    <title>CP-1820</title>
  <subtitle>InConcatenationTotalNumber must be greater than one if present</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/06/04</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                 xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/04/14</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1820</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: InConcatenationTotalNumber must be greater than one if present</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>If InConcatenationTotalNumber is less than one, then by definition there is no concatenation.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="C.7.6.16" status="4" xml:id="sect_C.7.6.16">
          <title>Multi-frame Functional Groups Module</title>
          <table frame="box" label="C.7.6.16-1" rules="all" xml:id="table_C.7.6.16-1">
            <caption>Multi-frame Functional Groups Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_efd07f13-1243-42bd-93c8-919ea56a1145">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_171895f0-b152-4207-973c-ffb9c84c6a01">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d1a5c9f2-8074-4a22-b652-84b9ddd75672">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_00043efd-ea22-4575-b910-c72be0c72adf">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_456f83f0-9370-4c89-b968-4f9b7b1d8c38">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d3bd8879-c474-4c60-b173-010e5162c5d8">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_61b72077-ca4b-4656-bef9-4004ba0f277b">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3aaa5c86-e6dd-41c0-9ba4-457b78303702">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_75b699b3-67f1-4db4-a88a-5f464521330f">Concatenation Frame Offset Number</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d0162595-fe77-4e63-af90-343be9459df0">(0020,9228)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_40353cf5-a23f-400c-bc2f-a398b2e298e6">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_304aea0f-4b16-4808-b33b-a0377c3fc1cc">Offset of the first frame in a multi-frame image of a concatenation. Logical frame numbers in a concatenation can be used across all its SOP instances. This offset can be applied to the implicit frame number to find the logical frame number in a concatenation. The offset is numbered from zero; i.e., the instance of a concatenation that begins with the first frame of the concatenation has a Concatenation Frame Offset Number (0020,9228) of zero.</para>
                  <para xml:id="para_1d00d051-1ffb-4329-9940-0f0b736da2c2">Required if Concatenation UID (0020,9161) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3e2125f6-5d1b-4c3d-ad74-e4bc80dd034c">Representative Frame Number</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_215068d0-ece4-4687-b181-5c65a0bfb712">(0028,6010)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e78f5152-2c6c-48fe-bf6d-6c86f462d78d">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d2ee13ab-5471-42d3-b8ab-75a7ae6d640b">The frame number selected for use as a pictorial representation (e.g., icon) of the multi-frame Image.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_86893204-d047-4fc0-9940-bc7611532fa6">Concatenation UID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_398ba4e7-d0c3-438b-aaa9-a0ad0b9ef17a">(0020,9161)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a8dc717a-70c1-421f-977a-acaafa586ceb">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_45114939-4b72-4abb-8a71-57c56955e258">Identifier of all SOP Instances that belong to the same concatenation.</para>
                  <para xml:id="para_e7080657-dbd8-410d-8d4e-acc6bc1cb2fa">Required if a group of multi-frame image SOP Instances within a Series are part of a Concatenation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ab069ad6-e365-41d1-9c86-1e50b1015eaf">SOP Instance UID of Concatenation Source</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6ad6d73a-9c80-4d67-ad11-c425ec2894e9">(0020,0242)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6eee96f6-ccd1-4109-87b2-b7628e279e89">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e795995f-62a0-44c8-98c1-cebf2bfa6abb">The SOP Instance UID of the single composite SOP Instance of which the Concatenation is a part. All SOP Instances of a concatenation shall use the same value for this Attribute, see Section C.7.6.16.1.3<!--<xref linkend="sect_C.7.6.16.1.3" xrefstyle="select: label"/>-->.</para>
                  <note>
                    <para xml:id="para_326f77dc-84d0-4948-bb19-c7374695e54f">May be used to reference the entire instance rather than individual instances of the concatenation, which may be transient (e.g., from a presentation state).</para>
                  </note>
                  <para xml:id="para_d8f04107-cc2f-475d-a868-cf78c1cf911f">Required if Concatenation UID (0020,9161) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9497fce2-bbd0-4ff7-a3cf-6075d60b798d">In-concatenation Number</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8f37c8d5-6cb6-4552-b46c-07e1738eeb13">(0020,9162)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_4b9f77fc-68a4-4e8e-830e-2a83d4dde4fe">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1289db75-3c06-41b5-aeee-408a197f08cb">Identifier for one SOP Instance belonging to a concatenation. See Section C.7.6.16.2.2.4<!--<xref linkend="sect_C.7.6.16.2.2.4" xrefstyle="select: label"/>--> for further specification. The first instance in a concatenation (that with the lowest Concatenation Frame Offset Number (0020,9228) value) shall have an In-concatenation Number (0020,9162) value of 1, and subsequent instances shall have values monotonically increasing by 1.</para>
                  <para xml:id="para_30bb36e1-a8b8-4b05-9c97-60a65ae20928">Required if Concatenation UID (0020,9161) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dc04e05a-5489-45f7-a9c8-31caba938e79">In-concatenation Total Number</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a6e2b9f2-a236-406a-b682-93c359c30a8d">(0020,9163)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_91e377df-42cf-4b85-9882-fb7957dd177d">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_613f3530-5308-453b-8bb8-fc6897b0fa46">The number of SOP Instances sharing the same Concatenation UID.</para>
              <para><emphasis role="bold"><emphasis role="underline">If present, shall have a value greater than one.</emphasis></emphasis></para>