<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2102</title>
    <subtitle>Add Whole Slide Microscopy Image Flavor of THUMBNAIL and retire LOCALIZER</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2021/07/04</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Nick Haarselhorst van den Goorbergh</para><para><link xl:href="mailto:nick.haarselhorst@philips.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/08/25</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2102</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add Whole Slide Microscopy Image Flavor of THUMBNAIL and retire LOCALIZER</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>There is a need to distinguish a THUMBNAIL image type, which is a single frame low resolution representation of the tissue content of a slide, and which may not be the same size, shape, content or Frame of Reference as a low resolution VOLUME pyramid layer, and which does not include the label as the OVERVIEW image does (per CP 1829, and there is still a need to identify the OVERVIEW image with the label).</para>
                    <para>The OVERVIEW image type, which is intended to be an overview of the whole slide, may include only part of the label, rather than the entire label, to achieve the objective of including all of the potential tissue area.</para>
                    <para>In reviewing the need for this, and its distinction from the existing LOCALIZER image type, which triggers the requirement to include the Multi-Resolution Navigation Module, which contains mandatory (non-empty sequence of items mapping frame numbers to physical position), it has been recognized that the LOCALIZER as originally conceived is not useful (replicates information that is either implicit in TILED_FULL and can be computed, or duplicates what is in Per-Frame Functional Groups Sequence of base layer) and unnecessarily increases size (and parsing time) of the image if just needed for viewing, so it is retired.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        <section label="A.32.8" status="3" xml:id="sect_A.32.8">
                <title>VL Whole Slide Microscopy Image IOD</title>
                <section label="A.32.8.3" status="4" xml:id="sect_A.32.8.3">
                    <title>VL Whole Slide Microscopy Image IOD Module Table</title>
                    <table frame="box" label="A.32.8-1" rules="all" xml:id="table_A.32.8-1">
                        <caption>VL Whole Slide Microscopy Image IOD Modules</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_eb33a04a-73c0-41fc-a4fe-7c8fb9316e64">IE</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7e383b2d-2788-4ee5-97e6-4bbf5959a5e8">Module</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0c39971c-5a9c-4692-9c43-97d2ca3b6275">Reference</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8fb856bb-dd94-4d2c-9d27-5ff66fca97fc">Usage</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="2">
                                    <para xml:id="para_320c6918-9632-4a00-ae8a-5fb4b28e4d1a">Patient</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4a504b48-eab1-4980-a5b1-14019f96c3b6">Patient</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_41913b79-c971-46d9-baaa-e77ca219a2db">
                                        <xref linkend="sect_C.7.1.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_74e501ee-5c6b-4c35-afa4-2bb1a787e3e1">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5f24067a-a59a-447f-bf04-122394e66f26">Clinical Trial Subject</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e60dd6db-f988-4b4f-be9f-57e925fada8b">
                                        <xref linkend="sect_C.7.1.3" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3534f6a1-531a-495d-b610-1d1dcc6e85e8">U</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="3">
                                    <para xml:id="para_aeb48a77-9cb3-43fe-bfd7-53fa15c55b28">Study</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3b675d7b-e7ec-448f-8279-e4a1b13bb774">General Study</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_68e8bbdb-c271-4881-9e2e-6b8c7336b0ef">
                                        <xref linkend="sect_C.7.2.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_be0da7ff-d024-4d35-b8b5-690a7122f8ed">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d4b5c158-89e5-4b7a-9d0b-c842f4228811">Patient Study</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fbf22541-795b-4447-8581-07f30ffc2a37">
                                        <xref linkend="sect_C.7.2.2" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_29c732e8-5042-4973-b160-e852dad1058d">U</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e8bb0c4c-6b28-4c08-b099-16a8be6a3aa5">Clinical Trial Study</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e44eae31-157d-44e4-ba12-8a7a7484e54b">
                                        <xref linkend="sect_C.7.2.3" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_96ca5359-cd14-454b-9a9d-723afde6720b">U</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="3">
                                    <para xml:id="para_33b60570-e731-4d59-8d44-275339bf4891">Series</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1f4d3383-6770-469d-b743-63e10616c06b">General Series</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ce2364eb-8390-4f0b-be8c-078a3d1c02af">
                                        <xref linkend="sect_C.7.3.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1c662022-0e24-448e-8cec-e9ea997ccd32">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dddf9296-1c3e-44d2-b8a8-45f6b15e6bd9">Whole Slide Microscopy Series</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9cfd56fe-cf65-47a0-9a54-dcdc6e0657e7">
                                        <xref linkend="sect_C.8.12.3" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_20cfb123-d696-41fc-92bf-fcf0acae9f62">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_95c0ab6b-7540-4e4d-9b95-337bb933a4b7">Clinical Trial Series</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_42678f44-4fd0-4fd1-a132-5f6557e1af4d">
                                        <xref linkend="sect_C.7.3.2" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6f950102-c4bb-4fe5-84ec-60a04683ca05">U</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d9957255-a01a-4661-8cd3-05818063f693">Frame of Reference</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1d08a784-8e38-45b4-917d-17875f6cf8e7">Frame of Reference</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c577f627-7be7-478b-bbd7-7eaed25fba03">
                                        <xref linkend="sect_C.7.4.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f108cea2-c2e5-47f9-a045-3a3e2aa4296e">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="2">
                                    <para xml:id="para_5a145364-25f8-4ed7-b90b-1ad6f5adca85">Equipment</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7b5fbb2b-65f7-4e83-a9d3-f106d6befa62">General Equipment</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8486f4a6-56f9-44e7-9d0f-92e42a405d25">
                                        <xref linkend="sect_C.7.5.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_91bab745-911c-4f5e-bb62-a6be7121a701">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_24325e10-ce95-4553-a3b8-b1d711c8bd51">Enhanced General Equipment</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c3d610bc-f1ae-4252-bae6-fece2cad0dc6">
                                        <xref linkend="sect_C.7.5.2" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_030bb10c-f0d6-4f07-85ae-258cdd38d92b">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="14">
                                    <para xml:id="para_7375a25a-5403-4de8-9a4c-45b30c5be07a">Image</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f20a33ce-5b09-4dd7-8c69-37b2e0cda0b6">General Image</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_790b103c-7a37-4517-b467-f6f1108da467">
                                        <xref linkend="sect_C.7.6.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6c162cc7-ae32-4a46-aa06-cb620d26a27d">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_97029079-24ab-4cd7-b101-6fdcd1b04311">General Reference</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b827e86f-6eab-4903-847c-7e7a930817bc">
                                        <xref linkend="sect_C.12.4" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b78b86fa-c8f3-4b34-b17d-a67d81538aa3">U</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2f883737-c2c7-4972-a4da-e70c019fb143">Image Pixel</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ca7e8f1a-ac02-49ed-bba3-289fe0435f40">
                                        <xref linkend="sect_C.7.6.3" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ced6afa9-37df-405c-bdba-a9c4b151166d">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d4596786-626f-462f-a32e-9c49a173f46c">Acquisition Context</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2f12e8ec-1ec0-41c6-999c-57d7a20cb2d0">
                                        <xref linkend="sect_C.7.6.14" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_84afd14d-b896-48db-88b3-b5530817cf37">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e15cd9e7-d68f-44eb-953d-540e0b4ca2b2">Multi-frame Functional Groups</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4f172328-99c9-4000-87c4-b2cb77ec3a3f">
                                        <xref linkend="sect_C.7.6.16" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_731c5a92-fd45-4ec9-8ac9-85d315abb765">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_746402a4-0af5-4630-a9d1-a654727e401c">Multi-frame Dimension</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5f895da2-3484-4b6b-8008-873c5d81f575">
                                        <xref linkend="sect_C.7.6.17" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_13e00028-3e42-4eeb-9157-802f96ca1b58">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_528f7ec0-77fc-4cf0-9971-08a704e580bf">Specimen</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0e13a989-972f-4c62-8eb0-96fee82f9373">
                                        <xref linkend="sect_C.7.6.22" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4e7e7ae9-7986-48d2-be21-07cd491fe8d3">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5e1d6075-346d-4a57-b0ec-52a2234b92d7">Whole Slide Microscopy Image</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1bf64cff-c4c1-4e6b-853f-855bad364810">
                                        <xref linkend="sect_C.8.12.4" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_611700fc-a735-4464-840b-c66d18d3e095">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4e3fc080-f518-4077-bb11-b7441c9a9925">Optical Path</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7cd0f9d8-1e96-48b0-8974-f999d7b841d6">
                                        <xref linkend="sect_C.8.12.5" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8b664085-1c72-494f-86b2-01c05d2f0f50">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_08f5b135-1390-4843-a4dd-835bfa088f93"><emphasis role="bold"><emphasis role="strikethrough">Multi-Resolution Navigation</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8bf48559-c208-43e6-ac2b-3dfbeda6f018">
                                        <emphasis role="bold"><emphasis role="strikethrough"><xref linkend="sect_C.8.12.7" xrefstyle="select: labelnumber"/></emphasis></emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3ae19513-f0f7-47b8-9fc4-279232b3aa73"><emphasis role="bold"><emphasis role="strikethrough">C - Required if Image Type (0008,0008) Value 3 is LOCALIZER</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0e07c6e5-9797-4dc6-9608-243b0158ab26">Slide Label</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9f1a66d7-35a2-437a-8173-f706055c2da0">
                                        <xref linkend="sect_C.8.12.8" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4d77831d-26a2-466b-bd25-e8d29fb6b64a">C - Required if Image Type (0008,0008) Value 3 is LABEL; may be present otherwise</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fc1ca4b1-fd53-4b9d-8561-fcb30736d654">SOP Common</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_61998379-8957-42bf-bebe-8e94117a357b">
                                        <xref linkend="sect_C.12.1" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1a13a5e7-473a-4374-ab8a-733114bd3046">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_eb5e5f65-414e-4d97-ae1c-0f9ed2c6d328">Common Instance Reference</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d0c8a88f-45dc-4b94-a7c1-a5359f24d54c">
                                        <xref linkend="sect_C.12.2" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dbbe92a1-082f-44a3-9342-abf8eb64720c">M</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_974f541e-478d-4fd4-a949-41ded223a630">Frame Extraction</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ff7e5cab-f99e-4551-9727-0755e020d958">
                                        <xref linkend="sect_C.12.3" xrefstyle="select: labelnumber"/>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f14b76a7-69e2-4628-b35e-b9425325544c">C - Required if the SOP Instance was created in response to a Frame-Level retrieve request</para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
            </section>

        <section label="C.8.12.4" status="4" xml:id="sect_C.8.12.4">
                    <title>Whole Slide Microscopy Image Module</title>
                    <table frame="box" label="C.8.12.4-1" rules="all" xml:id="table_C.8.12.4-1">
                        <caption>Whole Slide Microscopy Image Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_19b0b4dc-4e92-4c95-9988-804684315cb7">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_96a09a1d-8879-43be-97e5-121291459369">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3e3024bf-3c64-498a-ba60-ec15eea8abd4">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_18f4a093-c520-4639-ae1b-71059a30ef26">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7a877f73-4d9d-41b5-86d4-a8c3bbfedf52">Image Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_379f1037-05af-47e4-a7cc-263ccdb59552">(0008,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5e62e3f9-7cd2-48f1-898c-3cc78973e9f9">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8987cdf6-559b-4308-8bb6-886172153890">Image identification characteristics.</para>
                                    <para xml:id="para_6af97694-9c87-41bf-9a64-b1a1ab24272e">See <xref linkend="sect_C.8.12.4.1.1" xrefstyle="select: label"/> for specialization.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para>...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_305dffa2-cc70-4472-81da-73a738d8d1db">Number of Frames</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b35b2b79-4aeb-4806-88cc-eb002a89af1e">(0028,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_df7f6e9c-de29-4a8a-8fc0-c33eeb2a82d3">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_60b26676-3382-4932-b4fc-fdf092f8966d">Number of frames in a multi-frame image.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Image Type (0008,0008) Value 3 is LOCALIZER or LABEL or OVERVIEW<emphasis role="bold"><emphasis role="underline"> or THUMBNAIL</emphasis></emphasis>:</title>
                                        <varlistentry>
                                            <term>1</term>
                                            <listitem>
                                                <para xml:id="para_495140e7-94cc-4a17-aee0-9ae324754fea"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para>...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c0abbc80-e1e3-4b82-bedf-eea3f41d842c">Specimen Label in Image</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_41907a6c-a9af-43b1-96a1-a30dc3d2f54d">(0048,0010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8a2db66a-0db1-47bf-ad3c-17b6c58819d2">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b85b9cc4-4c6f-4568-a3c9-e94856b627a8">Indicates whether the specimen label is captured in the image.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>YES</term>
                                            <listitem>
                                                <para xml:id="para_af18122b-cea0-4201-b132-eac704fa487c"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>NO</term>
                                            <listitem>
                                                <para xml:id="para_e2814578-ee15-4106-a9bc-0b7b646d6950"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para><emphasis role="bold"><emphasis role="underline">Shall be YES if Image Type (0008,0008) Value 3 is OVERVIEW or LABEL.</emphasis></emphasis></para>
                                    <para><emphasis role="bold"><emphasis role="underline">Shall be NO if Image Type (0008,0008) Value 3 is THUMBNAIL.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_23c65759-8224-43f0-8108-57bff7f37706">Burned In Annotation</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b7dec658-348b-449e-a9a5-2f2f72bcd6b3">(0028,0301)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_663b9cbd-c4a7-4239-9523-e3f407e1886f">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e0d86012-53f3-40d8-a34e-f26162e14ae2">Indicates whether or not image contains sufficient burned in annotation to identify the patient.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>YES</term>
                                            <listitem>
                                                <para xml:id="para_1412cf14-c7ff-4f3b-b606-e392f36a318a"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>NO</term>
                                            <listitem>
                                                <para xml:id="para_177c1e45-14e0-4047-9c14-b9a5471ba8f3"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <note>
                                        <para xml:id="para_99d1052a-b130-4d1b-bb98-c56b3d10fe17">If Specimen Label in Image (0048,0010) value is YES, Burned In Annotation (0028,0301) might be NO if the label includes only a specimen identifier and not patient identifying data.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">...</td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_63510a3e-850c-468a-b922-27f8aab9ff9a"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bdd1c995-c5ae-4f73-beaa-e69b33786880"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7e2ffef0-9f59-41c4-b401-17681bd063d7"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.8.12.4.1" status="5" xml:id="sect_C.8.12.4.1">
                        <title>Whole Slide Microscopy Image Attribute Descriptions</title>
                        <section label="C.8.12.4.1.1" status="6" xml:id="sect_C.8.12.4.1.1">
                            <title>Image Type</title>
                            <para xml:id="para_570df186-e80e-495a-8c3f-a885b9326afc">Image Type (0008,0008) is specified to be Type 1 with the following constraints:</para>
                            <para xml:id="para_5d7c3916-3108-4847-b5d9-a9efd8c2719f">Value 1 shall have a value of ORIGINAL or DERIVED</para>
                            <para xml:id="para_64085fb2-c119-461d-b5ea-597a1b2f0c02">Value 2 shall have a value of PRIMARY</para>
                            <para xml:id="para_7a14bdbb-e4da-42ef-a9f0-42bf9c4a72d6">Value 3 (Image Flavor) shall have the Defined Terms in <xref linkend="table_C.8.12.4-2" xrefstyle="select: label"/>.</para>
                            <table frame="box" label="C.8.12.4-2" rules="all" xml:id="table_C.8.12.4-2">
                                <caption>Whole Slide Microscopy Image Flavors</caption>
                                <tbody>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_f4e0b9db-57d3-40c8-be18-c56a76e704e9"><emphasis role="bold"><emphasis role="strikethrough">LOCALIZER</emphasis></emphasis></para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9914348f-ddb2-4d12-9137-fbec1afa0557"><emphasis role="bold"><emphasis role="strikethrough">Collected for the purpose of planning or navigating other images; provides frame-level navigational features.</emphasis></emphasis></para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_7490619d-a47e-420d-bbab-aa7af0caad42">VOLUME</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9b847f99-2c22-472f-b17d-8f60eb58e765">Set of frames that define a regularly sampled volume</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_b553e150-7ca4-4380-933d-3ff577682c88">LABEL</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_32434d33-8596-4260-86d4-c66a6bf6994e">Purpose of image is to capture the slide label only; non-label area shall not be present; the label is shown in <xref linkend="figure_C.8-16" xrefstyle="select: label"/>.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_e0209857-046a-4c26-9cab-c409ce8c2cc8">OVERVIEW</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_e4393930-6e0d-41d4-b595-4b51fb20dbb6">Purpose of image is to provide an overview of the entire glass slide; shall include <emphasis role="bold"><emphasis role="underline"> at least part of </emphasis></emphasis>the label, <emphasis role="bold"><emphasis role="underline">all of the </emphasis></emphasis>the specimen and the empty space that are shown in <xref linkend="figure_C.8-16" xrefstyle="select: label"/>.</para>
                                            <note>
                                                <para xml:id="para_e3d5a499-8d49-4574-bca7-302449e0f320">One purpose of the OVERVIEW image is to allow visual quality control to assure that tissue fragments have not been omitted when selecting areas (manually or automatically) to digitize at high resolution. The OVERVIEW image is distinct from the lowest resolution VOLUME image(s) of WSI pyramids<emphasis role="bold"><emphasis role="underline">, and the THUMBNAIL image, neither of which includes the label</emphasis></emphasis>.</para>
                                            </note>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para><emphasis role="bold"><emphasis role="underline">THUMBNAIL</emphasis></emphasis></para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para><emphasis role="bold"><emphasis role="underline">Purpose of image is to provide an overview of the specimen; shall not include the label.</emphasis></emphasis></para>
                                        </td>
                                    </tr>
                                </tbody>
                            </table>
                            <note>
                                <para><emphasis role="bold"><emphasis role="underline">A Value 3 of LOCALIZER was previously been defined, which triggered the mandatory inclusion of the retired <xref linkend="sect_C.8.12.7" xrefstyle="select: label title"/>. See PS3.3 202nx.</emphasis></emphasis></para>
                            </note>
                            <para xml:id="para_fb8f0316-22e1-4eb4-bc32-cba178b99ea1">Value 4 (Derived Pixel) shall have the Defined Terms specified in <xref linkend="table_C.8.12.4-3" xrefstyle="select: label"/>.</para>
                            <table frame="box" label="C.8.12.4-3" rules="all" xml:id="table_C.8.12.4-3">
                                <caption>Whole Slide Microscopy Image Derived Pixels</caption>
                                <tbody>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_21cbe34a-c110-42e8-8b67-a67c8ab89150">NONE</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_4b1732c6-26fc-4014-aa1a-10101b0ac589">No derivation of pixels (original)</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9b1cc671-a848-44d5-87a6-6562644deb2a">RESAMPLED</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9bb22eb0-ea22-48a0-8a00-0350e6287192">Pixels were derived by down sampling a higher resolution image</para>
                                        </td>
                                    </tr>
                                </tbody>
                            </table>
                            <para xml:id="para_3a8971ef-39d9-413a-83a7-81a386515947">No additional values shall be present.</para>
                        </section>
                    </section>
                </section>
        
        <section label="C.8.12.7" status="4" xml:id="sect_C.8.12.7">
            <title>Multi-Resolution Navigation Module</title>
            <para xml:id="para_58233e8c-dd44-4ab6-aec7-11818387a78b">
                <emphasis role="bold"><emphasis role="strikethrough"><xref linkend="table_C.8.12.7-1" xrefstyle="select: label"/> specifies the Attributes that describe the Multi-Resolution Navigation Module, used in Whole Slide Microscopy LOCALIZER images.</emphasis><emphasis role="underline">Retired. See PS3.3 202nx.</emphasis></emphasis></para>
            <table frame="box" label="C.8.12.7-1" rules="all" xml:id="table_C.8.12.7-1">
                <caption><emphasis role="bold"><emphasis role="strikethrough">Multi-Resolution Navigation Module Attributes</emphasis></emphasis></caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8f9e9740-12d7-46c6-a747-ae32ff9149ea"><emphasis role="bold"><emphasis role="strikethrough">Attribute Name</emphasis></emphasis></para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8228eb86-be7d-442b-9df8-a66b555ebcd8"><emphasis role="bold"><emphasis role="strikethrough">Tag</emphasis></emphasis></para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a16ecfb9-0d51-498e-8f8e-a8332ddf4acc"><emphasis role="bold"><emphasis role="strikethrough">Type</emphasis></emphasis></para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2125580f-0ab8-494b-9d1f-4d06b97cab52"><emphasis role="bold"><emphasis role="strikethrough">Attribute Description</emphasis></emphasis></para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7fc5973d-4349-462f-b6df-cef1bd2a1966"><emphasis role="bold"><emphasis role="strikethrough">Referenced Image Navigation Sequence</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cdd55b3d-2b76-4a5a-aea7-fb9595e7296c"><emphasis role="bold"><emphasis role="strikethrough">(0048,0200)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b081b8d3-38e4-4541-9d25-ee1c57979c2d"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f044bd05-ad09-4ae0-822d-2e1aac41dfe5"><emphasis role="bold"><emphasis role="strikethrough">Identification of correspondence between areas of Pixel Data (07FE,0010) and frames of higher resolution images.</emphasis></emphasis></para>
                            <para xml:id="para_68e9434d-c09b-4a5d-8a7d-a75999d839e9"><emphasis role="bold"><emphasis role="strikethrough">One or more Items shall be included in this Sequence.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_0b838e3f-8efb-49c9-a18d-28f40ba04186">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">&gt;Include <xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>
                                        </emphasis></emphasis></emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2d504f52-fb19-412c-8cd0-91d9776183db">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">Referenced Image Instance</emphasis></emphasis></emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2f37eaf5-2dae-47ca-92b6-f7f127ea696b"><emphasis role="bold"><emphasis role="strikethrough">&gt;Referenced Frame Number</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_dffe74d9-90bb-40fc-8577-579f2b3d36dd"><emphasis role="bold"><emphasis role="strikethrough">(0008,1160)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f4bbdbd6-4513-4549-aacd-bec59ab5002c"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_98a08974-e6df-4766-a841-f78890f419fb"><emphasis role="bold"><emphasis role="strikethrough">Referenced Frame Number within Referenced Image. Only a single value shall be present.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cbf72753-ee2d-41e7-b7b5-b1bc4db24cd0"><emphasis role="bold"><emphasis role="strikethrough">&gt;Top Left Hand Corner of Localizer Area</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_26e105a6-02fa-4703-a018-5d74eab330e5"><emphasis role="bold"><emphasis role="strikethrough">(0048,0201)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ef33294f-873d-4301-9c12-d2dbdb52a5b1"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_229df27f-f087-463e-a373-2da1516c94bd"><emphasis role="bold"><emphasis role="strikethrough">Top Left Hand Corner of Localizer image pixel area corresponding to referenced frame, specified as column\row. Column is the horizontal offset and row is the vertical offset.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2be1d41a-e0c3-4a36-bea4-9e6f19912eda"><emphasis role="bold"><emphasis role="strikethrough">&gt;Bottom Right Hand Corner of Localizer Area</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6a12d5f7-fe92-4ca3-8812-dc2c49b6f0bc"><emphasis role="bold"><emphasis role="strikethrough">(0048,0202)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2f40d70f-d0a6-4c3f-897b-7769d4df0bf3"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9aa14a2a-1827-4e12-9477-18abc6ef4144"><emphasis role="bold"><emphasis role="strikethrough">Bottom Right Hand Corner of Localizer image pixel area corresponding to referenced frame, specified as column\row. Column is the horizontal offset and row is the vertical offset.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c4805921-94ad-45d0-a3fb-c322a99cb433"><emphasis role="bold"><emphasis role="strikethrough">&gt;Pixel Spacing</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_77fb486e-edf3-43f7-b02f-305dc653e835"><emphasis role="bold"><emphasis role="strikethrough">(0028,0030)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_19b5bba7-51db-4106-875b-d2a30c42a6b3"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bd26dfef-c22a-41ce-8ecc-80d325b92bb4"><emphasis role="bold"><emphasis role="strikethrough">Pixel Spacing in Referenced Image, specified as a numeric pair - adjacent row spacing \ adjacent column spacing, in mm.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a6442012-b802-4c54-b40a-9082bc00af72"><emphasis role="bold"><emphasis role="strikethrough">&gt;Z Offset in Slide Coordinate System</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c2d0cc05-e1e3-4336-8d07-ff9f4175ed49"><emphasis role="bold"><emphasis role="strikethrough">(0040,074A)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_151af56a-e819-4dd4-8f43-d7d6f613935e"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f003ffe8-2047-4070-a414-dd9af71b7445"><emphasis role="bold"><emphasis role="strikethrough">The Z offset of the referenced frame from the Slide Coordinates System reference plane, in microns</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2b99656f-69e1-49c8-9c34-80702bd988ee"><emphasis role="bold"><emphasis role="strikethrough">&gt;Samples per Pixel</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f9c09a4c-093a-4d3c-95f2-bc935011acfa"><emphasis role="bold"><emphasis role="strikethrough">(0028,0002)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cd6b6665-f291-44e9-8bba-9e109e728102"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_76afb7a9-02db-460b-adc5-c293040c2c8b"><emphasis role="bold"><emphasis role="strikethrough">The number of samples (color planes) containing information in the referenced image.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e9f20f14-b273-43b4-ae53-9d48ab810d5f"><emphasis role="bold"><emphasis role="strikethrough">&gt;Optical Path Identifier</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_dc350eb5-ee43-4c79-8bdb-38ae89e15ff9"><emphasis role="bold"><emphasis role="strikethrough">(0048,0106)</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0d8a59fa-bb45-4053-ba2b-d76ad96521a6"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ea5d120e-f438-44ae-8aab-b13deb51a43a"><emphasis role="bold"><emphasis role="strikethrough">Identifies the optical path used in acquiring the referenced Frame by reference to an Item in the Optical Path Sequence (0048,0105) of the current SOP Instance. See <xref linkend="sect_C.8.12.5" xrefstyle="select: label"/>.</emphasis></emphasis></para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged, DICOM PS3.3:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="C.8.12.2.1" status="5" xml:id="sect_C.8.12.2.1">
            <title>Slide Coordinates Attribute Descriptions</title>
            <section label="C.8.12.2.1.1" status="6" xml:id="sect_C.8.12.2.1.1">
                <title>Image Center Point Coordinates Sequence</title>
                <para xml:id="para_36d52c09-aef4-4f53-8b10-bc2e5b626b98">
                    <figure label="C.8-16" pgwide="1" xml:id="figure_C.8-16">
                        <title>Reference Slide Orientation</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp2102_AddWSIThumbnailAndRetireLocalizer/PS3.3_C.8-16.svg"/>
                                <!--<imagedata fileref="figures/PS3.3_C.8-16.svg"/>-->
                                <!--<imagedata fileref="part03_fromword_files/image241.png"/>-->
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
            </section>
        </section>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="6" status="1" xml:id="chapter_6">
            <title>Registry of DICOM Data Elements</title>
            <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
                <caption>Registry of DICOM Data Elements</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                                <emphasis role="bold">Tag</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                                <emphasis role="bold">Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                                <emphasis role="bold">Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                                <emphasis role="bold">VR</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                                <emphasis role="bold">VM</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_80f799cd-fb24-4567-90da-84a160c34564"><emphasis role="italic">(0048,0200)</emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_363dc5b9-08e5-406a-ac1b-19d5e070e651"><emphasis role="italic">Referenced Image Navigation Sequence</emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6bd910d5-2096-4b4b-bebc-df2610e137b3"><emphasis role="italic">Referenced​Image​Navigation​Sequence</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1f3f2ba6-0000-46e6-8718-fb40e18f6426"><emphasis role="italic">SQ</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4b65befe-2647-4a62-86a9-18e95c166487"><emphasis role="italic">1</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fddd1195-e5d0-42c5-9d94-de77cf2f413b"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET (202nx)</emphasis></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b6c507ce-692e-4c93-a012-03b9155775ad"><emphasis role="italic">(0048,0201)</emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a159bab1-3526-4283-b245-bd82e8e9b812"><emphasis role="italic">Top Left Hand Corner of Localizer Area</emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6c8f3f16-e4b7-46f0-9081-3b64d5eaecab"><emphasis role="italic">Top​Left​Hand​Corner​Of​Localizer​Area</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3e873347-45c2-47e7-9e9c-292701454a00"><emphasis role="italic">US</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_814e0d1f-e8a1-4b7f-b879-eaf1d17c389c"><emphasis role="italic">2</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6ce76b81-b87a-4f7b-ba8a-3f374a6b514a"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET (202nx)</emphasis></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_48d6dc35-596c-4f1f-a632-d1a7a3f6f873"><emphasis role="italic">(0048,0202)</emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_de73a3bd-1ea5-4564-b696-2d10b8eb40ca"><emphasis role="italic">Bottom Right Hand Corner of Localizer Area</emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_efb8b344-502e-4466-8c2e-34eb9bdb36e0"><emphasis role="italic">Bottom​Right​Hand​Corner​Of​Localizer​Area</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_690e0685-66a2-4a5e-8fe5-d945ea1c81f2"><emphasis role="italic">US</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d986a385-c6da-45e5-aa0c-6d4b40ecebfd"><emphasis role="italic">2</emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_63c018db-52da-431c-b32b-1e109366615e"><emphasis role="italic"><emphasis role="bold"><emphasis role="underline">RET (202nx)</emphasis></emphasis></emphasis></para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        
               
        
        
        
        
    </chapter>
</book>