    <title>CP-1693</title>
  <subtitle>Add study longitudinal temporal offset from event for clinical trials</subtitle>
    <chapter>
        <title/>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>June 2017 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/06/07</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>TCIA</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2017/01/25</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1693</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add study longitudinal temporal offset from event for clinical trials</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.6, PS3.15, PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>For clinical trials, especially for de-identified images from which the actual dates have been remove,
                    it is often important to record not just the sequence of time points but the time interval (usually in
                    days) between events. E.g., to check that imaging has been performed within the correct time window
                  specified in the clinical trial protocol, or to compute rate of change over time.</para>
          <para>This information may be required in image and SR attributes in the Clinical Trial Study Module, and also may be required in SR templates as local observation context when when multiple measurements in the same instance are from different time points.</para>
                  <para>The word "event" is used rather than "epoch" to describe the point to which offset times are relative, since the latter refers to a period (span) of time rather a single point, and is also overloaded in the context of clinical trials
                    (see <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C2347803"/> and <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C1948053"/>).</para>
               </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
  <chapter>
    <title/>
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <section label="C.7.2.3" status="4" xml:id="sect_C.7.2.3">
      <title>Clinical Trial Study Module</title>
      <table frame="box" label="C.7-4b" rules="all" xml:id="table_C.7-4b">
        <caption>Clinical Trial Study Module Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_038f2710-65b5-45f1-b2d8-57ce788e8157">Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f141f9ed-e583-498b-b9dd-044092792e28">Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_02d777ac-e532-4f1f-80d7-dea01d049359">Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c6956c31-d9eb-46e7-9b68-d9cb30c6f1a5">Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d2ee376-92b4-4f23-9560-4fd0f5af77ad">Clinical Trial Time Point ID</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a2274dba-a374-4711-8bab-51b459c6fa6e">(0012,0050)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_98e33c92-d06d-4aef-b0cd-eb018c86a4e1">2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_79ac94a9-c09d-4fbd-979e-f58ec6940298">An identifier specifying the one or more studies that are grouped together as a clinical time point or submission in a clinical trial or research. See <xref linkend="sect_C.7.2.3.1.1" xrefstyle="select: label"/>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad358363-68c7-45af-a2d0-7041aa1a60fc">Clinical Trial Time Point Description</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_220e62b8-b898-4556-ba0a-b2f2060129ca">(0012,0051)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a378088b-4647-42e9-92c3-8ec76953ca09">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d46fbc1-b662-4a10-9a16-019e8e6d310c">A description of a set of one or more studies that are grouped together to represent a clinical time point or submission in a clinical trial or research. See <xref linkend="sect_C.7.2.3.1.1" xrefstyle="select: label"/>.</para>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Offset from Event</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0012,eee1)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">An offset in days from a particular event of significance. May be fractional. In the context of a clinical trial, this is often the days since enrollment, or the baseline imaging study.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Event Type</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0012,eee2)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1C</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">The type of event to which Longitudinal Temporal Offset from Event (0012,eee1) is relative.</emphasis></emphasis></para>
              <variablelist spacing="compact">
                <title>Defined Terms:</title>
                <varlistentry>
                  <term>ENROLLMENT</term>
                  <listitem>
                    <para><emphasis role="bold"><emphasis role="underline">Relative to enrollment of the subject in the research activity or clinical trial.</emphasis></emphasis></para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>BASELINE</term>
                  <listitem>
                    <para><emphasis role="bold"><emphasis role="underline">Relative to the baseline imaging study.</emphasis></emphasis></para>
                  </listitem>
                </varlistentry>
              </variablelist>
              <para><emphasis role="bold"><emphasis role="underline">Required if Longitudinal Temporal Offset from Event (0012,eee1) is present.</emphasis></emphasis></para>
            </td>
          </tr>
          
          
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00f184f9-964a-48e8-a642-89d9e26c9225">Consent for Clinical Trial Use Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_5db2e521-cb5b-4379-8a97-65636c8ac8db">(0012,0083)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c7d50c71-d3f8-4da7-926c-c72001735b5c">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67d5ecb8-0767-4867-bda3-04518dbf2aec">A Sequence that conveys information about consent for Clinical Trial or research use of the composite instances within this Study.</para>
              <para xml:id="para_654f352b-0275-48ab-b564-db0f30186a39">One or more Items are permitted in this Sequence.</para>
              <para xml:id="para_44735c0a-0bd5-4356-a49c-3dd0f718fd7e">See <xref linkend="sect_C.7.2.3.1.2" xrefstyle="select: label"/>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_684e49cc-38d9-48f6-afeb-cb4a29fd64b2">...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_be162d00-2e5e-4cd6-862e-244fa709b253">...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_d2efd3cf-db32-45aa-a275-161c070bf756">...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7bdd4f7a-69f4-4a06-b0a2-10ea60f30cf6">...</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.6 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <section label="6" status="1" xml:id="chapter_6">
      <title>Registry of DICOM Data Elements</title>
      
      <table frame="box" label="6-1" rules="all" xml:id="table_6-1">
        <caption>Registry of DICOM Data Elements</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3e8c52d5-3582-4ec4-a7e6-acd7604caf32">
                <emphasis role="bold">Tag</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ac298c3b-bcdd-462b-9240-c38e738bfb7c">
                <emphasis role="bold">Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_99f6b0ff-e1b0-484e-83b5-c3003d61704c">
                <emphasis role="bold">Keyword</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_42b3621d-8b30-46eb-8992-10bb21310ac7">
                <emphasis role="bold">VR</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_551ea829-a949-4bfd-a201-bb102024e64c">
                <emphasis role="bold">VM</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3a18713f-2e81-4664-b9a1-e4e3d41957ab"/>
            </th>
          </tr>
        </thead>
        <tbody>
          
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1c943b1d-bc0f-44da-be4d-19f27571a611">(0012,0051)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dccb4f89-0d6a-4e9b-9bb0-ca37db586737">Clinical Trial Time Point Description</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3d14abc-67e4-4a4a-99bf-9b5061f3d0a0">Clinical​Trial​Time​Point​Description</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_ea90d05b-fa99-4de5-ab45-b1bc3fc0ee3b">ST</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c5e96586-85b3-451e-b8a0-d2a491c31ba9">1</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1871ce46-4cb1-4cb1-8803-74af7e3c524a"/>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0012,eee1)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Offset from Event</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal​Temporal​Offset​From​Event</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">FD</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">(0012,eee2)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Event Type</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal​Temporal​Event​Type</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">CS</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          
        </tbody>
      </table>
    </section>
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.15 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <section label="E.3.6" status="3" xml:id="sect_E.3.6">
      <title>Retain Longitudinal Temporal Information Options</title>
      
      <para xml:id="para_e7eb7427-5fea-4b7b-a219-d4e9a0f8ec56">Dates and times are recognized as having a potential for leakage of identity because they constrain the number of possible individuals that could be the imaging subject, though only if there is access to other information about the individuals concerned to match it against.</para>
      <para xml:id="para_030cf9a3-f89e-4c57-b94a-842db3aecea1">However, there are applications that require dates and times to be present to able to fulfill the objective. This is particularly true in therapeutic clinical trials in which the objective is to measure change in an outcome measure over time. Further, it is often necessary to correlate information from images with information from other sources, such as clinical and laboratory data, and dates and times need to be consistent.</para>
      <para xml:id="para_8bfa03c7-fa72-4888-ad34-c8ded393c10b">Two options are specified to address these requirements:</para>
      <itemizedlist>
        <listitem>
          <para xml:id="para_714a3d71-cfae-43a6-9b0a-b80d12c4dd50">Retain Longitudinal Temporal Information With Full Dates Option</para>
        </listitem>
        <listitem>
          <para xml:id="para_576d373f-bc21-4225-a5fd-45f2b31bed59">Retain Longitudinal Temporal Information With Modified Dates Option</para>
        </listitem>
      </itemizedlist>
      <para xml:id="para_389c7767-f03f-4716-970d-72a2766aabfa">When the Retain Longitudinal Temporal Information With Full Dates Option is specified in addition to an Application Level Confidentiality Profile, any dates and times present in the Attributes shall be retained, as described in <xref linkend="table_E.1-1" xrefstyle="select: label"/>. The Attribute Longitudinal Temporal Information Modified (0028,0303) shall be added to the Dataset with a value of "UNMODIFIED".</para>
      <para xml:id="para_b71081d0-569f-4426-aa01-b0765ce1ab32">When the Retain Longitudinal Temporal Information With Modified Dates Option is specified in addition to an Application Level Confidentiality Profile, any dates and times present in the Attributes listed in <xref linkend="table_E.1-1" xrefstyle="select: label"/> shall be modified. The modification of the dates and times shall be performed in a manner that:</para>
      <itemizedlist>
        <listitem>
          <para xml:id="para_04a9122d-b1f6-4f16-96e3-46b3f1ee6165">aggregates or transforms dates so as to reduce the possibility of matching for re-identification</para>
        </listitem>
        <listitem>
          <para xml:id="para_82eae777-e9eb-4445-9b6c-bd47b153329c">preserves the gross longitudinal temporal relationships between images obtained on different dates to the extent necessary for the application</para>
        </listitem>
        <listitem>
          <para xml:id="para_c4869466-f8bb-4da5-8a7c-dd59fbb7a974">preserves the fine temporal relationships between images and real-world events to the extent necessary for analysis of the images for the application</para>
        </listitem>
      </itemizedlist>
      <para xml:id="para_eb42e924-3cfc-4d91-aa01-0f153b41b88c">The Attribute Longitudinal Temporal Information Modified (0028,0303) shall be added to the Dataset with a value of "MODIFIED".</para>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_45e31969-dc33-493e-b413-36a407968cff">Aggregation of dates may be performed by various means such as setting all dates to the first day of the month, all months to the first month of the year, etc., depending on the precision required for the application.</para>
          </listitem>
          <listitem>
            <para xml:id="para_cee62190-61bc-47ba-bf0b-23e3aefc6f74">It is possible to modify all dates and times to dummy values by shifting them relative to an arbitrary <emphasis role="bold"><emphasis role="strikethrough">epoch</emphasis><emphasis role="underline">event</emphasis></emphasis>, and hence retain the precise longitudinal temporal relationships amongst a set of studies, when either de-identification of the entire set is performed at the same time, or some sort of mapping or database is kept to repeat this process on separate occasions.<emphasis role="bold"><emphasis role="underline"> It may also be desirable to record the type of event and the temporal offset from that event, and Attributes are provided for that purpose; see Longitudinal Temporal Offset from Event (0012,eee1) and Longitudinal Temporal Event Type (0012,eee2) in the PS3.3 Clinical Trial Study Module.</emphasis></emphasis></para>
          </listitem>
          <listitem>
            <para xml:id="para_022291d8-0c4d-4725-b43f-a195028dbdcd">Transformation of dates and times should be considered together, in order to address studies that span midnight.</para>
          </listitem>
          <listitem>
            <para xml:id="para_9a0d80b4-2672-4fa3-871b-4b5668016946">Any transformation of times should be performed in such a manner as to not disrupt computations needed for analysis, such as comparison of start of injection time to the acquisition time for PET SUV, or extraction of time-intensity values from dynamic contrast enhanced studies.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_cc45149e-ac5e-4b26-bb5f-b8af861cc54d">The manner of date modification shall be described in the Conformance Statement.</para>
    </section>
    
    <informaltable rules="all" frame="box">
      <tr>
        <td>
          <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
        </td>
      </tr>
    </informaltable>
    
    <section label="TID 1502" status="3" xml:id="sect_TID_1502">
      <title>Time Point Context</title>
      <table frame="box" label="TID 1502" rules="all" xml:id="table_TID_1502">
        <caption>Time Point Context</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ea5efbc3-ccba-4112-bb51-837409ec9c73"/>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_7c7dff37-8a3c-4b5f-8fe8-c39c2ecfded4">NL</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_61e9ba6f-f9eb-4f05-8483-cf98e82ef57f">Rel with Parent</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a82363a7-637e-4a4e-a035-8dd623abba70">VT</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c065aa2c-fdb6-4058-abdf-d692c1ee6f45">Concept Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_29ca6937-3ebc-45d1-898f-11f9d11f0834">VM</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0556075d-59b1-4d2f-aab5-ea8eab43788f">Req Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8056e9bf-f1bc-42cc-8a31-23b40301e16a">Condition</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_289df25e-28b3-448c-b01c-523c969498d3">Value Set Constraint</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2c40c4b8-d9df-414b-8a0e-c217bc6a0605">1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61c63d2a-6d70-46f2-a8d0-09ea870865a0"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9f2f334-f867-44b8-8b1f-083a5ff80e83">HAS OBS CONTEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ff955ad-524f-4fb8-8351-1f3c416202c7">TEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_09c009ba-8205-4d5e-ba24-6e096bc1bd9e">EV <link linkend="DCM_126070" xrefstyle="select: nopage">(126070, DCM, "Subject Time Point Identifier")</link>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c04653ed-190b-4719-939a-5a44f6a49631">1</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_897fd8ad-92c1-4b5b-8357-b7b27ca2bed5">U</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c229d8f-248d-43cb-9291-f8c7bd9d6718"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4b5f31c-d527-4e1b-9abf-f5c5a1586e9b"/>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e49fba1c-4d4e-4634-a30b-2c130820265a">2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c002fba-7c66-4c56-b923-e1ed6f92f02a"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71debeab-fad9-4399-ba0f-b6522e27359d">HAS OBS CONTEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0d5c1b3-0cf8-4815-93c9-4ef81403b9e4">TEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_25a86fab-7187-4deb-9c69-9fb0345f8647">EV <link linkend="DCM_126071" xrefstyle="select: nopage">(126071, DCM, "Protocol Time Point Identifier")</link>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_5585ec2c-3610-46ca-93ef-0640fcb5a7eb">1</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_55492822-b820-41ca-91b8-4eb9da5c7359">U</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60f23d07-bc54-4e61-a41e-1f701ffc193e"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fce40221-6859-4b0d-a6a4-03ebae958c99"/>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1eec7ed-3f52-4f76-9ec9-91beecbdb3a7">3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_027d4fb8-f8ca-40d2-a963-29d5c280839b"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6bda98e3-c204-451c-a6be-34b13119bb33">HAS OBS CONTEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd2c6aa6-2325-4be4-8a29-2634a97c9969">TEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d14044bb-cf56-48ee-b21f-c0cc5ea03088">EV <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C2348792" xl:show="new">(C2348792, UMLS, "Time Point")</link>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a62e2d49-e05b-4469-be57-f7a796406c55">1</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6ec7dea3-26cc-4e53-bef2-af0611375557">M</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36f9ee93-14af-4589-807c-34342420ebde"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3bc43a8c-fa7a-47f7-9892-d295175fba6c"/>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_72f6248a-2d3a-43e2-a0a5-9d789ba9d7a3">4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40e787eb-5e53-4b38-a1d3-99ab6c9c93b0"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81cd391e-c8cc-46ad-ba47-ffeb9373a8f7">HAS OBS CONTEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ffd091a9-2437-4934-8a2f-608959605cca">CODE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a03db627-3fee-4d3c-8b26-53420fd1769b">EV <link linkend="DCM_126072" xrefstyle="select: nopage">(126072, DCM, "Time Point Type")</link>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_e57e4da2-9051-4e5e-afcc-e567efd5da16">1-n</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_cfcf9df1-ca02-4d99-9c48-5798dbd7f761">U</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d323298-44d5-4e31-ba47-e546df2e78da"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5350f986-cf4b-437b-a2e7-b579aa160368">B<xref linkend="sect_CID_6146" xrefstyle="select: labelnumber quotedtitle nopage"/>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61aceef9-54e2-4b6c-bf97-b1820eb2cf26">5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4da11889-f9f6-42cf-af80-ff1284f92b2e"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4f83ce4-4477-4f6f-bf3f-892eb37e9366">HAS OBS CONTEXT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b6b2b05-c1cf-47e6-88ab-c02a2b083ae2">NUM</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f38f6d9-84db-4ac8-9f7f-204bdb639d4a">EV <link linkend="DCM_126073" xrefstyle="select: nopage">(126073, DCM, "Time Point Order")</link>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_db4c1259-50ec-40ca-b5f2-dee53e904650">1</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_17202178-632b-474a-99c2-43aad71960ed">U</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_25e53873-aa3b-408b-8c31-16d18616a4a5"/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bcf20b92-88e9-4a8e-8440-2d8057756e86">UNITS = EV (1, UCUM, "no units")</para>
            </td>
          </tr>
          
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">6</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">HAS OBS CONTEXT</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">NUM</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd001" xrefstyle="select: nopage">(ddd001, DCM, "Longitudinal Temporal Offset from Event")</link></emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">UNITS = DT (d, UCUM, "days")</emphasis></emphasis></para>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">7</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">HAS CONCEPT MOD</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd002" xrefstyle="select: nopage">(ddd002, DCM, "Longitudinal Temporal Event Type")</link></emphasis></emphasis>
              </para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">D<xref linkend="sect_CID_cccc1" xrefstyle="select: labelnumber quotedtitle"/></emphasis></emphasis></para>
            </td>
          </tr>
          
          
        </tbody>
      </table>
      <para xml:id="para_501fe256-5a06-4b04-b7ca-8f36c48ced12">
        <emphasis role="bold">Content Item Descriptions</emphasis>
      </para>
      <informaltable frame="box" rules="all">
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9c40ab2-80ec-4610-882d-d548df86a3e5">Row 1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e52703b-8819-43a1-80d0-f95d9a255c6d">Usually the same value as the Clinical Trial Time Point ID (0012,0050) attribute in the Clinical Trial Study Module, though not confined to clinical trial use. May or may not be human readable, and not required to be a DICOM UID.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_048e1ee0-b2a2-46a6-9bf8-01833c7f74e7">Row 2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1de2f279-7481-4ebd-9eba-40d77534a833">All of the subjects within a treatment protocol that are examined at a particular scheduled time point (e.g., "baseline", "pre-treatment", "first post-treatment") will have the same Protocol Time Point Identifier, but different Subject Time Point Identifiers. However, in different protocols, the Protocol Time Point Identifiers for the same conceptual “time point” will be different. E.g., the “baseline” time point will have different Protocol Time Point Identifiers in different protocols. May or may not be human readable, and not required to be a DICOM UID.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_90250868-2caa-40df-9a7a-6e572174325c">Row 3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_be58dfda-7187-4045-a2d8-3c4a4d902cba">Typically a short pre-defined label that has the same scope as Protocol Time Point Identifier (i.e., same conceptual time point within a treatment protocol) but is human-readable and understandable, e.g., "BASELINE" or "TP0", "TP1", etc. Usually the same value as Clinical Trial Time Point Description (0012,0051) attribute in the Clinical Trial Study Module, though not confined to clinical trial use. The Concept Name is selected as <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C2348792" xl:show="new">(C2348792, UMLS, "Time Point")</link> (which is <link xl:href="https://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&amp;code=C68568" xl:show="new">(C68568, NCIt, "Time Point")</link>, defined as "a specific point in the time continuum, including those established relative to an event") in order to be compatible with external terminologies.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c997e50-ff6d-4aba-aa6e-cd1e4d751bc1">Row 4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d93ce58-b017-4a7c-9cf2-764159761c48">More than one type is permitted, e.g., a time point may be "posttreatment" as well as "unscheduled" or "nadir", etc.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_292fa2b1-af58-4d99-84a0-7ddb2f475e0c">Row 5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2c729ab6-11f0-4ecd-b374-be68bd351aa4">The order is expected to be monotonically increasing within a particular scope of usage, but is not required to start at 0 or 1, nor required to increase in increments of 1 or even the same increment (e.g., to allow for retrospective insertion of unscheduled time points). In clinical usage, the Time Point Order would be expected to be temporally increasing, but in a clinical trial may be a randomized reading order rather than a temporal order.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Rows 6, 7</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal temporal information may be inherited from Longitudinal Temporal Offset from Event (0012,eee1) and Longitudinal Temporal Event Type (0012,eee2) in the PS3.3 Clinical Trial Study Module, or may be specified or overridden within this template (e.g., if different measurements in the same SR Instance were measured on different time points).</emphasis></emphasis></para>
            </td>
          </tr>
        </tbody>
      </informaltable>
    </section>
    
    <section label="CID cccc1" status="3" xml:id="sect_CID_cccc1">
      <title><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Event Types</emphasis></emphasis></title>
      <variablelist spacing="compact">
        <varlistentry>
          <term> Type: </term>
          <listitem>
            <para>
              <emphasis role="bold"><emphasis role="underline">Extensible</emphasis></emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term> Version: </term>
          <listitem>
            <para>
              <emphasis role="bold"><emphasis role="underline">yyyymmdd</emphasis></emphasis>
            </para>
          </listitem>
        </varlistentry>
      </variablelist>
      <table frame="box" label="CID cccc1" rules="all" xml:id="table_CID_cccc1">
        <caption><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Event Types</emphasis></emphasis></caption>
        <thead>
          
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Coding Scheme Designator</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code <emphasis role="underline">Value</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Code Meaning</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">SNOMED-CT Concept ID</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">UMLS Concept Unique ID</emphasis></emphasis>
                </para>
              </th>
            </tr>
          
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">NCIt</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline"><link xl:href="https://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&amp;ns=NCI_Thesaurus&amp;code=C37948" xl:show="new">C37948</link></emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Enrollment</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline"><link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C1516879" xl:show="new">C1516879</link></emphasis></emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline"><link linkend="DCM_121079" xrefstyle="select: nopage">121079</link></emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Baseline</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1"><para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para></td>
            <td align="center" colspan="1" rowspan="1"><para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para></td>
          </tr>
        </tbody>
      </table>
    </section>
    
    
    <section label="D" status="1" xml:id="chapter_D">
      <title>DICOM Controlled Terminology Definitions (Normative)</title>
      <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
        <caption>DICOM Controlled Terminology Definitions</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">
                <emphasis role="bold">Definition</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">
                <emphasis role="bold">Notes</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_121079">121079</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e702c247-eb3a-40be-90b1-c223ff556618">Baseline</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_903b904e-e25f-47b3-b956-d00f0a725bd3">Initial images used to establish a beginning condition that is used for comparison over time to look for changes. [Paraphrases NCI-PT <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C1442488" xl:show="new">(C1442488, UMLS, "Baseline")</link>, which is defined as "An initial measurement that is taken at an early time point to represent a beginning condition, and is used for comparison over time to look for changes. For example, the size of a tumor will be measured before treatment (baseline) and then afterwards to see if the treatment had an effect. A starting point to which things may be compared."]</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46165b45-8571-4311-aab6-e464c16d9255"/>
            </td>
          </tr>
          
                    
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_ddd001"><emphasis role="bold"><emphasis role="underline">ddd001</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Offset from Event</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">An offset in time from a particular event of significance. In the context of a clinical trial, this is often the time since enrollment, or the baseline imaging study.</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_ddd002"><emphasis role="bold"><emphasis role="underline">ddd002</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Longitudinal Temporal Event Type</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">The type of event to which a temporal offset is relative.</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          
        </tbody>
      </table>
    </section>
    
  </chapter>