<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2540</title>
    <subtitle>Clarify Concatenation Number of Frames</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Aug 2026 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2026/06/03</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Tamás Sárga</para><para><link
                    xl:href="mailto:tamas.sarga@3dhistech.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/04/10</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2540</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Clarify Concatenation Number of Frames</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>There is no question that the use of Concatenations depends on the Number of Frames value being the number of Frames encoded in the actual SOP Instance that is the part of the Concatenation.</para>
                    <para>Some may be confused that it is instead the total number of Frames in the Multi-frame Image that spans the Instances of the Concatenation re-combined.</para>
                    <para>Resolve any potential confusion by adding a new optional Attribute to reflect the total number of Frames, add notes clarifying the difference, improve the definition of the Number of Frames Attribute throughout (even for uses in which Concatenations are not applicable, for consistency), factor out the description of a Multi-frame Image into the definition, factor out the description of a Concatenation into a new definition, and remove mention of specific Attributes in the Information Entity description.</para>
                    <para>Also clarify extraneous mention of a contiguous stream of pixels, which is only applicable to an uncompressed encoding.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>


        <section label="3" status="1" xml:id="chapter_3">
            <title>Definitions</title>
            <para xml:id="para_a1b339eb-d0e9-4689-b2f0-3435fec8e155">For the purposes of this Standard the following definitions apply.</para>
            <glossary>
                <glossdiv label="3.8" status="2" xml:id="sect_3.8">
                    <title>DICOM Information Object</title>
                    
                    <glossentry xml:id="glossentry_Concatenation" xreflabel="Concatenation">
                        <glossterm><emphasis role="bold"><emphasis role="underline">Concatenation</emphasis></emphasis></glossterm>
                        <glossdef>
                            <para><emphasis role="bold"><emphasis role="underline">A Multi-frame Image that has been split into more than one SOP Instance.</emphasis></emphasis></para>
                        </glossdef>
                    </glossentry>
                    
                    <glossentry xml:id="glossentry_Frame" xreflabel="Frame">
                        <glossterm>Frame</glossterm>
                        <glossdef>
                            <para xml:id="para_cff67cef-f4b1-4271-ad07-5c634c5598e5">A single two-dimensional pixel plane of a Multi-frame Image.</para>
                        </glossdef>
                    </glossentry>
                    
                    <glossentry xml:id="glossentry_MultiFrameImage" xreflabel="Multi-frame Image">
                        <glossterm>Multi-frame Image</glossterm>
                        <glossdef>
                            <para xml:id="para_97cbf9b3-31cd-4be3-b7be-4aca4c74b783"><emphasis role="bold"><emphasis role="strikethrough">I</emphasis><emphasis role="underline">An i</emphasis></emphasis>mage <emphasis role="bold"><emphasis role="strikethrough">that contains</emphasis><emphasis role="underline">whose pixel data consists of</emphasis></emphasis> multiple two-dimensional pixel planes.<emphasis role="bold"><emphasis role="underline"> An uncompressed unencapsulated Multi-frame Image is transmitted as a single contiguous stream of pixels within the Instance in which it is encoded, though it may be split into Instances of a Concatenation (each of which is also a Multi-frame Image). Encapsulated Multi-frame Image pixel data may be delimited between frames depending on the Transfer Syntax.</emphasis></emphasis></para>
                        </glossdef>
                    </glossentry>
                </glossdiv>
            </glossary>
        </section>

        <section label="7.5" status="2" xml:id="sect_7.5">
            <title>Organizing Large Sets of Information</title>
            <para xml:id="para_122f19da-76a2-485a-85cf-0940635e183a">For the purpose of accommodating large sets of frames in Multi-frame Image SOP Instances the Real-World Entity Relationship Diagram has been extended to describe the relationships of these instances: Concatenation (see <xref linkend="sect_7.5.1" xrefstyle="select: label"/>) and Dimension Organization (see <xref linkend="sect_7.5.2" xrefstyle="select: label"/>). <xref linkend="figure_7.5-1" xrefstyle="select: label"/> depicts the additions to <xref linkend="figure_7-1a" xrefstyle="select: label"/>.</para>
            <figure label="7.5-1" pgwide="1" xml:id="figure_7.5-1">
                <title>Extension of the Real World Model with Concatenations and Dimensions</title>
                <mediaobject>
                    <imageobject>
                        <imagedata fileref="cp2540/PS3.3_7.5-1.svg"/>
                        <!--<imagedata fileref="figures/PS3.3_7.5-1.svg"/><!-->
                        <!--<imagedata fileref="part03_fromword_files/image022.png"/>-->
                    </imageobject>
                </mediaobject>
            </figure>
            <section label="7.5.1" status="3" xml:id="sect_7.5.1">
                <title>Concatenation</title>
                <para xml:id="para_6486fcca-b39a-4e28-998f-aaf419b3a37a">For implementation specific reasons (such as practical limits on the maximum size of an individual SOP Instance) <emphasis role="bold"><emphasis role="strikethrough">the content of </emphasis></emphasis>a Multi-frame Image may need to be split into more than one SOP Instance<emphasis role="bold"><emphasis role="strikethrough">. These SOP Instances</emphasis><emphasis role="underline">, each of which is also a Multi-frame Image, and all of which </emphasis></emphasis> together form a Concatenation<emphasis role="bold"><emphasis role="strikethrough">, which is a group of</emphasis><emphasis role="underline">. The</emphasis></emphasis> SOP Instances <emphasis role="bold"><emphasis role="underline">that comprise a Concatenation shall be </emphasis></emphasis>within a <emphasis role="bold"><emphasis role="underline">single </emphasis></emphasis>Series<emphasis role="bold"><emphasis role="strikethrough"> that is
                            uniquely identified by Concatenation UID (0020,9161)</emphasis></emphasis>.<emphasis role="bold"><emphasis role="underline"> Multiple different Concatenations can be present in the same Series. Different source Multi-frame Images shall not be combined into a single Concatenation (i.e., the Concatenation source shall be a single Multi-frame Image).</emphasis></emphasis></para>
            </section>
            <section label="7.5.2" status="3" xml:id="sect_7.5.2">
                <title>Dimension Organization</title>
                <para xml:id="para_f8a3a048-e854-45b0-ab60-f651aa760af3">...</para>
            </section>
        </section>
        
        <section label="C.7.6.6" status="4" xml:id="sect_C.7.6.6">
            <title>Multi-frame Module</title>
            <para xml:id="para_a3ab971a-a8b7-4205-b3a4-dae6fab9a931">
                <xref linkend="table_C.7-14" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.7.6.6" xrefstyle="select: title"/>, which describe a Multi-frame <emphasis role="bold"><emphasis role="strikethrough">pixel data </emphasis></emphasis>Image.</para>
            <table frame="box" label="C.7-14" rules="all" xml:id="table_C.7-14">
                <caption>Multi-frame Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0215757a-84e2-4bc2-9ee3-0bfde4e3e09a">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_30cf5c7b-d8a8-42fa-a344-25430cc63dcf">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0015bb9-e34e-40cc-8570-034f0aa9ffef">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99064f22-5951-46a0-9d99-cd80be1e069c">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c6798678-5f49-4bf5-8b42-465ea686c28c">Number of Frames</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_37c045d9-4757-4c9c-b4ae-e3a6eb483edb">(0028,0008)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_db393087-b151-4f48-976c-ccf996e74fc5">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cf13eb17-5479-4d78-bc50-1918596fd8f3">Number of frames in <emphasis role="bold"><emphasis role="strikethrough">a</emphasis><emphasis role="underline">the</emphasis></emphasis> Multi-frame Image<emphasis role="bold"><emphasis role="underline"> in this SOP Instance</emphasis></emphasis>. See <xref linkend="sect_C.7.6.6.1.1" xrefstyle="select: label"/> for further explanation.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ca4b7856-dda6-414a-a2aa-bd0d8dfe2ff8">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3441e9ce-46d7-4330-8f65-f69839efee02"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ce8fcc4d-aec2-4b82-b61f-6058d90b5728"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_107c2923-faa4-4342-b772-98382661c906"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a42c5caf-0873-42f4-b915-d8eef63dab70">Encapsulated Pixel Data Value Total Length</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b7ab9a7d-68f8-48a2-ab00-409ef522ac02">(7FE0,0003)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9ca5088d-0842-45d1-82e5-1440f78a81d0">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ab512734-5ea8-4d4a-9262-084fd439ecb4">The length of the pixel data bit stream encapsulated in Pixel Data (7FE0,0010), in bytes, when all the fragments have been combined, not including any trailing padding to even length in the last Fragment added for encapsulation.</para>
                            <note>
                                <para xml:id="para_b5c2f2ca-53f5-43ab-a7d1-881a30ca1f15">This value will depend on the Transfer Syntax in which the Pixel Data (7FE0,0010) is encoded, and may need to be updated depending on the Transfer Syntax negotiated and selected for a particular transfer. See <olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="template:PS3.5 Section %n “%t”"/>.</para>
                            </note>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.7.6.6.1" status="5" xml:id="sect_C.7.6.6.1">
                <title>Multi-frame Module Attribute Descriptions</title>
                <section label="C.7.6.6.1.1" status="6" xml:id="sect_C.7.6.6.1.1">
                    <title>Number of Frames</title>
                    <para xml:id="para_d3d36f49-02b9-40f8-bce7-c52aa298a3d3"><emphasis role="bold"><emphasis role="strikethrough">A Multi-frame Image is defined as a Image whose pixel data consists of a sequential set of individual Image Pixel frames. A Multi-frame Image is transmitted as a single contiguous stream of pixels. Frame headers do not exist within the data stream.</emphasis></emphasis></para>
                    <para xml:id="para_9f634dd5-ee84-4c51-8b3b-3f8ce3fa43b6">...</para>
                    <para xml:id="para_1d5f78ad-b288-49f7-87e7-9dcf7c9e3ad0">The total number of frames contained within a <emphasis role="bold"><emphasis role="underline">SOP Instance of a </emphasis></emphasis>Multi-frame Image is conveyed in the Number of Frames (0028,0008). Number of Frames (0028,0008) shall have a value greater than zero.</para>
                </section>
            </section>
        </section>
        
        <section label="C.7.6.16" status="4" xml:id="sect_C.7.6.16">
                    <title>Multi-frame Functional Groups Module</title>
                    <table frame="box" label="C.7.6.16-1" rules="all" xml:id="table_C.7.6.16-1">
                        <caption>Multi-frame Functional Groups Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_efd07f13-1243-42bd-93c8-919ea56a1145">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_171895f0-b152-4207-973c-ffb9c84c6a01">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d1a5c9f2-8074-4a22-b652-84b9ddd75672">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_00043efd-ea22-4575-b910-c72be0c72adf">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_456f83f0-9370-4c89-b968-4f9b7b1d8c38">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d3bd8879-c474-4c60-b173-010e5162c5d8"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_61b72077-ca4b-4656-bef9-4004ba0f277b"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3aaa5c86-e6dd-41c0-9ba4-457b78303702"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_57b15c28-50c9-4c02-8604-75f34911043d">Number of Frames</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_27542988-edb1-41a7-a5fc-9bcfe8c27414">(0028,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_00ba676a-0947-4fdb-b4cb-fabfd51a7f6b">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0e0f6da7-1fac-4eba-81e1-48c472fd359b">Number of frames in <emphasis role="bold"><emphasis role="strikethrough">a</emphasis><emphasis role="underline">the</emphasis></emphasis> Multi-frame Image<emphasis role="bold"><emphasis role="underline"> in this SOP Instance</emphasis></emphasis>. See <xref linkend="sect_C.7.6.6.1.1" xrefstyle="select: label"/> for further explanation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b5c77050-7297-43ee-b108-9d3b7772b2e3">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a309f3df-7c6e-4bf3-96ca-df033ace2507"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ce4497fa-d1eb-477e-a163-dbb7d233c840"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_eb1622f4-9f9c-4653-a898-f806fab29261"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_75b699b3-67f1-4db4-a88a-5f464521330f">Concatenation Frame Offset Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d0162595-fe77-4e63-af90-343be9459df0">(0020,9228)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_40353cf5-a23f-400c-bc2f-a398b2e298e6">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_304aea0f-4b16-4808-b33b-a0377c3fc1cc">Offset of the first frame in a Multi-frame Image of a Concatenation. Logical frame numbers in a Concatenation can be used across all its SOP Instances. This offset can be applied to the implicit frame number to find the logical frame number in a Concatenation. The offset is numbered from zero; i.e., the instance of a Concatenation that begins with the first frame of the Concatenation has a Concatenation Frame Offset Number (0020,9228) of zero.</para>
                                    <para xml:id="para_1d00d051-1ffb-4329-9940-0f0b736da2c2">Required if Concatenation UID (0020,9161) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3e2125f6-5d1b-4c3d-ad74-e4bc80dd034c">Representative Frame Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_215068d0-ece4-4687-b181-5c65a0bfb712">(0028,6010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e78f5152-2c6c-48fe-bf6d-6c86f462d78d">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d2ee13ab-5471-42d3-b8ab-75a7ae6d640b">The frame number selected for use as a pictorial representation (e.g., icon) of the Multi-frame Image.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_86893204-d047-4fc0-9940-bc7611532fa6">Concatenation UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_398ba4e7-d0c3-438b-aaa9-a0ad0b9ef17a">(0020,9161)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a8dc717a-70c1-421f-977a-acaafa586ceb">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_45114939-4b72-4abb-8a71-57c56955e258">Identifier of all SOP Instances that belong to the same Concatenation.</para>
                                    <para xml:id="para_e7080657-dbd8-410d-8d4e-acc6bc1cb2fa">Required if a group of Multi-frame Image SOP Instances within a Series are part of a Concatenation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ab069ad6-e365-41d1-9c86-1e50b1015eaf">SOP Instance UID of Concatenation Source</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6ad6d73a-9c80-4d67-ad11-c425ec2894e9">(0020,0242)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6eee96f6-ccd1-4109-87b2-b7628e279e89">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e795995f-62a0-44c8-98c1-cebf2bfa6abb">The SOP Instance UID of the single composite SOP Instance of which the Concatenation is a part. All SOP Instances of a Concatenation shall use the same value for this Attribute, see <xref linkend="sect_C.7.6.16.1.3" xrefstyle="select: label"/>.</para>
                                    <note>
                                        <para xml:id="para_326f77dc-84d0-4948-bb19-c7374695e54f">May be used to reference the entire Instance rather than individual Instances of the Concatenation, which may be transient (e.g., from a presentation state).</para>
                                    </note>
                                    <para xml:id="para_d8f04107-cc2f-475d-a868-cf78c1cf911f">Required if Concatenation UID (0020,9161) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9497fce2-bbd0-4ff7-a3cf-6075d60b798d">In-concatenation Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8f37c8d5-6cb6-4552-b46c-07e1738eeb13">(0020,9162)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4b9f77fc-68a4-4e8e-830e-2a83d4dde4fe">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1289db75-3c06-41b5-aeee-408a197f08cb">Identifier for one SOP Instance belonging to a Concatenation. See <xref linkend="sect_C.7.6.16.2.2.4" xrefstyle="select: label"/> for further specification. The first Instance in a Concatenation (that with the lowest Concatenation Frame Offset Number (0020,9228) value) shall have an In-concatenation Number (0020,9162) value of 1, and subsequent Instances shall have values monotonically increasing by 1.</para>
                                    <para xml:id="para_30bb36e1-a8b8-4b05-9c97-60a65ae20928">Required if Concatenation UID (0020,9161) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dc04e05a-5489-45f7-a9c8-31caba938e79">In-concatenation Total Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a6e2b9f2-a236-406a-b682-93c359c30a8d">(0020,9163)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_91e377df-42cf-4b85-9882-fb7957dd177d">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_613f3530-5308-453b-8bb8-fc6897b0fa46">The number of SOP Instances sharing the same Concatenation UID.</para>
                                    <para xml:id="para_1df40183-93b4-407d-9521-607f9dfb4313">If present, shall have a value greater than one, unless an IOD overrides this constraint to enumerate a value of 1 to prevent the use of Concatenations (e.g., ...).</para>
                                </td>
                            </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Total Number of Frames in Concatenation</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(0020,eee1)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Total number of frames in all SOP Instances that comprise the Concatenation that is a Multi-frame Image.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">If present, shall have a value greater than zero.</emphasis></emphasis></para>
                        </td>
                    </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e12cfd01-0e94-4d62-9682-34d4e4d8d187">Encapsulated Pixel Data Value Total Length</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8071ec1f-d9fe-462f-894b-536c1b046467">(7FE0,0003)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ee72c268-925c-4706-bef5-ac66b6d6bccd">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b06ba4e5-0fce-440b-ab4c-24362f93a1ae">The length of the pixel data bit stream encapsulated in Pixel Data (7FE0,0010), in bytes, when all the fragments have been combined, not including any trailing padding to even length in the last Fragment added for encapsulation.</para>
                                    <note>
                                        <para xml:id="para_85887344-c082-4e81-84dc-beafc1a85d73">This value will depend on the Transfer Syntax in which the Pixel Data (7FE0,0010) is encoded, and may need to be updated depending on the Transfer Syntax negotiated and selected for a particular transfer. See <olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="template:PS3.5 Section %n “%t”"/>.</para>
                                    </note>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.7.6.16.1" status="5" xml:id="sect_C.7.6.16.1">
                        <title>Multi-frame Functional Groups Module Attribute Description</title>
                        <section label="C.7.6.16.1.3" status="6" xml:id="sect_C.7.6.16.1.3">
                            <title>SOP Instance UID of Concatenation Source</title>
                            <para xml:id="para_6754c525-4d65-4921-a5ad-cf6f903f8e18">...</para>
                        </section>
                    </section>
<section label="C.7.6.16.2" status="5" xml:id="sect_C.7.6.16.2">
                        <title>Common Functional Group Macros</title>
                        
                        
                        
                        <section label="C.7.6.16.2.2" status="6" xml:id="sect_C.7.6.16.2.2">
                            <title>Frame Content Macro</title>
                            <para>...</para>
                            <section label="C.7.6.16.2.2.4" status="7" xml:id="sect_C.7.6.16.2.2.4">
                                <title>Concatenations and Stacks</title>
                                <para xml:id="para_75d22358-2aea-4b78-a0be-410561efc3c8"><emphasis role="bold"><emphasis role="strikethrough">Due to implementation specific reasons (such as maximum object size) the information of a Multi-frame Image may be split into more than one SOP Instance. These SOP Instances form together a Concatenation. This is a group of </emphasis><emphasis role="underline">The </emphasis></emphasis>SOP Instances <emphasis role="bold"><emphasis role="strikethrough">within a Series</emphasis><emphasis role="underline">that comprise a Concatenation shall be </emphasis></emphasis> uniquely identified by <emphasis role="bold"><emphasis role="underline">the same Value of </emphasis></emphasis>Concatenation UID (0020,9161).</para>
                                <para xml:id="para_408673e1-ec6d-4bbc-b3ea-b75d3db0d9f1">...</para>
                                <para xml:id="para_5d9639a2-43ae-4380-a978-aa9223bd8168">In a Concatenation the Dimension Index Sequence (0020,9222) Items of the Shared Functional Groups Sequence (5200,9229) shall be identical and have the same values for all individual SOP Instances. The Items of the Per-Frame Functional Groups Sequence (5200,9230) shall be identical for all individual SOP Instances but the values may change per frame. For all other Attributes of all the Modules of the IOD, the same Attributes shall be present and the values shall be identical, with the exception of the following Attributes:</para>
                                <itemizedlist>
                                    <listitem>
                                        <para xml:id="para_ad74c67a-498b-4289-b69f-bf15f6264cb7">Number of Frames (0028,0008)</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_ec656bee-3492-426a-a8a2-de8ee816df95">Concatenation Frame Offset Number (0020,9228)</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_7e477c8c-1bec-4cfc-a9f9-01925210039b">In-concatenation Number (0020,9162)</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_77629160-efc8-40ec-b4e4-579c814bb807">SOP Instance UID (0008,0018)</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_aa207c7b-57cb-4ecb-bcdf-81d8db47e3fb">Instance Creation Time (0008,0013)</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_1aad0924-10eb-41f5-9d51-d14e01002420">Extended Offset Table (7FE0,0001)</para>
                                    </listitem>
                                    <listitem>
                                        <para xml:id="para_6197e86d-5d39-4270-a6e8-72ae0985ae0e">Extended Offset Table Lengths (7FE0,0002)</para>
                                    </listitem>
                                </itemizedlist>
                                <note>
                                    <orderedlist>
                                        <listitem>
                                            <para xml:id="para_25fede90-cda3-4b8f-a677-96f133a55abd">The intent of Concatenations is to split what might have been encoded in a single SOP Instance into smaller fragments for more convenient storage or transmission.
                                                All the multiple SOP Instances of a Concatenation should be able to be assembled into a valid single SOP Instance
                                                Hence it is not permitted to change such Attributes as Photometric Interpretation (0028,0004), Rows (0028,0010), Columns (0028,0011), etc.</para>
                                        </listitem>
                                        <listitem>
                                            <para><emphasis role="bold"><emphasis role="underline">Within each SOP Instance that comprises the Concatenation, the Number of Frames (0028,0008) Value is the number of frames within that SOP Instance, not the total number of frames in the Concatenation, which may be described in Total Number of Frames in Concatenation (0020,eee1). </emphasis></emphasis></para>
                                        </listitem>
                                        <listitem>
                                            <para>When multiple SOP Instances of a Concatenation are combined into a single SOP Instance, the Basic Offset Table, if populated, and the Extended Offset Table (7FE0,0001) and Extended Offset Table Lengths (7FE0,0002), if present, need to be regenerated to match the new content of the Pixel Data attribute.</para>
                                        </listitem>
                                    </orderedlist>
                                </note>
                                <para xml:id="para_61473bf0-d40d-42ad-a8ee-9035a0ca7036">...</para>
                            </section>
                        </section>
                    </section>
        </section>
        
        <section label="C.7.6.29" status="3" xml:id="sect_C.7.6.29">
                    <title>Sparse Multi-frame Functional Groups Module</title>
                    <para xml:id="para_04c47292-da48-459a-a2ff-d64b6250411b">...</para>
                    <table frame="box" label="C.7.6.29-1" rules="all" xml:id="table_C.7.6.29-1">
                        <caption>Sparse Multi-frame Functional Groups Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e2406fc8-4312-4d3d-82af-0c7e35c8aa4a">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0fefd71b-008c-4e9d-8434-3c19e8f78863">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cacd60e8-ec07-4d55-a12e-e487153433c3">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_25a679b6-da70-4bca-aa7a-effe427eb4b7">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_17948b48-e195-44d5-a8fe-615ea6cf47de">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c45b54f7-aff7-457a-8011-196f0ccfbc3b"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5fb9bfb2-331c-4e9f-bb14-d177f7d77d77"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_122ec85b-a02c-4d30-aa7b-326ef2713eb2"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_040803c2-1c4c-45fa-b9f5-9b2cb01c7a2c">Number of Frames</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9590ec35-9e5d-4f28-af76-52af3b3ad0a6">(0028,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ce10435b-a50c-4928-83f8-d39974bfb710">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e3dec03d-a5e2-40eb-90f9-59053f93dfab">Number of frames in <emphasis role="bold"><emphasis role="strikethrough">a</emphasis><emphasis role="underline">the</emphasis></emphasis> Multi-frame Image<emphasis role="bold"><emphasis role="underline"> in this SOP Instance</emphasis></emphasis>. See <xref linkend="sect_C.7.6.6.1.1" xrefstyle="select: label"/> for further explanation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6887081d-cedd-4eb8-8791-803985d04922">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3a5828f0-6955-46a0-a331-bb147f240b07"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9a00c714-ad41-4e0f-adb2-24cec3b2d1a7"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4b857711-6698-49be-a05d-2dda72f446c9"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c19dc77b-4b2f-496a-8f79-d6c93a6ffddd">Concatenation Frame Offset Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a6a13646-b33c-46ab-9fba-083e746b8f8e">(0020,9228)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_21259bee-8b49-4670-8361-9b750508d35b">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c39a9ec7-4d50-4a3b-9079-f6915faddcb8">Offset of the first frame in a Multi-frame Image of a Concatenation. Logical frame numbers in a Concatenation can be used across all its SOP Instances. This offset can be applied to the implicit frame number to find the logical frame number in a Concatenation. The offset is numbered from zero; i.e., the instance of a Concatenation that begins with the first frame of the Concatenation has a Concatenation Frame Offset Number (0020,9228) of zero.</para>
                                    <para xml:id="para_6bb05670-7bbe-4763-9900-c96d8d81b820">Required if Concatenation UID (0020,9161) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_40544d1e-d2a5-4e1e-85ac-538f013c6d85">Representative Frame Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1cb41220-93fb-41b3-9bd9-31721f8bf1bb">(0028,6010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8bda7dac-16c4-4854-a789-02991c0742e2">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f1b4b69c-7628-4535-97f8-058708a9bf52">The frame number selected for use as a pictorial representation (e.g., icon) of the Multi-frame Image.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_05021d87-8b4f-4d7e-930f-7c840179292e">Concatenation UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e324378c-9b10-4094-bf0a-7ed46089d782">(0020,9161)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_29b2c56a-a488-4a87-b71a-62cfd2f7d1af">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_794c5903-4ab1-41fd-a796-8cee76940539">Identifier of all SOP Instances that belong to the same Concatenation.</para>
                                    <para xml:id="para_7678b801-a812-4db9-9115-acdba51772ed">Required if a group of Multi-frame Image SOP Instances within a Series are part of a Concatenation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6872d69d-c24f-474b-9cb8-b5cd5e834f80">SOP Instance UID of Concatenation Source</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c0eac0e3-a18e-436a-867f-2ff432e7234f">(0020,0242)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_78abcfbb-a771-491a-ae50-e009f6032644">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f69f278e-297f-449f-aaa2-a998c212900e">The SOP Instance UID of the single composite SOP Instance of which the Concatenation is a part. All SOP Instances of a Concatenation shall use the same value for this Attribute, see <xref linkend="sect_C.7.6.16.1.3" xrefstyle="select: label"/>.</para>
                                    <para xml:id="para_9616770a-b56d-4034-95d6-620679573e5e">Note</para>
                                    <para xml:id="para_6d76cc03-a6d5-4507-a8f9-56beb2139146">May be used to reference the entire Instance rather than individual Instances of the Concatenation, which may be transient (e.g., from a presentation state).</para>
                                    <para xml:id="para_a068653f-c9b6-47cd-a803-9b59519f78ae">Required if Concatenation UID (0020,9161) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bc796eb2-ed6d-4e93-869f-c9f30ce129c2">In-concatenation Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_10f5e872-8cfe-4ff5-96f4-81149f088106">(0020,9162)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_52023db8-da98-4767-a296-b0fff6ba6450">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b69f3799-57c5-4050-bb03-919ae74439f8">Identifier for one SOP Instance belonging to a Concatenation. See <xref linkend="sect_C.7.6.16.2.2.4" xrefstyle="select: label"/> for further specification. The first Instance in a Concatenation (that with the lowest Concatenation Frame Offset Number (0020,9228) value) shall have an In-concatenation Number (0020,9162) value of 1, and subsequent Instances shall have values monotonically increasing by 1.</para>
                                    <para xml:id="para_55cb434b-240f-4b5f-b97d-43dbe8d02cbf">Required if Concatenation UID (0020,9161) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6c10cde8-b851-476f-821d-e445cef698cf">In-concatenation Total Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_dec3e5b3-fb3b-4a9d-a0ab-85afbe1f649e">(0020,9163)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6b1a46e7-07d2-4481-8d48-2d553121d1bb">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c125d1aa-4503-4de8-8feb-8a3187fe7c4b">The number of SOP Instances sharing the same Concatenation UID.</para>
                                    <para xml:id="para_52385320-ec6f-42d2-bfa1-533f5c7dcf37">If present, shall have a value greater than one, unless an IOD overrides this constraint to enumerate a value of 1 to prevent the use of Concatenations (e.g., ...).</para>
                                </td>
                            </tr>
                            <tr>
                                <td align="left" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Total Number of Frames in Concatenation</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0020,eee1)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Total number of frames in all SOP Instances that comprise the Concatenation that is a Multi-frame Image.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">If present, shall have a value greater than zero.</emphasis></emphasis></para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.7.6.29.1" status="4" xml:id="sect_C.7.6.29.1">
                        <title>Sparse Multi-frame Functional Groups Module Attribute Descriptions</title>
                        <section label="C.7.6.29.1.1" status="5" xml:id="sect_C.7.6.29.1.1">
                            <title>Selected Frame Functional Groups Sequence</title>
                            <para xml:id="para_2e96f200-d4b1-4443-aebf-ccc871f2f427">...</para>
                        </section>
                    </section>
                </section>
        
        <section label="C.8.12.4" status="4" xml:id="sect_C.8.12.4">
                    <title>Whole Slide Microscopy Image Module</title>
                    <table frame="box" label="C.8.12.4-1" rules="all" xml:id="table_C.8.12.4-1">
                        <caption>Whole Slide Microscopy Image Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_19b0b4dc-4e92-4c95-9988-804684315cb7">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_96a09a1d-8879-43be-97e5-121291459369">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3e3024bf-3c64-498a-ba60-ec15eea8abd4">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_18f4a093-c520-4639-ae1b-71059a30ef26">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7a877f73-4d9d-41b5-86d4-a8c3bbfedf52">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_379f1037-05af-47e4-a7cc-263ccdb59552"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5e62e3f9-7cd2-48f1-898c-3cc78973e9f9"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8987cdf6-559b-4308-8bb6-886172153890">...</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_305dffa2-cc70-4472-81da-73a738d8d1db">Number of Frames</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b35b2b79-4aeb-4806-88cc-eb002a89af1e">(0028,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_df7f6e9c-de29-4a8a-8fc0-c33eeb2a82d3">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_60b26676-3382-4932-b4fc-fdf092f8966d">Number of frames in <emphasis role="bold"><emphasis role="strikethrough">a</emphasis><emphasis role="underline">the</emphasis></emphasis> Multi-frame Image<emphasis role="bold"><emphasis role="underline"> in this SOP Instance</emphasis></emphasis>.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Image Type (0008,0008) Value 3 is THUMBNAIL, LABEL or OVERVIEW:</title>
                                        <varlistentry>
                                            <term>1</term>
                                            <listitem>
                                                <para xml:id="para_495140e7-94cc-4a17-aee0-9ae324754fea"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <note>
                                        <para xml:id="para_9b062dec-aeb3-49ab-93b6-9a4d7d005b91">This Enumerated Value of 1 previously applied to Image Type (0008,0008) Value 3 of LOCALIZER, which has been retired. See <link xl:href="http://dicom.nema.org/medical/dicom/2021c/output/pdf/part03.pdf">PS3.3-2021c</link>.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cbf7b471-b8a3-4750-9c1d-2ca54d2bb4f1">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e47d8169-473c-46a3-a20a-4e3ca54bc3aa"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8abcf907-22b6-485b-86de-d562dba4db44"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c4d8ad7a-4df4-458c-86e7-6f1e897522bf"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>

        <section label="6" status="1" xml:id="chapter_6">
            <title>Registry of DICOM Data Elements</title>
            <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
                <caption>Registry of DICOM Data Elements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                                <emphasis role="bold">Tag</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                                <emphasis role="bold">Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                                <emphasis role="bold">Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                                <emphasis role="bold">VR</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                                <emphasis role="bold">VM</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_62be808e-622b-4053-948d-81983b3ab87b">(0020,9228)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_934bdb49-d119-4f25-9930-539105afcff5">Concatenation Frame Offset Number</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9c37e3ac-5147-4dd4-80f3-dc838f41b646">Concatenation​Frame​Offset​Number</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6e362079-86ee-4cb2-9a29-93d18b4e5b6e">UL</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_51c756cc-2a8b-444d-9c8f-909214e803aa">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d1bdabf3-cc76-4431-821b-bf145795eab2"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(0020,eee1)</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Total Number of Frames in Concatenation</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">TotalNumberOfFramesInConcatenation</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">UL</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4d114d50-1542-402c-a71a-8812e0221943">(0028,0008)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a8a709f1-cd24-47dc-9190-bb16d204003d">Number of Frames</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_63b911ad-fce6-4a9b-b4a2-ae0110b2356e">Number​Of​Frames</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_73fcfdd7-1ad8-4c6f-9216-e744237c17aa">IS</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0d734f8d-e852-4fa2-91bc-5da756ece5b5">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eb62a039-2e58-4098-ba49-4fbc992f42d6"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
    </chapter>
</book>
