  <title>CP-1758</title>
  <subtitle>Incorrect data element tag and name for Center Point Location in Slide Coordinates</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/03/25</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2017/09/24</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1758</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Incorrect data element tag and name for Center Point Location in Slide Coordinates</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>The description of the X, Y and Z offsets uses the wrong name and tag "Image Center Point Location (0040,073A)" rather than the "Image Center Point Coordinates Sequence (0040,071A)"; the tag used is actually Y Offset in Slide Coordinate System (0040,073A)).</para>
                    <para>Either reference the correct sequence, or don't reference the enclosing sequence at all and reference the specific X, Y and Z attributes.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
        <section label="C.8.12.2" status="4" xml:id="sect_C.8.12.2">
            <title>Slide Coordinates Module</title>
            <table frame="box" label="C.8-78" rules="all" xml:id="table_C.8-78">
                <caption>Slide Coordinates Module Attributes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3f959cb1-e121-41b5-88a3-73ec4b8fa9d3">Attribute Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cf29fc92-0651-4de5-b97a-31461e13a0ec">Tag</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_afc4f0b3-7761-4a1a-b502-343626da7a72">Type</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cf715e60-bae2-4726-9063-b7dd11bba732">Attribute Description</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_443381b0-6573-446d-83d4-888e2933fca0">Image Center Point Coordinates Sequence</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6776915a-3512-4105-b779-68c72c40db12">(0040,071A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_024d5c92-6623-4f7d-80c2-75b41d9ee72d">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_427c4f4f-04ac-4568-b41f-13fc204d9fd3">The coordinates of the center point of the Image in the Slide Coordinate System Frame of Reference.</para>
                            <para xml:id="para_78474887-9738-46d7-a94d-547420b2489e">Zero or one Item shall be included in this Sequence.</para>
                            <para xml:id="para_b22fa108-f6e9-4afe-a1f4-ff36149358fa">See <xref linkend="sect_C.8.12.2.1.1" xrefstyle="select: label"/> for further explanation.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a6a12828-f236-4666-a21c-18cc9df225f9">&gt;X Offset in Slide Coordinate System</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2753c59d-7521-4e86-b8bd-8ea8e0005c52">(0040,072A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fedaba77-7bc9-4b1e-846d-fe89590e781d">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f430bf45-5d9c-40cf-8ab9-747f793f34fb">The X offset in millimeters from the Origin of the Slide Coordinate System. See <xref linkend="figure_C.8-16" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_294b0719-0752-49e6-8bb6-728b4196b030">&gt;Y Offset in Slide Coordinate System</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_497ff905-84f2-494d-86b7-5d2f2acb7a67">(0040,073A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_bf5ba336-e772-474f-9dae-d9a131c52131">1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4c0aafc5-cfb6-4c09-b636-1a9b2e4ee779">The Y offset in millimeters from the Origin of the Slide Coordinate System. See <xref linkend="figure_C.8-16" xrefstyle="select: label"/>.</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_78383171-3389-4aff-8294-0efbf5854342">&gt;Z Offset in Slide Coordinate System</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_508470df-71f0-4aba-9c12-b946dd501fe3">(0040,074A)</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_acefe789-74f7-4bb9-9432-6f6659759f0e">2</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_680fba6a-c0c7-4976-8f17-3b2b23f5cd1e">The Z offset in microns from the image substrate reference plane (i.e., utilized surface of a glass slide).</para>
                        </td>
                    </tr>
                </tbody>
            </table>
            <section label="C.8.12.2.1" status="5" xml:id="sect_C.8.12.2.1">
                <title>Slide Coordinates Attribute Descriptions</title>
                <section label="C.8.12.2.1.1" status="6" xml:id="sect_C.8.12.2.1.1">
                    <title>Image Center Point Coordinates Sequence</title>
                    <para>...</para>
                    <para xml:id="para_832272ef-7802-428b-a0c1-06098988b172">
                        <xref linkend="figure_C.8-17" xrefstyle="select: label"/> depicts the Z-axis center point location.
                        The X-axis value of Image Center Point <emphasis role="bold"><emphasis role="strikethrough">Location</emphasis><emphasis role="underline">Coordinates Sequence</emphasis></emphasis> <emphasis role="bold"><emphasis role="strikethrough">(0040,073A)</emphasis><emphasis role="underline">(0040,071A)</emphasis></emphasis> shall increase from the Origin toward the Right Edge in Reference Slide Orientation.
                        The Y-axis value of Image Center Point <emphasis role="bold"><emphasis role="strikethrough">Location</emphasis><emphasis role="underline">Coordinates Sequence</emphasis></emphasis> <emphasis role="bold"><emphasis role="strikethrough">(0040,073A)</emphasis><emphasis role="underline">(0040,071A)</emphasis></emphasis> shall increase from the Origin toward the Label Edge in Reference Slide Orientation.
                        The Z-axis value of Image Center Point <emphasis role="bold"><emphasis role="strikethrough">Location</emphasis><emphasis role="underline">Coordinates Sequence</emphasis></emphasis> <emphasis role="bold"><emphasis role="strikethrough">(0040,073A)</emphasis><emphasis role="underline">(0040,071A)</emphasis></emphasis> shall be nominally referenced as zero at the image substrate reference plane
                        (i.e., the top surface of a glass slide) and shall increase in a positive fashion coincident with increased distance from the substrate surface.</para>
                    <informaltable rules="all" frame="box">
                        <tr>
                            <td>
                                <para><emphasis role="italic">Alternatively, since WSI uses the X, Y and Z attributes but not the enclosing sequence:</emphasis></para>
                            </td>
                        </tr>
                    </informaltable>
                    <para xml:id="para_832272ef-7802-428b-a0c1-06098988b172">
                        <xref linkend="figure_C.8-17" xrefstyle="select: label"/> depicts the Z-axis center point location. The <emphasis role="bold"><emphasis role="strikethrough">X-axis value of Image Center Point Location (0040,073A)</emphasis>
                            <emphasis role="underline">X Offset in Slide Coordinate System (0040,072A)</emphasis></emphasis> shall increase from the Origin toward the Right Edge in Reference Slide Orientation. The <emphasis role="bold"><emphasis role="strikethrough">Y-axis value of Image Center Point Location (0040,073A)</emphasis>
                                <emphasis role="underline">Y Offset in Slide Coordinate System (0040,073A)</emphasis></emphasis> shall increase from the Origin toward the Label Edge in Reference Slide Orientation. The <emphasis role="bold"><emphasis role="strikethrough">Z-axis value of Image Center Point Location (0040,073A)</emphasis>
                            <emphasis role="underline">Z Offset in Slide Coordinate System (0040,074A)</emphasis></emphasis> shall be nominally referenced as zero at the image substrate reference plane (i.e., the top surface of a glass slide) and shall increase in a positive fashion coincident with increased distance from the substrate surface.</para>
                </section>
            </section>
        </section>
      
    </chapter>