<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2335</title>
    <subtitle>Clinical Trial Identifiers need Issuers and may be multiple</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Jan 2024 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2023/11/16</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2023/08/28</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2335</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Clinical Trial Identifiers need Issuers and may be multiple</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6, PS3.15</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>Identifiers need issuers in order to define their scope.</para>
                    <para>Multiple identifiers may be needed the same entity, esp. the protocol, à la Other Patient IDs Sequence. A specific use case is the need to identify images from a specific trial or experiment by its well-known research collection ID as used in a shared data repository, as well as the <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="https://www.doi.org/">Digital Object Identifier (DOI)</link> for that collection.]</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>

        <section label="C.7.1.3" status="4" xml:id="sect_C.7.1.3">
            <title>Clinical Trial Subject Module</title>
            <table frame="box" label="C.7-2b" rules="all" xml:id="table_C.7-2b">
                <caption>Clinical Trial Subject Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f8f97ad5-c431-4fbf-804d-5dd6d11553d3">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_adb90c3f-5b1d-4deb-8f82-91c455994046">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7d27c212-3bee-40b1-9d75-15b75642575d">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_53c7a237-5999-43b3-b273-bfd074ec9f16">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c800b736-ce36-4484-bff1-666f74ad7532">Clinical Trial Sponsor Name</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7df8621b-147a-4fdf-a9cd-a534de3a5d25">(0012,0010)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7ec362ce-6c2f-42d4-8c47-c71b1b4839a8">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_97332098-9e7b-43bf-a273-6bfe7d3a8bc2">The name of the clinical trial or research sponsor. See <xref linkend="sect_C.7.1.3.1.1" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3a812bfe-b4fc-4313-9ff9-e067e60fe0df">Clinical Trial Protocol ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fa084632-668d-410b-a032-a2598dea4e9b">(0012,0020)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42e15f44-9483-43f4-b4b7-42d1e4bbfef2">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_268edb27-4ff5-4082-89ef-d4ac74319c5c">Identifier for the <emphasis role="bold"><emphasis role="strikethrough">noted </emphasis></emphasis>protocol. See <xref linkend="sect_C.7.1.3.1.2" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Protocol ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee1)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Protocol ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">Other Clinical Trial Protocol IDs Sequence</emphasis></emphasis></para></td>
                        <td align="center" colspan="1"><para><emphasis role="bold"><emphasis role="underline">(gggg,eee5)</emphasis></emphasis></para></td>
                        <td align="center" colspan="1"><para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para></td>
                        <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">Identification numbers or codes used to identify the protocol.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">One or more Items are permitted in this Sequence.</emphasis></emphasis></para></td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;Clinical Trial Protocol ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(0012,0020)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier for the protocol. See <xref linkend="sect_C.7.1.3.1.2" xrefstyle="select: label"/>.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee1)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Protocol ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6e26f1c0-ac00-45a5-98e6-7af33be5eee5">Clinical Trial Protocol Name</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6f0c1984-0b23-45b9-ab4e-6ee8540dd9c1">(0012,0021)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a14fbcd6-3b9d-41f1-b337-aaf41bad1dce">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3c25a265-48db-4507-8c0e-64470a99f7e6">The name of the clinical trial or research protocol. See <xref linkend="sect_C.7.1.3.1.3" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1eb28cf0-cce0-4e87-8bee-8069a9a07463">Clinical Trial Site ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0535bea4-657c-41fc-a9ca-96e5903b4281">(0012,0030)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f7ebe5e6-eb65-461b-88a5-87eb80e11bd7">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f8960118-70b3-4d01-b13d-513a7f7ce059">The identifier of the site responsible for submitting clinical trial or research data. See <xref linkend="sect_C.7.1.3.1.4" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Site ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee2)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Site ID.</emphasis></emphasis></para>
                        </td>
                    </tr>                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bd803aa7-0e0a-4bc4-a996-a4ebc44dd47b">Clinical Trial Site Name</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4539fbc2-decf-41fc-bc5b-507b82955e49">(0012,0031)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bb04fd02-1382-42c8-b83b-935030dab363">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_76680080-332f-4ff8-bbd6-d56a86ea178c">Name of the site responsible for submitting clinical trial or research data. See <xref linkend="sect_C.7.1.3.1.5" xrefstyle="select: label"/>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5d5c6749-0b65-495f-9955-67ac4cc8465b">Clinical Trial Subject ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_60c6da55-6761-4637-abf5-fa5edad3a76e">(0012,0040)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_10a2dabb-2970-42f5-b3be-ed1338b5a689">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cd98c453-d9d6-44fe-a5f8-72f02a835d58">The assigned identifier for the clinical trial or research subject. See <xref linkend="sect_C.7.1.3.1.6" xrefstyle="select: label"/>. Shall be present if Clinical Trial Subject Reading ID (0012,0042) is absent. May be present otherwise.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Subject ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee3)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Subject ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_86b0a59f-5c2c-491f-ad29-f354d14a86ae">Clinical Trial Subject Reading ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6bee4677-baa2-4031-9b91-2f93cf337cb8">(0012,0042)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_94edfa83-8ad2-428c-a9a9-3137d14770de">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_760de139-c125-4cd3-b643-938d7a1536fd">Identifies the subject for blinded evaluations. Shall be present if Clinical Trial Subject ID (0012,0040) is absent. May be present otherwise. See <xref linkend="sect_C.7.1.3.1.7" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Subject Reading ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee4)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Subject Reading ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c0fd6a2e-2b9c-4bc7-a90c-b34d6c7619e0">Clinical Trial Protocol Ethics Committee Name</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_daf4f512-4aa6-46ac-817d-1bfbeff008d0">(0012,0081)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_10ddccea-6e2d-414d-9787-8fc8b9d11fd0">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1871ce69-8a4c-4ae8-b33e-97a9ad734bb3">Name of the Ethics Committee or Institutional Review Board (IRB) or Institutional Animal Care and Use Committees (IACUC) responsible for approval of the Clinical Trial or research. Required if Clinical Trial Protocol Ethics Committee Approval Number (0012,0082) is present.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9e3fca4a-369d-4574-8cde-44f625b337de">Clinical Trial Protocol Ethics Committee Approval Number</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_51e2da3a-c655-4b8c-99e2-6f77984d8fd5">(0012,0082)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_088b6ad9-ed03-4f83-adfa-99a180c28977">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_91c118cd-4466-4a9b-b1e3-1a2ae1c56235">Approval number issued by committee described in Clinical Trial Protocol Ethics Committee Name (0012,0081).</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.7.1.3.1" status="5" xml:id="sect_C.7.1.3.1">
                <title>Clinical Trial Subject Module Attribute Descriptions</title>
                <para xml:id="para_7acfe0c6-679e-4ce6-8288-dbb6c3e07b43">Identification of subjects in clinical trials or research generally requires a combination of the following four Attributes:</para>
                <orderedlist numeration="arabic">
                    <listitem>
                        <para xml:id="para_6b11fe8b-d78b-4b7d-8346-ab93e7d1504a">Clinical Trial Sponsor Name (0012,0010), </para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_c6e25e70-4a1e-41b3-8407-5de0893de205">Clinical Trial Protocol ID (0012,0020), </para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_5fafd42f-922b-4f41-9619-83c97eeba9c5">Clinical Trial Subject ID (0012,0040) (or Clinical Trial Subject Reading ID (0012,0042) for blinded evaluations), and</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_5f0f0451-c25a-4da7-bcd0-f3894fdf6111">Clinical Trial Site ID (0012,0030).</para>
                    </listitem>
                </orderedlist>
                <para xml:id="para_1449d73c-2593-426d-9835-14c7f3666b03">For trials in which subject identifiers are unique within the scope of the Clinical Trial Protocol (e.g., if subject identifiers are centrally assigned or contain the site identifier) the Clinical Trial Site ID (0012,0030) is not required to identify subjects.</para>
                <section label="C.7.1.3.1.1" status="6" xml:id="sect_C.7.1.3.1.1">
                    <title>Clinical Trial Sponsor Name</title>
                    <para xml:id="para_38fb18cb-ad31-40da-9a14-ddc6b2d21f6a">The Clinical Trial Sponsor Name (0012,0010) identifies the entity responsible for conducting the clinical trial or research and for defining the Clinical Trial Protocol ID (0012,0020).</para>
                </section>
                <section label="C.7.1.3.1.2" status="6" xml:id="sect_C.7.1.3.1.2">
                    <title>Clinical Trial Protocol ID</title>
                    <para xml:id="para_182d9c41-afb8-42d5-9e3e-14e7db9b9843">The Clinical Trial Protocol ID (0012,0020) is the number or character sequence used by the Clinical Trial Sponsor to uniquely identify the investigational protocol in which the subject has been enrolled.</para>
                    <note>
                        <para><emphasis role="bold"><emphasis role="underline">If more than one identifier is needed to identify the protocol, one may be conveyed Clinical Trial Protocol ID (0012,0020) in the top level dataset and the others included in Other Clinical Trial Protocol IDs Sequence (gggg,eee5), and the source of each distinguished by their Issuer of Clinical Trial Protocol ID (gggg,eee1).</emphasis></emphasis></para>
                        <para><emphasis role="bold"><emphasis role="underline">Here is an example of identifying a completed trial whose data has been shared and assigned a digital object identifier:</emphasis></emphasis></para>
                        <itemizedlist mark="none">
                            <listitem><para><emphasis role="bold"><emphasis role="underline">Clinical Trial Protocol ID (0012,0020) = "TCGA-GBM"</emphasis></emphasis></para></listitem>
                            <listitem><para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Protocol ID (gggg,eee1) = "NCI"</emphasis></emphasis></para></listitem>
                            <listitem>
                                <para><emphasis role="bold"><emphasis role="underline">Other Clinical Trial Protocol IDs Sequence (gggg,eee5)</emphasis></emphasis></para>
                                <itemizedlist mark="none">
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Clinical Trial Protocol ID (0012,0020) = "doi:10.7937/K9/TCIA.2016.RNYFUYE9"</emphasis></emphasis></para></listitem>
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID (gggg,eee1) = "DOI"</emphasis></emphasis></para></listitem>
                                </itemizedlist>
                            </listitem>
                        </itemizedlist>
                        <para><emphasis role="bold"><emphasis role="underline">Here is an example of identifying a clinical trial that is potentially ongoing and not yet published or shared, and has multiple (primary and secondary) identifiers from the same issuer (which are enumerated but not otherwise distinguished from each other):</emphasis></emphasis></para>
                        <itemizedlist mark="none">
                            <listitem><para><emphasis role="bold"><emphasis role="underline">Clinical Trial Protocol ID (0012,0020) = "D6940C00002"</emphasis></emphasis></para></listitem>
                            <listitem><para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Protocol ID (gggg,eee1) = "NCI"</emphasis></emphasis></para></listitem>
                            <listitem>
                                <para><emphasis role="bold"><emphasis role="underline">Other Clinical Trial Protocol IDs Sequence (gggg,eee5)</emphasis></emphasis></para>
                                <itemizedlist mark="none">
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Clinical Trial Protocol ID (0012,0020) = "NCI-2018-00805"</emphasis></emphasis></para></listitem>
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID (gggg,eee1) = "NCI"</emphasis></emphasis></para></listitem>
                                </itemizedlist>
                                <itemizedlist mark="none">
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Clinical Trial Protocol ID (0012,0020) = "135803"</emphasis></emphasis></para></listitem>
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID (gggg,eee1) = "NCI"</emphasis></emphasis></para></listitem>
                                </itemizedlist>
                                <itemizedlist mark="none">
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Clinical Trial Protocol ID (0012,0020) = "2017-002451-28"</emphasis></emphasis></para></listitem>
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID (gggg,eee1) = "NCI"</emphasis></emphasis></para></listitem>
                                </itemizedlist>
                                <itemizedlist mark="none">
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Clinical Trial Protocol ID (0012,0020) = "NCT03423628"</emphasis></emphasis></para></listitem>
                                    <listitem><para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID (gggg,eee1) = "ClinicalTrials.gov"</emphasis></emphasis></para></listitem>
                                </itemizedlist>
                            </listitem>
                        </itemizedlist>
                    </note>
                </section>
                <section label="C.7.1.3.1.3" status="6" xml:id="sect_C.7.1.3.1.3">
                    <title>Clinical Trial Protocol Name</title>
                    <para xml:id="para_f96b04c5-834e-483d-9508-8f7af24c81f2">The Clinical Trial Protocol Name (0012,0021) contains the title of the investigational protocol in which the subject has been enrolled.</para>
                    <note>
                        <para xml:id="para_238c86a6-3af4-40cc-9f45-2d5339ffc848">It is recommended that the phase of the clinical trial or research be noted in the Clinical Trial Protocol Name, if applicable.</para>
                    </note>
                </section>
                <section label="C.7.1.3.1.4" status="6" xml:id="sect_C.7.1.3.1.4">
                    <title>Clinical Trial Site ID</title>
                    <para xml:id="para_93ce778b-e020-4fca-b7d1-35a7c85662be">The Clinical Trial Site ID (0012,0030) is the identification number or character string (issued by the entity identified by the Clinical Trial Sponsor Name (0012,0010)) used to identify the site responsible for submitting clinical trial or research data.</para>
                </section>
                <section label="C.7.1.3.1.5" status="6" xml:id="sect_C.7.1.3.1.5">
                    <title>Clinical Trial Site Name</title>
                    <para xml:id="para_33f511c7-1d7f-4140-8b38-3ffe8e8af435">The Clinical Trial Site Name (0012,0031) is a character string used to identify the site responsible for submitting clinical trial or research data.</para>
                </section>
                <section label="C.7.1.3.1.6" status="6" xml:id="sect_C.7.1.3.1.6">
                    <title>Clinical Trial Subject ID</title>
                    <para xml:id="para_ae810159-7ede-4ccb-9b6d-04ed241d721a">The Clinical Trial Subject ID (0012,0040) identifies the subject within the investigational protocol specified by Clinical Trial Protocol ID (0012,0020).</para>
                    <note>
                        <para xml:id="para_0ae33980-85da-45fd-b8ba-74ce58b410d2">The Clinical Trial Subject ID (0012,0040) may, but is not required to be, the same as Patient ID (0010,0020).</para>
                    </note>
                </section>
                <section label="C.7.1.3.1.7" status="6" xml:id="sect_C.7.1.3.1.7">
                    <title>Clinical Trial Subject Reading ID</title>
                    <para xml:id="para_f3b02334-545c-48e3-8988-ea4eac5d0545">The Clinical Trial Subject Reading ID (0012,0042) identifies the subject in the context of blinded evaluations.</para>
                </section>
            </section>
        </section>
        
        <section label="C.7.2.3" status="4" xml:id="sect_C.7.2.3">
            <title>Clinical Trial Study Module</title>
            <table frame="box" label="C.7-4b" rules="all" xml:id="table_C.7-4b">
                <caption>Clinical Trial Study Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_038f2710-65b5-45f1-b2d8-57ce788e8157">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f141f9ed-e583-498b-b9dd-044092792e28">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_02d777ac-e532-4f1f-80d7-dea01d049359">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c6956c31-d9eb-46e7-9b68-d9cb30c6f1a5">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4d2ee376-92b4-4f23-9560-4fd0f5af77ad">Clinical Trial Time Point ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a2274dba-a374-4711-8bab-51b459c6fa6e">(0012,0050)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_98e33c92-d06d-4aef-b0cd-eb018c86a4e1">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_79ac94a9-c09d-4fbd-979e-f58ec6940298">An identifier specifying the one or more Studies that are grouped together as a clinical time point or submission in a clinical trial or research. See <xref linkend="sect_C.7.2.3.1.1" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Time Point ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee5)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Time Point ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ad358363-68c7-45af-a2d0-7041aa1a60fc">Clinical Trial Time Point Description</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_220e62b8-b898-4556-ba0a-b2f2060129ca">(0012,0051)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a378088b-4647-42e9-92c3-8ec76953ca09">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6d46fbc1-b662-4a10-9a16-019e8e6d310c">A description of a set of one or more Studies that are grouped together to represent a clinical time point or submission in a clinical trial or research. See <xref linkend="sect_C.7.2.3.1.1" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7b329e45-8970-45fe-b9fb-03409e9214c2">Clinical Trial Time Point Type Code Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_da483f57-963d-4afc-8439-a70fdebf66b8">(0012,0054)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_31d61324-69dc-442d-a997-fe5ce7faf015">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8a715aa7-ec6d-44e9-a6e1-0b673d0e6537">A pre-defined type of a set of one or more Studies that are grouped together to represent a clinical time point or submission in a clinical trial or research. See <xref linkend="sect_C.7.2.3.1.1" xrefstyle="select: label"/>.</para>
                            <para xml:id="para_ec214fa8-50d3-4d5d-bacb-158e825262dc">One or more Items are permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_d952333f-7463-4d9f-b78c-4ee4cbc4dc4e">
                                <emphasis role="italic">&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_065469cb-ce67-4c8e-a915-d5539a34f999">
                                <emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_6146" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_48722df2-47cd-4223-937e-61f12abd8c15">Longitudinal Temporal Offset from Event</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_74af4d43-a8e1-4e1b-aea3-411a1ce16d1e">(0012,0052)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_de98defe-e023-4d2b-9c28-66f077ac1e5f">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7486c5a3-8464-472d-9a02-1aa182afa461">An offset in days from a particular event of significance. May be fractional. In the context of a clinical trial, this is often the days since enrollment, or the baseline imaging Study.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7a52e9c9-8167-40a2-aa0d-9ff6c04a960b">Longitudinal Temporal Event Type</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d053fa97-dccc-40af-ad0e-9b4a803d56e7">(0012,0053)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ba92078b-e9cf-44df-a46f-6c1d7e650bac">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1b1f6023-61b5-41c0-9af5-c37be13638a5">The type of event to which Longitudinal Temporal Offset from Event (0012,0052) is relative.</para>
                            <variablelist spacing="compact">
                                <title>Defined Terms:</title>
                                <varlistentry>
                                    <term>ENROLLMENT</term>
                                    <listitem>
                                        <para xml:id="para_044faf78-3226-4566-84f9-4524f4c6537c">Relative to enrollment of the subject in the research activity or clinical trial.</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>BASELINE</term>
                                    <listitem>
                                        <para xml:id="para_a2506d6c-47df-49ac-84d1-6235eb9c6c67">Relative to the baseline imaging Study.</para>
                                    </listitem>
                                </varlistentry>
                            </variablelist>
                            <para xml:id="para_8a2623f6-53ec-4da3-a8f0-62a136f78f11">Required if Longitudinal Temporal Offset from Event (0012,0052) is present.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_00f184f9-964a-48e8-a642-89d9e26c9225">Consent for Clinical Trial Use Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5db2e521-cb5b-4379-8a97-65636c8ac8db">(0012,0083)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c7d50c71-d3f8-4da7-926c-c72001735b5c">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_67d5ecb8-0767-4867-bda3-04518dbf2aec">A Sequence that conveys information about consent for Clinical Trial or research use of the Composite Instances within this Study.</para>
                            <para xml:id="para_654f352b-0275-48ab-b564-db0f30186a39">One or more Items are permitted in this Sequence.</para>
                            <para xml:id="para_44735c0a-0bd5-4356-a49c-3dd0f718fd7e">See <xref linkend="sect_C.7.2.3.1.2" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_50e5d02d-9568-4772-aa6c-86d9733c3977">&gt;Distribution Type</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_143607aa-afb6-4337-befa-b47dea29fbd0">(0012,0084)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8f741182-3637-4359-9769-497856e68612">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f1d1b98e-68a1-45b9-b08f-3f3e9972c8c4">The type of distribution for which consent to distribute has been granted.</para>
                            <variablelist spacing="compact">
                                <title>Defined Terms:</title>
                                <varlistentry>
                                    <term>NAMED_PROTOCOL</term>
                                    <listitem>
                                        <para xml:id="para_b26fa723-e8dc-44af-a9f6-dbd6bb9d4604"/>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>RESTRICTED_REUSE</term>
                                    <listitem>
                                        <para xml:id="para_55b0d401-e6bb-4dfb-ab27-1c9eca9ceb6f"/>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>PUBLIC_RELEASE</term>
                                    <listitem>
                                        <para xml:id="para_7d63d3da-01f6-4aa6-b2c8-efad114660c6"/>
                                    </listitem>
                                </varlistentry>
                            </variablelist>
                            <para xml:id="para_2d04f3f6-1fb1-46b0-b9c2-836144a5893d">See <xref linkend="sect_C.7.2.3.1.2" xrefstyle="select: label"/>.</para>
                            <para xml:id="para_84b4b829-536c-474e-8fa1-42964fb052f8">Required if Consent for Distribution Flag (0012,0085) equals YES or WITHDRAWN.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_684e49cc-38d9-48f6-afeb-cb4a29fd64b2">&gt;Clinical Trial Protocol ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_be162d00-2e5e-4cd6-862e-244fa709b253">(0012,0020)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d2efd3cf-db32-45aa-a275-161c070bf756">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7bdd4f7a-69f4-4a06-b0a2-10ea60f30cf6">The identifier of the protocol for which consent to distribute has been granted.</para>
                            <para xml:id="para_409c4578-cbff-450e-a370-af65a2525e9b">Required if Distribution Type (0012,0084) is NAMED_PROTOCOL and the protocol is not that which is specified in Clinical Trial Protocol ID (0012,0020) in the <xref linkend="sect_C.7.1.3" xrefstyle="select: title"/>.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">&gt;Issuer of Clinical Trial Protocol ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee1)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Protocol ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1478931d-0620-4176-be82-7a39d58b07b3">&gt;Consent for Distribution Flag</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7cfbd01b-e594-43a0-8e43-b4e3b272004a">(0012,0085)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3efbef76-dbb2-4ec0-8d53-287500a7b5eb">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_edf76443-9fb4-4b56-8084-046954b80b7f">Whether or not consent to distribute has been granted for the purpose described in Distribution Type (0012,0084).</para>
                            <variablelist spacing="compact">
                                <title>Enumerated Values:</title>
                                <varlistentry>
                                    <term>NO</term>
                                    <listitem>
                                        <para xml:id="para_335dd902-457e-4f68-8309-aef017646e06"/>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>YES</term>
                                    <listitem>
                                        <para xml:id="para_1fb0f725-2f0f-4b8a-bb09-0d199b3b63cf"/>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>WITHDRAWN</term>
                                    <listitem>
                                        <para xml:id="para_81cac961-d020-4343-af66-d2309739b438"/>
                                    </listitem>
                                </varlistentry>
                            </variablelist>
                            <para xml:id="para_1248e8ad-77ac-481f-997d-e2f803f30fe0">See <xref linkend="sect_C.7.2.3.1.2" xrefstyle="select: label"/>.</para>
                            <note>
                                <para xml:id="para_3a13f81e-d497-421f-aebd-f103e8f2243a">Under some circumstances, consent may be withdrawn. The purpose of encoding this is to warn receiving systems that further distribution may not be appropriate, but no semantics are defined by the Standard for what action is appropriate under such circumstances, such as what to do with previously received images that had a value of YES.</para>
                            </note>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.7.2.3.1" status="5" xml:id="sect_C.7.2.3.1">
                <title>Clinical Trial Study Module Attribute Descriptions</title>
                <section label="C.7.2.3.1.1" status="6" xml:id="sect_C.7.2.3.1.1">
                    <title>Clinical Trial Time Point</title>
                    <para xml:id="para_07ece038-2c13-4cdd-a486-dbcee4941e0d">The Clinical Trial Time Point ID (0012,0050) Attribute identifies an imaging Study within the context of an investigational protocol. This Attribute is used to define a set of Studies that are grouped together as a clinical time point or data submission in a clinical trial or research. The Clinical Trial Time Point Description (0012,0051) Attribute can be used to give a description of the Clinical Trial Time Point to which the set of Studies belongs. Clinical Trial Time Point Type Code Sequence (0012,0054) can be used to specify one or more pre-defined type of time point from a standard lexicon; more than one type is permitted, e.g., a time point may be "posttreatment" as well as "unscheduled" or "nadir", etc.</para>
                </section>
                <section label="C.7.2.3.1.2" status="6" xml:id="sect_C.7.2.3.1.2">
                    <title>Consent For Clinical Trial Use Sequence</title>
                    <para xml:id="para_b9ded7aa-0942-4f46-925b-41cc359b58e0">For applications such as clinical trials or research, the distribution of Composite Instances in Studies, whether containing identifying information or partially or completely de-identified, may need to be controlled. Permission for distribution is usually granted under the control of the Patient (through informed consent), the ethics committee or institutional review board responsible for the Study, and the sponsor of the Study through contractual means. The Consent for Clinical Trial Use Sequence (0012,0083) is intended to encode the result of the consent process to allow appropriate subsequent handling of the Instances.</para>
                    <para xml:id="para_30ab96d2-6c30-4a3e-b411-003d00d18c5f">The Defined Terms for Distribution Type (0012,0084) mean that consent has been issued to distribute for the following purposes:</para>
                    <variablelist spacing="compact">
                        <title>Defined Terms:</title>
                        <varlistentry>
                            <term>NAMED_PROTOCOL</term>
                            <listitem>
                                <para xml:id="para_055b9fc8-ed19-4b74-9834-2a8657eab70f">conducting the protocol named in Clinical Trial Protocol ID (0012,0020)</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>RESTRICTED_REUSE</term>
                            <listitem>
                                <para xml:id="para_6fd9d77e-01c7-47b9-80b5-1f28a43ae42a">re-use for restricted purposes (not specified here) other than those for which the Instances were originally created</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>PUBLIC_RELEASE</term>
                            <listitem>
                                <para xml:id="para_81882410-12e1-4eff-a248-bfa114e0ea64">release to the general public for re-use without restriction</para>
                            </listitem>
                        </varlistentry>
                    </variablelist>
                    <note>
                        <orderedlist numeration="arabic">
                            <listitem>
                                <para xml:id="para_920fbb65-8270-490c-aa3f-d1c6870c5d18">There is no intent to convey in this Sequence further details of the often complex consent and approval process. Further information about the protocol and ethics committee may be found in the Clinical Trials Modules, if present.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_204df6d2-34be-4457-8671-6b01d68c2f86">There is no identification of an individual responsible for the approval or granting consent, since in the case of a clinical trial or research subject granting informed consent, the presence of this information would breach de-identification requirements.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_4250d6ad-7a96-475e-bf69-5965b81832a1">Multiple Sequence Items may be present; for example a Study may be approved for distribution for conducting multiple explicitly named protocols.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_d8d58bb7-7289-4e2a-b75a-dbedbb31a17e">Whether or not the Instances have been adequately de-identified for any particular purpose of distribution is not defined by the Attributes in the Consent for Clinical Trial Use Sequence (0012,0083). Other Attributes address this, such as Patient Identity Removed (0012,0062), De-identification Method (0012,0063), De-identification Method Code Sequence (0012,0064) and Burned In Annotation (0028,0301). See also <olink targetdoc="PS3.15" targetptr="chapter_E" xrefstyle="template:Annex %n “%t” in PS3.15"/>.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_ab03dd8c-78a7-4760-bc94-11c4809f961e">It is possible that the list of Defined Terms for Distribution Type (0012,0084) may be extended in future for other purposes, not necessarily related to the conduct of clinical trials or research.</para>
                            </listitem>
                        </orderedlist>
                    </note>
                </section>
            </section>
        </section>
        
        <section label="C.7.3.2" status="4" xml:id="sect_C.7.3.2">
            <title>Clinical Trial Series Module</title>
            <table frame="box" label="C.7-5b" rules="all" xml:id="table_C.7-5b">
                <caption>Clinical Trial Series Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6aa2d484-4dde-464a-b13f-f0dd5c3304a5">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_aba7484e-c5ea-4c3f-b0d2-9036a8e41e76">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5d407023-86da-45b4-bec5-3e5ff18e4cb0">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f4a39cbb-7f3d-45a9-bf13-eae221a1cd9d">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b1503eb1-6425-4750-b6d8-d2a58edac9a9">Clinical Trial Coordinating Center Name</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5713253e-811e-4a06-a4ee-9d4059620fae">(0012,0060)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d9629711-e67e-4c60-91f7-e19dd57f7c01">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_362f8704-889a-4549-be92-cfb8c43fcd92">The name of the institution that is responsible for coordinating the medical imaging data for the clinical trial or research. See <xref linkend="sect_C.7.3.2.1.1" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_54a8982b-91dc-41b7-828f-d90ea92f56bb">Clinical Trial Series ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ea3c85b0-1ea6-4954-9227-9c1621a6f1e5">(0012,0071)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e3dc2590-570d-468d-a0d2-e898746356d6">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bf7c8ffd-9720-4e8e-80e7-e47b209a4115">An identifier of the Series in the context of a clinical trial or research. See <xref linkend="sect_C.7.3.2.1.2" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Issuer of Clinical Trial Series ID</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">(gggg,eee6)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Identifier of the Assigning Authority that issued the Clinical Trial Series ID.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_622dfb38-9ab3-453d-a577-e488cd86095f">Clinical Trial Series Description</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e665a6a6-db2e-47fb-bd6b-816ffb384c7d">(0012,0072)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_93c43671-e80c-49f5-b2d0-30ab22a03039">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6b450412-7bd9-4971-ba9d-9fb722b0d460">A description of the Series in the context of a clinical trial or research. See <xref linkend="sect_C.7.3.2.1.2" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.7.3.2.1" status="5" xml:id="sect_C.7.3.2.1">
                <title>Clinical Trial Series Module Attribute Descriptions</title>
                <section label="C.7.3.2.1.1" status="6" xml:id="sect_C.7.3.2.1.1">
                    <title>Clinical Trial Coordinating Center Name</title>
                    <para xml:id="para_9d359345-ab0d-4416-a31c-e37797b7d931">The Clinical Trial Coordinating Center Name (0012,0060) identifies the institution responsible for coordinating the collection of images and associated data for subjects enrolled in the clinical trial or research.</para>
                </section>
                <section label="C.7.3.2.1.2" status="6" xml:id="sect_C.7.3.2.1.2">
                    <title>Clinical Trial Series Identifier and Description</title>
                    <para xml:id="para_bd0b064c-2dc1-47be-9315-62289ed7ee43">The Clinical Trial Series ID (0012,0071) and Clinical Trial Series Description (0012,0072) Attributes can be used to identify and describe a Series within the context of a clinical trial or research without requiring the replacement of the values in Series Number (0020,0011) and Series Description (0008,103E) Attributes in the <xref linkend="sect_C.7.3.1" xrefstyle="select: title"/>, whose manufacturer or user provided values may be relevant and important to retain.</para>
                </section>
            </section>
        </section>
        
        
        
        
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="10.14" status="2" xml:id="sect_10.14">
            <title>HL7v2 Hierarchic Designator Macro</title>
            <para xml:id="para_5b652149-5151-433e-9ece-20f2832246f5">
                <xref linkend="table_10-17" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_10.14" xrefstyle="select: title"/>, which identify an entity (system, organization, agency, or department) that has responsibility for managing or assigning a defined set of instance identifiers (such as placer or filler number, Patient identifiers, provider identifiers, etc.). This entity could be a particular health care application such as a registration system that assigns Patient identifiers, a governmental entity such as a licensing authority that assigns professional identifiers or drivers' license numbers, or a facility where such identifiers are assigned.</para>
            <note>
                <para xml:id="para_3cc44496-2c43-4779-8531-49b1d028a91a">This definition is identical to HL7 v2.5, Section 2.A.33, with only minor changes for editorial style.</para>
            </note>
            <para xml:id="para_424a0f37-7642-457b-84e8-e7ab7d18c2bd">These Attributes are equivalent to the components of the HL7 v2 Hierarchic Designator (HD) and Entity Identifier (EI) data types (see HL7 v2 Chapter 2.A).</para>
            <para xml:id="para_e8f5e302-29a2-457a-8b26-a78c8ce7bf6d">If both Local Namespace Entity ID (0040,0031) and Universal Entity ID (0040,0032) are present, they shall refer to the same entity.</para>
            <table frame="box" label="10-17" rules="all" xml:id="table_10-17">
                <caption>HL7v2 Hierarchic Designator Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_43a03f46-3f87-4433-adb0-fb429bcedad5">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f224d45b-762c-4fe4-be4d-7ae549f52b9a">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_31036985-3cc0-4cf5-8299-d13958d0d56f">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e93b2bb-d87d-40dd-aeb7-aa688e41925b">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_df207a7c-a83a-494e-8736-2f91144eae83">Local Namespace Entity ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4b06af43-949f-49df-bc27-f10a4cf2edd0">(0040,0031)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b4f8d988-a48b-4e93-a7e0-e6a799a3c39b">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fc53739d-da3e-43ea-9f33-7e7e58279e85">Identifies an entity within the local namespace or domain. Required if Universal Entity ID (0040,0032) is not present; may be present otherwise.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_63c3f9d2-8473-41cc-a91c-936387453e66">Universal Entity ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e7b40a42-b85c-4141-a777-69266b608dc7">(0040,0032)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_19ad403a-055e-4dad-ac27-2303684875c6">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_29156532-a15e-4770-bd8b-e5d283d20720">Universal or unique identifier for an entity. Required if Local Namespace Entity ID (0040,0031) is not present; may be present otherwise.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eb8efe83-db51-43f7-933c-5f598fbcd440">Universal Entity ID Type</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6c54ec2c-d486-4fde-9597-ed8b37168ba4">(0040,0033)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b8e4d318-76a0-44ff-a0f3-32bd8bc89c1d">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_98293def-3b64-4e5e-8619-746a7b16e9fc">Standard defining the format of the Universal Entity ID. Required if Universal Entity ID (0040,0032) is present.</para>
                            <variablelist spacing="compact">
                                <title>Enumerated Values:</title>
                                <varlistentry>
                                    <term>DNS</term>
                                    <listitem>
                                        <para xml:id="para_eb080a2f-0aa1-46ef-a8a4-eab08149bf92">An Internet dotted name. Either in ASCII or as integers</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>EUI64</term>
                                    <listitem>
                                        <para xml:id="para_306bbe4d-1dcb-48a2-98c4-6baa909894e3">An IEEE Extended Unique Identifier</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>ISO</term>
                                    <listitem>
                                        <para xml:id="para_e12d79ff-da99-4207-82c7-ac496067994f">An International Standards Organization Object Identifier</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>URI</term>
                                    <listitem>
                                        <para xml:id="para_3d4d3471-641a-4496-8ce1-353ab94fa8dc">Uniform Resource Identifier</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>UUID</term>
                                    <listitem>
                                        <para xml:id="para_b7d026fd-8ebf-44bd-ad0e-f6bdc20f0c6d">The DCE Universal Unique Identifier</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>X400</term>
                                    <listitem>
                                        <para xml:id="para_710cb5bf-1677-498d-810e-310c501cbd8d">An X.400 MHS identifier</para>
                                    </listitem>
                                </varlistentry>
                                <varlistentry>
                                    <term>X500</term>
                                    <listitem>
                                        <para xml:id="para_bdfa28dc-d3d2-4f51-9601-0774783e5677">An X.500 directory name</para>
                                    </listitem>
                                </varlistentry>
                            </variablelist>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="10.15" status="2" xml:id="sect_10.15">
            <title>Issuer of Patient ID Macro</title>
            <para xml:id="para_337a45c5-b5ea-44ed-94ac-a25bc0941823">
                <xref linkend="table_10-18" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_10.15" xrefstyle="select: title"/>, which identify the source of Patient ID (0010,0020).</para>
            <para xml:id="para_e510c361-91d9-4afd-9d17-5e8ddee567fa">These Attributes are equivalent to components of the HL7 v2 Extended Composite ID with Check Digit (CX) data type (see HL7 v2 Chapter 2.A.14), as used in the HL7 v2 PID-3 Patient Identifier List field.</para>
            <table frame="box" label="10-18" rules="all" xml:id="table_10-18">
                <caption>Issuer of Patient ID Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_070e516d-4016-43f4-890c-6c6d03228c25">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_530d51d4-3fa3-42e7-88a9-7a2cb5bf2c49">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_39c6c5de-6e39-46bf-8a42-6f106cb5d629">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f5c66541-225f-44d0-8693-5c0563dfb131">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_df31bd9e-0fd7-4e14-82f6-0fd02f5ce8ee">Issuer of Patient ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4c145a34-0fce-46c7-bd80-ae376dada851">(0010,0021)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_706d8ecb-61ca-4a56-b192-31947917fe85">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6003b085-206b-4fae-a0ae-410844ff8980">Identifier of the Assigning Authority (system, organization, agency, or department) that issued the Patient ID.</para>
                            <note>
                                <para xml:id="para_fde39d36-34f5-4459-8bca-12b577233532">Equivalent to HL7 v2 CX component 4 subcomponent 1.</para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_02ce40f4-dc63-4aca-98d9-72c5274aa6f8">Issuer of Patient ID Qualifiers Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_857be70e-4dee-485a-a372-88d48be0ed60">(0010,0024)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3fdd9a00-eb31-4db8-af70-247fbef8daa2">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_113a690d-eb00-485b-8b86-6cf24804f9bc">Attributes specifying or qualifying the identity of the issuer of the Patient ID, or scoping the Patient ID.</para>
                            <para xml:id="para_25452064-cfd1-4b3b-9dfe-8a9c48d2deab">Only a single Item is permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_beff7d9b-9e55-40ee-bb70-4ad01fad4b57">&gt;Universal Entity ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f78c0251-9a34-4bbb-8af7-21e617d40b86">(0040,0032)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8cc7b297-caf8-429f-84c9-8debeb4858ad">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c8423c11-6e7c-46ea-8bec-1f773afcaede">Universal or unique identifier for the Patient ID Assigning Authority. The authority identified by this Attribute shall be the same as that of Issuer of Patient ID (0010,0021), if present.</para>
                            <note>
                                <para xml:id="para_8933ed06-5a13-4f47-a4fd-9e2d7cc913a6">Equivalent to HL7 v2 CX component 4 subcomponent 2 (Universal ID).</para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_980f2028-ed4a-4444-a22d-6243e6f94cf9">&gt;Universal Entity ID Type</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_431ce520-fab9-48be-a014-32866ae0c6f0">(0040,0033)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_377f21a6-d67e-4b2f-a79f-910cc4885f9a">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_db7d89fc-3b4b-4944-94af-9c10f0097055">Standard defining the format of the Universal Entity ID (0040,0032). Required if Universal Entity ID (0040,0032) is present.</para>
                            <note>
                                <para xml:id="para_cdc31146-8c35-4426-8c4c-509f0256519d">Equivalent to HL7 v2 CX component 4 subcomponent 3 (Universal ID Type).</para>
                            </note>
                            <para xml:id="para_b2a63f41-04dd-4ae5-b758-1cb173e955c3">See <xref linkend="sect_10.14" xrefstyle="select: label"/> for Defined Terms.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9e153931-ca14-4333-bdb5-0b4f4798ecd3">&gt;Identifier Type Code</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_49f7b3bf-28c4-4e1d-9866-566787d80298">(0040,0035)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d59242b3-8f89-42f7-af95-4b0f7e9d8b4e">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6d982790-a827-44d5-8fb8-cb5f09898398">Type of Patient ID. Refer to HL7 v2 Table 0203 for Defined Terms.</para>
                            <note>
                                <para xml:id="para_dc68c1a1-46db-4a55-bcba-f75b9d9a8ab4">Equivalent to HL7 v2 CX component 5 (Identifier Type Code).</para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f1a4c71e-5927-4256-b2c3-cdb335cd3aa6">&gt;Assigning Facility Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c11ff19c-cff9-4e26-8c4a-7d6419b8e621">(0040,0036)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_14ef5f75-aaea-4d7d-85ef-ec3349f743e2">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d5831995-e3a4-4a21-8649-c52c44a95f5a">The place or location identifier where the identifier was first assigned to the Patient. This component is not an inherent part of the identifier but rather part of the history of the identifier.</para>
                            <para xml:id="para_9ccf0428-d166-42b5-88ee-0e52aaaffbb0">Only a single Item is permitted in this Sequence.</para>
                            <note>
                                <para xml:id="para_6a16f809-6605-41ac-8d02-b6104b0bc0ad">Equivalent to HL7 v2 CX component 6 (Assigning Facility).</para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="4" rowspan="1">
                            <para xml:id="para_59d6f4bf-a1e3-4c7a-bd58-5889b64ce38f">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_10-17" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_40a1a958-8925-424d-93b5-2d39468d9392">&gt;Assigning Jurisdiction Code Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b65f6dce-6ee2-4b79-90d3-1833c1ba16d9">(0040,0039)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0c17bdb-117f-4064-9cbd-f4e07e11679d">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a4bbfe26-afa4-4d57-abf8-04fe86756a03">The geo-political body that assigned the Patient identifier. Typically a code for a country or a state/province. Only a single Item is permitted in this Sequence.</para>
                            <note>
                                <para xml:id="para_da6e8d69-e95e-4320-b5c5-be5b4cbb811f">Equivalent to HL7 v2 CX component 9 (Assigning Jurisdiction).</para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_424cd96d-af2b-4880-a1eb-339e040082d2">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f6b16376-83d6-4352-9dbf-6d50cb3397db">
                                <emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_5001" xrefstyle="select: labelnumber quotedtitle"/> for country codes.</emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_412b1f6c-0d0b-41e7-8d1c-6713d4acea63">&gt;Assigning Agency or Department Code Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_33248bfd-fa07-44cf-ae60-7060a45f85fb">(0040,003A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_60c3fe6f-2dbb-4722-9f8b-300d001207fd">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1275e635-bf32-4c55-91fd-83047fb19d80">The agency or department that assigned the Patient identifier. Only a single Item is permitted in this Sequence.</para>
                            <note>
                                <para xml:id="para_d5f050e9-c374-464c-bbc6-c76b57d065fd">Equivalent to HL7 v2 CX component 10 (Assigning Agency or Department).</para>
                            </note>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_98185bf3-300b-4c6a-90f1-34f0044a6eb4">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cf7077aa-b349-4f66-bd52-c854f50f3181">
                                <emphasis role="italic">No Baseline CID is defined.</emphasis>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
    <section label="C.7.1.1" status="4" xml:id="sect_C.7.1.1">
                    <title>Patient Module</title>
                    <table frame="box" label="C.7-1" rules="all" xml:id="table_C.7-1">
                        <caption>Patient Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b3c32db2-8602-4023-89bb-3856b3c51c8e">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2f7d731b-1fef-4474-81fb-bd2eb47efb40">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8d036e51-8ab1-43d3-8460-77d97c7a5ff9">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_53c874ba-0179-437b-86d0-406e872b2065">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5bf1f026-8432-4ae7-b5cd-1935b608d4cd">Patient's Name</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_353153a4-93d9-455d-9cd3-4f9c0e6e5c3a">(0010,0010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e4638255-baff-4e4c-aa74-ed52d49beb0e">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_427a23ce-994c-4eb2-9cab-0c07583ec129">Patient's full name.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dc134be9-9a8a-4643-b186-a8fd7173de7c">Patient ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_daf35f34-ae5e-4ea9-986b-a438e3f19f4b">(0010,0020)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f3197a9c-24e0-47d7-b206-18c482103507">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2fded90d-b5c8-439c-8931-a1889a1cec40">Primary identifier for the Patient.</para>
                                    <note>
                                        <para xml:id="para_bd933dd8-c95a-459e-a9cd-8b9c90650099">In the case of imaging a group of small animals simultaneously, the single value of this identifier corresponds to the identification of the entire group. See also <xref linkend="sect_C.7.1.4.1.1" xrefstyle="select: label"/>.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="4" rowspan="1">
                                    <para xml:id="para_bdf9392f-867f-4cc7-8307-325b3851c98f">
                                        <emphasis role="italic">Include <xref linkend="table_10-18" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6de1e32d-7a03-4346-9fed-5d99b555ce3e">Type of Patient ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0dda4790-6ba5-429c-b890-626b06ebc582">(0010,0022)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d8a174e0-c6d6-417c-9472-442f412ce8a2">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1fa400a7-fb08-4058-9357-712ff93b76d8">The type of identifier in the Patient ID (0010,0020).</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>TEXT</term>
                                            <listitem>
                                                <para xml:id="para_b228b166-ba66-46c4-a1fa-8336bf5d49d6"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>RFID</term>
                                            <listitem>
                                                <para xml:id="para_77e9d7b7-9aca-4bd2-b94b-e18f0aa660a9"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BARCODE</term>
                                            <listitem>
                                                <para xml:id="para_7ca38259-4766-4492-ac25-8a3e39db4a47"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <note>
                                        <orderedlist>
                                            <listitem>
                                                <para xml:id="para_83b3a1d4-c0ac-4746-9c7a-8fe85d6b7290">The identifier is coded as a string regardless of the type, not as a binary value.</para>
                                            </listitem>
                                            <listitem>
                                                <para xml:id="para_bef606c8-fcd8-4652-bc60-34ba3a239757">When this Attribute has a value of BARCODE, Patient ID (0010,0020) may or may not have the same value as Barcode Value (2200,0005) in the <xref linkend="sect_C.12.1" xrefstyle="select: title"/>, if present.</para>
                                            </listitem>
                                        </orderedlist>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2932b321-8aa7-43ae-9ad0-16914c4154e3">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8a40ce47-a847-44dd-921d-cc7ca49108d3"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f076e871-98f6-4db1-bd79-bf5b38e8e38b"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5021b797-1062-4e2b-9dbc-10f60699bf21"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3d9c6702-b2a3-4ccb-b825-525d9fb3be63">Other Patient IDs Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_90f18446-ac61-4629-8027-4f7dfa75d3a4">(0010,1002)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_05783582-39eb-4a8b-9a92-c2925762702b">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1a211856-4d95-46ba-b734-d19783bb9886">A Sequence of identification numbers or codes used to identify the Patient, which may or may not be human readable, and may or may not have been obtained from an implanted or attached device such as an RFID or barcode.</para>
                                    <para xml:id="para_1f93539c-2e2c-48b5-bc5d-1b1c9b50e0d0">One or more Items are permitted in this Sequence.</para>
                                    <note>
                                        <para xml:id="para_50865e13-ea26-46c5-aab5-de51877a7263">This Attribute replaces the use of Other Patient IDs (0010,1000), which did not specify an issuer for each other identifier, and which has been retired.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bcff3555-e810-4ce9-83e0-3e1a79f7eee4">&gt;Patient ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_980328b8-5e0b-463d-b5c8-0b6225e260ef">(0010,0020)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_71b8ca46-4ab8-4fe8-be66-f6d06bd7dfb5">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3dda8310-34e9-42d6-9cc0-9b12e38e6360">An identifier for the Patient.</para>
                                    <note>
                                        <para xml:id="para_f0c2cfdd-2ad9-4f36-9f8d-a8723d825eae">In the case of imaging a group of small animals simultaneously, the single value of this identifier corresponds to the identification of the entire group. See also <xref linkend="sect_C.7.1.4.1.1" xrefstyle="select: label"/>.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="4" rowspan="1">
                                    <para xml:id="para_7871a232-3f18-47ec-8e35-8094384347ac">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_10-18" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_efa5ca6b-4547-48f9-9cff-a5ea39b59bed">&gt;Type of Patient ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6dcbc467-adea-469d-b8ef-ee5ef47d9102">(0010,0022)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_86cc97a4-cc5f-4e9e-b029-512eb5e37cab">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9ba5f030-9c13-4a79-9bf1-e58a82c3eddb">The type of identifier in the Patient ID (0010,0020) in this Item.</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>TEXT</term>
                                            <listitem>
                                                <para xml:id="para_7f0458b8-0b8b-44cd-83cd-a1d2dfecff69"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>RFID</term>
                                            <listitem>
                                                <para xml:id="para_00901c07-cc34-41ce-aa3e-351fa7622965"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BARCODE</term>
                                            <listitem>
                                                <para xml:id="para_b41beb24-e11f-4687-978b-dbdc9a4079c0"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <note>
                                        <orderedlist>
                                            <listitem>
                                                <para xml:id="para_0c5465f7-fe2f-460e-8394-64ad149bdd60">The identifier is coded as a string regardless of the type, not as a binary value.</para>
                                            </listitem>
                                            <listitem>
                                                <para xml:id="para_bd67902a-37c7-42f3-9382-26fc5f7eab1a">When this Attribute has a value of BARCODE, Patient ID (0010,0020) may or may not have the same value as Barcode Value (2200,0005) in the <xref linkend="sect_C.12.1" xrefstyle="select: title"/>, if present.</para>
                                            </listitem>
                                        </orderedlist>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d355c3bc-4aff-407f-a9b3-b0755d03702a">Other Patient Names</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2fc0273a-2954-4b89-ab3f-a0f35210947f">(0010,1001)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_aa448bf3-7bd5-466e-9582-782bc5a2d3f5">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5488fac8-94b9-4b05-ac22-16b40e88078b">Other names used to identify the Patient.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1f7608c5-0f7c-4762-9200-04a602f2cb06">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a0c22dd7-0ab0-49a0-95e2-6c0c472d780d"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_79cb0f04-62bc-4b14-aaa0-21280ba10032"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b9146e82-c7a9-44ff-9775-e0d8482195dc"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <note>
                        <para xml:id="para_6de891fc-ddce-4d41-8dcf-e2b7b1152e23">Previously, Other Patient IDs (0010,1000) was included in this table. This Attribute have been retired. See <link xl:href="http://dicom.nema.org/MEDICAL/Dicom/2017a/output/pdf/part03.pdf">PS3.3-2017a</link>.</para>
                    </note>
                </section>    
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Add new data elements to DICOM PS3.6 as follows:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="6" status="1" xml:id="chapter_6">
            <title>Registry of DICOM Data Elements</title>
            
            <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
                <caption>Registry of DICOM Data Elements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                                <emphasis role="bold">Tag</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                                <emphasis role="bold">Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                                <emphasis role="bold">Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                                <emphasis role="bold">VR</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                                <emphasis role="bold">VM</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>(gggg,eee1)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer of Clinical Trial Protocol ID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer​Of​Clinical​Tria​lProtocol​ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>LO</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>(gggg,eee2)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer of Clinical Trial Site ID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer​Of​Clinical​Trial​Site​ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>LO</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>(gggg,eee3)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer of Clinical Trial Subject ID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer​Of​Clinical​Trial​Subject​ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>LO</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>(gggg,eee4)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer of Clinical Trial Subject Reading ID</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Issuer​Of​Clinical​Trial​Subject​Reading​ID</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>LO</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>(gggg,eee5)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Other Clinical Trial Protocol IDs Sequence</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Other​Clinical​Trial​Protocol​IDs​Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>SQ</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Add new Attributes to DICOM PS3.15 de-identification profile as follows:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="E.1.1" status="3" xml:id="sect_E.1.1">
            <title>De-identifier</title>
            
        <table frame="box" label="E.1-1" rules="all" xml:id="table_E.1-1">
            <caption>Application Level Confidentiality Profile Attributes</caption>
            <thead>
                <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Attribute Name</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Tag</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Retd. (from <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>)</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">In Std. Comp. IOD (from <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Basic Prof.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. Safe Priv. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. UIDs Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. Dev. Id. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. Inst. Id. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. Pat. Chars. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. Long. Full Dates Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Rtn. Long. Modif. Dates Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Clean Desc. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Clean Struct. Cont. Opt.</emphasis>
                        </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="bold">Clean Graph. Opt.</emphasis>
                        </para>
                    </th>
                </tr>
            </thead>
            <tbody>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para>Issuer of Clinical Trial Protocol ID</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>(gggg,eee1)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para>Issuer of Clinical Trial Site ID</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>(gggg,eee2)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para>Issuer of Clinical Trial Subject ID</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>(gggg,eee3)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para>Issuer of Clinical Trial Subject Reading ID</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>(gggg,eee4)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para>Other Clinical Trial Protocol IDs Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>(gggg,eee5)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>N</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>Y</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para>X</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                    <td align="center" colspan="1" rowspan="1"/>
                </tr>
                
            </tbody>
        </table>
            
        </section>
        
    </chapter>
</book>