    <title>CP-2331</title>
    <subtitle>Clarify Segment scan order for TILED_FULL</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2024/01/14</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Chris Bridge</para><para><link
                    xl:href="mailto:CBRIDGE@mgh.harvard.edu"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2023/07/26</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2331</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Clarify Segment scan order for TILED_FULL</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>CP 1984 extended the standard with attributes in the Segmentation IOD with the intention of supporting TILED_FULL in
                        addition to the TILED_SPARSE support that had been enabled by CP 1830.</para>
                    <para>However, it failed to specify the necessary implicit ordering of Segments, when more than one Segment is present.</para>
                    <para>In this proposal, to be consistent with the precedent set for the optical path, the order of the segments is specified in terms of Segment Number (0062,0004) as defined in the Segment Sequence (0062,0002) rather than Referenced Segment Number (0062,000B) as defined in the Segment Identification Sequence (0062,000A) as would be used in the (omitted) Per-Frame Functional Group Sequence Items, except that the order is specified by numeric value since each Sequence Item is so numbered, whereas the optical paths are not defined numerically but by a textual identifier.</para>
                    <para>CP 1830 also failed to relax the mandatory requirement for the Segmentation Macro (Segment Identification Sequence) to be present in the functional groups, which describes which segment a frame represents. Relax this constraint conditioned on being TILED_FULL.</para>
                    <para>The Frame Content Macro, which in most IODs is required to be in the Per-Frame Functional Groups is in the Segmentation IOD not so constrained in the IOD definition (despite the general coment in the Macro definition itself), is made consistent, and conditions on being TILED_FULL are added to its content so that it may be rendered empty and thence omitted as in other IODs in which its contents are constrained by SOP Class.</para>
                    <para>Add the note that typically accompanies those IODs in which the Per-Frame Functional Groups Sequence can be omitted.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                  >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        <section label="A.51.5" status="3" xml:id="sect_A.51.5">
            <title>Segmentation Functional Groups</title>
            <table frame="box" label="A.51-2" rules="all" xml:id="table_A.51-2">
                <caption>Segmentation Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f4dac09b-ec34-48a8-bc04-61d6eaeccf62">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3ebeea64-2c0d-4aec-8ac4-34ca38a578f5">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_07abb63b-8340-4a86-a957-b6129678667b">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d225a1f4-1580-4be6-8b8a-b0a0d94fc407">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d718e2c3-00cc-447d-8313-f46175fe2cb6"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_edf2f027-832b-4bbb-b1d9-3a0b61c50a47"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3e0ef8b9-c64d-417e-b072-4372f1322139">Plane Position (Slide)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6b261bbd-473c-409b-9b08-573396606725">
                                <xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_20e543a1-9c89-42ad-b619-c161f3322d35">C - Required if Derivation Image Functional Group (<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>) is not present and the Frame of Reference is defined in the Slide Coordinate System and Dimension Organization Type (0020,9311) is not TILED_FULL. May be present otherwise if the Frame of Reference is defined in the Slide Coordinate System. See <xref linkend="sect_A.51.5.1" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_91c06f6e-a310-4239-b87b-152f277e8bb2">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a526bf4f-3813-48b6-b5fe-82258a8f6126"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_59530188-7dc9-4e95-a726-ec244a4ffb19"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_12f90a9f-cba9-4b47-a388-469cca7cdd07">Frame Content</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0c59c690-b0f2-4164-872c-5c273eb4521e">
                                <xref linkend="sect_C.7.6.16.2.2" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_901e8b81-50fe-4fee-b851-028ef2058a47">M<emphasis role="bold"><emphasis role="underline"> - Shall not be used as a Shared Functional Group</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cb0c73da-b198-4fa8-b66c-f49d967a5244">Segmentation</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3be6feb6-0c95-4371-817b-313342a455bc">
                                <xref linkend="sect_C.8.20.3.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ec95e942-28e9-461b-8564-0a619598f9fb"><emphasis role="bold"><emphasis role="strikethrough">M</emphasis><emphasis role="underline">C - Required if Dimension Organization Type (0020,9311) is not TILED_FULL</emphasis></emphasis></para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <note>
                <para><emphasis role="bold"><emphasis role="underline">The Plane Position (Slide), Frame Content and Segmentation Macros are Type C, which allows the Per-Frame Functional Group Sequence (5200,9230) to be entirely omitted in those cases in which there are no other Per-Frame Functional Group Macros with content (i.e., the Frame Content Macro is empty or absent).</emphasis></emphasis></para>
            </note>
            <section label="A.51.5.1" status="4" xml:id="sect_A.51.5.1">
                <title>Segmentation Functional Groups Description</title>
                <para xml:id="para_31f0915e-245f-409a-8307-e332181a2e3c">When a Frame of Reference UID is present the segment shall be specified within that coordinate system, using the Pixel Measures and either the Plane Position (Patient) and Plane Orientation (Patient), or the Plane Position (Slide) Functional Groups. Since this defines the spatial relationship of the segment, the size of the segmentation frames need not be the same size, or resolution, as the image data used to generate the segment data. The Derivation Image Functional Group may also be present, to specify on which images the segmentation was actually performed (since there may be others in the same Frame of Reference that are spatially co-located, but were not used to perform the segmentation).</para>
                <para xml:id="para_d40a5ab6-c9c4-4a90-9da0-36f91fa8f599">If the Frame of Reference UID is not present, each pixel of the segmentation shall correspond to a pixel in a referenced image, using the Derivation Image Functional Group. Hence, the rows and columns of each referenced image will match the segmentation image. If both the Frame of Reference UID and the Derivation Image Functional Group are present, the segmentation and referenced image pixels need not correspond.</para>
                <para xml:id="para_4a034de9-2d78-47e1-b66b-663baca9730d">The value of Purpose of Reference Code Sequence (0040,A170) in the <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: title"/> shall be <olink targetdoc="PS3.16" targetptr="DCM_121322">(121322, DCM, "Source Image for Image Processing Operation")</olink>. The value of Derivation Code Sequence (0008,9215) shall be <olink targetdoc="PS3.16" targetptr="DCM_113076">(113076, DCM, "Segmentation")</olink>.</para>
                <note>
                    <para xml:id="para_a53a9c5c-79b1-416e-9c6c-ea96717b9f7e">Non-image source Instances used during segmentation, such as Real World Value maps, can be described in the top level Data Set in the Source Instance Sequence (0042,0013) of the <xref linkend="sect_C.12.4" xrefstyle="select: title"/> and are implied to have been used for the derivation of all frames. I.e., there is no mechanism for selectively specifying on a per-frame basis which non-Image Instances were used. Real World Value Map Instances already contain a means of selectively applying different scale factors to different frames.</para>
                </note>
            </section>
        </section>
        
        <section label="C.7.6.17" status="4" xml:id="sect_C.7.6.17">
                    <title>Multi-frame Dimension Module</title>
                    <para xml:id="para_fd19bb43-f33a-4a72-a467-b7f31e8b2ba0">...</para>
                    <table frame="box" label="C.7.6.17-1" rules="all" xml:id="table_C.7.6.17-1">
                        <caption>Multi-frame Dimension Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_52fedc83-3eba-4eb9-b710-42a6374511bd">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bb8b8a4e-884d-4d4a-8f5b-d85cd988f904">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a52154a0-b892-41ce-8072-c409e373472c">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6042c75f-8536-43cf-ada2-e55670bf8a93">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3c9fc2fc-f7cd-4f03-8913-9309af099d34">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b142daad-8ba6-43c0-861e-dda13e684c5b"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c93cfbc9-bf03-433a-9ec6-f956e76dc3db"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_05cc2719-dd8e-4ca0-869b-6e2f74b3ddde"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2ed4e229-86f0-49eb-a975-82e909a3bda2">Dimension Organization Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_057023b5-439a-476b-b653-b891e6760ac0">(0020,9311)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_eeded002-5f74-4378-87d7-f882f59cab4c">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c658e40b-70bb-4fdd-8d65-b7c387e1b921">Dimension organization of the Instance.</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>3D</term>
                                            <listitem>
                                                <para xml:id="para_7f8d5062-be96-4036-9f05-f0bcf74caad8">Spatial Multi-frame image of equally spaced parallel planes (3D volume set)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>3D_TEMPORAL</term>
                                            <listitem>
                                                <para xml:id="para_0d8b0c6d-db9c-4f89-8e1d-234082c4c824">Temporal loop of equally spaced parallel-plane 3D volume sets.</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>TILED_FULL</term>
                                            <listitem>
                                                <para xml:id="para_bbbacde5-9240-44c7-aaae-75f23d398fe2">Tiled image in which each frame represents a single tile and the <emphasis role="bold"><emphasis role="strikethrough">spatial </emphasis></emphasis>positions of the tiles are implicitly defined as per <xref linkend="sect_C.7.6.17.3" xrefstyle="select: label"/>.</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>TILED_SPARSE</term>
                                            <listitem>
                                                <para xml:id="para_ac40ae40-8977-4c43-b515-b181c773e2a3">Tiled image in which each frame represents a single tile and the spatial positions of tiles are explicitly defined by per-frame Functional Group Macro entries.</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9a948690-a5d2-499e-ab49-3354658f32eb">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c1d622aa-159e-4b26-8b77-adffb26e52e2"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fbbb37b9-09b7-4c7a-92c0-183eb09d18be"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f2c32ac7-6e56-44f9-9190-09774137ca81"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.7.6.17.3" status="5" xml:id="sect_C.7.6.17.3">
                        <title>Spatial Location<emphasis role="strikethrough"> and</emphasis><emphasis role="underline">,</emphasis> Optical Path<emphasis role="underline">, and Segment</emphasis> of Tiled Images</title>
                        <para xml:id="para_81c46a69-b4dc-428d-822c-2efeee711150">If Dimension Organization Type (0020,9311) is present with a value of TILED_FULL, then the Per-Frame Functional Group Macros that would otherwise describe the spatial location of each tile explicitly (e.g., the X, Y and Z offsets from the origin in the Slide Coordinate System Plane Position (Slide)), <emphasis role="bold"><emphasis role="strikethrough">and</emphasis><emphasis role="underline">,</emphasis></emphasis> the optical path<emphasis role="bold"><emphasis role="underline">, or segment</emphasis></emphasis> may be omitted.</para>
                        <para xml:id="para_08fcce55-df5f-49f2-b8a2-978d0468872c">A value of TILED_FULL indicates that the frames across all Instances of a Concatenation, or a single Instance in the absence of a Concatenation, comprise a non-sparse non-overlapping representation of an entire rectangular region, and are sequentially encoded as successive frames in Pixel Data (7FE0,0010) in an implicit order varying:</para>
                        <itemizedlist>
                            <listitem>
                                <para xml:id="para_f0dd20ae-8ade-46f2-ad35-2c50add983d5">first along the row direction from left to right, where the row direction is defined in the Slide Coordinate System by the first three values of Image Orientation (Slide) (0048,0102),</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_bd60c273-e1a2-4c77-89f4-538768f79537">then along the column direction from top to bottom, where the column direction is defined in the Slide Coordinate System by the second three values of Image Orientation (Slide) (0048,0102),</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_0ae08229-6360-49ad-a748-9e855f00a350">then along the depth direction from the glass slide towards the coverslip, where the depth direction is defined in the Slide Coordinate System from zero to positive,</para>
                            </listitem>
                            <listitem>
                        <para xml:id="para_e553a9fe-2e1b-42ed-aa5f-a27f3c7c00e6">then along optical paths, <emphasis role="bold"><emphasis role="underline">if applicable, </emphasis></emphasis>where the direction is defined by successive Items of the Optical Path Sequence (0048,0105) in the order in which they are listed in that Sequence<emphasis role="bold"><emphasis role="strikethrough">.</emphasis><emphasis role="underline">,</emphasis></emphasis></para>
                    </listitem>
                    <listitem>
                        <para><emphasis role="bold"><emphasis role="underline">then along the segments, if applicable, where the direction is defined by ascending numeric values of Segment Number (0062,0004) as defined in the Segment Sequence (0062,0002).</emphasis></emphasis></para>
                    </listitem>
                        </itemizedlist>
                        <para xml:id="para_6d73468e-28c6-4aee-a708-dab0b3a14702">If Dimension Organization Type (0020,9311) is absent or has a value of TILED_SPARSE, then the <emphasis role="bold"><emphasis role="strikethrough">spatial </emphasis></emphasis>location of each tile is explicitly encoded using information in the Per-Frame Functional Group Sequence, and the recipient shall not make any assumption about the spatial position or optical path<emphasis role="bold"><emphasis role="underline"> or segment</emphasis></emphasis> or order of the encoded frames but shall rely on the values of the relevant Per-Frame Functional Group Macro.</para>
                    </section>
                </section>
        
        <section label="C.7.6.16.2.2" status="6" xml:id="sect_C.7.6.16.2.2">
                            <title>Frame Content Macro</title>
                            <para xml:id="para_2797cdd6-abe6-4e70-a1f8-c1b965189d4c">
                                <xref linkend="table_C.7.6.16-3" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.7.6.16.2.2" xrefstyle="select: title"/>, which is used as a Functional Group Macro.</para>
                            <para xml:id="para_c05df9f4-9a80-4060-95e7-99b39556fd21">This Functional Group Macro may only be part of the Per-frame Functional Groups Sequence (5200,9230) Attribute.</para>
                            <table frame="box" label="C.7.6.16-3" rules="all" xml:id="table_C.7.6.16-3">
                                <caption>Frame Content Macro Attributes</caption>
                                <thead>
                                    <tr valign="top">
                                        <th align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_524ebe21-f60a-44d2-9648-e16379da4dce">Attribute Name</para>
                                        </th>
                                        <th align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_7fd19267-d36c-4cf5-a83c-2c238303ce38">Tag</para>
                                        </th>
                                        <th align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_f38a031a-bf90-45fb-86b0-4d9633671d73">Type</para>
                                        </th>
                                        <th align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_0467d156-3cf3-43ae-a201-93f7efa5c0d2">Attribute Description</para>
                                        </th>
                                    </tr>
                                </thead>
                                <tbody>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_2cbd10d3-95f9-469c-b795-4aae3e62f9a3">Frame Content Sequence</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_11f48380-d331-45fe-9341-78ddd8617369">(0020,9111)</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_3e3dafc7-4788-43af-a588-eff012eb150d">1</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_47f02e7a-77c5-4bf3-9dd4-6176e50f2782">Identifies general characteristics of this frame.</para>
                                            <para xml:id="para_8c19b021-9574-49de-9cea-6c45d8e229c5">Only a single Item shall be included in this Sequence.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_a24dce14-f92c-4ab9-b15d-88a91fff6f08">&gt;Frame Acquisition Number</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_1430cd99-6dfd-4950-b2b2-ba93a5fa6a34">(0020,9156)</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_a98dbfa7-d0aa-4f20-b848-168fec91a874">3</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_3ba6b5df-b9f2-4670-9987-687ecfd20f96">A number identifying the single continuous gathering of data over a period of time that resulted in this frame.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_aced64ba-6bc1-4605-bdc7-0ce3db227b65">&gt;Frame Reference DateTime</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_991c9dc8-b3f0-4308-b8f5-cd04d893ed41">(0018,9151)</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_44e259c4-5874-45df-98d9-d51d12eee3b5">1C</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_b4be9e58-cb0b-41d8-ab9b-c062befbf53c">The point in time that is most representative of when data was acquired for this frame. See <xref linkend="sect_C.7.6.16.2.2.1" xrefstyle="select: label"/> and <xref linkend="sect_C.7.6.16.2.2.2" xrefstyle="select: label"/> for further explanation.</para>
                                            <note>
                                                <para xml:id="para_0595ede0-6d6b-4144-979a-4d2c52ae8262">The synchronization of this time with an external clock is specified in the <xref linkend="sect_C.7.4.2" xrefstyle="select: title"/> in Acquisition Time Synchronized (0018,1800).</para>
                                            </note>
                                            <para xml:id="para_a33f713c-0957-4d73-abd2-4b97e12e5bdc">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL, <emphasis role="bold"><emphasis role="underline">and Dimension Organization Type (0020,9311) is not TILED_FULL, </emphasis></emphasis>and the SOP Class UID is not:</para>
                                            <itemizedlist>
                                                <listitem>
                                                    <para xml:id="para_a4a91ec1-2630-4513-8f6a-dd64d82648cd">"1.2.840.10008.5.1.4.1.1.2.2 (Legacy Converted Enhanced CT Image Storage)", or</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_30fa5743-c4bc-4a14-a5f7-355cb7c5e1c6">"1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage), or</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_7c5e5260-6601-4f7e-9088-b61cbfb937ea">"1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), or</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_536ee700-c037-48b6-aa58-05c23c5d0b33">"1.2.840.10008.5.1.4.1.1.77.1.6" (VL Whole Slide Microscopy Image Storage)</para>
                                                </listitem>
                                            </itemizedlist>
                                            <para xml:id="para_6b757ba7-fcbd-4223-a5f4-afcfb8e6d778">May be present otherwise.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_1f06faaa-748c-4ac6-bfe8-1db57aac87a6">&gt;Frame Acquisition DateTime</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_7c34ae9a-49c5-4a9e-ad76-33e2cdbbaaa0">(0018,9074)</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_7462965d-82e4-490f-a0fd-353de51ee00a">1C</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_54046302-5d66-4799-8c80-ebcf361902a7">The date and time that the acquisition of data that resulted in this frame started. See <xref linkend="sect_C.7.6.16.2.2.1" xrefstyle="select: label"/> for further explanation.</para>
                                            <para xml:id="para_2865964e-056f-42ca-acf9-0db9c54872e1">Required if</para>
                                            <itemizedlist>
                                                <listitem>
                                                    <para xml:id="para_fb473e02-43d5-4085-b1d9-64455a7ae88f">Frame Type (0008,9007) Value 1 of this frame is ORIGINAL, <emphasis role="bold"><emphasis role="underline">and Dimension Organization Type (0020,9311) is not TILED_FULL, </emphasis></emphasis>and the SOP Class UID is not:</para>
                                                    <itemizedlist>
                                                        <listitem>
                                                            <para xml:id="para_6b10da1f-ca01-4937-b119-b3832b70530c">"1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage), or</para>
                                                        </listitem>
                                                        <listitem>
                                                            <para xml:id="para_423b5562-5745-4acb-a6d4-11b326738fcb">"1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage), or</para>
                                                        </listitem>
                                                        <listitem>
                                                            <para xml:id="para_e5a668dc-7480-42bc-b42c-578dfd5ef34c">"1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), or</para>
                                                        </listitem>
                                                        <listitem>
                                                            <para xml:id="para_2f6bd53d-a51d-4c09-bb3a-e4f70e0bede4">"1.2.840.10008.5.1.4.1.1.77.1.6" (VL Whole Slide Microscopy Image Storage)</para>
                                                        </listitem>
                                                    </itemizedlist>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_42802844-b2d3-4fe1-9660-eea3160c55ae">or SOP Class UID (0008,0016) is "1.2.840.10008.5.1.4.1.1.6.3" (Photoacoustic Image Storage).</para>
                                                </listitem>
                                            </itemizedlist>
                                            <para xml:id="para_9b7d1ad3-8590-4896-accc-f4f0cf56f7d3">May be present otherwise.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_4e0948b3-a4e3-4de7-8757-a9a56557051c">&gt;Frame Acquisition Duration</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_df641104-eee2-4cbe-8ab0-0fdb97459ba8">(0018,9220)</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_d8afe1e8-80ce-45ab-8efe-dfaa337aaf8e">1C</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_25eec740-e3e5-4c3d-9869-475178a0f475">The actual amount of time [in milliseconds] that was used to acquire data for this frame. See <xref linkend="sect_C.7.6.16.2.2.1" xrefstyle="select: label"/> and <xref linkend="sect_C.7.6.16.2.2.3" xrefstyle="select: label"/> for further explanation.</para>
                                            <para xml:id="para_5717e579-92ca-4a89-bc96-403ded30d10a">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL, <emphasis role="bold"><emphasis role="underline">and Dimension Organization Type (0020,9311) is not TILED_FULL, </emphasis></emphasis>and the SOP Class UID is not:</para>
                                            <itemizedlist>
                                                <listitem>
                                                    <para xml:id="para_308248d7-3098-4adf-8729-af205e44669b">"1.2.840.10008.5.1.4.1.1.2.2 (Legacy Converted Enhanced CT Image Storage)", or</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_1a86c6aa-80e3-4e7d-9aa5-027e5b6484b8">"1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage), or</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_430f64ef-1b71-4445-b429-784ceb96d620">"1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), or</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_d657df92-0905-473c-9f49-6f7c7252ad04">"1.2.840.10008.5.1.4.1.1.77.1.6" (VL Whole Slide Microscopy Image Storage)</para>
                                                </listitem>
                                            </itemizedlist>
                                            <para xml:id="para_1b9bb038-37f4-4848-9bd5-700ecd1bad63">May be present otherwise.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_16007e1d-d00a-4188-aef8-2aadea90fd8c">...</para>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_0beb9dd5-c385-4cc8-91f1-4fbfd1463019"/>
                                        </td>
                                        <td align="center" colspan="1" rowspan="1">
                                            <para xml:id="para_0f587ad1-5890-4842-a83c-4d9e0fb1674b"/>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_5754c2c7-bd57-4a58-b841-7024d2e00748"/>
                                        </td>
                                    </tr>
                                </tbody>
                            </table>
                        </section>
        
        
        
        
        
        
        
        
        

        
        
        
    </chapter>