    <title>CP-1830</title>
    <subtitle>Use of Segmentation and Parametric Maps with Whole Slide Imaging</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/11/11</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para>
                    <para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
           <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para>
                   <para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
           </tr>
           <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>2018/06/02</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1830</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Use of Segmentation and Parametric Maps with Whole Slide Imaging</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Though Segmentation and Parametric Map objects can be used for WSI to encode annotations and the results of processing
                        by using the existing Derivation Image Functional Group to reference the source WS images when there is a 1:1 relationship
                        between segmentation and parametric map pixels and the pixels of the source images, just as for cross-sectional
                        imaging, this may not be the case, so there is a need for coordinate information,
                        except that this needs to be relative to the slide coordinate system.</para>
                    <para>Make the cross-sectional Plane Position (Patient) and Plane Orientation (Patient) Functional Groups conditional and add the Plane Position (Slide) Functional Group when appropriate, as well as the Image Orientation (Slide) Attribute.</para>
                   
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="A.32.8.4" status="4" xml:id="sect_A.32.8.4">
            <title>VL Whole Slide Microscopy Image Functional Group Macros</title>
            <table frame="box" label="A.32.8-2" rules="all" xml:id="table_A.32.8-2">
                <caption>VL Whole Slide Microscopy Image Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_661654ac-dde2-4b5b-b103-22db80f7df54">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0df0b22-b551-45da-b912-36013f2dec5a">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1d9c0115-1d06-4a82-9f8b-b435811646e1">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_eb389343-b64d-4efe-a32f-71e091518fb8">Pixel Measures</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ea1313a9-1968-4f04-8cf9-5f7e50434db1">
                                <xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b6971130-c9fd-4a1d-b939-8d43782d685c">M - Shall be used as a Shared Functional Group.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa7cff4f-5a24-4b9d-9643-63f2878c0b06">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_927096fd-ffe5-4351-a2e2-0dbf757b07dd"></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_fa450bb3-cf94-4482-9690-a8a0953834a5"></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_06fee48c-e13a-4a73-a2b0-3f7095a05e60">Referenced Image</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ac09cd80-0a48-4a55-b984-01ea10623663">
                                <xref linkend="sect_C.7.6.16.2.5" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_370b6b45-d0dc-410f-b49e-a39bbe0e4425">U</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_9248e796-3b6a-43c3-9d69-586501a1e427">Derivation Image</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b906be5b-b845-44e6-9724-a5a4f83f1afe">
                                <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e1a55ba2-e0d7-4c80-a9c9-5388750afd25">C - Required if the image or frame has been derived from another SOP Instance.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cf24da64-e959-4b71-aec1-87735508871e">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_697ef855-597d-464c-aa18-9bb3e0f2e5e7"> </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d4f78b1f-ad51-4f90-9a8f-22a269acc18a"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e9bebd70-7c3c-4d05-af0c-5393b389c5ea">Plane Position (Slide)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f0b762cb-8165-40f8-be2d-165a8c78dde2">
                                <xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b8936e04-89ca-4496-a36d-f7bfa6c6c7b5">C - Required if Dimension Organization Type (0020,9311) is not TILED_FULL; may be present otherwise.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a0eaee81-b5a4-4a0d-b33b-c95cf2b648c9">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_72be2f57-a60c-4710-b61d-24465bc4a311"> </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6ec616ba-a4ef-4b20-82c4-ca8bee1a3f30"/>
                        </td>
                    </tr>
                </tbody>
            </table>
            <note>
                <para xml:id="para_7bfc4a93-3e1d-41dd-80a0-43b1295f3dde">The Plane Position (Slide) and Optical Path Identification Macros are Type C, which allows the Per-Frame Functional Group Sequence (5200,9230) to be entirely omitted in those cases in which there are no other Per-Frame Functional Group Macros with content (i.e., the Frame Content Macro is empty).</para>
            </note>
            <section label="A.32.8.4.1" status="5" xml:id="sect_A.32.8.4.1">
                <title>VL Whole Slide Microscopy Image Functional Group Macros Content Constraints</title>
                <section label="A.32.8.4.1.1" status="6" xml:id="sect_A.32.8.4.1.1">
                    <title>Referenced Image</title>
                    <para xml:id="para_0e134aa2-5986-4abb-a0b0-c2a77e7db5ff">Defined CID for Purpose of Reference Code Sequence (0040,A170) in the Referenced Image Functional Group is CID 8120<!--<olink targetdoc="PS3.16" targetptr="sect_CID_8120" xrefstyle="select: labelnumber quotedtitle"/>-->.</para>
                </section>
                <section label="A.32.8.4.1.2" status="6" xml:id="sect_A.32.8.4.1.2">
                    <title>Plane Position (Slide)</title>
                    <para xml:id="para_bf4d5524-e59f-4f0d-80cf-a6d9dc80b5b8">Frames shall occupy plane positions in a regular tiling of the Total Image Matrix; the frames may extend beyond the edges of the Total Image Matrix. Therefore, the value of Column Position In Total Pixel Matrix (0048,021E) shall be an integer multiple of the value of Columns (0028,0011) plus a constant, and the value of Row Position In Total Pixel Matrix (0048,021F) shall be an integer multiple of the value of Rows (0028,0010) plus a (possibly different) constant.</para>
                    <para xml:id="para_48795c84-13e7-4133-b18d-49d7455ad924">...</para>
                </section>
            </section>
        </section>
        
       
        
        <section label="A.51.5" status="3" xml:id="sect_A.51.5">
            <title>Segmentation Functional Groups</title>
            <table frame="box" label="A.51-2" rules="all" xml:id="table_A.51-2">
                <caption>Segmentation Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f4dac09b-ec34-48a8-bc04-61d6eaeccf62">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3ebeea64-2c0d-4aec-8ac4-34ca38a578f5">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_07abb63b-8340-4a86-a957-b6129678667b">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d225a1f4-1580-4be6-8b8a-b0a0d94fc407">Pixel Measures</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d718e2c3-00cc-447d-8313-f46175fe2cb6">
                                <xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_edf2f027-832b-4bbb-b1d9-3a0b61c50a47">C - Required if Derivation Image Functional Group (<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>) is not present. May be present otherwise. See Section A.51.5.1<!--<xref linkend=sect_A.51.5.1" xrefstyle="select: label"/>-->
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_666fccea-1bd6-4436-af0f-b03a6b64d2d5">Plane Position (Patient)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2ac099dc-e14c-4c20-8b5f-05a98a7a0626">
                                <xref linkend="sect_C.7.6.16.2.3" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ec30dd1b-a5c6-41dd-a445-4423a8865a87">C - Required if Derivation Image Functional Group (<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>) is not present<emphasis role="bold"><emphasis role="underline"> and the Frame of Reference is defined in the patient-relative Reference Coordinate System</emphasis></emphasis>. May be present <emphasis role="bold"><emphasis role="strikethrough">otherwise</emphasis><emphasis role="underline"> if the Frame of Reference is defined in the patient-relative Reference Coordinate System</emphasis></emphasis>. See Section A.51.5.1<!--<xref linkend=sect_A.51.5.1" xrefstyle="select: label"/>-->
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8eb3193e-9456-45f6-9124-a167bce5f9b9">Plane Orientation (Patient)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f15036a1-4360-4ff4-a3be-6632cbb4c87a">
                                <xref linkend="sect_C.7.6.16.2.4" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_46755a9c-db45-40ba-9cf0-b13114a7bbc2">C - Required if Derivation Image Functional Group (<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>) is not present<emphasis role="bold"><emphasis role="underline"> and the Frame of Reference is defined in the patient-relative Reference Coordinate System</emphasis></emphasis>. May be present <emphasis role="bold"><emphasis role="strikethrough">otherwise</emphasis><emphasis role="underline"> if the Frame of Reference is defined in the patient-relative Reference Coordinate System</emphasis></emphasis>. See Section A.51.5.1<!--<xref linkend=sect_A.51.5.1" xrefstyle="select: label"/>-->
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Plane Position (Slide)</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline"><xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/></emphasis></emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">C - Required if Derivation Image Functional Group (<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>) is not present and the Frame of Reference is defined in the Slide Coordinate System. May be present if the Frame of Reference is defined in the Slide Coordinate System. See Section A.51.5.1<!--<xref linkend=sect_A.51.5.1" xrefstyle="select: label"/>-->.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_91c06f6e-a310-4239-b87b-152f277e8bb2">Derivation Image</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a526bf4f-3813-48b6-b5fe-82258a8f6126">
                                <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_59530188-7dc9-4e95-a726-ec244a4ffb19">C - Required if <emphasis role="bold"><emphasis role="strikethrough">any of </emphasis></emphasis>Pixel Measures (<xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>) or <emphasis role="bold"><emphasis role="underline">either </emphasis></emphasis>Plane Position (Patient) (<xref linkend="sect_C.7.6.16.2.3" xrefstyle="select: labelnumber"/>) or Plane Orientation (Patient) (<xref linkend="sect_C.7.6.16.2.4" xrefstyle="select: labelnumber"/>) <emphasis role="bold"><emphasis role="underline">(if the Frame of Reference is defined in the patient-relative Reference Coordinate System), or Plane Position (Slide) (if the Frame of Reference is defined in the Slide Coordinate System) </emphasis></emphasis>Functional Groups are not present. May be present otherwise. See Section A.51.5.1<!--<xref linkend=sect_A.51.5.1" xrefstyle="select: label"/>-->
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_12f90a9f-cba9-4b47-a388-469cca7cdd07">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0c59c690-b0f2-4164-872c-5c273eb4521e"> </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_901e8b81-50fe-4fee-b851-028ef2058a47"/>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="A.51.5.1" status="4" xml:id="sect_A.51.5.1">
                <title>Segmentation Functional Groups Description</title>
                <para xml:id="para_31f0915e-245f-409a-8307-e332181a2e3c">When a Frame of Reference UID is present the segment shall be specified within that coordinate system, using the Pixel Measures, <emphasis role="bold"><emphasis role="underline">and either the </emphasis></emphasis>Plane Position (Patient) and Plane Orientation (Patient) <emphasis role="bold"><emphasis role="underline">or the Plane Position (Slide) </emphasis></emphasis>Functional Groups. Since this defines the spatial relationship of the segment, the size of the segmentation frames need not be the same size, or resolution, as the image data used to generate the segment data. The Derivation Image Functional Group may also be present, to specify on which images the segmentation was actually performed (since there may be others in the same Frame of Reference that are spatially co-located, but were not used to perform the segmentation).</para>
                <para xml:id="para_d40a5ab6-c9c4-4a90-9da0-36f91fa8f599">If the Frame of Reference UID is not present, each pixel of the segmentation shall correspond to a pixel in a referenced image, using the Derivation Image Functional Group. Hence, the rows and columns of each referenced image will match the segmentation image. If both the Frame of Reference UID and the Derivation Image Functional Group are present, the segmentation and referenced image pixels need not correspond.</para>
                <para xml:id="para_4a034de9-2d78-47e1-b66b-663baca9730d">...</para>
            </section>
        </section>
        
        <section label="A.75.5" status="3" xml:id="sect_A.75.5">
            <title>Parametric Map Functional Groups</title>
            <table frame="box" label="A.75-2" rules="all" xml:id="table_PS3.3_A.75-2">
                <caption>Parametric Map Functional Group Macros</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_547ba2ad-5709-41d4-8277-d415a52b2492">Functional Group Macro</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_05ea90aa-2ea1-4793-bca8-04b6b99f35b8">Section</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_833424c0-4d32-461d-ab5d-c0291f7a5d3c">Usage</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_dca1fd4f-ad2d-445f-8e8a-6af52e58133f">Pixel Measures</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b2632bb3-d152-4530-8d3e-d267d88d9017">
                                <xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6b752c0d-cb20-4db3-8f14-d97f27b4d9bf">M</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f398bb40-c30e-46f9-8f53-8221e164ccee">Plane Position (Patient)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7c827d01-f0fb-4629-9b2b-cda9fbd8aee4">
                                <xref linkend="sect_C.7.6.16.2.3" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_32ae6baf-23ab-44ec-b197-631074c4d590"><emphasis role="bold"><emphasis role="strikethrough">M</emphasis><emphasis role="underline">C – Required if the Frame of Reference is defined in the patient-relative Reference Coordinate System.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f97d8197-7518-4a1f-a7fd-2c58c4ea6c23">Plane Orientation (Patient)</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_52d9990f-427d-4104-8749-e48b581a5cc3">
                                <xref linkend="sect_C.7.6.16.2.4" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_95a377b3-2c19-4b2f-9e0b-0f8b7eaccdcd"><emphasis role="bold"><emphasis role="strikethrough">M</emphasis><emphasis role="underline">C – Required if the Frame of Reference is defined in the patient-relative Reference Coordinate System.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Plane Position (Slide)</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline"><xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/></emphasis></emphasis>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">C - Required if the Frame of Reference is defined in the Slide Coordinate System.</emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6507bf5f-b78b-4e2c-a4b2-47d732739943">Referenced Image</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8fb6b93b-1d55-499c-b5d9-4067461fa46c">
                                <xref linkend="sect_C.7.6.16.2.5" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_09f04a09-ad2a-4d9c-b293-7757dc9f76c2">U</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_ece1354a-60f7-445b-a194-dae412b467cf">Derivation Image</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_49cd25b2-3038-4973-912d-33226db4927c">
                                <xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b35be938-40fa-4c34-8f63-e774d1277a35">C - Required if the image or frame has been derived from another SOP Instance.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7c9d1211-317b-4a79-87c8-e212d8b31c39">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_21ef828c-c552-4e16-8bef-4e7c31fc2e7e"> </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cae8e275-ae60-4d79-b6ed-0dd14c54df8b"/>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="C.8.12.4" status="4" xml:id="sect_C.8.12.4">
          <title>Whole Slide Microscopy Image Module</title>
          <table frame="box" label="C.8.12.4-1" rules="all" xml:id="table_C.8.12.4-1">
            <caption>Whole Slide Microscopy Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_19b0b4dc-4e92-4c95-9988-804684315cb7">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_96a09a1d-8879-43be-97e5-121291459369">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3e3024bf-3c64-498a-ba60-ec15eea8abd4">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_18f4a093-c520-4639-ae1b-71059a30ef26">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7a877f73-4d9d-41b5-86d4-a8c3bbfedf52">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_379f1037-05af-47e4-a7cc-263ccdb59552"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5e62e3f9-7cd2-48f1-898c-3cc78973e9f9"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8987cdf6-559b-4308-8bb6-886172153890"/>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cb610b4e-9a13-4eb4-928f-76a61e845887">Image Orientation (Slide)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_97222e38-02d5-413d-8482-35456170c383">(0048,0102)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e5057868-739b-4914-93a5-e360198f7c8e">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4c8acf7c-33d3-4db1-85f5-0b38dd6ebdc7">The direction cosines of the first row and the first column of the total pixel matrix with respect to the Slide Coordinate System Frame of Reference. See Section C.8.12.4.1.4<!--<xref linkend=sect_C.8.12.4.1.4" xrefstyle="select: label"/>-->
</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_04b27051-8d00-4970-a1c8-a7c03dd6dccd">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d8c8dbcf-a26d-4dbf-9d5e-d1c0fe71e199"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_4267c9b9-3db5-49f5-9865-44384a3de17a"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8119d7df-e33f-4f22-9f25-55b37d9988f5"/>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        
        <section label="C.8.20.2" status="4" xml:id="sect_C.8.20.2">
          <title>Segmentation Image Module</title>
          <table frame="box" label="C.8.20-2" rules="all" xml:id="table_C.8.20-2">
            <caption>Segmentation Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5ac516a3-1159-4ff7-b037-2e9d8f9240fd">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f09fb47c-78cd-4775-9d95-171f90cd203a">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a1bccd2a-d522-40b7-9120-db796ab8fc01">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b9de38ed-1b90-4723-afe4-5080da14ee7e">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_10a6e23d-6fee-4fc5-ba40-b1f0073f6ffc">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b697d7b7-6945-444b-a814-9e74c75a912a"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a376baac-a18b-44bd-b0f4-0fe4b3f9bc2f"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fea81a93-95f6-4da5-988a-84358398da66"/>
                </td>
              </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Image Orientation (Slide)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">(0048,0102)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">The direction cosines of the first row and the first column of the total pixel matrix with respect to the Slide Coordinate System Frame of Reference. See Section C.8.12.4.1.4<!--<xref linkend=sect_C.8.12.4.1.4" xrefstyle="select: label"/>-->.</emphasis></emphasis></para>
                        <para><emphasis role="bold"><emphasis role="underline">Required if Plane Position (Slide) Sequence (0048,021A) is present within a Functional Group Sequence.</emphasis></emphasis></para>
                    </td>
                </tr>
            </tbody>
          </table>
        </section>

        
      
       <section label="C.8.32.2" status="4" xml:id="sect_C.8.32.2">
          <title>Parametric Map Image Module</title>
          <table frame="box" label="C.8.32-2" rules="all" xml:id="table_PS3.3_C.8.32-2">
            <caption>Parametric Map Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_23fc7870-7a84-42b1-aaf4-f6c579c26f2b">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_149376b8-7372-4af7-8c1c-d51475cbd49d">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_114ade4b-c14b-4461-a182-5c2f0f76f53b">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6c3bbc9d-e1ad-4c40-83fe-cb7d41cf3ab0">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7015a685-5cab-47af-beb7-c5b449a8cbd6">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2a92a608-e41b-4d05-b19d-3fafa201cc67"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_15104eaa-bad1-4ca5-a5c6-031c3fdd0823"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_01bb239f-e2b6-4a27-a45b-302e49cfcdd6"/>
                </td>
              </tr>
                <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Image Orientation (Slide)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">(0048,0102)</emphasis></emphasis></para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">The direction cosines of the first row and the first column of the total pixel matrix with respect to the Slide Coordinate System Frame of Reference. See Section C.8.12.4.1.4<!--<xref linkend=sect_C.8.12.4.1.4" xrefstyle="select: label"/>-->.</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">Required if Plane Position (Slide) Sequence (0048,021A) is present within a Functional Group Sequence.</emphasis></emphasis></para>
                    </td>
                </tr>
            </tbody>
          </table>
        </section>
        
        
        
            <section label="C.7.6.16.2.1" status="6" xml:id="sect_C.7.6.16.2.1">
              <title>Pixel Measures Macro</title>
              <table frame="box" label="C.7.6.16-2" rules="all" xml:id="table_C.7.6.16-2">
                <caption>Pixel Measures Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3992f536-4458-4e2d-94b4-57afc8a475db">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_337b926f-fc29-4567-9e61-3a8b0f52be53">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0611a4b5-9e2d-45ca-97ea-4d03d4d6094f">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_08b7c6a8-f3e7-4fab-bce7-cdcc7d937f78">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_56ecd3d0-ba02-47a5-8912-b235590de93b">Pixel Measures Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_051d3a9f-2d16-4272-b227-00b9cd425a74">(0028,9110)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_29b518c6-c069-4886-a7e2-1166aa6a9a27">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d90b1b3b-3f9e-4c9f-8dae-7a0f7094df35">Identifies the physical characteristics of the pixels of this frame.</para>
                      <para xml:id="para_bf5e449f-5ea0-4966-af25-0fe69fddc0c4">Only a single Item shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_2e05f4d3-26bd-47fa-9afb-a5b4fbae7212">&gt;Pixel Spacing</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_1ce11d30-84bf-4c14-8ba3-97aa67e90836">(0028,0030)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_502fe013-a4a0-40cd-94e8-27da2139be84">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_7da3ffd0-d044-4f67-8d46-b89fa2480435">Physical distance in the imaging target (patient, specimen, or phantom) between the centers of each pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing in mm. See Section 10.7.1.3<!--<xref linkend=sect_10.7.1.3" xrefstyle="select: label"/>--> for further explanation of the value order.</para>
                      <note>
                        <para xml:id="para_455e363a-b31f-40c6-8e35-a278cff5c446">In the case of CT images with an Acquisition Type (0018,9302) of CONSTANT_ANGLE, the pixel spacing is that in a plane normal to the central ray of the diverging X-Ray beam as it passes through the data collection center.</para>
                      </note>
                      <para xml:id="para_91b814f3-4fb9-4854-b861-eae3971d5485">Required if:</para>
                      <itemizedlist>
                        <listitem>
                          <para xml:id="para_46ca9662-60b2-4003-96a2-6ee71bd6919f">Volumetric Properties (0008,9206) is other than DISTORTED or SAMPLED, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_4f60e859-7de7-490f-9539-9699548d655e">SOP Class UID is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4") and Frame of Reference UID (0020,0052) is present, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_9e39180f-3d47-41ab-98fd-2cce0ccedac6">SOP Class UID is Ophthalmic Tomography Image Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.4") and Ophthalmic Volumetric Properties Flag (0022,1622) is YES, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_5e272059-3f28-466d-8527-0cb3aeff5413">SOP Class UID is Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.8").</para>
                        </listitem>
                      </itemizedlist>
                      <para xml:id="para_f6668dcf-e253-4dec-a832-9204bfb4f8b8">May be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e5827725-4fab-480a-9f50-85d2c7004b41">&gt;Slice Thickness</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_955c0dba-b1d5-4d61-9373-6253d97e345b">(0018,0050)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_93a4f798-6c58-4a3d-89a1-701e16cb9330">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_39b10803-3e0d-47b9-8bcc-980653f14b8d">Nominal reconstructed slice thickness (for tomographic imaging) or depth of field (for optical non-tomographic imaging), in mm.</para>
                      <para xml:id="para_3e015ab0-05b9-4a55-ad40-f08477bd4ad7">See Section C.7.6.16.2.3.1<!--<xref linkend=sect_C.7.6.16.2.3.1" xrefstyle="select: label"/>--> for further explanation.</para>
                      <note>
                        <para xml:id="para_ed492080-b894-4488-aaa2-9b2554cdcb1b">Depth of field may be an extended depth of field created by focus stacking (see Section C.8.12.4<!--<xref linkend=sect_C.8.12.4" xrefstyle="select: label"/>-->).</para>
                      </note>
                      <para xml:id="para_5a4e9680-d607-458f-b068-b0a1b4aa003c">Required if:</para>
                      <itemizedlist>
                        <listitem>
                          <para xml:id="para_8e861a56-8d1d-489e-9ffb-91486caca8e3">Volumetric Properties (0008,9206) is VOLUME or SAMPLED, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_2d062c45-e834-47e3-aec8-eff93e80fbb0">SOP Class UID is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4") and Frame of Reference UID (0020,0052) is present, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_06408d93-de2d-4cce-b3ba-1033777db305">SOP Class UID is Ophthalmic Tomography Image Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.4") and Ophthalmic Volumetric Properties Flag (0022,1622) is YES, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_5265c600-6690-45c1-91b4-75e87c0db4f5">SOP Class UID is Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.8").</para>
                        </listitem>
                      </itemizedlist>
                      <para xml:id="para_6555345a-53b7-4e7f-8b1f-3370a2d19e61">May be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_6d5eed76-9ae2-4505-b13d-b4bd93b2080b">&gt;Spacing Between Slices</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0891b7bd-67fa-4963-ac40-5fea88778ee4">(0018,0088)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_65e86e48-bb45-4fdd-8624-0d7e1912ac27">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3874a405-c3b1-4f6b-a78c-8dbcbbacb876">Spacing between adjacent slices, in mm. The spacing is measured from the center-to-center of each slice, and if present shall not be negative.</para>
                      <para xml:id="para_2a27a0ca-e588-4dd3-8eec-5a225b88a3ab">Required if Dimension Organization Type (0020,9311) is TILED_FULL and Total Pixel Matrix Focal Planes (0048,0303) is greater than 1. May be present otherwise.</para>
                      <note>
                        <para xml:id="para_46d23721-f2aa-485f-a5a5-817268ce8789">In the case of Whole Slide Images, Spacing Between Slices (0018,0088) describes the spacing of focal planes separately encoded,
                          and is distinct from Distance Between Focal Planes (0048,0014),
                          which describes in what manner different focal planes were combined into a single encoded plane (focus stacking).</para>
                      </note>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
            <section label="C.7.6.16.2.3" status="6" xml:id="sect_C.7.6.16.2.3">
              <title>Plane Position (Patient) Macro</title>
              <table frame="box" label="C.7.6.16-4" rules="all" xml:id="table_C.7.6.16-4">
                <caption>Plane Position (Patient) Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_637cf64e-0b8a-47ef-bcd4-ca7e42634c91">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_5dfd7dbc-fa9e-44b7-90b0-1d92649efbf8">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8c1a4ee9-e5df-49f6-bf6b-e3162d8b9cd1">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8a742c75-9728-4561-a94c-e156416ea0c1">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_2c06b2ff-6ef9-44df-8ccc-8e608e93fde2">Plane Position Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_1471ba7c-6d18-422e-9a33-cb86b21a1be4">(0020,9113)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_b6b6f4e0-76a9-4db5-a659-284c99a7d7f0">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_56fc9f22-e504-4af2-bdf0-7328e12b6f61">Identifies the position of the plane of this frame.</para>
                      <para xml:id="para_5b5fae81-13fe-4b0c-b951-307365d0c16d">Only a single Item shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b0388894-c20a-488a-9971-75668bcd53a6">&gt;Image Position (Patient)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_04801b64-7750-41ae-baf3-97bd0b50ca29">(0020,0032)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0697a067-69f5-4550-9e46-5e8120ca49fe">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_aed7e3f5-bf0e-45d2-8cd9-7a4b0f30cc5d">The x, y, and z coordinates of the upper left hand corner (center of the first voxel transmitted) of the frame, in mm. See Section C.7.6.2.1.1<!--<xref linkend=sect_C.7.6.2.1.1" xrefstyle="select: label"/>--> and Section C.7.6.16.2.3.1<!--<xref linkend=sect_C.7.6.16.2.3.1" xrefstyle="select: label"/>--> for further explanation.</para>
                      <note>
                        <para xml:id="para_ba69d0b7-048d-4de7-bfde-b1b4100d17c8">In the case of CT images with an Acquisition Type (0018,9302) of CONSTANT_ANGLE the image plane is defined to pass through the data collection center and be normal to the central ray of the diverging X-Ray beam.</para>
                      </note>
                      <para xml:id="para_78ce9d13-b459-4402-b1fa-c36ab053ee4f">Required if:</para>
                      <itemizedlist>
                        <listitem>
                          <para xml:id="para_d99dc091-ccbd-47f1-a3d9-0c722f1cd249">Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and Volumetric Properties (0008,9206) of this frame is other than DISTORTED, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_f02e21ce-756d-407e-8825-9b4e3e9a458b">SOP Class UID is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4") and Frame of Reference UID (0020,0052) is present, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_c277e7b4-392d-4787-82eb-b985e627ef13">SOP Class UID is Ophthalmic Tomography Image Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.4") and Ophthalmic Volumetric Properties Flag (0022,1622) is YES, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_b116f203-fd97-459f-b53d-3199c19d449a">SOP Class UID is Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.8").</para>
                        </listitem>
                      </itemizedlist>
                      <para xml:id="para_03f651cf-c561-49ed-9703-f4589f54f7d9">May be present otherwise.</para>
                    </td>
                  </tr>
                </tbody>
              </table>
              <section label="C.7.6.16.2.3.1" status="7" xml:id="sect_C.7.6.16.2.3.1">
                <title>Position and Orientation for SAMPLED Frames</title>
                    <para>...</para>
              </section>
            </section>
            <section label="C.7.6.16.2.4" status="6" xml:id="sect_C.7.6.16.2.4">
              <title>Plane Orientation (Patient) Macro</title>
              <table frame="box" label="C.7.6.16-5" rules="all" xml:id="table_C.7.6.16-5">
                <caption>Plane Orientation (Patient) Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_a3ca5dee-ffe4-4e04-8f21-9b5517b2aa03">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_237796ac-bd86-4e10-b3a2-88145e5cbf09">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_23d4eb95-3b6b-47c2-b5b5-847391bca75f">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_18c51aed-bf5b-4ffb-93d4-c89d83e4a30c">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3033e5c2-a163-4cc0-af88-2078a74d50e8">Plane Orientation Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_ba5c0d92-9ec3-4655-8d77-9502eb4de1e7">(0020,9116)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_f698b59a-df11-45d4-90d3-87d158549aca">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_f549dfb8-5444-4505-9f0d-d32ef0edf061">Identifies orientation of the plane of this frame.</para>
                      <para xml:id="para_930cf617-7dbf-4a27-ae1b-3b59f7bf2ae2">Only a single Item shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3e872fe2-fa3f-47f8-9a47-ac3278d4168d">&gt;Image Orientation (Patient)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_d3c33485-64af-43bc-be44-cb382e4925a3">(0020,0037)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_130f7df4-bc28-4907-b3cb-131af952b528">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_872e4c29-5e52-4cab-b59c-2cdc29aa9421">The direction cosines of the first row and the first column with respect to the patient. See Section C.7.6.2.1.1<!--<xref linkend=sect_C.7.6.2.1.1" xrefstyle="select: label"/>--> and Section C.7.6.16.2.3.1<!--<xref linkend=sect_C.7.6.16.2.3.1" xrefstyle="select: label"/>--> for further explanation.</para>
                      <para xml:id="para_a74162da-8b5a-4b57-8112-38a998daf489">Required if:</para>
                      <itemizedlist>
                        <listitem>
                          <para xml:id="para_29a0ca1f-1c99-4a1e-af3e-429d63aeb43b">Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and Volumetric Properties (0008,9206) of this frame is other than DISTORTED, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_bdb0a5ba-960c-48ff-9287-42036df4a13b">SOP Class UID is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4") and Frame of Reference UID (0020,0052) is present, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_dfc60e8b-494a-406a-9c81-e880f9745df5">SOP Class UID is Ophthalmic Tomography Image Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.4") and Ophthalmic Volumetric Properties Flag (0022,1622) is YES, or</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_3315744c-05ad-4e03-9e95-92d0f697f1d9">SOP Class UID is Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Storage ("1.2.840.10008.5.1.4.1.1.77.1.5.8").</para>
                        </listitem>
                      </itemizedlist>
                      <para xml:id="para_3acb06e0-ff92-4477-a8fc-2f29959efa02">May be present otherwise.</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
            <section label="C.7.6.16.2.5" status="6" xml:id="sect_C.7.6.16.2.5">
              <title>Referenced Image Macro</title>
              <table frame="box" label="C.7.6.16-6" rules="all" xml:id="table_C.7.6.16-6">
                <caption>Referenced Image Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_6da3af5a-6740-420d-8c83-50088de7de32">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_643d1156-67f8-4c9d-ae2a-856eb05b4afa">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_1617e757-5a1a-4854-9258-491c5c24a313">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_40b0abd4-e1d2-455d-8f4f-438dfd626bae">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_9ea47561-5505-4bd5-8122-5cc2211ac857">Referenced Image Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_71cd104d-1ab4-411e-886a-8d2541a86613">(0008,1140)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_f6d117e0-dba1-466f-a9a3-670cd250cb2b">2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_8f98e7a5-6508-45a6-bd26-78a9e633fe7b">The set of images or other composite SOP Instances used to plan the acquisition, if any, and other significant related images. See Section C.7.6.16.2.5.1<!--<xref linkend=sect_C.7.6.16.2.5.1" xrefstyle="select: label"/>--> for further explanation. Zero or more Items shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_be291764-e4ed-45a7-90b0-7e6aa8c13388">
                        <emphasis role="italic">&gt;Include Table 10-3<!--<xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>-->
</emphasis>
                      </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_57849f9c-0a95-416f-bc87-e13fe33b607c"/>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d74edb80-e80b-4223-b740-9fc01df338c2">&gt;Purpose of Reference Code Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_a73187f6-af8e-4f32-ac14-c89653b09fba">(0040,A170)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_fd351f5f-097e-4aa5-8831-4edeb86f1f77">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d3002b7b-2b97-44f0-b899-b410ab297b90">Describes the purpose for which the reference is made.</para>
                      <para xml:id="para_b7e03f81-d808-4684-a46a-b990cec3e025">Only a single Item shall be included in this Sequence.</para>
                      <para xml:id="para_b1a049cf-453a-42b7-9ee7-5a3de769c20e">Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage) and not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage) and not "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), may be present otherwise.</para>
                      <para xml:id="para_951dde0e-b932-49e0-a9af-c2f86d1c72e0">See Section C.7.6.16.2.5.1<!--<xref linkend=sect_C.7.6.16.2.5.1" xrefstyle="select: label"/>--> for further explanation.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_798e3224-8b34-4b68-8bf7-563de1a27d64">
                        <emphasis role="italic">&gt;&gt;Include Table 8.8-1<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
</emphasis>
                      </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_41efa90f-b594-486f-ab67-af6819e61a43">Defined CID 7201<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7201" xrefstyle="select: labelnumber quotedtitle"/>-->, or as specified in the IOD invocation of this Functional Group.</para>
                    </td>
                  </tr>
                </tbody>
              </table>
              <section label="C.7.6.16.2.5.1" status="7" xml:id="sect_C.7.6.16.2.5.1">
                <title>Use of Referenced Image Macro</title>
                <para xml:id="para_3161b247-76d4-4ba2-8b82-7436fb6b1e20">...</para>
              </section>
            </section>
            <section label="C.7.6.16.2.6" status="6" xml:id="sect_C.7.6.16.2.6">
              <title>Derivation Image Macro</title>
              <table frame="box" label="C.7.6.16-7" rules="all" xml:id="table_C.7.6.16-7">
                <caption>Derivation Image Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_2ac3b7fe-6624-4256-a133-dec8c40607fc">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_9128492c-276c-4069-8891-6db1e5d438d8">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_b1aa53c0-c57b-4cce-918c-1dc7c988337d">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3adbfc2b-51c4-41a4-9af2-3f44b49558ab">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d69233c4-13da-4c68-8516-a043ccd4940d">Derivation Image Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_5ac27ef2-c4d9-4e1b-a9a0-d6e93cea0ac8">(0008,9124)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_66119f32-1e30-4666-b2c5-e3a311b2e777">2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_86967de3-0386-4fd0-995e-a8d23bce3101">The set of Images or other composite SOP Instances that were used to derive this frame.</para>
                      <para xml:id="para_c720dda0-7765-47e0-b3d6-e9e3c5a2c2d0">Zero or more Items shall be included in this Sequence.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ec35d0a9-7a09-448a-8ee8-973ab04979ed">&gt;Derivation Description</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3b8d54c6-72cd-4cbe-b1e6-26fd871e8c35">(0008,2111)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_4f97e423-7eea-44b1-8d6f-5ec569651210">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3732fc9f-fae2-40d8-a991-06a119e79024">A text description of how this frame data was derived. See Section C.12.4.1.1<!--<xref linkend=sect_C.12.4.1.1" xrefstyle="select: label"/>--> for further explanation.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_df50da99-906c-4bf2-b03e-7904644dc62e">&gt;Derivation Code Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_9b3e6f6f-e448-4c1c-852c-334f620eb68e">(0008,9215)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_490b633c-1dcb-4e5f-b7f4-d7c95a7a589d">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_c11ee7fa-7cad-4818-852f-09d54a0783b5">A coded description of how this frame was derived. See Section C.12.4.1.1<!--<xref linkend=sect_C.12.4.1.1" xrefstyle="select: label"/>--> for further explanation.</para>
                      <para xml:id="para_381b534a-2427-4187-87f1-4ee552c27557">One or more Items shall be included in this Sequence. More than one Item indicates that successive derivation steps have been applied.</para>
                      <para xml:id="para_b214d77b-5f9d-4c3c-8c34-b1e148d15ec2">Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage) and not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage) and not "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), may be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_9b13f626-4fd7-48d1-ae9a-2f35f344f8e1">
                        <emphasis role="italic">&gt;&gt;Include Table 8.8-1<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
</emphasis>
                      </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_779715a9-1411-41ea-be1f-47a9ab4216e1">
                        <emphasis role="italic">Defined CID 7203<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7203" xrefstyle="select: labelnumber quotedtitle"/>-->.</emphasis>
                      </para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_38842ac5-d9cb-4264-bf88-6f7d4748980c">&gt;Source Image Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_e4dc5f98-f26a-469c-981f-854dbd16bf60">(0008,2112)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_892a80d1-19e5-425e-a402-4a66d0707445">2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b122d178-47e0-44b8-b201-16868702bd55">The set of Images or other Composite SOP Instances that were used to derive this frame.</para>
                      <para xml:id="para_ada48db8-a193-4052-8cac-43731a5d36ba">Zero or more Items shall be included in this Sequence. See Section C.12.4.1.2<!--<xref linkend=sect_C.12.4.1.2" xrefstyle="select: label"/>--> for further explanation.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_78fdad00-5bff-42df-b965-0b0c055cff5c">
                        <emphasis role="italic">&gt;&gt;Include Table 10-3<!--<xref linkend="table_10-3" xrefstyle="select: label quotedtitle"/>-->
</emphasis>
                      </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_afad1d04-5991-4730-9b67-832074720213"/>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_aa7b20eb-afb5-4e5b-a2d7-37f7bc31e57f">&gt;&gt;Purpose of Reference Code Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_f928d8cd-3306-4411-ab4f-ec0f4f755a1c">(0040,A170)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_898c0fa2-6909-4b7b-a8ca-83c226adce4f">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e0340a5f-4bad-4257-beb4-6bb66c01995e">Describes the purpose for which the reference is made, that is what role the source image or frame played in the derivation of this image or frame.</para>
                      <para xml:id="para_67e2bb55-1b23-4b2e-87f7-17b04633df02">Only a single Item shall be included in this Sequence.</para>
                      <para xml:id="para_8d7d4a1a-3ee1-4347-a36e-59746a78678d">Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted Enhanced CT Image Storage) and not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted Enhanced MR Image Storage) and not "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted Enhanced PET Image Storage), may be present otherwise.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_beb90cf3-f6e9-4232-b4d4-2f62c7224576">
                        <emphasis role="italic">&gt;&gt;&gt;Include Table 8.8-1<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
</emphasis>
                      </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_30946779-c590-460d-acaf-c7827cb031d0">
                        <emphasis role="italic">Defined CID 7202<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7202" xrefstyle="select: labelnumber quotedtitle"/>-->.</emphasis>
                      </para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_1392a7a6-47ce-4682-84b5-04aec6523919">&gt;&gt;Spatial Locations Preserved</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_704357f9-bb95-4324-8b76-800b6a1bc45a">(0028,135A)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_215cb573-dc45-485a-b814-aac03f4e90bd">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_cc4c42a5-016b-410f-a0d8-f8529bf27787">The extent to which the spatial locations of all pixels are preserved during the processing of the source image that resulted in the current image or frame.</para>
                      <variablelist spacing="compact">
                        <title>Enumerated Values:</title>
                        <varlistentry>
                          <term>YES</term>
                          <listitem>
                            <para xml:id="para_7cc30ce6-9149-408d-bd76-5a4d4a4f51ab"/>
                          </listitem>
                        </varlistentry>
                        <varlistentry>
                          <term>NO</term>
                          <listitem>
                            <para xml:id="para_44d55035-81bd-47b9-8d6a-a159c22e44cb"/>
                          </listitem>
                        </varlistentry>
                        <varlistentry>
                          <term>REORIENTED_ONLY</term>
                          <listitem>
                            <para xml:id="para_e6084561-4aa3-4f79-aed7-359bf76522fd">A projection radiograph that has been flipped, and/or rotated by a multiple of 90 degrees</para>
                          </listitem>
                        </varlistentry>
                      </variablelist>
                      <note>
                        <orderedlist>
                          <listitem>
                            <para xml:id="para_59e61333-5d11-4c89-b1de-409d8221b2b8">This applies not only to images with a known relationship to a 3D space, but also to projection images.
                              For example, a projection radiograph such as a mammogram that is processed by a point image processing operation such as contrast enhancement, or a smoothing or edge enhancing convolution, would have a value of YES for this Attribute.
                              A projection radiograph that had been magnified or warped geometrically would have a value of NO for this Attribute.
                              A projection radiograph that has been flipped, and/or rotated by a multiple of 90 degrees, such that transformation of pixel locations is possible by comparison of the values of Patient Orientation (0020,0020) would have a value of REORIENTED_ONLY.
                              This Attribute is typically of importance in relating images with Presentation Intent Type (0008,0068) values of FOR PROCESSING and FOR PRESENTATION.</para>
                          </listitem>
                          <listitem>
                            <para xml:id="para_02e3e4a3-4ad1-4503-9d75-3f5ecf01c633">When the value of this Attribute is NO, it is not possible to locate on the current image any pixel coordinates that are referenced relative to the source image, such as for example,
                              might be required for rendering CAD findings derived from a referenced FOR PROCESSING image on the current FOR PRESENTATION image.</para>
                          </listitem>
                        </orderedlist>
                      </note>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_5855dee4-56fd-4dcb-bded-927ea41e8972">&gt;&gt;Patient Orientation</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8df6d030-f32f-45e4-a728-27114aea4f04">(0020,0020)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_e9b5e61d-2598-4f15-8ea8-6a18b35a896e">1C</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_872e4463-f77b-4ebc-bda8-1a77b01c7867">The Patient Orientation values of the source image.</para>
                      <para xml:id="para_fcd04da0-4438-4712-8058-c66e05a8fdd6">Required if the value of Spatial Locations Preserved (0028,135A) is REORIENTED_ONLY.</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
            
            
            <section label="C.8.12.6.1" status="5" xml:id="sect_C.8.12.6.1">
                <title>Plane Position (Slide) Macro</title>
                <table frame="box" label="C.8.12.6.1-1" rules="all" xml:id="table_C.8.12.6.1-1">
                    <caption>Plane Position (Slide) Macro Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_850332ba-1490-4793-bc12-861c321fcd99">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e7771af9-29e4-4c4b-80eb-b136651661c0">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_56b2f9ce-13d4-4aac-86bc-7b8113002fbf">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_280d277c-c0c6-4a1e-bf4e-5dba8fcec357">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6fc9e512-d859-414b-ab91-50b283757162">Plane Position (Slide) Sequence</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_4cf8f79a-1e2c-49dd-902e-ff44e2eb33e3">(0048,021A)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_df3d902e-3445-4d68-88a4-09ef5c6f43fa">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_78ead0e0-8b9f-4f0d-aacf-f1f12f328162">Describes position of frame in the Total Pixel Matrix and in the Slide Coordinate System Frame of Reference.</para>
                                <para xml:id="para_9ac9348e-b5d3-43af-9f71-0fa693f9e615">Only a single Item shall be included in this Sequence.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_42b5a3fb-466f-444f-9990-4e01f51e6245">&gt;Column Position In Total Image Pixel Matrix</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9b1b992d-f50f-48bf-bd21-b97d8910ac35">(0048,021E)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_e3294c3d-0fab-4aea-ba28-9e7b5d670e03">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5ed1659c-db53-4adc-a19b-e6e3281180c4">The column position of the top left hand pixel of the frame in the Total Pixel Matrix (see Section C.8.12.4.1.1<!--<xref linkend=sect_C.8.12.4.1.1" xrefstyle="select: label"/>-->). The column position of the top left pixel of the Total Pixel Matrix is 1.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7ab3379b-b016-4574-a6a8-4e41941558b3">&gt;Row Position In Total Image Pixel Matrix</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_47b4ae6e-3377-4255-a8f2-4cb78b232699">(0048,021F)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6d8ac9f0-e970-4643-9b65-ae4b20cdae7f">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ba282150-c64b-4fcd-bbe5-f301a450ad0f">The row position of the top left hand pixel of the frame in the Total Pixel Matrix (see Section C.8.12.4.1.1<!--<xref linkend=sect_C.8.12.4.1.1" xrefstyle="select: label"/>-->). The row position of the top left pixel of the Total Pixel Matrix is 1.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f2a755e1-37e9-4650-a9fd-0b1fc7e6163b">&gt;X Offset in Slide Coordinate System</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5431da5c-89ab-4159-994b-8c4860be13d6">(0040,072A)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_54f4f27a-1d7a-44b0-9908-bdfc2d4849a3">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_55433671-c313-485c-a2ca-50e0098d68ff">The X offset in mm from the Origin of the Slide Coordinate System. See Figure C.8-16<!--<xref linkend=figure_C.8-16" xrefstyle="select: label"/>-->.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_22234278-d134-4392-91ca-32d2074ea8e4">&gt;Y Offset in Slide Coordinate System</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_ac3833b2-5048-4645-bd21-e019a40ded28">(0040,073A)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_046934cc-fb91-445d-b6bc-c139cf07789b">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_553321a3-f758-4d03-bd9c-c5081787e0ef">The Y offset in mm from the Origin of the Slide Coordinate System. See Figure C.8-16<!--<xref linkend=figure_C.8-16" xrefstyle="select: label"/>-->.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_627e2068-4560-4f16-a96c-8ff113e8dfef">&gt;Z Offset in Slide Coordinate System</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9a947af2-8ef9-477d-a148-dd0fa8b86be2">(0040,074A)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_77c0a11f-79ad-43f0-8f36-7af1b5d5c5ac">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3c462443-6570-486e-8dc4-9f25b0cb85b4">The Z offset in µm from the Origin of the Slide Coordinate System, nominally the surface of the glass slide substrate. See Figure C.8-17<!--<xref linkend=figure_C.8-17" xrefstyle="select: label"/>-->
                                </para>
                                <note>
                                    <para xml:id="para_35c60974-5494-4736-97ab-adc076b1874a">Required even if only a single focal plane was acquired.</para>
                                </note>
                            </td>
                        </tr>
                    </tbody>
                </table>
            </section>
            
            
            
            
            
            
            
            
            
            
            
            
            
            
            
            
            
            
            
        
        
        
        
        
      
      