<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2035</title>
    <subtitle>Reference RT Structure Sets for Planar and Volumetric ROIs in DICOM SR TID 1500</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>August 2020 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/07/04</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie <link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>IDC</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/03/29</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2035</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Reference RT Structure Sets for Planar and Volumetric ROIs in DICOM SR TID 1500</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>RT Structure Set (RTSS) objects are frequently used to define regions of interest (ROI) for radiation oncology.</para>
                    <para>Rather than transcribing their 3D coordinates into an SR content item, it may be appropriate to reference an ROI within an RTSS object. This is possible, given the SR infrastructure, but not defined in the existing Planar and Volumetric templates.</para>
                    <para>References to Conceptual Volumes defined for 2nd Generation RT are outside the scope of this CP, and may be the subject of a different CP. This CP is intended to address the installed base of RTSS producing systems and re-use of archives of RTSS objects.</para>
                    <para><emphasis role="italic">[Ed.Note. The SNOMED code (410673009, SCT, "Region") is not used, because its position in the class hierarchy as a child of (10546003, SCT, "Site") implies that it is an anatomical (body) region, not a generic region in space.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        
                 <section label="TID 1410" status="3" xml:id="sect_TID_1410">
            <title>Planar ROI Measurements and Qualitative Evaluations</title>
            <para xml:id="para_47a24b81-ef09-41fc-b6aa-1cd54a1889dd">This Template provides a general structure to report measurements for some metric, e.g., density, flow, or concentration, and/or qualitative evaluations, over a planar region of interest in an image. The ROI may be specified by an SCOORD on an image, by a Segmentation Image, <emphasis role="bold"><emphasis role="strikethrough">or </emphasis></emphasis>by an SCOORD3D defining an area relative to a 3D Frame of Reference<emphasis role="bold"><emphasis role="underline">, or by a reference to an ROI defined in a radiotherapy Structure Set</emphasis></emphasis>.</para>
            <table frame="box" label="TID 1410" rules="all" xml:id="table_TID_1410">
               <caption>Planar ROI Measurements and Qualitative Evaluations</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_30b4aee0-ea2f-4145-8208-793586b42cc7"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e2114728-d791-431b-9191-bc8f10b33fb6">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ebe5dfb8-ae20-4bef-910d-06c15730b75b">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_11076f1d-2653-47e3-867d-98687bcf34d3">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0e28dd32-3fea-42af-ba33-fb3836638cf4">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b53dfca1-52dc-4c1a-86ae-6e7da6594631">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fdfbf0aa-9607-43d3-8888-66cb7b9ecd25">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_98188a1d-f8c2-4703-b93c-1bbbda3c1e56">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4191d21b-4639-4e6b-9acf-3961b18247e2">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bda96106-9fc7-43dd-a450-a91a0a6e6fca">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_71a85977-09a6-44a2-8acd-8142481e9370"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_489b564d-d244-477d-bd40-a3039fc941b6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f6aa1279-28cf-49b4-9778-3eec192f8b70">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f372076d-6d4d-4f09-bbdc-e10f7d314c85">EV <link linkend="DCM_125007" xrefstyle="select: nopage">(125007, DCM, "Measurement Group")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b6fcff3e-4c4d-4a1d-a7ea-d26a8443c9e3">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_da8a0eb6-8b06-4635-a02a-cba4d9ce5d6d">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_71881a64-73b5-48d6-b0b0-fbe9806900be"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_07f57d22-61b8-4875-bd89-d3283d0dbdcc"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fb0e53f7-4360-4bdc-90f3-8a0eceeb1ef5">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_156f7617-86c4-4f6c-b54b-caf1b3189fbb"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_276c3053-b014-48d2-a7a7-55cfc41f878f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_41a9d7d3-bf90-4341-8ad3-9837ad16c24f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1a7b3a52-1412-4e42-9d2d-0cfa9e7d62a0"> </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f859274b-a604-4548-87e7-eb0fd5b28d7b"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c455c16d-f3e3-455e-be7f-b95a4ac1edd5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ea4722b2-6f7c-4c8e-bb0d-8a806ebab8fc"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e3890255-829a-4c67-ac45-36643d998a30"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_409f6e6d-f887-4f49-8b11-989188aedec1">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ee6fd942-0371-4fde-9638-f2a8e67e6fcb">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_208032d8-b7e8-4af0-bd71-840b23278e24">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b68aa0a9-d0b1-475c-9630-fbbd48dc581d">SCOORD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_aeb2a487-642c-44a7-8b45-d6d50adb365f">EV <link linkend="DCM_111030" xrefstyle="select: nopage">(111030, DCM, "Image Region")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ae8bf884-64d7-40bc-97dc-e6a4d7a6db83">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c2d8f97e-e24e-4591-b38a-f69d71824f03">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7056160c-74ff-4830-8ffe-c048ab696afc">XOR Row 7<emphasis role="bold"><emphasis role="underline">, 7b, 8b</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9fc59e52-3212-40a3-aea4-5b89635d1073">GRAPHIC TYPE = not {MULTIPOINT}</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2fd976b5-86d6-402a-a25e-e1d3ad8c71df">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b61af4ca-9a37-482f-8823-7f1b73b7eaeb">&gt;&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61538b3d-d790-4596-b395-7e7f083e18c0">SELECTED FROM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5d898cf3-65ee-46c1-a19e-17ebe5d0def2">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3a46a4c1-9af4-401e-80fb-9f9265a270db"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_584a39b5-760d-4662-8675-3367175e20bf">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_45965ebc-e1d3-40dd-b11f-72e8e7169770">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8337874b-0930-408d-a290-518ba3f2656c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6bf3b656-2906-4808-a9d0-762fee69b54c"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1d5a0bc4-a020-44ab-a211-f17a9fa83246">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dafa9592-2b76-411a-9dfb-b07bf3f4bdb5">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_37c5f744-ec7b-4ecc-96c5-ce5e33d14f03">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1e873d36-d3af-4f1c-9fd1-79cc437ae752">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0840dc73-1210-4877-a6fe-5c287e00c5b6">EV <link linkend="DCM_121214" xrefstyle="select: nopage">(121214, DCM, "Referenced Segmentation Frame")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3c544407-674c-4954-98c4-405cae664c59">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2f37e245-08d0-46b3-b446-4b1b9955370b">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7b26463d-cb66-40bc-a179-c245a1aa6541">XOR Row 5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_24dd541b-fcbc-4cc8-9364-0a8fee6741cd">Reference shall be to a Segmentation Image, with a single value specified in Referenced Frame Number, and with a single value specified in Referenced Segment Number</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_0f9c960f-d44e-4440-9762-3fbf05b80de1">7b</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_4e882976-fa30-4757-8905-259512b4b924">&gt;</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_558c96de-5729-4bc5-950f-56bd85dd47b0">CONTAINS</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_1a723f9c-a6d6-4e75-b433-68c8c2724875">SCOORD3D</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_ac4f927a-6af9-48a5-bb6c-cf20f9ffaa8e">EV <link linkend="DCM_111030" xrefstyle="select: nopage">(111030, DCM, "Image Region")</link>
                           </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_29311599-70ad-48b6-8b6f-c6177681c7d0">1</para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_6208f3a4-b6a6-43c0-b1f8-eb799bb5d2fa">MC</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_17e1358f-3b6b-4611-9676-982717a42680">XOR Rows 5, 7<emphasis role="bold"><emphasis role="underline">, 8b</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_61b5c76f-1055-46bd-832f-9f5f1db0bf8d">GRAPHIC TYPE = not {MULTIPOINT, POLYLINE or ELLIPSOID}</para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fd6f4e6d-bfaf-469c-a569-678a3e28a45c">8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9eb47db7-af21-475d-9bd6-a2c3f16dfdde">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e07e9e71-53c6-4d5a-899e-6fd1ae56665b">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ff3d0c74-1592-4dfc-bc58-b7beb4433ffb">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_97ea6a5d-9ccb-4e07-b9e6-00d50cf38213">EV <link linkend="DCM_121233" xrefstyle="select: nopage">(121233, DCM, "Source image for segmentation")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ffa90db0-e98e-4257-ab76-0e13a9c0376f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cb697351-58d1-4567-ad0e-5ddcd6258d27">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fb70c3c6-b10e-410f-a1b4-71dce1ecdad9">IFF Row 7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d8d18206-d68b-49d1-8f4a-863567c5f41d"/>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">8b</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">COMPOSITE</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd001" xrefstyle="select: nopage">(ddd001, DCM, "Region in Space")</link></emphasis></emphasis>
                            </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">MC</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">XOR Row 5, 7, 7b</emphasis></emphasis></para>
                                </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">Reference shall be to an RT Structure Set Storage SOP Class Instance.</emphasis></emphasis></para>
                       </td>
                   </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">8c</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">HAS PROPERTIES</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">TEXT</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd002" xrefstyle="select: nopage">(ddd002, DCM, "Referenced Region of Interest Identifier")</link></emphasis></emphasis>
                            </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">Reference shall be to a single value of ROI Number (3006,0022) within within an Item of Structure Set ROI Sequence (3006,0020) of the parent RT Structure Set Storage SOP Class Instance.</emphasis></emphasis></para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_313a151c-67cd-49b7-846e-4d5a3e2c75b6">9</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4786b290-a3b9-480b-82d8-61da51f69902">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1e173176-b24a-43ba-8a9d-140b3018ae55">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0cbcfa96-4ad8-476e-b1e8-383174d68dc5">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d6c0f3f0-30c4-4ecf-a4bb-6e5a1c86e9e3">EV <link linkend="DCM_121200" xrefstyle="select: nopage">(121200, DCM, "Illustration of ROI")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ea521b1d-ad15-4728-a26d-849b43104bb3">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0d9737a8-771b-4229-a641-b9f86c4feee8">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a573c035-ba51-4957-8247-657d5af5f258"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b989e7e5-94ac-4903-9449-18b04d1fe108"/>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_cba23934-2c47-44b6-b834-d7ff6809e9ea">9b</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_f83a89d7-f094-41c9-820d-76cc398c9fc0">&gt;</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_c8c2a72c-7db8-4792-ae87-823523fbfd77">CONTAINS</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_f5759476-134c-4025-9859-b6b4bc1d5f09">IMAGE</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_eb85c5a6-5724-4504-9675-1737077a473f">EV <link linkend="DCM_130401" xrefstyle="select: nopage">(130401, DCM, "Visual explanation")</link>
                           </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_d3e88864-b76c-4b1c-9212-e423a7aa801b">1-n</para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_c2fbb35d-c8b0-4183-b814-9c2ff56cd55b">U</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_b2d95ee3-e445-4738-af9e-afa87b21e28d"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_6505d173-ea2f-47a7-ba7c-55faab18d15b"/>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ed99ed8f-5812-4207-be2e-ac6586ee4667">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_046533bc-8863-4fd7-a14f-642334b42b94"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61769706-ae87-40f2-927d-77450a617980"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f95c9649-525a-47c8-9944-441781b1347e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_52f6fcf1-9602-4df6-9456-549628536aa5"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cb90f7af-88ac-4016-82aa-a846603f7103"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0884c679-b70d-4a60-84be-71f1c936781d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3321af56-c75d-4230-a308-db95c12a3366"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dfbc48a7-79d2-496c-af44-3032222b4a7a"/>
                     </td>
                  </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0a284ad5-0379-4e48-bdfb-935fe207e896">13</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3f196fad-8ce5-4b00-b3e9-0bbc042a5078">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6adbf8e7-442a-4620-b275-a5b34eb2588c">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_65e0c51c-c749-4d27-83f8-efbfe1ffefde">TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_03aa607e-19a2-42b9-a606-c8581a633751">$QualType</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c12f8242-a497-480e-b036-2c52b60444da">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5ebab0ff-fde9-4176-b1b2-0e9611b5d9a0">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_637039fb-753c-4a98-a7b5-25c573296787"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3fbbf641-5108-4c3c-87bb-af431edfe500"/>
                     </td>
                  </tr>
               </tbody>
            </table>
            <para xml:id="para_12f34dea-c5d4-4127-b3fd-5f2adff0a8ef">
               <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c53d2751-c9e1-4f42-b8cb-2db8762be074">Row 5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_91888c98-d413-40db-985e-f81f83c2ebb0">To describe an infinitely small area, such as the center of a lesion, a Graphic Type of POINT may be used.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2babd6ee-8440-4df7-87f8-8cfc79a2f2e2">Rows 6, 7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_48197dd3-b070-41ae-b65f-61265a57f1b3">Referenced Frame Number (0008,1160) is an attribute of the IMAGE Content Item, and shall be present with a single value.</para>
                        <para xml:id="para_6dab3616-4bba-44c5-92fa-571941f4ada2">If the Referenced Segmentation SOP Instance has Segmentation Type (0062,0001) value BINARY, it identifies the area of defined (measured) region of interest by pixel values in the referenced frame with value 1. For Segmentation Type value FRACTIONAL, the area is computed by an implementation dependent method.</para>
                        <para xml:id="para_f5322076-0cc8-427a-ab7b-a6f22534be34">Frame number shall be specified even if the Segmentation SOP Instance has only a single frame.</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_9d0586d2-2dd6-4a95-a15a-520170d8f32c">Row 7b</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_30254e85-71f3-44fe-b526-84ca904c9817">The area may be defined independently of an image by reference to 3D coordinates in any type of DICOM Reference Coordinate System, whether it be patient-relative (Patient Based Coordinate System), volume-relative (including acquired and presentation state volumes), or whole slide relative (Slide Coordinate System).</para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f1ded33a-a26f-4d96-8c25-411ba89d6eed">Row 8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_983898e7-403a-43eb-ba65-ada0710d991d">Identifies the source image that was segmented to identify the ROI, and whose properties are described in this container.</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">Rows 8b, 8c</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">A reference to a single ROI that defines a planar ROI, within an RT Structure Set.</emphasis></emphasis></para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_da642821-8bef-4b68-b732-fa0a6c7208c4">Row 9</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7d9345d8-365c-4575-97af-38086b027daf">This referenced image may contain a "screen shot" illustrating a rendered version of the ROI.</para>
                     </td>
                  </tr>
               </tbody>
            </informaltable>
         </section>
         <section label="TID 1411" status="3" xml:id="sect_TID_1411">
            <title>Volumetric ROI Measurements and Qualitative Evaluations</title>
            <para xml:id="para_cbcf130c-59d0-43ea-b71c-7f7d5b19c70b">This Template provides a general structure to report measurements for some metric, e.g., density, flow, or concentration, and/or qualitative evaluations, over a volumetric region of interest in a set of images or a Frame of Reference. The volumetric ROI may be specified by a set of SCOORDs on an image set representing a volume, by a volumetric Segmentation Image, by a volume defined in a Surface Segmentation, <emphasis role="bold"><emphasis role="strikethrough">or </emphasis></emphasis>by a set of SCOORD3Ds defining a volume relative to a 3D Frame of Reference<emphasis role="bold"><emphasis role="underline">, or by a reference to an ROI defined in a radiotherapy Structure Set</emphasis></emphasis>.</para>
            <table frame="box" label="TID 1411" rules="all" xml:id="table_TID_1411">
               <caption>Volumetric ROI Measurements and Qualitative Evaluations</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_87e5605c-30ef-40ab-9e54-e5a160ac1e4c"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b792ca18-4b6b-4780-8c36-f0872b923a62">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4548b0a2-dafe-4ffe-b5ff-6f4b746b5f37">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0b48ed64-9046-488b-aaab-3e437093e793">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_037bfccf-fcbf-4646-a614-7d3edd7495e1">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a4416f1f-b2da-48e2-afc1-6d1916588c9f">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8d7ea4d1-7d0f-4df2-9d64-44c4151c48a8">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9c44f7e5-e932-432b-ae62-83dc4ba5d58f">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6daad6dd-d304-4d58-b580-ab3f55ee692f">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0bd991eb-9642-42a0-9c93-87848af2ac98">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9797c7cc-b481-4eec-8d93-ae08fb873fac"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eab96362-2bfa-41eb-9637-0bf9efdfb87c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0bcea660-1791-4428-9783-86bf458b9f04">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f2b51594-f8ef-4ee7-8448-f3dfaa51b3a9">EV <link linkend="DCM_125007" xrefstyle="select: nopage">(125007, DCM, "Measurement Group")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b8912904-0922-489a-85ba-f80c0b80db5e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_06cf9952-1268-4f7b-a9fe-915d8d26cd3c">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_91c8a03f-d2c8-4731-b118-5c6353303797"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ac09d88a-4385-4f34-9717-b507f6b6dc0e"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_877bb76f-4c64-41b3-bfab-0eea84a49504">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0b700d29-3e4b-4bd9-bd54-da1cbfda3b27"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_939f9e7a-0ff4-45cd-be67-8501659ec0a1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3c2a24f6-d223-4764-9bc5-f445a2517571"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e4e1a46f-5ec2-439f-8ac6-82a7ec89529f"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1bf8ef93-0eec-4a8d-a65d-95d52d0bb5a7"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0ecb62e6-4ce5-4d1a-9602-fd6d1a6f2b53"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3e1259d1-10b4-4705-822c-766dcb020a35"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eda1d8ee-c6de-417f-ab61-06a118c97fd2"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8a7f801e-885f-430a-bd16-b30d110a1688">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7af8c31f-2643-458d-8e14-dd1f3255bb05">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f8a1fdf0-a434-47d9-801f-d653ef323489">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1770bfca-c1c0-4d94-afef-585287c8fab9">SCOORD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5da4d3d-0d8e-4e1d-89ad-b0ff686e2821">EV <link linkend="DCM_111030" xrefstyle="select: nopage">(111030, DCM, "Image Region")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ad1a6260-eebb-4362-bb5d-36a4c358709c">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1433ce9d-9be0-4b8b-9cea-fac6e2457283">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d7c9f261-862f-4f4e-af48-c4991bad9403">XOR Rows 7, 10<emphasis role="bold"><emphasis role="underline">, 12b</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8a4e4c48-d4b8-4fd8-9916-f7c7dd6c68bf">GRAPHIC TYPE = not {MULTIPOINT}</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1922085b-ae2d-417f-884e-4ad99df2cfd1">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a2f1b3d-4d96-46a7-9e71-8c38f1923cc6">&gt;&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_162b8ae0-ceab-44ad-9e45-7e2d52b8a450">SELECTED FROM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc1348d7-32ba-4ebc-b33f-da5ab5244b1b">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_de799c4a-6157-4d21-b70f-5ceabe82e778"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f685cd2d-5961-43c7-b692-a8b66e591af8">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_26d503ce-2255-48b0-bc65-01ba0499077f">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_62913662-81a3-4c89-baa3-c8f9401ddb60"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3295597b-5673-4db5-8f89-e1635d935dce"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ad9a7ba2-af63-48fb-84d3-00c124017027">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_99b8c5b5-8655-41ee-86e3-092c9881c82d">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a3d526cd-9556-4905-a0be-00c13011cacb">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ae357b28-4114-42e0-8fcd-29e6cef183f3">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9a692999-4664-4d03-b6ee-3675f3f580c9">EV <link linkend="DCM_121191" xrefstyle="select: nopage">(121191, DCM, "Referenced Segment")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4d7b8acd-9ea7-4389-939a-37b5b8127b76">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6904a646-c0dd-4765-9d3a-0efaa128d777">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c1c1ce88-c21b-4e66-acf8-2d0862590ad2">XOR Rows 5, 10<emphasis role="bold"><emphasis role="underline">, 12b</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1e016cb9-0c46-4f8b-ac6b-0d5b36b83d8f">Reference shall be to a Segmentation Image or Surface Segmentation object, with a single value specified in Referenced Segment Number</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e4f3a8f6-2d92-49e9-ab18-d1750c31f1cb">10</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d5942475-caf5-4b5b-8ef1-572ab8b05778">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_08d967de-4460-4233-ad43-542afda76c50">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_953ef38b-a49f-4d4e-a791-e54c13480fe6">SCOORD3D</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f1082097-4877-4b87-be85-459ec1f45e57">EV <link linkend="DCM_121231" xrefstyle="select: nopage">(121231, DCM, "Volume Surface")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_783866ce-63e0-4021-85c3-bce5be35e4cc">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f19a0a1f-c92a-45ed-b6b6-c97eced455e1">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4ee17874-ef57-4cc0-bc0d-9ba6d1365733">XOR Rows 5, 7<emphasis role="bold"><emphasis role="underline">, 12b</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_0eabd745-2fbb-4d51-906c-a6872c0c3631">If one item, GRAPHIC TYPE = {ELLIPSOID or POINT}</para>
                         <para xml:id="para_94e71819-f1b8-4a7b-8194-f8fcf7ee244d">If more than one item, GRAPHIC TYPE = {POLYGON or ELLIPSE}</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_51ce42d9-348f-4b22-a1f3-c4eeeae13aea">11</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fbfdbb5f-2399-48cd-8a37-27cfd3ac93e0">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9ef769a4-b036-488a-9617-cbe524d65743">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_582c8232-dcd5-4aab-813e-3258dbc8f13c">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5041208f-c5d5-4906-87d1-92f4610d7f08">EV <link linkend="DCM_121233" xrefstyle="select: nopage">(121233, DCM, "Source image for segmentation")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6e888005-191c-4f37-b69d-091f05b75823">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fb51dcec-c372-4680-82e9-f14a4617538d">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ba907fb4-8b79-43d3-99fd-ab6d5d47c208">XOR Row 12 and IFF (Row 7 or Row 10)</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_17aefa54-5d49-42ca-b9e7-7e736ebbdcb0"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4330498d-5af9-47c5-8d78-bb4b0e579616">12</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7513186f-31e6-40c3-b827-b4ea1a357362">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b970e94e-0c56-4720-96dd-6aa4ba1b692f">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_277a3bf0-2f46-4870-9fad-39c31f5ce53b">UIDREF</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_72cde9df-be57-4d02-829c-edb25d574bb5">EV <link linkend="DCM_121232" xrefstyle="select: nopage">(121232, DCM, "Source series for segmentation")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0c538e29-5818-47d8-8bae-7a89ce13b8e3">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_41bc29f5-d975-4822-ab35-33cc8a34c59f">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f394ce81-6558-4814-8ed0-59cf7e54df6c">XOR Row 11 and IFF ((Row 7 or Row 10)</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_87e310ad-ee4f-4cbd-8478-4f33ad28764a"/>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">12b</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">COMPOSITE</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd001" xrefstyle="select: nopage">(ddd001, DCM, "Region in Space")</link></emphasis></emphasis>
                            </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">MC</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">XOR Row 5, 7, 10</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">Reference shall be to an RT Structure Set Storage SOP Class Instance.</emphasis></emphasis></para>
                       </td>
                   </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">12c</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">HAS PROPERTIES</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">TEXT</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd002" xrefstyle="select: nopage">(ddd002, DCM, "Referenced Region of Interest Identifier")</link></emphasis></emphasis>
                            </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">Reference shall be to a single value of ROI Number (3006,0022) within within an Item of Structure Set ROI Sequence (3006,0020) of the parent RT Structure Set Storage SOP Class Instance.</emphasis></emphasis></para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_48d6906f-d64b-47a6-a3a4-c6b1d2a8e17f">13</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d65bfe21-f6b6-433a-9868-8eee64a57b2b">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_19f678da-627c-4265-a695-4a07c9981220">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bed1f32e-09b6-4083-8f9a-87d8c997f7df">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7de6d9e4-466b-40ca-9eb2-57050058b0e2">EV <link linkend="DCM_121200" xrefstyle="select: nopage">(121200, DCM, "Illustration of ROI")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_081c16d4-8759-4063-bb24-f86082716b96">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_86df942c-5ae8-49ae-a72b-1d2f03d42485">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a8ee2840-b2c0-4b64-8d2b-fd4331239381"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4c9c4835-8892-4926-b8a8-16974eae8e20"/>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_1708792b-18ef-4f5c-bf2f-1ba13e8ccd04">...</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_262219e3-299a-424c-8505-99c0e823a501"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_edf9965f-3c20-4849-90e3-c366a64bbc14"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_eec8fe1e-7559-4584-a092-11271a3929e0"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_976a2eb3-f0ea-4338-b119-150f0f76fe7c"> </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_c26e509b-f7b4-4211-9c2d-79f710f0950a"/>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_e54c81ca-d4b1-47a9-b333-3dbe8754323d"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_67eef97c-8fc9-4853-95fc-a04ac08c8493"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_d8deeafb-9d48-49a3-acec-e8e01eb16cb9"/>
                       </td>
                   </tr>
               </tbody>
            </table>
            <para xml:id="para_b60fecd1-f50d-462d-ba6a-1c914b17ae75">
               <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7e52f281-af25-486f-b461-b626e2f2e688">Rows 5, 10</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_72f73806-17d2-492c-a070-95eb9bf8f3a6">To describe an infinitely small volume, such as the center of a lesion, a Graphic Type of POINT may be used.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7ad17dc6-1664-4b8b-a7d8-0a6bd547d0de">Rows 6, 7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a344ee4b-2603-4c07-9969-c29e565fd785">Referenced Segment Number (0062,000B) is an attribute of the IMAGE Content Item, and shall be present with a single value.</para>
                        <para xml:id="para_aea22b97-c31f-4d41-a51f-8b8dd95b6f87">If the Referenced SOP Instance is a Segmentation Image, it shall have a defined Frame of Reference. If it has Segmentation Type (0062,0001) value BINARY, it identifies the volume of defined (measured) region of interest by voxel values in the referenced segment with value 1. If it has Segmentation Type value FRACTIONAL, the volume is defined by an implementation dependent method.</para>
                        <para xml:id="para_2a348bb0-f614-4335-a72b-034d14684ac4">If the referenced SOP Instance is a Surface Segmentation, the referenced segment shall constitute a finite volume. It identifies the volume of the defined (measured) region of interest by the interior of the finite volume.</para>
                        <para xml:id="para_fdab5d3a-2a14-43fb-a1cd-f731294def29">Segment number shall be specified even if the Segmentation SOP Instance has only a single segment.</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_da291e85-6abb-4feb-a4b1-b18a89fbc13a">Row 10</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_10016efa-1951-4738-9cc9-8c960cc9f485">Either a single item describing a closed volumetric surface, or multiple items describing a set of parallel closed coplanar areas (contours) is specified.</para>
                       </td>
                   </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7f909f70-cd30-4868-bc4d-897aa81c362b">Row 11</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e946b4aa-e3f2-4d0a-873c-522eb2642c32">Identifies the source images that were segmented to identify the ROI, when, for example a subset of images in a series was used.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dd85fc21-38d8-4440-a684-b1f3da9fa6e7">Row 12</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a4cbdb4-42a8-4791-baef-fd84ab032a45">Identifies the source series of images that were segmented to identify the ROI, when, for example an entire set of images in a series was used.</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">Rows 12b, 12c</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">A reference to a single ROI that defines a volumetric ROI, within an RT Structure Set.</emphasis></emphasis></para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d05f88ff-b8b3-47d2-8237-915cb710bac2">Row 13</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2545a9fe-b061-42c5-b8cf-564ea638c248">These referenced images may contain "screen shot" illustrating rendered versions of the ROI.</para>
                     </td>
                  </tr>
               </tbody>
            </informaltable>
         </section>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Add new codes to PS3.16 Annex D:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding Scheme Version "01")</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">Definition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">Notes</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_ddd001">ddd001</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Region in Space</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>A continuous part of space, not necessarily associated with a particular image.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_ddd002">ddd002</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Referenced Region of Interest Identifier</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>A reference to an identifier of a Region of Interest.</para>
                            <para>For example, a reference to the ROI Number (3006,0022) within an Item of Structure Set ROI Sequence (3006,0020) of an RT Structure Set Storage SOP Class Instance.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>

        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged, PS3.3:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
                        <section label="C.8.8.5" status="4" xml:id="sect_C.8.8.5">
                    <title>Structure Set Module</title>
                    <para xml:id="para_e9b33d4a-4173-4815-8b40-b46df3b96c96">A structure set defines a set of areas of significance. Each area can be associated with a Frame of Reference and zero or more images. Information that can be transferred with each region of interest (ROI) includes geometrical and display parameters, and generation technique.</para>
                    <table frame="box" label="C.8-41" rules="all" xml:id="table_C.8-41">
                        <caption>Structure Set Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e0617fa8-7cbb-40a5-8c2b-4bba34953aa4">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b1147d05-c64b-4e1c-b7a9-d94752343350">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8a4afa99-60b9-48db-867c-ce69327e2c48">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_75c1df0e-5b6c-4078-ba10-77b003ebaf7a">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0851234e-fac7-4ef3-b47b-eb428dab7d67">Structure Set Label</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f1a41efc-d623-426b-b100-17450aea603c">(3006,0002)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fed1ea83-6fe3-46c3-87c1-de608ffd1cf3">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_00d17b21-fce9-4b68-a8b1-ab693be6fbd8">User-defined label for Structure Set.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4122b927-4dea-47ed-99d7-2ba67a1dd4e1">Structure Set Name</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ae257380-c267-455b-b4f2-02fe791f17a2">(3006,0004)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9a0ebcd1-8384-46b2-87ea-0894d4f44728">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4f96f752-8031-463d-9f37-0c1cd186548a">User-defined name for Structure Set.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fbc9a655-7c91-4db9-9e65-eba7f7278c17">Structure Set Description</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0dc8119c-8757-4999-97fa-a7e404f559a5">(3006,0006)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b722e0cf-489e-4dce-98ef-68807b73978d">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ee28b5e9-1191-4546-ae80-0f20c4c0e132">User-defined description for Structure Set.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_86fb1d09-7160-441f-86fa-c85caf423932">Instance Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_24ab3789-a895-4bae-aa51-f19301c31bed">(0020,0013)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_797863a8-1aaa-4029-8dd1-d55aa7837072">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_b17baa1e-d36b-4063-99f1-8f3a792b56b2">A number that identifies this object Instance.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_edfeecbe-66ec-4b09-8da6-9cff1d6ad55e">Structure Set Date</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ecffaeca-107f-4462-8903-e68cc26574c2">(3006,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_724d74ca-48b2-47b1-97d3-cf716cefc9f6">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cad362be-e225-4d8a-a6f5-608cfaae1735">Date at which Structure Set was last modified.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e789ea46-821e-452f-922e-149e06fb0a0b">Structure Set Time</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d83ddb84-900b-4608-8891-8edb62cd6d37">(3006,0009)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ed073be2-6ac7-4541-a10d-db8efc108a2c">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_31d7367d-1c45-4967-abe1-768def3e1f6f">Time at which Structure Set was last modified.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_46c42527-30b5-4019-8b64-87704fe57b45">Referenced Frame of Reference Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_53fe5cd0-ee67-4c6e-aef4-1089359835d7">(3006,0010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a5cf689a-49f3-420a-b762-b8e83cfe25ff">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a27bc8cd-5633-4a7f-8788-bf7b9fd90a1b">Sequence describing Frames of Reference in which the ROIs are defined. One or more Items are permitted in this Sequence. See <xref linkend="sect_C.8.8.5.1" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6aa087d8-f515-4633-8ec8-ec67c5e89439">&gt;Frame of Reference UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_46ac388c-3959-4904-99a3-a8f55cd25b5b">(0020,0052)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4b0197ed-9058-4365-861a-d34afe66d526">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7695f769-94f5-4344-b3cd-a21511deced0">Uniquely identifies Frame of Reference within Structure Set.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5034483f-58d2-4aab-9716-59f630bc685c">&gt;RT Referenced Study Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_716a57e9-2a00-4e6d-a4e4-ad224f4b3822">(3006,0012)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d9ec3b37-7a6f-48ed-8580-19d1cc416070">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c798bf74-f271-4143-8d98-ba9ef263b83f">Sequence of Studies containing Series to be referenced.</para>
                                    <para xml:id="para_f1dbd183-c291-486c-8c3a-281becd16831">One or more Items are permitted in this Sequence.</para>
                                    <para xml:id="para_50b787f6-f399-46de-a188-a977c0cebaa5">See <xref linkend="sect_C.8.8.5.4" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_91e806a8-536f-4cfd-ac7a-db72d497f83b">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4cde3da7-bd13-407c-9726-382fd82bb368"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_36e96508-d726-4c79-9bcf-6a0e1df310f0">&gt;&gt;RT Referenced Series Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_efe2e1a4-2292-4b53-a484-fa174c865e12">(3006,0014)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b6753306-a2ff-4d7f-9c4a-8064bb3a0b4e">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_22784924-11d6-40a2-9bf2-667efaac71f1">Sequence describing Series of images within the referenced Study that are used in defining the Structure Set.</para>
                                    <para xml:id="para_8779c86d-e8dc-419c-b91f-6002501929b8">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5d793484-8886-4ff1-ad25-5fd18b800036">&gt;&gt;&gt;Series Instance UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e17a180f-dd08-4f0b-958d-602db987e43d">(0020,000E)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_191da809-55a4-4f59-8ed0-46f83e9f5879">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1885613f-c52f-44cf-8a0b-db620431af28">Unique identifier for the Series containing the images.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_74a79cae-4684-4a63-b240-8c3fcfd4ff5e">&gt;&gt;&gt;Contour Image Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d2e7bd8a-ed91-4b93-b3c7-d72810d0dcb9">(3006,0016)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6dea1984-aebd-40ad-9e69-b3e88db573ea">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_efc397fd-7799-42ca-83c5-1a86b5b6e659">Sequence of Items describing images in a given Series used in defining the Structure Set (typically CT or MR images).</para>
                                    <para xml:id="para_4a4045e5-fecf-4fb4-b466-accbcc83e746">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_e11384c0-6597-4621-852c-92fe6c146b8b">
                                        <emphasis role="italic">&gt;&gt;&gt;&gt;Include <xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_58c8ebe8-1700-4986-b796-41840b27adbf"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_105805ca-2bb2-47e5-b470-bfda7736c89a">Structure Set ROI Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bc874c02-c039-40bf-a4af-1e083a5fa81a">(3006,0020)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_57f95d66-fd11-4e8b-b9e9-7c9a28d051aa">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3174e5c0-41a6-4d2c-a0b1-1486fcb0e76a">ROIs for current Structure Set.</para>
                                    <para xml:id="para_3d35ea3d-3bf8-4fa4-8cb2-fa4c70d20ea8">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bd7de259-3aaa-46b2-9ff6-e1bc697bdb36">&gt;ROI Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_20ccd60d-2911-4129-aa4f-6d11dca65b93">(3006,0022)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d9881665-b89d-4286-a769-08b5509a6ba4">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1facfcb2-f4a1-41a9-bcdf-0a85f4249e66">Identification number of the ROI. The value of ROI Number (3006,0022) shall be unique within the Structure Set in which it is created.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ffa4c3bf-fc2f-45b3-8e15-60058df20951">&gt;Referenced Frame of Reference UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a509f467-8559-4a00-8175-baf7b46a7484">(3006,0024)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_670eb8f6-9b2d-40a9-9784-7ad63415d236">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0da61efd-f1df-4e3b-9a5c-af6f68b4d60a">Uniquely identifies Frame of Reference in which ROI is defined, specified by Frame of Reference UID (0020,0052) in Referenced Frame of Reference Sequence (3006,0010).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e6a9c968-c9d6-4e2c-b49f-ad8d22d1f3a2">&gt;ROI Name</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c5724686-245d-4687-bf98-cc4306b35722">(3006,0026)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_55434209-4fba-4ac5-b585-10d9004170fb">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1326ebc7-2e1a-45fb-b2ce-aa351373e9c4">User-defined name for ROI.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d61ae0d1-811c-449f-ac3c-57d4ca66cefe">&gt;ROI Description</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ca151af4-955f-4f17-96f5-ea5c06662419">(3006,0028)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_106ad573-d0c0-4441-add1-d7fab235077c">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_81600e1d-27ca-4fdd-82c7-bca531c36bcd">User-defined description for ROI.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_36b8a214-286d-4a2a-8767-eae0cb688d7d">&gt;ROI Volume</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2d8b4edd-555b-47c4-8236-2d4eb6de19f3">(3006,002C)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_61450e58-639a-48f0-8582-739f0b793391">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6152ecdf-fb5e-423b-a13c-2c17f5408b83">Volume of ROI (cubic centimeters).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e9c8b760-ad70-4fda-ba9e-63aeb194272f">&gt;ROI Generation Algorithm</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_16d7a0cf-40de-4d0f-8182-cb7228efcf8e">(3006,0036)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fa493edb-302e-4ac0-b466-860a2eaf7459">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1cc35054-1a1b-4b76-9744-d58f3afd06b1">Type of algorithm used to generate ROI.</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>AUTOMATIC</term>
                                            <listitem>
                                                <para xml:id="para_60abfbec-3760-49d0-9721-3d7d65596121">calculated ROI</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>SEMIAUTOMATIC</term>
                                            <listitem>
                                                <para xml:id="para_8e748a6f-11d7-4c6f-8b97-2aed6aecfe0c">ROI calculated with user assistance</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>MANUAL</term>
                                            <listitem>
                                                <para xml:id="para_9077eff9-7435-4554-8389-3911fd6b25e3">user-entered ROI</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1ef95b45-358d-4cb9-bd1c-065a3bb0c077">&gt;ROI Generation Description</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b6fc5328-e5d6-4bcd-bcf4-dfd31af63094">(3006,0038)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_048c3aef-85ac-4bee-b13c-373ae0adcd94">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8dd7e884-754c-4d47-845f-4109392b9341">User-defined description of technique used to generate ROI.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_32f83dd5-6578-4a9e-89d3-05553bf9dae7">&gt;ROI Derivation Algorithm Identification Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8a25a32e-547c-4bb2-8a55-f863297986ce">(3006,0037)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7d9dd13f-9999-4ea2-ba75-b749060d56be">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_00159958-53a2-4740-bac7-bd9cb096d228">Software algorithm that derived the ROI.</para>
                                    <para xml:id="para_6425272b-3fa5-4dff-a540-67eb272d77c8">Only a single Item is permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_9c0c019a-452c-4be9-94ad-e68a3c13451b">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_10-19" xrefstyle="select: label quotedtitle"/>.</emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_40cea69c-8f7f-40e4-99ac-47b904a457d3"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9c7e35ee-b72c-4cad-bacc-83042feb7e19">&gt;Derivation Code Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_914aad56-db13-4b42-aaf2-1bd625f0010a">(0008,9215)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6b20ed81-0b38-4df6-8a7b-a7648f8915eb">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_02f11e9c-ecee-4d76-99e5-7e98c90ebd9f">A coded description of how this ROI was derived.</para>
                                    <para xml:id="para_31bb497a-8cbd-41b1-a7f5-7c7a4d74b1e2">One or more Items are permitted in this Sequence.</para>
                                    <para xml:id="para_d8476dac-2091-4af5-a441-ccd216f4cbef">See <xref linkend="sect_C.8.8.5.3" xrefstyle="select: label"/> for further explanation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_edda6142-cf9c-491b-9f04-07a960790591">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>.</emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_15bacdd5-c8d8-4d5b-a8d7-92208857104b">
                                        <emphasis role="italic">Enumerated Value <olink targetdoc="PS3.16" targetptr="DCM_113085">(113085, DCM, "Spatial resampling")</olink>.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_13cae73a-eb7c-48d1-9102-43c72d319ece">&gt;Definition Source Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8adcda71-5770-410e-8766-5f45879a7659">(0008,1156)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c71dc960-8613-4c9f-9cc2-a3161ed33bbb">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_353e4d90-8905-4ce8-852c-99e192639509">Instances containing the source of the ROI Contour information.</para>
                                    <para xml:id="para_64df4da9-6ccc-4936-8697-efd5396e70f3">Only a single Item is permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_2af4272b-bee9-4b6f-a563-9a9aecdd3235">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>.</emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f1a3d316-089c-45e6-ae04-ec7dabfcce73"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ec488a06-473a-43e7-b3f8-762704deaed6">&gt;&gt;Referenced Segment Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bd8f77db-9864-4891-a9c6-457d376254d6">(0062,000B)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ffcee7e2-64fa-4c25-95ea-323d95c66451">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5e4fb78f-eb38-4c00-9eeb-adb33abdd8a4">The value of Segment Number (0062,0004) in the referenced SOP Instance that identifies the segment that is the source of the ROI Contour information.</para>
                                    <para xml:id="para_52173d42-c557-43ba-9476-16c26858236a">Required if Referenced SOP Class UID (0008,1150) is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4").</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e6367d36-dc18-49bc-bdab-d23db28f20b2">&gt;&gt;Referenced Fiducial UID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5f7d7242-bfcc-4d1f-a0b0-38080a62e7fa">(0070,031B)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_305d23a4-1a00-455b-a8c4-915fdd271917">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0b883057-cbc9-4c99-b06d-d14f069d3c2e">The value of Fiducials UID (0070,031A) in the referenced SOP Instance that identifies the fiducial that is the source of the ROI Contour information.</para>
                                    <para xml:id="para_e7f57481-3fbe-4df6-9cdc-e26587817404">Required if Referenced SOP Class UID (0008,1150) is Spatial Fiducials Storage ("1.2.840.10008.5.1.4.1.1.66.2").</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3250482f-e157-4d79-b16a-6371d531a833">Predecessor Structure Set Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_188514ab-cb7f-4bac-b320-25890f4db9fe">(3006,0018)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0af9ec83-ee0c-4d8a-9ef7-341eb4a73be8">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1bc72c32-c03f-4f0f-83b1-5e65151007f1">The Structure Set that has been used to derive the current Structure Set.</para>
                                    <para xml:id="para_1111d3d3-c666-41fe-b1ad-892bfeac02c2">Only a single Item is permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_8f63e7dc-07c9-4683-96d9-1fd078ca8118">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0ed8f47d-68e2-4fba-bfa0-818c3def35b7"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.8.8.5.1" status="5" xml:id="sect_C.8.8.5.1">
                        <title>Frames of Reference</title>
                        <para xml:id="para_81635a73-9def-4b35-a97b-3d1e63e8dd7c">The Referenced Frame of Reference Sequence (3006,0010) describes a set of frames of reference in which some or all of the ROIs are expressed. Since the Referenced Frame of Reference UID (3006,0024) is required for each ROI, each Frame of Reference used to express the coordinates of an ROI shall be listed in the Referenced Frame of Reference Sequence (3006,0010) once and only once.</para>
                        <note>
                            <para xml:id="para_ff372086-83bf-4141-ac41-f4141069b3f9">As an example, a set of ROIs defined using a single image Series would list the image Series in a single Referenced Frame of Reference Sequence (3006,0010) Item, providing the UID for this referenced Frame of Reference (obtained from the source images), and listing all pertinent images in the Contour Image Sequence (3006,0016).</para>
                        </note>
                    </section>
                    <section label="C.8.8.5.2" status="5" xml:id="sect_C.8.8.5.2">
                        <title>Frame of Reference Relationship Sequence and Transformation Matrix</title>
                        <para xml:id="para_32a226c9-0845-4d86-b799-ead114bfe5d7">Retired. See <link xl:href="ftp://medical.nema.org/MEDICAL/Dicom/2011/11_03pu.pdf">PS3.3-2011</link>.</para>
                        <para xml:id="para_e7e0e65b-2169-4bc0-8d17-f9da6637ca55">The concept of definition of registered Frame of References using the Frame of Reference Relationship Sequence (3006,00C0) formerly present in the Standard is retired. The use of Registration IODs is advised since the introduction of Spatial Registration IOD, which is a much stronger and more general concept, and independent from the specifics of RT Structure Sets. Additionally it is of importance that registrations are decoupled from image and segmentation objects.</para>
                    </section>
                    <section label="C.8.8.5.3" status="5" xml:id="sect_C.8.8.5.3">
                        <title>ROI Derivation Sequence</title>
                        <para xml:id="para_18abacec-0365-4e24-bfef-5aaa049a09f9">If an ROI is created by re-sampling an existing ROI that is either (a) present in a different Frame of Reference (FoR) and has been resampled to align with the image Series referenced by the current structure set, or (b) present in the same FoR, but its contours are referenced to a different image Series, it is useful to be able to identify that the ROI is resampled. The Derivation Code Sequence (0008,9215) shall be used with code <olink targetdoc="PS3.16" targetptr="DCM_113085">(113085, DCM, "Spatial resampling")</olink> to indicate the resampled ROI.</para>
                    </section>
                    <section label="C.8.8.5.4" status="5" xml:id="sect_C.8.8.5.4">
                        <title>SOP Class UID in RT Referenced Study Sequence</title>
                        <para xml:id="para_a2104f85-679a-4a60-bd1f-d4468fce3fde">Since RT Referenced Study Sequence (3006,0012) is Type 3, the Attribute may be omitted.</para>
                        <para xml:id="para_06fe3735-3ddf-47f8-b186-ee45164079b4">If RT Referenced Study Sequence (3006,0012) is present with an Item, the SOP Class UID of the Detached Study Management SOP Class (Retired) may be used in Referenced SOP Class UID (0008,1150).</para>
                    </section>
                </section>
                <section label="C.8.8.6" status="4" xml:id="sect_C.8.8.6">
                    <title>ROI Contour Module</title>
                    <para xml:id="para_d5e99224-1d6e-4056-b725-85b11223bc59">In general, a ROI can be defined by either a sequence of overlays or a sequence of contours. This Module, if present, is used to define the ROI as a set of contours. Each ROI contains a sequence of one or more contours, where a contour is either a single point (for a point ROI) or more than one point (representing an open or closed polygon).</para>
                    <table frame="box" label="C.8-42" rules="all" xml:id="table_C.8-42">
                        <caption>ROI Contour Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f3653f23-e05d-40d1-95e1-065a0fb2bae6">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c75214f0-9953-4be2-93a5-596aa7387226">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_423d0c7e-0713-4110-96da-1bb5f5d401eb">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8463cf31-e9a2-490f-993f-9b67b0d0ceeb">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3fab2aee-8023-4adb-8591-c8d58fdd37c4">ROI Contour Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_30ba12d1-dd2c-47f4-b792-72d41e10c630">(3006,0039)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c774c1e4-6d47-41bd-aeac-246c3b75602a">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_de60780f-19d5-4974-b7d1-a53c8197f786">Sequence of Contour Sequences defining ROIs.</para>
                                    <para xml:id="para_5bfff524-a4c5-407a-8c7d-f3b6efaeafd3">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a5361281-0162-476a-a91a-b06f3fd4ac6a">&gt;Referenced ROI Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ea17c8bc-8bec-41a6-8afb-777f550401ec">(3006,0084)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bac1d044-878a-4260-b1e0-6b4056e7233b">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5a25ffa1-3edc-4619-9721-7f57db4c3027">Uniquely identifies the referenced ROI described in the Structure Set ROI Sequence (3006,0020).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_875a03e7-028e-421d-b014-ef12996953a7">&gt;ROI Display Color</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1586b43e-8f5a-4c82-9144-de7f8a35b447">(3006,002A)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_892d8e81-cc56-4a67-8f36-33af7bd476f1">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c68bc991-10c4-4651-b160-ae2a4fa66d14">RGB triplet color representation for ROI, specified using the range 0-255.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_79837df9-0598-461a-ad0e-20832b2615bc">&gt;Recommended Display Grayscale Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a82de3a0-004d-422b-a5cd-e9bc7877b31c">(0062,000C)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b2642ce9-c3f0-48a1-b6ca-48a44887bb0e">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d1ec2f96-4c83-497b-8164-f72a41ba753c">A default single gray unsigned value in which it is recommended that the contour be rendered on a monochrome display. The units are specified in P-Values from a minimum of 0000H (black) up to a maximum of FFFFH (white).</para>
                                    <note>
                                        <para xml:id="para_6113a007-b553-4634-9416-ccaeedafe8a0">The maximum P-Value for this Attribute may be different from the maximum P-Value from the output of the Presentation LUT, which may be less than 16 bits in depth.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bfb9b03d-63c9-4d31-b208-5d90d200530e">&gt;Recommended Display CIELab Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d928b3bd-9f60-410c-88b1-ba1b5ffc111b">(0062,000D)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f7c85e11-e1ce-4b5d-ab69-ec97c2c88aad">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c96c7816-67a4-4a96-9cef-c59e5806f8f8">A default triplet value in which it is recommended that the contour be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See <xref linkend="sect_C.10.7.1.1" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8df2ae10-15f6-498d-b728-bf1db5c17d01">&gt;Contour Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_32e5d070-c142-4396-b984-21acd3c1bbf8">(3006,0040)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_af62dfaa-d140-4332-b651-d4da0b8124fe">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1fc0a3e2-6e73-4da1-96fa-d1c59d88e666">Sequence of Contours defining ROI. One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4cce654e-16fe-4755-8df8-0461d943dfd8">&gt;&gt;Contour Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_27235739-4922-43af-a293-2ac98a3ec801">(3006,0048)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8ef7b0fd-f010-4c6b-a2c9-978747282911">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e43f6b09-90cd-4e51-b5f6-af357782adfe">Identification number of the contour. The value of Contour Number (3006,0048) shall be unique within the Contour Sequence (3006,0040) in which it is defined. No semantics or ordering shall be inferred from this Attribute.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9b9d1970-c506-47f5-80e1-42faf748a527">&gt;&gt;Attached Contours</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f56828d0-be5b-400e-956a-6b4cc3b63668">(3006,0049)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_26a7da3c-9605-4092-b013-c1496a7a74a3">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3a18f164-0bc6-4e3c-ad03-f517d6a71594">List of Contour Number (3006,0048) defining lower-numbered contour(s) to which the current contour is connected.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8e05d445-4adf-4f97-ab98-b186990d005b">&gt;&gt;Contour Image Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_27a768ce-6b31-4d8b-b427-a1092bfc0ff4">(3006,0016)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9bc88903-b61a-4707-9641-20ea6c6c2cd4">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cf846a41-9f07-49c2-aecc-a4aed7db6abb">Sequence of images containing the contour.</para>
                                    <para xml:id="para_ac5cb41b-13b3-48a9-96bd-1ca464853fb8">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_be87dfe9-6d09-451c-b532-72bb4adcc607">
                                        <emphasis role="italic">&gt;&gt;&gt;Include <xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8c810973-61a2-4118-ba06-4ab9fa2828fd"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fc1c7979-9d22-45e9-b8c7-9067e6d51165">&gt;&gt;Contour Geometric Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3d15b54d-3174-4c5f-aa58-bfc446ef300e">(3006,0042)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8cb6b935-865a-49f3-b171-6d09f840ba85">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6e5d0251-61f0-494d-af68-7d91e9e4b396">Geometric type of contour. See <xref linkend="sect_C.8.8.6.1" xrefstyle="select: label"/>.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>POINT</term>
                                            <listitem>
                                                <para xml:id="para_02eceb48-7c94-4ae8-bd79-1089b35d2c58">single point</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>OPEN_PLANAR</term>
                                            <listitem>
                                                <para xml:id="para_9e6c5b1b-0d3e-42c0-aebd-5a103b3f3a46">open contour containing coplanar points</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>OPEN_NONPLANAR</term>
                                            <listitem>
                                                <para xml:id="para_a8028517-a4cf-49ca-9eb1-09c3dbb39df4">open contour containing non-coplanar points</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>CLOSED_PLANAR</term>
                                            <listitem>
                                                <para xml:id="para_45b2eb68-f3a3-4cde-8b26-cb699f4a3c96">closed contour (polygon) containing coplanar points</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_190da16d-8231-4558-b225-750e80a37ce4">&gt;&gt;Contour Slab Thickness</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_79e5e8db-cc44-4d28-9a5d-57577b515641">(3006,0044)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_df66ecc4-79b2-4f21-b979-66fb058ba009">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0dbcb249-0fa1-417b-a721-5d5f363bcae0">Thickness of slab (in mm) represented by contour, where Contour Data (3006,0050) defines a plane in the center of the slab, offset by the Contour Offset Vector (3006,0045) if it is present. See <xref linkend="sect_C.8.8.6.2" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_87ca8c22-e0bb-43e7-b6da-2af9e98bf663">&gt;&gt;Contour Offset Vector</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0734ccb0-3a83-46f6-972b-45eee5263033">(3006,0045)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_225996fd-a7cf-44b7-8d65-f486f9e6c39a">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7b95673a-e114-4aa4-9790-b7661a1ddaf3">Vector (x,y,z) in the Patient-Based Coordinate System described in <xref linkend="sect_C.7.6.2.1.1" xrefstyle="select: label"/> that is normal to plane of Contour Data (3006,0050), describing direction and magnitude of the offset (in mm) of each point of the central plane of a contour slab from the corresponding original point of Contour Data (3006,0050). See <xref linkend="sect_C.8.8.6.2" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e9423481-47ea-4972-8008-cffb0baa8086">&gt;&gt;Number of Contour Points</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4612ca11-477f-48f0-87aa-d102e52634d0">(3006,0046)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9b138bde-064b-4661-8bad-6294b75a2287">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f1811bbb-d9eb-4d40-8108-958c2856e7e1">Number of points (triplets) in Contour Data (3006,0050).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_74f33656-9ca3-48da-9950-b5848204b25c">&gt;&gt;Contour Data</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a1cdf87f-08a6-4c4a-ab24-cb606d0b8dec">(3006,0050)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1ab22818-e605-4482-94b9-7967009f277e">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6c43c3f1-7783-4b16-8a12-81d330da3f3e">Sequence of (x,y,z) triplets defining a contour in the Patient-Based Coordinate System described in <xref linkend="sect_C.7.6.2.1.1" xrefstyle="select: label"/> (mm). See <xref linkend="sect_C.8.8.6.1" xrefstyle="select: label"/> and <xref linkend="sect_C.8.8.6.3" xrefstyle="select: label"/>.</para>
                                    <note>
                                        <para xml:id="para_23e73451-d6b6-436b-a13a-76d0e3b341b6">Contour Data may not be properly encoded if Explicit VR Transfer Syntax is used and the VL of this Attribute exceeds 65534 bytes.</para>
                                    </note>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.8.8.6.1" status="5" xml:id="sect_C.8.8.6.1">
                        <title>Contour Geometric Type</title>
                        <para xml:id="para_8cfa02b0-0fd6-4ef7-81d5-10a4f55bf357">A contour can be one of the following geometric types:</para>
                        <itemizedlist>
                            <listitem>
                                <para xml:id="para_e65e35c3-1b98-4b61-98f6-e5f27984ca7b">A Contour Geometric Type (3006,0042) of POINT indicates that the contour is a single point, defining a specific location of significance.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_6585f4e6-ef90-43d6-a373-b21dc9d82c22">A Contour Geometric Type (3006,0042) of OPEN_PLANAR indicates that the last vertex shall not be connected to the first point, and that all points in Contour Data (3006,0050) shall be coplanar.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_598b0539-d9f4-45b0-a051-29e8ba448d51">A Contour Geometric Type (3006,0042) of OPEN_NONPLANAR indicates that the last vertex shall not be connected to the first point, and that the points in Contour Data (3006,0050) may be non-coplanar. Contours having a Geometric Type (3006,0042) of OPEN_NONPLANAR can be used to represent objects best described by a single, possibly non-coplanar curve, such as a brachytherapy applicator.</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_04e4e58c-522d-44c8-9d35-c4e8a8dc6079">A Contour Geometric Type (3006,0042) of CLOSED_PLANAR indicates that the last point shall be connected to the first point, where the first point is not repeated in Contour Data (3006,0050). All points in Contour Data (3006,0050) shall be coplanar.</para>
                            </listitem>
                        </itemizedlist>
                    </section>
                    <section label="C.8.8.6.2" status="5" xml:id="sect_C.8.8.6.2">
                        <title>Contour Slab Thickness</title>
                        <para xml:id="para_e8a8aa7e-34a4-473e-b924-d11e9df4e0c4">A set of Contour slabs may define a multi-slab Volume of Interest. Contour Slab Thickness (3006,0044) shall specify the thickness of a slab, the central plane of which shall be defined by the set of points offset from Contour Data (3006,0050) by the value of Contour Offset Vector (3006,0045). One contour slab may contain one to many sets of Contour Data (3006,0050) that may define regions of one complex Volume of Interest. If no valid value of Contour Slab Thickness (3006,0044) is present, then the offset value shall be (0,0,0) and the original Contour Data (3006,0050) shall define the central plane of the Contour slab.</para>
                    </section>
                    <section label="C.8.8.6.3" status="5" xml:id="sect_C.8.8.6.3">
                        <title>Representing Inner and Outer Contours on an Image</title>
                        <para xml:id="para_4717f0b2-824c-4c35-a532-4c1b3f88d069">When a single ROI describes an excluded inner volume, this can be encoded with a single contour, using a "keyhole" technique. In this method, an arbitrarily narrow channel is used to connect the outer contour to the inner contour, so that it is drawn as a single contour. An example of such a structure is shown in <xref linkend="figure_C.8.8.6-1" xrefstyle="select: label"/>
                        </para>
                        <para xml:id="para_76e61d5a-affc-4165-87e8-68ee542e68fb">Points in space lying along the path defined by the contour are considered to be inside the ROI.</para>
                        <para xml:id="para_23450dac-0dcc-4498-a3d7-cd7f197cb70e">
                            <figure label="C.8.8.6-1" pgwide="1" xml:id="figure_C.8.8.6-1">
                                <title>Example of ROI with excluded inner volume</title>
                                <mediaobject>
                                    <imageobject>
                                        <imagedata fileref="figures/PS3.3_C.8.8.6-1.svg"/>
                                        <!--<imagedata fileref="part03_fromword_files/image195.png"/>-->
                                    </imageobject>
                                </mediaobject>
                            </figure>
                        </para>
                    </section>
                </section>
                <section label="C.8.8.7" status="4" xml:id="sect_C.8.8.7">
                    <title>RT Dose ROI Module</title>
                    <para xml:id="para_bb7a2259-e069-4ebe-b8d9-d69bdef4a79e">RT Dose ROI provides ancillary dose-related information to the ROI data defined within the <xref linkend="sect_C.8.8.5" xrefstyle="select: title"/> and <xref linkend="sect_C.8.8.6" xrefstyle="select: title"/>, which may be included in the <xref linkend="sect_A.18" xrefstyle="select: title"/> composite object. These Modules in combination provide for the definition of dose data in the form of <emphasis role="italic">isodose curves</emphasis> or named or unnamed dose points. Isodose curves in radiation oncology are simply contours identifying a set of points with the same dose value.</para>
                    <table frame="box" label="C.8-43" rules="all" xml:id="table_C.8-43">
                        <caption>RT Dose ROI Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c92edf72-728a-4ac4-acfe-ea6131450514">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5def89c9-f257-4927-bc96-659137e5ae79">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f7208e45-cd08-42a4-9982-d00799dc2814">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8286c100-c60d-4552-a66e-6494228874d0">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e09b4b00-43db-4c42-ac9a-5bf44fadead5">RT Dose ROI Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0591154e-ac63-4f6d-ad99-80b08ad9d005">(3004,0010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_124e2911-7394-4d31-ac75-3221b20c7f96">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c920d201-0fda-4cf8-9cae-34d5cd6d16cd">Sequence specifying dose levels for isodose curves or dose points described in the ROI Module.</para>
                                    <para xml:id="para_ca9d686d-fe43-4e51-842a-1d72b2675bed">One or more Items shall be included in this Sequence.</para>
                                    <para xml:id="para_44efca1e-c30a-4467-9535-d498f8b03ab9">See <xref linkend="sect_C.8.8.7.1" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a66fe3ec-96a9-4c5c-85e8-82c67adabfd3">&gt;Referenced ROI Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_97f49eaf-da67-4ec6-af3d-d451ce439ef8">(3006,0084)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_50b6e6a3-0f75-4a42-89aa-b0dc422df75b">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_57bd01c0-de2d-4d01-aec3-261b28a4d452">Uniquely identifies the referenced ROI within the current RT Dose. See Note 1 and <xref linkend="sect_C.8.8.7.2" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4e06493e-bf14-4161-858b-fea8b3f1e8a2">&gt;Dose Units</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_fce3020a-7c6a-4bb4-8aac-8784dbfe5830">(3004,0002)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_85c9e33c-281d-4389-8969-f74f64eee676">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_35beaee0-e020-4cd8-8dad-b9e62d6a4646">Units used for ROI Dose.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>GY</term>
                                            <listitem>
                                                <para xml:id="para_53004144-9530-455f-a872-495858b2b1aa">Gray</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>RELATIVE</term>
                                            <listitem>
                                                <para xml:id="para_47c2fddf-793c-4d86-a5fa-0789ce837445">dose relative to implicit reference value</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_727da382-c803-4629-9708-f4e4f5a36f0a">&gt;Dose Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_892ca9c4-d455-4888-a406-6364e5c453d3">(3004,0012)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c18c5fab-0196-41c0-8d6f-a839029cd03c">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a69fe7a5-7ba7-4d48-84e1-244b09b7ca13">Dose value for ROI, in units defined by Dose Units (3004,0002). See <xref linkend="sect_C.8.8.7.3" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <note>
                        <orderedlist numeration="arabic">
                            <listitem>
                                <para xml:id="para_fc27c749-6d5d-4207-911a-9e961052cacc">The Structure Set ROI Sequence (3006,0020) defining the dose point and surfaces is defined in the <xref linkend="sect_C.8.8.5" xrefstyle="select: title"/>. ROI Number (3006,0022) is unique within the Structure Set ROI Sequence, and is referenced from the <xref linkend="sect_C.8.8.7" xrefstyle="select: title"/> using Referenced ROI Number (3006,0084).</para>
                            </listitem>
                            <listitem>
                                <para xml:id="para_808e2f81-06dd-4a77-83d3-410ef6b41f68">The <xref linkend="sect_C.8.8.7" xrefstyle="select: title"/> defines the Attributes that describe references to ROIs contained within the associated <xref linkend="sect_C.8.8.5" xrefstyle="select: title"/> and <xref linkend="sect_C.8.8.6" xrefstyle="select: title"/>. Note that the RT Dose IOD Module table specifies that either all or none of the <xref linkend="sect_C.8.8.5" xrefstyle="select: title"/>, <xref linkend="sect_C.8.8.6" xrefstyle="select: title"/>, and <xref linkend="sect_C.8.8.7" xrefstyle="select: title"/> must be present in the <xref linkend="sect_A.18" xrefstyle="select: title"/>.</para>
                            </listitem>
                        </orderedlist>
                    </note>
                    <section label="C.8.8.7.1" status="5" xml:id="sect_C.8.8.7.1">
                        <title>Contour Geometric Type of Referenced ROI</title>
                        <para xml:id="para_bc3c6f5f-4d92-47ba-b1e9-02f3ac49fdfc">ROIs referenced in the <xref linkend="sect_C.8.8.7" xrefstyle="select: title"/> shall have a Contour Geometric Type (3006,0042) of POINT, OPEN_PLANAR or CLOSED_PLANAR.</para>
                    </section>
                    <section label="C.8.8.7.2" status="5" xml:id="sect_C.8.8.7.2">
                        <title>Referenced ROI Number</title>
                        <para xml:id="para_ece0ef1b-aa3f-44e0-96ac-3f5db2c07e8c">There shall be a one-to-one correspondence between Referenced ROI Number (3006,0084) and the Sequence of ROIs defined in the <xref linkend="sect_C.8.8.5" xrefstyle="select: title"/> and <xref linkend="sect_C.8.8.6" xrefstyle="select: title"/>. The <xref linkend="sect_C.8.8.7" xrefstyle="select: title"/> shall only contain references to structures that are dose-related (i.e., dose points and isodose curves).</para>
                    </section>
                    <section label="C.8.8.7.3" status="5" xml:id="sect_C.8.8.7.3">
                        <title>Dose Value</title>
                        <para xml:id="para_332f092f-9661-4311-90e6-7c3d8e44d981">Dose Value (3004,0012) shall be the dose value corresponding to the referenced isodose curve, named dose point, or unnamed dose point.</para>
                    </section>
                </section>
                <section label="C.8.8.8" status="4" xml:id="sect_C.8.8.8">
                    <title>RT ROI Observations Module</title>
                    <para xml:id="para_6d1e71f3-f40d-4f5c-960a-7073f749af83">The RT ROI Observations Module specifies the identification and interpretation of an ROI specified in the <xref linkend="sect_C.8.8.5" xrefstyle="select: title"/> and <xref linkend="sect_C.8.8.6" xrefstyle="select: title"/>.</para>
                    <table frame="box" label="C.8-44" rules="all" xml:id="table_C.8-44">
                        <caption>RT ROI Observations Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_eea39f9d-5f2f-4c95-b226-22e91892a21f">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a25cdae7-9c78-43e1-9c28-f2dca8500693">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_eb5d0f52-164d-416f-a838-27032f74cad3">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a03dbf46-49f3-4599-be6e-19d524a9fd51">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5e592f6e-3985-4fe8-8c4a-b8cdec5e50f8">RT ROI Observations Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_21b30a9d-9943-4659-98ff-248ffa1dad0e">(3006,0080)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_24620485-78ec-4097-8af7-52c0a09ac9d4">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_15c76e79-b5b6-4a40-a781-d8ef63d6b13e">Sequence of observations related to ROIs defined in the ROI Module.</para>
                                    <para xml:id="para_6d45fc27-329e-4971-85fd-cd66d46cce43">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3aeee1f9-9284-4015-99e5-d35b8776653a">&gt;Observation Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a0ab537a-03b4-4c87-bc1c-dffaf1229165">(3006,0082)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_860006ed-75fc-416e-8e69-3d1903afc4a6">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0d3d5c98-baeb-4690-ba2c-89babf6ff810">Identification number of the Observation. The value of Observation Number (3006,0082) shall be unique within the RT ROI Observations Sequence (3006,0080).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e67d12b9-1705-4e21-bfc9-05a5fb3ad858">&gt;Referenced ROI Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_146f3de5-4b46-4361-bc93-b371ebf8c226">(3006,0084)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ab95c138-1f07-4887-93af-165033ee5c3b">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1f0e99f4-5aa3-4ef5-8d59-a84de51e8c0b">Uniquely identifies the referenced ROI described in the Structure Set ROI Sequence (3006,0020).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_dbdf0947-748c-49e1-9933-c4cb667a327b">&gt;ROI Observation Label</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_17b3a888-0fa6-480d-9d8d-2a73c3d738ac">(3006,0085)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2f6e2bda-3600-4498-82cb-4c4530f7ba50">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ad5e7149-93fc-43b8-bc50-5f76b3ba6855">User-defined label for ROI Observation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bd555e98-282f-42f8-95d9-861b7805be6f">&gt;ROI Observation Description</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_047c5e04-5e65-483b-aa40-bca1bdbac7d5">(3006,0088)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bd50187f-3136-4baf-a87c-86c9656fb993">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c2ec7891-2c1e-4f5b-a821-6a04ac472ef6">User-defined description for ROI Observation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c612f63a-4d61-403b-90c9-cef0d6e0dbd9">&gt;RT Related ROI Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8843f340-0705-441e-bf7e-ef6ca77ddce0">(3006,0030)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7c3cc47e-5f69-452a-b7c7-800540b81b47">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_062f5098-2390-4c76-82c1-4a6973b6a8d7">Sequence of significantly related ROIs, e.g., CTVs contained within a PTV.</para>
                                    <para xml:id="para_f2e2898a-40b1-44b7-ad8c-ab3e53685626">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3d39d30e-86c0-4313-94af-71083798069f">&gt;&gt;Referenced ROI Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_59dfc81d-5ab9-4f20-8ebe-3ebe8b91c3fd">(3006,0084)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f22b289d-42b6-4609-8333-03b356a8af0d">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_42263d1b-0ac0-435c-9bb8-1533d22f63b1">Uniquely identifies the related ROI described in the Structure Set ROI Sequence (3006,0020).</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c7f5c641-effd-4365-9754-1189fa6ad340">&gt;&gt;RT ROI Relationship</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a6a93fc3-b282-4f72-982b-9fc05119c327">(3006,0033)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_30d59995-bef1-4a7e-92eb-3077ff8b8b2d">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_09dde0d6-3d8d-4761-a07e-4d1f58b334ac">Relationship of referenced ROI with respect to referencing ROI.</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>SAME</term>
                                            <listitem>
                                                <para xml:id="para_e3c51377-b8a7-4f03-9c08-40431a93c2d4">ROIs represent the same entity</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>ENCLOSED</term>
                                            <listitem>
                                                <para xml:id="para_75de755c-38c9-4807-82cb-3a396e56e7d8">referenced ROI completely encloses referencing ROI</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>ENCLOSING</term>
                                            <listitem>
                                                <para xml:id="para_fb90f5c9-3bda-4123-8bf8-8f8ec08376c5">referencing ROI completely encloses referenced ROI</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_341ad623-b05e-43f9-87de-21773543ebbb">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_10-7b" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2ff98cc2-1697-49bf-aa05-704987766e71">
                                        <emphasis role="italic">May not be necessary if the anatomy is implicit in the RT ROI Identification Code Sequence (3006,0086).</emphasis>
                                    </para>
                                    <para xml:id="para_9d15a87a-dcdf-45fa-ae87-e599e39d9959">
                                        <emphasis role="italic">More than one item in Anatomic Region Sequence (0008,2218) may be used when a region of interest spans multiple anatomical locations and there is not a single pre-coordinated code describing the combination of locations. There is no requirement that the multiple locations be contiguous.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_56b10514-1c91-496e-95bb-ccc493c64ee0">&gt;Segmented Property Category Code Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8a4cf5e1-809f-47da-87b4-403e3a3158ea">(0062,0003)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_96fcf903-21de-4ff4-a11d-3a0e8895ab4f">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e3047e90-4f38-46be-81a4-7f2830e11a53">Sequence defining the general category of the property this ROI represents.</para>
                                    <para xml:id="para_c15ed8d2-0c74-438e-8cb4-f9a274168737">Only a single Item is permitted in this Sequence.</para>
                                    <para xml:id="para_e8452b7f-558c-48ab-8e8a-2c662d9a9fbf">See Note <xref linkend="note_C.8.8.8_1"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_5e54981e-7b2f-46e0-9d15-1c67c1d0634d">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_53ba7a64-3763-4509-9fd9-7e5a57ee1ff9">
                                        <emphasis role="italic">No Baseline CID is defined.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_967a22fb-2c7e-40d3-8d25-076401ac9226">&gt;RT ROI Identification Code Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ad37d68e-2abb-47b8-8c97-e3f4f6a47e39">(3006,0086)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5ed62576-cddd-4b1b-b0fd-57417c3265a7">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_27a599df-276c-4830-a9a3-a5892bed369e">Sequence containing Code used to identify ROI.</para>
                                    <para xml:id="para_37c8171e-aa06-48c4-b29c-b77cdf191556">Only a single Item is permitted in this Sequence.</para>
                                    <para xml:id="para_434eda07-369c-48b9-9319-b51bd12c13e9">See Note <xref linkend="note_C.8.8.8_2"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_a0598b5e-4cec-442c-bd8e-fa692ab7de4c">
                                        <emphasis role="italic">&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cf304c99-551e-42f0-93e9-a34b7576be6c">
                                        <emphasis role="italic">For dose measurement devices Baseline <olink targetdoc="PS3.16" targetptr="sect_CID_7026" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                                    </para>
                                    <para xml:id="para_001c172b-dd6f-45bb-9bf4-e3fa4c40f88d">
                                        <emphasis role="italic">No Baseline CID is defined otherwise.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_38f3636a-5078-4356-9539-ef2cfc21cb89">&gt;&gt;Segmented Property Type Modifier Code Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_998942ec-026e-4243-8830-b74a04d99dd6">(0062,0011)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1692011c-f377-434e-8359-b696bdd545e0">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_17542c36-efc1-495d-b59e-4b714c8d5851">Sequence defining the modifier of the property type of this segment.</para>
                                    <para xml:id="para_77911bf8-4d2e-48c6-8785-43fddd8142c8">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_e95d1ed2-300f-4609-ade4-7fca220dfc4d">
                                        <emphasis role="italic">&gt;&gt;&gt;Include <xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a493e4fc-9509-4417-b78a-7ba8ce1492a5">
                                        <emphasis role="italic">D<olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e8aa860c-3387-4032-92d0-cce6c895fa9f">&gt;Related RT ROI Observations Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_db027b80-d3cf-487a-83e6-35b8f93b4f40">(3006,00A0)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_67adee22-b48e-40b6-8fa3-d5475419cf11">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_15bf840e-d432-40e2-832c-1c703dfb12a2">Sequence of related ROI Observations.</para>
                                    <para xml:id="para_03a9d37c-5449-4b62-8e24-c0c6f5da80d6">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8d65b478-0601-4e1e-bf58-9920dd503369">&gt;&gt;Observation Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5082599a-6af5-4538-ab63-6212f918b222">(3006,0082)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bea27541-441e-419a-8a06-c72a645db055">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d96f7bfa-4525-4722-95fd-8bcf1387d8c1">Uniquely identifies a related ROI Observation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5558a767-44a0-4508-a4bc-9852b0b29880">&gt;RT ROI Interpreted Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e7373dab-c410-4987-932f-43c62ac32308">(3006,00A4)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_eb16935d-7f8a-4047-9015-93c00f36be03">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7a3d82b2-5a09-485a-922a-e02e388dd7c3">Type of ROI. See <xref linkend="sect_C.8.8.8.1" xrefstyle="select: label"/>.</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>EXTERNAL</term>
                                            <listitem>
                                                <para xml:id="para_6137987d-0502-449d-b352-776b9fec168e">external patient contour</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>PTV</term>
                                            <listitem>
                                                <para xml:id="para_4eebdf41-9f3e-4831-b40f-9af2ad3f8953">Planning Target Volume (as defined in <xref linkend="biblio_ICRU50"/>)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>CTV</term>
                                            <listitem>
                                                <para xml:id="para_f3932cd1-3387-4efc-bea7-0eaedff08be7">Clinical Target Volume (as defined in <xref linkend="biblio_ICRU50"/>)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>GTV</term>
                                            <listitem>
                                                <para xml:id="para_05febe90-7dd1-4b81-8742-520468b50d98">Gross Tumor Volume (as defined in <xref linkend="biblio_ICRU50"/>)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>TREATED_VOLUME</term>
                                            <listitem>
                                                <para xml:id="para_568d5acd-157d-4418-b7d8-a0e19e71694f">Treated Volume (as defined in <xref linkend="biblio_ICRU50"/>)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>IRRAD_VOLUME</term>
                                            <listitem>
                                                <para xml:id="para_e0b1bb68-4b7b-4f07-8a53-16bbd4a4f433">Irradiated Volume (as defined in <xref linkend="biblio_ICRU50"/>)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BOLUS</term>
                                            <listitem>
                                                <para xml:id="para_f3145084-1c0f-4fdb-b500-1120fde2adde">patient bolus to be used for external beam therapy</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>AVOIDANCE</term>
                                            <listitem>
                                                <para xml:id="para_c3d53673-cdbc-483c-b231-356887807a05">region in which dose is to be minimized</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>ORGAN</term>
                                            <listitem>
                                                <para xml:id="para_d5e1e478-6a41-473a-8c00-2205020855f8">patient organ</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>MARKER</term>
                                            <listitem>
                                                <para xml:id="para_a0e69d3d-34ca-470e-9c1f-e3822c031aab">patient marker or marker on a localizer</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>REGISTRATION</term>
                                            <listitem>
                                                <para xml:id="para_70d6d3d7-98ad-4188-81c4-1e090b86b476">registration ROI</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>ISOCENTER</term>
                                            <listitem>
                                                <para xml:id="para_81f9c33e-eeb2-4ca6-8d6b-2765ab65668a">treatment isocenter to be used for external beam therapy</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>CONTRAST_AGENT</term>
                                            <listitem>
                                                <para xml:id="para_5746b35b-a75e-4a99-b951-c79078ba31ba">volume into which a contrast agent has been injected</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>CAVITY</term>
                                            <listitem>
                                                <para xml:id="para_c99da971-e707-42f1-9501-d312b7d475bd">patient anatomical cavity</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BRACHY_CHANNEL</term>
                                            <listitem>
                                                <para xml:id="para_92627cbd-d575-4d05-b70e-06383d475599">brachytherapy channel</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BRACHY_ACCESSORY</term>
                                            <listitem>
                                                <para xml:id="para_ffee4548-744d-44d6-8d1a-ddc00949e44f">brachytherapy accessory device</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BRACHY_SRC_APP</term>
                                            <listitem>
                                                <para xml:id="para_2b8ac04e-81f4-4174-b956-99e1a8d04435">brachytherapy source applicator</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>BRACHY_CHNL_SHLD</term>
                                            <listitem>
                                                <para xml:id="para_317c9cf0-2d69-4c75-88c7-957fe1d26a3e">brachytherapy channel shield</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>SUPPORT</term>
                                            <listitem>
                                                <para xml:id="para_6f36868f-4b9c-436c-a0e9-b4639672948c">external patient support device</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>FIXATION</term>
                                            <listitem>
                                                <para xml:id="para_3bd1a102-fdda-4902-87f2-e120e2548b32">external patient fixation or immobilization device</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>DOSE_REGION</term>
                                            <listitem>
                                                <para xml:id="para_2a3bce51-b14a-45a3-b7e8-0b9d801fe617">ROI to be used as a dose reference</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>CONTROL</term>
                                            <listitem>
                                                <para xml:id="para_94f58953-895d-40c0-b026-a2b8aeb37940">ROI to be used in control of dose optimization and calculation</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>DOSE_MEASUREMENT</term>
                                            <listitem>
                                                <para xml:id="para_1464d8aa-cd21-4b44-9c91-af145228fae6">ROI representing a dose measurement device, such as a chamber or TLD</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_59a7a879-5f21-4073-87ff-514414ea2f58">&gt;ROI Interpreter</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_41009fc1-7b4f-496b-9b2a-456c2d7c9edf">(3006,00A6)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0ed5a92f-692a-4fa3-9817-4d1af05b9e16">2</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9a89fe46-d5b2-49ff-b0a6-501f891c0ad0">Name of person performing the interpretation.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bdb23b1d-ebf9-4ccd-9640-3eeab94031b9">&gt;Material ID</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b298cb3e-97ef-4d3b-854e-735d28fc36fd">(300A,00E1)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_33edb85e-0764-412d-aa7c-50c556eb3e68">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1d26d4e4-539b-41b3-b1d4-0d0f9c4c4885">User-supplied identifier for ROI material.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5bbde7a1-2d27-4f98-a867-e03f818acbba">&gt;ROI Physical Properties Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f4092fbd-6ce0-44ae-b421-047bdfbbd24e">(3006,00B0)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_624a7acf-bc7c-497f-8c46-629be25937db">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d331a160-3d76-4826-a9a7-a5a81fcb81b5">Sequence describing physical properties associated with current ROI interpretation.</para>
                                    <para xml:id="para_24aabc4e-c198-4566-a721-ce790c26099c">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_20076978-0a7d-4413-8cce-bccce00b4013">&gt;&gt;ROI Physical Property</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_36df4947-9008-471c-a19c-7c41c3260710">(3006,00B2)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f0403a6b-1f60-4b4b-abdd-54f82d74abd4">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8265d015-4472-40a5-b4be-89376d320925">Physical property specified by ROI Physical Property Value (3006,00B4).</para>
                                    <variablelist spacing="compact">
                                        <title>Defined Terms:</title>
                                        <varlistentry>
                                            <term>REL_MASS_DENSITY</term>
                                            <listitem>
                                                <para xml:id="para_465ff7ac-6b55-4d97-92de-776991fe59d8">mass density relative to water</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>REL_ELEC_DENSITY</term>
                                            <listitem>
                                                <para xml:id="para_67c77e6e-a17a-4baf-a84d-66c0a4053f0d">electron density relative to water</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>EFFECTIVE_Z</term>
                                            <listitem>
                                                <para xml:id="para_176e2457-a73d-4146-9b6c-a2d1e7dec905">effective atomic number</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>EFF_Z_PER_A</term>
                                            <listitem>
                                                <para xml:id="para_90ca5e20-59a7-4187-83ba-57c841f6a845">ratio of effective atomic number to mass (AMU<superscript>-1</superscript>)</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>REL_STOP_RATIO</term>
                                            <listitem>
                                                <para xml:id="para_a3ff145e-6317-467c-99f9-9a9cdc572c5e">ratio of linear stopping power of material relative to linear stopping power of water</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>ELEM_FRACTION</term>
                                            <listitem>
                                                <para xml:id="para_63b43cdf-fc6b-4902-9349-4cc1e04edf0d">elemental composition of the material</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>MEAN_EXCI_ENERGY</term>
                                            <listitem>
                                                <para xml:id="para_d878b8aa-af22-4041-8a8b-7b97351a2176">Mean Excitation Energy of the material (eV)</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3da963f1-5f16-4f83-ae5d-bfd4a4b64d0c">&gt;&gt;ROI Elemental Composition Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6fb3821d-f518-4bf8-b0e1-044c95ee8d25">(3006,00B6)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_86c547af-61d3-4604-ac03-25f586b22e4c">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_76f24b2b-6266-4ef9-98c0-07fab1e41909">The elemental composition of the ROI and the atomic mass fraction of the elements in the ROI.</para>
                                    <para xml:id="para_5b507190-bee8-4b43-86da-94920c9dc707">Required if ROI Physical Property (3006,00B2) equals ELEM_FRACTION.</para>
                                    <para xml:id="para_8f8437fb-05f6-4490-8f87-0e4a032e1538">One or more Items shall be included in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_43e360b4-c35e-4fa4-9034-c36cb9efbc89">&gt;&gt;&gt;ROI Elemental Composition Atomic Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_102fbf0e-aa71-48e7-a577-248b4ee865b6">(3006,00B7)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d434a899-f990-4109-8c24-eb1e5500e771">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_e7e88db7-bf93-49d4-89e3-cef97b997323">The atomic number of the element for which the ROI Elemental Composition Sequence (3006,00B6) is present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_afe68961-12d8-4c34-99bb-646c82698e3e">&gt;&gt;&gt;ROI Elemental Composition Atomic Mass Fraction</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a8ee4a30-bd34-415e-aae8-679ea1035bf8">(3006,00B8)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_45fc77d0-9b67-4da9-8f73-4df9776c5fd9">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f17cbf7d-caa9-420d-b741-70a66c3e9d9c">The fractional weight of the element for which the ROI Elemental Composition Sequence (3006,00B6) is present.</para>
                                    <para xml:id="para_7ceccfeb-b2a4-44da-8e9b-5ecc423ffec9">The sum of all ROI Elemental Composition Atomic Mass Fractions (3006,00B8) within the ROI Elemental Composition Sequence (3006,00B6) shall equal 1.0 within acceptable limits of floating point precision.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_47d41bfc-6190-417e-ba00-08e81425ab8e">&gt;&gt;ROI Physical Property Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ddd38754-0a01-4a26-beb9-484e959db67e">(3006,00B4)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a1c5e3f7-36ed-444d-a534-d0e2ee624db6">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_2ca94cd1-0be4-4fbb-b82d-c8cbfdc616b0">User-assigned value for physical property.</para>
                                    <note>
                                        <para xml:id="para_5e13c03b-0463-4c50-b6a5-c719bebe45e1">The value has no meaning when ROI Physical Property (3006,00B2) has the value ELEM_FRACTION. Therefore this Attribute may contain any value and receivers may ignore that value.</para>
                                    </note>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <note>
                        <orderedlist>
                            <listitem xml:id="note_C.8.8.8_1">
                                <para xml:id="para_fc5e3924-a5e3-4649-a1fc-a853541890d7">This Attribute allows preserving information by copying the content of Segmented Property Category Code Sequence (0062,0003) in case a Segmentation object is re-encoded as an RT Structure Set or vice-versa.</para>
                            </listitem>
                            <listitem xml:id="note_C.8.8.8_2">
                                <para xml:id="para_5e6a4a4c-2c03-493c-98a3-3159a14ad5c8">In case of re-encoding a Segmentation object as an RT Structure Set or vice-versa it is suggested, that the Segmented Property Type Code Sequence (0062,000F) is mapped to RT ROI Identification Code Sequence (3006,0086).</para>
                            </listitem>
                        </orderedlist>
                    </note>
                    <section label="C.8.8.8.1" status="5" xml:id="sect_C.8.8.8.1">
                        <title>RT ROI Interpreted Type</title>
                        <para xml:id="para_878d026f-7ace-4251-9687-22848f079272">RT ROI Interpreted Type (3006,00A4) shall describe the class of ROI (e.g., CTV, PTV). Individual instances of each class of structure (e.g., CTV1, CTV2) can be distinguished using ROI Observation Label (3006,0085).</para>
                    </section>
                    <section label="C.8.8.8.2" status="5" xml:id="sect_C.8.8.8.2">
                        <title>Additional RT ROI Identification Code Sequence</title>
                        <para xml:id="para_8f54765c-7010-4773-af72-c4e891110369">Retired. See DICOM PS3.3 2016c.</para>
                    </section>
                </section>

        
        
        
        
        



        
        
        
        
        
    </chapter>
</book>
