<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2025</title>
    <subtitle>Require Whole Slide OVERVIEW images to be single frame</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/03/09</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/02/17</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2025</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Require Whole Slide OVERVIEW images to be single frame</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>CP 1829 added the OVERVIEW image type, similar to the LABEL and LOCALIZER types, but failed to require that it be single frame as intended.</para>
                    <para>Also, it may not be obvious that a Dimension Organization Type of TILED_FULL is not applicable to these single frame images whose spatial extent may not match the Total Pixel Matrix.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
<section label="C.8.12.4" status="4" xml:id="sect_C.8.12.4">
                    <title>Whole Slide Microscopy Image Module</title>
                    <table frame="box" label="C.8.12.4-1" rules="all" xml:id="table_C.8.12.4-1">
                        <caption>Whole Slide Microscopy Image Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_19b0b4dc-4e92-4c95-9988-804684315cb7">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_96a09a1d-8879-43be-97e5-121291459369">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_3e3024bf-3c64-498a-ba60-ec15eea8abd4">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_18f4a093-c520-4639-ae1b-71059a30ef26">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7a877f73-4d9d-41b5-86d4-a8c3bbfedf52">Image Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_379f1037-05af-47e4-a7cc-263ccdb59552">(0008,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5e62e3f9-7cd2-48f1-898c-3cc78973e9f9">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8987cdf6-559b-4308-8bb6-886172153890">Image identification characteristics.</para>
                                    <para xml:id="para_6af97694-9c87-41bf-9a64-b1a1ab24272e">See <xref linkend="sect_C.8.12.4.1.1" xrefstyle="select: label"/> for specialization.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3bf068f2-90e2-4038-9795-a5e533a7457d">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6fd29a76-4932-4bf1-bd94-1bc9e5e03526"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_0f85d412-d926-40c7-a05a-cd74c9d8391c"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0c0bca29-c4c8-4fba-bb7a-8d244fda6667"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_305dffa2-cc70-4472-81da-73a738d8d1db">Number of Frames</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b35b2b79-4aeb-4806-88cc-eb002a89af1e">(0028,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_df7f6e9c-de29-4a8a-8fc0-c33eeb2a82d3">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_60b26676-3382-4932-b4fc-fdf092f8966d">Number of frames in a multi-frame image.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Image Type (0008,0008) Value 3 is LOCALIZER or LABEL<emphasis role="underline"> or OVERVIEW</emphasis>:</title>
                                        <varlistentry>
                                            <term>1</term>
                                            <listitem>
                                                <para xml:id="para_495140e7-94cc-4a17-aee0-9ae324754fea"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cbf7b471-b8a3-4750-9c1d-2ca54d2bb4f1">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e47d8169-473c-46a3-a20a-4e3ca54bc3aa"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8abcf907-22b6-485b-86de-d562dba4db44"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_c4d8ad7a-4df4-458c-86e7-6f1e897522bf"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.8.12.4.1" status="5" xml:id="sect_C.8.12.4.1">
                        <title>Whole Slide Microscopy Image Attribute Descriptions</title>
                        <section label="C.8.12.4.1.1" status="6" xml:id="sect_C.8.12.4.1.1">
                            <title>Image Type</title>
                            <para xml:id="para_570df186-e80e-495a-8c3f-a885b9326afc">Image Type (0008,0008) is specified to be Type 1 with the following constraints:</para>
                            <para xml:id="para_5d7c3916-3108-4847-b5d9-a9efd8c2719f">Value 1 shall have a value of ORIGINAL or DERIVED</para>
                            <para xml:id="para_64085fb2-c119-461d-b5ea-597a1b2f0c02">Value 2 shall have a value of PRIMARY</para>
                            <para xml:id="para_7a14bdbb-e4da-42ef-a9f0-42bf9c4a72d6">Value 3 (Image Flavor) shall have the Defined Terms in <xref linkend="table_C.8.12.4-2" xrefstyle="select: label"/>.</para>
                            <table frame="box" label="C.8.12.4-2" rules="all" xml:id="table_C.8.12.4-2">
                                <caption>Whole Slide Microscopy Image Flavors</caption>
                                <tbody>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_f4e0b9db-57d3-40c8-be18-c56a76e704e9">LOCALIZER</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9914348f-ddb2-4d12-9137-fbec1afa0557">Collected for the purpose of planning or navigating other images; provides frame-level navigational features.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_7490619d-a47e-420d-bbab-aa7af0caad42">VOLUME</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9b847f99-2c22-472f-b17d-8f60eb58e765">Set of frames that define a regularly sampled volume</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_b553e150-7ca4-4380-933d-3ff577682c88">LABEL</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_32434d33-8596-4260-86d4-c66a6bf6994e">Purpose of image is to capture the slide label only; non-label area shall not be present; the label is shown in <xref linkend="figure_C.8-16" xrefstyle="select: label"/>.</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_e0209857-046a-4c26-9cab-c409ce8c2cc8">OVERVIEW</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_e4393930-6e0d-41d4-b595-4b51fb20dbb6">Purpose of image is to provide an overview of the entire glass slide; shall include the label, the specimen and the empty space that are shown in <xref linkend="figure_C.8-16" xrefstyle="select: label"/>.</para>
                                            <note>
                                                <para xml:id="para_e3d5a499-8d49-4574-bca7-302449e0f320">One purpose of the OVERVIEW image is to allow visual quality control to assure that tissue fragments have not been omitted when selecting areas (manually or automatically) to digitize at high resolution. The OVERVIEW image is distinct from the lowest resolution VOLUME image(s) of WSI pyramids.</para>
                                            </note>
                                        </td>
                                    </tr>
                                </tbody>
                            </table>
                            <para xml:id="para_fb8f0316-22e1-4eb4-bc32-cba178b99ea1">Value 4 (Derived Pixel) shall have the Defined Terms specified in <xref linkend="table_C.8.12.4-3" xrefstyle="select: label"/>.</para>
                            <table frame="box" label="C.8.12.4-3" rules="all" xml:id="table_C.8.12.4-3">
                                <caption>Whole Slide Microscopy Image Derived Pixels</caption>
                                <tbody>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_21cbe34a-c110-42e8-8b67-a67c8ab89150">NONE</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_4b1732c6-26fc-4014-aa1a-10101b0ac589">No derivation of pixels (original)</para>
                                        </td>
                                    </tr>
                                    <tr valign="top">
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9b1cc671-a848-44d5-87a6-6562644deb2a">RESAMPLED</para>
                                        </td>
                                        <td align="left" colspan="1" rowspan="1">
                                            <para xml:id="para_9bb22eb0-ea22-48a0-8a00-0350e6287192">Pixels were derived by down sampling a higher resolution image</para>
                                        </td>
                                    </tr>
                                </tbody>
                            </table>
                            <para xml:id="para_3a8971ef-39d9-413a-83a7-81a386515947">No additional values shall be present.</para>
                        </section>
                    </section>
                </section>
        
        <section label="C.7.6.17.3" status="5" xml:id="sect_C.7.6.17.3">
            <title>Spatial Location and Optical Path of Tiled Images</title>
            <para xml:id="para_81c46a69-b4dc-428d-822c-2efeee711150">If Dimension Organization Type (0020,9311) is present with a value of TILED_FULL, then the Per-Frame Functional Group Macros that would otherwise describe the spatial location of each tile explicitly (e.g., the X, Y and Z offsets from the origin in the Slide Coordinate System Plane Position (Slide)), and the optical path, may be omitted.</para>
            <para xml:id="para_08fcce55-df5f-49f2-b8a2-978d0468872c">A value of TILED_FULL indicates that the frames across all Instances of a Concatentation, or a single Instance in the absence of a Concatenation, comprise a non-sparse non-overlapping representation of an entire rectangular region, and are sequentially encoded as successive frames in Pixel Data (7FE0,0010) in an implicit order varying:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_f0dd20ae-8ade-46f2-ad35-2c50add983d5">first along the row direction from left to right, where the row direction is defined in the Slide Coordinate System by the first three values of Image Orientation (Slide) (0048,0102),</para>
                </listitem>
                <listitem>
                    <para xml:id="para_bd60c273-e1a2-4c77-89f4-538768f79537">then along the column direction from top to bottom, where the column direction is defined in the Slide Coordinate System by the second three values of Image Orientation (Slide) (0048,0102),</para>
                </listitem>
                <listitem>
                    <para xml:id="para_0ae08229-6360-49ad-a748-9e855f00a350">then along the depth direction from the glass slide towards the coverslip, where the depth direction is defined in the Slide Coordinate System from zero to positive,</para>
                </listitem>
                <listitem>
                    <para xml:id="para_e553a9fe-2e1b-42ed-aa5f-a27f3c7c00e6">then along optical paths, where the direction is defined by successive Items of the Optical Path Sequence (0048,0105) in the order in which they are listed in that Sequence.</para>
                </listitem>
            </itemizedlist>
            <para xml:id="para_6d73468e-28c6-4aee-a708-dab0b3a14702">If Dimension Organization Type (0020,9311) is absent or has a value of TILED_SPARSE, then the spatial location of each tile is explicitly encoded using information in the Per-Frame Functional Group Sequence, and the recipient shall not make any assumption about the spatial position or optical path or order of the encoded frames but shall rely on the values of the relevant Per-Frame Functional Group Macro.</para>
            <note>
                <para><emphasis role="bold"><emphasis role="underline">Dimension Organization Type (0020,9311) is not applicable to images with an Image Type (0008,0008) Value 3 of LOCALIZER, LABEL and OVERVIEW, which are single frame and may have a spatial extent that is not the same as the Total Pixel Matrix.</emphasis></emphasis></para>
            </note>
        </section>
        
        
    </chapter>
</book>
