<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2100</title>
    <subtitle>Clarify SMPTE Transfer Syntaxes for DICOM Web Services</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2021/07/04</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Mathieu Malaterre</para><para><link xl:href="mailto:mathieu.malaterre@gmail.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/04/16</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2100</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Clarify SMPTE Transfer Syntaxes for DICOM Web Services</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.18</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The video Transfer Syntax SMPTE ST 2110-20, added for RTV, is not mentioned as being supported or not.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.18 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        
<section label="8.7.3.1" status="4" xml:id="sect_8.7.3.1">
                    <title>The application/dicom Media Type</title>
                    <para xml:id="para_47a6547d-80ef-453a-82e6-2adc550f82db">The application/dicom media type specifies a representation of Instances encoded in the DICOM File Format specified in <olink targetdoc="PS3.10" targetptr="chapter_7" xrefstyle="template:Section %n “%t” in PS3.10"/>.</para>
                    <note>
                        <para xml:id="para_2db85632-05a8-4ea6-861c-c938c857bc96">The origin server may populate the <olink targetdoc="PS3.10" targetptr="PS3.10" xrefstyle="select: labelnumber"/> File Meta Information with the identification of the Source, Sending and Receiving AE Titles and Presentation Addresses as described in <olink targetdoc="PS3.10" targetptr="sect_7.1" xrefstyle="select: docname label"/>, or these Attributes may have been left unaltered from when the origin server received the objects. The user agent storing the objects received in the response may populate or coerce these Attributes based on its own knowledge of the endpoints involved in the transaction, so that they accurately identify the most recent storage transaction.</para>
                    </note>
                    <para xml:id="para_500a6fa6-1823-4929-905f-a9fdc2ff988c">
                        <xref linkend="table_8.7.3-2" xrefstyle="select: label"/> specifies the default and optional Transfer Syntax UID combinations for each DICOM Resource Category (see <xref linkend="table_8.7.2-1" xrefstyle="select: label"/>). The default media type for the Resource Category shall be returned when the origin server supports none of the Acceptable Media Types, unless the origin server has only access to the pixel data in lossy compressed form or the pixel data in a lossless compressed form that is of such length that it cannot be encoded in the Explicit VR Little Endian Transfer Syntax.</para>
                    <table frame="box" label="8.7.3-2" rules="all" xml:id="table_8.7.3-2">
                        <caption>Transfer Syntax UIDs for application/dicom Media Types</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7ba28fa3-4cb7-464c-afbe-0b4693771288">Category</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_95d35ffd-d5e7-43b2-a0bd-1a8d81e2c085">Transfer Syntax UID</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4f36c4f7-667f-49cc-af7f-7d3981022110">Transfer Syntax Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a9174cee-b820-4d78-8aa0-4de7f0f12936">Optionality</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="12">
                                    <para xml:id="para_81301422-2f37-4af8-a162-e0510c654a0f">Single Frame Image</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_262b6fc9-089e-4c9b-b4ce-82106b83879b">1.2.840.10008.1.2.1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ebe50c83-1e02-4bd1-a897-c14b8d9dce17">Explicit VR Little Endian</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a5cc9add-6e4f-4a41-8eff-d58dde41ca80">D</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_565fcd1f-97b4-4d79-9a41-ff6143b65616">1.2.840.10008.1.2.4.70</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_02ec73e5-2c33-4b1f-aee6-daad26c73ab6">JPEG Lossless, Non-Hierarchical, First-Order Prediction(Process 14 [Selection Value 1]): Default Transfer Syntax for Lossless JPEG Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_be656b08-d398-4b49-9c1f-6224305daafe">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6fbc9e71-a805-4f35-a4bf-52d6acb573a8">1.2.840.10008.1.2.4.50</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_99dedbab-077c-4149-8683-4d6d745bdbca">JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8 Bit Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2b0424e7-d8a7-454b-8816-288b1a9d5e08">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d87adb5b-44ef-49b3-8be9-4d3fb32d35c8">1.2.840.10008.1.2.4.51</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8bdff36f-d412-49f2-b303-94cd2cebfaa0">JPEG Extended (Process 2 &amp; 4): Default Transfer Syntax for Lossy JPEG 12 Bit Image Compression (Process 4 only)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c97a0fbe-f8c3-4247-a4bd-894a06a5a489">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_80c2e1ca-056a-43bd-90b2-5d00296d4051">1.2.840.10008.1.2.4.57</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_de9a9c75-c1a6-48b5-8531-3de78131d7cd">JPEG Lossless, Non-Hierarchical (Process 14)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9442f4ab-9310-4e46-8e08-5b18d5884f4d">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7ba426e6-82b8-45b8-9ef6-8060afbc0592">1.2.840.10008.1.2.5</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e5458d6b-a9e1-4fbb-a9fc-d2005a0f559f">RLE Lossless</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_019e4c35-7e1c-490e-a6a2-70d7832fef3c">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d9c28978-b593-4678-b98d-6b4557e43ca1">1.2.840.10008.1.2.4.80</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_684a91c5-8159-44f5-a88f-0a18f4ca7973">JPEG-LS Lossless Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c8b1228c-0778-44ec-ae19-ecb843a76fd0">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_573a987d-b3aa-43f9-afc9-ad4b23a9a5bf">1.2.840.10008.1.2.4.81</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2853615d-447d-4bfe-9326-84564a825ad4">JPEG-LS Lossy (Near-Lossless) Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e7cb497b-292f-456b-9308-e8d4cf73db6f">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1e2a7e28-47f6-410c-ac4c-92c7da200f0f">1.2.840.10008.1.2.4.90</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1ec81877-62e6-4381-b0e6-aa0f55f37ad0">JPEG 2000 Image Compression (Lossless Only)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_737d1a8f-5bab-4b2e-9028-54d175b1a60f">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a42b67cb-1e4e-422e-94c0-b48ab21062a4">1.2.840.10008.1.2.4.91</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7d9c58e3-f1f4-45ad-8f0b-949320a24e6d">JPEG 2000 Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_11fd1dba-8fff-4b5a-9a3f-a8d5a9ac498a">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9eb09bad-24e4-4e2d-90f7-640d477c84ec">1.2.840.10008.1.2.4.92</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_32eb08e1-c0c3-4180-9c0e-9ee6bd7c227f">JPEG 2000 Part 2 Multi-component Image Compression (Lossless Only)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d544af72-a31d-459c-ac22-8fa3fc4aa72f">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e019e212-a279-4ebf-9076-4bf2d672c17e">1.2.840.10008.1.2.4.93</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_082ca231-7f12-4a74-a13e-c3c036ac560e">JPEG 2000 Part 2 Multi-component Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6a281dc5-4769-4e4a-9e7b-848e80ad1bde">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="5">
                                    <para xml:id="para_0e238099-831d-4651-a463-db14a6e67cf6">Multi-frame Image</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0941aca4-dba5-42cc-8ab3-77ecfef272e5">1.2.840.10008.1.2.1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cb397a36-09e9-47a0-b149-dd686be9b429">Explicit VR Little Endian</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2ef08400-9d37-4ced-8c34-3bda87eadf87">D</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_76ab2c44-9160-4b50-94f4-bfb275aa4f3b">1.2.840.10008.1.2.4.90</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3c4f5a70-45b8-4cc0-9af8-386f519c44b1">JPEG 2000 Image Compression (Lossless Only)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3afca3a0-937a-4da8-9726-24cddcfa6b38">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bf0b6949-1956-4ce9-bc96-8e1874f04a37">1.2.840.10008.1.2.4.91</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f7a015ae-0e37-4b0f-a031-c132953246be">JPEG 2000 Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_988ce086-9923-4481-949a-9bd93d8b1dee">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0db0e2e7-e1e2-4a19-ad64-b050ce021c4a">1.2.840.10008.1.2.4.92</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1ef9d9c4-f985-479d-8611-14fb45716208">JPEG 2000 Part 2 Multi-component Image Compression (Lossless Only)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_02030129-438d-4462-a2b0-e0326ab580ec">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d64cb6d2-7b46-4bc8-8670-762ee25f2583">1.2.840.10008.1.2.4.93</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_39a92bca-59ac-4ad5-aa35-3f734dd0bd9f">JPEG 2000 Part 2 Multi-component Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_85572163-a653-4b40-a5da-45313e918f6f">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="10">
                                    <para xml:id="para_35596624-6724-4346-9666-43dea9252dcd">Video</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dba2dd58-c749-4f33-a94c-f5325c9f721a">1.2.840.10008.1.2.1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f383fc94-e036-4a12-942e-b3cca6fb5d63">Explicit VR Little Endian</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_91f03550-a63f-46d7-8804-537af132ad23">D</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2789b547-5be0-4e6e-9a51-21a53b2273ef">1.2.840.10008.1.2.4.100</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f2a2f8a7-629d-492b-b30d-e267a20aabf0">MPEG2 Main Profile @ Main Level</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fa80d456-c11e-4927-9ef4-2ef52fdc80d8">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ef46357c-7df6-447b-a9b0-38d004376cc3">1.2.840.10008.1.2.4.101</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6505b6b3-6324-41e5-887c-cf090a3f32f1">MPEG2 Main Profile @ High Level</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_36345536-da66-4bb3-b58f-dbfb18d0949a">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8113ab7e-e7e9-4334-a648-272b12966d10">1.2.840.10008.1.2.4.102</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_47b1b669-652a-41b8-8826-6dbe5333c431">MPEG-4 AVC/H.264 High Profile / Level 4.1</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9294657e-ad8b-40e3-8520-16eb0567dc53">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_510aeb03-1207-43d9-8e08-127b35962035">1.2.840.10008.1.2.4.103</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b1662c00-1691-4142-86eb-77133a2329bb">MPEG-4 AVC/H.264 BD-compatible High Profile / Level 4.1</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_07a60523-0337-4142-aeb2-02e1bb54e5a0">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f60b5531-7e4f-4d8d-b381-9d2aade6e9b4">1.2.840.10008.1.2.4.104</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_09e910a4-f027-444a-921d-f735498ccd29">MPEG-4 AVC/H.264 High Profile / Level 4.2 For 2D Video</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_279cb89f-873e-4241-8f33-9f71bc3a4f8e">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_81dce5e3-230d-49ba-b54d-2599fcd886dc">1.2.840.10008.1.2.4.105</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bc4be8b1-27ec-4da2-b760-70a9ec0e5fe7">MPEG-4 AVC/H.264 High Profile / Level 4.2 For 3D Video</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0d04a0c3-3637-4bf7-b257-a562c5d2d9a0">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4eb967a2-f2fb-406f-abc3-c088e26fba63">1.2.840.10008.1.2.4.106</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ed0918fe-1d22-44c2-926b-a5c2c23375e0">MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d4f78fbf-279b-4c3e-aade-61dcd7a09315">O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para>1.2.840.10008.1.2.4.107</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para>HEVC/H.265 Main Profile / Level 5.1</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para>O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para>1.2.840.10008.1.2.4.108</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para>HEVC/H.265 Main 10 Profile / Level 5.1</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para>O</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_255fe5b2-d3ac-42d7-8f10-b24d4be03f38">Text</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7328dfc5-ff13-4588-8329-a50a815ce651">1.2.840.10008.1.2.1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cdb0448e-8194-4a55-93f0-3862bc989e95">Explicit VR Little Endian</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8519a66a-613d-4037-96b4-355b6fb00985">D</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c204a964-0626-4a2f-936d-cf0592ce61c8">Other</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_26013b21-b5e4-4ba7-89a8-119696bc22f3">1.2.840.10008.1.2.1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6ead09c6-8278-461a-95a8-55c21a33c43d">Explicit VR Little Endian</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5ea162c4-a84f-47e3-a2e7-039cd3fda28f">D</para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <note>
            <para><emphasis role="bold"><emphasis role="underline">The SMPTE ST 2110-20 Uncompressed Progressive Active Video Transfer Syntax is not included, since it is only used in a DICOM-RTV Metadata Flow, rather than in a representation of an Instance encoded in the DICOM File Format.</emphasis></emphasis></para>
        </note>
        
        

        
        
               
        
        
        
        
    </chapter>
</book>