    <title>CP-2193</title>
    <subtitle>Add Encapsulated Document to de-identification profile</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2022/04/03</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie <link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Kevin O’Donnell <link
                    xl:href="mailto:KOdonnell@MRU.MEDICAL.CANON"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2021/10/20</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2193</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add Encapsulated Document to de-identification profile</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.15</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The Encapsulated Document Attribute may contain identifying information and needs to be explicitly addressed, whether the Attribute occurs  in IODs defined for encapsulated document storage, or in Standard Extended SOP Classes.</para>
                    <para>Specify that the Encapsulated Document Attribute is to be removed unless the Clean Graphics Option is implemented, and amend the description of the option accordingly.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.15 as follows, according to this pattern:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        
        <section label="E.1" status="2" xml:id="sect_E.1">
            <title>Application Level Confidentiality Profiles</title>
            <section label="E.1.1" status="3" xml:id="sect_E.1.1">
                <title>De-identifier</title>
                <table frame="box" label="E.1-1" rules="all" xml:id="table_E.1-1">
                    <caption>Application Level Confidentiality Profile Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c5e6e8ba-34e0-4e31-90b5-bd9ed7086ddd">
                                    <emphasis role="bold">Attribute Name</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d02d5580-8438-4a00-8c30-ba93530bcd31">
                                    <emphasis role="bold">Tag</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_705936dc-2c42-4198-b406-ba0e61b35e99">
                                    <emphasis role="bold">Retd. (from <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>)</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a8387cac-386f-4ad9-96b3-b3dcb1b1010c">
                                    <emphasis role="bold">In Std. Comp. IOD (from <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fc0ff9a1-5780-4d06-861a-24ae0ba38b47">
                                    <emphasis role="bold">Basic Prof.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_19ddb67a-a5d6-474d-a13f-8df27d28b707">
                                    <emphasis role="bold">Rtn. Safe Priv. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5b335431-5276-4cc0-bddf-93b6d107226e">
                                    <emphasis role="bold">Rtn. UIDs Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8317e0cf-0080-48f9-8040-1f49a4478c0f">
                                    <emphasis role="bold">Rtn. Dev. Id. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1ba661a2-a572-420a-95c1-bc1bf694a70e">
                                    <emphasis role="bold">Rtn. Inst. Id. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_796841da-6b27-4558-99c9-99503edccc6c">
                                    <emphasis role="bold">Rtn. Pat. Chars. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_edac8181-290b-4f9c-945d-5dfc7b1a43fc">
                                    <emphasis role="bold">Rtn. Long. Full Dates Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6305752d-56e7-4f4d-b9d8-e5052fcd6eea">
                                    <emphasis role="bold">Rtn. Long. Modif. Dates Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_84f3b128-8fe5-45bc-b04e-abc483377b06">
                                    <emphasis role="bold">Clean Desc. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0adc5d55-1196-4265-8ee1-b205103c96f1">
                                    <emphasis role="bold">Clean Struct. Cont. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5e74e547-b6fd-410e-9c09-b271915c3eeb">
                                    <emphasis role="bold">Clean Graph. Opt.</emphasis>
                                </para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Encapsulated Document</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">(0042,0011)</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">N</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Y</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">D</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        </section>
        <section label="E.3" status="2" xml:id="sect_E.3">
            <title>Basic Application Level Confidentiality Options</title>

            <para xml:id="para_54f1923e-319e-4dc7-a2c6-03b9eeae9085">Various Options are defined to be applicable to the Basic Application Level Confidentiality Profile. Some of these Options require removal of additional information, and some of these Options require retention of information that would otherwise be removed.</para>
            <para xml:id="para_31e04a88-0281-4e8d-b68f-d4569f529a22">The following Options are defined that require removal of additional information:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_70734f6e-18c4-4dde-89a0-39e8ba21c2be">Clean Pixel Data Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_b0870319-f6bb-4768-b83d-85dd1fd8c76d">Clean Recognizable Visual Features Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_b2e700d9-e51f-4bcb-9c89-2a154a88b810">Clean Graphics Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_f142ae05-3b96-4a03-95b1-deee279b2c50">Clean Structured Content Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_016380f3-4324-46eb-825e-4e11e1dc1d6e">Clean Descriptors Option</para>
                </listitem>
            </itemizedlist>
            <para xml:id="para_1f84fb8b-37fa-4e00-bc12-81adad03bd8e">The following Options are defined that require retention of information that would otherwise be removed but that is needed for specific uses:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_db0e82fb-5dea-4364-90c7-01f803528585">Retain Longitudinal Temporal Information with Full Dates Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_e18f5841-373d-4392-865f-1afeabdb9374">Retain Longitudinal Temporal Information with Modified Dates Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_fa7665b7-3b0d-443e-9aea-5ee5a05eaf9c">Retain Patient Characteristics Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_3888366f-98d3-4187-8949-727a73ff7127">Retain Device Identity Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_142897c1-3a82-4edb-8952-bfdbf457190d">Retain Institution Identity Option</para>
                </listitem>
                <listitem>
                    <para xml:id="para_c33124ca-9bca-4742-9e6b-54c48907c48c">Retain UIDs</para>
                </listitem>
                <listitem>
                    <para xml:id="para_ebd54700-c7dd-41a9-8ce3-d32547c2c1a1">Retain Safe Private Option</para>
                </listitem>
            </itemizedlist>
            <section label="E.3.3" status="3" xml:id="sect_E.3.3">
                <title>Clean Graphics Option</title>

                <para xml:id="para_5b5f4d1d-5e4e-4aea-9435-246aeb7389bf">Instances of various Standard and Standard Extended SOP Classes, including Images, Presentation States<emphasis role="bold"><emphasis role="underline">, Encapsulated Documents</emphasis></emphasis> and other Composite SOP Instances, may contain identification information encoded as graphics, text annotations<emphasis role="bold"><emphasis role="underline">, encapsulated documents</emphasis></emphasis> or overlays. This does not include information contained in Structured Report SOP Classes.</para>
                <para xml:id="para_628bbc9f-6ba4-4316-bfcd-fdc7bc2d3076">When this Option is specified in addition to an Application Level Confidentiality Profile, any information encoded in graphics, text annotations<emphasis role="bold"><emphasis role="underline">, encapsulated documents</emphasis></emphasis> or overlays corresponding to the Attribute information specified to be removed by the Profile and any other Options specified shall also be removed, as described in <xref linkend="table_E.1-1" xrefstyle="select: label"/>.</para>
                <para xml:id="para_503874e4-0dbf-49d2-b74d-4fb26d630d97">This may require intervention of a human operator.</para>
                <note>
                    <orderedlist numeration="arabic">
                        <listitem>
                            <para xml:id="para_09034aea-2f35-4d77-9b96-562b7587edf5">This capability is called out as a specific Option, since it may be more practical to simply remove all such graphics, text annotations<emphasis role="bold"><emphasis role="underline">, encapsulated document</emphasis></emphasis>s or overlays (as required by the Profile without this Option).</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_26150ea2-5502-4c4f-8949-b9377b987281">As with burned-in pixel data annotation, deciding whether or not text is identifying information or some other type of information may be non-trivial. It is not required that information specified to be retained in the non-pixel data by other Options (e.g., physical characteristics, dates or descriptors) also be retained in graphics, text annotations or overlays.</para>
                        </listitem>
                    </orderedlist>
                </note>
            </section>
        </section>
    
        
        
        
        
    
        