<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-1562</title>
    <subtitle>Correct conditions in Referenced and Derivation Image Macros when legacy converted</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/09/08</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                  xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>2015/12/06</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1562</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct conditions in Referenced and Derivation Image Macros when legacy converted</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3 2016c</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>The Referenced and Derivation Image Macros are required to be present and populated when the corresponding reference
                        sequences are present in the Classic Images that were converted, but the Macros themselves contain mandatory attributes
                        that may be absent or empty in the Classic Images and hence cannot be populated, such as Purpose of Reference and
                        Derivation Code Sequence.</para>
                     <para>Relax the conditions for the Attribute inclusion based on the SOP Class. This is preferrable to relaxing the condition on including the functional groups, since the information is useful even without the extra attributes and it is preferrable to have them in a standard location.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title></title>

        <section>
            <title></title>
            <informaltable rules="all" frame="box">
                <tr><td><para><emphasis role="italic">Amend DICOM PS3.3 as follows:</emphasis></para></td></tr>
            </informaltable>
        </section>
       
           <section label="A.70" status="2" xml:id="sect_A.70">
      <title>Legacy Converted Enhanced CT Image IOD</title>
      <section label="A.70.4" status="3" xml:id="sect_A.70.4">
        <title>Legacy Converted Enhanced CT Image Functional Group Macros</title>
        <table frame="box" label="A.70-2" rules="all" xml:id="table_A.70-2">
          <caption>Legacy Converted Enhanced CT Image Functional Group Macros</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_75473f7f-f39d-4ed9-b95f-8ec32e04a013">Functional Group Macro</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_7cffb27e-adf5-4ee4-aa6d-55acb34f358c">Section</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_00a26950-e058-495d-974c-4206a933d745">Usage</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b7dec3d-7550-4186-b17a-fd4d498030bd">Referenced Image</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d84846b8-3897-454b-9496-aedbe0d4defb">
                  <xref linkend="sect_C.7.6.16.2.5" xrefstyle="select: labelnumber"/>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7c7132a-6b70-468e-80da-8d317d05c75b">C - Required if Referenced Image Sequence (0008,1140) was present in any of the Classic Images that were converted.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e86d658-02ad-4444-9142-89382bf33baa">Derivation Image</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2853d8e-d910-4075-9e2c-a7c79faa40a5">
                  <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6e0057eb-a6df-4dfc-90e1-88c953a42204">C - Required if Source Image Sequence (0008,2112) was present in any of the Classic Images that were converted.</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
    </section>
    <section label="A.71" status="2" xml:id="sect_A.71">
      <title>Legacy Converted Enhanced MR Image IOD</title>
      <section label="A.71.4" status="3" xml:id="sect_A.71.4">
        <title>Legacy Converted Enhanced MR Image Functional Group Macros</title>
        <table frame="box" label="A.71-2" rules="all" xml:id="table_A.71-2">
          <caption>Legacy Converted Enhanced MR Image Functional Group Macros</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_76e6cf9a-6cba-47bc-b4a7-f5c065cfcef1">Functional Group Macro</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a3d8eedf-bee9-4fc4-a131-fcb505e5c1c2">Section</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_bba6bc58-c571-4ab2-9242-a80923bca5d9">Usage</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d376eca-994b-4d43-bb22-9cc870052250">Referenced Image</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4e686a1e-b7fa-400e-9430-0a4a7516f053">
                  <xref linkend="sect_C.7.6.16.2.5" xrefstyle="select: labelnumber"/>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36a7d201-53fc-47d4-b06b-3d228a3b24e8">C - Required if Referenced Image Sequence (0008,1140) was present in any of the Classic Images that were converted.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a2b13a62-214c-4dae-8019-336518190618">Derivation Image</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6aed151-9192-4e0b-9799-931cd19dc9aa">
                  <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d514d9d0-5b1b-4f73-9a35-e103103ae813">C - Required if Source Image Sequence (0008,2112) was present in any of the Classic Images that were converted.</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
    </section>
    <section label="A.72" status="2" xml:id="sect_A.72">
      <title>Legacy Converted Enhanced PET Image IOD</title>
      <section label="A.72.4" status="3" xml:id="sect_A.72.4">
        <title>Legacy Converted Enhanced PET Image Functional Group Macros</title>
        <table frame="box" label="A.72-2" rules="all" xml:id="table_A.72-2">
          <caption>Legacy Converted Enhanced PET Image Functional Group Macros</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e2110293-b8e8-4424-ab3c-8c5afe8ee519">Functional Group Macro</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ce5a564a-829c-4129-8f50-08f15c13869f">Section</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_f2f63195-244b-4e64-9647-a403eb1a3379">Usage</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d903f53-5a12-4753-8541-92d1d2464fe9">Referenced Image</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_99546e29-b218-49c2-a667-6dffa152fc31">
                  <xref linkend="sect_C.7.6.16.2.5" xrefstyle="select: labelnumber"/>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d34f0f1-d662-4432-96da-30155cea5745">C - Required if Referenced Image Sequence (0008,1140) was present in any of the Classic Images that were converted.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_27a9b477-0bfa-4112-afe9-7a8e9a605b82">Derivation Image</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6531856-31b2-4448-a882-ba58a431c18a">
                  <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ce6802d6-013e-4f81-b8b9-aa001c32a510">C - Required if Source Image Sequence (0008,2112) was present in any of the Classic Images that were converted.</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
    </section>

        <section label="C.7.6.16.2.5" status="6" xml:id="sect_C.7.6.16.2.5">
            <title>Referenced Image Macro</title>
            <table frame="box" label="C.7.6.16-6" rules="all" xml:id="table_C.7.6.16-6">
                <caption>Referenced Image Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6da3af5a-6740-420d-8c83-50088de7de32">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_643d1156-67f8-4c9d-ae2a-856eb05b4afa">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1617e757-5a1a-4854-9258-491c5c24a313">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_40b0abd4-e1d2-455d-8f4f-438dfd626bae">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9ea47561-5505-4bd5-8122-5cc2211ac857">Referenced Image Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_71cd104d-1ab4-411e-886a-8d2541a86613">(0008,1140)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f6d117e0-dba1-466f-a9a3-670cd250cb2b">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8f98e7a5-6508-45a6-bd26-78a9e633fe7b">The set of images or other composite SOP Instances used to plan the acquisition, if any, and other significant related images. See <xref linkend="sect_C.7.6.16.2.5.1" xrefstyle="select: label"/> for further explanation. Zero or more Items shall be included in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_be291764-e4ed-45a7-90b0-7e6aa8c13388">
                                <emphasis role="italic">&gt;Include <xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_57849f9c-0a95-416f-bc87-e13fe33b607c"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d74edb80-e80b-4223-b740-9fc01df338c2">&gt;Purpose of Reference Code Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a73187f6-af8e-4f32-ac14-c89653b09fba">(0040,A170)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fd351f5f-097e-4aa5-8831-4edeb86f1f77">1<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d3002b7b-2b97-44f0-b899-b410ab297b90">Describes the purpose for which the reference is made.</para>
                            <para xml:id="para_b7e03f81-d808-4684-a46a-b990cec3e025">Only a single Item shall be included in this Sequence.</para>
                            <para><emphasis role="bold"><emphasis role="underline">Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage) and not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage) and not "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), may be present otherwise.</emphasis></emphasis></para>
                            <para xml:id="para_951dde0e-b932-49e0-a9af-c2f86d1c72e0">See <xref linkend="sect_C.7.6.16.2.5.1" xrefstyle="select: label"/> for further explanation.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_798e3224-8b34-4b68-8bf7-563de1a27d64">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_41efa90f-b594-486f-ab67-af6819e61a43">Defined <olink targetdoc="PS3.16" targetptr="sect_CID_7201" xrefstyle="select: labelnumber quotedtitle"/>, or as specified in the IOD invocation of this Functional Group.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.7.6.16.2.5.1" status="7" xml:id="sect_C.7.6.16.2.5.1">
                <title>Use of Referenced Image Macro</title>
                <para xml:id="para_3161b247-76d4-4ba2-8b82-7436fb6b1e20">Referenced Image Sequence (0008,1140) shall be used to identify other data objects used to plan the acquisition of this image. Instances referenced for this purpose shall have the same Frame of Reference UID (0020,0052) as this image. For each Item that contains such a reference, the value of Purpose of Reference Code Sequence (0040,A170) shall be ("121311", DCM, "Localizer"). Applications can use the Referenced Image Sequence (0008,1140) in combination with data in Plane Position (Patient) and Plane Orientation (Patient) Macros to provide projections of the position of an image with respect to the referenced image.</para>
                <para xml:id="para_b6d2c1e4-48f7-4574-9018-20ef8bb538a1">The Referenced Image Sequence (0008,1140) may also be present when references to other images (or frames within other images) are required for other reasons, as specified by Purpose of Reference Code Sequence (0040,A170).</para>
                <note>
                    <para xml:id="para_d7deec63-f185-46d2-a0e1-bc5f0f2e8d71">An Image may contain references to itself (e.g., to other frames within itself).</para>
                </note>
            </section>
        </section>
        <section label="C.7.6.16.2.6" status="6" xml:id="sect_C.7.6.16.2.6">
            <title>Derivation Image Macro</title>
            <table frame="box" label="C.7.6.16-7" rules="all" xml:id="table_C.7.6.16-7">
                <caption>Derivation Image Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ac3b7fe-6624-4256-a133-dec8c40607fc">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9128492c-276c-4069-8891-6db1e5d438d8">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b1aa53c0-c57b-4cce-918c-1dc7c988337d">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3adbfc2b-51c4-41a4-9af2-3f44b49558ab">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d69233c4-13da-4c68-8516-a043ccd4940d">Derivation Image Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5ac27ef2-c4d9-4e1b-a9a0-d6e93cea0ac8">(0008,9124)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_66119f32-1e30-4666-b2c5-e3a311b2e777">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_86967de3-0386-4fd0-995e-a8d23bce3101">The set of Images or other composite SOP Instances that were used to derive this frame.</para>
                            <para xml:id="para_c720dda0-7765-47e0-b3d6-e9e3c5a2c2d0">Zero or more Items shall be included in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ec35d0a9-7a09-448a-8ee8-973ab04979ed">&gt;Derivation Description</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3b8d54c6-72cd-4cbe-b1e6-26fd871e8c35">(0008,2111)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4f97e423-7eea-44b1-8d6f-5ec569651210">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3732fc9f-fae2-40d8-a991-06a119e79024">A text description of how this frame data was derived. See <xref linkend="sect_C.7.6.1.1.3" xrefstyle="select: label"/> for further explanation.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_df50da99-906c-4bf2-b03e-7904644dc62e">&gt;Derivation Code Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9b3e6f6f-e448-4c1c-852c-334f620eb68e">(0008,9215)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_490b633c-1dcb-4e5f-b7f4-d7c95a7a589d">1<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1"><para xml:id="para_c11ee7fa-7cad-4818-852f-09d54a0783b5">A coded description of how this frame was derived. See <xref linkend="sect_C.7.6.1.1.3" xrefstyle="select: label"/> for further explanation.</para>One or more Items shall be included in this Sequence. More than one Item indicates that successive derivation steps have been applied.<para><emphasis role="bold"><emphasis role="underline">Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage) and not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage) and not "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), may be present otherwise.</emphasis></emphasis></para></td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_9b13f626-4fd7-48d1-ae9a-2f35f344f8e1">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_779715a9-1411-41ea-be1f-47a9ab4216e1">
                                <emphasis role="italic">Defined <olink targetdoc="PS3.16" targetptr="sect_CID_7203" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_38842ac5-d9cb-4264-bf88-6f7d4748980c">&gt;Source Image Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e4dc5f98-f26a-469c-981f-854dbd16bf60">(0008,2112)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_892a80d1-19e5-425e-a402-4a66d0707445">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b122d178-47e0-44b8-b201-16868702bd55">The set of Images or other Composite SOP Instances that were used to derive this frame.</para>
                            <para xml:id="para_ada48db8-a193-4052-8cac-43731a5d36ba">Zero or more Items shall be included in this Sequence. See <xref linkend="sect_C.7.6.1.1.4" xrefstyle="select: label"/> for further explanation.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_78fdad00-5bff-42df-b965-0b0c055cff5c">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_afad1d04-5991-4730-9b67-832074720213"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_aa7b20eb-afb5-4e5b-a2d7-37f7bc31e57f">&gt;&gt;Purpose of Reference Code Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f928d8cd-3306-4411-ab4f-ec0f4f755a1c">(0040,A170)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_898c0fa2-6909-4b7b-a8ca-83c226adce4f">1<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e0340a5f-4bad-4257-beb4-6bb66c01995e">Describes the purpose for which the reference is made, that is what role the source image or frame played in the derivation of this image or frame.</para>
                            <para xml:id="para_67e2bb55-1b23-4b2e-87f7-17b04633df02">Only a single Item shall be included in this Sequence.</para>
                            <para><emphasis role="bold"><emphasis role="underline">Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage) and not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage) and not "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), may be present otherwise.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_beb90cf3-f6e9-4232-b4d4-2f62c7224576">
                                <emphasis role="italic">&gt;&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_30946779-c590-460d-acaf-c7827cb031d0">
                                <emphasis role="italic">Defined <olink targetdoc="PS3.16" targetptr="sect_CID_7202" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1392a7a6-47ce-4682-84b5-04aec6523919">&gt;&gt;Spatial Locations Preserved</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_704357f9-bb95-4324-8b76-800b6a1bc45a">(0028,135A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_215cb573-dc45-485a-b814-aac03f4e90bd">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cc4c42a5-016b-410f-a0d8-f8529bf27787">The extent to which the spatial locations of all pixels are preserved during the processing of the source image that resulted in the current image or frame.</para>
                            <variablelist spacing="compact">
                                <title>Enumerated Values:</title>
                                <varlistentry>
                                    <term>YES</term>
                                    <listitem>
                                        <para xml:id="para_7cc30ce6-9149-408d-bd76-5a4d4a4f51ab"/>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>NO</term>
                                    <listitem>
                                        <para xml:id="para_44d55035-81bd-47b9-8d6a-a159c22e44cb"/>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>REORIENTED_ONLY</term>
                                    <listitem>
                                        <para xml:id="para_e6084561-4aa3-4f79-aed7-359bf76522fd">A projection radiograph that has been flipped, and/or rotated by a multiple of 90 degrees</para>
                                    </listitem>
                                </varlistentry>
                            </variablelist>
                            <note>
                                <orderedlist>
                                    <listitem>
                                        <para xml:id="para_59e61333-5d11-4c89-b1de-409d8221b2b8">This applies not only to images with a known relationship to a 3D space, but also to projection images.
                                            For example, a projection radiograph such as a mammogram that is processed by a point image processing operation such as contrast enhancement, or a smoothing or edge enhancing convolution, would have a value of YES for this attribute.
                                            A projection radiograph that had been magnified or warped geometrically would have a value of NO for this attribute.
                                            A projection radiograph that has been flipped, and/or rotated by a multiple of 90 degrees, such that transformation of pixel locations is possible by comparison of the values of Patient Orientation (0020,0020) would have a value of REORIENTED_ONLY.
                                            This attribute is typically of importance in relating images with Presentation Intent Type (0008,0068) values of FOR PROCESSING and FOR PRESENTATION.</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_02e3e4a3-4ad1-4503-9d75-3f5ecf01c633">When the value of this attribute is NO, it is not possible to locate on the current image any pixel coordinates that are referenced relative to the source image, such as for example,
                                            might be required for rendering CAD findings derived from a referenced FOR PROCESSING image on the current FOR PRESENTATION image.</para>
                                    </listitem>
                                </orderedlist>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5855dee4-56fd-4dcb-bded-927ea41e8972">&gt;&gt;Patient Orientation</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8df6d030-f32f-45e4-a728-27114aea4f04">(0020,0020)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e9b5e61d-2598-4f15-8ea8-6a18b35a896e">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_872e4463-f77b-4ebc-bda8-1a77b01c7867">The Patient Orientation values of the source image.</para>
                            <para xml:id="para_fcd04da0-4438-4712-8058-c66e05a8fdd6">Required if the value of Spatial Locations Preserved (0028,135A) is REORIENTED_ONLY.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        

    </chapter>
</book>