Second Moment (ASM)<emphasis role="bold"><emphasis role="strikethrough">)</emphasis></emphasis> of a Gray Level Co-occurrence Matrix (GLCM). A measure of orderliness.</para>
                            <para xml:id="para_69f50963-8921-4bc0-83e9-328bcfb3e286">See <link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sometimes referred to as "energy", "uniformity" or "uniformity of energy" but then potentially confused with ASM. Not defined in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126062">126062</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_270d5893-991e-47ec-96e5-d5b0b804eb6d"><emphasis role="bold"><emphasis role="strikethrough">Homogeneity</emphasis><emphasis role="underline">Inverse Difference Moment</emphasis></emphasis> of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_3ca85ace-c55a-4de1-ae72-0f51dab7e8dd">The Inverse Difference Moment <emphasis role="bold"><emphasis role="underline">(homogeneity) </emphasis></emphasis>of a Gray Level Co-occurrence Matrix (GLCM).<emphasis role="bold"><emphasis role="underline"> Abbreviated IDM.</emphasis></emphasis></para>
                            <para xml:id="para_092daca8-866b-47a3-a816-c1f35674f005">See<emphasis role="bold"><emphasis role="underline"> F<subscript>cm.inv.diff.mom</subscript> in</emphasis></emphasis>
                                <emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Other concepts are sometimes referred to as "homogeneity", e.g., the "inverse difference", which is calculated from the absolute value of differences rather than square of them.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126063">126063</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_53e7905c-42b5-4ad9-bc3a-d58db16ce500">Contrast of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_9dde55fd-eacd-4648-98dc-bb1480c79962">The sum of squares <emphasis role="bold"><emphasis role="strikethrough">variance </emphasis></emphasis>of a Gray Level Co-occurrence Matrix (GLCM).<emphasis role="bold"><emphasis role="underline"> A measure of gray level variations. Abbreviated CON.</emphasis></emphasis></para>
                            <para xml:id="para_3fd341de-f682-40c4-ad31-62fef52eb8fe">See<emphasis role="bold"><emphasis role="underline"> F<subscript>cm.contrast</subscript> in</emphasis></emphasis>
                                <emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Distinct from "joint (sum of squares) variance" and "dissimilarity".</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126064">126064</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_af26a1f5-4c60-44f9-acf3-47f7b55ae248">Dissimilarity of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_a6a66989-dce5-4dd7-ad15-19d2431e400e">The dissimilarity of a Gray Level Co-occurrence Matrix (GLCM).<emphasis role="bold"><emphasis role="underline"> Abbreviated DIS.</emphasis></emphasis></para>
                            <para xml:id="para_f939bdf5-3373-4037-8126-5e5deca720af">See<emphasis role="bold"><emphasis role="underline"> F<subscript>cm.dissimilarity</subscript> in</emphasis></emphasis>
                                <emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Distinct from "contrast", which uses square rather than absolute value of difference.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126065">126065</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_cf0eaa3f-3ff4-4fe3-aee5-6dd315c5696f"><emphasis role="bold"><emphasis role="strikethrough">ASM</emphasis><emphasis role="underline">Angular Second Moment</emphasis></emphasis> of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_4509dbfa-bf13-4b31-b8ea-eb00fd15b5df">The Angular Second Moment of a Gray Level Co-occurrence Matrix (GLCM).<emphasis role="bold"><emphasis role="underline"> Abbreviated ASM.</emphasis></emphasis></para>
                            <para xml:id="para_e3cf1c80-d4cc-4bf3-801c-c5ba201fd1c2">See <emphasis role="bold"><emphasis role="underline">F<subscript>cm.energy</subscript> in </emphasis></emphasis><emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sometimes referred to as "energy", "uniformity" or "uniformity of energy" but then potentially confused with square root of ASM.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126066">126066</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_9e7cef00-6579-4c23-91c6-c76859837294">Correlation of GLCM</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_48f4c2aa-a9d6-409e-b61e-e0ee5933d1e7">A measure of the linear dependency of <emphasis role="bold"><emphasis role="strikethrough">grey</emphasis><emphasis role="underline">gray</emphasis></emphasis> levels on those of neighbouring pixels of a Gray Level Co-occurrence Matrix (GLCM).<emphasis role="bold"><emphasis role="underline"> Abbreviated COR.</emphasis></emphasis></para>
                            <para xml:id="para_aac63cc2-fb0c-40ae-90ba-6b51a78430f7">See<emphasis role="bold"><emphasis role="underline"> F<subscript>cm.corr</subscript> in</emphasis></emphasis>
                                <emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Correlation is NaN for a constant image.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_126067">126067</para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_29b741b2-c77b-4e88-8f71-e816d0a7d47c">Gray Level Co-occurrence Matrix<emphasis role="bold"><emphasis role="strikethrough"> (GLCM)</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para xml:id="para_5abe135c-ad0c-4502-a1b4-110550d109df">A tabulation of how often different combinations of pixel values (<emphasis role="bold"><emphasis role="strikethrough">grey</emphasis><emphasis role="underline">gray</emphasis></emphasis> levels) occur in an image.<emphasis role="bold"><emphasis role="underline"> Abbreviated GLCM.</emphasis></emphasis></para>
                            <para xml:id="para_1532bbb9-943a-45a7-b228-bfb9e12c0e06">See <emphasis role="bold"><emphasis role="strikethrough"><link xl:href="http://www.fp.ucalgary.ca/mhallbey/the_glcm.htm"/></emphasis><emphasis role="underline"><xref linkend="biblio_IBSI_Features_v4"/></emphasis></emphasis>.</para>
                        </td>
                        <td align="left" colspan="1"/>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128776"><emphasis role="bold"><emphasis role="underline">128776</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Run Length Matrix</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The tabulation of gray level run lengths in a particular direction in an image. Abbreviated GLRLM.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128777"><emphasis role="bold"><emphasis role="underline">128777</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Size Zone Matrix</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A tabulation of counts of the number of groups of connected voxels with a specific discretized gray level value and size. Abbreviated GLSZM.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    
                    
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128781"><emphasis role="bold"><emphasis role="underline">128781</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Joint Maximum of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The probability corresponding to the most common gray level co-occurrence in the GLCM. Abbreviated MAX.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.joint.max</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128782"><emphasis role="bold"><emphasis role="underline">128782</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Joint Average of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The gray level weighted sum of joint probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.joint.avg</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128783"><emphasis role="bold"><emphasis role="underline">128783</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Joint Variance of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The sum of squares of the difference from the joint average of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.joint.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128784"><emphasis role="bold"><emphasis role="underline">128784</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Difference Average of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The average for the diagonal probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.diff.avg</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128785"><emphasis role="bold"><emphasis role="underline">128785</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Difference Variance of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The variance for the diagonal probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.diff.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128786"><emphasis role="bold"><emphasis role="underline">128786</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Difference Entropy of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The entropy for the diagonal probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.diff.entr</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128787"><emphasis role="bold"><emphasis role="underline">128787</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sum Average of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The average for the cross-diagonal probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.sum.avg</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128788"><emphasis role="bold"><emphasis role="underline">128788</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sum Variance of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The variance for the cross-diagonal probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.sum.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128789"><emphasis role="bold"><emphasis role="underline">128789</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sum Entropy of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The entropy for the cross-diagonal probabilities of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.sum.entr</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128790"><emphasis role="bold"><emphasis role="underline">128790</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Difference of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The inverse difference of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.inv.diff</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Sometimes referred to as "homogeneity" but that term is historically used to refer to the "inverse difference moment", which is calculated from the square of differences rather than absolute value of them.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128791"><emphasis role="bold"><emphasis role="underline">128791</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Difference Normalized of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The normalized inverse difference of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.inv.diff.norm</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The US not UK spelling of "normalized" is used to be consistent with the DICOM convention, rather than the IBSI spelling.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128792"><emphasis role="bold"><emphasis role="underline">128792</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Difference Moment Normalized of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The normalized inverse difference moment of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.inv.diff.mom.norm</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The US not UK spelling of "normalized" is used to be consistent with the DICOM convention, rather than the IBSI spelling.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128793"><emphasis role="bold"><emphasis role="underline">128793</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Inverse Variance of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The inverse variance of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.inv.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128794"><emphasis role="bold"><emphasis role="underline">128794</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Autocorrelation of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The autocorrelation of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.auto.corr</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128795"><emphasis role="bold"><emphasis role="underline">128795</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Cluster Tendency of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The cluster tendency of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.clust.tend</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128796"><emphasis role="bold"><emphasis role="underline">128796</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Cluster Shade of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The cluster shade of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.clust.shade</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128797"><emphasis role="bold"><emphasis role="underline">128797</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Cluster Prominence of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The cluster prominence of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.clust.prom</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128798"><emphasis role="bold"><emphasis role="underline">128798</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">First Measure of Information Correlation of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The first measure of information correlation of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.info.corr.1</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128799"><emphasis role="bold"><emphasis role="underline">128799</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Second Measure of Information Correlation of GLCM</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The second measure of information correlation of a Gray Level Co-occurrence Matrix (GLCM).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>cm.info.corr.2</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                        </td>
                    </tr>

                    
                    
                    
                    
                    
                    
                    
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128801"><emphasis role="bold"><emphasis role="underline">128801</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Short Runs Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the distribution of short runs in a gray level run length matrix. Abbreviated SRE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.sre</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128802"><emphasis role="bold"><emphasis role="underline">128802</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Long Runs Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the distribution of long runs in a gray level run length matrix. Abbreviated LRE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.lre</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128803"><emphasis role="bold"><emphasis role="underline">128803</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Low Gray Level Run Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the distribution of low gray level values in a gray level run length matrix. Abbreviated LGRE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.lgre</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128804"><emphasis role="bold"><emphasis role="underline">128804</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">High Gray Level Run Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the distribution of high gray level values in a gray level run length matrix. Abbreviated HGRE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.hgre</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128805"><emphasis role="bold"><emphasis role="underline">128805</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Short Run Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the joint distribution of short runs and low gray level values in a gray level run length matrix. Abbreviated SRLGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.srlge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128806"><emphasis role="bold"><emphasis role="underline">128806</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Short Run High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the joint distribution of short runs and high gray level values in a gray level run length matrix. Abbreviated SRHGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.srhge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128807"><emphasis role="bold"><emphasis role="underline">128807</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Long Run Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the joint distribution of long runs and low gray level values in a gray level run length matrix. Abbreviated LRLGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.lrlge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128808"><emphasis role="bold"><emphasis role="underline">128808</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Long Run High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the the joint distribution of long runs and high gray level values in a gray level run length matrix. Abbreviated LRHGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.lrhge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128809"><emphasis role="bold"><emphasis role="underline">128809</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity in Runs</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the similarity of gray level values throughout the image in a gray level run length matrix. Abbreviated RLM.GLNU.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.glnu</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128810"><emphasis role="bold"><emphasis role="underline">128810</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity in Runs Normalized</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A normalized measure of the similarity of gray level values throughout the image in a gray level run length matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.glnu.norm</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The US not UK spelling of "normalized" is used to be consistent with the DICOM convention, rather than the IBSI spelling.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128811"><emphasis role="bold"><emphasis role="underline">128811</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Length Nonuniformity</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the the similarity of the length of runs throughout the image in a gray level run length matrix. Abbreviated RLNU.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.rlnu</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128812"><emphasis role="bold"><emphasis role="underline">128812</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Length Nonuniformity Normalized</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A normalized measure of the the similarity of the length of runs throughout the image in a gray level run length matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.rlnu.norm</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The US not UK spelling of "normalized" is used to be consistent with the DICOM convention, rather than the IBSI spelling.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128813"><emphasis role="bold"><emphasis role="underline">128813</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Percentage</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A measure of the homogeneity and distribution of runs of an image in a specific direction in a gray level run length matrix. Abbreviated RPC.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.r.perc</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128814"><emphasis role="bold"><emphasis role="underline">128814</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Variance in Runs</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The variance in runs for the gray levels in a gray level run length matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.gl.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128815"><emphasis role="bold"><emphasis role="underline">128815</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Length Variance</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The variance in runs for run lengths in a gray level run length matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.rl.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128816"><emphasis role="bold"><emphasis role="underline">128816</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Run Entropy</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The entropy of runs in a gray level run length matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>rlm.rl.entr</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128821"><emphasis role="bold"><emphasis role="underline">128821</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Small Zone Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes small zones from a gray level size zone matrix. Abbreviated SZE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.sze</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128822"><emphasis role="bold"><emphasis role="underline">128822</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Large Zone Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes large zones from a gray level size zone matrix. Abbreviated LZE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.lze</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128823"><emphasis role="bold"><emphasis role="underline">128823</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Low Gray Level Zone Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes low gray level zones from a gray level size zone matrix. Abbreviated LGZE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.lgze</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128824"><emphasis role="bold"><emphasis role="underline">128824</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">High Gray Level Zone Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes high gray level zones from a gray level size zone matrix. Abbreviated LGZE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.hgze</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128825"><emphasis role="bold"><emphasis role="underline">128825</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Small Zone Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes small zone sizes and low gray levels from a gray level size zone matrix. Abbreviated SZLGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.szlge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128826"><emphasis role="bold"><emphasis role="underline">128826</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Small Zone High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes small zone sizes and high gray levels from a gray level size zone matrix. Abbreviated SZHGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.szhge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128827"><emphasis role="bold"><emphasis role="underline">128827</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Large Zone Low Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes large zone sizes and low gray levels from a gray level size zone matrix. Abbreviated LZLGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.lzlge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128828"><emphasis role="bold"><emphasis role="underline">128828</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Large Zone High Gray Level Emphasis</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">A feature that emphasizes large zone sizes and high gray levels from a gray level size zone matrix. Abbreviated LZHGE.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.lzhge</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128829"><emphasis role="bold"><emphasis role="underline">128829</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity of Zone Counts</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The distribution of zone counts over the gray values in a gray level size zone matrix. Abbreviated SZM.GLNU.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.glnu</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128830"><emphasis role="bold"><emphasis role="underline">128830</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Nonuniformity of Zone Counts Normalized</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The normalized distribution of zone counts over the gray values in a gray level size zone matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.glnu.norm</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The US not UK spelling of "normalized" is used to be consistent with the DICOM convention, rather than the IBSI spelling.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128831"><emphasis role="bold"><emphasis role="underline">128831</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Nonuniformity</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The distribution of zone counts over the different zone sizes in a gray level size zone matrix. Abbreviated ZSNU.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.zsnu</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128832"><emphasis role="bold"><emphasis role="underline">128832</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Nonuniformity Normalized</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The normalized distribution of zone counts over the different zone sizes in a gray level size zone matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.zsnu.norm</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The US not UK spelling of "normalized" is used to be consistent with the DICOM convention, rather than the IBSI spelling.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128833"><emphasis role="bold"><emphasis role="underline">128833</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Percentage</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The fraction of the number of realised zones relative to the maximum number of potential zones in a gray level size zone matrix. Abbreviated ZPERC.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.z.perc</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128834"><emphasis role="bold"><emphasis role="underline">128834</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Gray Level Variance in Zones</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The variance ithe variance in zone counts for the gray levels in a gray level size zone matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.gl.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128835"><emphasis role="bold"><emphasis role="underline">128835</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Variance</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The variance in zone counts for the different zone sizes in a gray level size zone matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.zs.var</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_128836"><emphasis role="bold"><emphasis role="underline">128836</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Zone Size Entropy</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">The entropy of zone sizes in a gray level size zone matrix.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">See F<subscript>szm.zs.entr</subscript> in <xref linkend="biblio_IBSI_Features_v4"/>.</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <table frame="box" label="A-3" rules="all" xml:id="table_A-3">
            <caption>Context Group UID Values</caption>
            <thead>
                <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3322bcc5-7ea4-44d1-9b17-499a6a5e512c">
                            <emphasis role="bold">Context UID</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_225c7fe6-91ef-424b-812b-249ecd55f194">
                            <emphasis role="bold">Context Identifier</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_25349958-e4d1-4bb0-aaa9-33d5cb03df93">
                            <emphasis role="bold">Context Group Name</emphasis>
                        </para>
                    </th>
                </tr>
            </thead>
            <tbody>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.1.1199</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold"><emphasis role="underline">CID 7475</emphasis></emphasis><!--<olink targetdoc="PS3.16" targetptr="sect_CID_7475" xrefstyle="select: labelnumber"/>-->
                        </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold"><emphasis role="underline">Gray Level Run Length Based Features</emphasis></emphasis><!--<olink targetdoc="PS3.16" targetptr="sect_CID_7475" xrefstyle="select: title"/>-->
                        </para>
                    </td>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.1.1200</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold"><emphasis role="underline">CID 7476</emphasis></emphasis><!--<olink targetdoc="PS3.16" targetptr="sect_CID_7476" xrefstyle="select: labelnumber"/>-->
                        </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold"><emphasis role="underline">Gray Level Size Zone Based Features</emphasis></emphasis><!--<olink targetdoc="PS3.16" targetptr="sect_CID_7476" xrefstyle="select: title"/>-->
                        </para>
                    </td>
                </tr>
            </tbody>