    <title>CP-1648</title>
    <subtitle>Correct requirement for conversion source in legacy converted images and tractography results</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/04/18</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>QIICR</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/08/17</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1648</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct requirement for conversion source in legacy converted images and tractography results</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3 2017a</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The Image Frame Conversion Source Macro is required to encoded the SOP Instance UIDs of the
                    images from which the converted instances were converted, but this requirement precludes the use
                    of the IOD for images that have been converted from another format. Make the requirement conditional.</para>
                    <para>The Tractography Results Module also requires the Referenced Instance Sequence, which precludes the use
                    of the IOD for tractography results that were not created from DICOM images (whereas tractography
                    results may be created from proprietary format images).</para>
                 </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="A.70.4" status="3" xml:id="sect_A.70.4">
            <title>Legacy Converted Enhanced CT Image Functional Group Macros</title>
            <table frame="box" label="A.70-2" rules="all" xml:id="table_A.70-2">
                <caption>Legacy Converted Enhanced CT Image Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_75473f7f-f39d-4ed9-b95f-8ec32e04a013">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7cffb27e-adf5-4ee4-aa6d-55acb34f358c">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_00a26950-e058-495d-974c-4206a933d745">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_36401360-2170-4509-8b53-088206f41404">Image Frame Conversion Source</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ebb3d294-cbd2-42f5-b4b4-082ef049bfe3">C.7.6.16.2.25.3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_888217dd-4041-408b-94d2-854e8032f8d4"><emphasis role="bold"><emphasis role="strikethrough">M</emphasis><emphasis role="underline">C</emphasis></emphasis> - <emphasis role="bold"><emphasis role="underline">Required if created by conversion from a DICOM source. </emphasis></emphasis>May not be used as a Shared Functional Group</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="A.71.4" status="3" xml:id="sect_A.71.4">
            <title>Legacy Converted Enhanced MR Image Functional Group Macros</title>
            <table frame="box" label="A.71-2" rules="all" xml:id="table_A.71-2">
                <caption>Legacy Converted Enhanced MR Image Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_76e6cf9a-6cba-47bc-b4a7-f5c065cfcef1">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a3d8eedf-bee9-4fc4-a131-fcb505e5c1c2">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bba6bc58-c571-4ab2-9242-a80923bca5d9">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1ccfb0ac-a060-4188-9281-fc612cc2c42e">Image Frame Conversion Source</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ca8886d2-abe5-406f-ba7c-7013af80b830">C.7.6.16.2.25.3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_95504258-3aec-487f-ba8b-7883148a22ee"><emphasis role="bold"><emphasis role="strikethrough">M</emphasis><emphasis role="underline">C</emphasis></emphasis> - <emphasis role="bold"><emphasis role="underline">Required if created by conversion from a DICOM source. </emphasis></emphasis>May not be used as a Shared Functional Group</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="A.72.4" status="3" xml:id="sect_A.72.4">
            <title>Legacy Converted Enhanced PET Image Functional Group Macros</title>
            <table frame="box" label="A.72-2" rules="all" xml:id="table_A.72-2">
                <caption>Legacy Converted Enhanced PET Image Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e2110293-b8e8-4424-ab3c-8c5afe8ee519">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ce5a564a-829c-4129-8f50-08f15c13869f">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f2f63195-244b-4e64-9647-a403eb1a3379">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8ef00a78-a8e7-421b-94a1-d45ace31a83c">Image Frame Conversion Source</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_876c3b36-8514-4c9e-9b30-7bab0861f821">C.7.6.16.2.25.3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_85fd8d00-2f3a-49be-a829-f48c40d45591"><emphasis role="bold"><emphasis role="strikethrough">M</emphasis><emphasis role="underline">C</emphasis></emphasis> - <emphasis role="bold"><emphasis role="underline">Required if created by conversion from a DICOM source. </emphasis></emphasis>May not be used as a Shared Functional Group</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

<section label="C.12.1" status="3" xml:id="sect_C.12.1">
        <title>SOP Common Module</title>
        <table frame="box" label="C.12-1" rules="all" xml:id="table_C.12-1">
          <caption>SOP Common Module Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_86caed0d-a734-44a9-b3fa-f2224d719bcc">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ec64be98-77ae-4fa5-9bac-b84253ffbf75">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9165c006-7f35-4ef0-864c-d2c66ab6cfb6">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_02261e99-408e-4d8c-86fa-91d6a19a4b74">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c80378a-5239-4c66-b52a-ad12912d22e6">Conversion Source Attributes Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_597053a9-3e88-4d36-97b1-fd3eac35ba15">(0020,9172)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_aabeb561-ad2e-4b79-8d10-fa48633e134f">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b7519bb-0289-4c1d-86c2-7afdd284e9fa">The set of images or other composite SOP Instances that were converted to this instance.</para>
                <para xml:id="para_49c133b8-1aa7-4f0f-8427-3d22db9faf27">If this instance was converted from a specific frame in the source instance, the reference shall include the Frame Number.</para>
                <para xml:id="para_72587af8-e659-4c34-97ab-d4b397b4e804">One or more Items shall be included in this Sequence.</para>
                <para xml:id="para_b3528cf4-dc5f-46bb-81a0-c1e3a1833e86">Required if this instance was created by conversion<emphasis role="bold"><emphasis role="underline"> from a DICOM source</emphasis></emphasis>, and Conversion Source Attributes Sequence (0020,9172) is not present in an Item of Shared Functional Groups Sequence (5200,9229) or Per-Frame Functional Groups Sequence (5200,9230).</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para xml:id="para_082f243f-aa4d-4003-b614-9668083efb0d">
                    <emphasis role="italic">&gt;Include Table 10-3 "Image SOP Instance Reference Macro Attributes"</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be8bf53c-3d4b-430e-964d-cf684d30f28a"/>
              </td>
            </tr>            
          </tbody>
        </table>
      </section>

<section label="C.8.33.2" status="4" xml:id="sect_C.8.33.2">
    <title>Tractography Results Module</title>
  
  

  <table frame="box" label="C.8.33-2" rules="all" xml:id="table_C.8.33-2">
    <caption>Tractography Results Module Attributes</caption>
    <thead>
      <tr valign="top">
        <th align="center" colspan="1" rowspan="1">
          <para xml:id="para_60a02d67-9594-42ee-b55c-bf7a54cd6920">Attribute Name</para>
        </th>
        <th align="center" colspan="1" rowspan="1">
          <para xml:id="para_303ecc1c-54b5-4107-a91f-8ee308ebc78d">Tag</para>
        </th>
        <th align="center" colspan="1" rowspan="1">
          <para xml:id="para_b2c5c465-1a98-4c7e-9e1a-3abc4f30dbc6">Type</para>
        </th>
        <th align="center" colspan="1" rowspan="1">
          <para xml:id="para_b8887c01-755b-4cf6-b32f-da9a1f54a2c4">Attribute Description</para>
        </th>
      </tr>
    </thead>
    <tbody>
        <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
                <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para>...</para>
            </td>
        </tr>
        <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3ac2fb7-4691-4b48-b535-a982af67dfb9">Referenced Instance Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_bee26847-e734-42a2-88d8-687c40b07022">(0008,114A)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_061bae21-3a68-41b7-8406-de184536a7ef">1<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccb7fcac-9ad2-47d0-9062-5e87e31ee503">The set of images used for tractography for all Track Sets within this Instance.</para>
                <para xml:id="para_881a20d7-9ba7-4190-90d6-c97d68cacdb3">One or more items shall be included in this Sequence.</para>
                            <para><emphasis role="bold"><emphasis role="underline">Required if this instance was created from DICOM images.</emphasis></emphasis></para>
            </td>
        </tr>
        <tr valign="top">
        <td align="left" colspan="3" rowspan="1">
          <para xml:id="para_1a2f3a84-00d5-4f12-ab9f-bb5322066d45">
              <emphasis role="italic">&gt;Include Table 10-3 "Image SOP Instance Reference Macro Attributes"</emphasis>
          </para>
        </td>
        <td align="left" colspan="1" rowspan="1"/>
      </tr>
    </tbody>
  </table>
  
  
  
  </section>
    </chapter>
    
    