    <title>CP-2506</title>
    <subtitle>Length limit of Detector ID</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/01/21</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Andrew Chaffer</para><para><link
                    xl:href="mailto:andy.chaffer1@gmail.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/01/14</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2506</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Length limit of Detector ID</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>Detector ID (0018,700A) gives a unique serial number for each detector.
                        This is useful as it allows the detector a clinical image has been taken on to be uniquely identified which can be very useful if there are issues such as artefacts.</para>
                    <para>Detector ID (0018,700A) is a short string, which is 16 characters in length, and is too short for the serial numbers of some detectors.</para>
                    <para>For some use cases (such as multi-energy CT) the problem has been solved by the UC VR X-Ray Detector ID (0018,9371).</para>
                    <para>Propagate the use of the longer attribute to all uses of the shorter attribute, allowing for mandatory uses of the shorter attribute.</para>
                    <para>Note that identification of the detector is distinct from identification of the entire equipment/device (so Device Serial Number (0018,1000), an LO VR data element, is not applicable).</para>
                    <para><emphasis role="italic">[Ed.Note. We could retire the use of Detector ID (0018,700A) but since it is used as Type 1 in the Breast Tomosynthesis Contributing Sources Module, this would theoretically be a breaking change.]</emphasis></para>
                    <para><emphasis role="italic">[Ed.Note. Is it worth harmonizing the description of the two Attributes, e.g., make both "The identifier or serial number of the physical X-Ray detector used to acquire this image"?]</emphasis></para>
                    <para><emphasis role="italic">[Ed.Note. Are we pushing the generalization here by re-using an X-Ray CT detector ID for general X-Ray purposes, given that one is a flat panel and the other is an assembly of multiple discrete detectors presumably identified as a group?]</emphasis></para>
                    <para><emphasis role="italic">[Ed.Note. What  about trying to re-use UDI at the detector level - probably not.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>        
        
        <section label="C.8.11.4" status="4" xml:id="sect_C.8.11.4">
                    <title>DX Detector Module</title>
                    <table frame="box" label="C.8-71b" rules="all" xml:id="table_C.8-71b">
                        <caption>Digital X-Ray Detector Macro Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6a879d21-ea75-4f7a-aff2-991c91525d32">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8bfa8450-be0f-49b0-bc7d-65263c8512e1">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bb673b08-f7a0-4741-b478-eb731e910af5">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d962e9af-401c-4897-aae3-329828b8574c">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fcfe3e48-28f8-4c92-b12c-3895804d49e7">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d0022743-b426-4f20-a3f9-fcdadada6ba2"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2e4def81-4fcc-40b7-808c-544dd55787bd"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4bb3fd10-ac3e-4390-8f52-b219b4b2fcab"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5eb7d82a-f55b-49f7-a373-4dbb385574de">Detector ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f9120f2b-7824-4b39-8d75-f0d6ded12c14">(0018,700A)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5c7b93d0-8e1e-4c1c-8345-cc201bb459c5">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b4238dd5-3645-4ca3-8fc8-ea6266e8bf87">The ID or serial number of the detector used to acquire this image.</para>
                                    <note>
                                        <para><emphasis role="bold"><emphasis role="underline">Due to the length limit imposed by the VR of this Attribute, a full length serial number may need to be truncated. If truncated, to the extent possible the remaining string value should be as unique as possible within the local (site) context of use, and hence perhaps be the tail rather than the head of the truncated string. The full length value can be encoded in X-Ray Detector ID (0018,9371).</emphasis></emphasis></para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">X-Ray Detector ID</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0018,9371)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Identifier of the physical X-Ray detector. This might be the serial number.</emphasis></emphasis></para>
                                    <note>
                                        <para><emphasis role="bold"><emphasis role="underline">If the value in Detector ID (0018,70)A) is truncated due to VR length limitations, this Attribute may be used to convey the full length value.</emphasis></emphasis></para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_35e9b188-dc05-4875-ab27-d5d984cb1f60">Date of Last Detector Calibration</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8390788a-d816-439f-b08d-81a3213696ed">(0018,700C)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_89a042cd-f6c7-495c-a559-3ace4088306b">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7cd2d2f1-6aed-4254-8de2-aaa9fe1d44b1">The date on which the detector used to acquire this image <emphasis role="bold"><emphasis role="strikethrough">as identified in Detector ID (0018,700A) </emphasis></emphasis>was last calibrated.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6dd35b27-cbf5-4c45-aac8-f1c28b735eab">Time of Last Detector Calibration</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5bef6d60-6d91-47bb-a8a4-505b6c966bc4">(0018,700E)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fb939b4d-1ee3-49a2-bec7-106fd7d0a5cd">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a832be95-7995-4ba9-8282-22c7a881013d">The time at which the detector used to acquire this image <emphasis role="bold"><emphasis role="strikethrough">as identified in Detector ID (0018,700A) </emphasis></emphasis>was last calibrated.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f5fee9f2-1636-4a5a-8c09-3836a8ded8ad">Exposures on Detector Since Last Calibration</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_10c5759c-2b41-4909-bca3-b584a0e5d2c9">(0018,7010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f7a57fa4-c519-4ea3-8d94-72fb572079a1">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dea00a78-7e88-4c56-911d-e241351abc9b">Total number of X-Ray exposures that have been made on the detector used to acquire this image <emphasis role="bold"><emphasis role="strikethrough">as identified in Detector ID (0018,700A) </emphasis></emphasis>since it was calibrated.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a3144c0e-3891-475a-8378-0f5cbdd61a17">Exposures on Detector Since Manufactured</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d42c1d20-0147-4a4f-89d6-ca7c600e6fe2">(0018,7011)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_21e05a24-090f-4700-8bfd-03e34bef487f">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_89de6587-16c3-4179-b81a-c9bc372928ad">Total number of X-Ray exposures that have been made on the detector used to acquire this image <emphasis role="bold"><emphasis role="strikethrough">as identified in Detector ID (0018,700A) </emphasis></emphasis>since it was manufactured.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_843e2170-87ae-4990-8f0c-cd7847662827">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1f34f118-d76a-4321-8a87-6c12ef94c643"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e415711f-e7a2-4b82-9c8c-f8915979d270"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d9b69301-4f81-4527-99e1-66c6dbe2540c"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <section label="C.8.21.2.3" status="5" xml:id="sect_C.8.21.2.3">
            <title>Breast Tomosynthesis Contributing Sources Module</title>
            <table frame="box" label="C.8.21.2.3-1" rules="all" xml:id="table_C.8.21.2.3-1">
                <caption>Breast Tomosynthesis Contributing Sources Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f38ef4ea-d4dc-441f-bac4-b45fdd2bd6fa">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_13407253-a815-4ef6-bcda-502d9e57ade1">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ff8ec414-df44-4981-b02a-d4102e0b84bf">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_23a3a928-8398-4d94-b7db-0515a16a4d86">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_66514ac0-dbd5-45fe-bf7c-8e1c6c345f39">Contributing Sources Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b00e02d1-a810-4c36-bb8e-1ba627440c01">(0018,9506)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4bcd446d-9073-4753-a8e1-cc766c286e13">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cd84c1dd-924a-4f79-b9f7-e3abf5b35dd6">A Sequence that describes characteristics of the sources that are used to create a derived SOP Instance.</para>
                            <para xml:id="para_140f221e-7293-4d6b-b947-00e3f44d2094">One or more Items shall be included in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_24313039-5c5b-4db3-b62c-0934d796c874">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f571a348-5f42-4ff8-9d64-5ff90b894034"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_80e87f22-979d-44c9-a98d-0839e6672543"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_87490e8c-55d4-4da1-96e1-d2ace7efeddc"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8ac2cd02-ab36-4125-b378-d0cc0fed7b8c">&gt;Detector ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eb1ec6f3-03ad-42eb-9fb8-4daeb55458f2">(0018,700A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_abc86cb1-8f5c-4afa-a099-ff3de302bb4a">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f15f1799-905c-4b14-9c6a-cbd47e3ac654">The ID or serial number of the detector used to acquire this image.</para>
                            <note>
                                <para><emphasis role="bold"><emphasis role="underline">Due to the length limit imposed by the VR of this Attribute, a full length serial number may need to be truncated. If truncated, to the extent possible the remaining string value should be as unique as possible within the local (site) context of use, and hence perhaps be the tail rather than the head of the truncated string. The full length value can be encoded in X-Ray Detector ID (0018,9371).</emphasis></emphasis></para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;X-Ray Detector ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(0018,9371)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the physical X-Ray detector. This might be the serial number.</emphasis></emphasis></para>
                            <note>
                                <para><emphasis role="bold"><emphasis role="underline">If the value in Detector ID (0018,70)A) is truncated due to VR length limitations, this Attribute may be used to convey the full length value.</emphasis></emphasis></para>
                            </note>
                         </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_24bc9afc-d419-4108-8f23-e612f5947583">&gt;Date of Last Detector Calibration</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_33603ba2-3a94-46d5-8dac-5e9d3d887263">(0018,700C)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4157c704-6ae0-41d0-bf83-08712bc5a92d">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_aa1bd515-d9a6-49bb-968f-9b0b27496463">The date on which the detector used to acquire this image <emphasis role="bold"><emphasis role="strikethrough">as identified in Detector ID (0018,700A) </emphasis></emphasis>was last calibrated.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bdee8b10-11da-4480-aa22-75a6df9d258e">&gt;Time of Last Detector Calibration</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_45bc98e8-13e7-4a0e-98b8-c6a8d8c40192">(0018,700E)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f97cc496-d61b-42f3-aed4-0528cbf8c85b">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d5679243-fc23-4ae7-b6a0-c0db97270e2e">The time at which the detector used to acquire this image <emphasis role="bold"><emphasis role="strikethrough">as identified in Detector ID (0018,700A) </emphasis></emphasis>was last calibrated.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3094144c-16c1-46a7-930f-7ef4c9f5ca58">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_954dbaa2-a594-4f36-9350-efc1f7d718e8"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8f9fa468-081c-41f1-a2f3-a53928591100"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9b08f392-a14d-4fd1-96e3-f9d5b9d2fcff"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged DICOM PS3.3:</emphasis></para>
                </td>
            </tr>
        </informaltable>        
        
        <section label="C.8.2.2.2" status="5" xml:id="sect_C.8.2.2.2">
            <title>Multi-energy CT X-Ray Detector Macro</title>
            <table frame="box" label="C.8.2.2-3" rules="all" xml:id="table_C.8.2.2-3">
                <caption>Multi-energy CT X-Ray Detector Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_836a5806-2f26-4a34-a736-33395d439203">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a1f392cb-906c-435b-977f-ece1d4c27aa3">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0f09fd19-4545-4203-a8d6-b3c8ba30149f">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4a7d0426-8473-40e8-b67d-b35a11e17a99">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f71f3f6a-da5c-4e06-a20d-32760481750d">Multi-energy CT X-Ray Detector Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_074d3791-fa4e-4280-b0df-0bab12343cb8">(0018,936F)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_96bf2fe7-29b3-43c1-9183-a77c8c7ab6c7">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>...</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_804defaf-7381-4748-85da-9b123659fb96">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c5601544-2126-412d-9ef7-91778c491976"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cc917b2a-3aa7-47b7-aeef-518c178271e0"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_18b33106-c103-4054-83eb-9d75208ceff2"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3fb224e7-1a4f-4740-9874-3695db34b0ea">&gt;X-Ray Detector ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_88958bf1-3cb3-4d51-b27a-87371ea800c9">(0018,9371)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_56640bea-391d-499f-9684-07c6b175c1a5">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8668ae8f-5d50-4a28-b0a9-78851b977fee">Identifier of the physical X-Ray detector. This might be the serial number.</para>
                            <para xml:id="para_8aa441fb-faa1-4adb-b9eb-4d85abc697f5">When a single detector discriminates different energies, the X-Ray Detector ID (0018,9371) will have the same value in different Items of Multi-energy CT X-Ray Detector Sequence (0018,936F).</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_28391467-9357-4ffb-96f9-2d4d759f66a4">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e96ffd00-dffa-4ec5-94c0-adc2a740ddf2"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_73b6de8a-6420-4ce4-9652-d04048b913ae"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_049562d6-be27-4134-965e-3524b5471710"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged DICOM PS3.6:</emphasis></para>
                </td>
            </tr>
        </informaltable>        
        
        <section label="6" status="1" xml:id="chapter_6">
            <title>Registry of DICOM Data Elements</title>
            <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
                <caption>Registry of DICOM Data Elements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                                <emphasis role="bold">Tag</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                                <emphasis role="bold">Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                                <emphasis role="bold">Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                                <emphasis role="bold">VR</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                                <emphasis role="bold">VM</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ee5350c-4b38-4b5e-bbbc-e8764ff5b8c5">(0018,700A)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c2ef2691-021e-4690-a93e-70f59ecb0153">Detector ID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5a145417-75a1-4372-a81a-928e917bc871">Detector​ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_30f90b1b-fd49-4a10-9007-426b742ea5b0">SH</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9d02ab41-d138-494b-af38-7894676d491d">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_502bd155-4a75-4269-a8fc-614b96092c76"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_70b82313-e2da-4436-afb4-f0d5893c4479">(0018,9371)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2b5e972c-37ca-4911-be70-b60f91427faf">X-Ray Detector ID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_603c9bc8-00b1-4e76-bf2c-5cd6d8f832c8">XRay​Detector​ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_385f2cb9-7b72-47b0-a5ea-a799177cbb05">UC</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9eff2c06-e14b-400f-acaa-29a5e8b8b972">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a2d3fd15-32f8-4e7b-a427-795295fa643f"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
</chapter>