<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-2022</title>
    <subtitle>Correct ratio units</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>August 2020 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/07/04</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2020/01/27</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-2022</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct ratio units</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Some occurrences of units for ratio in Templates and Context Groups are incorrect.</para>
                    <para><emphasis role="italic">[Ed.Note: Only selected examples of correct usage are provided for reference.]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="CID 7180" status="3" xml:id="sect_CID_7180">
         <title>Abstract Multi-dimensional Image Model Component Semantics</title>
         
         <table frame="box" label="CID 7180" rules="all" xml:id="table_CID_7180">
            <caption>Abstract Multi-dimensional Image Model Component Semantics</caption>
            <thead>
               <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_23e1cfdd-3bd0-48de-bce1-75f55fa3ab28">Coding Scheme Designator</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4ba8bc6e-507d-40e6-9b13-12ef14fd9194">Code Value</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_806f7e90-026c-4276-a9db-d429513c4040">Code Meaning</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_2308ba59-1ec9-46ad-8f07-911e936dc837">SNOMED-RT ID</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_194fa5d5-1a11-4bdb-8e01-c9748c94b11b">UMLS Concept Unique ID</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5377ff09-5498-45e0-9bd4-29fc6ceebdd9">Units</para>
            </th>
               </tr>
            </thead>
            <tbody>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_1be36c6c-7fb8-4b5f-a0d0-f3a2c746cca0">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_ba602f82-00fa-4874-991f-2684fd3e9aea">
                        <link linkend="DCM_113098" xrefstyle="select: nopage">113098</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6689ad87-de1f-42b6-8520-96d1bf928c3d">Magnetization Transfer Ratio</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_ce4cd494-56e6-4599-b9c7-1280e56bdb7e"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_b3907c0b-4074-490f-bf25-b6e2b2f41778"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_f387cdcb-ef79-4fc9-8b23-c9fda0d26cd2">DT ({ratio}, UCUM, "ratio")</para>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_3eaa788d-77d3-47be-b640-1ef748435252">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_6629b665-45dd-48ad-a438-ed76d2edcec2">
                        <link linkend="DCM_126396" xrefstyle="select: nopage">126396</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_f1744d8f-7334-45e4-8354-9383a4219e06">Magnetic Susceptibility</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_4c089cbb-f7ab-412e-aa66-ca88a15c6eca"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_ba06e82d-1289-466e-9dce-ff28b459098b"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_94d0230a-0a9f-4c2d-9cf9-47e1822133c8">DT ({ratio}, UCUM, "ratio")</para>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_ade024af-98e5-4a7d-aec2-f15c18b9f667">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_02005268-636f-4b16-86ff-5feef268f4a4">
                        <link linkend="DCM_130086" xrefstyle="select: nopage">130086</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_07539321-631f-46cb-a5af-b6f80aa55ada">Relative Linear Stopping Power</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_19ebfbcc-4ff9-418b-9a86-3a13b0b6429c"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_6566b03c-6346-4687-96af-8ac472d1fbd7"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6eacfd41-0236-451d-95bf-db590528daa3">DT (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                  </td>
               </tr>
                
            </tbody>
         </table>
      </section>
        
        <section label="TID 15301" status="2" xml:id="sect_TID_15301">
            <title>RT Segment Characteristics</title>
            <table frame="box" label="TID 15301" rules="all" xml:id="table_TID_15301">
               <caption>RT Segment Characteristics</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bd51e68a-eae1-44d7-9fd3-830a7bfb75b5"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a58e9aa7-8f3a-447b-8d08-cd975b6e394c">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_52e2a4aa-610f-4f80-9b35-a8e58e338639">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bbdf7ab1-eed7-4124-a5e5-141ab60c2a67">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d4445aca-51c3-4b37-bf18-f205e19b8383">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_55d4c344-97a3-44bb-b092-87bc0f33422a">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d33a6746-93fd-4142-b89d-59dde999bfb7">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_adb5dcbe-16ac-4b35-91e9-670e8cb0ffe4">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5393329-d6f1-40ea-a71c-70770a4e9de3">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bcb87fb7-395d-4373-b755-e11f7129b401">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d3742f71-edc5-43ef-a7bc-3959c386daf5">EV <link linkend="DCM_130082" xrefstyle="select: nopage">(130082, DCM, "Relative Mass Density")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_795ed627-322d-4d41-8e17-868af840aaa4">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f0c0c178-e3b9-4136-bfd5-dbda1ae9ff0d">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e5180af0-36e5-4ea9-bd27-05f825396a5c">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1402b870-ad4a-4cad-a539-db4def249e71">2</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a754cc9-0780-4321-8560-54f1c6e9987d">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_69b8cd7c-f65c-4900-83c8-ecfe9686a0c7">EV <link linkend="DCM_130083" xrefstyle="select: nopage">(130083, DCM, "Relative Electron Density")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_09570202-7a0a-48e7-8bd7-4304f160475c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1f52c809-6ae2-46a7-8d0f-aefeac177ce4">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f9f44d80-e4cf-4e9b-acfe-799d2740662f">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ba2895f-5edc-4772-aca7-3f972512cfab">3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dc1ae499-3cb0-4e8c-8be2-52c0883108be">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8bf2c960-8c46-48e2-afdd-969decf30148">EV <link linkend="DCM_130084" xrefstyle="select: nopage">(130084, DCM, "Effective Z")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_37aa1434-73d6-466e-885f-2b9d361adae9">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e1df8545-8502-42be-8cc5-cb6c54e4d81b">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a050ee87-1ac6-442b-a2a3-2aff3b5f93fb">Units = EV (1, UCUM, "no units")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_24241881-c721-4107-bc01-aa48e4ef2417">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d72738b5-bf77-412f-94fb-37efd90b3ce6">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a6d12dec-d5a9-47e7-8d80-c6d95fdf15bd">EV <link linkend="DCM_130086" xrefstyle="select: nopage">(130086, DCM, "Relative Linear Stopping Power")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dbd818bb-e0b9-4f10-b442-fd5a1f00c38b">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fe377536-f7d5-4ae3-9fc2-a5abab97ed9d">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4970991-7c32-423a-a2ae-5307a8a46c73">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a4a9e6e6-8b32-4f1c-b096-d251e249c003">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_583871c1-ea43-4287-8ec5-32a52293e477">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_18851616-bd8d-4e4c-86f5-a03a39a55787">EV <link linkend="DCM_130088" xrefstyle="select: nopage">(130088, DCM, "Linear Cell Kill Factor")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_87051988-275a-4490-a8bd-88f771dc28e6">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d78b0ed2-c15a-4cc7-9ced-3003af26028b">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7c6b00af-0044-4630-96e5-aa2b0eb5f902">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_574817db-90d9-49ee-9f3c-f179181365d3">8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0265abba-394f-4913-9b22-cff4efb049df">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b463df4c-1eaa-420f-a03f-a36c199c3e61">EV <link linkend="DCM_130089" xrefstyle="select: nopage">(130089, DCM, "Quadratic Cell Kill Factor")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_24074c19-a2a7-425e-b6c4-5fd375c9f3da">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a55b9f82-2c37-42aa-aa1c-b1efdfb72114">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b6525107-88d4-4fa9-b655-6930fb857741">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6294436c-4e06-4b76-9b30-697f66afb753">9</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b63adc29-8c15-4cdd-87f4-39ff89d21dbe">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2ab221e3-8e22-4fcf-944a-c698e238e2f2">EV <link linkend="DCM_130090" xrefstyle="select: nopage">(130090, DCM, "High Dose Fraction Linear Cell Kill Factor")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e2bd86cf-5c50-43d2-b9c6-25cf4f06e380">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_11e81468-b5a3-4473-8528-c8eafc05664c">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_98a58037-bf21-4645-a597-c65b83eb6931">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0c0987cf-8100-4b44-84cb-489c821515dd">12</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_46ddbb62-d617-4b7d-a132-396c18aa51f8">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2dc41b0c-7a37-4159-a9b0-f7dd82208460">EV <link linkend="DCM_130093" xrefstyle="select: nopage">(130093, DCM, "Atomic Number")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_03ef8f6a-34b7-48de-aded-d67b94532a2f">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_69597a52-eb7a-4c35-85e5-851176151786">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eb427b95-906a-4122-8f6d-9bd116779f56">Units = EV (1, UCUM, "no units")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c89693b0-9e8c-4b6d-b5fe-28a0f0547aab">13</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6cb43979-05ca-45ec-a427-69f5d85fa64d">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4cbd2ceb-04d3-465d-b736-2c6f5136a237">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_73601e3d-00c5-4222-a302-e8b96f032c9d">EV <link linkend="DCM_130094" xrefstyle="select: nopage">(130094, DCM, "Elemental Composition Atomic Mass Fraction")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7e7f71f7-270d-4ac0-8289-d1c8d90de503">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8cf45426-603b-45e6-82a8-3041ce847187">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_287ff738-0030-4334-8578-25f24bcfbb25">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2f864b33-abf0-41c3-bfb8-fe7f0adc600c">14</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_08ce5679-f4df-41d8-ba80-48c7b341753c">NUMERIC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_76b093f6-ffc4-49b4-b14f-d6617d244b7b">EV <link linkend="DCM_130095" xrefstyle="select: nopage">(130095, DCM, "alpha gEUD value")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4b3a001e-cb31-40ae-9409-a926c5502eaa">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_30d937e3-b931-46ce-9295-06bc19efe41f">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6fb66f31-0be4-4ce1-9671-8f5e62125973">Units = EV (<emphasis role="bold"><emphasis role="underline">{</emphasis></emphasis>ratio<emphasis role="bold"><emphasis role="underline">}</emphasis></emphasis>, UCUM, "ratio")</para>
                     </td>
                  </tr>
               </tbody>
            </table>
            <informaltable frame="box" rules="all">
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c50351f0-653f-4da8-a1fa-eb6c99897ccf">Rows 12, 13</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fca3099a-496f-419a-ae94-9fadf7afaa95">The value of <link linkend="DCM_130094" xrefstyle="select: nopage">(130094, DCM, "Elemental Composition Atomic Mass Fraction")</link> annotates the fractional weight of the elements identified by the <link linkend="DCM_130093" xrefstyle="select: nopage">(130093, DCM, "Atomic Number")</link> with respect to the total mass of the segment. The allowed value is in the range of [0, 1].</para>
                     </td>
                  </tr>
               </tbody>
            </informaltable>
         </section>
        
        
        
        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference unchanged in PS3.16 :</emphasis></para>
                </td>
            </tr>
        </informaltable>
        <section label="7.1" status="2" xml:id="sect_7.1">
            <title>Context Group Table Field Definition</title>
            <para xml:id="para_32b8ebee-184f-4386-a49d-ad2aa2b8c1e9">...</para>
            <para xml:id="para_be54812e-092a-4644-bc36-f3a1bc3c5707">An optional column may provide a normative baseline or defined set of units to use for numeric measurements using the concept,
                either as a single term (e.g., DT ({ratio}, UCUM, "ratio")), a list of such terms,
                or a reference to a Context Group (e.g., DCID 7277 "Units of Diffusion Rate Area Over Time").</para>
            <para>...</para>
        </section>
        
        <section label="7.2.2" status="3" xml:id="sect_7.2.2">
            <title>Units of Measurement</title>
            <para xml:id="para_a92ec74c-b0fb-4537-9e0d-0213bce91064">...</para>
            <para xml:id="para_319481f7-5126-45e6-974e-d307a8937bcb">The specialization of the UCUM standard as it is used in DICOM involves the following rules:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_ce59b92f-30e4-482d-81e8-470067f36249">...</para>
                </listitem>
                <listitem>
                    <para xml:id="para_5db73b35-62b5-46b4-9436-acb29489cab4">In the case of UCUM unity ("1", or curly braces expression) it is forbidden to use "1" as a Code Meaning. <xref linkend="chapter_G" xrefstyle="template:Annex %n"/> provides Code Meanings for a Code Value (0008,0100) of 1. A Template or Context Group may constrain the Code Meaning according to the following rules:</para>
                    <itemizedlist>
                        <listitem>
                            <para xml:id="para_6dfe7eff-908a-4aef-ac63-1500f97e001e">UCUM default unit 1 shall use one of the Code Meaning synonyms specified in <xref linkend="chapter_G" xrefstyle="template:Annex %n"/>
                            </para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_20e5493f-af13-42e3-ba3c-c21980801165">ratios of identically dimensioned values may use ({ratio}, UCUM, "ratio")</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_945ea154-8a14-4e68-a7f4-3df63a9f0e3d">unitless numeric scores may use ({M:N}, UCUM, "range: M:N") to specify the minimum and maximum value, for example, ({0:10}, UCUM, "range: 0:10")</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_bd032db3-228d-48d0-8570-f8b933073429">counts using UCUM annotation shall always use the text within the curly braces as the Code Meaning, for example, ({masses}, UCUM, "masses")</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_a5bf59e2-0125-4fb2-aa34-c3ce7113a3b9">compositions of a curly braces expression with other UCUM values may use a conventional clinical representation, for example, ({H.B.}/min, UCUM, "BPM")</para>
                        </listitem>
                    </itemizedlist>
                </listitem>
            </itemizedlist>
            <para xml:id="para_e16c21d5-586d-489f-9c74-f9fdfb25097e">...</para>
        </section>
        
        <section label="TID 3206" status="3" xml:id="sect_TID_3206">
            <title>VA Main Results</title>
            <table frame="box" label="TID 3206" rules="all" xml:id="table_TID_3206">
               <caption>VA Main Results</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ec82f963-bf96-428e-9bfd-43f69d1ddb26"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cc4436ae-e4cd-46bd-84a4-22b68f3b54fa">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_56cde304-4c5d-452b-9087-71e57829870c">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b1d963bb-bcbb-4e02-8fd3-80ccc215106d">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b8289a3-3d48-40d2-ad66-11f8d365d64f">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_307c5d1d-6d82-44dc-ae2a-1f6db0e16ab8">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1400ca5c-f44a-43ef-b67b-c0bdc3546199">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_faaf919b-6cc3-4211-baa4-aa7230706459">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_40f7b07a-2991-4da9-88e5-4885c82f3817">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9622854e-40b5-4a7b-8791-2bae5831a5a2">4</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_15a4314e-956f-4553-a3a9-364b2ee75749">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cdd9c79a-fbf7-4f1d-bd16-b492a9afa829">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_69507a0e-118b-438c-84d7-80beb330a308">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5eb219c-c83c-43d2-b602-7688d84ed379">EV <link linkend="DCM_122435" xrefstyle="select: nopage">(122435, DCM, "Regression Volume Exponent")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2bd7183f-271e-4a5c-bd9e-d82f52030269">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f6aa8bc8-f214-4bb7-9f38-04f03a7ce707">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_21f4d5fb-60d4-4b52-87f0-c750409eb33b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_934c8508-3e12-442f-978e-7ce23d6560ec">Unit = DT (1, UCUM, "no units")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d69d0860-a5c1-4a5b-a7fb-af01989ef77f">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_475440a8-6e6f-4eda-9154-e7a23d05c78b">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d8d142e7-0a1c-4a25-8acf-5d9720266d72">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0205038c-29e0-45ce-ae80-58473df1277c">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0d07d315-e99e-4485-99d7-44e997de020c">EV <link linkend="DCM_122431" xrefstyle="select: nopage">(122431, DCM, "Regression Slope ED")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_850430d7-fecf-4587-8aff-b7620150a96c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ac811624-6f38-4e50-8747-18db9aa88d84">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_48b1b221-d742-4348-92d9-582a6997455f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_929923d2-70c1-4df0-b174-0a60f4aff819">Unit = DT ({ratio}, UCUM, "ratio")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_93d2b516-e157-4e08-9843-e045a6dfb6d9">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6b40f397-8387-4f1b-93f0-f3f87b0fc9d9">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a3172de9-2990-4293-a372-fcd18b78960f">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f101c7f4-4420-489a-b378-c90dbdefbbfe">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e1b040a2-4ea4-4ee9-854b-c233a8f3a88d">EV <link linkend="DCM_122433" xrefstyle="select: nopage">(122433, DCM, "Regression Slope ES")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8a0d9d7c-4faf-4c6a-a791-d078a2274fb6">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b0cfbfb6-0940-46bf-9803-985d1bf97643">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1e540e94-f097-4fce-8dc2-3137b64af6f2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3e4c0ab9-3f97-4229-a1e1-bef1d38cd7f6">Unit = DT ({ratio}, UCUM, "ratio")</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
        
        <section label="CID 4107" status="3" xml:id="sect_CID_4107">
            <title>Tracer Kinetic Model Parameters</title>
            <table frame="box" label="CID 4107" rules="all" xml:id="table_CID_4107">
                <caption>Tracer Kinetic Model Parameters</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d199c90d-9b2f-4f27-b501-9997dbcadae4">Coding Scheme Designator</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_78542d71-df00-4b97-8d62-5e7ad3f23698">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7983e12d-1647-400e-ad29-ecd9b5989120">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_45a85dbe-bff6-4dca-b21e-17464b13c63b">Units</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_138b64b2-612b-4ef2-b712-2c91d7ed5fc9">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d95350c4-4620-4daf-8fd2-603e195a815d">
                                <link linkend="DCM_126314" xrefstyle="select: nopage">126314</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0fb0dfc9-da58-45df-8fcd-fc301a5cdfdf">ve</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b4755561-01c7-48d8-a0e8-b95bc5a26b6e">DT ({ratio}, UCUM, "ratio")</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3ca3fb25-d5fe-449a-8a33-2e08dce6849b">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_794870dd-fd97-445b-9bd9-4dab9c97c585">
                                <link linkend="DCM_126331" xrefstyle="select: nopage">126331</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9e5e4dee-9071-4a9a-95e7-f5547cf8be2b">vp</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6c4880b2-083c-47ff-afe1-3645cc2c1041">DT ({ratio}, UCUM, "ratio")</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="CID 7181" status="3" xml:id="sect_CID_7181">
         <title>Abstract Multi-dimensional Image Model Component Units</title>
         <table frame="box" label="CID 7181" rules="all" xml:id="table_CID_7181">
            <caption>Abstract Multi-dimensional Image Model Component Units</caption>
            <thead>
               <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_36984dd0-c4bb-4d60-9672-aeae10eeea63">Coding Scheme Designator</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4285321f-d1a2-4b02-b336-f75e9cf6cb9a">Code Value</para>
            </th>
                  <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_2d913b98-479c-4cbe-ad48-0f22e9153823">Code Meaning</para>
            </th>
               </tr>
            </thead>
            <tbody>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_722970fc-e047-4581-97da-b2abb064f1c9">UCUM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_386c7284-2a92-4796-ac94-c63d939ce008">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_b409b845-34d8-416e-99d4-80fce0cd1c9a">no units</para>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_ea8829d0-dbfe-4b53-93c3-4c175e9a9c02">UCUM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_e991ea98-0156-4774-a1dc-b7f9b6e884b4">{ratio}</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_da1a29ac-3a63-45e7-bf92-3753693f5e81">ratio</para>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_3a5ffef4-16ed-441a-a9d5-b706e7a74558">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_c347a968-557a-4d46-a477-d0d52f440abf"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a19faf0e-01cc-4239-9d76-55ebf368e807"/>
                  </td>
               </tr>
            </tbody>
         </table>
      </section>
        
        <section label="G" status="1" xml:id="chapter_G">
      <title>English Code Meanings of Selected Codes (Normative)</title>
      <table frame="box" label="G-1" rules="all" xml:id="table_G-1">
         <caption>English Code Meanings of Selected Codes</caption>
         <thead>
            <tr valign="top">
               <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_37309cee-1fc3-4a9c-94ea-8ef8cdbaa5bb">Coding Scheme Designator</para>
            </th>
               <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_149479f8-a549-4f0a-a8ed-d301cfbe996b">Code Value</para>
            </th>
               <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a650cc21-6a0a-4bce-8340-4c616facbb36">Code Meaning</para>
            </th>
            </tr>
         </thead>
         <tbody>
            <tr valign="top">
               <td align="center" colspan="1" rowspan="2">
                  <para xml:id="para_9dd7ad98-7a74-4744-b3b4-c0302dab220f">UCUM</para>
               </td>
               <td align="center" colspan="1" rowspan="2">
                  <para xml:id="para_10e1d94e-95d0-4e43-8b42-80fdca575788">1</para>
               </td>
               <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3d65c3bd-915e-4a2c-9b69-b87cadc277e4">unary</para>
               </td>
            </tr>
            <tr valign="top">
               <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_14ea7894-2856-462f-9a62-0ebee7805dff">no units</para>
               </td>
            </tr>
            <tr valign="top">
               <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2238e51d-474a-4ec9-8601-50a1f041bfe5">UCUM</para>
               </td>
               <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6a99a3a3-ab3e-418a-bdc6-fb3332b5264d">{ratio}</para>
               </td>
               <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e1d9de95-6c54-43b3-95a3-9566d84c77db">ratio</para>
               </td>
            </tr>
            <tr valign="top">
               <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7da0ecc8-a23f-4eca-874a-215216c20167">...</para>
               </td>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="center" colspan="1" rowspan="1"/>
            </tr>
         </tbody>
      </table>
   </section>
        
    </chapter>
</book>
