  <title>CP-1761</title>
  <subtitle>Correct Acquisition Context and DateTime in Confidentiality Profile Attributes</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2018/01/23</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>Regis Deshayes</para><para><link xl:href="mailto:regis.deshayes@zeiss.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2016/12/07</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1761</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct Acquisition Context and DateTime in Confidentiality Profile Attributes</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.15</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>There are are two inconsistencies in Part 15’s Basic Application Level Confidentiality Profile:</para>
                  <itemizedlist>
                    <listitem><para>The Action Code for the Acquisition Context Sequence (0040,0555) is X whereas this is a Type 2 sequence in Part3’s Table C.7.6.14-1. Acquisition Context Module Attributes, which is a mandatory Module in many IODs; its Action Code should be X/Z</para></listitem>
                    <listitem><para>The Action Code for the Acquisition DateTime attribute (0008,002A) is X/D whereas this is a Type 2 attribute in Part3’s Table C.24-2. Encapsulated Document Module Attributes; its Action Code should be X/Z/D </para></listitem>
                  </itemizedlist>
               </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.15 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table frame="box" label="E.1-1" rules="all" xml:id="table_E.1-1">
        <caption>Application Level Confidentiality Profile Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c5e6e8ba-34e0-4e31-90b5-bd9ed7086ddd">
                <emphasis role="bold">Attribute Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_d02d5580-8438-4a00-8c30-ba93530bcd31">
                <emphasis role="bold">Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_705936dc-2c42-4198-b406-ba0e61b35e99">
                <emphasis role="bold">Retired (from <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>)</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a8387cac-386f-4ad9-96b3-b3dcb1b1010c">
                <emphasis role="bold">In Std. Comp. IOD (from <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_fc0ff9a1-5780-4d06-861a-24ae0ba38b47">
                <emphasis role="bold">Basic Profile</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_19ddb67a-a5d6-474d-a13f-8df27d28b707">
                <emphasis role="bold">Retain Safe Private Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5b335431-5276-4cc0-bddf-93b6d107226e">
                <emphasis role="bold">Retain UIDs Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8317e0cf-0080-48f9-8040-1f49a4478c0f">
                <emphasis role="bold">Retain Device Ident. Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_796841da-6b27-4558-99c9-99503edccc6c">
                <emphasis role="bold">Retain Patient Chars. Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_edac8181-290b-4f9c-945d-5dfc7b1a43fc">
                <emphasis role="bold">Retain Long. Full Dates Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_6305752d-56e7-4f4d-b9d8-e5052fcd6eea">
                <emphasis role="bold">Retain Long. Modif. Dates Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_84f3b128-8fe5-45bc-b04e-abc483377b06">
                <emphasis role="bold">Clean Desc. Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0adc5d55-1196-4265-8ee1-b205103c96f1">
                <emphasis role="bold">Clean Struct. Cont. Option</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5e74e547-b6fd-410e-9c09-b271915c3eeb">
                <emphasis role="bold">Clean Graph. Option</emphasis>
              </para>
            </th>
          </tr>
          
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_045bc7be-238b-4ba2-b271-d80e1a8c9748">Acquisition Context Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_30794a69-1c9f-4803-983a-76e3201d5556">(0040,0555)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_679ed018-7598-4e71-89d8-13330257e27f">N</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_e4aecdf5-5073-4584-ab3c-6cbdb2345393">Y</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_defe6a5d-af72-4b40-8b08-2c88470ca485">X<emphasis role="bold"><emphasis role="underline">/Z</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_aca1ed20-6ad4-4a2b-b071-416eb914aedf">C</para>
            </td>
            <td align="center" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0794a3e-df29-480d-b536-7f75fe1cb37c">Acquisition DateTime</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_61d31ed8-d59f-4c95-bc78-1594a0670275">(0008,002A)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_8ab00502-6192-403c-b68d-c84abf87de0e">N</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c3e10274-64dd-48e5-b94e-5bd78d3adb47">Y</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_5eaf1c7b-52fe-4f39-b6ee-6bfba802dea0">X<emphasis role="bold"><emphasis role="underline">/Z</emphasis></emphasis>/D</para>
            </td>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_7363f436-5e67-478a-82b9-9cc69b9e4782">K</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_26d2d7c6-a725-4140-9922-0dc43d5abeab">C</para>
            </td>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
            <td align="center" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </table>
      