  <title>CP-1367</title>
  <subtitle>Correct errors in Colon CAD SR IOD Template Figures</subtitle>
  <chapter>
    <title></title>
    <informaltable rules="all" frame="box">
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>STATUS</para></td>
      <td rowspan="1" colspan="1" valign="top"><para>Final Text</para></td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
      <td rowspan="1" colspan="1" valign="top"><para>2014/11/11</para></td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
      <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
        xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
      <td rowspan="1" colspan="1" valign="top"><para>James J Rogers</para><para><link
        xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:jr@coastalcg.com"/></para></td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
      <td rowspan="1" colspan="1" valign="top"><para>2014/01/17</para></td>
    </tr>
  </informaltable>
    <informaltable rules="all" frame="box">
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1367</para></td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct errors in Colon CAD SR IOD Template Figures</para></td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top">
        <para>Name of Standard</para>
        <para>PS3.16 2014b</para>
      </td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top">
        <para>Rationale for Correction:</para><para>Figure
            A-9b Colon CAD SR IOD Template Structure is inconsistent with the template.
            Specifically:</para>
        <para>• The second instance of TID 4127 Single image finding does
            not indicate possible recursion.</para>
        <para>• The second instance of TID 4125 Composite
            feature has a subordinate TID 4019 CAD Algorithm Identification, however, the first
            instance of TID 4125 does not.</para>
        <para>• The first instance of TID 4127 Single Image
            finding has a subordinate TID 4019 CAD Algorithm Identification.</para>
        <para>• TID 4126
            Composite feature body does not have a subordinate TID 4019 CAD Algorithm
            Identification, which may be inconsistent with other SR IOD template
            structures.</para>
        <para>This CP updates the figure to correct these findings.</para>
      </td>
    </tr>
    <tr>
      <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
    </tr>
  </informaltable>
  </chapter>
  <chapter>
    <title></title>
      <section label="TID 4120" status="3" xml:id="sect_TID_4120">
          <title>Colon CAD Document Root Template</title>
          <para>...</para>
        <informaltable rules="all" frame="box">
          <tr><td><para><emphasis role="italic">Replace the existing Figure A-9b shown here:</emphasis></para></td></tr>
        </informaltable>
        <para>
          <mediaobject>
            <imageobject>
              <imagedata fileref="cp1367_coloncadfigure/PS3.16_A-9b_old.svg" format="SVG"/>
            </imageobject>
          </mediaobject>
        </para>
        <informaltable rules="all" frame="box">
          <tr><td><para><emphasis role="italic">with the new Figure A-9b shown here:</emphasis></para></td></tr>
        </informaltable>
        <para>
          <mediaobject>
            <imageobject>
              <imagedata fileref="cp1367_coloncadfigure/PS3.16_A-9b_new.svg" format="SVG"/>
            </imageobject>
          </mediaobject>
        </para>
        <para>...</para>