<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-1757</title>
  <subtitle>Patient Orientation should not be required for Whole Slide Images</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>March 2018 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/01/23</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2017/09/16</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1757</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Patient Orientation should not be required for Whole Slide Images</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>Patient Orientation has no meaning so should not be required for Whole Slide Images.</para>
               </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
        
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table frame="box" label="C.7-9" rules="all" xml:id="table_C.7-9">
            <caption>General Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8401777e-d9c1-4d22-8011-9a1bf0fbcea4">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_afde5dfc-2829-42bf-8491-b3bd6c747d8e">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6cd2c26f-14d7-4ff6-9db5-9a9f34361a8a">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_452e431a-5c40-49fc-be7c-dfb46dcd6833">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_73e3e3a4-664c-4b5c-b457-1b6eb0e5f91b">Patient Orientation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_97172d17-dec4-48ff-b0ae-48f7368b64f3">(0020,0020)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1802027f-66db-4bac-bf80-2c658805e5d6">2C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3cc05471-69c1-44fe-94ae-264330f3ca85">Patient direction of the rows and columns of the image. Required if image does not require Image Orientation (Patient) (0020,0037) and Image Position (Patient) (0020,0032)<emphasis role="bold"><emphasis role="underline"> or Image Orientation (Slide) (0048,0102)</emphasis></emphasis>. May be present otherwise. See Section C.7.6.1.1.1<!--<xref linkend="sect_C.7.6.1.1.1" xrefstyle="select: label"/>--> for further explanation.</para>
                  <note>
                    <para xml:id="para_09636c87-54b7-4fd6-ae07-683cdb9bf8a4">IODs may have attributes other than Patient Orientation, Image Orientation, or Image Position (Patient) to describe orientation in which case this attribute will be zero length.</para>
                  </note>
                </td>
              </tr>
              
            </tbody>
          </table>
      
    </chapter>
</book>