    <title>CP-2388</title>
    <subtitle>Extended Offset Table clarification</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2024/02/29</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2024/02/23</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2388</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Extended Offset Table clarification</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.5</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The Extended Offset Table introduced by <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://www.dclunie.com/dicom-status/status.html#CP1818">CP 1818</link> contains offsets and lengths and requires each frame be encoded in a single fragment.</para>
                    <para>Yet the description of the length attribute implies that the offsets may be present when frames are fragmented. Clarify that this is not the case (since the "and every frame is encoded as a single fragment" conditional clause is always true).</para>
                    <para>Also, note that lengths may be odd if the compressed bitstream is an odd length (i.e., padding is not included).</para>
                    <para>Also, clarify the description of when the Basic Offset Table is not present (i.e., the Item Tag is always present, but when it is zero value length, the table is absent).</para>
                    <para>Also, explain why offset table is only present for Encapsulated Format.</para>
                    <para><emphasis role="italic">[Open Issue: Is there value in relaxing the requirement tbat the BOT and EOT are mutually exclusive? Is there a use case for both? E.g., legacy readers might use the BOT, new readers might take advantage of the EOT being in the metadata, or having the additional length information.]</emphasis></para>
                    <para><emphasis role="italic">[Open Issue: Is there value in adding another optional Attribute that is the byte offset from the start of the PS3.10 file to the beginning of the Pixel Data 7FE0 group? This would allow readers to skip to the extended offset table and then the pixel data without fully parsng the header before that point. If assigned a low group,element tag number, it would be encoded early in the stream, e.g., after the retired (0008,0001) Length to End (ACR-NEMA message length); alternatively, it could be in the PS3.10 meta information (group 0x0002).]</emphasis></para>
                    <para><emphasis role="italic">[Ed.Note. Account for changes in <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://www.dclunie.com/dicom-status/status.html#CP2362">CP 2362</link>, which adds text to the Extended Offset Table description note.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
      <section label="C.7.6.3" status="4" xml:id="sect_C.7.6.3">
                    <title>Image Pixel Module</title>
                    
                    <table frame="box" label="C.7-11a" rules="all" xml:id="table_C.7-11a">
                        <caption>Image Pixel Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cc00db8b-4a44-4a7d-99c1-2d44223f9377">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_303060dd-f820-473b-a7dc-5eba1e969dd1">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_17de6799-3c80-4d5a-9bda-958f55868735">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_59b2ecab-7027-4266-8506-44ac0f643785">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3106edae-90d1-42aa-8e4e-a069bf8967e6">Extended Offset Table</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bd8aa189-7c34-4cc2-88aa-02724ec1f886">(7FE0,0001)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4959f701-64ce-49a0-a244-f1d5a9d946e8">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f9d1f2a4-7635-4f95-a8d0-aa20049cc97d">Byte offsets of the <emphasis role="bold"><emphasis role="underline">Item Tags of the </emphasis></emphasis>Frames in the Sequence of Items in Encapsulated Pixel Data encoded in Pixel Data (7FE0,0010).</para>
                                    <para xml:id="para_8ea72466-67a0-41b5-9c0b-3aecb92ca759">See <xref linkend="sect_C.7.6.3.1.8" xrefstyle="select: label"/>.</para>
                                    <para xml:id="para_1dcb67d8-3e1f-4b20-bd10-2a5b80c3607b">May only be present when:</para>
                                    <itemizedlist>
                                        <listitem>
                                            <para xml:id="para_5ccc75f3-9968-47f7-9629-b12ee85790b2">Pixel Data (7FE0,0010) is present, and</para>
                                        </listitem>
                                        <listitem>
                                            <para xml:id="para_a9f8dace-d28c-48ca-a9d7-8f7379b8fc54">the Transfer Syntax uses Encapsulated <emphasis role="bold"><emphasis role="underline">Format for the </emphasis></emphasis>Pixel Data<emphasis role="bold"><emphasis role="underline">  (7FE0,0010)</emphasis></emphasis>, and</para>
                                        </listitem>
                                        <listitem>
                                            <para xml:id="para_1df522d0-dc6f-45bd-b03b-98078d2fce25">the Transfer Syntax encodes Frames in separate Fragments, and</para>
                                        </listitem>
                                        <listitem>
                                            <para xml:id="para_74d3ae81-729e-4278-942e-f6a00b001cbd">the Basic Offset Table <emphasis role="bold"><emphasis role="underline">is not present (i.e., </emphasis></emphasis><emphasis role="bold"><emphasis role="strikethrough"> in </emphasis></emphasis>the first Item <emphasis role="bold"><emphasis role="underline">of Pixel Data (7FE0,0010)</emphasis></emphasis><emphasis role="bold"><emphasis role="strikethrough"> is not present (its Item </emphasis></emphasis>has zero length), and</para>
                                        </listitem>
                                        <listitem>
                                            <para xml:id="para_938b0375-972a-4339-a2cd-edd559abafdb">each Frame is entirely contained within one Fragment.</para>
                                        </listitem>
                                    </itemizedlist>
                                    <note>
                                        <para xml:id="para_07527651-b7eb-4dd5-b173-800c973824e3">Unlike a Basic Offset Table, an Extended Offset Table is not permitted to be empty.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9a926428-2659-4d86-a9f4-aca255e99dd5">Extended Offset Table Lengths</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1881728e-9fb9-464a-8641-4e341e1f26b2">(7FE0,0002)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e6a3f825-93ce-4108-b84c-234f6a0790f3">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_03fcf3a6-e4b6-47d5-838a-d91672a09bca">Byte lengths of the Frames in the Sequence of Items in Encapsulated Pixel Data encoded in Pixel Data (7FE0,0010).</para>
                                    <para xml:id="para_86698ea5-5b13-492b-829a-ad33fdb552f3">See <xref linkend="sect_C.7.6.3.1.8" xrefstyle="select: label"/>.</para>
                                    <para xml:id="para_6479b234-6071-4f8a-a1ed-8a49ce0d94f8">Required if Extended Offset Table (7FE0,0001) is present<emphasis role="bold"><emphasis role="strikethrough"> and every frame is encoded as a single fragment</emphasis></emphasis>.</para>
                                    <note>
                                        <orderedlist>
                                            <listitem>
                                                <para xml:id="para_b310dea9-9fab-4c27-9584-d1a04f3fb0cc"><emphasis role="bold"><emphasis role="strikethrough">The length information is not sent when there are multiple fragments per frame, since it is necessary to assemble the frame from fragments with delimiters rather than assuming a single contiguous span of bytes.</emphasis><emphasis role="underline">This Attribute is only sent when each Frame is entirely contained within one Fragment as a single contiguous span of bytes so that it is not necessary to assemble the frame from fragments with delimiters.</emphasis></emphasis></para>
                                            </listitem>
                                            <listitem>
                                                <para><emphasis role="bold"><emphasis role="underline">The length encoded in this Attribute may be an odd number if the compressed bitstream for the frame is an odd length; i.e., it does not include any trailing padding required to make the Item an even length.</emphasis></emphasis></para>
                                            </listitem>
                                        </orderedlist>
                                    </note>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.7.6.3.1" status="5" xml:id="sect_C.7.6.3.1">
                        <title>Image Pixel Module Attribute Descriptions</title>
                        
                        
                        
                        
                        
                        
                        
                        <section label="C.7.6.3.1.8" status="6" xml:id="sect_C.7.6.3.1.8">
                            <title>Extended Offset Table</title>
                            <para xml:id="para_50de02f7-de9b-421f-8b3a-640702a43200">The Extended Offset Table (7FE0,0001) Value shall contain byte offsets to the first byte of the Item Tag of the first fragment for every frame in the Pixel Data <emphasis role="bold"><emphasis role="underline">(7FE0,0010) </emphasis></emphasis>Sequence.</para>
                            <para xml:id="para_59952002-77a5-4d29-aa55-453eb6c24377">The byte offsets are measured from the first byte of the first Item Tag following the empty (zero length) Basic Offset Table item, i.e., the Item Tag of the first fragment of the first frame.</para>
                    <para><emphasis role="bold"><emphasis role="underline">If present, the first entry will always be 00000000H.</emphasis></emphasis></para>
                            <note>
                                <para><emphasis role="bold"><emphasis role="underline">The offset table is only present when the Pixel Data (7FE0,0010) is encoded in Encapsulated Format, since the position of uncompressed frames encoded in Native Format can be computed from their known fixed length (i.e., derived from Rows (0028,0010), Columns (0028,0011), Samples per Pixel (0028,0002) and Bits Allocated (0028,0100)).</emphasis></emphasis></para>
                            </note>
                        </section>
                    </section>
                    
                    
                </section>
        
        
        
        
 
 
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.5 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>      
        
        
        
        
        
        <section label="A.4" status="2" xml:id="sect_A.4">
      <title>Transfer Syntaxes For Encapsulation of Encoded Pixel Data</title>
      <orderedlist>
        <listitem/>
        <listitem/>
        <listitem>
          <para xml:id="para_b5e2af39-f545-4a58-bd8a-faef00bfac1c">The encoding of the Data Elements of the Data Set shall be as follows according to their Value Representations:</para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_2f69630f-745c-44ad-8013-d58d1e40abc4">...</para>
            </listitem>
            <listitem>
              <para xml:id="para_428d2f48-a706-43e0-bb3d-2bc775291d51">For the Value Representations OB, OL, OV and OW, the encoding shall meet the following specification depending on the Data Element Tag:</para>
              <itemizedlist>
                <listitem>
                  <para xml:id="para_d11882a9-2d96-4239-8809-1d7343f9f6f6">Pixel Data (7FE0,0010) may be encapsulated or native.</para>
                  <para xml:id="para_6666f793-ab61-4d3f-8686-142a1fd52846">It shall be encapsulated if present in the top-level Data Set (i.e., not nested within a Sequence Data Element).</para>
                  <para xml:id="para_0775e8a7-768d-4792-830a-d498233a93ef">...</para>
                  <para xml:id="para_5755baf8-0054-4d4e-a46b-0d8c389d2cf5">If encapsulated, it has the Value Representation OB and is an octet-stream resulting from one of the encoding processes. It contains the encoded pixel data stream fragmented into one or more Item(s). This Pixel Data Stream may represent a Single or Multi-frame Image. See <xref linkend="table_A.4-1" xrefstyle="select: label"/> and <xref linkend="table_A.4-2" xrefstyle="select: label"/>.</para>
                  <itemizedlist>
                    <listitem>
                      <para xml:id="para_24cbd6af-a8d0-4517-b73f-2fc42dae65ee">...</para>
                    </listitem>
                    <listitem>
                      <para xml:id="para_468b80cc-d9c5-4d78-83d6-867548ed4e61">The first Item in the Sequence of Items before the encoded Pixel Data Stream shall be a Basic Offset Table item. The Basic Offset Table Item Value, however, is not required to be present:</para>
                      <itemizedlist>
                        <listitem>
                          <para xml:id="para_f97a5189-b573-4b93-bfd0-0f32e4b46a2e">When the Item Value is not present, the Item Length shall be zero (00000000H) (see <xref linkend="table_A.4-1" xrefstyle="select: label"/>).</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_ad21ce9b-a763-414a-8f87-752be151ef6b">When the Item Value is present, the Basic Offset Table Item Value shall contain concatenated 32-bit unsigned integer values that are byte offsets to the first byte of the Item Tag of the first fragment for each frame in the Sequence of Items. These offsets are measured from the first byte of the first Item Tag following the Basic Offset Table item (see <xref linkend="table_A.4-2" xrefstyle="select: label"/>).</para>
                          <note>
                            <orderedlist numeration="arabic">
                              <listitem>
                                <para xml:id="para_b12823cc-c49d-4ca1-b7a2-301cfc093866">For a Multi-Frame Image containing only one frame or a Single Frame Image, the Basic Offset Table Item Value may be present or not. If present it will contain a single 00000000H value.</para>
                              </listitem>
                              <listitem>
                                <para xml:id="para_98b02555-b62d-4cc3-b2e4-37b1d94e2733">Decoders of encapsulated pixel data, whether Single Frame or Multi-Frame, need to accept both an empty Basic Offset Table (zero length) and a Basic Offset Table filled with 32 bit offset values.</para>
                              </listitem>
                                <listitem>
                                                                <para xml:id="para_c09259a9-9178-46b0-81d7-a6cf0811f3af">A Basic Offset Table Item Value is not permitted (i.e., the Item Length of the first Item will be zero) if Extended Offset Table (7FE0,0001) is present.</para>
                                                            </listitem>
                                                            <listitem>
                                                                <para><emphasis role="bold"><emphasis role="underline">If the Basic Offset Table Item Value is not zero length, the first entry will always be 00000000H.</emphasis></emphasis></para>
                                                            </listitem>
                            </orderedlist>
                          </note>
                        </listitem>
                      </itemizedlist>
                    </listitem>
                    <listitem>
                      <para xml:id="para_07969f96-31db-4ac2-8aa0-80fa310f4a9a">...</para>
                    </listitem>
                  </itemizedlist>
                </listitem>
              </itemizedlist>
            </listitem>
          </itemizedlist>
        </listitem>
      </orderedlist>
      <note/>
      <table frame="box" label="A.4-2" rules="all" xml:id="table_A.4-2">
        <caption>Examples of Elements for an Encoded Two-Frame Image Defined as a Sequence of Three Fragments with Basic Table Item Values</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c54c808b-359f-4855-b184-2264909d9c87">
                <emphasis role="bold">Pixel Data Element Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="2" rowspan="1">
              <para xml:id="para_0aaef2b0-a6a5-461d-833a-17768ffee597">
                <emphasis role="bold">Value Representation</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_acdcb1db-1dbf-445d-9802-3ccc586e18ad">
                <emphasis role="bold">Data Element Length</emphasis>
              </para>
            </th>
            <th align="left" colspan="6" rowspan="1">
              <para xml:id="para_e793274d-64c7-4369-99cb-805c3b980770">
                <emphasis role="bold">Data Element</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="3">
              <para xml:id="para_6c37e67d-86a1-4dec-a100-673fb09fe093">(7FE0, 0010) with VR of OB</para>
            </td>
            <td align="left" colspan="1" rowspan="3">
              <para xml:id="para_82a2e336-57b2-4d7b-9153-aa0615b30f93">OB</para>
            </td>
            <td align="left" colspan="1" rowspan="3">
              <para xml:id="para_9c9bc002-5f25-4582-9d4f-5144202c7162">0000H Reserved</para>
            </td>
            <td align="left" colspan="1" rowspan="3">
              <para xml:id="para_23d99773-ccfb-4b59-be4e-c1097c7babcd">FFFF FFFFH Undefined Length</para>
            </td>
            <th align="left" colspan="3" rowspan="1">
              <para xml:id="para_0fc6811c-ba79-42c0-acda-ebbbf108b576">
<emphasis role="bold">Basic Offset Table with Item Value</emphasis>
</para>
            </th>
            <th align="left" colspan="3" rowspan="1">
              <para xml:id="para_6057f99f-9063-4669-b379-4cdee43e0c44">
<emphasis role="bold">First Fragment (Frame 1) of Pixel Data</emphasis>
</para>
            </th>
          </tr>
          <tr valign="top">
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ca5745c-4ccf-4a6d-9684-609ca4a5509f">
<emphasis role="bold">Item Tag</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_4df20b72-8786-46e5-832b-910c98ae15c7">
<emphasis role="bold">Item Length</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_8442bf38-df8d-4622-a9a3-1220b376de0f">
<emphasis role="bold">Item Value</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1351808-a6eb-4b3f-9240-af496370c6eb">
<emphasis role="bold">Item Tag</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ccaa0d9-fdbd-462a-8118-ae2ad1fd66ae">
<emphasis role="bold">Item Length</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_38ebb5d7-3a4d-4041-a2b0-1cec1d316c0e">
<emphasis role="bold">Item Value</emphasis>
</para>
            </th>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6cc8755-4a94-4a48-9fe3-e4d23d0c211c">(FFFE, E000)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1692cbf-3159-4df9-a05b-fc33b0af50b1">0000 0008H</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a5a8b6c5-613e-42c1-b5b0-435e08c67db4">0000 0000H 0000 0646H</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_356bd571-7c4a-4b97-a5fd-1a4c6a4c7269">(FFFE, E000)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e6ac3fcd-6b21-47e4-be8e-5120b3838c40">0000 02C8H</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c8bb8c6-61b8-44b4-9cf8-a6c76dca1cc7">Compressed Fragment</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a32d1401-e58b-44a4-981d-95b2a34afd2e">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3423fc69-520b-4c1c-a547-35acbafe5d35">2 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_37c07eb3-d8a5-453d-ade4-f588c52168c0">2 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47919f02-49f6-48de-a421-344dab689e69">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36e2a063-7867-472f-9478-1b31f572fe9f">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_56d80a46-5e00-40a3-93e5-6a81e85789d2">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e189864-d82f-4fe4-be70-de4d8db76f6e">0008H bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_96c9cb43-7c87-4f3c-a5c2-ed201f5d5e55">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7440db0c-7978-4574-af18-74abc5c7ef24">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d7cb1dd-04d4-43d8-81e2-3525bb6ff96f">02C8H bytes</para>
            </td>
          </tr>
        </tbody>
      </table>
      <table frame="box" label="A.4-2b" rules="all" xml:id="table_A.4-2b">
        <caption>Examples of Elements for an Encoded Two-Frame Image Defined as a Sequence of Three Fragments with Basic Table Item Values (continued)</caption>
        <thead>
          <tr valign="top">
            <th align="left" colspan="8" rowspan="1">
              <para xml:id="para_1508d17d-f8eb-499a-b15d-3bfaced8828c">
                <emphasis role="bold">Data Element Continued</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <th align="left" colspan="3" rowspan="1">
              <para xml:id="para_7cf01bcb-9a06-4af7-af92-f7c74b3ebb4d">
<emphasis role="bold">Second Fragment (Frame 1) of Pixel Data</emphasis>
</para>
            </th>
            <th align="left" colspan="3" rowspan="1">
              <para xml:id="para_a48ee7a8-c9de-4518-af09-a2aee2c19b04">
<emphasis role="bold">Third Fragment (Frame 2) of Pixel Data</emphasis>
</para>
            </th>
            <th align="left" colspan="2" rowspan="1">
              <para xml:id="para_009ffbb4-a828-430f-85ec-d4ba8fdab8e4">
<emphasis role="bold">Sequence Delimiter Item</emphasis>
</para>
            </th>
          </tr>
          <tr valign="top">
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_10223bb7-698b-4c1a-ac14-986bb7ed7910">
<emphasis role="bold">Item Tag</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9c24cec-9717-446b-89d5-7d3ebb720dc6">
<emphasis role="bold">Item Length</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_94c93e32-2270-4a24-995c-a814b24ea0d8">
<emphasis role="bold">Item Value</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_a3582958-f850-45fa-a241-a2f3110dc8e1">
<emphasis role="bold">Item Tag</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f792e9f-f4de-4f45-9b12-94c56db1a9fd">
<emphasis role="bold">Item Length</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_4742f543-9b73-4797-90be-fbdbdf3538f1">
<emphasis role="bold">Item Value</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0594d2c-fb17-427b-8b3f-abc33ee1b58b">
<emphasis role="bold">Sequence Delimiter Tag</emphasis>
</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_3badc9eb-2dd6-49b3-af64-d3c1578a9ca1">
<emphasis role="bold">Item Length</emphasis>
</para>
            </th>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a1c56e0-7554-4e12-8313-99a2dd67218c">(FFFE, E000)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21b84c24-1d74-4264-bf35-f37e0d90052a">0000 036EH</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_14816b49-d5e7-4e88-8f47-55e024c3c6b7">Compressed Fragment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2156493-9dbd-425f-8415-0c355624be0f">(FFFE, E000)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b18b968-0cb5-4413-85ba-53c5d4e88cbb">0000 0BC8H</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2f46145-29ef-4d47-83ff-f1d1ed0dd3cb">Compressed Fragment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cbe9e0c9-4d2c-45e4-b9c3-88f75b9590c5">(FFFE, E0DD)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4fec14c1-3325-48f2-8e46-fd815e4914cc">0000 0000H</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5c6fd7e-f4d6-4a80-aab6-15a2b1ef51ee">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cb0e42f-1799-4a91-a00d-99dcdae89338">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_80194a9a-267a-419a-a749-cfc9da1893cb">036EH bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83126233-fe2a-42c4-993d-294b305e2bb3">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64077550-a2cd-4456-a90f-965a7bc1341c">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2c75c9f-6e5c-418c-ba3f-491601369baf">0BC8H bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_350dbf5b-0a48-4e21-9822-426684e0aeee">4 bytes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d384a1a9-55b9-4abf-b925-6a440b0c8d52">4 bytes</para>
            </td>
          </tr>
        </tbody>
      </table>    </section>
        
                    
            
        
        
        
        

        
        
        

        
        
        
        
        
    </chapter>