<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-1690</title>
  <subtitle>Query Key for Anatomic Regions in Study</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/09/14</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2017/01/19</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1690</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Query Key for Anatomic Regions in Study</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.4, PS3.6 2017c</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>It is useful when retrieving relevant priors to know what anatomic regions are present in a Study and this information should not require multiple Series level queries to obtain, as is presently the case.</para>
                  <para>It is proposed to add a new Study level query attribute similar to the existing Modalities in Study and SOP Classes in Study.</para>
               </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
  <chapter>
    <title/>

    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.4 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>

    <section label="C.3.4" status="3" xml:id="sect_C.3.4">
      <title>Additional Query/Retrieve Attributes</title>
      <para xml:id="para_9c754e44-5930-4974-9da8-29a7dc52a2d6">Some optional Attributes that may be used in Query/Retrieve Information Models that are not Attributes of an Information Object Definition and, therefore, are not defined in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>. These Attributes are defined in <xref linkend="table_C.3-1" xrefstyle="select: label"/>.</para>
      <table frame="box" label="C.3-1" rules="all" xml:id="table_C.3-1">
        <caption>Additional Query/Retrieve Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4f593ed3-cdf5-43a3-a48e-540b70cd2fe2">
                <emphasis role="bold">Attribute Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_d86c3996-8ff3-413d-957b-af01106d02c9">
                <emphasis role="bold">Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8ed8738d-6dee-47bb-bbd1-df4a265606e2">
                <emphasis role="bold">Attribute Description</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_134743bd-98ca-497e-9f9a-563cca1585a9">Number of Patient Related Studies</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_d28bc03f-76d8-4f12-9ff0-bee0b4510788">(0020,1200)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_510a6735-d55f-49ce-bf7c-007615424a03">The number of studies that match the Patient level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1fa021ca-518b-499b-9fff-b56cd59cb6de">Number of Patient Related Series</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_e03b0cd4-16e0-4cfc-8537-35af1bebfaf2">(0020,1202)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c9b88b2-0e5d-4d9b-b18c-c303e5468ed6">The number of series that match the Patient level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_75adfffe-918d-4328-897c-0f2fcf8bd6b9">Number of Patient Related Instances</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_5375cb8e-4afd-468a-84a3-94ba3e9a47b1">(0020,1204)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c3f1590-e378-472d-924f-e5d9f1d202ee">The number of Composite Object Instances that match the Patient level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a1cc99e-53e5-4c0b-8cf7-e6e5ea041e7a">Number of Study Related Series</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1160fa3f-f3e5-43c6-92f7-62af77509338">(0020,1206)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_175b74e4-4d7a-43d4-970b-a388458397b3">The number of series that match the Study level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dedfb6a9-f8bb-4a36-9960-a87960e47012">Number of Series Related Instances</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_53c0f2d2-f01f-43f9-9299-f07a3ce9987a">(0020,1209)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e737efe-711f-42be-a072-9a4453b34af5">The number of Composite Object Instances in a Series that match the Series level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_edde4832-e6f6-4562-b106-4c3b4a8ba3bb">Number of Study Related Instances</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_38a71ab4-7fc1-41cf-a686-1971b105d83b">(0020,1208)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3acd70c-959d-4821-9f46-f0977dd59753">The number of Composite Object Instances that match the Study level Query/Retrieve search criteria</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d96fa807-424c-4d08-9d2a-618c20b337e3">Modalities in Study</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a6f4eb70-fc9b-4ce7-b8c9-20d924cb5003">(0008,0061)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5dd7fcad-8057-4042-8a54-509f001fc02c">All of the distinct values used for Modality (0008,0060) in the Series of the Study.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b9563ee-2571-4bce-8f1b-640740d1abd1">SOP Classes in Study</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a477def1-6574-4172-bce1-8d92ac353ec5">(0008,0062)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_764be1cf-09f4-43c2-9857-ea03eb44fde9">The SOP Classes contained in the Study.</para>
            </td>
          </tr>
          <tr>
            <td align="left" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Anatomic Regions in Study Code Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0008,0063)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">The anatomic regions of interest in this Study (i.e., external anatomy, surface anatomy, or general region of the body).</emphasis></emphasis></para>
              <para><emphasis role="bold"><emphasis role="underline">One or more Items are permitted in this Sequence.</emphasis></emphasis></para>
              <note>
                <para><emphasis role="bold"><emphasis role="underline">If the Instances in the Study contain Anatomic Region Sequence (0008,2218), then the Items of this Sequence may be the union of the codes in the Instances. Alternatively, multiple (usually contiguous) anatomic regions might be combined into a single code. E.g., (T-D3000, SRT, "Chest"), (T-D4000, SRT, "Abdomen") and (T-D6000, SRT, "Pelvis") might be combined into (R-FAB56, SRT, "Chest, Abdomen and Pelvis").</emphasis></emphasis></para>
                <para><emphasis role="bold"><emphasis role="underline">If Instances in the Study do not contain Anatomic Region Sequence (0008,2218) but do contain Body Part Examined (0018,0015), then codes equivalent to recognized values of Body Part Examined (0018,0015) may be used. See PS3.16 Annex L for standard equivalent values. E.g., if an Instance contained Body Part Examined (0018,0015) with a value of "CHEST", then this Sequence might contain (T-D3000, SRT, "Chest").</emphasis></emphasis></para>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0234f7f-1550-42d7-9a2e-129edd81f739">Alternate Representation Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_922e0d42-ebde-4c99-b74f-16a15efa1dd1">(0008,3001)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_32613412-1537-477d-ad0c-67167c5bb97d">A Sequence of Items, each identifying an alternate encoding of an image that matches the Instance level Query/Retrieve search criteria (see <xref linkend="sect_C.6.1.1.5.1" xrefstyle="select: label"/>)</para>
            </td>
          </tr>
        </tbody>
      </table>
      <para xml:id="para_12d37be6-3667-4be0-a405-830e1a86f3fa">...</para>
    </section>
    
    <section label="C.6.1.1.3" status="5" xml:id="sect_C.6.1.1.3">
            <title>Study Level</title>
            <table frame="box" label="C.6-2" rules="all" xml:id="table_C.6-2">
              <caption>Study Level Keys for the Patient Root Query/Retrieve Information Model</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ccf2db48-6cf9-4574-9ab3-eb239149408f">
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0923eab6-a467-4a62-a809-55bd940aada1">
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_001823bf-1f76-43de-b4d5-9837959137f1">
                      <emphasis role="bold">Type</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_802fef70-7396-4484-b82f-d31638458674">Study Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_52287ddd-1995-4a4f-8ee8-54f053697863">(0008,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ed424087-8a3a-4c4f-bf4c-d77b2cb42c7d">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3ee47af7-ef06-4aa3-9937-c4cf52acf6b6">Study Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fc1b1a2e-0051-4393-925f-eeb66bb9eb1a">(0008,0030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4918ee5d-a63b-424f-af6d-2e324bfcc6b2">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_5ffdba34-762e-4ea4-aad1-22fec1d396b2">Accession Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fd6bf611-34a0-4acf-a333-4449b475f03d">(0008,0050)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5075bc34-1a14-43ed-bea6-51a2b642ae61">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6cffd2bb-9714-410d-9976-07a7ee4b2492">Study ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9d081ae0-76f9-4147-bd3d-2adb7fbfd9b8">(0020,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_eb4593dc-92b9-43f7-92a8-291c68434f7e">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_893b0b58-53aa-482f-b12c-0c34bd1ebcf3">Study Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e76e3e65-600b-4534-8670-9edc450a7afe">(0020,000D)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c772899a-079d-4994-a38b-1a2c3124d98a">U</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4f66ae7e-62e8-4cb2-9b9b-54c064cffa02">Modalities in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_96dc072c-1f55-4e96-bc72-82f98a8fabbb">(0008,0061)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7d1f490f-e333-4142-99a0-1b843bc72fa1">O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_01f965a6-cfa7-4a99-970a-678f0c472329">SOP Classes in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e5079371-1c6a-4337-ae27-9f26940b3507">(0008,0062)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ca395095-3599-4bd6-98e6-dc58a9ed6ca5">O</para>
                  </td>
                </tr>
          <tr>
            <td align="left" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Anatomic Regions in Study Code Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0008,0063)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">O</emphasis></emphasis></para>
            </td>
          </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="italic">All other Attributes at Study Level</emphasis>
                    </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para>O</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
<section label="C.6.2.1.2" status="5" xml:id="sect_C.6.2.1.2">
            <title>Study Level</title>
            <table frame="box" label="C.6-5" rules="all" xml:id="table_C.6-5">
              <caption>Study Level Keys for the Study Root Query/Retrieve Information Model</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_98a3ba2e-39b3-4633-8242-80eb2172f4d4">
                      <emphasis role="bold">Attribute Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4de39ea1-2ba6-42f0-92a4-1a78dd50ab7d">
                      <emphasis role="bold">Tag</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f71a8012-7951-488f-9a9c-b6480c784453">
                      <emphasis role="bold">Type</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c6384e34-d394-49c4-9705-e00bf5052cdf">Study Date</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c93ae5af-e997-403c-9cb5-3cf5c7f9c93d">(0008,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_070fab5b-e442-4156-8f0a-86ab90676b7a">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_94b16220-7ef1-41e4-8ab9-d8a3204be4eb">Study Time</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2c2aa045-fb84-4f06-b689-fc4c4002034f">(0008,0030)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d00ca5a8-85e3-47bf-ae0b-c3508e7c042c">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9ba9d006-81e4-4b86-888e-3902ec7f88d6">Accession Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e41bdc99-8d79-4c74-8f24-438f2106fe21">(0008,0050)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_326caf0c-e5d8-4e02-834c-06164b40429b">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_01c14e01-71eb-4f33-a3e9-b9e33cc7c79d">Patient's Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2df6963f-5d59-4fc1-b8a2-0c747ab1dcc1">(0010,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_073f4dcd-15c3-4c24-b65f-564ba6c63ea5">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0984dd2b-32ee-45e3-bdee-e518eadfe737">Patient ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cb811614-1ec1-4932-ba85-08bb028966a5">(0010,0020)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0c17974c-68d6-45bc-8d6b-9cdd3bdf612a">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0b7ce340-3fa9-4875-8cb8-8149f0b8777b">Study ID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_dff65679-1c0e-48e7-85d1-b3a768fb4e74">(0020,0010)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aa606621-79e0-450b-b082-1c48ea0f182b">R</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7ff67948-e05d-4033-a3b7-bf5b6c5bc129">Study Instance UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_bee0b55f-86d7-4d3b-929d-af26c0af5462">(0020,000D)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0c673f4d-17c1-4464-8442-30c6a7d92282">U</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e61e7de1-bed7-4ca9-81fc-5b798c9f83da">Modalities in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4ce256d7-de0c-4523-a7d2-50dc945a0eb9">(0008,0061)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c244abf9-4c92-4889-80fe-4ad7d35cba8e">O</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_62165677-abf2-457c-8a30-2dd202c9db25">SOP Classes in Study</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1d19ccfc-c7bc-4be6-b7d7-62540b30a947">(0008,0062)</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cbb04151-a145-4606-a425-fce7eae26e89">O</para>
                  </td>
                </tr>
                <tr>
                  <td align="left" colspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">Anatomic Regions in Study Code Sequence</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">(0008,0063)</emphasis></emphasis></para>
                  </td>
                  <td align="center" colspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">O</emphasis></emphasis></para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para>...</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_d58500e7-ec9f-4b51-9f14-8b9435d959d7">
                      <emphasis role="italic">All other Attributes at Study Level</emphasis>
                    </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1"/>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e6ed04b3-d5d9-431a-a123-bdc3c824ed80">O</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>

    <section label="6" status="1" xml:id="chapter_6">
      <title>Registry of DICOM Data Elements</title>

      <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
        <caption>Registry of DICOM Data Elements</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                <emphasis role="bold">Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                <emphasis role="bold">Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                <emphasis role="bold">Keyword</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                <emphasis role="bold">VR</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                <emphasis role="bold">VM</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_f8f35f9d-1c46-400a-883b-d99cc79ce7f4">(0008,0062)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73566307-ca74-4d16-a340-b4b0d2457db9">SOP Classes in Study</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08f5f21a-5f51-4eb2-a268-a0f691e17a4e">SOP​Classes​In​Study</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_01729de3-f4bb-4147-80c9-ead96350034a">UI</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_b0df0da6-4eda-4e61-8d3e-5dd1916dea0d">1-n</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_f807fa48-fe34-4896-90cd-bd92aa8b74fa"/>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0008,0063)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Anatomic Regions in Study Code Sequence</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Anatomic​Regions​In​StudyC​ode​Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">SQ</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para/>
            </td>
          </tr>
        </tbody>
      </table>

    </section>

  </chapter>
</book>