    <title>CP-2484</title>
    <subtitle>Analysis Performed content item in TID 3905 should have HAS CONCEPT MOD relationship with Finding parent</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>JLetter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2025/06/20</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2024/12/02</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2484</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Analysis Performed content item in TID 3905 should have HAS CONCEPT MOD relationship with Finding parent</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The Analysis Peformed content item in TID 3905 should have HAS CONCEPT MOD relationship with Finding parent rather than CONTAINS, to be consistent with all other uses in the family of cardiovascular analysis result.</para>
                    <para><emphasis role="italic">[Ed.Note. The Analysis Performed code is in a different manner in Mammo CAD TID 4018 CAD Analysis Performed, not as a modifier of a Findings parent.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>     
        
        <section label="TID 3905" status="3" xml:id="sect_TID_3905">
                <title>Calcium Scoring Results</title>
                <table frame="box" label="TID 3905" rules="all" xml:id="table_TID_3905">
                    <caption>Calcium Scoring Results</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e1c57370-c5dc-4e18-a2b5-a3443d46e1f4"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_458818ef-0429-4e8c-aa99-419ae704e696">NL</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d1c83fc3-c91e-49bd-8ab5-a093db7905c1">Rel with Parent</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1257fa56-b137-4366-9b91-72f5a32dd3ed">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f15ff7d4-c1ce-4e9f-9235-e73d28424e6b">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7ffd96ed-52ac-4482-846d-927d93dc0aea">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_42e90442-a216-412f-ac5b-747014f839b4">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_afb76a84-e34d-4cb5-92d0-bf56db96294a">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_28bfafa8-a94a-41eb-b3e9-36956f93de56">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a416defe-c731-4bc4-b7b5-2a6a0d0941f5">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d5410d3b-db4c-4520-b4c2-a890da212ce5"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0196ca1d-5070-4266-a514-4d566b131e98"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_41a9cf32-2b9c-4678-8ff8-fe3bd5cbbaa4">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e78c658f-c403-4a31-9a8d-eedf0a196c2f">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8ee9d87a-3517-48fd-9701-5783225d28ec">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a1d7b750-bfaa-4fbe-a016-6f6b0d905daa">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6c015e9c-dedc-448e-9a3a-256e6e169677"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f6195128-ee57-45c1-ad37-565a815b13c7"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_68f58140-75bc-4124-ae5b-1c6ded0c6741">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a10f2d8e-8786-4c7d-91dd-b38ca57f5830">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_96244a2e-08c9-4981-926f-e2bd196f5f6c"><emphasis role="bold"><emphasis role="strikethrough">CONTAINS</emphasis><emphasis role="underline">HAS CONCEPT MOD</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_41d28b6e-f136-4cd9-b7c4-112004df2dbb">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7027a850-9158-41a2-b09b-ba0538aeee8f">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_69d614c0-2d4e-4094-adec-45429c9f77c7">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f724b84d-f3ef-41c2-9b20-e7a70111dcb7">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d23b4ba0-2ff9-4aea-a2f7-9d9b1627540a"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_244e22ff-755d-4ff0-b75f-bc5e6f30806e">EV <link linkend="DCM_122603" xrefstyle="select: nopage">(122603, DCM, "Calcium Scoring Analysis")</link>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7836ac98-ee32-4ed7-bfea-0830a3fc737a">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2ba136c3-3891-4e1e-bc53-662be618b70e">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_23b1b3f3-8695-48a2-b469-af7c144de5e1">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5ddee210-92dd-45bf-ab72-2bab4f119bbe">NUM</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_08716687-2ac0-46c5-b355-5d968d8e35f1">EV <link linkend="DCM_122657" xrefstyle="select: nopage">(122657, DCM, "Agatston Score Threshold")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0def24b6-8d72-455e-9fe5-69902c832440">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c5388003-48f9-49a4-8a1f-19c2da1293f0">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c5250886-4334-4698-b8b0-68af6e4bd518"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_272af562-67b8-461c-854d-c06916ffceb7">UNITS = DT ([hnsf'U], UCUM, "Hounsfield unit")</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d88c85d1-283c-4619-acb4-3f87a849f899">...</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_01f9a3d4-4f5c-48b9-aedc-ff1366ac87e6"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_77299da5-f09b-45f1-ac0d-714b31e1ed69"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b674fe88-e7e3-4180-b709-cd28e57e8e6f"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_08c9be6f-226d-4837-ba19-2b796ec6dd29"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_406392ce-e7a7-47b7-bf51-3d0d3e9ebb1e"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_05ec2c2f-f086-4506-93d8-791b86d3051b"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_47638495-063b-4256-946d-72eaef874ba9"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fdf8549e-1506-4582-a5cd-fb2948cfca07"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        


                
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <section label="TID 3902" status="3" xml:id="sect_TID_3902">
                <title>Vascular Analysis</title>
                <table frame="box" label="TID 3902" rules="all" xml:id="table_TID_3902">
                    <caption>Vascular Analysis</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_112fdc5d-f36c-4fcc-925b-4cd9a948d879"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_40f793ed-0108-414c-95ae-120bceb0a4e2">NL</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ed8cc5ce-0107-45c5-bc2a-2584c37197c5">Rel with Parent</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b74df532-832a-452a-af7f-fd0909664186">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e3dd22cb-0ec6-49e1-a8dd-86681f504d0d">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f4002115-8700-4f68-af42-448d32125c85">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9c1d2bd0-5c1f-4db7-a9f4-59102f5fa5b9">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bd06156a-0031-4fa0-a0b4-ac8244c74191">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ad9e2219-ab3a-4d4b-976a-a91e4f9ad618">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7842b6d4-7bc1-4489-8acc-b7864daa9b4e">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cc20ade8-efba-4d68-8233-abdc7075b8a5"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_65bd2942-a874-4c25-bb3d-c2b402efe5ac"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_28f23669-915d-4636-95c7-1ed94e451570">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c97e7b2b-7510-40e9-b9c9-566bafd41864">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fd8f1b9b-eece-44bc-a70e-ca752b8e076a">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac1004c7-e974-485b-8052-94c3cccd9649">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6447a5f2-e5bc-4c70-af0a-27d8f014a2cb"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_be9e7e0b-49ea-474a-b6dc-fb09455f8f4b"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_30f16e97-70be-4283-a54b-c258ef729afe">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_020dfeb7-89ca-4f4f-a6af-ca5d52abb80e">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_805c9ec5-edb7-419c-8805-52fc760956ac">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_44690b4f-bfed-4cb2-ac8a-fba26b74ef27">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f743ea64-e3e1-430b-a52b-352f363c285c">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8c67bac0-977a-4d13-8ef9-93608f93fef8">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_182d83f0-7c3e-453b-b858-7de4997ee210">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d5cc342b-be79-459c-b0e5-c62d58b4cf90"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3fc86661-fc54-44f7-8823-21cd927cbb92">$AnalysisPerformed</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_07d6c670-0f30-4b44-9dda-803627d49b81">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_835c587a-2bdf-487f-aabd-c0f162da4352">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_316bb7bd-b27a-4678-bc73-1bfaf3b17910">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_eee6430c-eef7-45eb-b04b-96392bb9d121">INCLUDE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b2fab699-9e9a-4fe4-b31e-2cdcfa037175">D<xref linkend="sect_TID_3905" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b269016c-df76-4162-9f66-263942b17ffa">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_07485efb-217c-4a24-b1a9-b09a08347814">UC</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2f94cbce-c534-4c7b-b3c0-2bc444626305">IFF the value of Row 2 is EV <link linkend="DCM_122605" xrefstyle="select: nopage">(122605, DCM, "Vascular Morphological Analysis")</link>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_010b3c57-cd0d-42cc-a8c8-3def2df0381f"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_910a307d-e488-43da-a86d-5c2281b2cd24">...</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9a37797d-a131-4903-a44a-cd31c8f1766d"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c203aa2c-0afe-42fa-a49e-4070629ced89"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_94291be5-f420-449b-b741-018e6986f08d"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7a99e95f-6eed-48bc-bd83-aee5901afabc"> </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b01343b0-55fb-4540-9082-11ce4c34022c"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_de747cf4-3916-428d-a344-6245cb452c94"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fa7077a6-fbb8-42c5-b128-1f42e9d66df5"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_084c9292-557a-4949-a2fa-b14ffe51ddb1"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        <section label="TID 3910" status="3" xml:id="sect_TID_3910">
                <title>Flow Quantification</title>
                <table frame="box" label="TID 3910" rules="all" xml:id="table_TID_3910">
                    <caption>Flow Quantification</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2d0a27bf-853a-41da-81c5-5342fcb0c016"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_62477605-f703-4037-9660-df4bb913bcd3">NL</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a4c6b3a6-a6eb-4998-a69c-d0285dbaf393">Rel with Parent</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d4acd748-66de-4037-b620-25bb455f0b4e">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ae9978cd-78ad-4e0d-ad80-45c1dda9002e">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5b989dce-4d38-4c5c-ad18-5815400d01fd">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bba9bb41-6aa9-4554-9901-ec78cecf08a4">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bddefeb2-f8a7-4ae8-a640-5d927a580080">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_68f6e65d-c38b-4f3c-92ad-2abf3c88f156">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4ca72945-33ea-4a41-8e5f-fd1d2382856b">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b5d23f82-912e-4827-ae7e-2299b1e94fac"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f9334a10-d733-40cd-a02e-428b1adee02d">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1fa4c0a3-d38d-4e0d-9f8b-a0f2f57d5a12">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4273691f-9b2e-4eba-9509-ad0b3bc0e5dc">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c1022a9a-a8ce-4d19-b07a-a45289b26056">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f43c7390-7a39-4e83-9b6c-a6f708c83ff4">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7652ad26-4977-4448-8a48-543b2bb2a29e"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_eb1b5bb8-e764-490c-958a-15cd42e4fa6e"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_070dbccc-0510-47e9-8a5a-e05e47626666">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ff9a4dd5-2d9a-4e39-895a-f25159ed585e">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9cb0dcdc-730b-45fd-8106-8f7a9921792c">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b26d51e1-906f-4b6c-9358-2bdebb176cb6">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9c16a220-bf4f-4868-80c3-c3ac8c6ac41d">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_10b8ffc3-def4-4702-a769-c1b4a14bd859">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a728046b-1258-4a6d-b5a9-519b0dfb2b59">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9149efb3-736a-4a6b-9a8d-4fcaa502c827"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cc553f56-059f-4941-a1e1-c0610ebc9bcd">EV <link linkend="DCM_122604" xrefstyle="select: nopage">(122604, DCM, "Flow Quantification")</link>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c1e7fb75-5398-4a1a-a3c4-701d8f733e46">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_32922320-c0d0-465e-9556-6f689a472754">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_555d0d8e-3e28-4d58-9d1b-2a9f6c2bf025">HAS OBS CONTEXT</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d53c72ba-1df2-4039-8cf1-2480ef4d7903">INCLUDE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3bb7dc3b-8361-4db4-99a5-c379d28cf8ec">D<xref linkend="sect_TID_3929" xrefstyle="select: labelnumber quotedtitle nopage"/>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_30619879-5d16-42af-909c-7d904c8b6542">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a7912dc4-d90f-4baf-b2cf-7b74838766a4">U</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_58d1b3e9-59d6-4991-bffa-16ef0eb1dfd7"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1cb5ed36-2b27-4bcf-8bd2-546467b5d432"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_842f07ab-011b-42b7-b07f-02748f3fdd63">...</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_10fc00e3-df1d-4219-aae3-6315e80c2886"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7ffea243-e60c-4de0-bf90-c748bee290f2"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8e192595-c692-4d89-bd6e-292f0fcea09a"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5095079d-a227-41f6-85e9-e651ebde7e56"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_41b20129-5404-4181-97b2-35a53a07a62f"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f1988fa5-dae9-4034-98e4-67408b37a6a4"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_853182db-210a-42ec-a3af-b974bf367c92"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_222ce8f0-a3e5-4672-b3d9-ab8d3450633a"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        <section label="TID 3920" status="3" xml:id="sect_TID_3920">
            <title>Ventricular Analysis Results</title>
            <table frame="box" label="TID 3920" rules="all" xml:id="table_TID_3920">
                <caption>Ventricular Analysis Result</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a562384b-adf3-4ea9-a6cb-d95a7da75fda"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_77ffe718-4a2c-42cc-b7a9-b2f1bbedd41e">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8b2bd98e-ef8d-44f1-ad3a-798262fabd8d">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_63a85b16-236e-4f19-839e-3d26dc865f5a">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7dcb85f1-8191-4116-89c1-ab7cecef8e2a">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d8136bb2-e5ff-4355-8fa2-a0dfeca101bd">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c6746345-2c09-4b89-8680-cccff25b9bfa">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a3801587-bbc8-4519-8afd-e8d185ac2c37">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9ebc95f4-599e-4ce8-83f7-20db0950f876">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eccbf709-5d9d-4da5-a1be-d202e5d305c8">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_286fa34a-69b5-48c2-9343-ebacf199193a"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6a708a5a-bc52-4c68-99d2-5ac5662936a2"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cf7c0afd-04b7-4e6b-982d-370dca4cc046">CONTAINER</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2cad54e8-ef89-4599-871d-0d7c863a49eb">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af33498-8cd1-4c7e-b24b-ace43ff0e6e7">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_20b0e6aa-3514-48c0-8770-4e46f8c4aeca">M</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ab773431-cc2b-453b-bdc2-bd4c6e238517"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_41f60803-1a36-4b70-88b0-b9dc7a712a70"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_10fb74c3-4643-410b-9c4f-b80b5cc2889a">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f1d666cd-b637-469a-87f5-18c382590c99">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_481c8139-66f4-4352-a888-0a188e1525ea">HAS CONCEPT MOD</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ab271be9-2fe5-4d01-8f90-5ad547c77647">CODE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5d486a2b-6f9e-4b6d-b28d-da4047eab488">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9a0738b6-858a-4031-bac6-e4b71d542c64">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_68196b8d-d09a-4f7b-99ae-7fc9d5592340">M</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e6921a53-b7ee-4bc1-b77f-79eb4b6fb32f"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_22a6f9fd-12ac-4e9f-a92e-506017ed05ef">EV <link linkend="DCM_122601" xrefstyle="select: nopage">(122601, DCM, "Ventricular Analysis")</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7c02e811-6660-43de-b65d-bb56a5b3de4c">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5d351c77-5402-49f7-8e95-86900953cdf7"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_61bd1f70-a70e-450d-b831-79361cab2391"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_36888f34-58a3-4110-b95a-0cf6fc799f80"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1a5d9c13-be2a-490d-aab8-bbb009ef74a5"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_aacd2d61-b0cd-4c55-8338-97e0b56bd416"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9aab9f87-e99a-4905-9768-36e5df65da66"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa9962bc-2bfa-4946-b40e-e5f9fdd00502"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ae5cd2f1-65fe-49b2-ac98-4dbe50a44979"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="TID 3925" status="3" xml:id="sect_TID_3925">
                <title>Ventricular Thickening Analysis</title>
                <table frame="box" label="TID 3925" rules="all" xml:id="table_TID_3925">
                    <caption>Thickening Analysis</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_41603abe-69cd-4967-a977-3493db444acf"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e0fa04d0-4f49-4b21-b91e-6fee15536554">NL</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_24368ee5-64a7-40c1-b107-004049b5ca39">Rel with Parent</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_14fc7f61-b9e6-424f-aa6c-b8081dabe20e">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_afac3559-b161-4779-995e-3b701cda3f90">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5bda1ae0-9f93-43da-a79a-7d29d79bd8ae">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cac9911a-d87f-49a2-8ac8-79cf9aaf2e1f">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_647a5d1d-514f-409c-9ba3-bbc9e0a95a26">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e232bc19-90b0-4d08-8050-fb6c79cff586">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_85cf5f88-d78a-402d-afac-c6527aeab344">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_de5fc35f-3845-4678-a62c-f4bc93280237"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a6bf5f9e-f6df-45ad-86bf-1213b4a0d8f2">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d8f129d3-bff5-4c28-aaef-d433646a198c">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fdebe778-7641-4249-9a7a-fb753aeff55c">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0cb754c5-a39b-4fce-953e-3785a6f074ce">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_715c8510-8d90-4260-abfb-ea1758c0c631">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_85896cf8-ad64-40cb-b266-57ac72be3d7f"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_be58e36d-93f9-4fe0-ba7d-6ad88d4cd492"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_78cbefee-a934-4fae-9931-18a4f9a6d875">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1aa98427-af22-4f70-8901-4afd113dc0cf">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9b3904f8-248a-4b39-b296-e9e3ff018de8">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9a49862c-6662-4204-8649-098c7b934e7d">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_039916fb-809d-4577-b397-12cfff6c4496">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_34b25927-986d-407b-b8ba-d2234038975c">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5c2727c5-9660-4eed-9ba7-9b2e27099994">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8997b5a9-7652-42c8-ac23-c91b909571e0"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6b0015c5-627b-4924-a341-6f521c40228a">EV <link linkend="DCM_122607" xrefstyle="select: nopage">(122607, DCM, "Thickening Analysis")</link>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6a31fb9e-8fe1-4441-947a-998afc733c37">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6558693c-ea60-4177-acfc-045867adb70b">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_43086f80-ddcc-45ba-8e6d-460ca6d27bfa">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3c163a2b-0e5a-4a36-9475-fbe9854c54d3">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_508b26fd-1e70-4205-a2bb-f56675f42e90">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d5afbf59-078f-4b5c-9364-e443447437e9">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_835699fa-f4c4-4514-8e0d-41f596d24961">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_29f5026e-80fc-4308-a1b9-f2fb4d47fd73"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5263fc55-6100-4cc8-af23-a622704e76f7"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_236be1d1-50b3-492d-af07-e8b80653df94">...</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8b8b82d1-ab7d-4a0a-9124-86c4cb2ceba6"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_609299f6-79ca-4b80-9edd-3a0f4937c8a5"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7c2e2437-5587-4583-b401-c4757c767e26"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7ac93508-7955-456e-aaf0-314afbe230ec"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_98473913-75c8-4150-9deb-cf37ac01e6e6">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ab900f9b-6c9c-443d-a737-7fd7e3f3bab0"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1e88c1e9-6bfa-43a0-94a9-98af3d4cb408"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b6710455-689e-43eb-be89-0c18eff6c72a"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        <section label="TID 3926" status="3" xml:id="sect_TID_3926">
                <title>Myocardial Perfusion Analysis</title>
                <table frame="box" label="TID 3926" rules="all" xml:id="table_TID_3926">
                    <caption>Myocardial Perfusion Analysis</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6e96ae8c-dd6c-4fa0-93a5-204d52ee893f"/>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_86d506ee-4bd4-439b-991f-000a0fe6b0bb">NL</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_866d9b4a-0552-49c7-86fd-7442409c8bc7">Rel with Parent</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ea1338e7-ac71-4118-b95e-7a7a08919362">VT</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac8891af-eb1c-41e6-b217-88ee73d7f8c7">Concept Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_01f03a15-b7e0-4727-a83d-b4654bba007a">VM</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_94adeccc-e167-4dcc-8e44-1cd222cd03c7">Req Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ffbe1fd0-5e56-4526-bc53-e713dd1dee2f">Condition</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a712f39f-7a38-4c93-96c8-374870b4d48e">Value Set Constraint</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_dce88753-8f3d-4e4e-9e18-15ae8ad8f99c">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_01aa7ac1-7015-4e55-9c00-3c54606c8b9a"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ff5ae340-fb5f-43b1-9d48-3b2c359b0b36">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_81254521-a366-4038-b141-a6d73da3b9c0">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fedaf06a-6a7f-4fdf-a1ed-2a65e380a9fb">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6498644a-b323-4fef-a176-4a4fc18b725d">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a3231aae-483d-4256-856f-260f667fdcf1">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6e66d335-b972-4acd-9130-eefa7fa0e76a"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_20061213-b072-4697-80e2-d3e6b8e33e88"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ca9f0f90-bbcb-4106-9c86-624da353e697">2</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d2204d83-e202-4c46-b4d6-c3afb10457ca">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a0754c03-6369-4bbf-9c8b-e79523b3ef51">HAS CONCEPT MOD</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cac1bd1e-da73-412e-93a9-e1e3a910c7c9">CODE</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_caa80131-b66b-44ae-83a9-0d9aaa5d6c19">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a9f2a3f0-e5f4-47b5-a931-1ef20c496f21">1</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_de004032-9d9a-4191-899e-fe45b9043cb0">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8014f58a-1779-46a7-8fab-77f675901501"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_95928f0f-8746-4fd4-8b53-930b27210c5a">EV <link linkend="DCM_122602" xrefstyle="select: nopage">(122602, DCM, "Myocardial Perfusion Analysis")</link>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2b0e4327-a171-4f65-a744-081c53f49fce">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7b76c615-05f9-4502-85a4-919b738a3fab">&gt;</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_72c89d91-53aa-46fc-b516-7f1f5c93f965">CONTAINS</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c0268628-829a-4ffb-a671-48276269adfd">CONTAINER</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c2686cc5-c55b-48c7-b6bb-d06e92694c20">EV <link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LN, "Findings")</link>
                                </para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b3b0bb84-04ab-43cf-b431-0cf2d4913025">1-n</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b60f1b2d-5ed8-45a1-aa95-8d02425806ef">M</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8a4557aa-1375-4e50-9110-6c25297ffca8"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9de94f3c-6bd1-4079-8cd1-fc6987b61a5e"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d8612760-2dbd-495b-8b06-44afb137f72c">...</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bca58dd6-e8d4-4479-9bb2-f8a167bc74a9"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6f024213-e50f-4d10-88a4-ba8ad3a7e48d"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_20995782-73cc-445e-a446-69b6355dadfd"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c6cd4b99-0e62-481c-8c1b-9537ff98efd9"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c1d0d9b1-6d65-4ea7-a929-11f8f09dcccb"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e9d5437b-18a4-4c5f-bdf5-80c4a2684797"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_00e73ca2-586a-4500-a178-65d8cbdb8801"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_65b7eec9-5e39-4868-bd2b-c2dadeb50c29"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged, hyperlink targets:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="TID 3929" status="3" xml:id="sect_TID_3929">
            <title>Cardiovascular Analysis Observation Context</title>
        </section>
        
        
        
        
        
        
            
            
            
            
            
            
            
            

        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        