    <title>CP-1885</title>
    <subtitle>Correct conflict in number of values between General Anatomy Optional and Required Macros causing conflict in DX introduced by CP 1744</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2019/02/23</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2019/01/27</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1885</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct conflict in number of values between General Anatomy Optional and Required Macros causing conflict in DX introduced by CP 1744</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>CP 1744 added the General Anatomy Optional Macro to the General Image Module
                    (Anatomic Region Sequence is Type 3 so if present must have a value) without
                    realizing that this would cause a conflict in DX family IODs that use the DX Anatomy Module that invokes
                    the General Anatomy Required Macro (Anatomic Region Sequence is Type 2 and may be empty).</para>
                    <para>Resolve the conflict by instatinating the Attributes of the Macro inline and using a type and condition on Anatomic Region Sequence that allows an empty value when using in an IOD that invokes the General Anatomy Required Macro.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        <section label="10.5" status="2" xml:id="sect_10.5">
      <title>General Anatomy Macros</title>
      <table frame="box" label="10-5" rules="all" xml:id="table_10-5">
        <caption>General Anatomy Mandatory Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_28acfe33-f5ea-4aa2-afec-b8efd6f83b3f">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_78e1da86-8dcf-4369-8e97-162b08fa542e">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f6b72f2f-672d-45fd-ab11-d39e77375b96">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a3bda21f-9999-4fab-8861-f039b5e17178">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3978d2a-6089-48c9-b67a-3198f37a4a45">Anatomic Region Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_233438ca-da73-47e7-8d4f-20d4894a5254">(0008,2218)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1b555d3d-f9c9-49a9-861d-0f19f78e6324">1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d8af205-401a-4cb1-979f-6cab0d17946a">Sequence that identifies the anatomic region of interest in this Instance (i.e., external anatomy, surface anatomy, or general region of the body).</para>
              <para xml:id="para_d7b57ebe-a0cd-461b-94fe-f44877c0ad2b">Only a single Item shall be included in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_d527ca56-4ebd-4ade-aafc-ae066953e55c">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a629542-70bf-4ef9-ab7c-aa8c06e3c102">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95245b0c-c835-4aaa-a652-effe2344ea1d">&gt;Anatomic Region Modifier Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_da7f877b-4b3a-462f-9e51-f99c081745b2">(0008,2220)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_ba7f2313-8a29-42b1-b424-aca7f8ef81bc">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1060f4e7-6687-4682-a27a-264d1afbb592">Sequence of Items that modifies the anatomic region of interest of this Instance.</para>
              <para xml:id="para_e6e2125e-010f-455c-85d5-263d3219d586">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_37c840bc-0954-464c-a793-6e334d78121f">
                <emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5eecdaef-8559-48d9-aaa4-a1755850366d">
<emphasis role="italic">DCID 2<!--<olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>-->, unless otherwise defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_e1cc678d-d7fb-4933-b842-a9ba4d2437aa">
                <emphasis role="italic">Include <xref linkend="table_10-8" xrefstyle="select: label quotedtitle"/>
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59ee8593-4ca4-4d91-89c5-5f9a2f774a04">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
        </tbody>
      </table>
      <table frame="box" label="10-6" rules="all" xml:id="table_10-6">
        <caption>General Anatomy Required Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0d170364-f4ac-4ca8-b472-b3252913f804">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ce3160aa-7ecd-4d0c-8297-cc1c947d3798">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_72f46f2a-9276-4dc3-9ee1-e3807d2908dd">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ec7d151e-f803-49ba-b1c9-8384135bcf2d">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e416165-236e-4d21-bed9-3e2925a564a9">Anatomic Region Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_7e93223e-629c-469f-b346-26530fcc2737">(0008,2218)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_af11ff3f-afac-4a51-b129-8d1706d72a5a">2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aef062d7-462f-4ea6-9efe-d6b161b16273">Sequence that identifies the anatomic region of interest in this Instance (i.e., external anatomy, surface anatomy, or general region of the body).</para>
              <para xml:id="para_4fc5ddf0-c3b6-4a81-a874-f76189e911b6">Zero or one Item shall be included in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_19b4b677-b208-4c05-9630-00e16bdf9f87">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d7e14df-4933-4426-8d62-3976f4eb2066">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9a00f064-5405-496b-9f97-aa200f8e5e91">&gt;Anatomic Region Modifier Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_f015e59b-9fc0-46d9-a915-eb753300ea4e">(0008,2220)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_162b14fd-a4cb-4c80-8793-135f041285b8">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fb3d74a-2234-4e8d-9056-6ee301f454e9">Sequence of Items that modifies the anatomic region of interest of this Instance</para>
              <para xml:id="para_81a23d0e-5e62-4d4a-9682-304600c10c5b">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_2297000c-7269-47e5-8c9e-db0795f399e9">
                <emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c24af139-a2dd-4f8b-94bb-56531aafd7b3">
<emphasis role="italic">DCID 2<!--<olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>-->, unless otherwise defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_93a89af7-a54e-4495-a830-b9a61f46822e">
                <emphasis role="italic">Include <xref linkend="table_10-8" xrefstyle="select: label quotedtitle"/>
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b10c3f5e-852a-4fed-acc8-fa9f3865711f">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
        </tbody>
      </table>
      <table frame="box" label="10-7" rules="all" xml:id="table_10-7">
        <caption>General Anatomy Optional Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_6193ccef-2e89-45ce-ab8f-76b31dcd7002">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c7a9e509-99f1-4558-b1d0-4f9a4e242f67">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_282b2777-0033-4af3-be31-8ae9c579cc34">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_66cdf02f-68e6-45df-8786-662e68dcdefc">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bcacde1c-235c-45e9-8dc8-ef2e24efd493">Anatomic Region Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_40df9dab-8bc8-42a9-8353-9650d740ed92">(0008,2218)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1acd0337-be6a-4853-b136-db305df14a85">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47377de6-60aa-42b9-8694-5c94bc39a671">Sequence that identifies the anatomic region of interest in this Instance (i.e., external anatomy, surface anatomy, or general region of the body).</para>
              <para xml:id="para_28d437ed-a1e3-456d-9c54-65fd06202952">Only a single Item is permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_e309f960-ebcf-406d-ad60-1df7e21370b3">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c333e8b-bd61-47f8-b9ee-3f7e9033fad4">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4761952-636e-4653-8f10-aadbb9624568">&gt;Anatomic Region Modifier Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_ff080979-52a3-48a0-8e0c-bfde404aabc5">(0008,2220)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_47ff12bc-f016-42c0-a78a-cd084a7e94b5">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29ee33e9-1d79-467f-8418-5e3ba1903d41">Sequence of Items that modifies the anatomic region of interest of this Instance</para>
              <para xml:id="para_fc9dbdd5-5a60-48fc-ae5a-f4a55882ba6c">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_d5ed1915-e8da-438f-ad70-0d0c45ef8d36">
                <emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b07cffaf-f43d-4829-ad3a-0855545e8d23">
<emphasis role="italic">DCID 2<!--<olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>-->, unless otherwise defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_381f8374-2a87-4f87-a326-9c7c3a207f9e">
                <emphasis role="italic">Include <xref linkend="table_10-8" xrefstyle="select: label quotedtitle"/>
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21452d3c-3844-4521-bc0e-687c945ed648">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
        </tbody>
      </table>
      <table frame="box" label="10-7b" rules="all" xml:id="table_10-7b">
        <caption>Multiple Site General Anatomy Optional Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3c3047e0-cc6f-42de-99cf-d671d4a25715">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_817b6d28-1357-4fbf-8930-7ee92cb6970d">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c61c4b68-7106-4323-89b7-2afe7063dc1d">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_31aab592-8a09-4b60-a0c4-78c8c184168a">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc5998d3-c023-401e-84e6-f73b69a9702e">Anatomic Region Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_f23c2d4a-0ec1-4675-a2ff-e11f4112ad9e">(0008,2218)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_adfd34aa-3041-4afa-bbce-4430f4b6b204">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64a62dfa-4734-4b5d-b0f1-1afb0d116422">Sequence that identifies the anatomic region of interest in this Instance (i.e., external anatomy, surface anatomy, or general region of the body).</para>
              <para xml:id="para_a6580a55-9ec4-43a5-a6be-931797093b88">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_c49bd235-2b8b-4815-bfef-57daf0f13a41">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_585c7260-d7de-46ac-91e3-cd4237109703">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a79bc06f-85b2-4a4b-83b1-8cdd9f35ffa2">&gt;Anatomic Region Modifier Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_59f3c31c-c720-4e37-82dd-37196c952458">(0008,2220)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_43c2a975-8409-45e5-be3e-cbbab6033ff3">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_139955a6-a2cc-4000-ab27-fa6a41477489">Sequence of Items that modifies the anatomic region of interest in this Instance</para>
              <para xml:id="para_5914c33a-ab3b-4467-8245-4b42feb4043e">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_9f5facba-a9f3-40b7-b6f0-3dd3acedf8be">
                <emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e98566ed-5531-4bbc-85d4-50c2fa361c9a">
<emphasis role="italic">DCID 2<!--<olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>-->, unless otherwise defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_0a7cd5e0-738f-47f4-8de5-c7250baa4483">
                <emphasis role="italic">Include <xref linkend="table_10-8" xrefstyle="select: label quotedtitle"/>
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e66eefb-ca70-41cc-bc16-e01ae6b5eaaf">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
        </tbody>
      </table>
      <table frame="box" label="10-8" rules="all" xml:id="table_10-8">
        <caption>Primary Anatomic Structure Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_188962e4-09ab-416c-ad41-bd4b926a1808">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_01d3aeb1-2aba-40e4-94bf-3abbf6374d3b">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5350aa94-ac62-4152-9d3c-d55b95265b75">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_725e3311-fdda-4ae6-b401-cfabdccb1726">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e32ffbcf-5658-45e5-b1e1-4ad2a782695c">Primary Anatomic Structure Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6a6be467-e6f0-483e-a236-5f98375fe3e5">(0008,2228)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_80b4886c-aee3-4452-84e0-4f4eb8d59a86">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc7003cc-8039-4fde-b4bd-3c3d4ed02f9c">Sequence of Items that identifies the primary anatomic structure(s) of interest in this Instance.</para>
              <para xml:id="para_9e678547-6183-4a79-a3d5-5c9566e34e95">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_76726d45-5476-4add-8558-9b2e9fd7d512">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cdb338bd-9a36-45c6-a1fb-e246b321b08d">
<emphasis role="italic">CID may be defined in the Macro invocation.</emphasis>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d9289b5-91cd-4892-92c6-889a9792caba">&gt;Primary Anatomic Structure Modifier Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_740ceb97-e57f-466c-bccb-76dfd8bbe4e5">(0008,2230)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c6da1105-4259-4bed-afc5-38612dbf241c">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29da9abf-f3de-4030-8fb2-ec3b93c4c67c">Sequence of Items that modifies the primary anatomic structure of interest in this Instance.</para>
              <para xml:id="para_9713a8de-250f-4496-83ba-eff55bce7ca7">One or more Items are permitted in this Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_50e6ad5e-e8bd-4abe-bd29-eb0ae8abde61">
                <emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a5541adf-4ae2-4f05-9c98-e583627e3deb">
                <emphasis role="italic">DCID 2<!--<olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>-->.</emphasis>
              </para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
        
        <section label="C.7.6.1" status="4" xml:id="sect_C.7.6.1">
          <title>General Image Module</title>
          <table frame="box" label="C.7-9" rules="all" xml:id="table_C.7-9">
            <caption>General Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8401777e-d9c1-4d22-8011-9a1bf0fbcea4">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_afde5dfc-2829-42bf-8491-b3bd6c747d8e">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6cd2c26f-14d7-4ff6-9db5-9a9f34361a8a">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_452e431a-5c40-49fc-be7c-dfb46dcd6833">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_de917bfa-cab7-45b3-8c3c-c05a25abba1e">
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="strikethrough">Include <xref linkend="table_10-7" xrefstyle="select: label quotedtitle"/>
                                        </emphasis></emphasis></emphasis>
                            </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1c6b3fe-3dc1-44b8-aab6-1c7489b03bb1">
                    <emphasis role="italic"><emphasis role="bold"><emphasis role="strikethrough">Baseline CID for the Anatomic Region Sequence is CID 4031<!--<olink targetdoc="PS3.16" targetptr="sect_CID_4031" xrefstyle="select: labelnumber quotedtitle"/>--> for humans and CID 7483<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7483" xrefstyle="select: labelnumber quotedtitle"/>--> for animals.</emphasis></emphasis></emphasis>
                            </para>
                </td>
              </tr>
                
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Anatomic Region Sequence</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">(0008,2218)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">2C</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Sequence that identifies the anatomic region of interest in this Instance (i.e., external anatomy, surface anatomy, or general region of the body).</emphasis></emphasis></para>
                        <para><emphasis role="bold"><emphasis role="underline">Zero or one Item shall be included in this Sequence.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">Required if the IOD invokes the General Anatomy Required Macro in the top level Data Set. May be present otherwise.</emphasis></emphasis></para>
                    </td>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                        <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                            </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">Baseline CID for the Anatomic Region Sequence is CID 4031<!--<olink targetdoc="PS3.16" targetptr="sect_CID_4031" xrefstyle="select: labelnumber quotedtitle"/>--> for humans and CID 7483<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7483" xrefstyle="select: labelnumber quotedtitle"/>--> for animals.</emphasis></emphasis></emphasis>
                        </para>
                    </td>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">&gt;Anatomic Region Modifier Sequence</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">(0008,2220)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Sequence of Items that modifies the anatomic region of interest of this Instance.</emphasis></emphasis></para>
                        <para><emphasis role="bold"><emphasis role="underline">One or more Items are permitted in this Sequence.</emphasis></emphasis></para>
                    </td>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                        <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis></emphasis></emphasis>
                            </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                            <emphasis role="italic">DCID 2<!--<olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>-->, unless otherwise defined in the Macro invocation.</emphasis>
                        </para>
                    </td>
                </tr>
                <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                        <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">Include <xref linkend="table_10-8" xrefstyle="select: label quotedtitle"/>
                                        </emphasis></emphasis></emphasis>
                            </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para>
                                <emphasis role="italic"><emphasis role="bold"><emphasis role="underline">No Baseline CID is defined.</emphasis></emphasis></emphasis>
                            </para>
                    </td>
                </tr>
            </tbody>
          </table>
        </section>
          
        <section label="C.8.11.2" status="4" xml:id="sect_C.8.11.2">
            <title>DX Anatomy Imaged Module</title>
            <table frame="box" label="C.8-69" rules="all" xml:id="table_C.8-69">
                <caption>DX Anatomy Imaged Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3d0518db-109b-42a5-b0e0-e9f80ef14160">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_897a745c-8d9b-4add-8ed6-be9276503b9e">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1fa0134c-c789-49aa-99a9-0e15e1b617c8">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fb4882bc-0981-4543-b5b3-fd14371c263c">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="3" rowspan="1">
                            <para xml:id="para_256621e0-ae20-4741-bb04-2af5ea7d62e8">
                                <emphasis role="italic">Include <xref linkend="table_10-6" xrefstyle="select: label quotedtitle"/>
                                </emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_18fa5774-acd2-4586-ac90-6fea0729ed15">
                                <emphasis role="italic">Defined CID for the Anatomic Region Sequence is CID 4009<!--<olink targetdoc="PS3.16" targetptr="sect_CID_4009" xrefstyle="select: labelnumber quotedtitle"/>--> for humans, and CID 7482<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7482" xrefstyle="select: labelnumber quotedtitle"/>--> for animals. This is the anatomic region that is placed on the table or bucky for examination.</emphasis>
                            </para>
                            <note>
                                <para xml:id="para_890e3b94-c82d-4f5f-bdc4-08d2c2dad2f0">It is strongly recommended that Anatomic Region Sequence (0008,2218) be present with a value, in order to ensure that images may be positioned correctly relative to one another for display.</para>
                            </note>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.8.11.2.1" status="5" xml:id="sect_C.8.11.2.1">
                <title>DX Anatomy Imaged Attribute Descriptions</title>
                <para xml:id="para_1e185d66-2278-4025-8799-e620f5870481">The Attributes in this Module extend the function of Body Part Examined (0018,0015) as used in other IODs, and are intended to be used to facilitate the management of images and Series in terms of routing, storage and display, as well as to dictate certain Conditions on Attributes and Modules in the DX IOD.</para>
            </section>
        </section>
         
    </chapter>
</book>