<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2031</title>
    <subtitle>Add Frame of Reference Module to Surface Scan Mesh and Point Cloud IODs</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/09/08</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie <link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Kari Jyrkkälä <link xl:href="mailto:kari.jyrkkala@gmail.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2019/11/25</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2031</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add Frame of Reference Module to Surface Scan Mesh and Point Cloud IODs</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>The Surface Scan Mesh and Surface Scan Point Cloud IODs need to contain the Frame of Reference Module, since the points they include are described in terms of an FoR UID. Add it.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <table frame="box" label="A.1-2" rules="all" xml:id="table_A.1-2">
          <caption>Composite Information Object Modules Overview - Non-Images</caption>
          <thead>
            <tr valign="top">
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_25535feb-a835-4595-b165-4bc34190bdf6">
                  <emphasis role="bold">IODs</emphasis>
                </para>
                <para xml:id="para_eca824ef-681e-4e95-8363-4d2a7f33bee6">
                  <emphasis role="bold">Modules</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0ee91eed-1c5e-454d-8384-0b745827e414">
                                <emphasis role="bold">MR Spectroscopy</emphasis>
                            </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_81f35238-0f65-4f7a-bbd1-dbe47f59f0ae">
                                <emphasis role="bold">Raw Data</emphasis>
                            </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4a16ba28-b38d-4861-8a50-8cceb084124a">
                                <emphasis role="bold">Surface Scan Point Cloud</emphasis>
                            </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b283d8e2-5d67-44fe-b618-f5f2649d8772">
                                <emphasis role="bold">Surface Scan Mesh</emphasis>
                            </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a77df48b-8838-4c6c-b9b1-e99b8bf5b843">
                                <emphasis role="bold">Content Assessment Result</emphasis>
                            </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8487b16-3d8c-476e-a312-f2e56a10c51c">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_948d2aec-7425-46eb-85e2-8eb50eab6dff"/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c79cc6f6-79da-40b1-b81e-13bf4242dcf4"/>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e305c64a-a843-4b94-ae16-cc86fe321822">Frame of Reference</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5ef69945-eef3-4fb1-b379-2efc1ca009b5">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f3aaf2da-1ef7-4522-b3ef-a26752736af8">U</para>
              </td>
                <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
                </td>
              <td align="left" colspan="1" rowspan="1">
                  <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54a15c30-2a7f-451c-ae4c-b9aa3026e305">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43cf6169-5836-4957-9d27-28bb05ddd89e"/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6e085b9-11b9-45da-8df2-9bf65ee60023"/>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </table>

        <section label="A.68" status="2" xml:id="sect_A.68">
            <title>Surface Scan Mesh IOD</title>
            <section label="A.68.1" status="3" xml:id="sect_A.68.1">
                <title>Surface Scan Mesh IOD Description</title>
            </section>
            <section label="A.68.3" status="3" xml:id="sect_A.68.3">
                <title>Surface Scan Mesh IOD Module Table</title>
                <table frame="box" label="A.68-1" rules="all" xml:id="table_A.68-1">
                    <caption>Surface Scan Mesh IOD Modules</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a235a55a-ab9a-470b-bb56-5f0780d4772d">IE</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f190b506-46e8-4c2b-b6da-fb0a6fac71fa">Module</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2e2f2259-a565-4044-980c-45702bf55663">Reference</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b272c8bc-660d-4561-b59a-16a0462cfa44">Usage</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="2">
                                <para xml:id="para_29d5cfd8-f64c-4183-8b1f-9f2ed578f8bd">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9efde7af-0374-4a4a-8ed8-c95698989141">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_984d93a7-4bdb-43db-93bd-c66e32a1fe7e">
                                    <xref linkend="sect_C.7.1.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0f02bb9c-0305-4f9e-bbea-e79fa94dc611">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_03dd774f-b7fc-4f3b-b241-4c89f066ae65">Clinical Trial Subject</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0432e6aa-b29a-4244-a59d-13c663910614">
                                    <xref linkend="sect_C.7.1.3" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4590cfe3-bd3a-4f97-85ed-f3ef2053e955">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="3">
                                <para xml:id="para_732e62aa-7a8a-496f-ab87-fb9cec0cab46">Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_86f7def9-6538-43d0-b959-9ad6f649824b">General Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c95f06c3-23c3-4389-a38f-669e3647a756">
                                    <xref linkend="sect_C.7.2.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_71d92394-b2d6-48c2-b8be-d2a9a2380340">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9b5837f1-9f5f-419e-abd0-4c5fb62148a2">Patient Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0d1c21f0-736b-41c8-9ae2-ea277d4a060e">
                                    <xref linkend="sect_C.7.2.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f666f758-b896-47c1-a474-eb249a893246">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_543ccb47-a9cd-4b84-a91c-645d644988c3">Clinical Trial Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b986530b-dad6-4b85-8c8b-3263316759cc">
                                    <xref linkend="sect_C.7.2.3" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0cb9ae25-b232-4eb1-8540-059f8714bf80">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="3">
                                <para xml:id="para_70b68db4-049d-417a-b5fa-7a1903d715d5">Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_325c41fc-3261-4fc0-9422-8b0837973b48">General Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ccaeed98-6362-4c56-a8eb-a4616bf220d6">
                                    <xref linkend="sect_C.7.3.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3922a353-5041-434a-9dac-833f05e1c8ba">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2b5efa10-7cf5-48b2-ba00-0e35ce6c93a0">Clinical Trial Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c2f4f2ea-56ca-4eaa-8fb5-4086b077df76">
                                    <xref linkend="sect_C.7.3.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e4fbca18-0f5e-4390-9215-b661ad92ecf2">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_915dc481-be44-4cd6-93ab-ad54ed466399">Optical Surface Scanner Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_551ff82d-29b6-4867-a3e6-e966515f1918">
                                    <xref linkend="sect_C.8.29.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5e9ae7c2-f9a7-4121-bb5d-d0a20d5a31ba">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Frame of Reference</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Frame of Reference</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline"> C.7.4.1</emphasis></emphasis><!--<xref linkend="sect_C.7.4.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="2">
                                <para xml:id="para_78eb3a48-80b1-4422-8f62-a362cb9f388d">Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9b7e5dae-4b1d-4057-9b8d-0c1012c57339">General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7a07e334-68f8-4f4c-9225-216a6fbb7e1c">
                                    <xref linkend="sect_C.7.5.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_32dd93f3-8027-479e-b28f-0b8ea743e24e">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_52404f02-d009-4a6c-bfa0-2b5832c1d7b7">Enhanced General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e4247d0e-8da3-4a12-baef-9c82c1a5d365">
                                    <xref linkend="sect_C.7.5.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d02ed319-1f4b-4310-bd84-04a44fb5c458">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="5">
                                <para xml:id="para_8f8cd86d-fb3f-4c2d-baa5-c3e92c48fbbd">Surface</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8e5308ab-c5b4-4e8e-9c7e-cf319b1db31a">Surface Mesh</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_485999d5-1189-4d4e-8bf5-c4ae06c5e8aa">
                                    <xref linkend="sect_C.27.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_91a2283d-d772-4d34-967a-a28b3e29f131">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0339d5cf-dd70-4907-82cd-c308973e5a7f">UV Mapping</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8cf604aa-225c-4dcc-a7c9-fbb55542fd48">
                                    <xref linkend="sect_C.27.6" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_55982588-49cf-4c45-ab44-0a3e2329356c">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_79419b1d-6d64-433c-a8e6-25693e571797">Scan Procedure</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_10c72317-f392-452a-bfb0-b3529e0df80f">
                                    <xref linkend="sect_C.8.29.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e67d317c-aaf3-45dd-92ff-01e18474de21">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ec7f52d0-412b-497d-b294-c683ebd01ff8">Specimen</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7dfcb000-fd7b-4daa-bf87-c91fb20e3c21">
                                    <xref linkend="sect_C.7.6.22" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3170eb4d-590b-451b-ad8a-882326025ddc">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f3a6a73a-d8f6-4b17-ae3d-2d09fc90d54d">SOP Common</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_02550d04-3a1a-41eb-a2c9-f7414a4d7461">
                                    <xref linkend="sect_C.12.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c211e07c-d015-4178-8587-1615f4f27e63">M</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        </section>
        <section label="A.69" status="2" xml:id="sect_A.69">
            <title>Surface Scan Point Cloud IOD</title>
            <section label="A.69.1" status="3" xml:id="sect_A.69.1">
                <title>Surface Scan Point Cloud IOD Description</title>
            </section>
            <section label="A.69.3" status="3" xml:id="sect_A.69.3">
                <title>Surface Scan Point Cloud IOD Module Table</title>
                <table frame="box" label="A.69-1" rules="all" xml:id="table_A.69-1">
                    <caption>Surface Scan Point Cloud IOD Modules</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5f50745f-40e7-4d77-8716-242d6a436927">IE</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_aaaceb71-3863-4627-bef6-ea79aa162de0">Module</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5fd3656e-e80c-4c22-ad07-7ba6f84f98bf">Reference</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b190509e-14be-47cd-a6e1-b3b214f38ec2">Usage</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="2">
                                <para xml:id="para_2b7eba58-b93a-45c0-b7aa-b1ba8422ce29">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d7dd8ce9-ec29-4929-a69a-717b7e13c756">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_262a0fd4-c6f4-4cf5-bcd6-3a052c2c8c76">
                                    <xref linkend="sect_C.7.1.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0a412e93-28cf-4d63-9658-46d6bd033a59">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1610105d-1795-4bd0-8c31-9f6f7c260c6c">Clinical Trial Subject</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_83bf546c-ad61-4e5e-a493-ccfb6e9e2516">
                                    <xref linkend="sect_C.7.1.3" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d135c167-0d79-4be4-b546-ff8e253b18ee">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="3">
                                <para xml:id="para_b476b4b7-add9-45f6-91da-7d9c7bbbca82">Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_212f823d-0312-47dc-8e86-80a26aa41880">General Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b21d523b-ee53-4af7-aac4-ff7698eacb61">
                                    <xref linkend="sect_C.7.2.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_11778690-af67-4f79-ad67-18003abbd120">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fbed6f88-817c-471d-8f64-cf4167661270">Patient Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_eeeff174-df3e-4cf0-b667-9e28e0fbfab4">
                                    <xref linkend="sect_C.7.2.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_95383141-ecd5-4140-a28c-59c7d74e22e3">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_74565b68-333b-4657-9518-24e3a1098327">Clinical Trial Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bc52ca6c-b4a7-4a43-acf3-9b4066bab455">
                                    <xref linkend="sect_C.7.2.3" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_94aed56f-95b1-463a-9a9a-d51a40249b23">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="3">
                                <para xml:id="para_10332f21-6c04-4777-9d0e-b4e4791ed150">Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_08ce7a24-bafb-49fe-835f-7e21f1bf5cf5">General Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_17d00900-3b86-4ac0-a0d7-d24425d77720">
                                    <xref linkend="sect_C.7.3.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_792502ac-3221-4b73-9aa2-82a209b0799b">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cf3f4064-16ec-495c-92ed-211b6051f61b">Clinical Trial Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c05fb3a0-3b81-44e5-84f6-4f449c97072b">
                                    <xref linkend="sect_C.7.3.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_df504022-1be8-4a59-b5dc-46f90c1ef6e4">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2a835176-579b-4235-828a-c6a167141213">Optical Surface Scanner Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ec79bbcf-65d0-430c-a230-376ee379bea4">
                                    <xref linkend="sect_C.8.29.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5038a86d-0197-4209-a541-813c9d3eca5b">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Frame of Reference</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Frame of Reference</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline"> C.7.4.1</emphasis></emphasis><!--<xref linkend="sect_C.7.4.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="2">
                                <para xml:id="para_396f6e9f-7b1a-42b7-8b4c-47e68ed93c3c">Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_36bc2722-4bb2-47a7-aaeb-e3fb2b30d85c">General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5bc210b7-f848-48c7-a363-2f04e8baf37c">
                                    <xref linkend="sect_C.7.5.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6609c3d8-bb20-4aa0-85b0-b4b732b0be39">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bcb8826b-dae0-4129-bfde-e14667161fd6">Enhanced General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b2749b50-a196-451e-8500-64b86f06e1e2">
                                    <xref linkend="sect_C.7.5.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a37317be-40b6-4eee-bebb-d9f03d3c2ab8">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="5">
                                <para xml:id="para_b75e6380-4987-4176-b5f2-104b5fa446db">Surface</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e6336681-6bb5-48f9-896a-ac07d2a1d8ba">Point Cloud</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2a7c13e6-9f3a-442b-ade2-06ec9422e2c2">
                                    <xref linkend="sect_C.27.5" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0cd281cc-823e-4bc0-b1ee-a05ef8fb5fbe">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_77641f07-8823-40ce-9a9a-750bd5bad850">UV Mapping</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3d686428-ca3e-426c-badf-85b89aaf8f62">
                                    <xref linkend="sect_C.27.6" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a7bae59e-f2af-4493-ad8b-4dba9787b186">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f343006a-6b6b-4653-9df9-4080f55c7a92">Scan Procedure</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cf9464fe-37cf-4521-a853-c23dc00f2358">
                                    <xref linkend="sect_C.8.29.2" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b92e6326-bab5-4827-81ff-a265377c2928">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_67ff2d86-4825-431c-a357-981f0f1cad44">Specimen</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e96e9437-6ec5-4b03-a73b-13b0fd94b674">
                                    <xref linkend="sect_C.7.6.22" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ef0840c9-3e17-4451-962d-0e59a4dde53c">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2aadb4fd-5f23-4fc1-a600-32f9ed6b5869">SOP Common</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_72205020-cc43-4412-9a0b-5ca4ee143e16">
                                    <xref linkend="sect_C.12.1" xrefstyle="select: labelnumber"/>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bb99ef2e-ac6f-41be-a974-44cfc5bbadb1">M</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        </section>

<section label="C.27.1" status="3" xml:id="sect_C.27.1">
                <title>Surface Mesh Module</title>
                <table frame="box" label="C.27-1" rules="all" xml:id="table_C.27-1">
                    <caption>Surface Mesh Module Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1526b339-e025-475b-b255-774b56d283e4">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_33d9577f-24b7-4997-a991-71c4c53b6055">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_572c97d9-ec66-42b7-b969-b532aa4e931c">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6788959a-b909-4707-a2e9-fd3ae25860cf">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_53af7e48-61dc-475d-882d-b09286a5b9ac">Number of Surfaces</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_230cd160-661c-445d-b3fd-88fbb89ba1ec">(0066,0001)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2d712256-4aaf-4648-be82-e1eb217b7ef6">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2ff18a2f-c44b-4e1d-80a9-c3741b9859d9">Number of surfaces contained in the Instance. Shall be 1 or more. Shall be the same as the number of Items in Surface Sequence (0066,0002).</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c211b583-a55f-4ed2-bf01-b8850495d071">Surface Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_73811982-359f-4d12-bb08-d23e36e45614">(0066,0002)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e3e4d98d-462c-485d-9d0d-02a7d3bb4b2a">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_083f87fb-c751-48aa-b356-6ef933575fbf">The surfaces that are described within the data.</para>
                                <para xml:id="para_08c30f15-d39d-4893-98eb-c4f5f5cf99ee">One or more Items shall be included in this Sequence.</para>
                                <para xml:id="para_7e48b12e-59b2-45a6-8fbf-e97d9e722751">There shall be Number of Surfaces (0066,0001) Items in the Sequence.</para>
                                <para xml:id="para_a860a648-7e19-4bf5-af7f-1ce1c1021cc8">See <xref linkend="sect_C.27.1.1.1" xrefstyle="select: label"/>.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9a6f7816-a1f9-4b69-9c22-5507f6593109">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_68b0be35-261f-4953-97e7-03054eb35ea6"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c10bab62-6f89-481d-ac98-cf3b93f47d18"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_81b7799c-19a3-4e56-9377-f4c9117bdc1b"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8b66ddef-8c72-4ce1-bc6c-81c0b1fac04a">&gt;Surface Points Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c32b99f6-e995-4fb9-96ab-6e900ce3922f">(0066,0011)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7bee88be-d73d-4e3e-8873-032b4c17e4f5">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9c7b8e47-9eff-4d83-ae85-9e374746c6e3">The point positions representing vertices of the surface.</para>
                                <para xml:id="para_ee9d891c-4c8f-47a4-a2cc-9c11a1e99c41">Only a single Item shall be included in this Sequence.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="3" rowspan="1">
                                <para xml:id="para_30954ae6-54bc-41b4-9ab7-25a0bc146dc0">
                                    <emphasis role="italic">&gt;&gt;Include <xref linkend="table_C.27-2" xrefstyle="select: label quotedtitle"/>
                                    </emphasis>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4fc1b38a-ca6f-4277-8c71-80baa6ee6809"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a71e5467-923e-460a-8428-a9a3d7f2b3a6">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c873e5bf-37cd-4fbb-8858-be4103d0d8dc"/>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_25ed7cc7-a842-406b-87fc-1f06d5c31d6f"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_46609dc0-3fd2-4f24-9848-339ca3ca60e2"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        <section label="C.27.2" status="3" xml:id="sect_C.27.2">
            <title>Points Macro</title>
            <table frame="box" label="C.27-2" rules="all" xml:id="table_C.27-2">
                <caption>Points Macro Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e5534c26-6176-40e1-b508-caffbb25c046">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_dadd4054-935e-4866-a5e3-85131478bfe2">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_609c25ac-0fe8-44ce-b3bb-fb7df9743d4a">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_607770d8-0069-415b-9cc6-343ee3ae8949">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_72a40fdd-ff7a-4e25-a68c-c23f3385a59a">Number Of Surface Points</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b0f36394-60ec-4a4e-a51c-66299e2900ed">(0066,0015)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2c31fbb6-50d5-42c5-bb2f-4018be51f257">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7ae984ff-2373-4c4c-ba77-c4e5b5b4692c">Specifies the number of points in the point set. See <xref linkend="sect_C.27.2.1.1" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_900eda10-60c4-40d4-a57b-8c9b083bc078">Point Coordinates Data</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cf7eb4ad-44d4-400e-83c9-9871406928ec">(0066,0016)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cd4d8b5d-fd82-417b-b7f1-741b54565055">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_89342794-e756-4a7a-b50c-7d27247dd736">See <xref linkend="sect_C.27.2.1.1" xrefstyle="select: label"/>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3aab4a2b-2c8d-48f0-8362-f0537ac36358">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_54558e5e-0d41-4283-9dc4-893363ef5d5d"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_71acb92a-0195-4a77-a8a6-68604f4edb85"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b8a5b115-d6ef-4c21-a09b-a2d0a9fdd9fc"/>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.27.2.1" status="4" xml:id="sect_C.27.2.1">
                <title>Points Macro Attribute Descriptions</title>
                <para xml:id="para_455c00e0-5fc2-4fd1-b168-e3c222d1ce23">All Attributes within this <emphasis role="bold"><emphasis role="strikethrough">Module</emphasis><emphasis role="underline">Macro</emphasis></emphasis> containing points or vectors are in x-y-z order. If multiple elements are encoded, the ordering is x<subscript>1</subscript>,y<subscript>1</subscript>,z<subscript>1</subscript>,…,x<subscript>n</subscript>,y<subscript>n</subscript>,z<subscript>n</subscript>.</para>
                <para xml:id="para_c00269be-081c-441f-b682-bd955139192f">The points are in the coordinate system identified by the Frame of Reference UID (0020,0052). To map these points into the coordinate system of another SOP Instance a Spatial Registration Instance can be used.</para>
            </section>
        </section>
        
        <section label="C.27.5" status="3" xml:id="sect_C.27.5">
            <title>Point Cloud Module</title>
            <table frame="box" label="C.27.5-1" rules="all" xml:id="table_C.27.5-1">
                <caption>Point Cloud Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_27a515db-040a-4d7b-afb3-1bf033f3eac8">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f85db903-76f3-44ff-8447-c00e02a24eee">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_48ee14f9-e559-48f4-854f-cf7106b0ae11">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_172549d7-a5f0-4841-a599-136cf2b08ca3">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7f1eb560-7584-4cc6-ae5f-bb8a64b67bb3">Surface Points Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8034c226-cc6a-4ea4-9206-805beea0c8a6">(0066,0011)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6552ca29-428a-4530-acda-204b1cf7ad3f">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_08845e1b-1822-4e40-ab1c-77294fe6bdf9">The point positions representing vertices of the surface.</para>
                            <para xml:id="para_39578417-f7b6-49b6-9d35-881dd501ae39">Only one Item shall be included in the Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_c4d62fe5-ce4e-4c01-9c76-2b4a9dcad45d">
                                <emphasis role="italic">&gt;Include <xref linkend="table_C.27-2" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_33766149-e6b1-438e-97cd-6a838c9bd793"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ab8aaf35-491b-4f09-9eaf-d08ca9bcda82">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3eff4eb6-d27a-41a5-afcc-c7b290eedc2e"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e003d574-d605-4906-9448-c466710df2a6"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f87cfcf2-657c-49fe-84e6-7e0f336e5631"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        
    </chapter>
</book>
