    <title>CP-2563</title>
    <subtitle>Allow TILED_FULL for Float and Double Float Pixel Data</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Jan 2026 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/11/10</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>Chris Bridge</para><para><link
                   xl:href="mailto:cbridge@mgh.harvard.edu"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/08/15</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2563</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Allow TILED_FULL for Float and Double Float Pixel Data</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>Currently TILED_FULL is only described for images with Frames sequentially encoded as successive Frames in Pixel Data (7FE0,0010), though there was no intent to forbid its use for Float Pixel Data and Double Float Pixel Data, such as might be used in Parametric Maps for whole slide images (as introduced in CP 1830 and CP 1984).</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        <section label="C.7.6.17.3" status="5" xml:id="sect_C.7.6.17.3">
            <title>Spatial Location and Optical Path of Tiled Images</title>
            <para xml:id="para_81c46a69-b4dc-428d-822c-2efeee711150">If Dimension Organization Type (0020,9311) is present with a Value of TILED_FULL, then the Per-Frame Functional Group Macros that would otherwise describe the spatial location of each tile explicitly (e.g., the X, Y and Z offsets from the origin in the Slide Coordinate System Plane Position (Slide)), and the optical path or segment, may be omitted.</para>
            <para xml:id="para_08fcce55-df5f-49f2-b8a2-978d0468872c">A Value of TILED_FULL indicates that the Frames across all Instances of a Concatenation, or a single Instance in the absence of a Concatenation, comprise a non-sparse non-overlapping representation of an entire rectangular region, and are sequentially encoded as successive Frames in Pixel Data (7FE0,0010)<emphasis role="bold"><emphasis role="underline">, Float Pixel Data  (7FE0,0008), or Double Float Pixel Data (7FE0,0009)</emphasis></emphasis> in an implicit order varying:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_f0dd20ae-8ade-46f2-ad35-2c50add983d5">first along the row direction from left to right, where the row direction is defined in the Slide Coordinate System by the first three Values of Image Orientation (Slide) (0048,0102),</para>
                </listitem>
                <listitem>
                    <para xml:id="para_bd60c273-e1a2-4c77-89f4-538768f79537">then along the column direction from top to bottom, where the column direction is defined in the Slide Coordinate System by the second three Values of Image Orientation (Slide) (0048,0102),</para>
                </listitem>
                <listitem>
                    <para xml:id="para_0ae08229-6360-49ad-a748-9e855f00a350">then along the depth direction from the glass slide towards the coverslip, where the depth direction is defined in the Slide Coordinate System from zero to positive,</para>
                </listitem>
                <listitem>
                    <para xml:id="para_e553a9fe-2e1b-42ed-aa5f-a27f3c7c00e6">then along optical paths, if applicable, where the direction is defined by successive Items of the Optical Path Sequence (0048,0105) in the order in which they are listed in that Sequence,</para>
                </listitem>
                <listitem>
                    <para xml:id="para_9e9bcf28-f958-4303-838f-262a3ec10aa6">then along the segments, if applicable, where the direction is defined by ascending numeric Values of Segment Number (0062,0004) as defined in the Segment Sequence (0062,0002).</para>
                </listitem>
            </itemizedlist>
            <para xml:id="para_6d73468e-28c6-4aee-a708-dab0b3a14702">If Dimension Organization Type (0020,9311) is absent or has a Value of TILED_SPARSE, then the location of each tile is explicitly encoded using information in the Per-Frame Functional Groups Sequence, and the recipient shall not make any assumption about the spatial position or optical path or segment or order of the encoded Frames, and whether or not any tiles overlap, but shall rely on the values of the relevant Per-Frame Functional Group Macro.</para>
            <note>
                <para xml:id="para_443d234b-b810-4ec2-b0df-f82ccdaef2f5">Images with an Image Type (0008,0008) Value 3 of THUMBNAIL, LABEL or OVERVIEW are single Frame and may have a spatial extent that is not the same as the Total Pixel Matrix, so Dimension Organization Type (0020,9311) is not applicable.</para>
                <para xml:id="para_3cc348a9-3606-4d1f-947e-7f5a132f8fe4">The same previously applied to images with an Image Type (0008,0008) Value 3 of LOCALIZER, which has been retired. See <link xl:href="http://dicom.nema.org/medical/dicom/2021c/output/pdf/part03.pdf">PS3.3-2021c</link>.</para>
            </note>