    <title>CP-2045</title>
    <subtitle>Use Content Item with Modifier Macro in Acquisition Context</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/06/15</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie <link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>IDC</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/04/11</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2045</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Use Content Item with Modifier Macro in Acquisition Context</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>Over time, the Content Item usage outside SR in various roles such as worklists and protocols has improved, but these changes (including the use of a level of nesting for modifiers) have not been propagated to Acquisition Context as they should have been.</para>
                    <para>This change allows for additional value types (including DATETIME as well as references to objects including UIDREF, COMPOSITE, IMAGE and WAVEFORM) as well as modifiers.</para>
                    <para>The Content Item macro is also extended to include WAVEFORMS, since there is no reason to allow imags but not waveforms.</para>
                    <para>The Content Item may be applicable to a subset of not only frames, but also segments or waveforms channels.</para>
                    <para>Originally, the Value Type was Type 3 not Type 1, but given the negligible adoption of Acquisition Context, it is appropriate to make it required.</para>
                    <para><emphasis role="italic">[Open Issue: Should we add an observation start datetime?]</emphasis></para>
                    <para><emphasis role="italic">[Open Issue: Does Acquisition Context need a TCCORD mechanism?]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.3 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
                 

        
        <section label="C.7.6.14" status="4" xml:id="sect_C.7.6.14">
                    <title>Acquisition Context Module</title>
                    <para xml:id="para_a73662c2-3f5f-4299-b8c9-51f960083a14">
                        <xref linkend="table_C.7.6.14-1" xrefstyle="select: label"/> specifies Attributes for description of the conditions present during data acquisition.</para>
                    <para xml:id="para_c6d756b4-39c5-451a-a6e5-19a40c1b6f72">This Module shall not contain descriptions of conditions that replace those that are already described in specific Modules or Attributes that are also contained within the IOD that contains this Module.</para>
                    <note>
                        <orderedlist numeration="arabic">
                            <listitem>
                                <para xml:id="para_9291707c-db24-4bb2-b188-3a10da00634a"><emphasis role="bold"><emphasis role="strikethrough">Each Item of the Acquisition Context Sequence (0040,0555) contains one Item of the Concept Name Code Sequence (0040,A043) and one of the mutually-exclusive Observation-value Attributes: Concept Code Sequence (0040,A168), the pair of Numeric Value (0040,A30A) and Measurement Units Code Sequence (0040,08EA), Date (0040,A121), Time (0040,A122), Person Name (0040,A123) or Text Value (0040,A160).</emphasis></emphasis></para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_bba9ef6b-aef2-409c-b080-4efee7a6344f">If this SOP Instance is a Multi-frame SOP Instance, each Item of the Acquisition Context Sequence (0040,0555) may be configured to describe one frame, all frames, or any specifically enumerated subset set of frames of the Multi-frame SOP Instance. Previously, Referenced Frame Numbers (0040,A136) was used to enumerate which frames were described, but its length was limited, so it has been replaced by Referenced Frame Number (0008,1160). <emphasis role="bold"><emphasis role="underline">Similarly, applicability to a subset of Segments or Waveform Channels may be specified.</emphasis></emphasis></para>
                            </listitem>
                        </orderedlist>
                    </note>
                    <table frame="box" label="C.7.6.14-1" rules="all" xml:id="table_C.7.6.14-1">
                        <caption>Acquisition Context Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_44b06e91-f65e-48b1-a80e-01c3fe205332">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a6f80096-6cbe-4009-9d63-867f30ad209a">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_abeb9f86-e6a0-4e64-8a30-a39193d7773e">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_aa82345c-df19-450a-92c0-7d51e666ff55">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b0780935-0a90-403a-8b30-493c4e4bedd0">Acquisition Context Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5b377077-b5c4-471d-913f-9f8e672489a8">(0040,0555)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1d890624-a126-45ef-acc9-fafac636e8e6">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d1c0c51c-b808-43cb-b329-46da35fbb1be">A Sequence of Items that describes the conditions present during the acquisition of the data of the SOP Instance.</para>
                                    <para xml:id="para_bac0f92e-0bd9-4623-91bc-b872b697bfef">Zero or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                    <tr>
                        <td align="left" colspan="4">
                            <para><emphasis role="italic">&gt;Include <xref linkend="table_10.2.1-1" xrefstyle="select: label quotedtitle"/></emphasis></para>
                        </td>
                    </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d4ae484c-1b35-474c-a45a-0820797faee9"><emphasis role="bold"><emphasis role="strikethrough">&gt;Value Type</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_cb3c1e95-9fc7-431f-b4b1-571b9e16f67d">(0040,A040)</para>
                                </emphasis></emphasis></td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_7ffc6359-b6da-44dd-8f35-403d8c79cb67">3</para>
                                </emphasis></emphasis></td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_428b8ab6-5317-4a76-b384-cf09ce58cec5"><emphasis role="bold"><emphasis role="strikethrough">The type of the value encoded in this Item.</emphasis></emphasis></para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term><emphasis role="bold"><emphasis role="strikethrough">TEXT</emphasis></emphasis></term>
                                            <listitem>
                                                <para xml:id="para_76e96784-c4c1-4b16-9c1f-7944e68c05db"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="bold"><emphasis role="strikethrough">NUMERIC</emphasis></emphasis></term>
                                            <listitem>
                                                <para xml:id="para_001da833-67b9-4f39-89de-c2a8b731a8f3"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="bold"><emphasis role="strikethrough">CODE</emphasis></emphasis></term>
                                            <listitem>
                                                <para xml:id="para_3bb4d82d-20ab-4601-ae4b-e388e725a664"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="bold"><emphasis role="strikethrough">DATE</emphasis></emphasis></term>
                                            <listitem>
                                                <para xml:id="para_5cc2fc68-4a48-411b-b2d8-9583c7bd3a8e"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="bold"><emphasis role="strikethrough">TIME</emphasis></emphasis></term>
                                            <listitem>
                                                <para xml:id="para_2de32129-6762-4e79-ba51-5100949d14b3"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="bold"><emphasis role="strikethrough">PNAME</emphasis></emphasis></term>
                                            <listitem>
                                                <para xml:id="para_2d1079fd-b08c-4158-bcce-7f9e2945e1df"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para xml:id="para_f0ef2c2a-5ab4-46b8-80bd-0c9cca2157e1"><emphasis role="bold"><emphasis role="strikethrough">See <xref linkend="sect_10.2" xrefstyle="select: label"/>.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7b136dc0-a00f-4eb0-922b-96a466918bfb"><emphasis role="bold"><emphasis role="strikethrough">&gt;Observation DateTime</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6c4ccdb8-acf1-4623-b10e-d46fe5ce1cc9"><emphasis role="bold"><emphasis role="strikethrough">(0040,A032)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b9509587-2457-4720-839a-30ac95dfb92f"><emphasis role="bold"><emphasis role="strikethrough">3</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e0f3521d-9bd9-41d5-a20b-0edae6acafcb"><emphasis role="bold"><emphasis role="strikethrough">The date and time on which this Item was completed. For the purpose of recording measurements or logging events, completion time is defined as the ending time of data acquisition of the measurement, or the ending time of occurrence of the event.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4d806211-4718-424b-9112-3cea930e342d"><emphasis role="bold"><emphasis role="strikethrough">&gt;Concept Name Code Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e9b0eac1-6269-4eca-8132-8e3b1b9ad1b5"><emphasis role="bold"><emphasis role="strikethrough">(0040,A043)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_660c7236-b5b5-43cc-8f7e-4727bb8d3dda"><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3c957764-7088-4342-9d9a-6fc4d34ee242"><emphasis role="bold"><emphasis role="strikethrough">A concept that constrains the meaning of (i.e., defines the role of) the Observation Value. The "Name" component of a Name/Value pair.</emphasis></emphasis></para>
                                    <para xml:id="para_df45a918-a9b2-44e7-a80d-a2445c67f940"><emphasis role="bold"><emphasis role="strikethrough">Only a single Item shall be included in this Sequence.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_181c7b37-0dee-4eda-8e1b-5263a0e9a389">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                            </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_469a088e-5bfd-4fd3-95f9-fe1c6171d45d">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">No Baseline CID is defined.</emphasis></emphasis></emphasis>
                            </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f2899877-dd46-490b-b3ad-4bd729941424">&gt;Referenced Frame Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_055a8812-e96b-4d88-9d39-ffdb46a34dc1">(0008,1160)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fbc39d50-4108-407c-8341-046371c04c98">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6d70950f-90ee-40c0-83aa-358e130294ef">References one or more frames in a Multi-frame SOP Instance. The first frame shall be denoted as frame number one.</para>
                                    <para xml:id="para_343dd750-60ea-4c07-9774-fad2819dd287">Required if this SOP Instance is a Multi-frame SOP Instance and the values in this Sequence Item do not apply to all frames<emphasis role="bold"><emphasis role="underline">, and Referenced Segment Number (0062,000B) is not present</emphasis></emphasis>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;Referenced Segment Number</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0062,000B)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para>Identifies the segments to which the reference applies identified by Segment Number (0062,0004).</para>
                                    <para>Required if this SOP Instance is a Segmentation or Surface Segmentation and the values in this Sequence Item do not apply to all segments and Referenced Frame Number (0008,1160) is not present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">&gt;Referenced Waveform Channels</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">(0040,A0B0)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">List of channels in Waveform to which the reference applies. See <xref linkend="sect_C.18.5.1.1" xrefstyle="select: label"/></emphasis></emphasis>
                                    </para>
                                    <para><emphasis role="bold"><emphasis role="underline">Required if this SOP Instance is a Waveform that contains multiple Channels and the values in this Sequence Item do not apply to all Waveform Channels.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c7e8c8f6-6c96-425b-8bf6-4fabc03427d6"><emphasis role="bold"><emphasis role="strikethrough">&gt;Numeric Value</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ebb22b0b-0c03-4a5c-b79e-7a503d8a1289"><emphasis role="bold"><emphasis role="strikethrough">(0040,A30A)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_589f2de2-cc6c-4597-9a91-0f23a4c8131e"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_897a3d35-c8cb-43e8-9c6e-f595dea3dcc2"><emphasis role="bold"><emphasis role="strikethrough">This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a set of one or more numeric values.</emphasis></emphasis></para>
                                    <para xml:id="para_4833dffa-ed3b-4461-bbca-907cf02914b9"><emphasis role="bold"><emphasis role="strikethrough">Required if the value that Concept Name Code Sequence (0040,A043) requires (implies) is a set of one or more integers or real numbers. Shall not be present otherwise.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3d95dbc2-4f4e-4526-a648-b819a95c1a50"><emphasis role="bold"><emphasis role="strikethrough">&gt;Floating Point Value</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d9402f07-d2a6-49f1-a1df-1ed0fb412d93"><emphasis role="bold"><emphasis role="strikethrough">(0040,A161)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2ee0afca-d77f-4a3f-b360-7dcbc9f56c8a"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_edc34807-f455-4f46-a278-ce535a369850"><emphasis role="bold"><emphasis role="strikethrough">The floating point representation of Numeric Value (0040,A30A). The same number of values as Numeric Value (0040,A30A) shall be present.</emphasis></emphasis></para>
                                    <para xml:id="para_dbf6035d-61d8-49b8-a959-e3f3b47c855a"><emphasis role="bold"><emphasis role="strikethrough">Required if Numeric Value (0040,A30A) has insufficient precision to represent the value as a string. May be present otherwise.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4e1302fc-07bd-4219-bfb1-a59c1b6f1f08"><emphasis role="bold"><emphasis role="strikethrough">&gt;Rational Numerator Value</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f753607e-9878-4f99-ab40-78d74e3b92fe"><emphasis role="bold"><emphasis role="strikethrough">(0040,A162)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d0dd386f-2b18-4b60-a16e-49d7c1b4ccf3"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_cbaa0a2c-1965-4baf-ac5c-c0fa4f89e91f">The integer numerator of a rational representation of Numeric Value (0040,A30A). Encoded as a signed integer value. The same number of values as Numeric Value (0040,A30A) shall be present.</para>
                                </emphasis></emphasis><para xml:id="para_9bd6c3b6-ca1e-44fc-9dd8-0a0f5da0b03e"><emphasis role="bold"><emphasis role="strikethrough">Required if Numeric Value (0040,A30A) has insufficient precision to represent a rational value as a string. May be present otherwise.</emphasis></emphasis></para></td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_926cb3aa-2fc9-48db-8b80-c2a05392f164"><emphasis role="bold"><emphasis role="strikethrough">&gt;Rational Denominator Value</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8ba2703b-7cb7-4884-8a93-5f80ccc5750a"><emphasis role="bold"><emphasis role="strikethrough">(0040,A163)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ba0e3cba-d6b9-4dbe-ad7e-2417b5a49d1d"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_74371d56-cdba-4bcf-ab73-94e7facef005"><emphasis role="bold"><emphasis role="strikethrough">The integer denominator of a rational representation of Numeric Value (0040,A30A). Encoded as a non-zero unsigned integer value. The same number of values as Numeric Value (0040,A30A) shall be present.</emphasis></emphasis></para>
                                    <para xml:id="para_45fdad96-31f9-42d4-9cf0-a245af5309e2"><emphasis role="bold"><emphasis role="strikethrough">Required if Rational Numerator Value (0040,A162) is present.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6c543c44-274a-45c0-baaf-be6d6cba148d"><emphasis role="bold"><emphasis role="strikethrough">&gt;Measurement Units Code Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2c5d408a-8b18-4f6a-9b5b-f2ee520ea19b"><emphasis role="bold"><emphasis role="strikethrough">(0040,08EA)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_153d5b16-b207-4e19-a755-3a3c7590ba06"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_86f9404e-493b-45b4-a809-e5496cb59f76"><emphasis role="bold"><emphasis role="strikethrough">Units of measurement.</emphasis></emphasis></para>
                                    <para xml:id="para_4602d913-70e5-4dba-b9ce-a4b5e1c491bc"><emphasis role="bold"><emphasis role="strikethrough">Only a single Item shall be included in this Sequence.</emphasis></emphasis></para>
                                    <para xml:id="para_b143a92c-76b1-4552-87dc-8bfaafe03a8a"><emphasis role="bold"><emphasis role="strikethrough">Required if Numeric Value (0040,A30A) is present. Shall not be present otherwise.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_301ded9f-062f-4ffe-b0b5-87dda8afc7f8">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                            </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_542e455c-2659-453d-beb0-0dc39c73f4ae">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_82" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis></emphasis></emphasis>
                            </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3477adcb-d68c-4d4e-ba4e-1e9c87fcc7a2"><emphasis role="bold"><emphasis role="strikethrough">&gt;Date</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_2b6f779b-ee4e-4c06-906a-d3e29e446244">(0040,A121)</para>
                                </emphasis></emphasis></td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0d35594c-da87-4206-a7ce-cdb1bc6a50b1"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_abb886c1-f3a2-4078-a1cf-77290e822406"><emphasis role="bold"><emphasis role="strikethrough">This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a date.</emphasis></emphasis></para>
                                    <note>
                                        <para xml:id="para_3da4ec3b-efe8-40a6-9c46-16bb7c8b79ca"><emphasis role="bold"><emphasis role="strikethrough">The purpose or role of the date value could be specified in Concept Name Code Sequence (0040,A043).</emphasis></emphasis></para>
                                    </note>
                                    <para xml:id="para_02f7ad1c-65c5-4fcd-89bf-f65c10f2d00f"><emphasis role="bold"><emphasis role="strikethrough">Required if the value that Concept Name Code Sequence (0040,A043) requires (implies) is a date. Shall not be present otherwise.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a4263653-e18c-44b6-88fb-aab230ba1dd8"><emphasis role="bold"><emphasis role="strikethrough">&gt;Time</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_aa046eb8-6752-4db6-9e42-a476c30ae4fc">(0040,A122)</para>
                                </emphasis></emphasis></td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_74a28c5b-07f8-429b-bb1c-d8de5a4a7d4d">1C</para>
                                </emphasis></emphasis></td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c538ff7d-c490-4b2e-8afb-76c5d42b91b0"><emphasis role="bold"><emphasis role="strikethrough">This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a time.</emphasis></emphasis></para>
                                    <note>
                                        <para xml:id="para_090933a4-c189-49db-a4ad-938040f8b381"><emphasis role="bold"><emphasis role="strikethrough">The purpose or role of the time value could be specified in Concept Name Code Sequence (0040,A043).</emphasis></emphasis></para>
                                    </note>
                                    <para xml:id="para_41802b01-2f5b-4d9c-a60e-ac04e088cef5"><emphasis role="bold"><emphasis role="strikethrough">Required if the value that Concept Name Code Sequence (0040,A043) requires (implies) is a time. Shall not be present otherwise.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f793b030-6e76-4e91-b8a1-304b83fd0937"><emphasis role="bold"><emphasis role="strikethrough">&gt;Person Name</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ae49dfa0-cca8-4b91-a59f-b9f93a8e48d6"><emphasis role="bold"><emphasis role="strikethrough">(0040,A123)</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e6913c1e-65f7-48ed-9bea-97de900a2831"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0ef38c62-ab6d-46cb-acf6-a3d65b645c70"><emphasis role="bold"><emphasis role="strikethrough">This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a Person Name.</emphasis></emphasis></para>
                                    <note>
                                        <para xml:id="para_a8c76dec-9c64-427d-9278-7ac6f369d5a7"><emphasis role="bold"><emphasis role="strikethrough">The role of the person could be specified in Concept Name Code Sequence (0040,A043).</emphasis></emphasis></para>
                                    </note>
                                    <para xml:id="para_1e3eb2cf-6b7a-47c6-a22f-29fe4e2ce4c1"><emphasis role="bold"><emphasis role="strikethrough">Required if the value that Concept Name Code Sequence (0040,A043) requires (implies) is a person name. Shall not be present otherwise.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_50b86a72-4b0d-429b-991b-a2b2f7e22f58"><emphasis role="bold"><emphasis role="strikethrough">&gt;Text Value</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_77fa08c1-40a8-452f-ad9f-e3a77e0bb7bc">(0040,A160)</para>
                                </emphasis></emphasis></td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_b27adb69-dd40-4fcf-9228-7fd810eb9f21">1C</para>
                                </emphasis></emphasis></td>
                                <td align="left" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_a67d9f27-6d29-4dfa-a66f-f3c4ab93a378">This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a Text Observation Value.</para>
                                </emphasis></emphasis><para xml:id="para_d3eb390b-62fd-44f6-bd0f-697c1acdefb7"><emphasis role="bold"><emphasis role="strikethrough">Required if Date (0040,A121), Time (0040,A122), and Person Name (0040,A123) do not fully describe the concept specified by Concept Name Code Sequence (0040,A043). Shall not be present otherwise.</emphasis></emphasis></para></td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7ef1e9bb-3e7c-4480-b141-f852adee49bf"><emphasis role="bold"><emphasis role="strikethrough">&gt;Concept Code Sequence</emphasis></emphasis></para>
                                </td>
                                <td align="center" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">
                                    <para xml:id="para_b29decaf-44c6-4270-bf5a-433c39c46f81">(0040,A168)</para>
                                </emphasis></emphasis></td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_eb14ab5f-e576-49aa-b6e6-8cf71bde6b06"><emphasis role="bold"><emphasis role="strikethrough">1C</emphasis></emphasis></para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_18b32665-0d3a-4cae-8cb8-3c642060c966"><emphasis role="bold"><emphasis role="strikethrough">This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a Coded Value.</emphasis></emphasis></para>
                                    <para xml:id="para_598eefb4-2f14-405d-8f4c-131af0880d17"><emphasis role="bold"><emphasis role="strikethrough">Only a single Item shall be included in this Sequence.</emphasis></emphasis></para>
                                    <para xml:id="para_43b54320-c145-46d2-881f-9fe8c1d201aa"><emphasis role="bold"><emphasis role="strikethrough">Required if Date (0040,A121), Time (0040,A122), Person Name (0040,A123), Text Value (0040,A160), and the pair of Numeric Value (0040,A30A) and Measurement Units Code Sequence (0040,08EA) are not present.</emphasis></emphasis></para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_e36c2e5e-1321-4018-9a69-ade9286e4045">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                            </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3bd73b01-4f73-424b-ae67-d94e11c5be66">
                                <emphasis role="bold"><emphasis role="strikethrough"><emphasis role="italic">No Baseline CID is defined.</emphasis></emphasis></emphasis>
                            </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_03fff138-ae8e-4dbd-84f6-e6562e2545e6">Acquisition Context Description</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_704eaa21-d7ca-49d3-b172-73233c61981b">(0040,0556)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_efcd0b05-582b-4ae6-bb4e-84c4ee9467b7">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bb506c03-9c19-4b49-b5d4-a48755c207b9">Free-text description of the image-acquisition context.</para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
                 
         
<section label="10.2" status="2" xml:id="sect_10.2">
      <title>Content Item Macro</title>
      <para xml:id="para_56f4881b-f853-4bc0-b24d-97609576bea3">A Content Item is a flexible means of encoding attribute identifiers and attribute values using the Code Sequence Macro (see <xref linkend="chapter_8" xrefstyle="template:Section %n"/>) for coded terminology defined by a coding scheme. The Content Item provides a name-value pair, i.e., a Concept Name, encoded as a Code Sequence, and a Concept Value. The Concept Value may be encoded by any of a set of generic Attributes, as specified by a Value Type, including text, personal name, numeric, and coded concept (Code Sequence) values.</para>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_f66f9b16-8a08-4a1b-b464-eece0e169312">Comparing a Content Item to a native DICOM Data Element, the Concept Name Code Sequence corresponds to the Data Element Tag and Attribute Name, the Value Type to the Value Representation, and the Concept Value to the Data Element Value Field. See <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>.</para>
          </listitem>
          <listitem>
            <para xml:id="para_9ce69581-009f-4fdb-8ec4-e57ef7564b3f">The IMAGE Value Type of this Macro does not include the Type 3 Attributes of the IMAGE Value Type defined in <xref linkend="sect_C.17.3" xrefstyle="select: label"/>, as they are not required for Acquisition Context or Protocol Context Content Items.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_20a81c02-8c2e-47e9-9a2f-88035129c1cf">Specific uses of the Content Item may invoke the Content Item Macro defined in this Section, the Document Content Macro of <xref linkend="sect_C.17.3" xrefstyle="select: label"/>, or another similar construct. An invocation of the Content Item Macro may constrain the allowed values of Value Type (0040,A040).</para>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_929250e1-1057-4869-a809-ca33a0a27f8d">The NUMERIC Value Type of this Macro differs from the NUM Value Type defined in <xref linkend="sect_C.17.3" xrefstyle="select: label"/>, since the encoding of the Concept Value is different.</para>
          </listitem>
          <listitem>
            <para xml:id="para_fe5344df-271f-4a9a-8b99-7dbcb03c5a73">The Value Type uses Enumerated Values so as to assure that non-standard Value Types are not used, and to prevent the nefarious use, for example, of a CONTAINER Value Type in an SR-like manner to create nested content, which is not the intent.</para>
          </listitem>
          <listitem>
            <para xml:id="para_09bfe572-9d23-45d1-ab4f-901e5e5c887b">Some invocations of this Macro may use the Content Item Modifier Sequence (0040,0441) to achieve a single level of "nesting". That Attribute is not included in this Macro itself, to prevent recursive inclusion.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_f615928a-ea2c-4401-a36c-c031595525d7">See <xref linkend="sect_5.4" xrefstyle="select: label"/> for the meaning of the Type column in this Macro when applied to Normalized IODs.</para>
      <table frame="box" label="10-2" rules="all" xml:id="table_10-2">
        <caption>Content Item Macro Attributes Description</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a1ae3b7f-7a70-467d-b25b-d14051b2355b">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f9d09816-a1cd-4bc5-ad0b-f678f03d8d81">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ab537a4a-89ae-4c36-a2aa-b6b9ab3efd68">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_7fa5a25c-af99-4e56-9313-0ac90f930498">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22306318-5576-4c7b-b5b0-30b4460e1db6">Value Type</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1db72e92-01b2-4fac-9991-9896839be997">(0040,A040)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_e50005bd-af68-4806-96cc-32a3d74efb6c">1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7666b6cb-1d40-487f-86ed-d60ff7ac3657">The type of the value encoded in this name-value Item.</para>
              <variablelist spacing="compact">
                <title>Enumerated Values:</title>
                <varlistentry>
                  <term>DATE</term>
                  <listitem>
                    <para xml:id="para_9037ade1-2c70-46c6-ba40-886642d9bbda"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>TIME</term>
                  <listitem>
                    <para xml:id="para_8af133c8-d719-4527-af71-95c13037b5d8"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>DATETIME</term>
                  <listitem>
                    <para xml:id="para_9c96cfd4-5228-4d58-8900-a569d0e86620"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>PNAME</term>
                  <listitem>
                    <para xml:id="para_830586a4-5d0a-462f-ab83-e9862046dee8"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>UIDREF</term>
                  <listitem>
                    <para xml:id="para_4b399761-8325-42d9-b7b0-0bad2e501ce7"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>TEXT</term>
                  <listitem>
                    <para xml:id="para_5bffa125-a196-4f11-9d0d-eb912abd718f"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>CODE</term>
                  <listitem>
                    <para xml:id="para_956b33d0-fb41-4ef4-9159-202cda0e98b6"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>NUMERIC</term>
                  <listitem>
                    <para xml:id="para_36307d59-1880-47f1-9c8f-794512fd0e88"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>COMPOSITE</term>
                  <listitem>
                    <para xml:id="para_c7f547ce-a0db-42f4-8543-e8ece5743861"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>IMAGE</term>
                  <listitem>
                    <para xml:id="para_bca62f4b-333d-49f5-92a3-04f896bfc193"/>
                  </listitem>
                </varlistentry>
                  <varlistentry>
                      <term><emphasis role="bold"><emphasis role="underline">WAVEFORM</emphasis></emphasis></term>
                      <listitem>
                          <para/>
                      </listitem>
                  </varlistentry>
              </variablelist>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_815bf631-fc36-4ab1-875c-4668ff689337">Observation DateTime</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_3ebfcd85-2abe-45f2-bf62-3f17e3d4a90e">(0040,A032)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_903b5950-83b8-4ada-895c-59e6327917cf">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7678c02a-135d-4aa5-9499-abcc9993294b">The date and time on which this Item was completed. For the purpose of recording measurements or logging events, completion time is defined as the ending time of data acquisition of the measurement, or the ending time of occurrence of the event.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e979376f-46ee-4ea5-adf1-88149578b46c">Concept Name Code Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_10424561-7c14-4d66-a171-cf1f4fad8824">(0040,A043)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a4dca359-f820-4ee9-ab77-3b8ca9529a5f">1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a884240d-69e6-40ab-81b9-3cd7ee4701d6">Coded concept name of this name-value Item.</para>
              <para xml:id="para_20722ee5-c6dc-4867-be2b-b4bdab901c20">Only a single Item shall be included in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_523aa1a5-2a76-402f-b232-63e5983b1f69">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84769935-4466-4326-961a-588667f2a997">
                <emphasis role="italic">No Baseline CID is defined.</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ae63f9e-e356-4381-a6fd-ddbbe1146062">DateTime</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_432a1d01-1721-4000-8996-600e7f0ddeb3">(0040,A120)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a33c1398-f3f6-4022-8e74-ac7eda52565d">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5fb02dc2-e74a-4095-843c-6071cd6f9f98">DateTime value for this name-value Item.</para>
              <para xml:id="para_b411d11c-b787-4bf1-b0bd-12d52ef4744a">Required if Value Type (0040,A040) is DATETIME.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba31a7af-abd1-454c-9d1d-b6b5127b8f38">Date</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_b9077010-9669-4065-acd5-607b3050c742">(0040,A121)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1b925370-0526-49e5-b31a-f0d6ed87112b">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_82ed72ae-b003-4324-a3ca-ef4f66b6e2d0">Date value for this name-value Item.</para>
              <para xml:id="para_f2faa53d-15ce-4986-9c65-e06d80397c84">Required if Value Type (0040,A040) is DATE.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47d923e0-d079-4aa3-8441-d7729dc4f083">Time</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_30065328-c8f7-42bd-bc49-4244d75235be">(0040,A122)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_580500a7-2ebf-4a42-bb5d-6166bbe93387">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac36e13e-bd06-4540-be4d-4849550826f2">Time value for this name-value Item.</para>
              <para xml:id="para_ef93f298-fff5-4d11-ac55-278010cde035">Required if Value Type (0040,A040) is TIME.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c66c61c0-361d-4406-8110-ad59065cbb4d">Person Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_191d8de1-a026-4da0-bcea-6bcecb50c090">(0040,A123)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_00802f3d-bdf9-428c-936b-7f56b3bbba94">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f2292bbd-f738-4ac6-bf7e-95826fcf6b5d">Person name value for this name-value Item.</para>
              <para xml:id="para_e263ad76-caf9-489f-a231-dbf6c9f9bb06">Required if Value Type (0040,A040) is PNAME.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bc1019db-96b8-4b6f-9742-cefbc8791037">UID</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_ba16d6ed-4a6d-43aa-ac5b-bc26434e67f6">(0040,A124)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_ada087a6-2335-4f9b-a82d-38fcbaf3ccd0">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_336fd9e3-690f-4c95-ae5d-a49f8f57f2f8">UID value for this name-value Item.</para>
              <para xml:id="para_2ac44603-a5d4-4ba4-8283-8fc051c24a4b">Required if Value Type (0040,A040) is UIDREF.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89120c2c-ed2c-4b91-b37e-1a4e408b8819">Text Value</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_945e1f92-f52d-4db9-a6b9-a5293f0d38fa">(0040,A160)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6c7ff11d-3305-4d03-abf0-3ef20ace5c84">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_316b0d8d-9939-495e-b31b-c0d129bbb17a">Text value for this name-value Item.</para>
              <para xml:id="para_6b40c5f0-8b6c-49c6-88c0-0a4373905b58">Required if Value Type (0040,A040) is TEXT.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f793bb0-d7a3-40d5-bb80-5b019798fb35">Concept Code Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1852ac09-f8d1-4051-be45-a79aa80032a9">(0040,A168)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_ab89854a-7439-45a5-b5ef-83379a837fb5">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d4a1bbc-469e-4a65-96ab-347193a04253">Coded concept value of this name-value Item.</para>
              <para xml:id="para_bcf4789d-57db-43df-a9ca-bf67ceb13297">Only a single Item shall be included in this Sequence.</para>
              <para xml:id="para_c36c3848-2137-4047-aa77-262f681ed639">Required if Value Type (0040,A040) is CODE.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_197fabf5-778b-452c-91c3-63bbba0ea9c1">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b5d557a0-53f2-491e-98fd-9872445f5417">
                <emphasis role="italic">No Baseline CID is defined.</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1881527-87ac-4caf-bc15-9c35e21630fd">Numeric Value</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_55084fd7-2a4c-4347-9791-783211607660">(0040,A30A)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_fec34e64-da00-4484-910a-ca0d7a9d07c1">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8ff30b04-69ce-4870-ae35-0e937ac14bc7">Numeric value for this name-value Item. Only a single value shall be present.</para>
              <para xml:id="para_580b1fda-317b-4c76-8a77-cffb08257554">Required if Value Type (0040,A040) is NUMERIC.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_57a14e73-8053-4473-98f9-5c6e8086098e">Floating Point Value</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_fdc7138a-bdcf-402f-b0ee-c059d5ca332f">(0040,A161)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_fb2090f6-e75e-4732-8f6d-430411fc41d7">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3117416b-3122-48b3-a382-09a68d43c34f">The floating point representation of Numeric Value (0040,A30A). The same number of values as Numeric Value (0040,A30A) shall be present.</para>
              <para xml:id="para_36235ebc-1b68-4e25-9261-cd14a264dabb">Required if Numeric Value (0040,A30A) has insufficient precision to represent the value as a string. May be present otherwise.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39c93077-caa9-476c-a27d-11a65c014770">Rational Numerator Value</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_2345397e-a9f4-4020-91ca-b7eae613e787">(0040,A162)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_daeba188-6c9c-4ac0-899d-03bd1f4aa7b3">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca64e9e5-4680-44b4-8fe5-7209c77ddbbd">The integer numerator of a rational representation of Numeric Value (0040,A30A). Encoded as a signed integer value. The same number of values as Numeric Value (0040,A30A) shall be present.</para>
              <para xml:id="para_fbbcc1c4-dcb7-405c-bf40-b6c3d307739a">Required if Numeric Value (0040,A30A) has insufficient precision to represent a rational value as a string. May be present otherwise.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e729a836-61ab-425f-8b26-4134770b66f3">Rational Denominator Value</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_11b7ab32-a907-4b16-9363-6945199aad90">(0040,A163)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_e9c30f63-b1e4-4c4a-bd9f-71fe70928c30">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_09362d46-4166-4535-a4ab-36403ac90176">The integer denominator of a rational representation of Numeric Value (0040,A30A). Encoded as a non-zero unsigned integer value. The same number of values as Numeric Value (0040,A30A) shall be present.</para>
              <para xml:id="para_c789fa67-b91f-4871-bc53-8806a40a922b">Required if Rational Numerator Value (0040,A162) is present.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3cf6508a-ff0d-4aba-8e92-459889355b2d">Measurement Units Code Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_7a6cb5c8-0da2-4ae1-8f7c-ef987c56305b">(0040,08EA)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_05d39636-c44e-4d0f-80a9-4ebd81b8c30d">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc035e6a-0ef2-49e9-8866-ccc571eac53a">Units of measurement for a numeric value in this name-value Item.</para>
              <para xml:id="para_28789102-5c2f-4c54-afa2-c3255e38084e">Only a single Item shall be included in this Sequence.</para>
              <para xml:id="para_701daee6-5df6-44f7-9461-9cf3f64d9375">Required if Value Type (0040,A040) is NUMERIC.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_7f68e8a3-c2b3-4d5f-bf9a-2a3ee15b8036">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_80580a55-23db-4e64-afc6-764dc32fbb8e">
                <emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_82" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb1fe82c-1475-441e-8ae5-46fc1cd04241">Referenced SOP Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_3bba6a22-f309-45a9-b9ce-9c51d5af2ef3">(0008,1199)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_d830d750-d09b-4df7-ad0a-0335080c23ea">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f912cbb-2451-4244-919c-9b8e6fbdc84e">Composite SOP Instance Reference value for this name-value Item.</para>
              <para xml:id="para_1b873a9f-e333-44ec-8faf-4870ac84e2d2">Only a single Item shall be included in this Sequence.</para>
              <para xml:id="para_cf02749f-c86a-427c-b083-6b49f1d2187f">Required if Value Type (0040,A040) is COMPOSITE or IMAGE<emphasis role="bold"><emphasis role="underline"> or WAVEFORM</emphasis></emphasis>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_91233428-7eb2-4efd-9ca1-46155bea7f2f">
                <emphasis role="italic">&gt;Include Table 10-11 "SOP Instance Reference Macro Attributes"<!--<xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf16498e-e81e-404f-9eed-6eec71c1c695"/>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_668b013c-10e7-4d7b-9099-99956476da29">&gt;Referenced Frame Number</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_87f1b33e-c198-42ec-b705-4556e598f4f1">(0008,1160)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_82461c1a-166d-47a6-b87a-f731ce57857d">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fd95aa7-dbbb-4e8e-84ca-757929e8b0a4">Identifies the frame numbers within the Referenced SOP Instance to which the reference applies. The first frame shall be denoted as frame number 1.</para>
              <note>
                <para xml:id="para_e06e8e0c-e051-4fbc-bfa9-82d4a32dd706">This Attribute may be multi-valued.</para>
              </note>
              <para xml:id="para_730b926a-9fd7-40d0-ade0-2880da297272">Required if the Referenced SOP Instance is a multi-frame image and the reference does not apply to all frames, and Referenced Segment Number (0062,000B) is not present.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f957fcb6-cf6a-45b7-ae0d-2914c6a908ae">&gt;Referenced Segment Number</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_337456b7-54f9-4ab9-8dcc-84d139f75976">(0062,000B)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_b51d7f7b-439e-4069-88d3-58191db445a6">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_204fe166-92db-45aa-b202-7e347d793689">Identifies the segments to which the reference applies identified by Segment Number (0062,0004).</para>
              <para xml:id="para_fb2d3862-8f78-49c1-b853-58f898004de8">Required if the Referenced SOP Instance is a Segmentation or Surface Segmentation and the reference does not apply to all segments and Referenced Frame Number (0008,1160) is not present.</para>
            </td>
          </tr>
            <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">&gt;Referenced Waveform Channels</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">(0040,A0B0)</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">List of channels in Waveform to which the reference applies. See <xref linkend="sect_C.18.5.1.1" xrefstyle="select: label"/></emphasis></emphasis>
                            </para>
                    <para><emphasis role="bold"><emphasis role="underline">Required if the Referenced SOP Instance is a Waveform that contains multiple Channels and not all Channels in the Waveform are referenced.</emphasis></emphasis></para>
                </td>
            </tr>
        </tbody>
      </table>
        <section label="10.2.1" status="2" xml:id="sect_10.2.1">
            <title>Content Item With Modifiers Macro</title>
            <para xml:id="para_39c41a70-d068-4802-abbe-3e577e68a475">Content Item with Modifiers is a means of describing structured content which needs a Content Item with single optional level of modifiers,
                i.e. a two-level structure of Content Items.
                An invocation of the Content Item with Modifiers Macro will usually specify the allowed values using a Protocol Context Template in <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>,
                which allows a single Nesting Level (see in <olink targetdoc="PS3.16" targetptr="sect_6.1.2" xrefstyle="select: docname label quotedtitle"/>).
                Constraints on the use of this Macro may be specified in <olink targetdoc="PS3.16" targetptr="chapter_C" xrefstyle="template:PS3.16 Annex %n"/>, which may be invoked in IODs in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>.</para>
            <table frame="box" label="10.2.1-1" rules="all" xml:id="table_10.2.1-1">
                <caption>Content Item with Modifiers Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0e5e077f-e05a-4d09-9ff8-fab37b223d5a">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5bd16b08-f31f-4422-8fda-99b09c3e0896">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3a86ff7a-148f-498b-83f0-6d51307f213f">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_45425ab9-731f-43c5-ac66-3f80542eac1a">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_9751d0cd-4246-46a0-8887-0e75503665c0">
                <emphasis role="italic">Include <xref linkend="table_10-2" xrefstyle="select: label quotedtitle"/>
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_52761ef2-6131-4731-b8a0-3af2551aabb8">
                                <emphasis role="italic">No Baseline TID is defined.</emphasis>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_344896a5-21e5-4c58-a8e7-7dfc5a14063c">Content Item Modifier Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_32447748-7742-46b0-aea6-38e4b6522a8e">(0040,0441)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_af97c074-0136-4432-9ac6-26d317564153">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_470d82ef-4a51-4992-964d-c73419628ee3">Specifies modifiers for the Content Item.</para>
                            <para xml:id="para_d81c963b-3451-4b9e-ba3d-7bfb4197260e">One or more Items are permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_00282c0a-9d44-49d4-9e03-b49db3d51191">
                <emphasis role="italic">&gt;Include <xref linkend="table_10-2" xrefstyle="select: label quotedtitle"/>
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bb041cef-0b30-4795-8e17-00f048a36f8b">
                                <emphasis role="italic">No Baseline TID is defined.</emphasis>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
    </section>



        
        <section label="C.18.3" status="3" xml:id="sect_C.18.3">
            <title>Composite Object Reference Macro</title>
            <para xml:id="para_524ab697-1ba1-4cc3-a936-ca62aba9a7e5">
                <xref linkend="table_C.18.3-1" xrefstyle="select: label"/> specifies the Attributes that convey a reference to a DICOM Composite Object that is not a DICOM Image or Waveform (such as an SR Document), or to an HL7 Structured Document.</para>
            <note>
                <orderedlist numeration="arabic">
                    <listitem>
                        <para xml:id="para_0029a6d2-7519-4870-b038-4a888e0f0e4a">If a Softcopy Presentation State is to be applied to an Image, it should be referenced by an Image Reference Macro.</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_a34717df-ce10-4e12-90c5-fca82fdbbffa">Other SR Documents may be referenced by this Macro, but there is no facility to reference individual Content Items within those reports.</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_00994dcc-2d14-4ac6-af9d-a7c92811b880">HL7 Structured Documents include, in particular, those conforming to the Clinical Document Architecture (CDA). See <xref linkend="sect_C.12.1.1.6" xrefstyle="select: label"/> for further details about this type of referenced object.</para>
                    </listitem>
                </orderedlist>
            </note>
            <table frame="box" label="C.18.3-1" rules="all" xml:id="table_C.18.3-1">
                <caption>Composite Object Reference Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8f4297d9-1c0a-4a4c-8e38-068307b2e8f7">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_da4e51f9-f802-4d2d-9421-c8d2ffcb079c">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ae8f2f0-d3a9-4367-9d14-1ea97c2b59ca">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d25a02b9-becd-42b0-9ff2-c999b738db09">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_566107c7-d029-426a-9921-2955fbb586bd">Referenced SOP Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7838824a-efb9-4024-b7f7-7ffa53a39deb">(0008,1199)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fcb5fc7e-4c0f-4656-a9ee-3f639eb9dda2">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_26472c12-47fb-40f1-9a53-29791a830825">References to Composite Object SOP Class/SOP Instance pairs.</para>
                            <para xml:id="para_a69eb9b5-0964-4397-a8a8-94494e9520c3">Only a single Item shall be included in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_a1c8b0ed-8e8c-40ba-b3f7-43cf35c50eb5">
                                <emphasis role="italic">&gt;Include Table 10-11 "SOP Instance Reference Macro Attributes"<!--<xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>-->
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_05a183e0-be6b-41a6-a887-5c45b96b3ad5"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        <section label="C.18.4" status="3" xml:id="sect_C.18.4">
            <title>Image Reference Macro</title>
            <para xml:id="para_84a7c6dc-c0f8-4b04-a1d0-49d5573df9c7">
                <xref linkend="table_C.18.4-1" xrefstyle="select: label"/> specifies the Attributes that convey a reference to a DICOM image.</para>
            <table frame="box" label="C.18.4-1" rules="all" xml:id="table_C.18.4-1">
                <caption>Image Reference Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d6dbb1f9-b774-4cb8-ad53-90dac3721e42">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_817c708d-7e35-4a14-8e03-f47121a866ee">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2a72ef80-b107-42dd-bbdb-e5ad3aff8815">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6c90b364-b67c-4f7e-84c2-a57e0dc0f525">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_0dc1f7ef-8e26-4665-b95f-d7c92d7ab07a">
                                <emphasis role="italic">Include <xref linkend="table_C.18.3-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e3583281-f23b-43ce-b922-1f0330296cf8"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0dc8172e-fa14-4d62-9381-c6467e544221">&gt;Referenced Frame Number</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c8213cca-8a47-49a6-b8e0-fb2f27260725">(0008,1160)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b1d4a2d2-469f-4439-b479-70937708d4f3">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_97bd56ed-4444-45bd-ac90-66a56b03ff61">Identifies the frame numbers within the Referenced SOP Instance to which the reference applies. The first frame shall be denoted as frame number 1.</para>
                            <note>
                                <para xml:id="para_c572337d-ddf5-4f67-a971-4f9655f578db">This Attribute may be multi-valued.</para>
                            </note>
                            <para xml:id="para_87de9970-31fc-4115-a5bb-55961ee24f35">Required if the Referenced SOP Instance is a multi-frame image and the reference does not apply to all frames, and Referenced Segment Number (0062,000B) is not present.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_54bcc5f8-8eac-4867-921c-db320e4b718b">&gt;Referenced Segment Number</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d84e9066-39ad-4a86-b544-3bac9ee8e2ea">(0062,000B)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5709abb1-7009-4676-918d-964fe3f378c8">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e619d844-cbd4-45c2-9b17-03ac8ea515f1">Identifies the segments to which the reference applies identified by Segment Number (0062,0004). Required if the Referenced SOP Instance is a Segmentation and the reference does not apply to all segments and Referenced Frame Number (0008,1160) is not present.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_dd2abdb7-ce9e-42b2-a0b7-bd7a8ccac9b7">&gt;Referenced SOP Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7e33545b-5d71-4b8b-92ba-8d6e947cb718">(0008,1199)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0bed1081-1921-4900-99ba-85f6190c27b9">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0bce67a9-7d9f-4ce7-b1a7-c872428a6ad5">Reference to a Softcopy Presentation State SOP Class/SOP Instance pair.</para>
                            <para xml:id="para_78449297-e885-4c12-8526-24b6b6278d25">Only a single Item is permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_0d8edcf1-2f20-4e0a-adb3-4b6caa2c183a">
                                <emphasis role="italic">&gt;&gt;Include Table 10-11 "SOP Instance Reference Macro Attributes"<!--<xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>-->
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a44f1091-e180-44ec-8d71-ed9430506080"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7113faa5-c2f2-49cf-b317-c1291022c0d0">&gt;Referenced Real World Value Mapping Instance Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b75a952e-b8b7-4f8b-a08d-27442b3a192d">(0008,114B)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4406b31f-510c-40d2-bcde-de9e4247db6e">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_dff88456-f4e3-4f74-b753-ea57b007ae81">Reference to a Real World Value Mapping SOP Class/SOP Instance pair.</para>
                            <para xml:id="para_ba0e3613-3c98-4c8f-a4b0-c9b34a95b440">Only a single Item is permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_0ced5b7d-7578-4219-8530-3670ebc51518">
                                <emphasis role="italic">&gt;&gt;Include Table 10-11 "SOP Instance Reference Macro Attributes"<!--<xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>-->
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a297a8e3-4445-4a3d-8ed6-e8d149cc1ce3"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_234d526b-d5c0-4c06-b6e7-6e2701433365">&gt;Icon Image Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_50d10993-aff0-492f-83a1-ce8c5edb849b">(0088,0200)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a31e85ae-02c9-4c32-95b6-ef955afed1c0">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9522a6e6-06a9-4901-8637-a660e1839642">This Icon Image is representative of the Image.</para>
                            <para xml:id="para_8b207e7e-d936-4b32-8d9f-64fc5164f5f2">Only a single Item is permitted in this Sequence.</para>
                            <para xml:id="para_9594ce18-71f5-40ce-929e-28c37dab2d9c">The Icon Image shall be no greater than 128 rows by 128 columns.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_598196fa-415e-49b1-940f-3e81c4636488">
                                <emphasis role="italic">&gt;&gt;Include <xref linkend="table_C.7-11b" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1de21502-4f17-4dd6-a90d-4f211ec92368">
                                <emphasis role="italic">See <xref linkend="sect_C.7.6.1.1.6" xrefstyle="select: label"/>
                                </emphasis>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        <section label="C.18.5" status="3" xml:id="sect_C.18.5">
            <title>Waveform Reference Macro</title>
            <para xml:id="para_5cbdacff-056f-4ca2-9aae-b48fc606da69">
                <xref linkend="table_C.18.5-1" xrefstyle="select: label"/> specifies the Attributes that convey a reference to a DICOM waveform.</para>
            <table frame="box" label="C.18.5-1" rules="all" xml:id="table_C.18.5-1">
                <caption>Waveform Reference Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b39cbab7-a2e8-48cc-b679-3431588959c4">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f55b47ef-be06-47ff-90e2-0240736b002e">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ba8301bb-d1cc-47a5-8677-5a69b4dfac81">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1272c819-cef6-449a-99d6-3aecbe0f0c8f">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_5b2fe6ce-aac0-4d74-933c-2b3240aeae1e">
                                <emphasis role="italic">Include <xref linkend="table_C.18.3-1" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5a2e66dc-ee50-4752-b6de-c728d72b546e"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2b92db12-0499-4c81-83d0-0139039a336b">&gt;Referenced Waveform Channels</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_19bad1fc-24d4-4dc9-afe5-240d5b0ffbad">(0040,A0B0)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4ac64a23-fc9a-406d-90cc-e5c7dc747d48">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c18ddec2-87ba-4a4e-af8f-ed953879062b">List of channels in Waveform to which the reference applies. See <xref linkend="sect_C.18.5.1.1" xrefstyle="select: label"/>
                            </para>
                            <para xml:id="para_78e8ba3a-2749-40ac-9e18-ba36ba999e24">Required if the Referenced SOP Instance is a Waveform that contains multiple Channels and not all Channels in the Waveform are referenced.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.18.5.1" status="4" xml:id="sect_C.18.5.1">
                <title>Waveform Reference Macro Attribute Descriptions</title>
                <section label="C.18.5.1.1" status="5" xml:id="sect_C.18.5.1.1">
                    <title>Referenced Waveform Channels</title>
                    <para xml:id="para_77a7ef9c-2233-4f5c-8807-af7f2be95dfb">Referenced Waveform Channels (0040,A0B0) is a multi-value Attribute that lists the channels referenced. Each channel is specified as a pair of values (M,C), where the first value is the Item number of Waveform Sequence (5400,0100) in the referenced object (i.e., the Multiplex Group Number), and the second value is the Item number of Channel Definition Sequence (003A,0200) (i.e., the Channel Number) within the multiplex group.</para>
                    <para xml:id="para_89776678-4541-46d7-a05c-1169eddd4ac1">If the specified channel number is 0, the annotation applies to all channels in the multiplex group.</para>
                    <note>
                        <para xml:id="para_4114272c-4711-42fb-b94e-2e680c7a8d17">As an example, an annotation that applies to the entire first multiplex group and channels 2 and 3 of the third multiplex group would have Referenced Waveform Channels (0040,A0B0) value 0001 0000 0003 0002 0003 0003.</para>
                    </note>
                </section>
            </section>
        </section>
        
        <section label="C.18.7" status="3" xml:id="sect_C.18.7">
            <title>Temporal Coordinates Macro</title>
            <para xml:id="para_1e8cdc31-a05d-4820-a981-3ee4bba8d2e8">
                <xref linkend="table_C.18.7-1" xrefstyle="select: label"/> specifies the Attributes that convey TCOORD Content Items. A TCOORD Content Item shall be the Source Content Item of one or more SELECTED FROM relationships with one or more SCOORD Content Items, one or more IMAGE Content Items, or one or more WAVEFORM Content Items.</para>
            <note>
                <orderedlist numeration="arabic">
                    <listitem>
                        <para xml:id="para_105fadc3-9a61-49df-bd7d-47353892d981">The same set of temporal coordinates may be selected from more than one single-frame image, or more than one frame of a multi-frame image, or from images and waveforms when the purpose of reference is applicable to multiple objects. For example, the definition of a sampling period at different spatial locations on multiple images and a synchronously acquired waveform.</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_a9719a18-7175-44d7-b55e-a6f534809389">Temporal coordinates may refer to spatial coordinates, which in turn refer to one or more frames or images, for example to indicate a region localized in both time and space.</para>
                    </listitem>
                </orderedlist>
            </note>
            <table frame="box" label="C.18.7-1" rules="all" xml:id="table_C.18.7-1">
                <caption>Temporal Coordinates Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9fa35409-7d80-4327-8ebc-767f1dc17ed0">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ea326b35-edc4-4725-ad48-56d7f513454f">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c276f436-9e8e-4460-8ba2-f883cbf7bfa7">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cf580889-d196-4cda-a210-bcceee683590">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e218b636-1484-4178-aef7-27c3a85c0a30">Temporal Range Type</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_992c64fb-4737-4587-b513-6f8724f8d82b">(0040,A130)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_932738e5-6951-4bff-b86f-7a84f55fb2c1">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f8d5538c-25c5-433b-a296-9654178433c3">See <xref linkend="sect_C.18.7.1.1" xrefstyle="select: label"/> for Enumerated Values.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cab47121-677f-43b2-afbd-d53a24053c2e">Referenced Sample Positions</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a4b3ba4a-7b56-4a5b-af0c-d3ac48a9f6d1">(0040,A132)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fc2cb23e-bfb5-4f3f-a3da-94c0947242ea">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_43929073-d0ba-4237-b12a-0d9d3a7d8022">List of samples within a multiplex group specifying temporal points of the referenced data. Position of first sample is 1.</para>
                            <para xml:id="para_0dea3a37-a62d-40b2-b8af-5886816551c8">Required if the Referenced SOP Instance is a Waveform and Referenced Time Offsets (0040,A138) and Referenced DateTime (0040,A13A) are not present.</para>
                            <para xml:id="para_d98b338c-9dd1-4ead-8b02-b4a6be6df7d6">May be used only if Referenced Channels (0040,A0B0) refers to channels within a single multiplex group.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e5c9dc67-03f5-4d28-8f81-987494933157">Referenced Time Offsets</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6b9c4262-896b-4ed6-9dae-700e0747ba6d">(0040,A138)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_78e8a803-0b9c-4cba-a0d1-1e680843f178">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_02f103a3-a642-446f-9a69-049adbedee47">Specifies temporal points for reference by number of seconds after start of data.</para>
                            <para xml:id="para_2b411dce-96cf-4f0b-9a13-902a7b1051fe">Required if Referenced Sample Positions (0040,A132) and Referenced DateTime (0040,A13A) are not present.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_db5f6bf3-8485-445a-8bd6-b9fe48bd74eb">Referenced DateTime</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_aeb4f072-d49f-4346-8866-a23dc8c70a13">(0040,A13A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_18f14f55-a1c2-44f7-9017-ad2af3706fcb">1C</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_273396d3-8892-467e-9947-5e364866bc03">Specifies temporal points for reference by absolute time.</para>
                            <para xml:id="para_aad576e7-5d17-40f7-a4ea-fbf59fe3fe22">Required if Referenced Sample Positions (0040,A132) and Referenced Time Offsets (0040,A138) are not present.</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.18.7.1" status="4" xml:id="sect_C.18.7.1">
                <title>Temporal Coordinates Macro Attribute Descriptions</title>
                <section label="C.18.7.1.1" status="5" xml:id="sect_C.18.7.1.1">
                    <title>Temporal Range Type</title>
                    <para xml:id="para_7da5ea4f-94bd-4b8c-b75d-c25dcca1abf4">This Attribute defines the type of temporal extent of the region of interest. A temporal point (or instant of time) may be defined by a waveform sample offset (for a single waveform multiplex group only), time offset, or absolute time.</para>
                    <variablelist spacing="compact">
                        <title>Enumerated Values:</title>
                        <varlistentry>
                            <term>POINT</term>
                            <listitem>
                                <para xml:id="para_acef985b-a054-4c6a-8311-2a6ad6b0e129">a single temporal point</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>MULTIPOINT</term>
                            <listitem>
                                <para xml:id="para_80203c8b-21fd-422a-a3bc-ff2eab9a848d">multiple temporal points</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>SEGMENT</term>
                            <listitem>
                                <para xml:id="para_f09ebd8f-2dc4-4c83-95c2-b31acbed8011">a range between two temporal points</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>MULTISEGMENT</term>
                            <listitem>
                                <para xml:id="para_23635790-f5b7-4d69-b6a4-9fe5f6c94d82">multiple segments, each denoted by two temporal points</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>BEGIN</term>
                            <listitem>
                                <para xml:id="para_cc4effc5-9b6b-4178-beb3-5f9b48c3ecf3">a range beginning at one temporal point, and extending beyond the end of the acquired data</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>END</term>
                            <listitem>
                                <para xml:id="para_7a8b8f95-edfc-4129-a904-a4ce6e915892">a range beginning before the start of the acquired data, and extending to (and including) the identified temporal point</para>
                            </listitem>
                        </varlistentry>
                    </variablelist>
                </section>
            </section>
        </section>
        
         

    </chapter>