    <title>CP-1876</title>
    <subtitle>TID 1501 should pass ImagePurpose and WavePurpose parameters to TID 300</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>June 2019 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2019/02/23</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>QIICR</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/11/13</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1876</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: TID 1501 should pass ImagePurpose and WavePurpose parameters to TID 300</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>TID 1501 lists as parameters ImagePurpose and WavePurpose but does not pass them to TID 300 as intended.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
       <section label="TID 1501" status="3" xml:id="sect_TID_1501">
             <title>Measurement and Qualitative Evaluation Group</title>
            <table frame="box" label="TID 1501" rules="all" xml:id="table_TID_1501_Parameters">
               <caption>Parameters</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3aed916a-5ecd-4156-864c-342338106c58">Parameter Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_10711228-3f87-414f-adee-cb3b75652b81">Parameter Usage</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_811e1292-2dcb-4c96-9321-abd8782d58d7">$ImagePurpose</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_afa350c7-6c39-4686-9cfd-c9e3bfa3d1e1">Purpose of Reference for an image used as a source of the measurement</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_24a9b682-2345-40fd-8c2b-d3eee882e0e1">$WavePurpose</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_bf8d79a5-6232-4639-bd0b-c77e8f4b85a4">Purpose of Reference for a waveform used as a source of the measurement</para>
                     </td>
                  </tr>
               </tbody>
            </table>
            <table frame="box" label="TID 1501" rules="all" xml:id="table_TID_1501">
                <caption>Measurement and Qualitative Evaluation Group</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6cf3eeee-bde3-41f2-9cd2-d2a7c2fd15de"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a753304a-333e-44a2-b1de-6a357d827b35">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9b361704-3515-44f8-bc8d-8b4c7bbd3bd6">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3f88f263-0608-4fda-9eb2-629f8f00ab90">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_472206d9-066c-4c71-9bbd-2114d729f75b">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8720acfb-54df-451c-b098-dc55f146f6bd">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_789f5c1b-14da-47e9-8b2d-91afa2c9f6ce">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e618c06c-2ed8-4cc8-86d8-570049b0290b">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_50c97dbc-38b3-4780-8666-e80d0e960f39">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9d5a36cf-23f5-4c7d-a5d7-7ee4322f1b03">10</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e58ce1bd-d0ac-448d-9b93-01f903241340">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6177297b-681a-4e6b-b8c7-2881cd7e0ba7">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_262cd8dd-5c62-46dc-9228-61b78178bd26">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4f382802-3d6d-4d33-a37d-5208d882708c">D<xref linkend="sect_TID_300" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8e58ba2c-c810-4780-b9f5-3f7391d17037">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_680211e9-9f16-46b4-830e-f947d05bb711">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5faad4f-a65b-4bee-8c06-4b271b8f4559"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f39a1c62-ce64-4943-a3b6-165b144f07e7">$Measurement = $Measurement</para>
                        <para xml:id="para_793d2241-8a62-4744-babe-23d6f78f652a">$Units = $Units</para>
                        <para xml:id="para_92e6f6ab-8253-4199-b880-698e0b3764e8">$ModType = $ModType</para>
                        <para xml:id="para_51b7a3f0-cef1-41be-ac3c-7328c2a23cf5">$ModValue = $ModValue</para>
                        <para xml:id="para_0195d2d8-4704-46aa-847f-888bf183efef">$Method = $Method</para>
                        <para xml:id="para_a598a4cf-16b8-41e9-8da5-e45fee215ba8">$Derivation = $Derivation</para>
                        <para xml:id="para_a520ab94-54af-48bc-94ad-1c0c012b35a8">$TargetSite = $TargetSite</para>
                        <para xml:id="para_9136f165-0728-41a9-846b-3c467174949e">$TargetSiteMod = $TargetSiteMod</para>
                         <para><emphasis role="bold"><emphasis role="underline">$ImagePurpose = $ImagePurpose</emphasis></emphasis></para>
                         <para><emphasis role="bold"><emphasis role="underline">$WavePurpose = $WavePurpose</emphasis></emphasis></para>
                         <para xml:id="para_618a925c-b9aa-4e68-9be9-69035614020c">$Equation = $Equation</para>
                        <para xml:id="para_d16f9c46-d989-48e8-8655-1fc0d42caced">$RefAuthority = $RefAuthority</para>
                        <para xml:id="para_874803a4-4c0f-45a0-b22b-6ee636108a9d">$RangeAuthority = $RangeAuthority</para>
                        <para xml:id="para_5b19ea79-97de-4f5e-87c4-8a04ddcc3740">$DerivationParameter = $DerivationParameter</para>
                        <para xml:id="para_9b9c9301-94db-4c83-af7d-b1da9193fccf">$DerivationParameterUnits = $DerivationParameterUnits</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section> 
 
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">DICOM PS3.16 for reference unchanged:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="TID 300" status="3" xml:id="sect_TID_300">
            <title>Measurement</title>
            <table frame="box" label="TID 300" rules="all" xml:id="table_TID_300_Parameters">
               <caption>Parameters</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e78884e8-2fba-44cb-9543-7b13a3f3d74f">Parameter Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_87aed517-66ec-469b-848a-051450840c1d">Parameter Usage</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_311f48b5-37c5-440e-8687-ff61aac2c7e2">$ImagePurpose</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0291ccff-9df5-4b97-a117-5f245e2d5d4d">Purpose of Reference for an image used as a source of the measurement</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_9972894f-aa6a-4168-83c4-0a734f412a88">$WavePurpose</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_82836268-0d2c-4de6-823f-d567f7fc929e">Purpose of Reference for a waveform used as a source of the measurement</para>
                       </td>
                   </tr>
               </tbody>
            </table>
            <table frame="box" label="TID 300" rules="all" xml:id="table_TID_300">
               <caption>Measurement</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_72040574-5d06-4448-92d3-65d64029aec2"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6dc1a5bb-799d-4ce7-b116-2de6f00399b0">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ada3f541-03e1-4c32-aca3-f7f3d0de93e9">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_34201b00-0939-4acb-975d-63b09c385b07">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_104a2ec7-1391-48f6-b7dc-c5f94a8bb57a">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0699ebc9-be03-4846-a7b1-e8c504fd0ae9">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_40849f85-f55e-4dd9-bc80-7af5c0e2a872">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cd9bf78e-2ee9-47a9-8a55-5cf98072a759">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9856b597-aeef-4056-a98c-25b47f0aabcf">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8e932889-0764-40b5-9f5e-ac3845d6f933">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4b175a93-7c5c-48bc-a5e5-380db6283009"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6ec6fe93-5cb0-4758-8adc-deef5678010b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5fe8320f-931e-4eec-91a8-f8d7ae6d7f00">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_48165cbf-5ff1-4353-9c7b-c05569a0066b">$Measurement</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a6d1fd64-a1d1-4245-8dc5-88b325d3094c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5ab6a36e-54ef-41db-a7c7-de35a9843624">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5043dfda-6fa9-4044-b601-cb8254bff80b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b0e7cc06-ed22-40d7-ae42-3e8831a76b76">UNITS = $Units</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c515a4c2-545c-43c4-8e41-ade857100b6a">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c94a32ec-4213-4104-9bab-c3f6db880f93"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_689822ca-4490-4ddc-ab4b-fa40eaaf2ee0"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7fdc1a0e-c8a0-46d7-96d9-6c3d8ee2531c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1b0eebe5-a3b6-4489-8519-995ab0fb2387"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_72a99565-a981-43c6-b839-2ea70017869b"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9f7c2de4-d8f4-4da7-b290-b85c8d6374a9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f35b74d0-e5b2-43a6-aa54-fe99d4fb9621"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e9202f3c-77aa-4877-8002-f52a780921c3"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_770d16b5-21b5-4887-96bd-6aeaf6afd2c8">13</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10c8009f-cdf8-481e-a705-02fa7087fc41">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b3918619-21e2-4ed1-b694-45f56b4953aa"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_161efed5-07f7-4bc4-9c2f-baffaed7e8f1">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bf78b3b7-4ed9-47bf-88a9-18efa573270c">D<xref linkend="sect_TID_320" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e1a45d45-0533-4a14-8fba-e85f2a151d8f">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_72e4002f-993f-410a-99d7-d699a37f8340">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eaaa3c9a-40ef-4fa0-82e2-39369442ba24"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_22d6c3ff-aaca-45d7-b0a6-d0109f903293">$Purpose = $ImagePurpose</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_0f1f348a-5796-4a87-9ba3-f2889749259b">14</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_451e74d0-aac0-48d2-99fe-d88c8f19c95e">&gt;</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_efe5adcc-e9c9-4812-badf-d6d44e61aa4b"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_c40f07c9-ca79-4b2f-8065-7325242d588b">INCLUDE</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_e79da970-e41b-4528-b581-450ab8af39d1">D<xref linkend="sect_TID_321" xrefstyle="select: labelnumber quotedtitle nopage"/>
                           </para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_c274ef43-d146-4065-9802-fbe6c4f48ca2">1-n</para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_3e6d8f0a-39ac-4c87-972c-e242109f5fe1">U</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_8ea7f370-a961-4816-867e-e6d3d41763d2"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_594b4882-e970-47db-9edb-c64c94f3421c">$Purpose = $WavePurpose</para>
                       </td>
                   </tr>
               </tbody>
            </table>
         </section>
                 
        <section label="TID 320" status="3" xml:id="sect_TID_320">
            <title>Image or Spatial Coordinates</title>
            <table frame="box" label="TID 320" rules="all" xml:id="table_TID_320_Parameters">
                <caption>Parameters</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ebd96938-c3b9-4a1f-bbfc-d6e2140302b5">Parameter Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_89248ed1-fd4f-41c5-acae-258078e91ccc">Parameter Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_147b3698-eb79-4c4b-8091-57a5e86c3a50">$Purpose</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fbcddfc1-e820-4df5-991a-1f9270ee8482">Purpose of Reference for an image used as a source of the measurement</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <table frame="box" label="TID 320" rules="all" xml:id="table_TID_320">
                <caption>Image or Spatial Coordinates</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e5f8bc8f-75e5-4644-88e4-d925e2427406"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_333bd5e2-7629-4c4e-8037-3629257e3f2e">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_8f0beaed-eea4-481c-97da-68077db0157e">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_23fb0cc2-5f00-4082-adb0-57061734c895">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_89a0aa3e-f8b7-4ee5-9ddc-2351e9934939">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0dd1ad58-9adc-45d9-90e5-51d90b0aa683">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_756a86ea-2029-4ade-94ce-ff5c3d0fd533">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f39253c8-c566-4914-a066-96fbba0740a3">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_30c31544-a1e3-4254-8a17-2ef2b1fbbdd7">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_caf417ca-507f-4096-9d79-46df005553ff">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_86376ae4-4426-4900-8fd9-b726051dc125"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_63be4664-2031-443e-a119-84603e9d7aba">INFERRED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8ef550a4-77b3-4c8f-902e-afa764d03179">IMAGE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1eca1136-5edc-4a15-a64a-190637f5ec0c">$Purpose</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ef1ab5cb-27b8-4b45-a527-4b2893bc4c6b">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4f86962b-c32a-4be7-b103-327b495de1e4">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4ddab8be-9d01-42e9-9dd1-09394ea1079e">XOR Rows 2, 3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b2dd3f6c-4280-41f3-a386-65063f83351e"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3c44ea90-1048-444d-aa82-c74e1b481099">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_26a28c7d-a02b-4085-9f99-fa24a514ea67"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4e837ba1-3f6f-4aa1-855e-97fcfed4867f">R-INFERRED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1463cc45-fd45-4e9d-ad38-3f46fcf5d6a4">IMAGE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7ea17b4f-cdde-4b81-b79d-39af735de2a6"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d15871e7-06e0-49b6-848c-b216c80fded6">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ec65b768-3373-414d-b5a8-fecb38956134">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2edb8274-04da-4083-9e6e-1428f981d3c0">XOR Rows 1, 3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8b70a03e-1daa-4c62-a28e-432e2f6a65cd"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_1ed28e47-c8db-491f-a314-44175372a0d3">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0664dea2-011d-4824-8030-26dcc6596bf2"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_75cea5a6-f615-4a68-90fb-ee0209ca10e9">INFERRED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4e621a2a-9387-423f-8aca-638a0b75932a">SCOORD</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c7138e34-b551-41ab-bc33-39d38e2947be">$Purpose</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bc4b8124-3eb7-49ea-9fba-17a9d5cc5353">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6ee9fa0a-d9ba-4797-906d-20d63e15a305">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c724e5e2-ad3c-4530-b978-ed19b07eef7e">XOR Rows 1, 2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7b057dd2-b9d2-43ff-839f-e8d0651c27fe"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2a78fd17-12e5-4947-92a8-5cb96644e39b">4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_256ec7f8-d499-4428-96da-d81d8ad92208">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2032108f-4677-4719-9386-f196e010fd11">SELECTED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3f547296-0472-4e1b-8c4e-45a92b35645c">IMAGE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_04bd368c-1689-4d64-8f6b-e62e041c27d7"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f5c6efa4-6319-4435-8466-bddda077f19a">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_74b90508-968e-43b5-be17-b965e856342c">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_aeb0ea69-01eb-48ae-a7e1-922b71b704ae">XOR Row 5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7eac4ea7-f3ac-4b03-9004-a336241011ef"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bf6ebe1b-356a-4df1-a064-5d562238bc2f">5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e84ddfe6-14b4-4e06-b13f-acf34fe65c8e">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b9973c22-8f4a-4db0-8895-61d1ee409963">R-SELECTED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4d7adb4a-0c19-4c64-a114-b722207bebca">IMAGE</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_75a56d6b-fb2c-40ce-9517-1f9f6d92b639"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ff3ff7a4-cf80-486c-9d56-d47767d9fa89">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_801a4820-fdd0-445a-b0c3-775a5467993f">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_76afe243-3c68-4f0b-b8a8-6547a0654090">XOR Row 4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4717b023-e2aa-40d4-9c67-dfc3bef1d5ce"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
         
        <section label="TID 321" status="3" xml:id="sect_TID_321">
            <title>Waveform or Temporal Coordinates</title>
            <table frame="box" label="TID 321" rules="all" xml:id="table_TID_321_Parameters">
                <caption>Parameters</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4fc0d7fd-f6a0-463a-9bad-73bd297ac300">Parameter Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_820425ea-3512-4005-9918-bbed89f262ab">Parameter Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_10d8a235-4a96-4a49-90e4-71880bd67fb7">$Purpose</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f008d5b8-10dd-4450-8954-0679e9d9f08b">Purpose of Reference for a waveform used as a source of the measurement</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <table frame="box" label="TID 321" rules="all" xml:id="table_TID_321">
                <caption>Waveform or Temporal Coordinates</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_85f4b0de-69cb-4cb0-9c9c-155e74ea90d6"/>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_af143f7b-6b65-4e8a-bbca-baa091de04e2">NL</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cb6cbf98-62b5-4ddc-9de8-2bb24291877c">Rel with Parent</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_34f344b4-b416-4ee7-9e2b-fa32af4f3100">VT</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0b68f4ee-3fab-4bdb-8907-31853975ca28">Concept Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_98820300-76da-47af-b72f-5c6ca5d39839">VM</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6ec58d21-4ef3-4ca0-bf73-fdd41a4447fe">Req Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bf809437-ccff-42fa-93e0-d8abc2eaa3b8">Condition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_29671141-ca40-4b18-b2f4-ba95f4bdca59">Value Set Constraint</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ac0fb94a-d0d3-4b06-a702-ec652580983e">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_79a39b69-2cdd-45a2-9858-8c2091b62dff"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9e574468-bed4-411f-89a3-2f84bc8ecc21">INFERRED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_113cd5c6-d27b-4a80-99d1-5f56955264d1">WAVEFORM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_54c693e7-f74c-4ea2-886b-6512f9c67682">$Purpose</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fad8d8e8-d639-424d-9ab2-c7c449e6be61">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_91f647af-84bc-43c9-a726-582a5e05df3b">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7f28f708-776d-441a-adaa-e532bf362155">XOR Rows 2, 3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8b3ebcbc-3b55-4f6a-bd65-5812056c6ce6"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e710cf00-3d41-4f50-81f6-9433f72a2db0">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3aea2952-194c-465f-ae05-024e08294ccf"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ab65a1d4-f32c-4621-b2bf-0dec6bf3a846">R-INFERRED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d9193e03-0479-4fbd-a72e-2e68a02efdae">WAVEFORM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_aa418de8-4034-453e-ac3c-c5991f1a71e3"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fd93aff8-4f31-4abb-9315-64841bd4dacc">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a0546d70-6aaa-41aa-92ee-185f417eac53">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d1e4ba49-996d-452c-8d2a-933c041c0404">XOR Rows 1, 3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e7c06286-aec2-4874-9096-aa2664263f4f"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3cb25d62-8a73-47dc-9ab7-209ffec5d131">3</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cba8f261-3fb6-45d8-b1d4-4270cab50d0a"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8dc67cd0-ee9d-4fce-b179-382dc7bdcbd7">INFERRED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_322f393d-e282-4305-be08-0d4ebe3d3e50">TCOORD</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6cb7a734-3037-4ca0-8e88-92e15db7f3a9">$Purpose</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3443dc49-4ea3-4848-bdcf-03458016c14f">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fa4074d7-e00c-4606-a911-fdd2dbbfa92b">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8e37aa2a-faff-41d8-b6b6-3b6bd10f8f21">XOR Rows 1, 2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c6cc009f-4981-44c5-9730-ddc19c02b769"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_200e2050-6259-44eb-90d3-2a21d5b88ad1">4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a6c8a13c-faf8-4b7f-af55-d40fc54cf502">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_38dcec74-ff32-4a7a-a844-22c46ab2d430">SELECTED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0e0cca61-800e-4c39-af6d-7a1aee3244a4">WAVEFORM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_36aecfbb-cc13-49a7-aabb-57fc44ad04fa"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d56198ef-b279-42aa-b626-b129c11fb45c">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_90f5cff2-b254-4330-a19b-aebc18095c6c">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_66062a81-5a8a-41e9-a6a7-b22760f1ea1b">XOR Row 5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0eb82f78-a498-4ce4-b8c2-968b8e042c5f"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b2aea49b-df60-4553-a506-366a223977df">5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3d0851e6-252d-4d09-a96a-acd8a3d132bd">&gt;</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3dcc0886-e92c-42d5-8e41-00c5c2183604">R-SELECTED FROM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_17be6630-8d0e-484e-80ce-0c145b0ee59f">WAVEFORM</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0e251192-45a9-4cba-9d2d-ffc676e80793"/>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ea1e62d9-82dc-4f31-9b05-534193cc66bf">1</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_512c2801-d3e6-495b-aa14-a2ed7358b536">MC</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_608fb84a-c7bb-4c3e-aa0e-b409a3b0c3e1">XOR Row 4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_811532c8-53f3-41a0-9268-473b8b248b2f"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
         

        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
      
        
        
        
    </chapter>
</book>