<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2628</title>
    <subtitle>Add KOS concepts related to WSI scan quality</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Nov 2026 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2026/09/13</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Bart Diricx</para><para><link
                    xl:href="mailto:bart.diricx@barco.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2026/03/17</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2628</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add KOS concepts related to WSI scan quality</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>In a typical workflow, a scan tech or algorithm will check the quality of a WSI (has all tissue been scanned,
                        is there not too much of the tissue area out-of-focus, ...) and either approve or disapprove that slide.</para>
                    <para>There is a need to document the result of that analysis.</para>
                    <para>These two scenarios are examples:</para>
                    <orderedlist>
                        <listitem>
                            <para>QC performed and passed (QC OK)</para>
                            <para>A positive QC decision should be conveyed explicitly using a Document Title from CID 7010.</para>
                            <para>Document Title: (ddd0000, DCM, "No quality issues detected")</para>
                            <para>This is distinct from the concept of (113034, DCM, "Complete Acquisition Content"), which may be conveyed in a separate KOS created by the scanner</para>
                            <para>No Document Title Modifier is needed.</para>
                            <para>The person or device Observer Context will identify who/what performed the assessment.</para>
                            <para>The presence of this KOS instance explicitly indicates that QC has been performed and that the WSI is acceptable, allowing it to be distinguished from a slide that has not yet been reviewed.</para>
                        </listitem>
                        <listitem>
                            <para>QC performed and failed (QC NOK / Rejected)</para>
                            <para>A negative QC decision should be conveyed explicitly using a Document Title from CID 7010.</para>
                            <para>Document Title: (113001, DCM, "Rejected for Quality Reasons")</para>
                            <para>Document Title Modifiers - similar to the radiology use but WSI-specific and re-using the region segmentation pathology QC set [CP 2435].</para>
                        </listitem>
                    </orderedlist>
                    <para>The Observer Content (TID 1002) can describe the human or device (or both) that were involved in the QC decision.</para>
                    <para>Optionally, a more detailed description of the QC findings can be included in the  Key Object Description (optional): e.g., "Automated QC detected focus degradation and air bubble artifact".</para>
                    <para>By using the KOS, the granularity of the finding (whether it applies to a single image of a set of images for an image acquisition, or the entire set, and whether or not the determination was made on a single image (e.g., the OVERVIEW rather than a layer of the pyramid (VOLUME or THUMBNAIL) but applies to the whole set (e.g., a missing tissue fragment in the pyramid that is apparent in the OVERVIEW), is not distinguished, since the intent is to provide actionable information, not the provenance of the determination. Nor does this CP provide a means of spatially locating on the OVERVIEW the missing tissue fragment (which could be encoded in an SR or MBSA or SEG).</para>
                    <para>The definitions used related to findings that were previously introduced for segmentation/localization purposes are generalized to be applicable for both localization and categorical quality issues.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        
        
        
        
        
        
        

        

        

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <section label="CID 7011" status="3" xml:id="sect_CID_7011">
            <title>Rejected for Quality Reason</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Resources:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_ff5ca6a9-5d6e-4ce4-9af8-fb999c02484d">
                            <emphasis role="bold">
                                <link xl:href="http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_7011.html" xl:show="new">HTML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/json/ValueSet-dicom-cid-7011-RejectedForQualityReason.json" xl:show="new">FHIR JSON</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/xml/ValueSet-dicom-cid-7011-RejectedForQualityReason.xml" xl:show="new">FHIR XML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/ihe/svs/1.2.840.10008.6.1.491.xml" xl:show="new">IHE SVS XML</link>
                            </emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_930794c3-124a-45e6-a250-836c8a10d52d">
                            <emphasis role="bold">RejectedForQualityReason</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">FHIR Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_0cc9e771-7297-458d-ab0c-c955502fdeb0">
                            <emphasis role="bold">dicom-cid-7011-RejectedForQualityReason</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_b5b6855f-e5b5-49ad-99ec-dc958d4b1e50">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_e2e26106-c7d7-4768-aed3-7f1c4dde2042">
                            <emphasis role="bold"><emphasis role="strikethrough">20020904</emphasis><emphasis role="underline">yyyymmdd</emphasis></emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">UID:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_75b785b3-4b2d-4475-aa97-d8e93d7eabe1">
                            <emphasis role="bold">1.2.840.10008.6.1.491</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7011" rules="all" xml:id="table_CID_7011">
                <caption>Rejected for Quality Reason</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a06a3871-7fc0-4d31-9957-1ff71ab4b9ad">Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_94324dc7-ad2a-4f70-9360-060c6e77a63b">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b4fb24e6-8651-4f87-a9ee-e8191765d1e9">Code Meaning</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0ebb9a7-59ac-4794-8145-16da9869331f">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6ed99ae5-1597-413d-8c37-c73f9a37e4c1">
                                <link linkend="DCM_111207" xrefstyle="select: nopage">111207</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_feebd2a3-a2af-4966-910c-cd43dfa3b14d">Image artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e4a9bf18-a020-4fc3-b34e-675b608573d4"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_792472eb-4b4e-4cb4-ab40-6d6a373c22c5"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111208" xrefstyle="select: nopage">111208</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_50a8dba8-9c78-49c6-950d-0e4179b3a6d7"><emphasis role="bold"><emphasis role="strikethrough">Grid artifact(s)</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af9b902-d6f9-46db-bc8f-d5faa9b9ab86"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ab7bdcb-c506-43de-a0a0-b862cddb9faf"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111209" xrefstyle="select: nopage">111209</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_10566e2f-71fd-48f5-8764-2aaac48d2c84"><emphasis role="bold"><emphasis role="strikethrough">Positioning</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_39774aad-6ebf-41dc-bd07-0515390150bb"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42699deb-511e-4b9f-accd-0468564f6f50"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111210" xrefstyle="select: nopage">111210</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_354837cd-6925-476e-bbb4-7212f10063fa"><emphasis role="bold"><emphasis role="strikethrough">Motion blur</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2086b79c-262f-4cdc-8676-64511220debd"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_84853ccd-053d-4aa7-8b83-be678c2285d3"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111211" xrefstyle="select: nopage">111211</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_01ce9134-24bb-4c61-918d-91028199d121"><emphasis role="bold"><emphasis role="strikethrough">Under exposed</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_efd9008b-d56d-40b3-a7d7-53cff4ad43fc"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b0ecf752-750f-4b58-8e68-77e8bb85c379"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111212" xrefstyle="select: nopage">111212</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3b8ba995-9203-4d64-8424-455bab7860f7"><emphasis role="bold"><emphasis role="strikethrough">Over exposed</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b9e22841-f0ab-4d9d-9862-a8f1c72b3ca3">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3b5d66d3-1bf1-40ef-82c8-98607899cb79">
                                <link linkend="DCM_111213" xrefstyle="select: nopage">111213</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a30b2051-8f27-4dea-9e1c-317d5213afa3">No image</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_26b8e179-d7a9-4b81-95b9-47fe548156ed">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4e9102cb-54a4-4060-880f-d455f2d76768">
                                <link linkend="DCM_111214" xrefstyle="select: nopage">111214</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e919003e-4800-48be-b669-de22dd5cf415">Detector artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_26b8e179-d7a9-4b81-95b9-47fe548156ed"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4e9102cb-54a4-4060-880f-d455f2d76768"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111214" xrefstyle="select: nopage">111214</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e919003e-4800-48be-b669-de22dd5cf415"><emphasis role="bold"><emphasis role="strikethrough">Detector artifact(s)</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_71a690dd-378e-4c3e-a0a6-ec8587521077"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7fe70faf-76c1-4344-adde-e5c2db2c936a"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_111215" xrefstyle="select: nopage">111215</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8918233a-3da4-42fe-83cb-1e4716ff9ef3"><emphasis role="bold"><emphasis role="strikethrough">Artifact(s) other than grid or detector artifact</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6cee293b-0a37-41cd-a0f2-aa951074e168">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_acd2078b-0daf-4f61-8b81-f4c3125791f9">
                                <link linkend="DCM_111216" xrefstyle="select: nopage">111216</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d425028c-db9a-4dec-a759-3c825b469a64">Mechanical failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e7cca2cd-67fc-40dc-a42a-53b9a15c90b3">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_30245332-9564-4e72-b341-f9acf3ab2252">
                                <link linkend="DCM_111217" xrefstyle="select: nopage">111217</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8f246218-97bd-49b0-bb2a-a35093106f55">Electrical failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f596c8e3-93a6-415a-8d8d-7f50e8ac55ae">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c6219a10-292d-47da-9134-142caf9a5bf3">
                                <link linkend="DCM_111218" xrefstyle="select: nopage">111218</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_206ea16c-3d8d-4f94-a6ac-1fa02e822917">Software failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4cc1684f-c6f7-4e8b-bb9e-fb10704a09e8">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fe6376a8-ec07-41a8-bd1d-daa9b3fb99da">
                                <link linkend="DCM_111219" xrefstyle="select: nopage">111219</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_417bd8ac-27bc-44c4-8623-a2991e65c40f">Inappropriate image processing</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_55a77f6a-7c6d-478e-9122-6f92a5100afa">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_890b77de-6d3f-47a2-ab42-22f8918af10d">
                                <link linkend="DCM_111220" xrefstyle="select: nopage">111220</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_87bf0ebe-430c-4d5e-8d0a-181c273729f8">Other failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1ba94aa3-b5ef-4af4-a20c-5796ecf13286">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e5b77798-88aa-45aa-a187-726d01ecc0ce">
                                <link linkend="DCM_111221" xrefstyle="select: nopage">111221</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_794bafc7-546c-421c-b8aa-7ad37a96291a">Unknown failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3018a27a-8806-4447-b712-6f602ef625c8"><emphasis role="bold"><emphasis role="strikethrough">DCM</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0aeb6408-f8a0-472c-9c24-52d7f494bddd"><emphasis role="bold"><emphasis role="strikethrough">
                                        <link linkend="DCM_113026" xrefstyle="select: nopage">113026</link>
                                    </emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8692f0b7-89c3-46a4-bc23-26f534343bef"><emphasis role="bold"><emphasis role="strikethrough">Double exposure</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">Include <xref linkend="sect_CID_newCIDnnn1" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </emphasis></emphasis></emphasis>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">Include <xref linkend="sect_CID_newCIDnnn2" xrefstyle="select: labelnumber quotedtitle nopage"/>
                            </emphasis></emphasis></emphasis>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Add new Context Groups to DICOM PS3.16 as follows :</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="CID newCIDnnn1" status="3" xml:id="sect_CID_newCIDnnn1">
            <title>Rejected for Quality Reason - Radiology</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Resources:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">
                                <link xl:href="http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_newCIDnnn1.html" xl:show="new">HTML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/json/ValueSet-dicom-cid-newCIDnnn1-RejectedForQualityReasonRadiology.json" xl:show="new">FHIR JSON</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/xml/ValueSet-dicom-cid-newCIDnnn1-RejectedForQualityReasonRadiology.xml" xl:show="new">FHIR XML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/ihe/svs/1.2.840.10008.6.1.newCIDuuu1.xml" xl:show="new">IHE SVS XML</link>
                            </emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">RejectedForQualityReasonRadiology</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">FHIR Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">dicom-cid-newCIDnnn1-RejectedForQualityReasonRadiology</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">yyyymmdd</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">UID:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">1.2.840.10008.6.1.newCIDuuu1</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID newCIDnnn1" rules="all" xml:id="table_CID_newCIDnnn1">
                <caption>Rejected for Quality Reason - Radiology</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a06a3871-7fc0-4d31-9957-1ff71ab4b9ad">Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_94324dc7-ad2a-4f70-9360-060c6e77a63b">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b4fb24e6-8651-4f87-a9ee-e8191765d1e9">Code Meaning</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e4a9bf18-a020-4fc3-b34e-675b608573d4">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_792472eb-4b4e-4cb4-ab40-6d6a373c22c5">
                                <link linkend="DCM_111208" xrefstyle="select: nopage">111208</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_50a8dba8-9c78-49c6-950d-0e4179b3a6d7">Grid artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af9b902-d6f9-46db-bc8f-d5faa9b9ab86">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ab7bdcb-c506-43de-a0a0-b862cddb9faf">
                                <link linkend="DCM_111209" xrefstyle="select: nopage">111209</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_10566e2f-71fd-48f5-8764-2aaac48d2c84">Positioning</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_39774aad-6ebf-41dc-bd07-0515390150bb">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42699deb-511e-4b9f-accd-0468564f6f50">
                                <link linkend="DCM_111210" xrefstyle="select: nopage">111210</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_354837cd-6925-476e-bbb4-7212f10063fa">Motion blur</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2086b79c-262f-4cdc-8676-64511220debd">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_84853ccd-053d-4aa7-8b83-be678c2285d3">
                                <link linkend="DCM_111211" xrefstyle="select: nopage">111211</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_01ce9134-24bb-4c61-918d-91028199d121">Under exposed</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_efd9008b-d56d-40b3-a7d7-53cff4ad43fc">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b0ecf752-750f-4b58-8e68-77e8bb85c379">
                                <link linkend="DCM_111212" xrefstyle="select: nopage">111212</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3b8ba995-9203-4d64-8424-455bab7860f7">Over exposed</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_26b8e179-d7a9-4b81-95b9-47fe548156ed">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4e9102cb-54a4-4060-880f-d455f2d76768">
                                <link linkend="DCM_111214" xrefstyle="select: nopage">111214</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e919003e-4800-48be-b669-de22dd5cf415">Detector artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_71a690dd-378e-4c3e-a0a6-ec8587521077">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7fe70faf-76c1-4344-adde-e5c2db2c936a">
                                <link linkend="DCM_111215" xrefstyle="select: nopage">111215</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8918233a-3da4-42fe-83cb-1e4716ff9ef3">Artifact(s) other than grid or detector artifact</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3018a27a-8806-4447-b712-6f602ef625c8">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0aeb6408-f8a0-472c-9c24-52d7f494bddd">
                                <link linkend="DCM_113026" xrefstyle="select: nopage">113026</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8692f0b7-89c3-46a4-bc23-26f534343bef">Double exposure</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="CID newCIDnnn2" status="3" xml:id="sect_CID_newCIDnnn2">
            <title>Rejected for Quality Reason - Slide Microscopy</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Resources:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">
                                <link xl:href="http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_newCIDnnn2.html" xl:show="new">HTML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/json/ValueSet-dicom-cid-newCIDnnn2-RejectedForQualityReasonSlideMicroscopy.json" xl:show="new">FHIR JSON</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/xml/ValueSet-dicom-cid-newCIDnnn2-RejectedForQualityReasonSlideMicroscopy.xml" xl:show="new">FHIR XML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/ihe/svs/1.2.840.10008.6.1.newCIDuuu2.xml" xl:show="new">IHE SVS XML</link>
                            </emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">RejectedForQualityReasonSlideMicroscopy</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">FHIR Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">dicom-cid-newCIDnnn2-RejectedForQualityReasonSlideMicroscopy</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">yyyymmdd</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">UID:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">1.2.840.10008.6.1.newCIDuuu2</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID newCIDnnn2" rules="all" xml:id="table_CID_newCIDnnn2">
                <caption>Rejected for Quality Reason - Slide Microscopy</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para>Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>SNOMED-RT ID</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>UMLS Concept Unique ID</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131502" xrefstyle="select: nopage">131502</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Unusable tissue</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131504" xrefstyle="select: nopage">131504</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Out of focus</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131505" xrefstyle="select: nopage">131505</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Cover slip defect</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131506" xrefstyle="select: nopage">131506</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Folded tissue</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131508" xrefstyle="select: nopage">131508</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Inadequately stained</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131509" xrefstyle="select: nopage">131509</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Cut off tissue</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131510" xrefstyle="select: nopage">131510</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Coverslip edge</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131511" xrefstyle="select: nopage">131511</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Slide marking</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131512" xrefstyle="select: nopage">131512</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Stitching</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131513" xrefstyle="select: nopage">131513</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Stripes</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131514" xrefstyle="select: nopage">131514</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Foreign material</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131515" xrefstyle="select: nopage">131515</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Scratch</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131516" xrefstyle="select: nopage">131516</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Tissue marking</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link linkend="DCM_131517" xrefstyle="select: nopage">131517</link></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Unscanned tissue</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>SCT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link xl:href="http://snomed.info/id/56117007" xl:show="new">56117007</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Air bubble</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link xl:href="http://snomed.info/id/56117007" xl:show="new">A-80232</link></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><link xl:href="http://uts.nlm.nih.gov/uts/umls/concept/C0001863" xl:show="new">C0001863</link></para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
<section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding Scheme Version "01")</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">Definition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">Notes</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131501">131501</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Usable tissue</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that contains t</emphasis><emphasis role="underline">T</emphasis></emphasis>issue that is usable.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131502">131502</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Unusable tissue</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that contains t</emphasis><emphasis role="underline">T</emphasis></emphasis>issue that is not usable.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131503">131503</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>In focus</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that i</emphasis><emphasis role="underline">I</emphasis></emphasis>s in focus.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131504">131504</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Out of focus</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that i</emphasis><emphasis role="underline">I</emphasis></emphasis>s not in focus.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131505">131505</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Cover slip defect</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample where t</emphasis><emphasis role="underline">T</emphasis></emphasis>here is a defect in the slide cover slip (such as a crack or chip).</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131506">131506</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Folded tissue</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains folded tissue.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131507">131507</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Adequately stained</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that i</emphasis><emphasis role="underline">I</emphasis></emphasis>s adequately stained.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131508">131508</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Inadequately stained</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that i</emphasis><emphasis role="underline">I</emphasis></emphasis>s not adequately stained.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131509">131509</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Cut off tissue</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample where t</emphasis><emphasis role="underline">T</emphasis></emphasis>issue is cut off at the edge of the viewable area.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr><tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131510">131510</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Coverslip edge</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains the edge of the coverslip.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr><tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131511">131511</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Slide marking</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains pen markings applied to the slide.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr><tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131512">131512</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Stitching</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains tile stitching discontinuities.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131513">131513</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Stripes</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains a periodic illumination non-uniformity pattern.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr><tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131514">131514</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Foreign material</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains foreign material (such as dust, hair, thread, fingerprint).</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr><tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131515">131515</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Scratch</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that c</emphasis><emphasis role="underline">C</emphasis></emphasis>ontains a scratch on the glass slide or coverslip.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131516">131516</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Tissue marking</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region within a sample that is c</emphasis><emphasis role="underline">C</emphasis></emphasis>olo<emphasis role="bold"><emphasis role="strikethrough">u</emphasis></emphasis>red with a marker applied to the tissue (such as surgical ink).</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_131517">131517</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Unscanned tissue</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="strikethrough">A region outside a sample that corresponds to t</emphasis><emphasis role="underline">T</emphasis></emphasis>issue not digitized at high resolution.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference DICOM PS3.16:</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="TID 2010" status="3" xml:id="sect_TID_2010">
                <title>Key Object Selection</title>
                <para xml:id="para_8f0100c0-8fda-48e1-8709-ca2aa5f78198">The Key Object Selection Template is intended for flagging one or more significant images, waveforms, or other composite SOP Instances. Key Object Selection contains:</para>
                <itemizedlist>
                    <listitem>
                        <para xml:id="para_76b69bac-1fd5-4bb9-8b82-b2c7c601a9b8">coded document title stating the reason for significance of the referenced objects in the Key Object Selection,</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_4d3593f7-a81a-48ca-8b3f-0ff02d0a4b4b">optional free form text comment in an explicitly identified language, and</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_8a97f5aa-54fe-499d-abc3-bb3d871ba4ff">optional identification of the observer (device or person) that created the Key Object Selection.</para>
                    </listitem>
                </itemizedlist>
                <note>
                    <orderedlist>
                        <listitem>
                            <para xml:id="para_ed721201-28e1-4871-bfe6-eaf667b1e577">For instance, when this Template is used to identify images rejected for quality reasons, the device or person performing the quality assessment is identified in observation context items (invoked through <xref linkend="sect_TID_1002" xrefstyle="select: labelnumber quotedtitle nopage"/>). The reason for rejection can be included both as a code used as a concept modifier for the document title, and as text description.</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_61c14115-e4bb-46df-ba14-18ef5edd151b">The order of object references may be significant, e.g., when the title concept is "For Conference".</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_c08be380-68b6-44f2-a078-eccce3dc13df">Instances referenced in a Key Object Selection Document may be securely referenced by Digital Signature or MAC mechanisms within the <olink targetdoc="PS3.3" targetptr="sect_C.17.2" xrefstyle="select: title"/> (see <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>).</para>
                        </listitem>
                    </orderedlist>
                </note>
                <para xml:id="para_5fd4b522-d007-44da-8e2d-e27abd9ca4f4">The Template can only be instantiated at the root node and cannot be included in other Templates. The Template is not extensible; that is, no other Content Items may be added to this Template, or the Templates that are included, recursively.</para>
                <variablelist spacing="compact">
                    <varlistentry>
                        <term>
                            <emphasis role="bold">Type:</emphasis>
                        </term>
                        <listitem>
                            <para xml:id="para_221627cc-b8bb-47ff-a350-07629aa7962a">
                                <emphasis role="bold">Non-Extensible</emphasis>
                            </para>
                        </listitem>
                    </varlistentry>
                    <varlistentry>
                        <term>
                            <emphasis role="bold">Order:</emphasis>
                        </term>
                        <listitem>
                            <para xml:id="para_4d0a34dc-63da-4fba-9b53-cc6e34457880">
                                <emphasis role="bold">Non-Significant</emphasis>
                            </para>
                        </listitem>
                    </varlistentry>
                    <varlistentry>
                        <term>
                            <emphasis role="bold">Root:</emphasis>
                        </term>
                        <listitem>
                            <para xml:id="para_84925155-bf88-4558-8b4f-974fc4e8c20f">
                                <emphasis role="bold">Yes</emphasis>
                            </para>
                        </listitem>
                    </varlistentry>
                </variablelist>
                <table frame="box" label="TID 2010" rules="all" xml:id="table_TID_2010">
                    <caption>Key Object Selection</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1a9a5865-3c13-4a6d-aadb-bec8b96224b9"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d843a45e-d8f6-4422-aad1-f63843b0889a">NL</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bcb247d9-2352-4c73-a696-7b6d00eba95b">Rel with Parent</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_15fad090-959c-4e82-8d95-e30f987f877e">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3362e379-751f-4b4d-bc70-8fe9fa61e52b">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_700fca3f-7bc7-437d-b597-f9fdbee76f6d">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_4dcdcec3-b151-4407-b795-e54c0d0f7bc0">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_77c7fed3-8787-49c1-bc9b-6ee72cf3c0b7">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c38ee96e-ae23-4c24-bb07-48e0edc0b8ce">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3d02bcc5-2843-4ba9-aad8-241dca5d1715">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_808aa828-5364-45f6-a189-290946007b63"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c668b594-dabd-49e5-a176-c84963d15b3b"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_339e44e0-c5ba-440d-9694-082f13dad390">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ce4aba56-4d29-418f-804d-bfe49cb87b79">D<xref linkend="sect_CID_7010" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_407fe943-292f-46db-8118-c288ec3a5a19">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c4abe852-99e4-4bdf-8464-2a8aba74dc71">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8cc279e7-e38a-4a59-9741-3ee230ce37a1"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0b092c07-80f0-4819-aafa-9cf72a6d39b1">Root node</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bb5f3820-9cab-47c6-81f3-a1b382749f27">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9a140bd6-7aa1-4cd6-99a2-11e59c54ed12">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_65cb6067-a183-4591-a29c-03b46660fc7d">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_13dcbaa4-3007-45dd-8cc5-17e801200860">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d8fc4a46-9199-481a-988d-935f9c7541d5">EV <link linkend="DCM_113011" xrefstyle="select: nopage">(113011, DCM, "Document Title Modifier")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c920d0a9-c4e0-427d-81e9-c5d322a4964e">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_86799f94-0e8c-49bc-9e09-537be8864f35">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fa48f3be-c248-4d8e-8d3a-2c240e280246"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_42c06426-36c2-4da8-8bae-7d6b70584211"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1377d767-e9a8-427b-a10e-db8ab8b9caab">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6c703e20-af07-44be-aa2d-b73bcd7d703d">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_08b3666c-d09d-4c89-86b0-0409c671e1fd">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d3c5d5ac-c9c9-44f5-8471-e47688f25de0">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_17acace2-32ff-467d-9021-8499f5d6b4a7">EV <link linkend="DCM_113011" xrefstyle="select: nopage">(113011, DCM, "Document Title Modifier")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_91c6db19-af50-4c7e-8391-7cdcf02839d1">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f3a2b8b1-f1af-4eca-bc43-02852f48a1d9">UC</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ef0c9d17-60b3-429b-86fa-fcf2a66a36ab">IF Row 1 Concept Name = <link linkend="DCM_113001" xrefstyle="select: nopage">(113001, DCM, "Rejected for Quality Reasons")</link> or <link linkend="DCM_113010" xrefstyle="select: nopage">(113010, DCM, "Quality Issue")</link>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7c95ce81-4b47-4aa5-96f1-bceac319ed7e">D<xref linkend="sect_CID_7011" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0c403633-88e1-4f56-90a0-bef48798ab02">4</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1641b290-d080-4e0e-8079-26a6d5f281e6">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a7180eb8-9b40-4a66-8702-5228e808dfac">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c855b490-10f3-47df-91af-e360d0cf04cc">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8b07b6b5-0de6-4b4d-8e58-163beaa01cc4">EV <link linkend="DCM_113011" xrefstyle="select: nopage">(113011, DCM, "Document Title Modifier")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c7b8d021-e30f-4584-8ad2-623cdf243168">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5856a0d0-a2c1-425a-9ddd-cbcdf6c7ac22">MC</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d1f1ea1e-96a1-4b2e-9345-eef8770dbae3">IF Row 1 Concept Name = <link linkend="DCM_113013" xrefstyle="select: nopage">(113013, DCM, "Best In Set")</link>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_28bbb45a-4c07-466a-ac91-5438fd26133c">D<xref linkend="sect_CID_7012" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ef3a50b0-398c-44ce-85b0-e9c3f33b7545">5</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4abaa80b-f83f-4fa9-8a9b-b1679d774ac9">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_08df3c4a-76ba-44c8-a241-896d972203a3">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0928053e-3467-4704-b503-a7643feaf0a4">INCLUDE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f6c4bbe0-b212-4e90-97ef-c6ba532e8b56">D<xref linkend="sect_TID_1204" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bc674930-5947-464b-b13c-4821a61754de">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d4fc946e-c877-498c-98cf-e14d88e72adb">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_63f611a5-f82f-451e-9f38-1d7ddb568fbf"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9d19707a-0f38-4eb5-ad56-b3ed233ff5e5"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b56dd490-11e0-4852-bbed-c313461b158c">6</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7da41906-dbff-4767-8255-f32474601331">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c98b3d3f-bc2e-4ded-b2de-8f9ebf37cdab">HAS OBS CONTEXT</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ed9d7109-893c-4aae-8c3e-54d1563562d1">INCLUDE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_68498a7f-c613-4b81-b1c1-925b2aea6aff">D<xref linkend="sect_TID_1002" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bd49badd-0cc8-4f55-8b44-d789d5aa7885">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1e2e43bf-4c0a-4662-a917-320bb8596427">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_84ef6ade-b319-42ef-a878-c6fe0a67e428"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b0b8409a-507a-4d70-97f0-478ff8b8f254"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_27327617-e20d-4f68-a712-848049c43deb">7</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_eead6a14-5a92-45c0-bc71-22d094c42592">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_68250331-3243-4272-a34d-69ecf9a9cae4">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c1bcfe00-4c5c-4173-93bd-93b1a28d4032">TEXT</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d07c34d2-5bdb-491d-b2a3-56ea379ba0d6">EV <link linkend="DCM_113012" xrefstyle="select: nopage">(113012, DCM, "Key Object Description")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f9d8505e-8c5b-42d4-a7cd-9c94f0f10317">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_846050ca-34b8-4260-af5d-d3939422a356">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7bc06e39-1958-455d-85a0-dcc676cdd512"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_00969451-f844-4875-9099-6539f1fa3349"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0cb2f3a4-65ba-4fe7-b2e1-28131d37fa6c">8</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_83b6dcee-27d1-4322-be67-5e3d7016179d">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e241ab2f-dde7-44c3-8943-fd954c0c25b5">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2c33ceea-c67f-41ee-bcd8-19904caa2983">IMAGE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e7597194-045f-42e3-85ed-bc177acda8ec"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3a3dd1f2-a93c-4c79-9204-d49f33e054f9">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac8096f4-d6ae-479c-898b-4acbffae2f9f">MC</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f71f48ff-e509-46b9-a2ab-4f1c5ba919ab">At least one of Rows 8, 9 and 10 shall be present</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_87e92975-171c-4cb7-ab56-f00c039656d9"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_92193464-88a1-4b0e-8438-b89f7a56d6df">9</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ed2b8642-c233-4c1d-8f2a-5bca895e78e0">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_24e191aa-557d-4ad4-976d-ea1b074292d3">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8a63c5ba-fd25-4f05-8867-43743950c5ed">WAVEFORM</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3835083e-c897-46fd-838e-ed05c0602b6f"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_4801b35a-de7c-486f-9752-fcc3cd3df549">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_129975d2-697e-40c4-841b-c22496d8d214">MC</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9b322886-d3a3-43ca-b8ec-e02a547c03d9">At least one of Rows 8, 9 and 10 shall be present</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fd135221-8647-4112-b576-7b5edd8fbba0"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_307161ce-26af-4619-94a2-099a8ca1bcc4">10</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1b062cf6-81cf-476b-a2e7-8e9e530d863b">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8ef651f6-e645-4248-8086-5b491f200339">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f7c35d1c-b6f7-4edb-90e8-a76ad2e904fb">COMPOSITE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_34250e04-b007-4d81-8393-77b68ffc4ad6"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_51eb1249-8ac9-4929-87a3-af44aaa12077">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_488ce58e-be3b-4cd8-bd45-f2a1d8110832">MC</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_57c5d30f-fcf6-4804-b890-5624d5545f7e">At least one of Rows 8, 9 and 10 shall be present</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_52036fd8-43df-4a6e-9374-a96982fda5e7"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
                <para xml:id="para_2a4538b0-0d1f-47a1-ab64-0a3ff21b67c0">
                    <emphasis role="bold">Content Item Descriptions</emphasis>
                </para>
                <informaltable frame="box" rules="all">
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6c31355a-11fa-446d-94a2-e94e70fe779e">Rows 8, 9, 10</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6dae9531-8309-425e-8d25-73d1e43d0e2c">Purpose of reference shall not be present.</para>
                            </td>
                        </tr>
                    </tbody>
                </informaltable>
            </section>


        <section label="CID 7010" status="3" xml:id="sect_CID_7010">
            <title>Key Object Selection Document Title</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Resources:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_f9360ca0-2902-4d1d-8f8c-7ef02938c21c">
                            <emphasis role="bold">
                                <link xl:href="http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_7010.html" xl:show="new">HTML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/json/ValueSet-dicom-cid-7010-KeyObjectSelectionDocumentTitle.json" xl:show="new">FHIR JSON</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/fhir/xml/ValueSet-dicom-cid-7010-KeyObjectSelectionDocumentTitle.xml" xl:show="new">FHIR XML</link> | <link xl:href="ftp://medical.nema.org/medical/dicom/resources/valuesets/ihe/svs/1.2.840.10008.6.1.490.xml" xl:show="new">IHE SVS XML</link>
                            </emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_3892ee30-598b-46ad-93fa-f022feab7fff">
                            <emphasis role="bold">KeyObjectSelectionDocumentTitle</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">FHIR Keyword:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_a2a21302-cf88-48d9-8a4d-cf008a649bd0">
                            <emphasis role="bold">dicom-cid-7010-KeyObjectSelectionDocumentTitle</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_d80f7133-a5c4-4e59-96d7-fcb96eb52949">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_0aa35b1a-bc66-4618-b09c-dba31c996ac5">
                            <emphasis role="bold">20251111</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">UID:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_2718a374-a0ea-4e30-af7f-00ab89e1c0ea">
                            <emphasis role="bold">1.2.840.10008.6.1.490</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7010" rules="all" xml:id="table_CID_7010">
                <caption>Key Object Selection Document Title</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5309fa4c-2160-4140-a0e5-618234e6bf0e">Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7b9d2d26-9979-4658-9c46-08be13a7a4e0">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_882c98f4-8f06-4a33-a89a-1995fba52e56">Code Meaning</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_90255361-c414-47be-a4a1-cccbe23703f6">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3611295e-d993-4567-b6c4-c1419d5e1953">
                                <link linkend="DCM_113000" xrefstyle="select: nopage">113000</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3731bbcc-f9da-4b53-90bf-e9b8f1898589">Of Interest</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c81d44a3-1123-413e-ba1f-c6d5828dca93">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2d822395-bc54-46d6-9ed8-28070e135429">
                                <link linkend="DCM_113001" xrefstyle="select: nopage">113001</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a4406f07-dbcd-4a77-b31b-2a1b6d40f175">Rejected for Quality Reasons</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7ddacbab-c627-4d7f-a3b8-cad11dd1599c">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8562ff2b-1f35-4fe0-a615-788b240d928f"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7fbe3fad-650f-45a4-95f6-5e7c46e3b85b"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c5167447-4f65-4363-bc5e-73050724619f">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cbabf0ba-aa2b-4f13-87d4-77072bf4c3d6">
                                <link linkend="DCM_113010" xrefstyle="select: nopage">113010</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c75a70c1-d0e3-487e-b512-a946e4ddeccc">Quality Issue</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_55ff0fec-365c-4876-a748-ca841b570f65">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d0e32741-c369-4be2-9cfe-bed4d14273ce"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_80b9f237-59d4-4bc7-87d6-654cbca7bb0f">...</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a16e41b4-e842-470a-aa11-a0686330b87f">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f1469b40-1af4-41e7-b8e7-a74cb5e547ca">
                                <link linkend="DCM_113032" xrefstyle="select: nopage">113032</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_dba915ef-2fbb-46d0-84bd-7d01e457cf3d">Complete Study Content</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_47952dd9-b3c9-4be6-9f08-ec97e8bc75cc">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_50416a86-b9fa-469f-8137-fb59d0ed1dff">
                                <link linkend="DCM_113033" xrefstyle="select: nopage">113033</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e340ac50-722a-4e78-b82c-372ae45155c6">Signed Complete Study Content</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d912e6b8-d869-4e32-bb87-30b1fc40cad2">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eb548db1-2c89-4348-877c-97b22b0f8ade">
                                <link linkend="DCM_113034" xrefstyle="select: nopage">113034</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1b0daa90-61e2-40a2-8425-6b40a1927c3f">Complete Acquisition Content</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d31d8f12-86c2-4d8f-adbb-96e95dac7ed6">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_02c1de0b-8bea-488c-9efc-771f3c41d18a">
                                <link linkend="DCM_113035" xrefstyle="select: nopage">113035</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d37e65cb-7041-4521-bfd9-861a143d8d09">Signed Complete Acquisition Content</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0506064c-2ba5-4281-8210-cd6fda001d37">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0b28b133-f054-4088-aa27-a675ee9e2217"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_33aefb5d-8105-4333-80f1-3fa8e180f038"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="CID 7012" status="3" xml:id="sect_CID_7012">
            <title>Best in Set</title>
        </section>
        
        
        <section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding Scheme Version "01")</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">Definition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">Notes</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111207">111207</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8344032d-3774-4278-a437-45e2635fed69">Image artifact(s)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a76516c8-7158-4339-8068-d26607f008af">Signals that do not faithfully reproduce actual anatomic structures because of distortion or of addition or deletion of information.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2d3928bb-0b7f-4d69-a1de-4d9aa17dd1e5"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111208">111208</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2c63f7cd-5c64-4582-a3c3-f6829ff462e4">Grid artifact(s)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_62a5148b-a84d-4c20-8109-650c47c0bb7c">Feature(s) arising from the acquisition unit's anti-scatter grid mechanism. For two-dimensional systems, such features include those of mechanically damaged or incorrectly positioned grids. For moving or Bucky grids, artifacts may result from intentional grid motion that is inadequate in duration or velocity uniformity.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ba005c43-196c-45e1-b939-0904ed610025"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111209">111209</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ad0261e7-98ee-4227-8c06-00caab967152">Positioning</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f5455290-e49d-4cd1-a228-1beb20ce0a94">Inadequate arrangement of the anatomy of interest with respect to the X-Ray field and image detector sensitive area. Examples: 1) positioning is "cutoff" when the projection of anatomy of interest falls outside the sensitive area of the detector; 2) "cone cut", in which the X-Ray field does not adequately cover the anatomy of interest; 3) detector's sensitive surface is too small to cover the projection of the anatomy of interest; 4) improper angular orientation or "rotation" of anatomy of interest with respect to the X-Ray source, or detector; 5) projection of other anatomy or clothing over the anatomy of interest in the image.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a60f26e0-4dab-41a0-8a95-d7de32c0eb66"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111210">111210</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_088a7ece-cd2d-43a1-b317-1d41164c572e">Motion blur</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_60528387-dcf3-4f2f-8180-0713a3c039c6">Unacceptable image blur resulting from motion of the anatomy of interest during exposure or the inadequately compensated motion of X-Ray source with respect to the image detector during exposure.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_69610eff-c152-451f-ad0f-5d98f129f4ee"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111211">111211</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_088c1150-ffbf-4731-ac65-f3222a4c995f">Under exposed</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b484ff03-0b9d-4fb3-94bf-8e792cc3d018">Inadequate number of quanta reached the detector during exposure. Reasons for under exposed images include low kVp, low mAs product, excess Source Image Distance. Under exposed images have inadequate signal and higher noise in the areas of interest.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_dafd1a0b-2fa6-42c4-a527-acfaf6f86c49"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111212">111212</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_61a8f322-b962-4466-8468-3a2f5795a270">Over exposed</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7056ce01-8b70-4938-a9d7-aed37204d576">An excess number of quanta reached the detector during exposure. Reasons for over exposed images include high kVp, high mAs product, short Source Image Distance. Over exposed images have high signal and lower noise in the areas of interest. Over exposed area may demonstrate lack of contrast from over saturation of the detector.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e07af948-6312-4f31-9899-30c1d9a0c5e4"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111213">111213</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a86f0a32-fa40-46ee-9047-5972157c5408">No image</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a810a56a-9c9d-4c1f-825a-178b13a8a905">No evidence of a patient exposure.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3dd54f64-122a-4730-ab0b-b9207feb89fa"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111214">111214</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_171ad525-1a86-40ee-8cbd-b50c1c540f53">Detector artifact(s)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a57a4660-83c3-47f1-b98c-e8cc8ba027bf">Superposed features or flaws of the detector.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a7917cf8-54b1-4013-a442-e4ba34cf90d5"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111215">111215</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d1435375-104a-4188-b8c9-044cbc273945">Artifact(s) other than grid or detector artifact</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c3f8dfee-2507-4f44-ac9b-e76c0ef9f65d">Features or discontinuities arising from causes other than the anti-scatter grid and image detector.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ab1d42f5-62b7-404c-9430-bbc2d7214abe"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111216">111216</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_806fb018-f2d8-47f6-b310-498bdd20156f">Mechanical failure</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_19d84656-1282-495c-84e9-df800ea9ae03">Failure of the device to operate according to mechanical design specifications.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_24a2f068-fd61-4e0f-b703-bea81606fee7"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111217">111217</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_117ae76c-5b23-4030-8d34-bc668df2c6ed">Electrical failure</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fd4e5523-78b8-4cce-8d55-7bb24fdb0f00">Failure of a device to operate according to electrical design specifications.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6e522bc9-5f47-49b0-b148-61e609cfb3ab"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111218">111218</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eadfc5ae-1595-4619-90c7-0d841a4ff82a">Software failure</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d77370c1-9e47-44b5-92da-3daa862e484e">Attributable to software used in generation or handling of image.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_66e5c1ad-0240-443a-8906-e858e5cf81fa"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111219">111219</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3fe15103-0cae-4da6-a543-51e7bba4277e">Inappropriate image processing</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cf3c019b-4e57-4ce2-84b7-96331cbdfb08">Images processed inappropriately, not following appropriate protocol.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_07b16778-7d75-445e-b031-3ecbc07f239f"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111220">111220</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c75a644a-9a04-47c5-b378-92fdcfbc31b4">Other failure</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8f896be1-c5b3-49d7-b22e-92494e25b80e">Failure that is not mechanical or electrical or otherwise described.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_72231bba-3ca7-4783-bf0a-f8e99588e7a8"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111221">111221</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_06604f37-4169-46a6-82bb-19bf1ac278d5">Unknown failure</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_57a6732b-6316-4e54-a7eb-36048bf3420e">Unidentified or unknown cause of failure.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_25e9333a-5b27-4186-b223-7f61266e331b"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111222">111222</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1a729746-0364-4cf6-ac6f-25232b4d8056">Succeeded</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1a09ad54-99cd-44d5-8425-662c8d832b74">The attempted process was completely successful.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1b8b9d9e-ea93-45f9-a85f-ef3df0ec1c2d"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111223">111223</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c5eaf8da-fceb-43f6-87f6-82261899778d">Partially Succeeded</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5779f802-729f-421c-92a5-21ff39725840">The attempted process succeeded in some ways, but failed in others.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6cf705c9-c579-4824-8c84-ab2145c798ed"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111224">111224</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_edff61fc-8629-4b0f-809b-50ae47d24487">Failed</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5d43c2cd-4da3-425b-a67c-7706f5eb5226">The attempted process completely failed.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_363f98f9-8f66-4fc1-826a-161be6e03959"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111225">111225</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2565d7b6-1f2f-4837-80ea-11cfabd0a594">Not Attempted</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e0a56df1-f198-4a51-be6d-16c018bebcc3">No process was performed.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4ae4364c-249c-4897-8277-6903c4fc44cf"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
    </chapter>
</book>