<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-1743</title>
   <subtitle>Update Conformance Statement example to be consistent with IHE Scheduled Workflow</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2018/01/23</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para>
                    <para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
           <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>Vladimir Kharam</para>
                 <para><link xl:href="mailto:vkharam@ramsoft.com"/></para></td>
           </tr>
           <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2017/08/23</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1743</para></td>
            </tr>
            <tr>
               <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Update Conformance Statement example to be consistent with IHE Scheduled Workflow</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.2</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                  <para>IHE SWF specifies that Requested Procedure Code be mapped from OBR-44, not OBR-4.</para>
               <para>The example in PS3.2 Annex C uses OBR-4.</para>
               <para>Also CPT-4 is mentioned as the codes to use, and this is inappropriate, since it is a billing not ordering scheme. Suggest LOINC instead (which is harmonized withg RadLex and the preferred wire encoding).</para>
                   <para>See also the IHE Code Mapping WP <link xl:href="http://www.ihe.net/uploadedFiles/Documents/Radiology/IHE_RAD_White-Paper_Codes_Rev2.0_2014-03-07.pdf"/>.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend DICOM PS3.2 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
      
      
      <section label="C.4.2.1.4.1" xml:id="sect_C.4.2.1.4.1">
              <title>Activity - Configured AE Requests MWL Query</title>
              <section label="C.4.2.1.4.1.1" xml:id="sect_C.4.2.1.4.1.1">
                <title>Description and Sequencing of Activities</title>
                <para xml:id="para_91772009-e760-467c-afa1-971a318a5913">When Modality Worklist SCUs query DICOMSRV the queries run against the Scheduled Procedure Step Worklist (referred to hereafter as the 'SPS Worklist' or 'Worklist') in the DICOMRis database. There is a configurable mapping between the Universal Service ID contained in the HL7 Order messages (see <xref linkend="table_C.8.1-4" xrefstyle="select: label"/>) and one or more Requested Procedures within the DICOMRis database. A Requested Procedure may, in turn, map to 1 or more Scheduled Procedure Steps though the relation is usually 1-to-1. This mapping is also site-configurable. The relation between Accession Number (0008,0050) and Requested Procedure ID (0040,1001) is 1-to-1 within DICOMRis and these attributes have the same value in all MWL responses. Scheduled Procedure Step entries are added and removed from the Worklist as follows:</para>
              </section>
            </section>
      
      
      <section label="C.8" status="2" xml:id="sect_C.8">
      <title>Annexes</title>
      <section label="C.8.1" xml:id="sect_C.8.1">
        <title>IOD Contents</title>
        <section label="C.8.1.3" xml:id="sect_C.8.1.3">
          <title>Attribute Mapping</title>
          <para xml:id="para_249f6b60-e6eb-4cc5-8e7e-39989da64ff8">The mapping between attributes received via HL7 from the HIS and those supplied in Modality Worklist is configurable. The default mapping is contained in the table below. Empty cells indicate that there is no mapping for the specific attribute</para>
          <table frame="box" label="C.8.1-2" rules="all" xml:id="table_C.8.1-2">
            <caption>HL7/Modality Worklist Attribute Mapping</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e19129cd-9128-4a03-a962-bba1be8b624e">DICOM Attribute</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2825399e-4234-406f-9409-e57517a4b583">DICOM Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9162c280-c663-4f23-9cec-658e9f4f9639">HL7 Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f2875dc8-4cf2-467c-a6c5-ae807861b395">HL7 Segment</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_77fe623a-68dc-41ae-b70f-3b1c464f8879">Notes</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6c5cbd46-d483-4be3-9d47-142b301b90db">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_01667585-3b19-40f5-b804-a0b6430ca7a9">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bc7bce80-66fd-4658-9d56-ebf2af8a883d">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e523507e-2d04-461c-90ff-9bd9957556c3">Requested Procedure Code Sequence</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_504077f1-9d00-4471-a8a1-c11ae34ed69a">(0032,1064)</para>
                </td>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="center" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d69563ec-fb3e-4bba-9c84-70ac485428cb">&gt;Code Value</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b4cfc2f6-0e14-4dab-93f2-9a2cdba85379">(0008,0100)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fa5ae9b0-4919-4b8b-a097-46e0814dff40">Universal Service Id</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cf5b760a-db77-492c-9d25-ecdbacfe6fff">ORM OBR:4<emphasis role="bold"><emphasis role="underline">4</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7660a3b8-3de2-4d13-a7e5-be2c7ee72d01">The value in the HL7 attribute is mapped to one or more procedure codes in the DICOMRis database. The mapping is configurable</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4548c286-8e56-49fe-b4f8-1807f43686c3">&gt;Coding Scheme Designator</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1a1c1e55-bfa2-4490-aca1-ac797af17d05">(0008,0102)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bfc8b4bc-2eed-4038-bec2-7a464c4a2a4a">Universal Service Id</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d1d91a83-d558-49fb-940e-82bce5c1af39">ORM OBR:4<emphasis role="bold"><emphasis role="underline">4</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3bc923e0-53dc-4ee7-8825-4a828177e082">Maps to a site-defined Coding Scheme, the <emphasis role="bold"><emphasis role="strikethrough">CPT</emphasis><emphasis role="underline">LOINC</emphasis></emphasis> Coding Scheme or the DICOMRis internal Coding Scheme</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e5528350-85a6-4f83-afc6-c6fdcc2cbd27">&gt;Code Meaning</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8cf38e1a-92d5-4e61-bc49-09d42c972a17">(0008,0008)</para>
                </td>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2a6d3239-b1f0-4493-8f5a-a44aa219d895">DICOMRis generated</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_053bc3b2-73a3-4d92-81d6-e25813d10aca">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="center" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a6a3b617-684b-4e3a-bcca-acb7a70f3c9a">...</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      </section>
      <section label="C.8.3" xml:id="sect_C.8.3">
        <title>Coded Terminology and Templates</title>
        <para xml:id="para_f8631fd6-1cfd-44ff-9c1b-aaa9a2df62c7">DICOMRis's usage of Coding Schemes is specified in the table below. This table lists the Coding Schemes used by DICOMRis for attributes it originates. Usage of Controlled Terminology by Applications sending IODs to DICOMRis is discussed in the relevant SOP Specific Conformance sections above. The Procedure and Protocol Codes in the DICOMRis database can be exported to files and transferred across the network using the Configuration Utility. This allows Modalities to access and incorporate these codes if so desired.</para>
        <table frame="box" label="C.8.1-4" rules="all" xml:id="table_C.8.1-4">
          <caption>DICOMRis Controlled Terminology Usage</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4edd3bc0-eee5-4e42-a46f-e9c8660dff8b">SOP Class/Service</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_abf12701-0b3e-40b1-8748-ee2e70efc84e">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_92b4a50f-e6f6-42de-8937-547340fb32ca">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a6366fff-308a-466c-af3a-c42469567818">Baseline Context ID</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_cd622a9a-a62d-4116-aa70-baea3559ff87">Coding Scheme</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e297c9c9-5ca2-4e04-9050-bc5a636e563e">Remarks</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="6" rowspan="1">
                <para xml:id="para_9ef32066-262a-4c78-b35c-8dfe889330e5">Scheduled Procedure Step Module</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fbd37d13-43f6-4817-9de7-912aaa9b19db">MWL/ C-FIND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_22781901-19af-493e-aa6a-465c2b4e37b5">&gt;Scheduled Protocol Code Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_31ad428e-7765-4256-94c7-bf0fcb521b7a">(0040,0008)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f9821a9-1a20-4ba8-a6fb-db39f971fe5c">None</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5fc30c17-b1e3-4968-954f-dec7a029863a"><emphasis role="bold"><emphasis role="strikethrough">CPT-4</emphasis><emphasis role="underline">LOINC</emphasis></emphasis>, DICOMRis Procedure, site-supplied procedure codes or site-supplied protocol codes</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0114d02a-89f6-4cf3-8acc-47e86e2296d8">At the option of the site, DICOMRis may be configured to associate <emphasis role="bold"><emphasis role="strikethrough">CPT-4</emphasis><emphasis role="underline">LOINC</emphasis></emphasis>, DICOMRis Internal codes or site-supplied procedure codes with the various procedures represented in their Item master file. The configured procedure code will be passed in this attribute unless the site has supplied and configured protocol codes to be associated with the respective procedures in addition to procedure codes. In this case the configured protocol code will be passed</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="6" rowspan="1">
                <para xml:id="para_b4b5d967-7f7a-4574-8732-b58ff29a6262">Requested Procedure Module</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2fdb8534-24d8-4ed3-a585-9d3d3c640100">Requested Procedure Code Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_74cd867a-3108-4bc0-8c03-6cac339f4e8e">(0032,1064)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea745927-7f6c-4607-8548-742b67597e48">None</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_06ef049e-3408-4c39-91de-9b1af4936cfc"><emphasis role="bold"><emphasis role="strikethrough">CPT-4</emphasis><emphasis role="underline">LOINC</emphasis></emphasis>, DICOMRis Procedure, site-supplied procedure codes</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d0095ce2-515f-4586-aaa1-82bfa211b0ad">See remarks for Scheduled Protocol Code Sequence (0040,0040). The difference is that a procedure code is always passed in this attribute rather than a protocol code</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
    </section>
      
      
      
      
      
      
      
      
      
      
      
      
      
      
      
      

    </chapter>
   
</book>