<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-1875</title>
    <subtitle>Two Dimensional Measurement Graph in SR has no concept for data values</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>March 2019 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2019/01/17</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Kevin ODonnell</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/11/09</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1875</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Two Dimensional Measurement Graph in SR has no concept for data values</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>TID 3990 Two Dimensional Measurement Graph row 4 CONTAINER says "no concept name".</para>
                    <para>One alternative would be to use $MeasurementGraph like the other rows.</para>
                    <para>The other alternative, proposed here, since we never say "no concept name", is to just leave it empty and add an explanation in the content item descriptions. It is permitted to have empty Concept Names for container content items that are not root content items.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        
        <section label="TID 3990" status="3" xml:id="sect_TID_3990">
            <title>Two Dimensional Measurement Graph</title>
            <para xml:id="para_f1da522b-2555-48c7-acc1-20726ef01308">Generic Template representing arbitrary two-dimensional graphs.</para>
            <table frame="box" label="TID 3990" rules="all" xml:id="table_TID_3990_Parameters">
               <caption>Parameters</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_22e89e70-41d2-40ea-9938-fc1528b8a46f">Parameter Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_96ee9a58-98d4-4b26-bb7d-0a3b7ceb7eb2">Parameter Usage</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9cb0db3c-b8a8-471d-8732-95c1f2956156">$MeasurementGraph</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5660be67-2913-4949-872a-1c7accfbeb8e">Describes what the graph is about</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_07c7e0b3-3989-48c1-8514-678e489190a4">$X-Concept</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7be9994a-3a84-4e65-955b-39643f5a70e7">Concept of the X-Axis of the graph</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7bf99bd2-aec8-443e-88aa-0d9f3bb08684">$Y-Concept</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8314956c-ddff-4e08-9c27-fcd13b47ff8e">Concept of the Y-Axis of the graph</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f2818537-d3d2-4bf3-99e3-9f8706e04ef9">$X-AxisUnit</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61b10ea1-aac7-4f96-8317-9b41f174dbce">Unit of the x-axis data elements</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_45b1b08a-0856-4d5a-aad6-ac57efd4eda0">$Y-AxisUnit</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f1026b8a-b195-475a-8560-7c9b18c7b266">Unit of the y-axis data elements</para>
                     </td>
                  </tr>
               </tbody>
            </table>
            <variablelist spacing="compact">
               <varlistentry>
                  <term>
                     <emphasis role="bold">Type:</emphasis>
                  </term>
                  <listitem>
                     <para xml:id="para_d805226f-1e06-4f5a-9b76-29592b80a383">
                        <emphasis role="bold">Extensible</emphasis>
                     </para>
                  </listitem>
               </varlistentry>
               <varlistentry>
                  <term>
                     <emphasis role="bold">Order:</emphasis>
                  </term>
                  <listitem>
                     <para xml:id="para_a7259faf-7e84-4ba2-88ee-6fce66f280b5">
                        <emphasis role="bold">Significant</emphasis>
                     </para>
                  </listitem>
               </varlistentry>
               <varlistentry>
                  <term>
                     <emphasis role="bold">Root:</emphasis>
                  </term>
                  <listitem>
                     <para xml:id="para_983feb47-df65-4fd1-9323-b9ceca43a5a1">
                        <emphasis role="bold">No</emphasis>
                     </para>
                  </listitem>
               </varlistentry>
            </variablelist>
            <table frame="box" label="TID 3990" rules="all" xml:id="table_TID_3990">
               <caption>Two Dimensional Measurement Graph</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c2e212be-4da4-4e93-ac73-6056f05f83a9"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e6ee47b0-8c83-4762-a0c7-a1ca5cef3def">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5e11f452-2ded-4eb7-87f5-4fd39804bd56">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_645f4adb-e016-4bda-87cd-d33639088b50">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e9d7c822-a1c0-4763-99d9-d658b278f4af">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5c7eb563-5241-4521-83c9-9005adc48f47">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_71040705-b35c-4ca9-9f2c-01e6a5f268ea">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b8df23ee-5f0c-4d23-9bcb-cea79f38b28d">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_94a0ba8c-eed0-4ca8-9a2b-81b1dc3f8179">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_546c6c97-cec8-43ea-9b37-897680686bed">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f4858b8b-6848-464f-a442-f86da3f6f3d5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b4577627-617d-41bf-b58e-4dff1064e94b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_716a02d6-7157-4caf-9931-fb4c6c66af28">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd6c55d1-a4a4-48a6-aa34-adff88460d28">$MeasurementGraph</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a25f16a4-5785-4167-86eb-56cb916e8c74">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_70646876-907b-49d8-91c7-c87bbcca650f">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c710dc21-a46e-4c04-a196-e6d6996f1dc1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0b9ae462-dd68-49fa-9ac4-43e022649ca6"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a1751f3c-7149-4633-aa01-7bde2c931162">2</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_aada9b58-b3ac-4036-bf5e-14178642c973">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1db6b7e6-823e-43dd-8a98-fcec9b5df512">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1ab1b23f-6f6f-46af-b110-3d54507b6492">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3cea0914-de71-4bcb-a48e-180ce50b6940">EV <link linkend="DCM_122698" xrefstyle="select: nopage">(122698, DCM, "X-Concept")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a9407b0e-ff97-4467-853d-c6f53ec2e152">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f884fd72-1d94-4376-a20d-8923d3d0c47d">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_04f7d842-386f-4c31-a3dd-0ab942730419"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0f6ce590-077b-48a0-9b72-97fe1a0bb6a1">$X-Concept</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7446fe15-906c-4dd0-a47c-52670b87cb6a">3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_37b01871-2c7d-4332-b55a-5eacc2855d5a">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b55b8f63-f209-41eb-9503-19a64c67908b">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e5c95423-b0cc-4add-bee0-1a04acab8c97">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a98206c5-6aa9-4e15-996f-537d94e2a0bc">EV <link linkend="DCM_122699" xrefstyle="select: nopage">(122699, DCM, "Y-Concept")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_01582591-76f7-4e48-b5f3-ab6226481cf4">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4bb3d0a5-61e8-4863-a1de-f82bd6bd5dd8">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c5241581-51a4-40cc-897f-72f3bdd9f6f9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6ea1abd5-26b2-4bf2-a33d-94e1b81c8fed">$Y-Concept</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b8e24f40-b77b-415a-8d74-8da26c683183">4</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4bf37587-99f3-4206-9b68-d1488422a3ff">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_68028e22-0220-4773-af10-ced6be227ddc">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5b5babb5-5b80-461c-8794-a68c47bf8a13">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5889abda-e05a-491e-8c2f-12a1b0cc3f9f"><emphasis role="bold"><emphasis role="strikethrough">no concept name</emphasis></emphasis></para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6a2067da-5ac1-431a-9fac-7bb24ba9c309">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_940ca0fb-44cb-46bb-a023-3f732f5449ed">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a3adebd-f0cd-4d9b-b72c-4dd78693b53c">IF Row 7, 8, or 9 not present</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5270981-43e2-4454-ab3d-56ffeab7c49e"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3ccaa68c-abc1-4d2e-8e84-035a3552c3df">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_41824cb4-8016-47fa-80b8-a6ebc2370bea">&gt;&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a3297bf7-3de4-4a32-ae0e-4e880c33fa1b">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_19d00973-69b5-4c46-b031-6da939344a4d">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7a1fecc8-dfea-4233-938b-2443fa4fb089">$X-Concept</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_03c03f67-1dbd-4661-9cd8-756e5df0ecd0">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a0f37adf-5acb-4497-b6f0-d540bab74310">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5517191c-cf9b-454c-a9c6-7c5489219914"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3163e8bd-0b6d-4a76-9861-c3468e47efe6">UNITS = $X-AxisUnit</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_26c2ffc1-4e5d-4567-9fc2-efaeb6b568fe">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_282c1119-ef07-4e45-ba19-f2fa5a1e8c7b">&gt;&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10d73d54-b50e-4660-a86f-4cb56d6f2c76">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cb27346d-ce8e-4bc3-babc-83b4d6a97854">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e87ed0b3-7d66-42ec-b5d7-c45a64273bc1">$Y-Concept</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b74bd3ae-8e70-4a61-bfd5-6714df8e058c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_029dc2ef-5e56-4449-a6d8-615a80fae527">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a3666d10-c25c-4d65-9833-3ec9ca4b2b36"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c9f07008-b303-4d7c-9b90-3ed1e1df51b7">UNITS = $Y-AxisUnit</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bc9c4793-bcb5-4a4e-9803-8bb184324fd0">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8365d2e1-0f81-4834-8b35-31743794acad">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_115d14be-1546-456a-ac8d-dd4c814a1a73">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c794e740-1077-407c-b080-f675f6ce490c">IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ead68f04-3624-40bd-9805-0e9066195af9">$MeasurementGraph</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e518868d-5ddb-4f51-a9cb-86fc897b557f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_247c3868-a260-464e-8908-06a685671d61">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1b2ff135-be5a-474f-8151-f286ecea26a6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1a7cdf39-b2f2-4e05-aef5-5ebe9818cc3f"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_99e36528-58c1-4f12-89a7-9fe13ac50e23">8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_315c60c8-12b1-4c92-802b-d59ed5724895">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d083c051-ea9e-4d9e-8156-f92d2f3ca243">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_934e0573-ee30-4218-be91-b08f8ad68523">WAVEFORM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_47c431f2-1cb1-4c7f-a9f7-5e6cdf4b1ece">$MeasurementGraph</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0b81f380-39e8-4885-aa6f-674fd327769e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fa2793ba-fbaa-47c9-a3f4-4061e4ee6436">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_632ed293-2471-4864-8736-6c1d754b2547"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ccf6eb17-5e07-4eec-91bc-f698a99ec5fe"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b8d7fc26-46e0-4098-a5e4-062be98d3afb">9</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_384ee369-2fcf-4772-bf7b-df50be70b756">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_82db54e4-5a20-4c7b-8956-adc79750115c">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc78a304-481d-4ed2-8a41-ed444c549c08">COMPOSITE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bdd72ee7-bfed-403e-8882-d974c1e2538c">$MeasurementGraph</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_852bbd7a-5537-4d7f-91fe-d97bb5a49732">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8cca6c4c-68db-445c-b7f2-f7013109e976">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8b02238c-1ca8-4c7a-8b11-bb40bd69ed7c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2fd9eb73-e88d-4dd0-87a3-b4735a25e3b6"/>
                     </td>
                  </tr>
               </tbody>
            </table>
            <para xml:id="para_a17e5708-8236-4e12-b942-caa12db652ad">
               <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Row 4</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">No Concept Name is used for this container.</emphasis></emphasis></para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_a60a5a0e-b392-4fed-9bd2-e8b24aca1afd">Rows 5-6</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_29a200ba-686b-43bd-b471-611b69b098b5">The X-Concept values shall be monotonically increasing.</para>
                       </td>
                   </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_51bcee53-91e5-47d2-a373-6cc798308a32">Row 7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3738ef4b-0f91-40c0-9e1f-3097b8c8e76e">Secondary Capture Image containing a bitmap representation of the graph</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_12f6bef7-beb6-4865-816f-2819a5131a4c">Row 8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_396cf00f-60eb-4fff-b008-d3f70fe2f8ac">Waveform containing a representation of the graph</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4aa61c05-3b46-42a4-b315-c7d9c31c01ad">Row 9</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_660525f2-95e5-4eb1-badd-6152ddc77c52">Composite Object containing a rendered representation of the graph</para>
                     </td>
                  </tr>
               </tbody>
            </informaltable>
         </section>
        
                 
         
         
         
         

        
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">For reference, unchanged DICOM PS3.16:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="TID 3910" status="3" xml:id="sect_TID_3910">
            <title>Flow Quantification</title>
            <para xml:id="para_4dd07736-8981-46f6-883d-c5a603089358">Contains the flow quantification measurement results</para>
            <variablelist spacing="compact">
               <varlistentry>
                  <term>
                     <emphasis role="bold">Type:</emphasis>
                  </term>
                  <listitem>
                     <para xml:id="para_ac72d67b-0493-4c58-ae7f-786bfe86b554">
                        <emphasis role="bold">Extensible</emphasis>
                     </para>
                  </listitem>
               </varlistentry>
               <varlistentry>
                  <term>
                     <emphasis role="bold">Order:</emphasis>
                  </term>
                  <listitem>
                     <para xml:id="para_2b6839d3-d53b-4e39-874b-b98d891c17c6">
                        <emphasis role="bold">Significant</emphasis>
                     </para>
                  </listitem>
               </varlistentry>
               <varlistentry>
                  <term>
                     <emphasis role="bold">Root:</emphasis>
                  </term>
                  <listitem>
                     <para xml:id="para_37e8eed7-01d6-414c-be71-1462ee9938ed">
                        <emphasis role="bold">No</emphasis>
                     </para>
                  </listitem>
               </varlistentry>
            </variablelist>
            <table frame="box" label="TID 3910" rules="all" xml:id="table_TID_3910">
               <caption>Flow Quantification</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2d0a27bf-853a-41da-81c5-5342fcb0c016"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_62477605-f703-4037-9660-df4bb913bcd3">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a4c6b3a6-a6eb-4998-a69c-d0285dbaf393">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d4acd748-66de-4037-b620-25bb455f0b4e">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ae9978cd-78ad-4e0d-ad80-45c1dda9002e">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b989dce-4d38-4c5c-ad18-5815400d01fd">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bba9bb41-6aa9-4554-9901-ec78cecf08a4">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bddefeb2-f8a7-4ae8-a640-5d927a580080">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_68f6e65d-c38b-4f3c-92ad-2abf3c88f156">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4ca72945-33ea-4a41-8e5f-fd1d2382856b">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b5d23f82-912e-4827-ae7e-2299b1e94fac"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f9334a10-d733-40cd-a02e-428b1adee02d">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1fa4c0a3-d38d-4e0d-9f8b-a0f2f57d5a12">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4273691f-9b2e-4eba-9509-ad0b3bc0e5dc">EV <link linkend="DCM_121070" xrefstyle="select: nopage">(121070, DCM, "Findings")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c1022a9a-a8ce-4d19-b07a-a45289b26056">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f43c7390-7a39-4e83-9b6c-a6f708c83ff4">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7652ad26-4977-4448-8a48-543b2bb2a29e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eb1b5bb8-e764-490c-958a-15cd42e4fa6e"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_070dbccc-0510-47e9-8a5a-e05e47626666">2</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ff9a4dd5-2d9a-4e39-895a-f25159ed585e">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9cb0dcdc-730b-45fd-8106-8f7a9921792c">HAS CONCEPT MOD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b26d51e1-906f-4b6c-9358-2bdebb176cb6">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9c16a220-bf4f-4868-80c3-c3ac8c6ac41d">EV <link linkend="DCM_111004" xrefstyle="select: nopage">(111004, DCM, "Analysis Performed")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_10b8ffc3-def4-4702-a769-c1b4a14bd859">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a728046b-1258-4a6d-b5a9-519b0dfb2b59">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9149efb3-736a-4a6b-9a8d-4fcaa502c827"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cc553f56-059f-4941-a1e1-c0610ebc9bcd">EV <link linkend="DCM_122604" xrefstyle="select: nopage">(122604, DCM, "Flow Quantification")</link>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c1e7fb75-5398-4a1a-a3c4-701d8f733e46">3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_32922320-c0d0-465e-9556-6f689a472754">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_555d0d8e-3e28-4d58-9d1b-2a9f6c2bf025">HAS OBS CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d53c72ba-1df2-4039-8cf1-2480ef4d7903">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3bb7dc3b-8361-4db4-99a5-c379d28cf8ec">D<xref linkend="sect_TID_3929" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_30619879-5d16-42af-909c-7d904c8b6542">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a7912dc4-d90f-4baf-b2cf-7b74838766a4">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_58d1b3e9-59d6-4991-bffa-16ef0eb1dfd7"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1cb5ed36-2b27-4bcf-8bd2-546467b5d432"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_842f07ab-011b-42b7-b07f-02748f3fdd63">4</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10fc00e3-df1d-4219-aae3-6315e80c2886">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7ffea243-e60c-4de0-bf90-c748bee290f2">HAS OBS CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8e192595-c692-4d89-bd6e-292f0fcea09a">DATETIME</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5095079d-a227-41f6-85e9-e651ebde7e56">EV <link xl:href="http://snomed.info/id/398201009" xl:show="new">(G-D321, SRT, "Start DateTime")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_41b20129-5404-4181-97b2-35a53a07a62f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f1988fa5-dae9-4034-98e4-67408b37a6a4">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_853182db-210a-42ec-a3af-b974bf367c92"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_222ce8f0-a3e5-4672-b3d9-ab8d3450633a"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_858bbefe-08eb-40a1-87af-11417dc2608a">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c8cf7bc7-7eca-4463-8ff9-8d43134b7e56">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fe52bea2-784b-4bdb-b158-59512fb9f624">HAS OBS CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f3f17629-cabe-4c82-b83d-86e70408c2e7">DATETIME</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f99a9802-6048-4f01-8d39-b825e3844083">EV <link xl:href="http://snomed.info/id/397898000" xl:show="new">(G-D320, SRT, "Stop DateTime")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cf45d5bb-34eb-48ec-925a-46c3968d0fab">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bf35fb5d-0298-4645-9d2e-b53442992c47">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2d0c9892-647c-4972-88e3-92fd3f62ec05"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e28b6f67-c09e-411c-bad3-83525f2b2cd8"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5545b145-5529-46eb-a886-77e13cad0e20">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5b51cb9a-dc07-4b11-a91e-34e686521889">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b2a64fa6-f962-4aeb-9f2b-73fc45388318">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4648ee14-90d1-4fb3-a9d5-4bf7bca4edd8">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a7d71ebe-a089-4c06-9acf-66901144b345">D<xref linkend="sect_TID_3990" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bbd66303-6f39-407d-a546-b13349a1fd80">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fa15832f-4131-4b31-99ce-0233b5d028fe">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_590d6dee-a011-40e0-b439-902586c7cfe4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c66d6029-b4f3-4949-9943-5c6cb6352d7b">$MeasurementGraph = EV <link linkend="DCM_122667" xrefstyle="select: nopage">(122667, DCM, "Blood velocity vs. time of cardiac cycle")</link>
                        </para>
                        <para xml:id="para_4b89de9c-3408-41a0-80e3-0cf62b5ace4d">$X-Concept = EV <link linkend="DCM_122666" xrefstyle="select: nopage">(122666, DCM, "Time relative to R-wave peak")</link>
                        </para>
                        <para xml:id="para_1d68acbb-174e-407d-a75d-ad78194705a4">$Y-Concept = EV <link xl:href="http://snomed.info/id/252064005" xl:show="new">(F-0319E, SRT, "Arterial Velocity")</link>
                        </para>
                        <para xml:id="para_9c1233e5-23d5-44a9-966d-6bc018aab018">$X-AxisUnits = DT (ms, UCUM, "ms")</para>
                        <para xml:id="para_c95beffa-8bf7-4058-86b2-a298317e248f">$Y-AxisUnits = DT (cm/s, UCUM, "cm/s")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ef5f765e-79bf-48bb-9807-86bd2c215797">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_93c4edf3-fa45-4852-bd7f-2ad45bb8e6ac">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_73793419-cdf3-4913-8c28-6ffe2784b6f0">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_84e898cc-4a58-4a9c-8028-60484b561fb7">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b2ddb167-7223-406e-b9f7-9b37ea1a9331">EV <link linkend="DCM_122642" xrefstyle="select: nopage">(122642, DCM, "Velocity Encoding Minimum Value")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d27c5cd5-6822-4add-ad19-d10eb76efe2c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b7c4996e-6b47-4a94-8e9b-51c2a0b0283d">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3e4c5f64-da1f-44f8-bea2-27349919fbe6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3f647cf5-82cf-4857-a51f-d567037f9f55">UNITS = DT (cm/s, UCUM, "cm/s")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_82155a1d-9127-4e89-a6cc-becdfe45656e">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a0e0da1b-dd20-4eb7-a7b9-ddba15f4b013">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_39e400f1-c8ad-46a6-9387-953395562acb">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b7fc3485-ee98-432a-adf6-8e28dfd0092d">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c0b491b2-daaf-4592-8fa1-450b58459b43">EV <link linkend="DCM_122643" xrefstyle="select: nopage">(122643, DCM, "Velocity Encoding Maximum Value")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1b24a440-1982-4e0b-b28a-90060084349f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5d915dcd-e2da-42df-be5b-98e11874d2bb">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_76ecfe2c-23c9-4019-b04e-6f8801c5b8ad"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d47230e3-7817-4893-9dec-aa10317a7de7">UNITS = DT (cm/s, UCUM, "cm/s")</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
        
      
        
        
        
    </chapter>
</book>