    <title>CP-2024</title>
    <subtitle>De-identify Scheduled Procedure Step ID</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>August 2020 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/07/24</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Wyatt Tellis</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/02/07</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2024</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: De-identify Scheduled Procedure Step ID</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, 3.15</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Scheduled Procedure Step ID is not listed as being required to be de-identified, yet all similar ID attributes are.</para>
                    <para>The only standard use in a composite IOD is within an optional sequence, Request Attributes Sequence, which is marked for removal or cleaning anyway.</para>
                    <para>However, it could be used elsewhere by an implementer creating a Standard Extended SOP Class, so it should be listed just in case.</para>
                    <para>The action suggested is to remove it, even though it is 1C, since the condition within Request Attributes Sequence is real world so, its absence in a de-deidentified object if Request Attributes Sequence was cleaned rather than removed would be harmless.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.15 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <table frame="box" label="E.1-1" rules="all" xml:id="table_E.1-1">
                    <caption>Application Level Confidentiality Profile Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c5e6e8ba-34e0-4e31-90b5-bd9ed7086ddd">
                                    <emphasis role="bold">Attribute Name</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d02d5580-8438-4a00-8c30-ba93530bcd31">
                                    <emphasis role="bold">Tag</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_705936dc-2c42-4198-b406-ba0e61b35e99">
                                    <emphasis role="bold">Retd. (from <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>)</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a8387cac-386f-4ad9-96b3-b3dcb1b1010c">
                                    <emphasis role="bold">In Std. Comp. IOD (from <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fc0ff9a1-5780-4d06-861a-24ae0ba38b47">
                                    <emphasis role="bold">Basic Prof.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_19ddb67a-a5d6-474d-a13f-8df27d28b707">
                                    <emphasis role="bold">Rtn. Safe Priv. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5b335431-5276-4cc0-bddf-93b6d107226e">
                                    <emphasis role="bold">Rtn. UIDs Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8317e0cf-0080-48f9-8040-1f49a4478c0f">
                                    <emphasis role="bold">Rtn. Dev. Id. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1ba661a2-a572-420a-95c1-bc1bf694a70e">
                                    <emphasis role="bold">Rtn. Inst. Id. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_796841da-6b27-4558-99c9-99503edccc6c">
                                    <emphasis role="bold">Rtn. Pat. Chars. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_edac8181-290b-4f9c-945d-5dfc7b1a43fc">
                                    <emphasis role="bold">Rtn. Long. Full Dates Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6305752d-56e7-4f4d-b9d8-e5052fcd6eea">
                                    <emphasis role="bold">Rtn. Long. Modif. Dates Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_84f3b128-8fe5-45bc-b04e-abc483377b06">
                                    <emphasis role="bold">Clean Desc. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0adc5d55-1196-4265-8ee1-b205103c96f1">
                                    <emphasis role="bold">Clean Struct. Cont. Opt.</emphasis>
                                </para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5e74e547-b6fd-410e-9c09-b271915c3eeb">
                                    <emphasis role="bold">Clean Graph. Opt.</emphasis>
                                </para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7656f7bc-287c-47b2-ad02-914b402b6b51">Accession Number</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7138d32a-2d94-4948-a80e-08f7a7ceca23">(0008,0050)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_44726b33-0b55-4c8c-9217-035cc42bdda6">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c92bf277-400a-4b75-9563-db0562225519">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6409f35b-fca0-40bb-9cb3-4697885ed819">Z</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7030df90-c342-4de7-94a5-dd1835a009cc">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c4127628-e3f8-4e9c-8f8d-fa886b2fc597">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cde84e69-3a73-416d-8fc8-5a7401d8ccba"></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ae3f824e-e3ea-4d9b-9373-3fd0148a6231"></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cedec952-f5e2-4d6f-bb95-0b5661517ff7"></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_27a20d18-a532-4534-a0a1-4ccb7169769a">Performed Procedure Step Description</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_908ae34f-eca2-40ac-9448-353f2486bceb">(0040,0254)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1d295ce7-f826-40fb-88c8-6dcf00abdc37">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_314a4505-9988-4388-b9cb-11c94cee8c3c">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7ece95c9-8540-4d27-8a30-e741b8a30090">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c3968b3a-8a52-4ef5-9f10-2af96f712d40">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6f443be9-70be-4d7f-87e7-efc08ba1545f">Performed Procedure Step End Date</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_991e0ab4-426b-49cb-95a7-969d6a66799e">(0040,0250)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_16d4fc42-f72e-4326-9775-b5f965652782">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a7d354d8-0918-429b-9c1a-b3dae69c3ff4">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_eb7f2d53-4ff3-4386-8ab9-a0ed46c75425">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ccfa1eb7-b23c-4f1c-9819-52767693cc9b">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cde30f24-b0c7-4964-a3a4-b6dc45ba19d6">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_914c160a-d4ae-4796-b326-4702dc12d891">Performed Procedure Step End DateTime</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_055c238d-2516-47b0-86b9-84b6fd564f68">(0040,4051)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7892819c-9ec0-44e2-a504-4738750bd805">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_68de4b9f-a517-406b-ac8d-e7379d7dec4e">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_44ef4766-0bcd-44e9-9315-2537a4677542">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a2225da5-0190-43af-8556-7e4b6800472b">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_29022661-0c3a-4d0d-a1a7-a0462e0b5485">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3b5d113b-82f7-4b95-aae8-c10dfd92cb8b">Performed Procedure Step End Time</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2efc516e-17c2-4bd0-984f-bebae6145239">(0040,0251)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fb7e4759-3e22-40e5-aff7-996f5c1da510">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_86fbc104-822e-41cc-99ee-cbfb91d9c316">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9003b36e-efe2-4086-b606-79d04b80c091">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2d5b3618-3b45-449e-a38d-f20ab0781ef4">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_55fc7900-cb26-4160-9f5f-a4017641c424">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ddcc2d2b-601c-4b4d-a310-956c3d341e4d">Performed Procedure Step ID</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_64aed264-4748-4d79-8e79-68d8f2c06638">(0040,0253)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a93df7c0-9a74-4439-ae6e-59a66da51f3f">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ee0e1b0a-da2d-4561-945a-4fde87062830">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d6a36ad9-164e-42d0-a03a-89128be6b3ed">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3aa1e1e8-007b-443e-bb84-ebe2add60134">Performed Procedure Step Start Date</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9fce19f3-3875-457d-af65-ac3448b65d27">(0040,0244)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_921c6415-9f0e-48f4-ac26-f901d3526a11">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3ab11a1f-8342-47e4-a306-f989524424fe">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ba6dd0b6-ac34-41b7-92a3-fec980d45962">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fe92fa97-926c-4b57-96a5-94e32b73a0e0">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3100d817-4bfd-4211-900c-a11a15288885">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8d26c1c8-7b43-4e7b-aae1-ccd0cfa3568a">Performed Procedure Step Start DateTime</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e8556d3a-f1aa-435f-9998-1f9ec0d9f1a0">(0040,4050)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3684cbc9-f7f9-4577-a421-eb0e6b016908">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1696300f-29ee-4e1c-bfe3-ccb00d1745f2">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_908151d6-44d4-4888-ae12-4ddb70a8b3ac">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_830e5187-0965-4080-96c1-bfb27e63f0d6">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_92ef4898-bf74-4f12-8719-7cd814281cb2">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a5e96434-5d98-4a95-b885-9e44a2732057">Performed Procedure Step Start Time</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_4be8ca7b-2981-402f-a831-26dcf3390f9d">(0040,0245)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c4f7372e-df05-4e17-984a-230553f788c4">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac0e489d-f985-48fb-ae08-a58ce7d174d1">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c51c9548-6161-4c39-bddd-ea30a0bcd6f1">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ff56834a-8b0c-494c-8595-8f2b45977e18">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_691f1679-276c-4704-bb7b-ac37125b188a">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_10af9e2f-9eda-4cc1-8133-223e334f030d">Performed Station AE Title</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1692f3bb-306e-4b27-a963-79b4663727a6">(0040,0241)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_62c2034d-7ddf-4b21-b011-cbb16d99c977">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_415b1592-9b12-433c-abcd-d9eccb76fdf8">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3f446a2a-24fa-4527-b3bf-3f2886eefd55">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_31d3dd71-ea88-482b-a319-769ea1934686">K</para>
                            </td>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d6315b92-c6b8-44eb-afcb-1aff48d24287">Performed Station Geographic Location Code Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6ea97dfc-aaef-495f-9d47-a6b07e31e317">(0040,4030)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_75464ecc-87b2-44eb-9c74-9c47e20b2ad5">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ef07bc2e-8dc0-43e4-8727-e411242a32b0">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2946ed41-7ea9-4df5-9b8e-67beee94cf94">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b0e0ed49-9525-4257-830b-d5fad6d44161">K</para>
                            </td>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b9010ae1-7e78-470d-9fba-a23db6e3934f">Performed Station Name</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac385a68-d655-453c-982c-b0ca8dc51963">(0040,0242)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0e676ba2-1f3a-4a5d-900d-0037083b31f4">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0cc61362-7626-4465-911c-56e4d33b7ec1">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e578fb27-05e4-4e8c-a138-99e0f49393ce">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8e51728e-0f9f-4ab2-87f6-78df3e8dedfb">K</para>
                            </td>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e3b213cd-47b5-40e6-ab91-d69b3ee339a1">Performed Station Name Code Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_16a1bf08-6c77-45f9-966b-d2d22b847c51">(0040,4028)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e8b27a05-a048-4cd8-afcc-bda1a9ea5a55">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bc32c45d-3f93-4b01-b8e4-b04b89d2272b">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b6508339-bc4b-4e0c-817f-899f972fa69b">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6d0bd4d1-f75e-4fd5-8dd6-860ac5cedf29">K</para>
                            </td>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para>...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para>...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_615d917f-f7d7-4eff-9d45-7cef74a9c4f0">Request Attributes Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0086eb24-2dbd-4a7a-8183-82b2324af681">(0040,0275)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0a0a0570-e359-440c-a749-880b61f9b874">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8260db7c-7260-45aa-a784-9281120094d0">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a72d6ef5-730b-40ce-8ff1-80c215f0952b">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_593ccd4b-a2d1-41bb-a7de-c95828ff9c2e">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para>...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para>...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f3a8e384-dd95-4d8f-bfa1-35d653664e20">Scheduled Procedure Step Description</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1edd2cf7-c70e-4c0d-86bb-a4c9cf5ccff5">(0040,0007)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_63ba7745-af8a-4822-8124-1f0a8b39077a">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9b7e65d4-0486-465a-8a3c-b4d1195fab32">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_77ae1255-172b-4f46-b630-b682fbc68dce">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cf9a5b53-5414-413e-91b9-89961f186500">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f38c8605-c3c4-4fdf-8816-0edc6b3e5091">Scheduled Procedure Step End Date</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_defc52d8-f50d-491e-8fc6-a12b1059f8c4">(0040,0004)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3728a926-9788-4c5c-b056-31cfc7a8ad6d">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_4884db6e-eb60-40ce-8093-510d48d824eb">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_59c0893e-d864-414d-adb5-fedd33e2e7a0">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_46aab857-effd-4045-806d-ceb71fbddb98">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_46a1c927-dc85-4a94-bbb3-82c52d85fea7">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_621d9ef6-4dfd-4d51-a5f7-334d725f9be6">Scheduled Procedure Step End Time</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_810c085a-6cb1-451c-ba56-19e74ad43cbd">(0040,0005)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_018d9369-bb8e-4038-b9ed-067a6ada802d">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d9a0b20b-e325-4728-993f-84a2ca0e17bb">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_54c03956-ebc9-47ee-905e-d162a21938a2">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7cf6694c-3fbf-4ee7-8141-9eadb67eea29">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6495d72d-e849-40b4-a4f8-be75e118c38a">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_977cb42e-a0d6-4647-b355-1102723ff596">Scheduled Procedure Step Expiration DateTime</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5a7bd997-842c-40f8-923f-db697ad76b67">(0040,4008)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bd48b9b6-4938-4623-bf6f-fe35a0ae3746">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_408c5416-0df0-492d-bc3f-eb7d888ae3f9">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_24922ce9-499b-40bd-bb22-7e91a2af15c9">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ddab7d29-bbb4-4445-8f5b-90910b594c81">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e41faa63-b25e-4605-aa15-d069b84924b2">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Scheduled Procedure Step ID</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">(0040,0009)</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">N</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Y</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">X</emphasis></emphasis></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_86d57a37-81f5-470e-bc00-0bfe4b8dd068">Scheduled Procedure Step Location</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fb39ae54-b05c-42bc-9fa1-fc57cee0a2b1">(0040,0011)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f2192ee8-fcdb-4fdf-8723-f49c0e5649a7">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3ba7b1be-edd7-4182-aee5-cb665d294be7">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_fa311cca-5e29-4c35-be1d-2553d3260595">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2d1b320d-dace-4be8-8368-c1c4a0a41db9">K</para>
                            </td>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_59e28d89-08d2-4453-8c80-eb5f600e487a">Scheduled Procedure Step Modification DateTime</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d440ae9b-24c6-4a81-a5e7-ed3af3e50b6a">(0040,4010)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d25125c3-ed91-4c94-b36b-569e1a73ce69">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2c68d786-d938-4c6d-a3c7-3c888f958967">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8001824a-3b29-45da-85b1-69d286ff8ddf">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_56a6d7c9-b044-4095-b5fe-c9d9d5e22ab9">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_05ef5caf-d0d7-445f-ab1e-e5a3458fd66b">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_22453877-9706-4cea-a45d-24855d295ebe">Scheduled Procedure Step Start Date</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6d060032-8d37-4507-9508-bb1760933c58">(0040,0002)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6735e044-c0a4-4626-bce6-a74a6c5214c6">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_49a90fac-3dca-45a1-aad0-9ec5dd34500e">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_74662d34-e4a1-4891-a49c-4e1e68296e99">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a9af2905-7fec-4c55-a68c-86f25ab64ed7">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a00327a4-e547-40c3-91b6-fa673ede7feb">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_51609877-72d3-4930-b08f-68345ef16630">Scheduled Procedure Step Start DateTime</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_046e3978-e868-4c26-9351-92b5ba8082e7">(0040,4005)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cfc01141-8a22-44a9-b6f3-0de6e0251281">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_31773cfe-d6ee-4d7f-8536-995f785f2f2b">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1bb562d3-9b3a-4ae3-ba16-d22b993d064b">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac52abe9-1790-436a-b9b4-77477521bdf1">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_af13798e-0676-4675-a22e-c361974b82ac">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9c0e5215-2559-4fcc-aae2-cc95d593210f">Scheduled Procedure Step Start Time</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8167971f-f1f6-4a34-91bc-2852c3d0c1f5">(0040,0003)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1513a891-c98f-4918-813b-816195c66145">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c6fb97a6-0491-4cd3-959c-f3969a239b13">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1efbf09a-e63f-440d-b47e-b995967416ce">X</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9b05a461-2075-482f-a6ed-9a51ed487f5c">K</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a6c23343-0f58-4fec-9590-0778211cbb9a">C</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cedf035a-b3bd-4b8a-b96f-35a2c30548fd">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3275f691-4d5b-4dc7-ad3b-a3c51e2d16a3">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_7ed492e8-dcc5-4678-ad11-eed3284774d5"></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6bc09a4b-b4b9-43e3-8a70-a5187c15c662"></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0dc91233-51e7-45e4-b95b-9dc77286ef86"></para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8e14b01e-10db-4fd5-9b3a-dcb0553efbb4">X-Ray Source ID</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c774e946-2a17-42ce-a1ec-fd0aa7d91de1">(0018,9367)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b78f2763-715b-4308-a781-6d037b636e7d">N</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_44e81e80-625d-4d25-9b9a-549dd1b5cbc3">Y</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ddb2c9f5-dc18-4227-8757-d1a3351e26a7">D</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_8a81c351-940b-423e-9f8d-99173481323e">K</para>
                            </td>
                            <td/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                            <td align="center" colspan="1" rowspan="1"/>
                        </tr>
                    </tbody>
                </table>
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged in PS3.3:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="10.6" status="2" xml:id="sect_10.6">
      <title>Request Attributes Macro</title>
      <table frame="box" label="10-9" rules="all" xml:id="table_10-9">
        <caption>Request Attributes Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5f751dce-aec4-4e47-a310-e87e889bd68b">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_77a7e01b-5873-4811-ab30-296955089494">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8ac3bdcd-e0d6-45b3-b11b-40e616d95dc2">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1ea4c264-8ad8-4005-953d-81f098b9eff2">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d7ccc21-0809-4551-861e-da750df178e0">...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_864d1034-2ffd-4414-83a5-f3d150b98704"/>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6c18929c-5d9d-47ff-925f-52af8ddbcc87"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1b951899-662c-49f0-beb7-fc6483a40231"/>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4683cb3e-8d2e-416a-8615-5aa45a0dcfce">Scheduled Procedure Step ID</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_cd679bb8-1ee3-4afe-9472-4cf9daf3b139">(0040,0009)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_d9faa9d7-d272-434b-b42e-54e32cb27187">1C</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c5edf4a-10fb-44cc-bce7-02662ddfe840">Identifier that identifies the Scheduled Procedure Step.</para>
              <para xml:id="para_1390d1c2-28f7-47e5-9a30-7ff9a25e0cd3">Required if procedure was scheduled.</para>
              <note>
                <para xml:id="para_36129a24-d644-48ab-914d-1f3fbe916df6">The condition is to allow the contents of this Macro to be present (e.g., to convey the reason for the procedure, such as whether a mammogram is for screening or diagnostic purposes) even when the procedure step was not formally scheduled and a value for this identifier is unknown, rather than making up a dummy value.</para>
              </note>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e585a02d-95f8-4292-b84e-adc8436b91ca">...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1c6a29c5-8032-42c5-be52-469bd1066b87"/>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_31b6e9a2-53c1-4f19-89be-1ec6058ab1cc"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad1809fc-ae3c-4321-851b-aea45f4fef34"/>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
        
        <section label="C.4.10" status="3" xml:id="sect_C.4.10">
                <title>Scheduled Procedure Step Module</title>
                <table frame="box" label="C.4-10" rules="all" xml:id="table_C.4-10">
                    <caption>Scheduled Procedure Step Module Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_bad07647-2114-4683-aa84-dc7864c0a7f5">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1e4ff9e6-bbd6-4704-b5ca-1db252dc8a24">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a801b5d9-dfac-4c65-8a19-1a71039c7839">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a2b4e090-eb24-425b-ba62-6e08c960d3a7">Scheduled Procedure Step Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_255ee85e-ec0e-48ce-8e13-4a46915f3e2a">(0040,0100)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_36b95f60-4092-4e88-867d-b2ca8295850f">One or more Scheduled Procedure Steps for one Requested Procedure.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7fbae099-d07f-40ce-82ed-7342fa4bd509">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_2669e173-ff41-432f-a2eb-8b2c8d03b149"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_aaa8c7c3-94ba-4027-bf2a-16bd2f9a59ed"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3c861c73-9c01-4c42-94af-57ad8ccde6e8">&gt;Scheduled Procedure Step ID</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a096b0d9-dbf6-40f1-9bca-bcfa155cdfdf">(0040,0009)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c4246fe5-831c-4015-9255-abb79e2ba082">Identifier that identifies the Scheduled Procedure Step.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5a116c0f-a8fd-43ae-8ce4-c37efee48e44">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_86ee4c17-03ee-45a4-91a8-8c93e71fa86d"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1329984c-fd1d-4542-838b-8f2fd688345a"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        <section label="C.4.13" status="3" xml:id="sect_C.4.13">
                <title>Performed Procedure Step Relationship</title>
                <table frame="box" label="C.4-13" rules="all" xml:id="table_C.4-13">
                    <caption>Performed Procedure Step Relationship Module Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d3135801-a071-4562-ac4c-e4735aacba47">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e5d0aafc-2037-4c56-a246-828570148351">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_a8b53f63-8cbd-4d30-94c1-82b8d787543b">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">...</td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5a6b8989-46a0-475f-a485-9eba2367ad2d"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_066a0a0d-8489-4695-8ebd-d3440f4554a5"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e2d8a7be-5701-49ee-860d-41ea883c339a">Scheduled Step Attributes Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3eeafdee-90bf-43fb-9e42-7d3a0a6dacc3">(0040,0270)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c428c48a-514c-4e17-b281-d43129bb33b6">Sequence containing Attributes that are related to the scheduling of the Procedure Step.</para>
                                <para xml:id="para_c7e5d307-26a3-4bc3-8878-f338324af61f">One or more Items shall be included in this Sequence.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_91452c01-140a-4833-b503-0470d8108a03">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_43f17670-ff39-4bf0-904d-353624869945"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_41f2cf8f-99fa-43ec-95c4-9d57d4097a18"/>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8fd43936-6ad9-4a64-9410-a25340ceccfe">&gt;Scheduled Procedure Step ID</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f54bde6b-1995-4c85-984d-bf668e9142e0">(0040,0009)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_97529fb1-c303-4d2e-b746-19a67af6ec5b">Identifier of the related Scheduled Procedure Step.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_36e2ac88-652a-4db4-b065-9f1c783e7309">...</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_dbe6635a-3f7f-4fdf-8762-2053c08a930a"/>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_eddb4a63-de0e-4345-aa2a-98a3891cd8b5"/>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
        
        
        
        <section label="C.7.3.1" status="4" xml:id="sect_C.7.3.1">
                    <title>General Series Module</title>
                    <table frame="box" label="C.7-5a" rules="all" xml:id="table_C.7-5a">
                        <caption>General Series Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_04100f31-5424-4105-b397-e81ea26cab1a">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9f6ee2bc-0c72-4acc-bd8a-3b63e075406c">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7be1655a-1548-4c66-a4a3-2f62a8898dc2">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_28b73666-093e-4904-b24e-ab760754d7a0">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cdb6f0e9-188b-4fda-bc33-00244b78db42">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_573048aa-6706-44a7-b7a9-56c1d9580215"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d64d0571-7bc4-4d9d-b327-23b378fde0d4"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_181afb38-9c62-4435-a027-a593e36049d8"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_6ca6b609-40a5-4309-a4ff-f5c15c292732">Request Attributes Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e17554f9-3b0e-4dfe-83a4-596bbd449e69">(0040,0275)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5a3d841d-b686-41ba-8bc9-b7eb9715fb35">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_493a6755-2cda-4480-807a-4a8e1987a12a">Sequence that contains Attributes from the Imaging Service Request.</para>
                                    <para xml:id="para_93ae0d07-e678-4c04-badd-05686d175771">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_7713cdc8-6fb3-4c5e-a773-9d70acfc1a83">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_10-9" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_1799100c-5076-4512-b80f-b9f0caa40a77">
                                        <emphasis role="italic">No Baseline CID is defined.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_50419f38-e31f-4b18-b847-970d294809c3">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_28845226-24b9-459a-b83a-e32ac7bdf673"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_21ce5082-0651-4f97-a865-81444d445192"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8c63303a-ef18-472f-abd7-42c4b0e63de2"/>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <section label="C.8.8.1" status="4" xml:id="sect_C.8.8.1">
                    <title>RT Series Module</title>
                    <table frame="box" label="C.8-37" rules="all" xml:id="table_C.8-37">
                        <caption>RT Series Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f90c1239-740c-4707-b148-b4d9bca27f8e">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_bb611acb-8e8e-48a6-90da-6d00e8957a94">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b4e8a2ec-7ce0-43c5-b41f-acb4bc940d90">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_335b90a9-ca11-4fe9-a27c-0d2692fca963">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d65d4a7c-3c8d-4f72-a2e1-1852479b32cd">...</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5af2fa1c-6c71-4f49-9d21-a95fa4449d15"/>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7e391522-89a6-415c-98a5-a98354dbc87e"/>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7e294953-9fc7-4b08-b476-263bbd26437e"/>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_23a2710a-3482-44f7-8776-25c539d5d711">Request Attributes Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_76e796ff-3b5d-4f1e-8967-0fd06b33b899">(0040,0275)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6c96e35b-7ff1-45ba-8a0a-bf7da6bd31ab">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_117d3809-6101-4926-8fd5-1d7f46ac5bcf">Sequence that contains Attributes from the Imaging Service Request.</para>
                                    <para xml:id="para_75e1220b-11f8-41d8-88e6-52785573252c">One or more Items are permitted in this Sequence.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_9132a4ff-74b4-494f-80e2-4173a50cba52">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_10-9" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d9a4a2a0-8801-40fa-95df-5843e5e8d9f2">
                                        <emphasis role="italic">No Baseline CID is defined.</emphasis>
                                    </para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
        
        <section label="C.8.11.6" status="4" xml:id="sect_C.8.11.6">
            <title>Mammography Series Module</title>
            <table frame="box" label="C.8-73" rules="all" xml:id="table_C.8-73">
                <caption>Mammography Series Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c49daaa4-5af2-4f80-bb35-0dc686eb5409">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f6ba8ce0-8602-4421-8f59-41199ba56d31">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_70fadfd9-5019-46ef-bad8-b5db77713b37">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d5e02b8a-aa16-43f8-9519-8448db6c9a0e">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c89d6540-3086-4591-9007-7dd0257aa62b">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5c82de8e-c99f-4f76-948d-b65ca7b466f9"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_04bcc0d8-e827-4d1f-b833-3ec89adaab4b"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8545214c-27f6-4da6-98d5-27eb0161b62b"/>
                            <para xml:id="para_c01aee01-5c63-4101-a4bc-3a9791cf8331">See <xref linkend="sect_C.7.3.1.1.1" xrefstyle="select: label"/> for further explanation.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8e095a40-e981-455d-a3b4-b6c5ff40809e">Request Attributes Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_73c04488-d035-415b-8ba7-e5514985793a">(0040,0275)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e33dec79-6775-4e71-827a-1761cca67238">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a9090577-a343-48c7-8afc-a54e1f3b6c9a">Sequence that contains Attributes from the Imaging Service Request.</para>
                            <para xml:id="para_d9b81d54-3847-416b-8dcb-178f4b0a6be7">One or more Items are permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_61786132-469f-4d00-b392-d6f331a0d719">
                                <emphasis role="italic">&gt;Include <xref linkend="table_10-9" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_51bd8334-75ba-467b-a3d6-1294b002bf61">
                                <emphasis role="italic">For Reason for Requested Procedure Code Sequence (0040,100A) the Baseline CIDs are <olink targetdoc="PS3.16" targetptr="sect_CID_6051" xrefstyle="select: labelnumber quotedtitle"/> and <olink targetdoc="PS3.16" targetptr="sect_CID_6055" xrefstyle="select: labelnumber quotedtitle"/>. No Baseline CIDs are defined for other Attributes.</emphasis>
                            </para>
                            <note>
                                <para xml:id="para_d4bbbd77-1abf-467b-aaba-f8deb1dabf3b">For example, Reason for Requested Procedure Code Sequence (0040,100A) may be used to convey whether the images in the Series were taken for screening or diagnostic purposes, if the information cannot be found in Reason For Performed Procedure Code Sequence (0040,1012) in the <xref linkend="sect_C.7.2.1" xrefstyle="select: title"/>.</para>
                            </note>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="C.8.11.10" status="4" xml:id="sect_C.8.11.10">
            <title>Enhanced Mammography Series Module</title>
            <table frame="box" label="C.8-76b" rules="all" xml:id="table_C.8-76b">
                <caption>Enhanced Mammography Series Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8b1288ab-e897-412e-ae11-4968d44ee4f6">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9b31ce1e-ded6-4555-b5e4-bfc8b34bd22d">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_039ba6d0-d22a-4b3e-a20e-e59dd33ef42c">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bdb07562-3bb5-4590-be4e-c7d67b2bff5d">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">...</td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7544e08f-8fb5-4743-b046-ca8716ba47f4"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6b700b73-c958-4ae9-8a4a-9b03ef3a17d2"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8975d44e-4c49-437e-915c-9d7bdac6746a"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f6e8ab3f-4a89-45bd-8124-6b7d48a3e5b3">Request Attributes Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fecea620-49ad-4fe9-a7f5-fdeb8e16fe6f">(0040,0275)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_71899d48-8459-4199-bc0b-9f647af7ae65">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0eb1a010-75f1-446c-b9a5-195cfde120d4">Sequence that contains Attributes from the Imaging Service Request.</para>
                            <para xml:id="para_4512ed51-f983-4ea9-af3b-02eaa2ea68b2">One or more Items are permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_a2223f7a-d284-4cb3-bba7-7d9eaa525cdd">
                                <emphasis role="italic">&gt;Include <xref linkend="table_10-9" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9644ec30-ee9b-4e73-be80-854eac9baf78">
                                <emphasis role="italic">For Reason for Requested Procedure Code Sequence (0040,100A) the Baseline CIDs are <olink targetdoc="PS3.16" targetptr="sect_CID_6051" xrefstyle="select: labelnumber quotedtitle"/> and <olink targetdoc="PS3.16" targetptr="sect_CID_6055" xrefstyle="select: labelnumber quotedtitle"/>. No Baseline CIDs are defined for other Attributes.</emphasis>
                            </para>
                            <note>
                                <para xml:id="para_c76071d6-f805-4ea7-9ec7-ea88c28ef35c">For example, Requested Procedure Code Sequence (0040,100A) may be used to convey whether the images in the Series were taken for screening or diagnostic purposes.</para>
                            </note>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="C.8.26.1" status="4" xml:id="sect_C.8.26.1">
            <title>Visual Field Static Perimetry Measurements Series Module</title>
            <table frame="box" label="C.8.26.1-1" rules="all" xml:id="table_C.8.26.1-1">
                <caption>Visual Field Static Perimetry Measurements Series Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2155b6d2-78de-4c77-b049-4728a1674d9e">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d4112817-11b3-46c6-9c2d-cbd492f1b0b3">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_88aa271c-f5a2-4400-9936-3a15cf7ad147">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8898cd93-19a7-4fc8-975b-f053c4149cf0">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1dc633ef-66c5-4d41-915b-64ab6e10dc01">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_58c39247-96ca-4f17-b4d3-24c9266bb55c"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0bf1f32-29c8-4163-94d4-1e1ed0a26294"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e458658d-faeb-4149-bb9b-1cf56ed9eda1">...</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a30ba666-af4a-4e50-8800-753c2e0cc2de">Request Attributes Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9be60d38-ed4e-4552-838e-17a09ffc0f24">(0040,0275)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_eb994d9e-94bb-4a55-97ba-e4a0b5c007cd">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0eeab8a7-496b-4d9d-b6fc-3c70c341fdc3">Sequence that contains Attributes from the Imaging Service Request.</para>
                            <para xml:id="para_62c6a7d4-0305-4aac-bc7a-4134ce2fdb64">One or more Items are permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_db89c889-1600-4497-8a0b-e6804ab595f4">
                                <emphasis role="italic">&gt;Include <xref linkend="table_10-9" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a5f7df7e-ab47-4728-b307-3984c25510e5">
                                <emphasis role="italic">In this Module Reason for Requested Procedure Code Sequence (0040,100A) has the Defined <olink targetdoc="PS3.16" targetptr="sect_CID_4256" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                            </para>
                            <note>
                                <para xml:id="para_64b9dcdf-b62c-412e-9817-186a4ec3813a">Reason for Requested Procedure Code Sequence (0040,100A) may be used to provide whether the test in the Series was taken for screening or diagnostic (quantitative) purposes.</para>
                            </note>
                            <para xml:id="para_8726cf63-90fa-481a-93c7-83b83cdb21ec">
                                <emphasis role="italic">In this Module Scheduled Protocol Code Sequence (0040,0008) has the Defined CIDs 4250 and 4251.</emphasis>
                            </para>
                            <note>
                                <para xml:id="para_5730db8e-b627-48b0-b566-e1e05bb71f8a">Scheduled Protocol Code Sequence (0040,0008) may be used to provide test pattern and test strategy.</para>
                            </note>
                            <para xml:id="para_42ec4f50-fdb7-4fd8-89bf-ec7fad8524e1">
                                <emphasis role="italic">No Baseline CIDs are defined for other Attributes.</emphasis>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="C.24.1" status="3" xml:id="sect_C.24.1">
            <title>Encapsulated Document Series Module</title>
            <table frame="box" label="C.24-1" rules="all" xml:id="table_C.24-1">
                <caption>Encapsulated Document Series Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_89c6b4aa-492c-4dc0-aabf-c6b934c6c4f1">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_55340e98-0c3c-4b69-9d29-86e5445911ed">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_28b3f499-8187-430a-a182-8818ec287a13">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ba09d1c7-d3df-4f09-bda2-48bad762c054">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c7df71fb-ca76-46d2-ae27-48f595654509">...</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1f9c1b01-6836-4dca-936f-cecc23a59922"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f98f1dbd-30aa-4254-b8a6-b50fcd6d13f3"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8e91732c-5381-4e0c-9db2-369b9d4d105f"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4f5188d3-7129-4b9d-a9a1-fdce841b61ff">Request Attributes Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f60aa845-0e2b-466f-84b9-d6ce376e62ee">(0040,0275)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_efdb5b76-c5a9-4d2b-8174-820712f117de">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2afc1c97-0376-4f8a-88cd-dbaa565c1475">Sequence that contains Attributes from the Imaging Service Request.</para>
                            <para xml:id="para_13de99be-eea4-49ad-82ca-958e9e17b1a7">One or more Items are permitted in this Sequence.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_0b78246b-201b-4192-8119-8183297ceccc">
                                <emphasis role="italic">&gt;Include <xref linkend="table_10-9" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_18bfe27e-2136-4392-80f6-c3ad90d849a7">
                                <emphasis role="italic">No Baseline CID is defined.</emphasis>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        
        
        
        

        
        
        
        
        
        
        

        
        
        
        
        
        
    </chapter>