<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-1653</title>
  <subtitle>Clarify Photometric Interpretations Permitted for Decompressed and Compressed Transfer Syntaxes</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/01/23</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>Dave Harvey</para><para><link
                xl:href="mailto:dave@medicalconnections.co.uk"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>2016/06/09</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1653</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Clarify Photometric Interpretations Permitted for Decompressed and Compressed Transfer Syntaxes</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3, PS3.5, PS3.11</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para><para>Some photometric interpretations are defined primarily for use with compressed data, YBR_FULL_422 for use with (lossy) JPEG, YBR_RCT &amp; YBR_ICT for use with JPEG 2000, and YBR_PARTIAL_420 for use with MPEG.</para><para>YBR_FULL_422 has also been encountered with RLE compressed and uncompressed data (particularly in Ultrasound applications). Though there are issues with using YBR_FULL_422 (e.g., the chrominance downsampling and upsampling methods are not formally defined even though the cositing mechanism is, and different choices by sender and recipient can cause aliasing artifacts), it cannot be forbidden because it is encountered in the field. It is also included in the Ultrasound Media Application profiles. Accordingly, some clarifications and limitations are proposed. </para><para>YBR_PARTIAL_422, though permitted by the Ultrasound Media Application profiles, has apparently never been used.</para><para>YBR_RCT, YBR_ICT and YBR_PARTIAL_420
            have problems and ambiguities if used for uncompressed data, very few, if any, recipients support such use, so it is proposed to disallow their use for uncompressed data.</para>
          <para>Some existing color IODs explicitly constrain the permitted Photometric Interpretation values to those appropriate for the Transfer Syntax, either in the IOD or in IOD-family-specific Modules. Expand the list to recognize the use of YBR_FULL for RLE when permitted (and note if not). Apply similar constraints to VL, WSI, Ophthalmology IODs by adding a similar constraint to the corresponding Modules.</para><para>The issues with the Photometric Interpretations that are to be restricted from use for uncompressed formats are as follows:</para><para>
            <emphasis role="bold">YBR_RCT:</emphasis>
          </para><para>The defined conversion takes 3xN bit data, and produces 3 components that are N, N+1 &amp; N+1 bits respectively. This is because the B &amp; R components need to be signed, and can exceed the range of an N-bit signed number. For instance, pure red, in 8 bit form (R=255, G=0, B=0) transforms to (Y=63, CB= -63, CR = 192) , and there is no unambiguous way to represent both -63 and 192 in single bytes, unless we explicitly define the reconstructing algorithm to use an unsigned addition with explicitly defined/accepted wraparound. This is not a problem when encoded in JPEG 2000, which unlike the rules for DICOM uncompressed data, specifically allows differing bit depths for different components.</para><para>As there is no means in DICOM to specify different numbers of bits for different samples, this situation could not be handled without expanding the number of bits for all samples, which would make no sense, and which would break the specificities for many of the
            SOP classes which use such color data.</para><para>
            <emphasis role="bold">YBR_ICT:</emphasis>
          </para><para>In most respects this photometric interpretation is the same as YBR_FULL, but the CB &amp; CR components do not have the +128 offset to bring then from the signed byte range to the unsigned range. It is therefore possible to encode and decode this data within the ranges allowed by 3x8 bit data, but only by making an assumption (unstated in the standard) that the 2 <superscript>nd</superscript>and 3 <superscript>rd</superscript>components should be interpreted as signed data.</para><para>
            <emphasis role="bold">YBR_PARTIAL_420:</emphasis>
          </para>
                  <para>The definition in PS3.3 C.7.6.3.1.2 does not provide sufficient information to indicate how the 4 Y values, and the CB &amp; CR should be interleaved - e.g. as YYYYBRYYYYBR or as alternating lines of YYBRYYBR and YYYY. Neither solution is really satisfactory (especially if row or column numbers were odd) , and as there is no significant use case for this format in uncompressed form, it is proposed that the format should be disallowed rather than generating a lot of work to define the form of "cositing".</para><para><emphasis role="italic">[Ed.Note. Should we actually forbid uncompressed YBR_FULL_422? ... it is very rarely seen and problematic when it is.]</emphasis></para><para><emphasis role="italic">[Ed. Note. Should we more aggressively restrict the values permitted for Planar Configuration of compressed data? Also could make consistent the placement/use of the "... is irrelevant" notes in MPEG-family TS]</emphasis></para>
                  <para><emphasis role="italic">[Ed.Note. Should we try and make the pattern for IOD and Modality-specific Module (MFSC, VL, WSI, US, MR, Ophthalmology) restrictions on Photometric Interpretation more generally applicable? I.e., we could add it to the Image Pixel Macro definition of Photometric Interpretation.]</emphasis></para>
          </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
      
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td><para><emphasis role="italic">Amend DICOM PS 3.3:</emphasis></para></td>
          </tr>
        </informaltable>
      </section>
      
      
      
      <section label="A.8.5.4" status="4" xml:id="sect_A.8.5.4">
        <title>Multi-frame True Color SC Image IOD Content Constraints</title> 
        <para xml:id="para_aebd2293-14b9-49dd-9a20-849417ddc3ef">In the <xref linkend="sect_C.7.6.3" xrefstyle="select: title"/>, the following constraints apply:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_609ed3a4-a36d-4963-bffa-556b0ff252da">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_7c596bcb-cf26-4d80-a56b-c48c96cfc03d">Photometric Interpretation (0028,0004) shall be RGB for uncompressed or lossless compressed Transfer Syntaxes that do not involve color space transformations, YBR_ICT for irreversible JPEG 2000 Transfer Syntaxes, YBR_RCT for reversible JPEG 2000 Transfer Syntaxes, YBR_PARTIAL_420 for MPEG2, MPEG-4 AVC/H.264 and HEVC/H.265 Transfer Syntaxes<emphasis role="bold"><emphasis role="underline">,</emphasis><emphasis role="strikethrough"> and</emphasis></emphasis> YBR_FULL_422 for <emphasis role="bold"><emphasis role="strikethrough">other</emphasis><emphasis role="underline">JPEG</emphasis></emphasis> lossy compressed Transfer Syntaxes<emphasis role="bold"><emphasis role="underline"> and YBR_FULL for RLE Transfer Syntaxes.</emphasis></emphasis></para>
            <note>
              <para xml:id="para_36f43a08-0225-4b43-bb87-b1e1246ae3dd">Future lossless and lossy Transfer Syntaxes may lead to the need for new definitions and choices for Photometric Interpretation.</para>
            </note>
          </listitem>
          <listitem>
            <para xml:id="para_f00a1cf9-e152-4d4e-b537-20e8e3b9ac5f">Bits Allocated (0028,0100) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_094aac78-6284-438f-a87f-4776db2afa2e">Bits Stored (0028,0101) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_4532aff2-01cd-4223-b8fd-9f825c8f3d0b">High Bit (0028,0102) shall be 7</para>
          </listitem>
          <listitem>
            <para xml:id="para_86a00dd7-7c7d-4b03-81eb-e48dc3dc781b">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_fbb97425-14d1-4743-ab15-c29edba43d83">Planar Configuration (0028,0006) shall be 0 (color-by-pixel) if Photometric Interpretation (0028,0004) is RGB</para>
          </listitem>
        </itemizedlist>
      </section>
      
      <section label="A.36.4.3.1" status="5" xml:id="sect_A.36.4.3.1">
        <title>Enhanced MR Color Image IOD Content Constraints</title>
        <para xml:id="para_662d046b-30e9-4cd3-9f1b-07f2748ddec4">Photometric Interpretation (0028,0004) shall be RGB for uncompressed or lossless compressed Transfer Syntaxes that do not involve color space transformations, YBR_ICT for irreversible JPEG 2000 Transfer Syntaxes, YBR_RCT for reversible JPEG 2000 Transfer Syntaxes, YBR_PARTIAL_420 for MPEG2, MPEG-4 AVC/H.264 and HEVC/H.265 Transfer Syntaxes<emphasis role="bold"><emphasis role="underline">,</emphasis><emphasis role="strikethrough"> and</emphasis></emphasis> YBR_FULL_422 for <emphasis role="bold"><emphasis role="strikethrough">other</emphasis><emphasis role="underline">JPEG</emphasis></emphasis> lossy compressed Transfer Syntaxes<emphasis role="bold"><emphasis role="underline"> and YBR_FULL for RLE Transfer Syntaxes</emphasis></emphasis>.</para>
      </section>
      
      <section label="A.76.4" status="3" xml:id="sect_A.76.4">
        <title>Wide Field Ophthalmic Photography Stereographic Projection Image IOD Content Constraints</title>
        <para xml:id="para_c0808d5e-ec35-4a37-b9f3-2c3274d6827c">The following constraints on Series and Image attributes take precedence over the descriptions given in the Module Attribute Tables.</para>
        <section label="A.76.4.1" status="4" xml:id="sect_A.76.4.1">
          <title>Bits Allocated, Bits Stored, and High Bit</title>
          <para xml:id="para_e085a189-3eef-4ba1-ab30-29de17bc2ae6">These Attributes shall be determined based upon the Photometric Interpretation (0028,0004):</para>
          <informaltable frame="box" rules="all">
            <thead>
              <tr valign="top">
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9985780e-9f2d-485c-9fa9-6226904b3b67">Photometric Interpretation (0028,0004)</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7e354366-bb7d-4ac5-b367-4c06922deda1">Bits Allocated (0028,0100)</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_75c155c6-f078-4b20-8e35-0ef56782c9d2">Bits Stored (0028,0101)</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_911843e0-d1fb-4218-a2c3-df4cc5732269">High Bit (0028,0102)</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="2">
                  <para xml:id="para_0c6cd17c-da4d-4bc5-8650-bd3b7c8fb2da">MONOCHROME2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f833401f-cbee-4cf5-9c90-bcf07bc36ec6">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_94b8a9c1-d5ec-4578-ba36-2689113b30e2">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4f25722d-30ac-47a7-8bca-33e6b38d09f3">7</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fd3fc04a-c4a2-4636-968c-37bf18357d29">16</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d4833241-2dd5-4b1f-9fb4-1d48b9bbe212">16</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0c3e543e-f874-49f2-bb16-7221fa4e848c">15</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dca8c1fc-1434-4d15-828e-b2c5afa8dac5">RGB</para>
                  <para xml:id="para_292f0ccf-f367-4fce-aa65-f3722f4bb29e">YBR_FULL_422</para>
                  <para xml:id="para_c2993267-cc06-4f83-94e6-c9cf22683c9a">YBR_PARTIAL_420</para>
                  <para xml:id="para_5551b2f8-ab6f-420f-83eb-ff46838a8058">YBR_ICT</para>
                  <para xml:id="para_ec83dfd5-30c4-41ac-ae9c-e25a62e2d9d5">YBR_RCT</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b21ca0ee-3f17-4be6-bff4-61e6e5a342f8">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9e116d53-37fc-40ac-a441-98452fcfcea1">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e432ba8e-5525-4d53-8c17-48cd90449f59">7</para>
                </td>
              </tr>
            </tbody>
          </informaltable>
        </section>
      </section>
      
      <section label="A.77.4" status="3" xml:id="sect_A.77.4">
        <title>Wide Field Ophthalmic Photography 3D Coordinates Image IOD Content Constraints</title>
        <para xml:id="para_9b9ab69c-f85b-482e-a287-2c680ab66647">The following constraints on Series and Image attributes take precedence over the descriptions given in the Module Attribute Tables.</para>
        <section label="A.77.4.1" status="4" xml:id="sect_A.77.4.1">
          <title>Bits Allocated, Bits Stored, and High Bit</title>
          <para xml:id="para_3831d1fd-f8dd-429d-bec9-049d22a5f224">These Attributes shall be determined based upon the Photometric Interpretation (0028,0004) :</para>
          <informaltable frame="box" rules="all">
            <thead>
              <tr valign="top">
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1bac4e99-2296-4b05-8be0-92976f9850c4">Photometric Interpretation (0028,0004)</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_604b3e95-5f0a-4671-bc77-608fb5debb2c">Bits Allocated (0028,0100)</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c7a88e3c-9dc1-411a-a009-79de9b748546">Bits Stored (0028,0101)</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7aa55a87-8aee-43cf-9804-0ee0e902fb06">High Bit (0028,0102)</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="2">
                  <para xml:id="para_4e44031a-1ad6-4747-a1cb-8e24d25ff18e">MONOCHROME2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c1306297-ae3a-4ef2-bdec-cf598c81ffa0">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0e8cce75-a533-40b5-9fb6-2905379fe751">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_89099152-5153-4c67-8b3c-49bf77651f27">7</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_714797a8-f5b1-4236-8e1a-aebea8430172">16</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6791ce0a-2224-4fce-b7bc-1d7d68c8d36e">16</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_60b5b7af-27fc-43df-bec8-c9c9c9566359">15</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7ee00cfa-7e1f-4a31-beb5-ebc458ff3716">RGB</para>
                  <para xml:id="para_978c3c3a-1a77-4a94-bfc3-d7fe9710d5e5">YBR_FULL_422</para>
                  <para xml:id="para_a8b33d7d-e6c6-40c1-8f9b-c715e997a855">YBR_PARTIAL_420</para>
                  <para xml:id="para_c795ad36-3598-4d07-89e5-d655fd6957ce">YBR_ICT</para>
                  <para xml:id="para_195245fa-f2fe-443d-b855-55dd69f4ae79">YBR_RCT</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_43768e3d-2a0d-49e5-89a3-5c7b3fcae97e">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6d540aff-97f6-41e9-952f-cac3a0c99ab8">8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3dec00bf-5834-4d28-a14a-511bc7494ebb">7</para>
                </td>
              </tr>
            </tbody>
          </informaltable>
        </section>
      </section>
      
      <section label="C.7.6.3" status="4" xml:id="sect_C.7.6.3">
          <title>Image Pixel Module</title>
          <table frame="box" label="C.7-11c" rules="all" xml:id="table_C.7-11c">
            <caption>Image Pixel Description Macro Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2cdeece3-29b7-4f94-9ffc-6da19ee39eef">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ce1e96de-7531-45e6-821c-7e7b6934bb55">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_4ec31088-493d-4af5-8489-15f740e3fbdf">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_eab5ad53-0d93-407c-a912-5cefe1d30314">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7b60f64f-e6f3-426c-8197-cebcf8752c5e">Samples per Pixel</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3b03cf41-8973-47f2-895d-421e651e3451">(0028,0002)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b23c16ba-974b-4684-9a27-32d9479beb43">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cb0f5461-d46c-4022-b457-b7480a90e490">Number of samples (planes) in this image. See <xref linkend="sect_C.7.6.3.1.1" xrefstyle="select: label"/> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_09ff4c11-8bd6-4fca-a646-d8a632369d1a">Photometric Interpretation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6bc58ca3-a8ec-4459-a651-047c0b0f232b">(0028,0004)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1d8adee4-5966-4751-84ee-c0cdae4ba541">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e86b5217-5343-4fb3-9860-91926c908a63">Specifies the intended interpretation of the pixel data. See <xref linkend="sect_C.7.6.3.1.2" xrefstyle="select: label"/> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ce407029-c083-45c4-a755-a53b7ffcb8e1">Rows</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d2d76b5e-dd2d-421b-a02a-edc159db06b5">(0028,0010)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_15549346-5ea9-4a3b-8c7b-5a0afecfc5e2">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d7447378-3ce9-48b7-bb28-d8c3be47354e">Number of rows in the image.</para>
              <para><emphasis role="bold"><emphasis role="underline">Required to be an exact multiple of the horizontal downsampling factor if any of the samples (planes) are encoded downsampled in the horizontal direction for pixel data encoded in a Native (uncompressed) format. E.g., required to be an even value for a Photometric Interpretation (0028,0004) of YBR_FULL_422.</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4c4538de-13d4-4aca-b675-1163a8e72924">Columns</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_daea10fd-db12-4a49-b85f-30c08912a080">(0028,0011)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c0bb3b6d-7b7d-4aa1-83d0-1d0ed834dcdb">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6cc096f5-8430-41be-ace5-2731a56a5c2f">Number of columns in the image<emphasis role="bold"><emphasis role="underline">.</emphasis></emphasis></para>
              <para><emphasis role="bold"><emphasis role="underline">Required to be an exact multiple of the vertical downsampling factor if any of the samples (planes) are encoded downsampled in the vertical direction for pixel data encoded in a Native (uncompressed) format.</emphasis></emphasis></para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_aa7643f9-8b02-41ab-8d77-ef88769760d6">Bits Allocated</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_79d4778b-01cf-4c05-98cf-fa84bee56307">(0028,0100)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ca9c6604-30a0-40b2-8de0-3ff8c993d70b">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f5c1c3f3-19dc-4dc8-a150-62824ed991db">Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. Bits Allocated (0028,0100) shall be either 1, or a multiple of 8. See PS3.5<!--<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>--> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c968d09c-7c7e-422b-a59d-c5107f22714f">Bits Stored</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0ec29841-c8fa-43f1-b94e-8df88843397f">(0028,0101)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_65ebbb3f-553f-46e2-83e0-35302acf4ca5">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_facdc70e-f693-42d2-9f44-24fb9e71f868">Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. See PS3.5<!--<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>--> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8d1cda25-70af-468a-b492-56d715a45e5c">High Bit</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_25575ab1-5339-499e-9215-72b6301b785d">(0028,0102)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0121d43b-88b4-48dd-9a14-ada5b8fed254">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fbaff261-93e4-485e-ae26-6b8cf9c22053">Most significant bit for pixel sample data. Each sample shall have the same high bit. High Bit (0028,0102) shall be one less than Bits Stored (0028,0101). See PS3.5<!--<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>--> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3c42c86a-7ba6-4fd4-9724-ac16585edaa0">Pixel Representation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1e4975b5-7d67-4e10-917e-17c9aea71fc7">(0028,0103)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_75919236-355e-464f-a4b0-dee2fb352d8e">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ee444489-0d46-45ca-9650-139c4ead14d0">Data representation of the pixel samples. Each sample shall have the same pixel representation.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>0000H</term>
                      <listitem>
                        <para xml:id="para_c16a71d4-ba6c-4dca-9766-93c0e4986ef2">unsigned integer.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>0001H</term>
                      <listitem>
                        <para xml:id="para_8c88599e-1314-47ef-b33f-b9aeb2c2601d">2's complement</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ebf579a3-cc59-41f7-a971-84c98ab24f96">Planar Configuration</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_cda57cb7-6b9f-42b8-8c53-86d20405087c">(0028,0006)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3cba2ef5-36de-452f-8f7f-222bd8c1e58d">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ee7ce3ac-5667-4430-9070-501e82cee7cd">Indicates whether the pixel data are sent color-by-plane or color-by-pixel. Required if Samples per Pixel (0028,0002) has a value greater than 1. See <xref linkend="sect_C.7.6.3.1.3" xrefstyle="select: label"/> for further explanation.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dde268cd-efa0-419c-a5e5-ffe98af14547">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8e3f56fa-e64b-4871-b3f7-302285672045">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ea8fed14-81a7-4faa-a3ee-810aad8dd497">...</para>
                </td>
              </tr>
            </tbody>
          </table>

          
          
          <section label="C.7.6.3.1" status="5" xml:id="sect_C.7.6.3.1">
            <title>Image Pixel Attribute Descriptions</title>
            <section label="C.7.6.3.1.1" status="6" xml:id="sect_C.7.6.3.1.1">
              <title>Samples Per Pixel</title>
              <para xml:id="para_f697a89a-e79a-4383-b293-5c2041ed45e5">Samples per Pixel (0028,0002) is the number of separate planes in this image. One and three image planes are defined. Other numbers of image planes are allowed, but their meaning is not defined by this Standard.</para>
              <para xml:id="para_538b2d8c-baaf-4e40-af78-b6f18fe166fc">For monochrome (gray scale) and palette color images, the number of planes is 1. For RGB and other three vector color models, the value of this attribute is 3.</para>
              <note>
                <para xml:id="para_b2137fa4-1d70-48f4-97dd-58cbb5e34ef2">The use of a value of 4 was previously described, but the Photometric Interpretations that used it have been retired.</para>
              </note>
              <para xml:id="para_9efe0faa-7b11-4bd9-9518-10e9a0dc6d47">All image planes shall have the same number of Rows (0028,0010), Columns (0028,0011), Bits Allocated (0028,0100), Bits Stored (0028,0101), High Bit (0028,0102), Pixel Representation (0028,0103), and Pixel Aspect Ratio (0028,0034).</para>
          <note>
            <para><emphasis role="bold"><emphasis role="underline">A special case is when chrominance planes of a color Photometric Interpretation are downsampled, e.g., for a Photometric Interpretation (0028,0004) of YBR_FULL_422. In such cases, Samples per Pixel (0028,0002) describes the nominal number of channels (i.e., 3), and does not reflect that two chrominance samples are shared between four luminance samples. For YBR_FULL_422, Rows (0028,0010) and Columns (0028,0011) describe the size of the luminance plane, not the downsampled chrominance planes.</emphasis></emphasis></para>
          </note>
              <para xml:id="para_798aae68-fbc3-4de5-8bc1-cc1c8159c3c1">The data in each pixel may be represented as a "Composite Pixel Code". If Samples Per Pixel is one, the Composite Pixel Code is just the "n" bit pixel sample, where "n" = Bits Allocated. If Samples Per Pixel is greater than one, Composite Pixel Code is a "k" bit concatenation of samples, where "k" = Bits Allocated multiplied by Samples Per Pixel, and with the sample representing the vector color designated first in the Photometric Interpretation name comprising the most significant bits of the Composite Pixel Code, followed in order by the samples representing the next vector colors, with the sample representing the vector color designated last in the Photometric Interpretation name comprising the least significant bits of the Composite Pixel Code. For example, for Photometric Interpretation = "RGB", the most significant "Bits Allocated" bits contain the Red sample, the next "Bits Allocated" bits contain the Green sample, and the least significant "Bits Allocated"
              bits contain the Blue sample.</para>
            </section>
            <section label="C.7.6.3.1.2" status="6" xml:id="sect_C.7.6.3.1.2">
              <title>Photometric Interpretation</title>
              <para xml:id="para_8ea9c814-7320-42b2-afeb-001d2b215dbd">The value of Photometric Interpretation (0028,0004) specifies the intended interpretation of the image pixel data.</para>
              <para xml:id="para_3d7cf56b-f9c6-44df-a10f-d1c3c101c814">See PS3.5<!--<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>--> for <emphasis role="bold"><emphasis role="underline">additional</emphasis></emphasis>restrictions imposed by compressed Transfer Syntaxes.</para>
              <para xml:id="para_962639f9-f777-4fc0-bd94-8860e59f5ecb">The following values are defined. Other values are permitted <emphasis role="bold"><emphasis role="underline">if supported by the Transfer Syntax</emphasis></emphasis>but the meaning is not defined by this Standard.</para>
              <variablelist>
                <title>Defined Terms:</title>
                <varlistentry>
                  <term>MONOCHROME1</term>
                  <listitem>
                    <para xml:id="para_c5edb722-1cb3-4c49-8f0c-548e71e371ed">Pixel data represent a single monochrome image plane. The minimum sample value is intended to be displayed as white after any VOI gray scale transformations have been performed. See PS3.4<!--<olink targetdoc="PS3.4" targetptr="PS3.4" xrefstyle="select: labelnumber"/>-->. This value may be used only when Samples per Pixel (0028,0002) has a value of 1.<emphasis role="bold"><emphasis role="underline"> May be used for pixel data in a Native (uncompressed) or Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>MONOCHROME2</term>
                  <listitem>
                    <para xml:id="para_689f7260-aa02-41c4-9712-eb9139762527">Pixel data represent a single monochrome image plane. The minimum sample value is intended to be displayed as black after any VOI gray scale transformations have been performed. See PS3.4<!--<olink targetdoc="PS3.4" targetptr="PS3.4" xrefstyle="select: labelnumber"/>-->. This value may be used only when Samples per Pixel (0028,0002) has a value of 1.<emphasis role="bold"><emphasis role="underline"> May be used for pixel data in a Native (uncompressed) or Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>PALETTE COLOR</term>
                  <listitem>
                    <para xml:id="para_6df2a768-f435-4a02-8d25-b8d9482b98d0">Pixel data describe a color image with a single sample per pixel (single image plane). The pixel value is used as an index into each of the Red, Blue, and Green Palette Color Lookup Tables (0028,1101-1103&amp;1201-1203). This value may be used only when Samples per Pixel (0028,0002) has a value of 1.<emphasis role="bold"><emphasis role="underline"> May be used for pixel data in a Native (uncompressed) or Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis> When the Photometric Interpretation is Palette Color; Red, Blue, and Green Palette Color Lookup Tables shall be present.</para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>RGB</term>
                  <listitem>
                    <para xml:id="para_88bb278f-860b-462f-8313-c017f15e4fc2">Pixel data represent a color image described by red, green, and blue image planes. The minimum sample value for each color plane represents minimum intensity of the color. This value may be used only when Samples per Pixel (0028,0002) has a value of 3.<emphasis role="bold"><emphasis role="underline"> Planar Configuration (0028,0006) may be 0 or 1. May be used for pixel data in a Native (uncompressed) or Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>HSV</term>
                  <listitem>
                    <para xml:id="para_672a8b5d-cc07-4dd3-9272-661cfc501c80">
<emphasis role="italic">Retired.</emphasis>
</para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>ARGB</term>
                  <listitem>
                    <para xml:id="para_589cb6d1-efe5-4dbe-b0a4-a99aab13f016">
<emphasis role="italic">Retired.</emphasis>
</para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>CMYK</term>
                  <listitem>
                    <para xml:id="para_c7c7289c-658c-462b-940d-6e070b254def">
<emphasis role="italic">Retired.</emphasis>
</para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_FULL</term>
                  <listitem>
                    <para xml:id="para_28ad2fb0-4d52-47ed-9d4f-3a88de11c906">Pixel data represent a color image described by one luminance (Y) and two chrominance planes (CB and CR).</para>
                <para>This photometric interpretation may be used only when Samples per Pixel (0028,0002) has a value of 3.<emphasis role="bold"><emphasis role="underline"> Planar Configuration (0028,0006) may be 0 or 1. May be used for pixel data in a Native (uncompressed) or Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                    <para><emphasis role="bold"><emphasis role="underline">This Photometric Interpretation is primarily used with RLE compressed bit streams, for which the Planar Configuration (0028,0006) is required to be 1; see PS3.5 <xref linkend="sect_8.2.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2.2" xrefstyle="select: docname label"/>-->. When used in the US Image Module, the Planar Configuration (0028,0006) is required to be 1; see <xref linkend="sect_C.8.5.6.1.16" xrefstyle="select: label quotedtitle"/>.</emphasis></emphasis></para>
                <para>Black is represented by Y equal to zero. The absence of color is represented by both CB and CR values equal to half full scale.</para>
                    <note>
                      <para xml:id="para_d63ebf2f-9363-4b46-bf67-3db76a39310c">In the case where Bits Allocated (0028,0100) has value of 8 half full scale is 128.</para>
                    </note>
                    <para xml:id="para_0a80eafe-b06b-49bf-887e-a9ee4db9ee20">In the case where Bits Allocated (0028,0100) has a value of 8 then the following equations convert between RGB and YCBCR Photometric Interpretation.</para>
                    <para xml:id="para_6d856863-9ffb-4dca-abbe-89eb5e021a0a">Y = + .2990R + .5870G + .1140B</para>
                    <para xml:id="para_f0d67560-d2b8-4323-be43-d0dad1fa3e18">CB= - .1687R - .3313G + .5000B + 128</para>
                    <para xml:id="para_1f88a61c-5517-4897-943e-27cc8ab2b1e1">CR= + .5000R - .4187G - .0813B + 128</para>
                    <note>
                      <para xml:id="para_7ac69d6d-1a7e-46c1-8ede-1b7a17b43069">The above is based on CCIR Recommendation 601-2 dated 1990.</para>
                    </note>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_FULL_422</term>
                  <listitem>
                    <para xml:id="para_10f0e14b-ff96-40af-b5a6-070f05125b83">The same as YBR_FULL except that the CB and CR values are sampled horizontally at half the Y rate and as a result there are half as many CB and CR values as Y values.</para>
                    <para xml:id="para_4347d9f2-c4d9-43ce-81e7-0261e12813c8"><emphasis role="bold"><emphasis role="strikethrough">This Photometric Interpretation is only allowed with </emphasis></emphasis>Planar Configuration (0028,0006) <emphasis role="bold"><emphasis role="strikethrough">equal to</emphasis><emphasis role="underline">shall be</emphasis></emphasis> 0.<emphasis role="bold"><emphasis role="underline"> May be used for pixel data in a Native (uncompressed) or Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                    <note>
                      <orderedlist>
                        <listitem>
                          <para><emphasis role="bold"><emphasis role="underline">This Photometric Interpretation is primarily used with JPEG compressed bit streams, but is also occasionally used for pixel data in a Native (uncompressed) format.</emphasis></emphasis></para>
                        </listitem>
                        <listitem>
                          <para><emphasis role="bold"><emphasis role="underline">Though the chrominance channels are downsampled, there are still nominally three channels, hence Samples per Pixel (0028,0002) has a value of 3, not 2. I.e., for pixel data in a Native (uncompressed) format, the Value Length of Pixel Data (7FE0,0010) is not:</emphasis></emphasis></para>
                          <para><emphasis role="bold"><emphasis role="underline">Rows (0028,0010) * Columns (0028,0011) * Number of Frames (0028,0008) * Samples per Pixel (0028,0002) * (</emphasis></emphasis><emphasis role="underline">⌊</emphasis><emphasis role="bold"><emphasis role="underline">(Bits Allocated (0028,0100)-1)/8</emphasis></emphasis><emphasis role="underline">⌋</emphasis><emphasis role="bold"><emphasis role="underline">+1)</emphasis></emphasis></para>
                      <para><emphasis role="bold"><emphasis role="underline">padded to an even length, as it would otherwise be, but rather is:</emphasis></emphasis></para>
                          <para><emphasis role="bold"><emphasis role="underline">Rows (0028,0010) * Columns (0028,0011) * Number of Frames (0028,0008) * 2 * (</emphasis></emphasis><emphasis role="underline">⌊</emphasis><emphasis role="bold"><emphasis role="underline">(Bits Allocated (0028,0100)-1)/8</emphasis></emphasis><emphasis role="underline">⌋</emphasis><emphasis role="bold"><emphasis role="underline">+1)</emphasis></emphasis></para>
                      <para><emphasis role="bold"><emphasis role="underline">padded to an even length, since Columns (0028,0011) is also required to be a multiple of two in this case.</emphasis></emphasis></para>
                        </listitem>
                        <listitem>
                          <para><emphasis role="bold"><emphasis role="underline">When used to describe JPEG compressed bit streams, the chrominance sub-sampling in the JPEG bit stream may differ from this description. E.g., though many JPEG codecs produce only horizontally sub-sampled chrominance components, some sub-sample vertically as well. Though inaccurate, the use of YBR_FULL_422 to describe both has proven harmless.</emphasis></emphasis></para>
                        </listitem>
                      </orderedlist>
                    </note>
                <para>Two Y values shall be stored followed by one CB and one CR value. The CB and CR values shall be sampled at the location of the first of the two Y values. For each Row of Pixels, the first CB and CR samples shall be at the location of the first Y sample. The next CB and CR samples shall be at the location of the third Y sample etc.</para>
                    <note>
                      <para xml:id="para_f95aac28-98ac-42e1-b7db-f5a1a95ed6be">This subsampling is often referred to as cosited sampling.</para>
                    </note>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_PARTIAL_422</term>
                  <listitem>
                    <para xml:id="para_c2c3c216-0657-4e77-8313-2cf4ef501b94"><emphasis role="bold"><emphasis role="strikethrough">The same as YBR_FULL_422 except that:</emphasis><emphasis role="underline"><emphasis role="italic">Retired.</emphasis></emphasis></emphasis></para>
                    <orderedlist numeration="arabic">
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="strikethrough">black corresponds to Y = 16;</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="strikethrough">Y is restricted to 220 levels (i.e., the maximum value is 235);</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="strikethrough">CB and CR each has a minimum value of 16;</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="strikethrough">CB and CR are restricted to 225 levels (i.e., the maximum value is 240);</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="strikethrough">lack of color is represented by CB and CR equal to 128.</emphasis></emphasis></para>
                      </listitem>
                    </orderedlist>
                    <para><emphasis role="bold"><emphasis role="strikethrough">In the case where Bits Allocated (0028,0100) has value of 8 then the following equations convert between RGB and YBR_PARTIAL_422 Photometric Interpretation</emphasis></emphasis></para>
                    <para><emphasis role="bold"><emphasis role="strikethrough">Y = + .2568R + .5041G + .0979B + 16</emphasis></emphasis></para>
                    <para><emphasis role="bold"><emphasis role="strikethrough">CB= - .1482R - .2910G + .4392B + 128</emphasis></emphasis></para>
                    <para><emphasis role="bold"><emphasis role="strikethrough">CR= + .4392R - .3678G - .0714B + 128</emphasis></emphasis></para>
                    <note>
                      <para><emphasis role="bold"><emphasis role="strikethrough">The above is based on CCIR Recommendation 601-2 dated 1990.</emphasis></emphasis></para>
                    </note>
                  </listitem>
                </varlistentry>
                
                <varlistentry>
                  <term>YBR_PARTIAL_420</term>
                  <listitem>
                    <para xml:id="para_af2c2d32-0f4b-46c8-8086-41e62f7d9382"><emphasis role="bold"><emphasis role="underline">Pixel data represent a color image described by one luminance (Y) and two chrominance planes (CB and CR).</emphasis></emphasis></para>
                <para><emphasis role="bold"><emphasis role="underline">This photometric interpretation may be used only when Samples per Pixel (0028,0002) has a value of 3. </emphasis></emphasis>The <emphasis role="bold"><emphasis role="strikethrough">same as YBR_PARTIAL_422 except that the </emphasis></emphasis>CB and CR values are sampled horizontally and vertically at half the Y rate and as a result there are four times less CB and CR values than Y values<emphasis role="bold"><emphasis role="strikethrough">, versus twice less for YBR_PARTIAL_422</emphasis></emphasis>.</para>
                    <para xml:id="para_d99b3757-3218-4f30-88d8-a672a65e3f13"><emphasis role="bold"><emphasis role="strikethrough">This Photometric Interpretation is only allowed with </emphasis></emphasis>Planar Configuration (0028,0006) <emphasis role="bold"><emphasis role="strikethrough">equal to</emphasis><emphasis role="underline">shall be</emphasis></emphasis> 0.<emphasis role="bold"><emphasis role="underline"> Shall only be used for pixel data in an Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                <note>
                  <para><emphasis role="bold"><emphasis role="underline">This Photometric Interpretation is primarily used with MPEG compressed bit streams,</emphasis></emphasis></para>
                </note>
                    <para><emphasis role="bold"><emphasis role="underline">Luminance and chrominance values are represented as follows:</emphasis></emphasis></para>
                    <orderedlist numeration="arabic">
                      <listitem>
                        <para xml:id="para_68c2df27-d754-415e-bdaf-943c0774480b"><emphasis role="bold"><emphasis role="underline">black corresponds to Y = 16;</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para xml:id="para_29e5e871-34d5-45c2-9a86-3074d295b122"><emphasis role="bold"><emphasis role="underline">Y is restricted to 220 levels (i.e., the maximum value is 235);</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para xml:id="para_b3898e3b-005b-4631-a964-d95d415abf9a"><emphasis role="bold"><emphasis role="underline">CB and CR each has a minimum value of 16;</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para xml:id="para_bfdf57d8-12e3-4958-a742-9c1d675230c8"><emphasis role="bold"><emphasis role="underline">CB and CR are restricted to 225 levels (i.e., the maximum value is 240);</emphasis></emphasis></para>
                      </listitem>
                      <listitem>
                        <para xml:id="para_00cfc386-7699-4642-949c-c743329b8681"><emphasis role="bold"><emphasis role="underline">lack of color is represented by CB and CR equal to 128.</emphasis></emphasis></para>
                      </listitem>
                    </orderedlist>
                    <para xml:id="para_937c01e6-81aa-4080-a50f-bbcbd3e4d09a"><emphasis role="bold"><emphasis role="underline">In the case where Bits Allocated (0028,0100) has value of 8 then the following equations convert between RGB and YBR_PARTIAL_420 Photometric Interpretation</emphasis></emphasis></para>
                    <para xml:id="para_c877d543-a629-4f40-9b45-9a73cebf59e2"><emphasis role="bold"><emphasis role="underline">Y = + .2568R + .5041G + .0979B + 16</emphasis></emphasis></para>
                    <para xml:id="para_5b6d75aa-ea01-4dcd-9b94-0694d3a6b7f9"><emphasis role="bold"><emphasis role="underline">CB= - .1482R - .2910G + .4392B + 128</emphasis></emphasis></para>
                    <para xml:id="para_fd20139a-26cd-4c0d-9f4a-5d706d46bbf7"><emphasis role="bold"><emphasis role="underline">CR= + .4392R - .3678G - .0714B + 128</emphasis></emphasis></para>
                    <note>
                      <para xml:id="para_a035976e-e8bb-4971-934d-cc01779a044f"><emphasis role="bold"><emphasis role="underline">The above is based on CCIR Recommendation 601-2 dated 1990.</emphasis></emphasis></para>
                    </note>
                <para>The CB and CR values shall be sampled at the location of the first of the two Y values. For the first Row of Pixels (etc.), the first CB and CR samples shall be at the location of the first Y sample. The next CB and CR samples shall be at the location of the third Y sample etc. The next Rows of Pixels containing CB and CR samples (at the same locations than for the first Row) will be the third etc.</para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_ICT</term>
                  <listitem>
                    <para xml:id="para_d8353953-87c9-46a0-b863-1b0d57001701">Irreversible Color Transformation:</para>
                    <para xml:id="para_0f411b52-056e-4808-81ce-9cccdbe8c34c">Pixel data represent a color image described by one luminance (Y) and two chrominance planes (CB and CR).</para>
                <para>This photometric interpretation may be used only when Samples per Pixel (0028,0002) has a value of 3. <emphasis role="bold"><emphasis role="underline"> Planar Configuration (0028,0006) may be 0 or 1. Shall only be used for pixel data in an Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                    <note>
                      <para><emphasis role="bold"><emphasis role="underline">This Photometric Interpretation is primarily used with JPEG 2000 compressed bit streams.</emphasis></emphasis></para>
                    </note>
                <para>Black is represented by Y equal to zero. The absence of color is represented by both CB and CR values equal to zero.</para>
                    <para xml:id="para_de17b0cc-7ca1-4257-9eb8-1a0ce448796f">Regardless of the value of Bits Allocated (0028,0100), the following equations convert between RGB and YCBCR Photometric Interpretation.</para>
                    <para xml:id="para_9025e6d0-ef50-44ef-bb3d-7093c981480c">Y = + .29900R + .58700G + .11400B</para>
                    <para xml:id="para_06f4bf7d-994f-4d10-bb28-6d9576f03214">CB= - .16875R - .33126G + .50000B</para>
                    <para xml:id="para_b793919c-c941-4cfc-a581-9fbfd3fa478f">CR= + .50000R - .41869G - .08131B</para>
                    <note>
                      <orderedlist numeration="arabic">
                        <listitem>
                          <para xml:id="para_66a41f1c-b53a-45ab-84fe-1f9b262f59da">The above is based on <xref linkend="biblio_ISOIEC15444-1"/> (JPEG 2000).</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_41a1f240-9a76-4645-8437-6543dd73dfe9">In a JPEG 2000 bit stream, DC level shifting (used if the untransformed components are unsigned) is applied before forward color transformation, and the transformed components may be signed (unlike in JPEG ISO/IEC 10918-1).</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_54f14642-5650-4ffa-92fe-89b51fab0ada">In JPEG 2000, spatial down-sampling of the chrominance components, if performed, is signaled in the JPEG 2000 bit stream.</para>
                        </listitem>
                      </orderedlist>
                    </note>
                  </listitem>
                </varlistentry>
                
                <varlistentry>
                  <term>YBR_RCT</term>
                  <listitem>
                    <para xml:id="para_94300c5e-92bc-4c38-b349-20d2af4bfe22">Reversible Color Transformation:</para>
                    <para xml:id="para_5dd8f16b-896b-429c-a166-e5576d5ce3e2">Pixel data represent a color image described by one luminance (Y) and two chrominance planes (CB and CR).</para>
                <para>This photometric interpretation may be used only when Samples per Pixel (0028,0002) has a value of 3.<emphasis role="bold"><emphasis role="underline"> Planar Configuration (0028,0006) may be 0 or 1. Shall only be used for pixel data in an Encapsulated (Compressed) format; see PS3.5 <xref linkend="sect_8.2" xrefstyle="select: label quotedtitle"/><!--<olink targetdoc="PS3.5" targetptr="sect_8.2" xrefstyle="select: docname label"/>-->.</emphasis></emphasis></para>
                    <note>
                      <para><emphasis role="bold"><emphasis role="underline">This Photometric Interpretation is primarily used with JPEG 2000 compressed bit streams.</emphasis></emphasis></para>
                    </note>
                    <para>Black is represented by Y equal to zero. The absence of color is represented by both CB and CR values equal to zero.</para>
                    <para xml:id="para_d7c5bcce-bf7f-444e-8638-0fa93b157419">Regardless of the value of Bits Allocated (0028,0100), the following equations convert between RGB and YBR_RCT Photometric Interpretation.</para>
                    <para xml:id="para_e5cf3a5e-063e-4d09-af30-f386b137745c">Y = ⌊(R + 2G +B) / 4⌋ (Note: ⌊…⌋ mean floor)</para>
                    <para xml:id="para_70070abb-39ae-44c0-a611-0fd6ae6d3684">CB= B - G</para>
                    <para xml:id="para_ceb05edc-b162-430b-bffb-afb269c29c6f">CR= R - G</para>
                    <para xml:id="para_f3deeb00-3ef5-4836-816e-efb41ace56fd">The following equations convert between YBR_RCT and RGB Photometric Interpretation.</para>
                    <para xml:id="para_50635fe3-6000-4472-890c-6dc302bd0b73">G = Y - ⌊ (CR+ CB) / 4⌋</para>
                    <para xml:id="para_7a3e22b7-6492-4ba6-b77a-b5cb46b0ea24">R = CR+ G</para>
                    <para xml:id="para_015901ac-7676-49a4-96ae-48b8b1e14d4b">B = CB+ G</para>
                    <note>
                      <orderedlist numeration="arabic">
                        <listitem>
                          <para xml:id="para_40112d2d-6cdb-40a7-8845-4601f70c7337">The above is based on <xref linkend="biblio_ISOIEC15444-1"/> (JPEG 2000).</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_65ff255f-3d89-495e-838b-5ced1c459ee5">In a JPEG 2000 bit stream, DC level shifting (used if the untransformed components are unsigned) is applied before forward color transformation, and the transformed components may be signed (unlike in JPEG ISO/IEC 10918-1).</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_48bd9bf7-9999-4713-a6e4-4bae29eaa018">This photometric interpretation is a reversible approximation to the YUV transformation used in PAL and SECAM.</para>
                        </listitem>
                      </orderedlist>
                    </note>
                  </listitem>
                </varlistentry>
              </variablelist>
            </section>
            <section label="C.7.6.3.1.3" status="6" xml:id="sect_C.7.6.3.1.3">
              <title>Planar Configuration</title>
              <para xml:id="para_80a38ce3-6b56-42e6-9941-a3db04694c7d">Planar Configuration (0028,0006) indicates whether the color pixel data are sent color-by-plane or color-by-pixel. This Attribute shall be present if Samples per Pixel (0028,0002) has a value greater than 1. It shall not be present otherwise.</para>
              <variablelist spacing="compact">
                <title>Enumerated Values:</title>
                <varlistentry>
                  <term>0</term>
                  <listitem>
                    <para xml:id="para_d8963355-e071-451b-8fb3-f8ec32dc8097">The sample values for the first pixel are followed by the sample values for the second pixel, etc. For RGB images, this means the order of the pixel values sent shall be R1, G1, B1, R2, G2, B2, …, etc.</para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>1</term>
                  <listitem>
                    <para xml:id="para_8eab4794-0f34-46d8-ba6d-118d39ee9f1a">Each color plane shall be sent contiguously. For RGB images, this means the order of the pixel values sent is R1, R2, R3, …, G1, G2, G3, …, B1, B2, B3, etc.</para>
                  </listitem>
                </varlistentry>
              </variablelist>
              <note>
                <para xml:id="para_a1dcc4bc-a0cd-40a6-914e-4f25a5f52e2a">Planar Configuration (0028,0006) is not meaningful when a compression Transfer Syntax is used that involves reorganization of sample components in the compressed bit stream. In such cases, since the Attribute is required to be sent, then an appropriate value to use may be specified in the description of the Transfer Syntax in PS3.5<!--<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>-->, though in all likelihood the value of the Attribute will be ignored by the receiving implementation.</para>
              </note>
            </section>
          </section>
        </section>
      
      <section label="C.8.5.6" status="4" xml:id="sect_C.8.5.6">
          <title>US Image Module</title>
          <table frame="box" label="C.8-18" rules="all" xml:id="table_C.8-18">
            <caption>US Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f075818a-7a0f-4350-a9f1-496627ef1065">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f87749c3-9467-49a5-83df-8fb3554ac1f8">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_70cff4e0-a976-4e60-bebf-bf5fc398447b">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_871fc0b3-f126-47c6-9481-28105508e42f">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_44f84f2d-56a2-40dc-8c25-562be70c537b">Samples Per Pixel</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e5b92728-3358-4444-aa8d-d4d554fab17a">(0028,0002)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d71196c6-cd85-4b39-afb2-11ad308fb718">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dec68064-b89d-4581-96f2-030cbc9313d7">Number of samples (planes) in this image.</para>
                  <para xml:id="para_7a263ef1-bf85-4e78-b1a7-7306b03d0e13">See <xref linkend="sect_C.8.5.6.1.12" xrefstyle="select: label"/> for specialization</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_18d37967-e6a2-4372-9f6a-76469c290999">Photometric Interpretation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5e58bcbe-050d-4434-9cb5-2001f3fedc69">(0028,0004)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_cef791e5-9c70-407b-beac-152ae0c58eae">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_134e5e43-a9d8-4f6c-a7b7-c55c6860a59e">Specifies the intended interpretation of the pixel data.</para>
                  <para xml:id="para_5460f010-c348-4db8-8af8-6f66c9974408">See <xref linkend="sect_C.8.5.6.1.2" xrefstyle="select: label"/> for specialization.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_660229ac-b803-4907-8c6e-1341b7948c0e">Bits Allocated</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_cf4d9ed9-56f3-4817-a1d0-3698625af4dc">(0028,0100)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_69d4b589-d19c-4fb0-a075-ae5f3bd30019">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_84188e3d-325c-4ae8-9af8-0045460ad584">Number of bits allocated for each pixel sample.</para>
                  <para xml:id="para_ea98fc9d-3b21-4893-bd16-c47a508297f2">See <xref linkend="sect_C.8.5.6.1.13" xrefstyle="select: label"/> for specialization.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fb3669fd-648c-43fb-8927-3ad41b4f379a">Bits Stored</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_af38865c-3277-43d5-800e-bb39172fc750">(0028,0101)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9ced4ab6-2936-4452-b317-14087f55304f">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_62d27d1d-b41d-4e56-965b-77b638689dd6">Number of bits stored for each pixel sample.</para>
                  <para xml:id="para_f492ff8a-85e5-48b1-81f3-bb5db2b90c4e">See <xref linkend="sect_C.8.5.6.1.14" xrefstyle="select: label"/> for specialization.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dff2a6a2-ffb8-43f9-b5af-33ef113895ed">High Bit</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_84f534cc-f5f2-4b60-a89e-094518fc8ca0">(0028,0102)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a3ae3002-d735-4972-a8db-60ccf2c1604d">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8b0c0ea1-2e6a-4544-bac2-5969ec841292">Most significant bit for pixel sample data.</para>
                  <para xml:id="para_015324aa-c49e-444c-8fc3-b0be9cc1bab5">See <xref linkend="sect_C.8.5.6.1.15" xrefstyle="select: label"/> for specialization.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_470391c6-d6ff-4edc-a2ab-0c014e01042d">Planar Configuration</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_395f22ce-3363-4efd-8126-5d1db2153859">(0028,0006)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_152a4e50-c39d-4307-a4bc-d84491668ef3">1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_47337884-efd0-4e42-8a32-f3523801ad21">Indicates whether the pixel data are sent color-by-plane or color-by-pixel.</para>
                  <para xml:id="para_07f90ebc-a91a-45fd-90e5-746f376f3958">Required if Samples per Pixel (0028,0002) has a value greater than 1.</para>
                  <para xml:id="para_7e871d72-9437-4f42-a35e-c9cda829d6f3">See <xref linkend="sect_C.8.5.6.1.16" xrefstyle="select: label"/> for specialization.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_108c03c6-7f11-4f68-add4-ce2780581128">Pixel Representation</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9acc483b-347b-4182-84b7-e48b4e4965e9">(0028,0103)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_15ced955-9480-4c5d-b17f-ca031c204090">1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fa2fbb2b-00d0-472e-a74a-49181552b331">Data representation of pixel samples.</para>
                  <para xml:id="para_35cae0ad-49b9-4531-a677-87c4a5f10bd3">See <xref linkend="sect_C.8.5.6.1.3" xrefstyle="select: label"/> for specialization.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5310cbe6-df7e-4189-9f83-e8f5aaafa2b8">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0b17a744-41aa-4553-974c-67a9a479229c">...</para>
                </td>
              </tr>
            </tbody>
          </table>
          <section label="C.8.5.6.1" status="5" xml:id="sect_C.8.5.6.1">
            <title>US Image Attribute Descriptions</title>
            <section label="C.8.5.6.1.2" status="6" xml:id="sect_C.8.5.6.1.2">
              <title>Photometric Interpretation</title>
              <variablelist spacing="compact">
                <title>Defined Terms:</title>
                <varlistentry>
                  <term>MONOCHROME2</term>
                  <listitem>
                    <para xml:id="para_00efda0f-1bda-41bb-96ec-e8cb0af1e13d"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>PALETTE COLOR</term>
                  <listitem>
                    <para xml:id="para_ebbc70b6-bdf1-4bf1-a2fa-dd6741656ce1"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>RGB</term>
                  <listitem>
                    <para xml:id="para_87ac70c1-80fb-47ac-b023-65035827d702"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term><emphasis role="bold"><emphasis role="strikethrough">ARGB</emphasis></emphasis></term>
                  <listitem>
                    <para xml:id="para_5b00a64a-9a47-4161-a010-715915f47683"><emphasis role="bold"><emphasis role="strikethrough">(retired)</emphasis></emphasis></para>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_FULL</term>
                  <listitem>
                    <para xml:id="para_c14a4455-521b-405c-b5f4-fc4e705f1618"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_FULL_422</term>
                  <listitem>
                    <para xml:id="para_0cdd1ffa-8ce7-4d5c-8e80-abae04efdce6"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term><emphasis role="bold"><emphasis role="strikethrough">YBR_PARTIAL_422</emphasis></emphasis></term>
                  <listitem>
                    <para xml:id="para_d01d70e4-c600-44b4-9d55-65103a0adf93"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_RCT</term>
                  <listitem>
                    <para xml:id="para_0171d040-f7f8-47e1-bf8a-e8fbf24dd05c"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_ICT</term>
                  <listitem>
                    <para xml:id="para_c1f22539-210e-466d-bba8-2124f6c63258"/>
                  </listitem>
                </varlistentry>
                <varlistentry>
                  <term>YBR_PARTIAL_420</term>
                  <listitem>
                    <para xml:id="para_ebf563fb-8f05-4158-af8f-b4aa9351f24c"/>
                  </listitem>
                </varlistentry>
              </variablelist>
              <note>
                <para xml:id="para_780d750d-4c25-4927-acd6-941ec1bff4a0"><emphasis role="bold"><emphasis role="strikethrough">It is recommended that future implementations should not use</emphasis><emphasis role="underline">The</emphasis></emphasis> ARGB <emphasis role="bold"><emphasis role="underline">and YBR_PARTIAL_422 </emphasis></emphasis>photometric interpretation<emphasis role="bold"><emphasis role="underline">s were previously defined but are retired</emphasis></emphasis>.</para>
              </note>
          <para><emphasis role="bold"><emphasis role="underline">When Samples per Pixel (0028,0002) is greater than 1, Photometric Interpretation (0028,0004) shall be RGB for uncompressed or lossless compressed Transfer Syntaxes that do not involve color space transformations, YBR_ICT for irreversible JPEG 2000 Transfer Syntaxes, YBR_RCT for reversible JPEG 2000 Transfer Syntaxes, YBR_PARTIAL_420 for MPEG2, MPEG-4 AVC/H.264 and HEVC/H.265 Transfer Syntaxes, YBR_FULL_422 for JPEG lossy compressed Transfer Syntaxes and YBR_FULL for RLE Transfer Syntaxes.</emphasis></emphasis></para>
              <para xml:id="para_51089819-41c5-4e4c-824e-655bb9d3078b">See PS3.5<!--<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>--> for <emphasis role="bold"><emphasis role="underline">additional</emphasis></emphasis>restrictions imposed by compressed Transfer Syntaxes.</para>
            </section>
            <section label="C.8.5.6.1.3" status="6" xml:id="sect_C.8.5.6.1.3">
              <title>Pixel Representation</title>
              <para xml:id="para_f0d8a321-cb13-4274-9e4e-fbff6211a16b">For US Images, Pixel Representation (0028,0103) is specified to use the following.</para>
              <variablelist spacing="compact">
                <title>Enumerated Values:</title>
                <varlistentry>
                  <term>0000H</term>
                  <listitem>
                    <para xml:id="para_e55cf9dc-51dd-4841-a27d-0c70b0e77ae8">unsigned integer</para>
                  </listitem>
                </varlistentry>
              </variablelist>
            </section>
            <section label="C.8.5.6.1.12" status="6" xml:id="sect_C.8.5.6.1.12">
              <title>Samples Per Pixel</title>
              <para xml:id="para_7746fb3c-4af4-43c8-a3ee-b3ffe22d3f83">For US Images, Samples Per Pixel (0028,0002) is specified to use the following values for specific Photometric Interpretations:</para>
              <table frame="box" label="C.8-19" rules="all" xml:id="table_C.8-19">
                <caption>US Samples Per Pixel</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_4beffd63-f736-4b19-a1f0-9d933b0c6a4e">
                        <emphasis role="bold">Photometric Interpretation</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8ec9adf4-74d4-46e1-9bbc-ccbb8903976e">
                        <emphasis role="bold">Samples Per Pixel Value</emphasis>
                      </para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e36dd55f-a786-4d40-a058-474dfbe73e0e">MONOCHROME2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_f62cbb94-cbe1-46e6-81ee-563d876abda5">1</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e4320faf-fe8c-4442-9945-d963348ec3c6">RGB</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_854cf4ac-81ae-4e4f-aae8-099f4a7c1d9e">3</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_f3607900-2e28-4d18-b770-33e25bf616af">YBR_FULL</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_193646af-a884-43c3-886c-eb5ac9cb04be">3</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_aa41c8b5-b2ab-4a14-8ac9-34d364b0270a">YBR_FULL_422</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_89956a79-7b0d-4599-b440-5d4fec666094">3</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e1c786f4-b5f0-4a24-b245-f6233ee41c65"><emphasis role="bold"><emphasis role="strikethrough">YBR_PARTIAL_422</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ff19d0fc-236f-476a-b043-8a9fce017556"><emphasis role="bold"><emphasis role="strikethrough">3</emphasis></emphasis></para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_fd08639f-6891-4a37-a6d7-e517a5307638">YBR_RCT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_87e2ae1a-3e55-4eae-8392-25b73225fd10">3</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_be763909-f570-4251-a661-bb2ac83d9f6b">YBR_ICT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_addbad1d-3994-4607-894e-bfb7177c9351">3</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_538f9577-5d30-46f2-a937-ffa4a0f536c1">YBR_PARTIAL_420</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_850e6afc-b2b9-4f79-9420-5557dc819e23">3</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_5b86019c-71fa-427f-9c57-8ba47377001e">PALETTE COLOR</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3acc0024-95b4-4b53-97c8-ba8abdc69eb5">1</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
            <section label="C.8.5.6.1.13" status="6" xml:id="sect_C.8.5.6.1.13">
              <title>Bits Allocated</title>
              <para xml:id="para_393fb6de-da19-4a7e-887b-955e34daf16c">For US Images, Bits Allocated (0028,0100) is specified to use the following values for specific Photometric Interpretations:</para>
              <table frame="box" label="C.8-20" rules="all" xml:id="table_C.8-20">
                <caption>US Bits Allocated</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3d7c7022-df9b-4bf0-9d4e-1d986f3469b5">
                        <emphasis role="bold">Photometric Interpretation</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_cfd1e29e-2302-4245-b28e-1156a81c29d5">
                        <emphasis role="bold">Bits Allocated Value</emphasis>
                      </para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_1065de20-f5ef-430f-811f-77b87ab439ec">MONOCHROME2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a2df5777-56a3-4c97-bd51-c1af56fcc40f">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_c6f3ed45-04ee-4444-a2cc-437c891b5188">RGB</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_8306f461-2ecb-49e1-8043-ecfa52c21d3b">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_91630d80-5647-413f-b01d-9fe0664ee99f">YBR_FULL</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_17d72975-256a-4e4c-8491-078708de8fe6">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_2db8ba2b-b9be-4a0c-b20f-0a99c1f84894">YBR_FULL_422</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e55a5072-cc2f-4858-866b-79bcb725ee44">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_96813cad-4e37-407f-94c3-8abaabfe6a10"><emphasis role="bold"><emphasis role="strikethrough">YBR_PARTIAL_422</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_bdfbd795-7afb-4fc6-b66d-83c352ce9e7d"><emphasis role="bold"><emphasis role="strikethrough">8</emphasis></emphasis></para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_9c11d72a-5301-4845-80c7-c755b35d5065">YBR_RCT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_792e03a0-47ff-4d37-9351-84961c6eadc0">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a6ffee32-2b16-4d27-b931-4ccb48c18e40">YBR_ICT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_0763cb5b-36f7-444e-b70a-ae8e4a43fca4">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_4b90744f-509f-4482-acfb-d35c83a21e2d">YBR_PARTIAL_420</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_63bec542-dbb5-40d9-bcde-8dd4e52576e2">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_80a6cddb-e837-49b9-bc7e-5a5b40c9acaf">PALETTE COLOR</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_314cde87-f239-4c96-802b-df893d8a80e6">8 - 8 bit palette, or</para>
                      <para xml:id="para_e2d42d9c-f640-4ffe-9763-97c3c668b9ea">16 - 16 bit palette</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
            <section label="C.8.5.6.1.14" status="6" xml:id="sect_C.8.5.6.1.14">
              <title>Bits Stored</title>
              <para xml:id="para_c259efe6-9f3d-4c36-8249-7f5ae4c566cb">For US Images, Bits Stored (0028,0101) shall be equal to Bits Allocated (0028,0100).</para>
            </section>
            <section label="C.8.5.6.1.15" status="6" xml:id="sect_C.8.5.6.1.15">
              <title>High Bit</title>
              <para xml:id="para_5d686ba3-afac-490e-bd97-4dfb9e719376">For US Images, High Bit (0028,0102) shall be one less than Bits Stored (0028,0101).</para>
            </section>
            <section label="C.8.5.6.1.16" status="6" xml:id="sect_C.8.5.6.1.16">
              <title>Planar Configuration</title>
              <para xml:id="para_e6482045-1b8f-4a38-83d3-905165a7ebd7">For US Images, Planar Configuration (0028,0006) is specified to use the following values for specific Photometric Interpretations:</para>
              <table frame="box" label="C.8-23" rules="all" xml:id="table_C.8-23">
                <caption>US Planar Configuration</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_c3991e48-e96f-446a-8740-189a5518d5be">
                        <emphasis role="bold">Photometric Interpretation</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_bd64ab14-ad84-4aec-829b-394eeb45c01a">
                        <emphasis role="bold">Planar Configuration Value</emphasis>
                      </para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d37ef974-6f6a-4a9b-99b2-e4bf909bc799">RGB</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ee37d0ed-e83f-44bc-b968-439091cb7c96">0 - color-by-pixel, or</para>
                      <para xml:id="para_023163f6-ddd5-4b40-869e-57bb3ae0d5e4">1 - color-by-plane</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ee93d8e8-b1bb-4dbe-9202-720a1d32feb4">YBR_FULL</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b589f6c2-7397-47f2-8f9e-0cc0f5c22ef2">1</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_f87afa49-3aa7-4ece-bceb-b9c2a9fd8917">YBR_FULL_422</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_c6c5809c-cbfb-43c0-b7d7-3a53fee8df27">0</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b791b571-ff82-4c01-bff9-1e9f698ba6b0"><emphasis role="bold"><emphasis role="strikethrough">YBR_PARTIAL_422</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_80dee116-3edd-4ea4-8957-97fca768032b"><emphasis role="bold"><emphasis role="strikethrough">0</emphasis></emphasis></para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_fe7ec753-0580-4afb-9402-c30bccc1f068">YBR_RCT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_45c9674d-f0a8-4cac-a5f6-c7fd7624a78e">0</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_6e06d97e-e3a1-4929-bff4-5f71a8a0f462">YBR_ICT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a572a1a1-010f-4c3f-b794-ab75e5fb384e">0</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_5243007b-3f35-4f15-bf2e-67642695f58c">YBR_PARTIAL_420</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_33f42dc2-60f0-46b0-a146-dc663233ad24">0</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
          </section>
        </section>
      
      <section label="C.8.12.1.1" status="5" xml:id="sect_C.8.12.1.1">
        <title>VL Image Module Attribute Descriptions</title>
        <section label="C.8.12.1.1.1" status="6" xml:id="sect_C.8.12.1.1.1">
          <title>Photometric Interpretation</title>
          <variablelist spacing="compact">
            <title>Enumerated Values:</title>
            <varlistentry>
              <term>MONOCHROME2</term>
              <listitem>
                <para xml:id="para_8126bbbf-b798-40f9-a2bc-2a3a58438ead"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>RGB</term>
              <listitem>
                <para xml:id="para_f15970a5-3508-4a01-80f2-e26861e84a08"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_FULL_422</term>
              <listitem>
                <para xml:id="para_62c09ac1-8709-4597-9d50-c89820ce6120"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_PARTIAL_420</term>
              <listitem>
                <para xml:id="para_627278cb-0a73-4c70-8063-31b679605c62"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_RCT</term>
              <listitem>
                <para xml:id="para_7ae79dbc-e29d-4cd5-9f40-c0cccf5962c4"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_ICT</term>
              <listitem>
                <para xml:id="para_87ddaa53-be94-4855-855f-4efd06a0aa1b"/>
              </listitem>
            </varlistentry>
          </variablelist>
          <para><emphasis role="bold"><emphasis role="underline">Photometric Interpretation (0028,0004) shall be RGB for uncompressed or lossless compressed Transfer Syntaxes that do not involve color space transformations, YBR_ICT for irreversible JPEG 2000 Transfer Syntaxes, YBR_RCT for reversible JPEG 2000 Transfer Syntaxes, YBR_PARTIAL_420 for MPEG2, MPEG-4 AVC/H.264 and HEVC/H.265 Transfer Syntaxes and YBR_FULL_422 for JPEG lossy compressed Transfer Syntaxes.</emphasis></emphasis></para>
          <note>
            <orderedlist>
              <listitem>
                <para><emphasis role="bold"><emphasis role="underline">The YBR_FULL Photometric Interpretation (0028,0004), such as might be used with the RLE Transfer Syntax, is not permitted.</emphasis></emphasis></para>
              </listitem>
              <listitem>
                <para xml:id="para_11828793-23a8-4d99-a6d0-d66c0f5f341a">There is no formal color space defined, hence "false" color applications that encode near-visible light images may be encoded, for example, as RGB.</para>
              </listitem>
            </orderedlist>
           </note>
        </section>
        <section label="C.8.12.1.1.2" status="6" xml:id="sect_C.8.12.1.1.2">
          <title>Bits Allocated, Bits Stored, and High Bit</title>
          <variablelist spacing="compact">
            <title>Enumerated Values of Bits Allocated (0028,0100):</title>
            <varlistentry>
              <term>8</term>
              <listitem>
                <para xml:id="para_0626ceb2-6590-4207-a1cf-9d1c2ce2527f"/>
              </listitem>
            </varlistentry>
          </variablelist>
          <variablelist spacing="compact">
            <title>Enumerated Values of Bits Stored (0028,0101):</title>
            <varlistentry>
              <term>8</term>
              <listitem>
                <para xml:id="para_5fab8ceb-ca07-40fb-a07d-2c2f8690e8d9"/>
              </listitem>
            </varlistentry>
          </variablelist>
          <variablelist spacing="compact">
            <title>Enumerated Values of High Bit (0028,0102):</title>
            <varlistentry>
              <term>7</term>
              <listitem>
                <para xml:id="para_7de9cc7c-8ee3-4858-a7b0-53d49752a56c"/>
              </listitem>
            </varlistentry>
          </variablelist>
        </section>
        <section label="C.8.12.1.1.3" status="6" xml:id="sect_C.8.12.1.1.3">
          <title>Pixel Representation</title>
          <variablelist spacing="compact">
            <title>Enumerated Values:</title>
            <varlistentry>
              <term>0</term>
              <listitem>
                <para xml:id="para_cf6d78b0-2274-4837-a728-2def45eb22b2"/>
              </listitem>
            </varlistentry>
          </variablelist>
        </section>
        <section label="C.8.12.1.1.4" status="6" xml:id="sect_C.8.12.1.1.4">
          <title>Samples Per Pixel</title>
          <variablelist spacing="compact">
            <title>Enumerated Values when Photometric Interpretation (0028,0004) is MONOCHROME2:</title>
            <varlistentry>
              <term>1</term>
              <listitem>
                <para xml:id="para_88a9b571-a8fd-490d-9a5c-2da86dcef515"/>
              </listitem>
            </varlistentry>
          </variablelist>
          <variablelist spacing="compact">
            <title>Enumerated Values when Photometric Interpretation (0028,0004) is RGB or YBR_FULL_422 or YBR_PARTIAL_420 or YBR_RCT or YBR_ICT:</title>
            <varlistentry>
              <term>3</term>
              <listitem>
                <para xml:id="para_df99d4a2-fc19-4736-84fe-5e17673ce3b3"/>
              </listitem>
            </varlistentry>
          </variablelist>
        </section>
        <section label="C.8.12.1.1.5" status="6" xml:id="sect_C.8.12.1.1.5">
          <title>Planar Configuration</title>
          <para xml:id="para_f44a6e32-ca7a-4943-9521-f7481a2d4253">This value shall be present if Samples per Pixel (0028,0002) has a value greater than 1.</para>
          <variablelist spacing="compact">
            <title>Enumerated Values:</title>
            <varlistentry>
              <term>0</term>
              <listitem>
                <para xml:id="para_a64e32c4-e71b-4fff-951b-8c36eec5cef5"/>
              </listitem>
            </varlistentry>
          </variablelist>
          <note>
            <para><emphasis role="bold"><emphasis role="underline">The prohibition of a value of 1 for Planar Configuration (0028,0006) prevents the use of the RLE Transfer Syntax.</emphasis></emphasis></para>
          </note>
        </section>
      </section>
      
      <section label="C.8.12.4.1.5" status="6" xml:id="sect_C.8.12.4.1.5">
        <title>Photometric Interpretation and Samples Per Pixel</title>
        <para xml:id="para_56895ea3-86e6-4c57-b92c-284a40a61c76">See <xref linkend="sect_C.7.6.3.1.2" xrefstyle="select: label"/>.</para>
        <variablelist spacing="compact">
          <title>Enumerated Values for Photometric Interpretation (0028,0004):</title>
          <varlistentry>
            <term>MONOCHROME2</term>
            <listitem>
              <para xml:id="para_7a7f7d7f-3c23-4ec2-990e-a3fad60a5863"/>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>RGB</term>
            <listitem>
              <para xml:id="para_2ba62380-d35e-429a-9f39-f2af10ca8dfb"/>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>YBR_FULL_422</term>
            <listitem>
              <para xml:id="para_ac059e64-c865-431e-8215-0e8ab5171439"/>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>YBR_ICT</term>
            <listitem>
              <para xml:id="para_69b8ed8a-f1a0-4923-b6b9-f3a043a27316"/>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>YBR_RCT</term>
            <listitem>
              <para xml:id="para_9803d2c1-5de8-4fcf-8d0d-afa795455453"/>
            </listitem>
          </varlistentry>
        </variablelist>
        <para xml:id="para_1e7b1466-5988-40c6-aef9-eb8724c622f3">The value shall be appropriate to the compression Transfer Syntax used, if any, and shall be MONOCHROME2 or RGB for uncompressed or lossless compressed Transfer Syntaxes that do not involve color space transformations, YBR_ICT for irreversible JPEG 2000 Transfer Syntaxes, YBR_RCT for reversible JPEG 2000 Transfer Syntaxes, and YBR_FULL_422 for <emphasis role="bold"><emphasis role="strikethrough">other</emphasis><emphasis role="underline">JPEG</emphasis></emphasis> lossy compressed Transfer Syntaxes.</para>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_e090e831-1242-4f28-af43-447e04d0ae88">Future lossless and lossy Transfer Syntaxes may lead to the need for new definitions and choices for Photometric Interpretation. The Enumerated Values may therefore be extended with additional Photometric Interpretation values directly associated with new Transfer Syntaxes that are negotiated, and hence do not render existing implementations non-conformant.</para>
            </listitem>
            <listitem>
              <para><emphasis role="bold"><emphasis role="underline">Motion compression Transfer Syntaxes are not expected to be used for Whole Slide Imaging, so the use of YBR_PARTIAL_420 for MPEG2, MPEG-4 AVC/H.264 and HEVC/H.265 Transfer Syntaxes is not permitted.</emphasis></emphasis></para>
            </listitem>
            <listitem>
              <para><emphasis role="bold"><emphasis role="underline">The prohibition of a value of 1 for Planar Configuration (0028,0006) prevents the use of the RLE Transfer Syntax.</emphasis></emphasis></para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_908e5dfc-1bcb-4e36-9579-4321b584d7a9">Multi-spectral images may be encoded as a single wavelength band (color) in each frame using MONOCHROME2, or with up to three bands in each frame using one of the color Photometric Interpretations.</para>
        <variablelist spacing="compact">
          <title>Enumerated Values for Samples per Pixel (0028,0002) when Photometric Interpretation (0028,0004) is MONOCHROME2:</title>
          <varlistentry>
            <term>1</term>
            <listitem>
              <para xml:id="para_4eef8f1d-9f61-4bb9-8d57-b5e987cdd413"/>
            </listitem>
          </varlistentry>
        </variablelist>
        <variablelist spacing="compact">
          <title>Enumerated Values for Samples per Pixel (0028,0002) when Photometric Interpretation (0028,0004) is not MONOCHROME2:</title>
          <varlistentry>
            <term>3</term>
            <listitem>
              <para xml:id="para_502ba4c0-30a4-4341-833f-f73c2cde930d"/>
            </listitem>
          </varlistentry>
        </variablelist>
      </section>
      
      <section label="C.8.13.1.1" status="5" xml:id="sect_C.8.13.1.1">
            <title>Enhanced MR Image Module Attribute Description</title>
            <section label="C.8.13.1.1.2" status="6" xml:id="sect_C.8.13.1.1.2">
              <title>Photometric Interpretation, Pixel Representation, Samples Per Pixel, Planar Configuration, Bits Allocated and Bits Stored</title>
              <para xml:id="para_310c3276-5781-4152-8f3a-749b8720d403">
<xref linkend="table_C.8-82" xrefstyle="select: label"/> specifies the Enumerated Values and allowed combinations of Samples per Pixel (0028,0002), Planar Configuration (0028,0006), Pixel Representation (0028,0103), Bits Allocated (0028,0100) and Bits Stored (0028,0101) for each allowable Photometric Interpretation allowed by the IOD that invokes this Module.</para>
              <table frame="box" label="C.8-82" rules="all" xml:id="table_C.8-82">
                <caption>Allowed Combinations of Attribute Values for Photometric Interpretation, Samples Per Pixel, Planar Configuration, Pixel Representation, Bits Allocated and Bits Stored</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_6d1a2b1f-c531-4a6d-ba3f-8428aeb69abf">
                        <emphasis role="bold">Photometric Interpretation</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_cbb44440-f319-4fb6-b829-88fdee66f1f4">
                        <emphasis role="bold">Samples per Pixel</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_ff293a76-74ef-444b-ac3c-5d27f949aa63">
                        <emphasis role="bold">Planar Configuration</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8a7ca290-f69c-4bae-a953-7548c7afd519">
                        <emphasis role="bold">Pixel Representation</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_5c1873ac-bf44-4966-a6c1-06a10b0d2fdb">
                        <emphasis role="bold">Bits Allocated</emphasis>
                      </para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_62d7bcb4-0178-455d-9f3b-b05735581727">
                        <emphasis role="bold">Bits Stored</emphasis>
                      </para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a7295525-b046-42ad-ab9d-169231e42106">MONOCHROME2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e0546659-7861-46a6-bc62-287e7b7ec3cf">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_6582e79c-bd33-422c-b711-f6046ccc7521">-</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_96c47867-9854-4162-9f37-d1d63deac667">0 or 1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_c2b058eb-03ce-4f68-81ae-22cda891fd32">8</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_8ccb545f-e66a-451b-a8d8-8c02b9bfa4de">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_746d49c0-1620-48fc-af0a-eeb8710c2920">MONOCHROME2</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_bcceae04-1c02-43ea-a79d-bdcb52f87b0e">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_6b0ab1dc-6bae-45e7-aa5c-8a17d0159b1e">-</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_840baad1-8476-48b3-8fc2-8d3bb20d5d1e">0 or 1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_bd9fbf3f-02e4-454c-886b-2595f279f401">16</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_64ea14f0-9d75-4530-afc7-4c14f5206d90">12, 16</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_50f53a87-f973-467f-b901-ee45954cef2e">RGB</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_24bc6149-0580-4d55-8998-2a528bee731a">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b3ba7bbf-75a3-4fe5-b166-28e6cbf80592">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a9a9160b-53fd-4cb6-ab36-b19fe8b72b49">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_14f2fece-a67b-4be7-a964-9cfb0a28a584">8</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_596cdf90-39de-427e-9826-6a469922b25c">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_af29b421-6532-4928-824a-00cf4b410645">YBR_ICT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_20eaf89d-b722-451a-82d0-b1bfa74f6fa3">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_930da09e-c20e-443e-97ef-39ee8bcb27df">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_23d4ee0e-017a-4f0a-b64e-b02344ad3081">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_8281d337-6a45-4755-b68e-d8ec9283432e">8</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e96fd3f0-049d-447b-8f5c-1eec0ee2a4db">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_7afbbc9b-44de-4fbe-92c9-31eb924e2fee">YBR_RCT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_f04afa0a-ddb4-46eb-a703-26909f01f8c9">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d233bb7e-7890-416e-9158-50374fa76672">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_57df8cc8-3d25-444a-a41b-37d7c2679cf1">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_46909d8d-7abb-4996-b759-15c5719a3720">8</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_49a391bb-77b3-492f-88ed-040fc057efac">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_dbb92341-3a4e-43d9-bc5b-acf910f1a0e0">YBR_PARTIAL_420</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_06590220-af1c-4ca0-86de-5fe1c877c8a4">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_399473c8-72c5-4e89-84cf-966053a8ce43">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3ca9100e-f5b4-4940-91ef-266ee4841fb9">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3f85dc6c-4442-4a11-b46f-f74e9ac8d224">8</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_5043d311-aa81-4f7d-87e0-5500a2fae1a0">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ac644e47-cbf4-4292-8ca1-24389a428b4c">YBR_FULL_422</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_82ade6d5-178f-4cba-aa12-d9effb1dda7e">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a3f88e98-1967-43c8-83e7-6dca889e0b2d">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b1f56789-583d-47c9-a489-315c8bb88ecf">0</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d0513ca4-abbe-4afd-9962-c222114dce65">8</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d3cf021f-7188-482d-9295-ef3a5c6cd295">8</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">YBR_FULL</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis></para>
                  </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis></para>
                  </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis></para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </section>
          </section>
      
      <section label="C.8.17.2.1" status="5" xml:id="sect_C.8.17.2.1">
        <title>Ophthalmic Photography Image Module Attribute Descriptions</title>
        <section label="C.8.17.2.1.2" status="6" xml:id="sect_C.8.17.2.1.2">
          <title>Samples Per Pixel and Samples Per Pixel Used</title>
          <para xml:id="para_9406cdab-74ae-4b44-a18a-280833b6d043">Samples per Pixel (0028,0002) shall be 1 or 3.</para>
          <para xml:id="para_e0ddbb0d-05f1-4ed7-843d-7565ac835541">Cameras producing 2-color images are required to use a value of 3 for Samples per Pixel (0028,0002) and a value of 2 for Samples per Pixel Used (0028,0003). For 2-color images with a RGB Photometric Interpretation, the R and G channel shall be used and the B channel shall have all values set to zero.</para>
          <note>
            <para xml:id="para_0d6d4a11-15ff-4368-ae79-40afeaf66715">In the case of Photometric Interpretations typically used for compression such as YBR_FULL_422, the encoding will be as if the RGB values were transformed to YCbCr.</para>
          </note>
        </section>
        <section label="C.8.17.2.1.3" status="6" xml:id="sect_C.8.17.2.1.3">
          <title>Photometric Interpretation</title>
          <para xml:id="para_ccc881c7-c7c8-41d8-8300-5ea9dd72b528">Specifies the intended interpretation of the pixel data.</para>
          <variablelist spacing="compact">
            <title>Enumerated Values:</title>
            <varlistentry>
              <term>MONOCHROME2</term>
              <listitem>
                <para xml:id="para_f3498835-c567-4b69-9f64-066959bee4af"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>RGB</term>
              <listitem>
                <para xml:id="para_3e2738a9-5964-4c19-9025-2b098d7ef4af"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_FULL_422</term>
              <listitem>
                <para xml:id="para_b3a08a4a-159c-485d-87aa-fd3186378efb"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_PARTIAL_420</term>
              <listitem>
                <para xml:id="para_d7aced0a-1b82-428f-a869-a67779035124"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_ICT</term>
              <listitem>
                <para xml:id="para_14535b46-9980-4ee7-a049-7acb3ca6593e"/>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>YBR_RCT</term>
              <listitem>
                <para xml:id="para_b0c814b2-b119-4c3d-a78b-286829f5458a"/>
              </listitem>
            </varlistentry>
          </variablelist>
          <para><emphasis role="bold"><emphasis role="underline">When Samples per Pixel (0028,0002) is greater than 1, Photometric Interpretation (0028,0004) shall be RGB for uncompressed or lossless compressed Transfer Syntaxes that do not involve color space transformations, YBR_ICT for irreversible JPEG 2000 Transfer Syntaxes, YBR_RCT for reversible JPEG 2000 Transfer Syntaxes, YBR_PARTIAL_420 for MPEG2, MPEG-4 AVC/H.264 and HEVC/H.265 Transfer Syntaxes and YBR_FULL_422 for JPEG lossy compressed Transfer Syntaxes.</emphasis></emphasis></para>
          <note>
            <para><emphasis role="bold"><emphasis role="underline">The prohibition of a value of 1 for Planar Configuration (0028,0006) prevents the use of the RLE Transfer Syntax.</emphasis></emphasis></para>
          </note>
        </section>
      </section>
      
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td><para><emphasis role="italic">Amend DICOM PS 3.5:</emphasis></para></td>
          </tr>
        </informaltable>
      </section>
      
        <section label="2" status="1" xml:id="chapter_2">
    <title>Normative References</title>
    <para xml:id="para_d61da3a1-0191-4b41-b1c2-873f49cb9d53">The following standards contain provisions that, through references in this text, constitute provisions of this standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this standard are encouraged to investigate the possibilities of applying the most recent editions of the standards indicated below.</para>
    <bibliography>
      <title/>
        <biblioentry xml:id="biblio_ECMA_TR-098">
          <abbrev>ECMA TR-098</abbrev>
          <author>
            <orgname>European Computer Manufacturers Association</orgname>
           </author>
           <date>2009</date>
            <title>JPEG File Interchange Format (JFIF)</title>
           <bibliosource>
            <link xl:href="http://www.ecma-international.org/publications/techreports/E-TR-098.htm"/>
           </bibliosource>
        </biblioentry>
        <biblioentry xml:id="biblio_ETSI_TS_102_366">
          <abbrev>ETSI TS 102 366</abbrev>
          <author>
            <orgname>ETSI</orgname>
          </author>
          <date>Feb. 2005</date>
          <title>Audio Compression (AC-3, Enhanced AC-3) Standard</title>
        </biblioentry>
       <biblioentry xml:id="biblio_ISOIEC10918-1">
          <abbrev>ISO/IEC 10918-1</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1994</date>
          <title>JPEG Standard for digital compression and encoding of continuous-tone still images. Part 1 - Requirements and implementation guidelines</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC10918-2">
          <abbrev>ISO/IEC 10918-2</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1995</date>
          <title>JPEG Standard for digital compression and encoding of continuous-tone still images. Part 2 - Testing</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC10918-5">
          <abbrev>ISO/IEC 10918-5</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2013</date>
          <title>JPEG Standard for digital compression and encoding of continuous-tone still images. Part 5 - JPEG File Interchange Format (JFIF)</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC11172-3">
          <abbrev>ISO/IEC 11172-3</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1993</date>
          <title>Information technology -- Coding of moving pictures and associated audio for digital storage media at up to about 1,5 Mbit/s -- Part 3: Audio</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC13818-1">
          <abbrev>ISO/IEC 13818-1</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2000</date>
          <title>Information technology -- Generic coding of moving pictures and associated audio information - Part 1: Systems</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC13818-2">
          <abbrev>ISO/IEC 13818-2</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2000</date>
          <title>Information technology -- Generic coding of moving pictures and associated audio information - Part 2: Video</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC13818-3">
          <abbrev>ISO/IEC 13818-3</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1998</date>
          <title>Information technology -- Generic coding of moving pictures and associated audio information - Part 3: Audio</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC13818-4">
          <abbrev>ISO/IEC 13818-4</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1998</date>
          <title>Information technology -- Generic coding of moving pictures and associated audio information - Part 4: Conformance testing</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC13818-7">
          <abbrev>ISO/IEC 13818-7</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1997</date>
          <title>Information technology -- Generic coding of moving pictures and associated audio information - Part 7: Advanced Audio Coding (AAC)</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC14495-1">
          <abbrev>ISO/IEC 14495-1</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>1997</date>
          <title>Lossless and near-lossless coding of continuous tone still images (JPEG-LS)</title>
        </biblioentry>      
        <biblioentry xml:id="biblio_ISOIEC14496-3">
          <abbrev>ISO/IEC 14496-3</abbrev>
          <author>
           <orgname>ISO/IEC</orgname>
          </author>
          <date>2009</date>
          <title>Information technology - Coding of audio-visual objects - Part 3: Audio</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC14496-10">
          <abbrev>ISO/IEC 14496-10</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2009</date>
          <title>Information technology - Coding of audio-visual objects - Part 10: Advanced Video Coding</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC14496-12">
          <abbrev>ISO/IEC 14496-12</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2003</date>
          <title>Information technology - Coding of audio-visual objects - Part 12: ISO base media file format</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC14496-14">
          <abbrev>ISO/IEC 14496-14</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2003</date>
          <title>Information technology - Coding of audio-visual objects - Part 14: MP4 file format</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC15444-1">
          <abbrev>ISO/IEC 15444-1</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2004</date>
          <title>JPEG 2000 Image Coding System</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC15444-2">
          <abbrev>ISO/IEC 15444-2</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2004</date>
          <title>JPEG 2000 Image Coding System: Extensions</title>
        </biblioentry>
        <biblioentry xml:id="biblio_ISOIEC15444-9">
          <abbrev>ISO/IEC 15444-9</abbrev>
          <author>
            <orgname>ISO/IEC</orgname>
          </author>
          <date>2005</date>
          <title>Information technology - JPEG 2000 image coding system: Interactivity tools, APIs and protocols</title>
        </biblioentry>
      <biblioentry xml:id="biblio_ISOIEC23008-2">
        <abbrev>ISO/IEC 23008-2</abbrev>
        <author>
          <orgname>ISO/IEC</orgname>
        </author>
        <title>Information technology - High efficiency coding and media delivery in heterogeneous environments - Part 2: High efficiency video coding</title>
        <bibliosource>
          <link xl:href="http://​www.iso.org/​iso/​iso_catalogue/​catalogue_tc/​catalogue_detail.htm?​csnumber=67660"/>
        </bibliosource>
      </biblioentry>
      
    </bibliography>
  </section>

      
      <section label="8.2" status="2" xml:id="sect_8.2">
      <title>Native or Encapsulated Format Encoding</title>
      <para xml:id="para_f9b54410-31ad-4044-9be0-651dd86c84b3">Pixel data conveyed in the Pixel Data (7FE0,0010) may be sent either in a Native (uncompressed) Format or in an Encapsulated Format (e.g., compressed) defined outside the DICOM standard.</para>
        
        
        <para>
        <emphasis role="bold">
          <emphasis role="underline">If Pixel Data (7FE0,0010) is sent in a Native Format, then the Photometric Interpretation (0028,0004) shall be other than:</emphasis>
        </emphasis>
      </para>
        <itemizedlist>
          <listitem>
            <para>
            <emphasis role="bold">
              <emphasis role="underline">YBR_RCT</emphasis>
            </emphasis>
          </para>
          </listitem>
          <listitem>
            <para>
            <emphasis role="bold">
              <emphasis role="underline">YBR_ICT</emphasis>
            </emphasis>
          </para>
          </listitem>
          <listitem>
            <para>
            <emphasis role="bold">
              <emphasis role="underline">YBR_PARTIAL_420</emphasis>
            </emphasis>
          </para>
          </listitem>
        </itemizedlist>
        <note>
          <para><emphasis role="bold"><emphasis role="underline">These values are not permitted because they are not encodable in an uncompressed form.</emphasis></emphasis></para>
        </note>
      <para xml:id="para_7bc14864-77a5-4409-8ec0-c246cace4681">Pixel Data conveyed in the Float Pixel Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009) shall be in a Native (uncompressed) Format if encoded in a Standard Transfer Syntax.</para>
      <note>
        <orderedlist>
          <listitem>
            <para xml:id="para_19247c30-979c-41f0-8ef0-f8174def91a2">In future, if Standard Transfer Syntaxes are defined for compression of Float Pixel Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009), this constraint may be relaxed and Encapsulated Format permitted.</para>
          </listitem>
          <listitem>
            <para xml:id="para_8f83d270-3f31-4a61-9325-e6f2a8e8c7f9">This constraint does not apply to Private Transfer Syntaxes.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_cdbe477b-a4f5-417d-bee0-82e5a0b1a9ad">If Pixel Data (7FE0,0010) is sent in a Native Format, the Value Representation OW is most often required. The Value Representation OB may also be used for Pixel Data (7FE0,0010) in cases where Bits Allocated has a value less than or equal to 8, but only with Transfer Syntaxes where the Value Representation is explicitly conveyed (see Annex A<!--see <xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->.).</para>
      <note>
        <para xml:id="para_ec814ff7-cbc9-4530-a0b6-a580a48d6cf4">The DICOM default Transfer Syntax (Implicit VR Little Endian) does not explicitly convey Value Representation and therefore the VR of OB may not be used for Pixel Data (7FE0,0010) when using the default Transfer Syntax.</para>
      </note>
      <para xml:id="para_dec19bac-823e-44b1-ba85-3bc875c2010d">Float Pixel Data (7FE0,0008) is sent
        in Native Format; the Value Representation shall be OF, Bits Allocated (0028,0100)
        shall be 32, Bits Stored (0028,0101), High Bit (0028,0102) and Pixel Representation
        (0028,0103) shall not be present.</para>
      <para xml:id="para_c0645d98-0279-4962-b94a-7481a0897cf2">Double Float Pixel Data (7FE0,0009)
        is sent in Native Format; the Value Representation shall be OD, Bits Allocated
        (0028,0100) shall be 64, Bits Stored (0028,0101) and High Bit (0028,0102) and Pixel
        Representation (0028,0103) shall not be present.</para>
      <para xml:id="para_3085b522-c695-4899-b4fd-9740e6e97fcc">It is not permitted to have more than
        one of Pixel Data Provider URL (0028,7FE0), Pixel Data (7FE0,0010), Float Pixel Data
        (7FE0,0008) or Double Float Pixel Data (7FE0,0009) in the top level Data
        Set.</para>
      <note>
        <para xml:id="para_a530b9e7-2346-42df-9816-f085515ad011">Pixel Data encoded in Float Pixel
          Data (7FE0,0008) or Double Float Pixel Data (7FE0,0009) can be considered as
          consisting of Pixel Cells that entirely occupy the allocated bits, and therefore do
          not cross word boundaries.</para>
      </note>
      <para xml:id="para_d6734089-67bc-433e-87f7-40f00a859a9d">Native format Pixel Cells are encoded as the direct concatenation of the bits of each Pixel Cell, the least significant bit of each Pixel Cell is encoded in the least significant bit of the encoded word or byte, immediately followed by the next most significant bit of each Pixel Cell in the next most significant bit of the encoded word or byte, successively until all bits of the Pixel Cell have been encoded, then immediately followed by the least significant bit of the next Pixel Cell in the next most significant bit of the encoded word or byte. The number of bits of each Pixel Cell is defined by the Bits Allocated (0028,0100) Data Element Value. When a Pixel Cell crosses a word boundary in the OW case, or a byte boundary in the OB case, it shall continue to be encoded, least significant bit to most significant bit, in the next word, or byte, respectively (see Annex D<!--<xref linkend="chapter_D" xrefstyle="template:Annex %n"/>-->). For Pixel Data (7FE0,0010) encoded with the Value Representation OW, the byte ordering of the resulting 2-byte words is
  defined by the Little Endian Transfer Syntaxes negotiated at the Association Establishment (see Annex A<!--see <xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->.).</para>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_1e6dec20-1376-4b1e-b909-255ba6abcf59">For Pixel Data (7FE0,0010) encoded with the Value Representation OB, the Pixel Data (7FE0,0010) encoding is unaffected by byte ordering.</para>
          </listitem>
          <listitem>
            <para xml:id="para_754c31cf-0a9a-484e-bfba-2130fc81252e">If encoding Pixel Data (7FE0,0010) with a Value for Bits Allocated (0028,0100) not equal to 16 be sure to read and understand Annex D<!--<xref linkend="chapter_D" xrefstyle="template:Annex %n"/>-->.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_6287384f-06a5-47f6-ab74-625074661b31">If sent in an Encapsulated Format (i.e., other than the Native Format) the Value Representation OB is used. The Pixel Cells are encoded according to the encoding process defined by one of the negotiated Transfer Syntaxes (see Annex A<!--see <xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->.). The encapsulated pixel stream of encoded pixel data is segmented into one or more Fragments, each of which conveys its own explicit length. The sequence of Fragments of the encapsulated pixel stream is terminated by a delimiter, thus allowing the support of encoding processes where the resulting length of the entire pixel stream is not known until it is entirely encoded. This Encapsulated Format supports both Single-Frame and Multi-Frame images (as defined in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->).</para>
      <note>
        <para xml:id="para_b50189ee-a299-4737-bea1-c0510c91ffe1">Depending on the Transfer Syntax, a frame may be entirely contained within a single fragment, or may span multiple fragments to support buffering during compression or to avoid exceeding the maximum size of a fixed length fragment. A recipient can detect fragmentation of frames by comparing the number of fragments (the number of Items minus one for the Basic Offset Table) with the number of frames. Some performance optimizations may be available to a recipient in the absence of fragmentation of frames, but an implementation that fails to support such fragmentation does not conform to the Standard.</para>
      </note>
      <section label="8.2.1" status="3" xml:id="sect_8.2.1">
        <title>JPEG Image Compression</title>
        <para xml:id="para_d1f3d3c5-fcdf-4dc2-b38b-0bd7fbf79491">DICOM provides a mechanism for supporting the use of JPEG Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a number of Transfer Syntaxes that reference the JPEG Standard and provide a number of lossless (bit preserving) and lossy compression schemes.</para>
        <note>
          <para xml:id="para_8cd648e8-6c76-459a-ad3a-f3c2b1c8fb4c">The context where the usage of lossy compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g., compression ratio) for JPEG lossy compression is also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_9d46a7e0-da15-452f-b8d4-02d7a4b6b25b">In order to facilitate interoperability of implementations conforming to the DICOM Standard that elect to use one or more of the Transfer Syntaxes for JPEG Image Compression, the following policy is specified:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_4d36fc54-f14b-4417-ada8-8b839895c1d3">Any implementation that conforms to the DICOM Standard and has elected to support any one of the Transfer Syntaxes for lossless JPEG Image Compression, shall support the following lossless compression: The subset (first-order horizontal prediction [Selection Value 1) of JPEG Process 14 (DPCM, non-hierarchical with Huffman coding) (see Annex F<!--<xref linkend="chapter_F" xrefstyle="template:Annex %n"/>-->).</para>
          </listitem>
          <listitem>
            <para xml:id="para_f7127df5-ae83-4a2f-ac8d-fd3356d5be0c">Any implementation that conforms to the DICOM Standard and has elected to support any one of the Transfer Syntaxes for 8-bit lossy JPEG Image Compression, shall support the JPEG Baseline Compression (coding Process 1).</para>
          </listitem>
          <listitem>
            <para xml:id="para_3df4846f-15b7-45ca-8705-582504847dfe">Any implementation that conforms to the DICOM Standard and has elected to support any one of the Transfer Syntaxes for 12-bit lossy JPEG Image Compression, shall support the JPEG Compression Process 4.</para>
          </listitem>
        </itemizedlist>
        <note>
          <para xml:id="para_80e3247d-ca00-45fb-9c75-56d6805533ad">The DICOM conformance statement shall differentiate whether or not the implementation is capable of simply receiving or receiving and processing JPEG encoded images (see PS3.2<!--<olink targetdoc="PS3.2" targetptr="PS3.2" xrefstyle="select: labelnumber"/>-->).</para>
        </note>
        <para xml:id="para_4bcb841e-c6bf-4e26-82a5-3fad3c942da0">The use of the DICOM Encapsulated Format to support JPEG Compressed Pixel Data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream.</para>
        <para><emphasis role="bold"><emphasis role="underline">The requirements when using a Standard Photometric Interpretation (i.e., a Defined Term from PS.3. C.7.6.3.1.2) are specified in <xref linkend="table_8.2.1-1" xrefstyle="select: label"/> and <xref linkend="table_8.2.1-2" xrefstyle="select: label"/>. No other Standard Photometric Interpretation values shall be used.</emphasis></emphasis></para>
        <table frame="box" label="8.2.1-1" rules="all" xml:id="table_8.2.1-1">
          <caption><emphasis role="bold"><emphasis role="underline">Valid Values of Pixel Data Related Attributes for JPEG Lossy Transfer Syntaxes using Standard Photometric Interpretations</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Photometric Interpretation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Transfer Syntax</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Transfer Syntax UID</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Samples per Pixel</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Planar Configuration</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Pixel Representation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Allocated</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Stored</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">High Bit</emphasis></emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Baseline</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.50</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">7</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Extended)</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.51</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">7</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Extended</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.51</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">12</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">11</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">YBR_FULL_422</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
               <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Baseline</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
               <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.50</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">7</emphasis></emphasis>
                </para>
              </td>
            </tr>
          </tbody>
        </table> 
        <table frame="box" label="8.2.1-2" rules="all" xml:id="table_8.2.1-2">
          <caption><emphasis role="bold"><emphasis role="underline">Valid Values of Pixel Data Related Attributes for JPEG Lossless Transfer Syntaxes using Standard Photometric Interpretations</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Photometric Interpretation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Transfer Syntax</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Transfer Syntax UID</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Samples per Pixel</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Planar Configuration</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Pixel Representation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Allocated</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Stored</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">High Bit</emphasis></emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Lossless, Non-Hierarchical</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Lossless, Non-Hierarchical, SV1</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.57</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.70</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0 or 1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">YBR_FULL</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">RGB</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Lossless, Non-Hierarchical</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG Lossless, Non-Hierarchical, SV1</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.57</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.70</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
          </tbody>
        </table>
        
        
        
        
        <para>The Pixel Data characteristics included in the JPEG Interchange Format shall be used to decode the compressed data stream.</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_651e0fda-0843-4bef-b448-30495fd6330f">These requirements were formerly specified in terms of the "uncompressed pixel data from which the compressed data stream was derived". However, since the form of the "original" uncompressed data stream could vary between different implementations, this requirement is now specified in terms of consistency with what is encapsulated.</para>
              <para xml:id="para_07617cc6-a4a9-4751-9435-7b496910168d">When decompressing, should the characteristics explicitly specified in the compressed data stream (e.g., spatial subsampling or number of components or planar configuration) be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
            </listitem>
            <listitem>
              <para xml:id="para_6424f9ef-dfaa-40ff-9ca9-109d42a93363">Those characteristics not explicitly specified in the compressed data stream (e.g., the color space of the compressed components, which is not specified in the JPEG Interchange Format), or implied by the definition of the compression scheme (e.g., always unsigned in JPEG), can therefore be determined from the DICOM Data Element in the enclosing Data Set. For example a Photometric Interpretation of "YBR_FULL_422" would describe the color space that is commonly used to lossy compress images using JPEG. It is unusual to use an RGB color space for lossy compression, since no advantage is taken of correlation between the red, green and blue components (e.g., of luminance), and poor compression is achieved.</para>
            </listitem>
            <listitem>
              <para xml:id="para_e639eb95-8d71-43a2-9a21-017e0bd8eb56">The JPEG Interchange Format is distinct from the JPEG File Interchange Format (JFIF). The JPEG Interchange Format is defined in <xref linkend="biblio_ISOIEC10918-1"/> section 4.9.1, and refers to the inclusion of decoding tables, as distinct from the "abbreviated format" in which these tables are not sent (and the decoder is assumed to already have them). The JPEG Interchange Format does NOT specify the color space. The JPEG File Interchange Format, not part of the original JPEG standard, but defined in [ECMA TR-098]<!--<xref linkend="biblio_ECMA_TR-098"/>--> and [ISO/IEC 10918-5]<!--<xref linkend="biblio_ISOIEC10918-5"/>-->, is often used to store JPEG bit streams in consumer format files, and does include the ability to specify the color space of the components. The JFIF APP0 marker segment is NOT required to be present in DICOM encapsulated JPEG bit streams, and should not be relied upon to recognize the color space. Its presence is not forbidden (unlike the JP2 information for JPEG 2000 Transfer Syntaxes), but it is recommended that it be absent.</para>
            </listitem>
            <listitem>
              <para xml:id="para_b217610e-776a-4c22-88c2-6c765f687a4b">Should the compression process be incapable of encoding a particular form of pixel data representation (e.g., JPEG cannot encode signed integers, only unsigned integers), then ideally only the appropriate form should be "fed" into the compression process. However, for certain characteristics described in DICOM Data Elements but not explicitly described in the compressed data stream (such as Pixel Representation), then the DICOM Data Element should be considered to describe what has been compressed (e.g., the pixel data really is to be interpreted as signed if Pixel Representation so specifies).</para>
            </listitem>
            <listitem>
              <para xml:id="para_c141d2a1-75f6-426e-85fe-e99495d48030">DICOM Data Elements should not describe characteristics that are beyond the capability of the compression scheme used. For example, JPEG lossy processes are limited to 12 bits, hence the value of Bits Stored should be 12 or less. Bits Allocated is irrelevant, and is likely to be constrained by the Information Object Definition in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>--> to values of 8 or 16. Also, JPEG compressed data streams are always color-by-pixel and should be specified as such (a decoder can essentially ignore this element however as the value for JPEG compressed data is already known).</para>
            </listitem>
            <listitem>
              <para xml:id="para_e47ec72b-e48e-4484-8cfb-62153e734c28">If JPEG Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_FULL_422 to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
      <section label="8.2.2" status="3" xml:id="sect_8.2.2">
        <title>Run Length Encoding Compression</title>
        <para xml:id="para_85ca0802-972c-4db3-bc67-7f344dc8aeb8">DICOM provides a mechanism for supporting the use of Run Length Encoding (RLE) Compression, which is a byte oriented lossless compression scheme through the encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>--> of this Standard). Annex G<!--<xref linkend="chapter_G" xrefstyle="template:Annex %n"/>--> defines RLE Compression and its Transfer Syntax.</para>
        <note>
          <para xml:id="para_68100f7a-9340-4d87-af3f-2f6950af6a3f">The RLE Compression algorithm described in Annex G<!--<xref linkend="chapter_G" xrefstyle="template:Annex %n"/>--> is the compression used in the TIFF 6.0 specification known as the "PackBits" scheme.</para>
        </note>
        <para xml:id="para_f46aa917-6020-441c-a144-02b3a06c3e39">The use of the DICOM Encapsulated Format to support RLE Compressed Pixel Data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the compressed data.</para>
        <para><emphasis role="bold"><emphasis role="underline">The requirements when using a Standard Photometric Interpretation (i.e., a Defined Term from PS.3. C.7.6.3.1.2) are specified in <xref linkend="table_8.2.2-1" xrefstyle="select: label"/>. No other Standard Photometric Interpretation values shall be used.</emphasis></emphasis></para>
        <table frame="box" label="8.2.2-1" rules="all" xml:id="table_8.2.2-1">
          <caption><emphasis role="bold"><emphasis role="underline">Valid Values of Pixel Data Related Attributes for RLE Compression using Standard Photometric Interpretations</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Photometric Interpretation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Samples per Pixel</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Planar Configuration</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Pixel Representation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Allocated</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Stored</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">High Bit</emphasis></emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0 or 1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">PALETTE COLOR</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">YBR_FULL</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-8</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-7</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">RGB</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
          </tbody>
        </table>
        
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_6f9c3153-4628-4f7e-a842-a7d81a6845ab">These requirements were formerly specified in terms of the "uncompressed pixel data from which the compressed data was derived". However, since the form of the "original" uncompressed data stream could vary between different implementations, this requirement is now specified in terms of consistency with what is encapsulated.</para>
            </listitem>
            <listitem>
              <para xml:id="para_d1bb8f9b-f160-4538-b2db-6717374398e2">Those characteristics not implied by the definition of the compression scheme (e.g., always color-by-plane in RLE), can therefore be determined from the DICOM Data Element in the enclosing Data Set. For example a Photometric Interpretation of "YBR_FULL" would describe the color space that is commonly used to losslessly compress images using RLE. It is unusual to use an RGB color space for RLE compression, since no advantage is taken of correlation between the red, green and blue components (e.g., of luminance), and poor compression is achieved (note however that the conversion from RGB to YBR_FULL is itself lossy. A new photometric interpretation may be proposed in the future that allows lossless conversion from RGB and also results in better RLE compression ratios).</para>
            </listitem>
            <listitem>
              <para xml:id="para_7844c3d7-4b28-4f14-bcf3-19476a836b6b">DICOM Data Elements should not describe characteristics that are beyond the capability of the compression scheme used. For example, RLE compressed data streams (using the algorithm mandated in the DICOM Standard) are always color-by-plane.</para>
            </listitem>
            <listitem>
              <para xml:id="para_57b858fe-3b23-47e7-bfc9-f67233f1188f">If RLE Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_FULL to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding. It is permitted, however, to leave the YBR_FULL color components unconverted but decompressed in the Native format, in which case the Photometric Interpretation in the Data Set with the Native encoding would be YBR_FULL.</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
      <section label="8.2.3" status="3" xml:id="sect_8.2.3">
        <title>JPEG-LS Image Compression</title>
        <para xml:id="para_eccda448-0381-4c4d-90d9-7e251c4a4e1e">DICOM provides a mechanism for supporting the use of JPEG-LS Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a number of Transfer Syntaxes that reference the JPEG-LS Standard and provide a number of lossless (bit preserving) and lossy (near-lossless) compression schemes.</para>
        <note>
          <para xml:id="para_c0ca469e-6703-4121-a744-14e04ae8ce0f">The context where the usage of lossy (near-lossless) compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g., compression ratio) for JPEG-LS lossy (near-lossless) compression is also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_573ea007-d4e1-4c07-a82a-2d6b13d109eb">The use of the DICOM Encapsulated Format to support JPEG-LS Compressed Pixel Data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream. The Pixel Data characteristics included in the JPEG-LS Interchange Format shall be used to decode the compressed data stream.</para>
        <para><emphasis role="bold"><emphasis role="underline">The requirements when using a Standard Photometric Interpretation (i.e., a Defined Term from PS.3. C.7.6.3.1.2) are specified in <xref linkend="table_8.2.3-1" xrefstyle="select: label"/>. No other Standard Photometric Interpretation values shall be used.</emphasis></emphasis></para>
        <table frame="box" label="8.2.3-1" rules="all" xml:id="table_8.2.3-1">
          <caption><emphasis role="bold"><emphasis role="underline">Valid Values of Pixel Data Related Attributes for JPEG-LS Compression using Standard Photometric Interpretations</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Photometric Interpretation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Samples per Pixel</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Planar Configuration</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Pixel Representation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Allocated</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Stored</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">High Bit</emphasis></emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0 or 1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">PALETTE COLOR</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">RGB</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8 or 16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-16</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-15</emphasis></emphasis>
                </para>
              </td>
            </tr>
          </tbody>
        </table>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_e40960fc-63dd-45b0-8ed7-993b40cbd763">See also the notes in <xref linkend="sect_8.2.1" xrefstyle="select: label"/>.</para>
            </listitem>
            <listitem>
              <para xml:id="para_7eacfba3-42dd-43d0-b47f-45efb2fb200b">No <emphasis role="bold"><emphasis role="underline">color transformation </emphasis></emphasis>Photometric Interpretation specific for JPEG-LS is currently defined in DICOM. Annex F of ISO 14495-2 describes a <foreignphrase xml:lang="en-GB">"Sample transformation for inverse colour transform"</foreignphrase> and a marker segment to encode its parameters, but this is not known to have been implemented. Common practice is to compress the RGB components unconverted, which sacrifices compression performance, and send the Photometric Interpretation as RGB. Though the YBR_RCT Photometric Interpretation and component conversion could theoretically be used, in the absence of DC shifting it results in signed values to be encoded, which are not supported by JPEG-LS.</para>
            </listitem>
            <listitem>
              <para xml:id="para_1510f0e0-33b0-4e0e-9d66-73f03399a37b">If JPEG-LS Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from any other Photometric Interpretation to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
            </listitem>
          </orderedlist>
         </note>
      </section>
      <section label="8.2.4" status="3" xml:id="sect_8.2.4">
        <title>JPEG 2000 Image Compression</title>
        <para xml:id="para_e87e36c7-4f0c-49e5-a267-38982870ce88">DICOM provides a mechanism for supporting the use of JPEG 2000 Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a number of Transfer Syntaxes that reference the JPEG 2000 Standard and provide lossless (bit preserving) and lossy compression schemes.</para>
        <note>
          <para xml:id="para_80ad309d-34ab-4927-9e16-8720489de547">The context where the usage of lossy compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g., compression ratio) for JPEG 2000 lossy compression are also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_b8bb4636-3e99-4465-86d3-207414bb06a5">The use of the DICOM Encapsulated Format to support JPEG 2000 Compressed Pixel Data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream. The Pixel Data characteristics included in the JPEG 2000 bit stream shall be used to decode the compressed data stream.</para>
        <para><emphasis role="bold"><emphasis role="underline">The requirements when using a Standard Photometric Interpretation (i.e., a Defined Term from PS.3. C.7.6.3.1.2) are specified in <xref linkend="table_8.2.4-1" xrefstyle="select: label"/>. No other Standard Photometric Interpretation values shall be used.</emphasis></emphasis></para>
        <table frame="box" label="8.2.4-1" rules="all" xml:id="table_8.2.4-1">
          <caption><emphasis role="bold"><emphasis role="underline">Valid Values of Pixel Data Related Attributes for JPEG 2000 Transfer Syntaxes using Standard Photometric Interpretations</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Photometric Interpretation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Transfer Syntax</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Transfer Syntax UID</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Samples per Pixel</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Planar Configuration</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Pixel Representation</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Allocated</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">Bits Stored</emphasis></emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">High Bit</emphasis></emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME1</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">MONOCHROME2</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression (Lossless Only)</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​90</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​91</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0 or 1</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8, 16, 24, 32 or 40</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-38</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-37</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">YBR_RCT</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression (Lossless Only)</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​90</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​91</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">absent</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8, 16, 24, 32 or 40</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-38</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-37</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">YBR_ICT</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.4.91</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8, 16, 24, 32 or 40</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-38</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-37</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">RGB</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression (Lossless Only)</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​90</emphasis></emphasis>
                </para>
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​91</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8, 16, 24, 32 or 40</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-38</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-37</emphasis></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">YBR_FULL</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">JPEG 2000 Image Compression</emphasis></emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1.2.840.10008.1.2.​4.​91</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">3</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">8, 16, 24, 32 or 40</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">1-38</emphasis></emphasis>
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold"><emphasis role="underline">0-37</emphasis></emphasis>
                </para>
              </td>
            </tr>
          </tbody>
        </table> 
        <note>
          <para xml:id="para_0e5c196d-7b04-42b4-9c18-22c0162d12ca">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
        </note>
        <para xml:id="para_307b3170-88ee-4604-b4c4-958a8cc15009">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
        <para xml:id="para_db847fdd-ca67-4516-bb90-45d161167bee">The JPEG 2000 bit stream specifies whether or not a reversible or irreversible multi-component (color) transformation [ISO 15444-1 Annex G], if any, has been applied. If no multi-component transformation has been applied, then the components shall correspond to those specified by the DICOM Attribute Photometric Interpretation (0028,0004). If the JPEG 2000 Part 1 reversible multi-component transformation has been applied then the DICOM Attribute Photometric Interpretation (0028,0004) shall be YBR_RCT. If the JPEG 2000 Part 1 irreversible multi-component transformation has been applied then the DICOM Attribute Photometric Interpretation (0028,0004) shall be YBR_ICT.</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_08197219-168c-4756-88fc-be63881c1cbf">For example, single component may be present, and the Photometric Interpretation (0028,0004) may be MONOCHROME2.</para>
            </listitem>
            <listitem>
              <para xml:id="para_832bcbc5-922e-43d1-9f35-4cba78b53059">The application of a JPEG 2000 Part 1 reversible multi-component transformation is signaled in the JPEG 2000 bitstream by a value of 1 rather than 0 in the SGcod Multiple component transformation type of the COD marker segment [ISO 15444-1 Table A.17]. No other value of Photometric Interpretation than YBR_RCT or YBR_ICT is permitted when SGcod Multiple component transformation type is 1.</para>
            </listitem>
            <listitem>
              <para xml:id="para_0a77cfd2-db8f-40f4-8d23-3a10092406af">Though it would be unusual, would not take advantage of correlation between the red, green and blue components, and would not achieve effective compression, a Photometric Interpretation of RGB could be specified as long as no multi-component transformation [ISO 15444-1 Annex G] was specified by the JPEG 2000 bit stream. Alternative methods of decorrelation of the color components than those specified in [ISO 15444-1 Annex G] are permitted as defined in PS3.3, such as a Photometric Interpretation of YBR_FULL; this may be useful when converting existing YBR_FULL Pixel Data (e.g., in a different Transfer Syntax) without further loss.</para>
              <para xml:id="para_30071ffc-c6fa-43f8-84a7-56424cb8f633">In either case (Photometric Interpretation of RGB or YBR_FULL), the value of SGcod Multiple component transformation type would be 0.</para>
              <para xml:id="para_1d6dc6c4-c9e9-4b2b-8175-94181e7c3deb">
PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>--> may constrain the values of Photometric Interpretation for specific IODs.</para>
            </listitem>
            <listitem>
              <para xml:id="para_9146fe8f-7760-48f7-8a52-9c1dbf1ca927">Despite the application of a multi-component color transformation and its reflection in the Photometric Interpretation attribute, the "color space" remains undefined. There is currently no means of conveying "standard color spaces" either by fixed values (such as sRGB) or by ICC profiles. Note in particular that the JP2 file header is not sent in the JPEG 2000 bitstream that is encapsulated in DICOM.</para>
            </listitem>
            <listitem>
              <para xml:id="para_e8f7c114-fe18-4f0d-a9f3-6b8367bfb3b2">If JPEG 2000 Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_ICT or YBR_RCT to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
            </listitem>
            <listitem>
              <para><emphasis role="bold"><emphasis role="underline">The upper limit of 40 on Bits Allocated (0028,0100) and 38 on Bits Stored (0028,0101) reflects the maximum JPEG 2000 sample precision of 38 and the DICOM requirement to describe Bits Allocated (0028,0100) as multiples of bytes (octets).</emphasis></emphasis></para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_a89ecd90-af2d-401e-99e6-e7ac76600d56">The JPEG 2000 bitstream is capable of encoding both signed and unsigned pixel values, hence the value of Pixel Representation (0028,0103) may be either 0 or 1 <emphasis role="bold"><emphasis role="underline">for monochrome Photometric Interpretations </emphasis></emphasis>depending on what has been encoded (as specified in the SIZ marker segment in the precision and sign of component parameter).</para>
        <para xml:id="para_2b48c289-0ee9-4ec6-9685-99e6a1a9a16f">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the JPEG 2000 standard, hence it shall be set to 0.</para>
      </section>
      <section label="8.2.5" status="3" xml:id="sect_8.2.5">
        <title>MPEG2 Main Profile / Main Level Image Compression</title>
        <para xml:id="para_1f5ce4d4-99d8-4263-8e07-fe67a269ef76">DICOM provides a mechanism for supporting the use of MPEG2 Main Profile / Main Level Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a Transfer Syntax that references the MPEG2 Main Profile / Main Level Standard.</para>
        <note>
          <para xml:id="para_c7cb7067-a7ee-4fc4-ba28-177d332921ba">MPEG2 compression is inherently lossy. The context where the usage of lossy compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g., compression ratio) for MPEG2 Main Profile / Main Level are also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_1cbdb168-629e-4145-99a9-4767b0dec92a">The use of the DICOM Encapsulated Format to support MPEG2 Main Profile / Main Level compressed pixel data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream, with some specific exceptions noted here. The Pixel Data characteristics included in the MPEG2 Main Profile / Main Level bit stream shall be used to decode the compressed data stream.</para>
        <note>
          <para xml:id="para_afa32905-415a-4f71-98b1-67e8ea4aba49">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
        </note>
        <para xml:id="para_9d54ce83-fbe6-44d9-a910-4c3eae7443be">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
        <para xml:id="para_c268bcf7-3ff6-4626-af47-8aab8feb28ff">The MPEG2 Main Profile / Main Level bit stream specifies whether or not a reversible or irreversible multi-component (color) transformation, if any, has been applied. If no multi-component transformation has been applied, then the components shall correspond to those specified by the DICOM Attribute Photometric Interpretation (0028,0004). MPEG2 Main Profile / Main Level applies an irreversible multi-component transformation, so DICOM Attribute Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420 in the case of multi-component data, and MONOCHROME2 in the case of single component data (even though the MPEG2 bit stream itself is always encoded as three components, one luminance and two chrominance).</para>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_2ba2146f-321d-499e-931c-17f41cb4a462">If MPEG2 Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_PARTIAL_420 to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
            </listitem>
            <listitem>
              <para xml:id="para_1657a855-c8c2-4d9c-a5a6-5c6bf4093180">MPEG2 proposes some video formats. Each of the standards specified is used in a different market, including: ITU-R BT.470-2 System M for SD NTSC and ITU-R BT.470-2 System B/G for SD PAL/SECAM. A PAL based system should therefore be based on ITU-BT.470 System B for each of Color Primaries, Transfer Characteristic (gamma) and matrix coefficients and should take a value of 5 as defined in <xref linkend="biblio_ISOIEC13818-2"/>.</para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_6572622d-e549-4279-9ff6-b1c29e90d98c">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the MPEG2 Main Profile / Main Level standard, hence it shall be set to 0.</para>
        <para xml:id="para_c52e14b1-3f92-476c-be1f-21528a35ce09">In summary:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_08e7f576-d52b-48c5-8b1c-b98d0f33a040">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_d4dfb074-7f84-49c0-a087-e6076c3985f5">Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420</para>
          </listitem>
          <listitem>
            <para xml:id="para_d9d8a93a-8456-4796-9b59-4fd175d26ef6">Bits Allocated (0028,0100) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_f84dcdf0-217a-4004-bb83-eeffd64aa049">Bits Stored (0028,0101) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_5d0902d5-9c2d-4a80-8131-a35062b83e92">High Bit (0028,0102) shall be 7</para>
          </listitem>
          <listitem>
            <para xml:id="para_89c83e3d-a690-412a-8525-d485f6d3254f">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_79bedcd7-f3d4-43a8-97df-ce86d89600e9">Planar Configuration (0028,0006) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_9af789da-aa17-43e7-91c3-2a122dd06ad3">Rows (0028,0010), Columns (0028,0011), Cine Rate (0018,0040) and Frame Time (0018,1063) or Frame Time Vector (0018,1065) shall be consistent with the limitations of Main Profile / Main Level, as ...</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="8.2.6" status="3" xml:id="sect_8.2.6">
        <title>MPEG2 Main Profile / High Level Image Compression</title>
        <para xml:id="para_2bd0d799-f7d2-4866-83d4-fd1bbb3819cb">MPEG2 Main Profile / High Level corresponds to what is commonly known as HDTV ('High Definition Television'). DICOM provides a mechanism for supporting the use of MPEG2 Main Profile / High Level Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a Transfer Syntax that references the MPEG2 Main Profile / High Level Standard.</para>
        <note>
          <para xml:id="para_2c1d1db8-092d-4e8a-a1c4-d30dc53fdfb2">MPEG2 compression is inherently lossy. The context where the usage of lossy compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g., compression ratio) for MPEG2 Main Profile / High Level are also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_99b66f2f-df69-4543-9d01-21907b1f3918">The use of the DICOM Encapsulated Format to support MPEG2 Main Profile / High Level compressed pixel data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream, with some specific exceptions noted here. The Pixel Data characteristics included in the MPEG2 Main Profile / High Level bit stream shall be used to decode the compressed data stream.</para>
        <note>
          <para xml:id="para_49d5c0fa-e261-4974-8d53-d8428ad857b0">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
        </note>
        <para xml:id="para_550767c3-9efa-4014-be77-38e3779c1b6b">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
        <note>
          <para xml:id="para_893ba67a-05cf-46bd-ad44-72ec2989f1e3">If MPEG2 Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_PARTIAL_420 to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
        </note>
        <para xml:id="para_39989b63-b4d8-46ee-ab01-0f10be3905b3">The requirements are:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_5901446e-9730-4bab-b6e0-c24def204714">Planar Configuration (0028,0006) shall be 0</para>
            <note>
              <para xml:id="para_5f4e2958-bb4e-46e7-b3ba-3e5cf5ee3980">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the MPEG2 standard, hence it is set to 0.</para>
            </note>
          </listitem>
          <listitem>
            <para xml:id="para_6ec13661-b33d-4064-a74a-a46268699647">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_fe60ba81-e0d8-4097-a998-470a8a156835">Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420 or MONOCHROME2</para>
          </listitem>
          <listitem>
            <para xml:id="para_2eab63a8-e0da-4b56-ad07-efa8d8b65ee5">Bits Allocated (0028,0100) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_5cf26856-81c3-4dc1-8ea7-98157fc23a38">Bits Stored (0028,0101) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_8084b705-3cb4-4372-8d5f-82624aebd6fa">High Bit (0028,0102) shall be 7</para>
          </listitem>
          <listitem>
            <para xml:id="para_48974a0b-0c63-40eb-9712-c6455bdf7bad">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_8978cac2-9467-4516-b17a-fa1fa359b009">Rows (0028,0010) shall be either 720 or 1080</para>
          </listitem>
          <listitem>
            <para xml:id="para_de5b75f8-58d4-4f22-9e7b-3f3373641996">Columns (0028,0011) shall be 1280 if Rows is 720, or shall be 1920 if Rows is 1080.</para>
          </listitem>
          <listitem>
            <para xml:id="para_44cea47b-02b1-4cfc-b4d2-e8af3a65c684">The value of MPEG2 aspect_ratio_information shall be 0011 in the encapsulated MPEG2 data stream corresponding to a 'Display Aspect Ratio' (DAR) of 16:9.</para>
          </listitem>
          <listitem>
            <para xml:id="para_86544910-25d3-4cd9-8de4-81c8323f5715">The DICOM attribute Pixel Aspect Ratio (0028,0034) shall be absent. This corresponds to a 'Sampling Aspect Ratio' (SAR) of 1:1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_e83e3223-14a5-4d0b-8baf-84f6fa172171">Cine Rate (0018,0040) and Frame Time (0018,1063) or Frame Time Vector (0018,1065) shall be consistent with the limitations of Main Profile / High Level, as ...</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="8.2.7" status="3" xml:id="sect_8.2.7">
        <title>MPEG-4 AVC/H.264 High Profile / Level 4.1 Video Compression</title>
        <para xml:id="para_65917ac3-a992-42eb-9354-ba19938a0b2b">MPEG-4 AVC/H.264 High Profile / Level 4.1 corresponds to what is commonly known as HDTV ('High Definition Television'). DICOM provides a mechanism for supporting the use of MPEG-4 AVC/H.264 Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a Transfer Syntax that references the MPEG-4 AVC/H.264 Standard.</para>
        <note>
          <para xml:id="para_3472a19e-4744-4448-ae6d-5f3cfbe425ec">MPEG-4 AVC/H.264 compression / High Profile compression is inherently lossy. The context where the usage of lossy compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g., compression ratio) for MPEG-4 AVC/H.264 High Profile / Level 4.1 are also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_382fd973-3a0c-4bca-9be2-a12b55d74976">The use of the DICOM Encapsulated Format to support MPEG-4 AVC/H.264 compressed pixel data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream, with some specific exceptions noted here. The Pixel Data characteristics included in the MPEG-4 AVC/H.264 bit stream shall be used to decode the compressed data stream.</para>
        <note>
           <para xml:id="para_189b7ade-5a0e-4b48-a83c-a2735266ee5a">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
        </note>
        <para xml:id="para_1e5e3340-6432-49d9-8cdf-21c44d306380">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
        <note>
          <para xml:id="para_10b66283-59e0-410b-bd94-c1ca64a0eec1">If MPEG-4 Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_PARTIAL_420 to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
        </note>
        <para xml:id="para_475237f7-b060-4238-a2ea-c4194c317a0a">The requirements are:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_1cbbe360-2540-4190-91ee-b15ad4889d03">Planar Configuration (0028,0006) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_45526acd-9d8d-4fae-9f2c-78789ead39a8">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_0e6d3a02-d2c9-436f-92bf-abf5e66426d2">Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420</para>
          </listitem>
          <listitem>
            <para xml:id="para_dfd4f6f7-f3f4-45ac-875d-22992ce31dfe">Bits Allocated (0028,0100) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_903f97bc-0eb4-41c7-8d04-c372431d599f">Bits Stored (0028,0101) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_77874c28-b409-45c5-b097-f4abb46f5a5c">High Bit (0028,0102) shall be 7</para>
          </listitem>
          <listitem>
            <para xml:id="para_346c4771-ce79-4e2a-a3a5-c0c2d76676fe">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_5a76a9a9-04c1-4998-92fd-5fb065e9df25">The value of MPEG-4 AVC/H.264 sample aspect_ratio_idc shall be 1 in the encapsulated MPEG-4 AVC/H.264 bit stream if aspect_ratio_info_present_flag is 1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_cd8a94cf-2a14-481e-919c-bea2000d8d02">Pixel Aspect Ratio (0028,0034) shall be absent. This corresponds to a 'Sampling Aspect Ratio' (SAR) of 1:1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_9f6199c0-53da-449c-93e0-19c616ebbcc6">The possible values for Rows (0028,0010), Columns (0028,0011), Cine Rate (0018,0040), and Frame Time (0018,1063) or Frame Time Vector (0018,1065) depend on the used Transfer Syntax.</para>
            <itemizedlist>
              <listitem>
                <para xml:id="para_2d08b7a4-ee6b-4e49-9045-36a0497567c1">....</para>
              </listitem>
            </itemizedlist>
          </listitem>
        </itemizedlist>
        <para>...</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_a10eb607-0ee8-402c-9f77-de73d5c8d0ad">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the MPEG-4 AVC/H.264 standard, hence it is set to 0.</para>
            </listitem>
            <listitem>
              <para xml:id="para_39e69a18-bf46-4d5d-bdeb-d14d4f83641b">...</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
      <section label="8.2.8" status="3" xml:id="sect_8.2.8">
        <title>MPEG-4 AVC/H.264 High Profile / Level 4.2 Video Compression</title>
        <para xml:id="para_0857d3a8-89d3-4fa6-99df-d010578a607b">DICOM provides a mechanism for supporting the use of MPEG-4 AVC/H.264 Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines Transfer Syntaxes that reference the MPEG-4 AVC/H.264 Standard.</para>
        <note>
          <para xml:id="para_82d93ddc-5e67-4256-8e12-f7e272fdd797">MPEG-4 AVC/H.264 compression / High Profile compression is inherently lossy. The context where the usage of lossy compression of medical images is clinically acceptable is beyond the scope of the DICOM Standard. The policies associated with the selection of appropriate compression parameters (e.g. compression ratio) for MPEG-4 AVC/H.264 High Profile / Level 4.2 are also beyond the scope of this standard.</para>
        </note>
        <para xml:id="para_1cd0da6b-94af-4eda-9f4e-3a4b71727169">The use of the DICOM Encapsulated Format to support MPEG-4 AVC/H.264 compressed pixel data requires that the Data Elements that are related to the Pixel Data encoding (e.g. Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream, with some specific exceptions noted here. The Pixel Data characteristics included in the MPEG-4 AVC/H.264 bit stream shall be used to decode the compressed data stream.</para>
        <note>
           <para xml:id="para_a11f89e6-4849-4d7e-a930-62ab9e4fb12c">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
        </note>
        <para xml:id="para_ec946e84-eec7-4e63-91e3-4ab0581cda27">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed data set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
        <note>
          <para xml:id="para_6627d6a9-29a0-42ad-b465-1769e704f609">If MPEG-4 Compressed Pixel Data is decompressed and re-encoded in Native (uncompressed) form, then the Data Elements that are related to the Pixel Data encoding are updated accordingly. If color components are converted from YBR_PARTIAL_420 to RGB during decompression and Native re-encoding, the Photometric Interpretation will be changed to RGB in the Data Set with the Native encoding.</para>
        </note>
        <para xml:id="para_f88bec93-d7a1-4443-a09b-d7b40a577317">The requirements are:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_4260fe92-7704-4986-8093-d94bfeb75c21">Planar Configuration (0028,0006) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_bf79123f-5908-421d-8582-b1dae38d79d6">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_dffaa66b-6d25-4a04-8c8c-89f3646823ab">Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420</para>
          </listitem>
          <listitem>
            <para xml:id="para_05a405b0-ff19-4015-b0eb-1629897b4365">Bits Allocated (0028,0100) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_dadf1b11-cf43-4b74-82c5-9435d1c702f9">Bits Stored (0028,0101) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_b750379b-d6cd-4537-9adf-420360b8a5ed">High Bit (0028,0102) shall be 7</para>
          </listitem>
          <listitem>
            <para xml:id="para_f3e1fe3d-deb9-4670-bdad-1a6276ad414c">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_fb67109c-80c2-4541-a75d-93198451455a">The value of MPEG-4 AVC/H.264 sample aspect_ratio_idc shall be 1 in the encapsulated MPEG-4 AVC/H.264 bit stream if aspect_ratio_info_present_flag is 1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_5c75cd5b-7ebb-4329-b345-ac81d5e391fc">Pixel Aspect Ratio (0028,0034) shall be absent. This corresponds to a 'Sampling Aspect Ratio' (SAR) of 1:1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_d588f9c8-c387-4c86-9f20-1060d4e3b5f8">The values for Rows (0028,0010), Columns (0028,0011), Cine Rate (0018,0040), and Frame Time (0018,1063) or Frame Time Vector (0018,1065) shall be compliant with the High Profile / Level 4.2 of the MPEG-4 AVC/H.264 standard (<xref linkend="biblio_ISOIEC14496-10"/>) and restricted to a square pixel aspect ratio.</para>
          </listitem>
        </itemizedlist>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_a224bd50-bbf3-4621-97f6-556ce086c889">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the MPEG-4 AVC/H.264 standard, hence it is set to 0.</para>
            </listitem>
            <listitem>
              <para xml:id="para_1d1bf2e2-69c7-498f-a782-4fd269a68f15">...</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
      <section label="8.2.10" status="3" xml:id="sect_8.2.10">
        <title>HEVC/H.265 Main Profile / Level 5.1 Video Compression</title>
        <para xml:id="para_ecbac5ae-1a29-45b9-97e7-de9a978c5864">HEVC/H.265 Main Profile / Level 5.1 Main tier is designed for the compression of 4:2:0 video formats up to 4k at 60 frames per second with a bit depth of 8 bits. DICOM provides a mechanism for supporting the use of HEVC/H.265 Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a Transfer Syntax that references the HEVC/H.265 Standard.</para>
        <para xml:id="para_401c12f8-ceff-4359-8382-e21ab367006f">The use of the DICOM Encapsulated Format to support HEVC/H.265 compressed pixel data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream, with some specific exceptions noted here. The Pixel Data characteristics included in the HEVC/H.265 bit stream shall be used to decode the compressed data stream.</para>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_7dd1c8b4-a8e0-4185-8af2-5f7e4dc4ad24">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
            </listitem>
            <listitem>
              <para xml:id="para_7089107c-7682-48e4-a7da-9863253c0142">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_51e74998-4588-4d0c-8004-64bbae248f7f">The requirements are:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_2a41359a-3421-4281-8480-49a617f23e85">Planar Configuration (0028,0006) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_48bc1fda-aad6-4485-8bf4-f66a7bc17ea8">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_23472eee-7d5d-441c-8431-045e40e29393">Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420</para>
          </listitem>
          <listitem>
            <para xml:id="para_56e56027-10e8-4b63-8bb4-1da922214860">Bits Allocated (0028,0100) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_11dc76d4-37d9-40bc-8e4c-b213ed0f0bd2">Bits Stored (0028,0101) shall be 8</para>
          </listitem>
          <listitem>
            <para xml:id="para_8c1e008d-4961-454e-9681-c6c068710c8c">High Bit (0028,0102) shall be 7</para>
          </listitem>
          <listitem>
            <para xml:id="para_f5286466-3f22-487f-8808-843fbb05aede">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_8501793a-f114-49db-a71d-f087976a5cfe">The value of HEVC/H.265 sample aspect_ratio_idc shall be 1 in the encapsulated HEVC/H.265 bit stream if aspect_ratio_info_present_flag is 1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_3a4c97e0-33c4-447d-89b0-7c7ce84a4462">Pixel Aspect Ratio (0028,0034) shall be absent. This corresponds to a 'Sampling Aspect Ratio' (SAR) of 1:1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_c423a46f-9a74-4ee1-9f35-c00103d9f01e">The values for Rows (0028,0010), Columns (0028,0011), Cine Rate (0018,0040) and Frame Time (0018,1063) or Frame Time Vector (0018,1065) shall be compliant with the Main Profile / Level 5.1 of the HEVC/H.265 standard <xref linkend="biblio_ISOIEC23008-2"/> and restricted to a square pixel aspect ratio.</para>
          </listitem>
        </itemizedlist>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_8b885190-3ea8-453f-98fb-24ea8dbb034b">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the HEVC/H.265 standard, hence it is set to 0.</para>
            </listitem>
            <listitem>
              <para xml:id="para_3b8460ea-6b85-45a4-a345-bc274d4494d6">...</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
      <section label="8.2.11" status="3" xml:id="sect_8.2.11">
        <title>HEVC/H.265 Main 10 Profile / Level 5.1 Video Compression</title>
        <para xml:id="para_ea60a87d-627c-4294-aed3-2cae565709e4">HEVC/H.265 Main 10 Profile / Level 5.1 Main tier is designed for the compression of 4:2:0 video formats up to 4k at 60 frames per second with a bit depth of 10 bits. DICOM provides a mechanism for supporting the use of HEVC/H.265 Image Compression through the Encapsulated Format (see PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->). Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> defines a Transfer Syntax that references the HEVC/H.265 Standard.</para>
        <para xml:id="para_bd767978-4c77-40ab-9f1c-96a4aad381e1">The use of the DICOM Encapsulated Format to support HEVC/H.265 compressed pixel data requires that the Data Elements that are related to the Pixel Data encoding (e.g., Photometric Interpretation, Samples per Pixel, Planar Configuration, Bits Allocated, Bits Stored, High Bit, Pixel Representation, Rows, Columns, etc.) shall contain values that are consistent with the characteristics of the compressed data stream, with some specific exceptions noted here. The Pixel Data characteristics included in the HEVC/H.265 bit stream shall be used to decode the compressed data stream.</para>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_c81d7049-de34-42fe-a614-1c154d3850e2">These requirements are specified in terms of consistency with what is encapsulated, rather than in terms of the uncompressed pixel data from which the compressed data stream may have been derived.</para>
            </listitem>
            <listitem>
              <para xml:id="para_1693618b-1de9-4c36-84de-c945b03033bb">When decompressing, should the characteristics explicitly specified in the compressed data stream be inconsistent with those specified in the DICOM Data Elements, those explicitly specified in the compressed data stream should be used to control the decompression. The DICOM data elements, if inconsistent, can be regarded as suggestions as to the form in which an uncompressed Data Set might be encoded<emphasis role="bold"><emphasis role="underline">, subject to the general and IOD-specific rules for uncompressed Photometric Interpretation and Planar Configuration, which may require that decompressed data be converted to one of the permitted forms</emphasis></emphasis>.</para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_f9359985-4512-44ae-8648-6c8b8f8dde52">The requirements are:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_026e918f-5e12-4df0-9f06-a0fb993c852a">Planar Configuration (0028,0006) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_ccb0f8a4-325c-40e1-a911-1745aa24056c">Samples per Pixel (0028,0002) shall be 3</para>
          </listitem>
          <listitem>
            <para xml:id="para_d3d07c69-1ef8-4fde-bcef-ad8ac395ca5a">Photometric Interpretation (0028,0004) shall be YBR_PARTIAL_420</para>
          </listitem>
          <listitem>
            <para xml:id="para_a368515b-5fb5-4749-8e53-cf43fdbf5ee0">Bits Allocated (0028,0100) shall be 16</para>
          </listitem>
          <listitem>
            <para xml:id="para_2e5b0075-b543-40a8-94a6-b18b3b2d221f">Bits Stored (0028,0101) shall be 10</para>
          </listitem>
          <listitem>
            <para xml:id="para_2346aecb-d609-4d13-a9b0-94e8ad9ebce2">High Bit (0028,0102) shall be 9</para>
          </listitem>
          <listitem>
            <para xml:id="para_1210a07f-4078-4534-bcb1-7d0ac13056f0">Pixel Representation (0028,0103) shall be 0</para>
          </listitem>
          <listitem>
            <para xml:id="para_fcc3f829-2fe4-4227-98cd-e1a51fb14eff">The value of HEVC/H.265 sample aspect_ratio_idc shall be 1 in the encapsulated HEVC/H.265 bit stream if aspect_ratio_info_present_flag is 1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_a613bf7f-43e7-448b-8c4d-04c7c9aa9bd7">Pixel Aspect Ratio (0028,0034) shall be absent. This corresponds to a 'Sampling Aspect Ratio' (SAR) of 1:1.</para>
          </listitem>
          <listitem>
            <para xml:id="para_09e08cee-c861-4025-9f02-a724b4a7b7ff">The values for Rows (0028,0010) , Columns (0028,0011), Cine Rate (0018,0040) , and Frame Time (0018,1063) or Frame Time Vector (0018,1065) shall be compliant with the Main 10 Profile / Level 5.1 of the HEVC/H.265 standard <xref linkend="biblio_ISOIEC23008-2"/> and restricted to a square pixel aspect ratio.</para>
          </listitem>
        </itemizedlist>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_b0a49b3b-7eae-4bb0-a0a3-cdb86e6d2ce1">The value of Planar Configuration (0028,0006) is irrelevant since the manner of encoding components is specified in the HEVC/H.265 standard, hence it is set to 0.</para>
            </listitem>
            <listitem>
              <para xml:id="para_93c3447c-db1d-4165-b2cc-9a942229f41d">...</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
</section>
      
      <section label="A" status="1" xml:id="chapter_A">
    <title>Transfer Syntax Specifications (Normative)</title>
    <section label="A.4" status="2" xml:id="sect_A.4">
      <title>Transfer Syntaxes For Encapsulation of Encoded Pixel Data</title>
      <para xml:id="para_637df8e8-3ec9-4054-accf-ac88bdd73f9d">...</para>
      <section label="A.4.1" status="3" xml:id="sect_A.4.1">
        <title>JPEG Image Compression</title>
        <para xml:id="para_1d105a31-528c-40e5-9872-b0852ffb6548">The International Standards Organization ISO/IEC JTC1 has developed an International Standard, ISO 10918-1 (JPEG Part 1) and an International Standard, ISO 10918-2 (JPEG Part 2), known as the JPEG Standard, for digital compression and coding of continuous-tone still images (see Annex F<!--<xref linkend="chapter_F" xrefstyle="template:Annex %n"/>--> for further details).</para>
        <para xml:id="para_50e5756a-84bc-43b9-9def-60847d29e2f0">A DICOM Transfer Syntax for JPEG Image Compression shall be identified by a UID value, appropriate to its JPEG coding process, chosen from <xref linkend="table_A.4-3" xrefstyle="select: label"/>.</para>
        <table frame="box" label="A.4-3" rules="all" xml:id="table_A.4-3">
          <caption>DICOM Transfer Syntax UIDs for JPEG</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5188fdac-1789-4071-a17a-c94e036d32ba">
                  <emphasis role="bold">DICOM Transfer Syntax UID</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_95aa5711-434a-4737-8f7a-a0812680f119">
                  <emphasis role="bold">JPEG coding process</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_921ea54c-bc6f-4f80-9453-6e45bf7c2b9c">
                  <emphasis role="bold">JPEG description</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f016033d-c682-4251-954e-8b452161dbf5">1.2.840.10008.1.2.4.50</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_e6005b7f-9401-4ba3-889c-129e48bcf005">1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_374928a9-fade-4b53-8df7-fd2990976ce2">baseline</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_9318b17d-e725-48e2-b06f-0e95f8b7d828">1.2.840.10008.1.2.4.51</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_dc483a68-beff-47d0-a4fb-5d212a8b333b">2(8-bit),4(12-bit)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_a1e704de-93ff-4c09-9995-55f0fbd8e196">extended</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_897bd126-b8eb-48ad-88ae-f24d4f09c6bc">1.2.840.10008.1.2.4.57</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_c7d8c844-9699-4362-a0bf-3cd68ae0e61f">14</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_4785a156-d763-4c91-acf6-0d580d5b4a2c">lossless, non-hierarchical</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_ce6bcb2e-1c04-4e51-bbdc-cf7140b26ee8">1.2.840.10008.1.2.4.70</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_abec0a8d-2572-4bac-bebe-bd39a433a92f">14</para>
                <para xml:id="para_ee2659b1-d789-406b-849d-e78fc537ddee">(Selection Value 1)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f8b8e3d3-f9f0-4843-8ca1-99520a5a0300">lossless, non-hierarchical, first-order prediction</para>
              </td>
            </tr>
          </tbody>
        </table>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_00ab89fb-06d9-4835-a863-b3f35e1ae27c">DICOM identifies, to increase the likelihood of successful association, three Transfer Syntaxes for Default JPEG Compression Image processes (see <xref linkend="sect_8.2.1" xrefstyle="select: label"/> and Section 10<!--<xref linkend="chapter_10" xrefstyle="template:Section %n"/>-->).</para>
            </listitem>
            <listitem>
              <para xml:id="para_38e67967-2754-41fe-9100-946f1a9b329b">Different JPEG processes may use different SOF marker segments. E.g., the baseline JPEG process 1 used with the 1.2.840.10008.1.2.4.50 Transfer Syntax uses the SOF0 marker, whereas the extended process 2 used with the 1.2.840.10008.1.2.4.51 Transfer Syntax uses the SOF1 marker. Accordingly, even though both bit streams encode 8 bit images using DCT and Huffman coding, the bit streams are not identical. Further, the extended process 2 may (but is not required to) use more AC and DC tables (up to 4 of each, rather than 2, per ISO 10918-1 Section F.1.3).</para>
              <para xml:id="para_98c18195-612e-4698-bc65-bc370730e227">It is not compliant to send bit streams with the SOF0 marker using the 1.2.840.10008.1.2.4.51 Transfer Syntax, but it is recommended that receivers of the 1.2.840.10008.1.2.4.51 Transfer Syntax be able to decode bit streams with the SOF0 marker (this asymmetry is consistent with ISO 10918-2 requirements; see <xref linkend="sect_A.4.1" xrefstyle="select: labelnumber"/>).</para>
            </listitem>
            <listitem>
              <para xml:id="para_bb9bb8f3-c89e-4032-8eaa-6fa560460ff0">It is recommended that lossy compressed 8 bit images be encoded with the 1.2.840.10008.1.2.4.50 Transfer Syntax rather than the 1.2.840.10008.1.2.4.51 Transfer Syntax, unless the additional features of the extended process are required. Support of the 1.2.840.10008.1.2.4.50 Transfer Syntax is required for 8 bit images anyway (as described in <xref linkend="sect_8.2.1" xrefstyle="select: labelnumber"/>) and to avoid confusion with the use of 12 bit images encoded with Process 4 in the 1.2.840.10008.1.2.4.51 Transfer Syntax (defined as a DICOM Default Transfer Syntax for 12 bit images in 10.3<!--<xref linkend="sect_10.3" xrefstyle="select: labelnumber"/>-->).</para>
            </listitem>
          </orderedlist>
         </note>
        <para xml:id="para_e5359a25-4c4f-47df-8c46-c54ba1882f10">If the object allows multi-frame images in the pixel data field, then each frame shall be encoded separately. Each fragment shall contain encoded data from a single-frame image.</para>
        <note>
          <para xml:id="para_dd8c4e91-11bf-428c-a004-f48d04eb8f60">Though a fragment may not contain encoded data from more than one frame, the encoded data from one frame may span multiple fragments. See note in <xref linkend="sect_8.2" xrefstyle="select: label"/>.</para>
        </note>
        <para xml:id="para_b4c579fd-5d9f-4c45-a18e-a7e377e6ee57">For all images, including all frames of a multi-frame image, the JPEG Interchange Format shall be used (the table specification shall be included).</para>
        <note>
          <para xml:id="para_7e177ec0-1d10-4178-9781-10569107c160">This refers to the <xref linkend="biblio_ISOIEC10918-1"/> "interchange format", not the <xref linkend="biblio_ISOIEC10918-5"/> JPEG File Interchange Format (JFIF).</para>
        </note>
        <para xml:id="para_822bfd5b-306d-43cf-95fe-9123126b697d">If images with Photometric Interpretation (0028,0004) YBR_FULL_422 or YBR_PARTIAL_422, are encoded with JPEG coding Process 1 (non hierarchical with Huffman coding), identified by DICOM Transfer Syntax UID "1.2.840.10008.1.2.4.50"the minimum compressible unit is YYCBCR, where Y, CB, and CR are 8 by 8 blocks of pixel values. The data stream encodes two Y blocks followed by the corresponding CB and CR blocks.</para>
      </section>
      <section label="A.4.2" status="3" xml:id="sect_A.4.2">
        <title>RLE Compression</title>
        <para xml:id="para_51630298-8d0b-4278-a3cb-1d87f0f09fe9">
Annex G<!--<xref linkend="chapter_G" xrefstyle="template:Annex %n"/>--> defines a RLE Compression Transfer Syntax. This transfer Syntax is identified by the UID value "1.2.840.10008.1.2.5". If the object allows multi-frame images in the pixel data field, then each frame shall be encoded separately. Each frame shall be encoded in one and only one Fragment (see <xref linkend="sect_8.2" xrefstyle="select: label"/>).</para>
      </section>
      <section label="A.4.3" status="3" xml:id="sect_A.4.3">
        <title>JPEG-LS Image Compression</title>
        <para xml:id="para_bb588e44-6c19-409a-bc82-eeadf45b36c7">The International Standards Organization ISO/IEC JTC1 has developed an International Standard, <xref linkend="biblio_ISOIEC14495-1"/> (JPEG-LS Part 1), for digital compression and coding of continuous-tone still images (see Annex F<!--<xref linkend="chapter_F" xrefstyle="template:Annex %n"/>--> for further details).</para>
        <para xml:id="para_50ad5f9e-67ec-405b-8aba-15ec24c1bb59">A DICOM Transfer Syntax for JPEG-LS Image Compression shall be identified by a UID value, appropriate to its JPEG-LS coding process.</para>
        <para xml:id="para_28e21f5e-7477-472d-a7f4-8e27bbdc9cc9">Two Transfer Syntaxes are specified for JPEG-LS:</para>
        <orderedlist>
          <listitem>
            <para xml:id="para_ee657e08-42f2-4e42-acf1-56c8d5bc0aae">A Transfer Syntax with a UID of "1.2.840.10008.1.2.4.80 ", which specifies the use of the lossless mode of JPEG-LS. In this mode the absolute error between the source and reconstructed images will be zero.</para>
          </listitem>
          <listitem>
            <para xml:id="para_9d5a32c3-3d43-4288-8a3e-1ad75a0a4978">A Transfer Syntax with a UID of "1.2.840.10008.1.2.4.81 ", which specifies the use of the near-lossless mode of JPEG-LS. In this mode, the absolute error between the source and reconstructed images will be constrained to a finite value that is conveyed in the compressed bit stream. Note that this process can, at the discretion of the encoder, be used to compress images with an error constrained to a value of zero, resulting in no loss of information.</para>
          </listitem>
        </orderedlist>
        <para xml:id="para_0cab9253-7b23-4273-b98d-aa585f2a1fa5">If the object allows multi-frame images in the pixel data field, then each frame shall be encoded separately. Each fragment shall contain encoded data from a single-frame image.</para>
        <note>
          <para xml:id="para_9180ee35-0d95-41e0-8bcf-738de183a0ed">Though a fragment may not contain encoded data from more than one frame, the encoded data from one frame may span multiple fragments. See note in <xref linkend="sect_8.2" xrefstyle="select: label"/>.</para>
        </note>
        <para xml:id="para_295a61f5-66d2-4bb3-a331-989257db23e0">For all images, including all frames of a multi-frame image, the JPEG-LS Interchange Format shall be used (all parameter specifications shall be included).</para>
      </section>
      <section label="A.4.4" status="3" xml:id="sect_A.4.4">
        <title>JPEG 2000 Image Compression</title>
        <para xml:id="para_83279ae5-ca69-460a-a33a-94159c991927">The International Standards Organization ISO/IEC JTC1 has developed an International Standard, <xref linkend="biblio_ISOIEC15444-1"/> (JPEG 2000 Part 1), for digital compression and coding of continuous-tone still images (see Annex F<!--<xref linkend="chapter_F" xrefstyle="template:Annex %n"/>--> for further details).</para>
        <para xml:id="para_2bf4d123-b0c1-4cad-b569-f453bac514c1">A DICOM Transfer Syntax for JPEG 2000 Image Compression shall be identified by a UID value, appropriate to the choice of JPEG 2000 coding process.</para>
        <para xml:id="para_06500b71-3629-459f-b21e-5e6512edd819">Two Transfer Syntaxes are specified for JPEG 2000 Part 1:</para>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_71f243ae-6db3-40cf-8034-73c5a5df8d55">A Transfer Syntax with a UID of "1.2.840.10008.1.2.4.90 ", which specifies the use of the lossless (reversible) mode of JPEG 2000 Part 1 (<xref linkend="biblio_ISOIEC15444-1"/>) (i.e., the use of a reversible wavelet transformation and a reversible color component transformation, if applicable, and no quantization).</para>
          </listitem>
          <listitem>
            <para xml:id="para_ef1b3945-ad5a-4bb9-95ad-195285762483">A Transfer Syntax with a UID of "1.2.840.10008.1.2.4.91", which specifies the use of either:</para>
            <orderedlist numeration="loweralpha">
              <listitem>
                <para xml:id="para_5442f79b-cf8e-45ae-ba12-b51ea018181b">the lossless (reversible) mode of JPEG 2000 Part 1 (<xref linkend="biblio_ISOIEC15444-1"/>) (i.e., the use of a reversible wavelet transformation and a reversible color component transformation, if applicable, and no quantization or code stream truncation), or</para>
              </listitem>
              <listitem>
                <para xml:id="para_f82bebad-2ba4-4cf9-a750-931c3d95c1b9">the lossy (irreversible) mode of JPEG 2000 Part 1 (<xref linkend="biblio_ISOIEC15444-1"/>) (i.e., the use of an irreversible wavelet transformation and an irreversible color component transformation, if applicable, and optionally quantization, or the use of a reversible wavelet transformation and a reversible color component transformation, if applicable, followed by code stream truncation).</para>
              </listitem>
            </orderedlist>
            <para xml:id="para_f08e27a7-a490-4cdc-a5a4-ca1e66cc5d64">The choice reversible versus irreversible is at the discretion of the sender (SCU or FSC/FSU).</para>
            <note>
              <para xml:id="para_6b41cb28-8c1f-4d95-9220-15598a2533e0">When using the irreversible wavelet transformation and an irreversible color component transformation, if applicable, even if no quantization is performed, some loss will always occur due to the finite precision of the calculation of the wavelet and multi-component transformations.</para>
            </note>
          </listitem>
        </orderedlist>
        <para xml:id="para_525df055-a70d-441c-b1fa-c99e89a05d2c">Only the features defined in JPEG 2000 Part 1 (<xref linkend="biblio_ISOIEC15444-1"/>) are permitted for these two Transfer Syntaxes. Additional features and extensions that may be defined in other parts of JPEG 2000 shall not be included in the compressed bitstream unless they can be decoded or ignored without loss of fidelity by all Part 1 compliant implementations.</para>
        <para xml:id="para_6a575c82-97de-48ef-9064-e55217019356">If the object allows multi-frame images in the pixel data field, then for these JPEG 2000 Part 1 Transfer Syntaxes, each frame shall be encoded separately. Each fragment shall contain encoded data from a single frame.</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_7233e21f-66f7-4341-9f8a-17fbd72047dd">That is, the processes defined in <xref linkend="biblio_ISOIEC15444-1"/> shall be applied on a per-frame basis. The proposal for encapsulation of multiple frames in a non-DICOM manner in so-called "Motion-JPEG" or "M-JPEG" defined in 15444-3 is not used.</para>
            </listitem>
            <listitem>
              <para xml:id="para_55f854ab-dee3-49b8-aac8-5436a43fb5ee">Though a fragment may not contain encoded data from more than one frame, the encoded data from one frame may span multiple fragments. See note in <xref linkend="sect_8.2" xrefstyle="select: label"/>.</para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_bcca5500-399e-49de-af34-42a0062b49c5">For all images, including all frames of a multi-frame image, the JPEG 2000 bitstream specified in <xref linkend="biblio_ISOIEC15444-1"/> shall be used. The optional JP2 file format header shall NOT be included.</para>
        <note>
          <para xml:id="para_46c6bcc3-6c74-4385-bb7d-5d3edfeb0f86">The role of the JP2 file format header is fulfilled by the non-pixel data attributes in the DICOM Data Set.</para>
        </note>
        <para xml:id="para_50ce9107-5cf2-43ff-a7f5-371626bc3094">The International Standards Organization ISO/IEC JTC1 has also developed JPEG 2000 Part 2 (<xref linkend="biblio_ISOIEC15444-2"/>), which includes Extensions to the compression techniques described in Part 1 of the JPEG 200 Standard. Annex J of JPEG 2000 Part 2 describes extensions to the ICT and RCT multiple component transformations allowed in Part 1. Two types of multiple component transformations are defined in Annex J of Part 2 of JPEG 2000:</para>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_36180de9-4f92-412d-8c8e-b7ded7c249d3">Array based multiple component transforms that form linear combinations of components to reduce the correlation between components. Array based transforms include prediction based transformations such as DPCM as well as more complicated transformations such as the KLT. These array based transformations can be implemented reversibly or irreversibly.</para>
          </listitem>
          <listitem>
            <para xml:id="para_b114d630-72bb-4708-8768-d130794baf0b">Wavelet based multiple component transformations using the same two wavelet filters as used in Part 1 of JPEG 2000 (5-3 reversible wavelet and 9-7 irreversible wavelet).</para>
          </listitem>
        </orderedlist>
        <para xml:id="para_abb32b74-82ad-4620-9945-c725785decde">Annex J of JPEG 2000 Part 2 also describes a flexible mechanism to allow these techniques to be applied in sequence. Furthermore, it provides mechanisms that allow components to be re-ordered and grouped into component collections. Different multiple component transformation can then be applied to each component collection.</para>
        <para xml:id="para_fb5e7612-d1ee-444d-b5cc-6eb46c0fff1d">Two additional Transfer Syntaxes are specified for Part 2 JPEG 2000:</para>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_1bbb145c-dcfb-4e18-a4cd-8ddb1610a6f1">A Transfer Syntax with a UID of 1.2.840.10008.1.2.4.92, which specifies the use of the lossless (reversible) mode of JPEG 2000 Part 2 (<xref linkend="biblio_ISOIEC15444-2"/>) multiple component transformation extensions, as defined in Annex J of JPEG 2000 Part 2 (i.e., the use of a reversible wavelet transformation and a reversible multiple component transformation, and no quantization or code stream truncation).</para>
          </listitem>
          <listitem>
            <para xml:id="para_523aa425-ad50-4025-9ff8-08f936a45836">A Transfer Syntax with a UID of 1.2.840.10008.1.2.4.93, which specifies the use of either:</para>
            <orderedlist numeration="loweralpha">
              <listitem>
                <para xml:id="para_77d4f2ff-63fd-4e5e-8335-c8f1a3198a15">the lossless (reversible) mode of JPEG 2000 Part 2 (<xref linkend="biblio_ISOIEC15444-2"/>) multiple component transformation extensions, as defined in Annex J of JPEG 2000 Part 2 (i.e., the use of a reversible wavelet transformation and a reversible multiple component transformation, and no quantization), or</para>
              </listitem>
              <listitem>
                <para xml:id="para_949ef798-bf33-4702-96c8-9645e7ea6854">the lossy (irreversible) mode of JPEG 2000 Part 2 (<xref linkend="biblio_ISOIEC15444-2"/>) multiple component transformation extensions, as defined in Annex J of JPEG 2000 Part 2 (i.e., the use of an irreversible wavelet transformation and an irreversible multiple component transformation, and optionally quantization, or the use of an reversible wavelet transformation and a reversible multiple component transformation, followed by code stream truncation).</para>
              </listitem>
            </orderedlist>
          </listitem>
        </orderedlist>
        <para xml:id="para_60ff91a3-5ce4-4b1d-b0e9-647cc0b4a2c3">Only the multiple component transformation extensions defined in Annex J of JPEG 2000 Part 2 (<xref linkend="biblio_ISOIEC15444-2"/>) are permitted for these two Transfer Syntaxes. Additional features and extensions that may be defined in other Annexes of JPEG 2000 Part 2 shall not be included in the compressed bitstream.</para>
        <note>
          <para xml:id="para_e69aa706-2eb4-4225-a499-482bd9cfef68">the arbitrary wavelet transformations, as defined in Annex H of JPEG 2000 Part 2 (<xref linkend="biblio_ISOIEC15444-2"/>) are not allowed for these two Transfer Syntaxes. The only wavelet transformations that are allowed to be used as multiple component transformations are the reversible 5-3 wavelet transformation and the irreversible 9-7 wavelet transformation, as defined in Annex F of JPEG 2000 Part 1 (<xref linkend="biblio_ISOIEC15444-1"/>).</para>
        </note>
        <para xml:id="para_bf909fe2-4e37-4c6b-bbed-f7b5d17fb4b1">If the object allows multi-frame images in the pixel data field, then, for these JPEG 2000 Part 2 Transfer Syntaxes, the frames in the object are first processed using the multi-component transformation. After the multiple component transformation has been applied, the transformed frames are encoded using the process described in JPEG 2000 Part 1.</para>
        <para xml:id="para_989f22fc-145e-46a9-910a-f058ae903c4c">Optionally, the frames can be grouped into one or more component collections. The multiple component transformations are then applied to each component collection independently. The use of component collections can be used to reduce computational complexity and to improve access to specific frames on the decoder. If component collections are used, each fragment shall contain encoded data from a single component collection.</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem>
              <para xml:id="para_304f83b1-70ff-412d-9b90-d4beaec20e83">The 3<superscript>rd</superscript> dimension transformations that are described in this Supplement are treated in Part 2 of JPEG 2000 as direct extensions to the color component transformations (RGB to YUV) that are described in Part 1 of JPEG 2000. For this reason, each image or frame in the sequence is called a "component". Although the term component is used as a generic term to identify an element of the 3<superscript>rd</superscript> dimension, no restriction is made or implied that the transformations in this Supplement apply only to multi-component (or multiple color channel) data. To compress a volumetric data set using this Transfer Syntax, each frame of the DICOM image is treated as a component of a multi-component image.</para>
            </listitem>
            <listitem>
              <para xml:id="para_372c4bc2-fdd0-475a-904a-aad205b8320d">The progressive nature of the JPEG 2000 code stream allows for the decompression of the image before the complete image has been transferred. If a Storage SCP truncates the code stream by aborting the association, the instance has not been completely transferred and hence should not persist unless different UIDs are assigned (even though it may have been transiently used for display purposes).</para>
            </listitem>
            <listitem>
              <para xml:id="para_42e65f47-4a8b-400f-b5b3-b3ae86921527">It has been shown that the use of component collections does not significantly affect the compression efficiency (for details, see <link xl:href="http://medical.nema.org/Dicom/minutes/WG-04/2004/2004-02-18/3D_compression_RSNA_2003_ver2.pdf"/>).</para>
            </listitem>
            <listitem>
              <para xml:id="para_95528047-a116-4b3c-970f-4c2b2457866b">Though a fragment may not contain encoded data from more than one component collection, the encoded data from one component collection may span multiple fragments.</para>
            </listitem>
          </orderedlist>
        </note>
      </section>
      <section label="A.4.5" status="3" xml:id="sect_A.4.5">
        <title>MPEG2 Image Compression</title>
        <para xml:id="para_3eba3581-f657-4587-8f95-8650f5249d78">The International Standards Organization ISO/IEC MPEG2 has developed an International Standard, <xref linkend="biblio_ISOIEC13818-2"/> 'Information Technology - Generic coding of moving pictures and associated audio information: video -- part 2', referred to as "MPEG-2".</para>
        <para xml:id="para_c0b400f3-aec8-4109-80dc-791633140232">A DICOM Transfer Syntax for MPEG2 Image Compression shall be identified by a UID value of either:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_5fae2fb4-e642-40de-b57f-80ffe546020b">1.2.840.10008.1.2.4.100 corresponding to MPEG2 Main Profile / Main Level option of the ISO/IEC MPEG2 Video standard</para>
          </listitem>
          <listitem>
            <para xml:id="para_e008514a-6a97-4d12-ab09-e702e7368093">1.2.840.10008.1.2.4.101 corresponding to the MPEG2 Main Profile / High Level option of the ISO/IEC MPEG2 Video standard.</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="A.4.6" status="2" xml:id="sect_A.4.6">
        <title>MPEG-4 AVC/H.264 High Profile / Level 4.1 Video Compression</title>
        <para xml:id="para_f58b29a0-d09e-44df-9fc1-ddcbf283db5a">The International Standards Organization ISO/IEC MPEG4 has developed an International Standard, <xref linkend="biblio_ISOIEC14496-10"/> (MPEG-4 Part 10), for the video compression of generic coding of moving pictures and associated audio information. This standard is jointly maintained and has identical technical content as the ITU-T H.264 standard.</para>
        <para xml:id="para_e3741454-6f7f-488a-98e9-0cca205dd497">A DICOM Transfer Syntax for MPEG-4 AVC/H.264 Image Compression shall be identified by a UID value of either:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_d2a74ea5-fce5-4810-b874-9e501cbd1f85">1.2.840.10008.1.2.4.102 corresponding to the MPEG-4 AVC/H.264 High Profile / Level 4.1 of the ITU-T H.264 Video standard</para>
          </listitem>
          <listitem>
            <para xml:id="para_b1f5554c-4943-48f0-b891-1216547478dc">1.2.840.10008.1.2.4.103 corresponding to the MPEG-4 AVC/H.264 BD-compatible High Profile / Level 4.1 of the ITU-T H.264 Video standard with the temporal and spatial resolution restrictions defined in Table 8-4 <!--<xref linkend="table_8-4" xrefstyle="select: label"/>-->.</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="A.4.7" status="2" xml:id="sect_A.4.7">
        <title>MPEG-4 AVC/H.264 High Profile / Level 4.2 Video Compression</title>
        <para xml:id="para_f250018c-01d5-4180-8279-1f792cc8b654">The International Standards Organization ISO/IEC MPEG4 has developed an International Standard, <xref linkend="biblio_ISOIEC14496-10"/> (MPEG-4 Part 10),
          for the video compression of generic coding of moving pictures and associated audio information. This standard is jointly maintained
          and has identical technical content as the ITU-T H.264 standard.</para>
        <para xml:id="para_998a7c3a-d89e-4540-8a5f-a970bcc533c0">A DICOM Transfer Syntax MPEG-4 AVC/H.264 High Profile / Level 4.2 for 2D Image Compression shall be identified by a UID value of:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_c8faa9b3-a055-433a-b1fc-bb33c093d329">1.2.840.10008.1.2.4.104 corresponding to the MPEG-4 AVC/H.264 High Profile / Level 4.2 of the ITU-T H.264 Video standard
              with the restriction that frame packing for stereoscopic 3D content shall not be used as defined in
              Table 8-8<!--<xref linkend="table_8-8" xrefstyle="select: label"/>-->.</para>
          </listitem>
        </itemizedlist>
        <para xml:id="para_5c446da5-2c9a-4ebf-987c-2f3725d80fe8">A DICOM Transfer Syntax MPEG-4 AVC/H.264 High Profile / Level 4.2 for 3D Image Compression shall be identified by a UID value of:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_561b9e67-0279-4bcf-bb6a-58f392b0b865">1.2.840.10008.1.2.4.105 corresponding to the MPEG-4 AVC/H.264 High Profile / Level 4.2 of the ITU-T H.264 Video standard.
              It should be used for transmitting stereoscopic 3D content with frame packing formats as defined in Table 8-8<!--<xref linkend="table_8-8" xrefstyle="select: label"/>-->.</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="A.4.8" status="2" xml:id="sect_A.4.8">
        <title>MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2 Video Compression</title>
        <para xml:id="para_f0c12be8-7e68-4d13-853c-8ec3d2dd3c84">The International Standards Organization ISO/IEC MPEG4 has developed an International Standard, <xref linkend="biblio_ISOIEC14496-10"/> (MPEG-4 Part 10),
          for the video compression of generic coding of moving pictures and associated audio information.
          This standard is jointly maintained and has identical technical content as the ITU-T H.264 standard.</para>
        <para xml:id="para_fd72e768-7953-42e2-8bad-e750becf724f">A DICOM Transfer Syntax for MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2 Image Compression shall be identified by a UID value of:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_1b8e40d3-a935-4258-b8e8-a8e22af8110b">1.2.840.10008.1.2.4.106 corresponding to the MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2 of the ITU-T H.264 Video standard.</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="A.4.9" status="3" xml:id="sect_A.4.9">
        <title>HEVC/H.265 Main Profile / Level 5.1 Video Compression</title>
        <para xml:id="para_f63fecff-47fe-4e4f-a706-1bd9ebf4f6c5">The International Standards Organization ISO/IEC MPEG has developed an International Standard, <xref linkend="biblio_ISOIEC23008-2"/> (HEVC), for the video compression of generic coding of moving pictures and associated audio information. This standard is jointly maintained and has identical technical content as the <xref linkend="biblio_ISOIEC23008-2"/> HEVC standard.</para>
        <para xml:id="para_b28f5b2d-734a-48ec-ad9d-1aed38bee69f">A DICOM Transfer Syntax for HEVC/H.265 Main Profile / Level 5.1 Image Compression shall be identified by a UID value of:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_6d3ecb08-d93f-47e6-a3c4-13aa0815db57">1.2.840.10008.1.2.4.107 corresponding to the HEVC/H.265 Main Profile / Level 5.1 of the <xref linkend="biblio_ISOIEC23008-2"/> HEVC Video standard.</para>
          </listitem>
        </itemizedlist>
      </section>
      <section label="A.4.10" status="3" xml:id="sect_A.4.10">
        <title>HEVC/H.265 Main 10 Profile / Level 5.1 Video Compression</title>
        <para xml:id="para_b28b604d-591f-4633-adc2-eb17b8a1b2c0">The International Standards Organization ISO/IEC MPEG has developed an International Standard, <xref linkend="biblio_ISOIEC23008-2"/> (HEVC), for the video compression of generic coding of moving pictures and associated audio information. This standard is jointly maintained and has identical technical content as the <xref linkend="biblio_ISOIEC23008-2"/> HEVC standard.</para>
        <para xml:id="para_ac367f71-0941-4c74-8442-501bd7e79ef2">A DICOM Transfer Syntax for HEVC/H.265 Main 10 Profile / Level 5.1 Image Compression shall be identified by a UID value of:</para>
        <itemizedlist>
          <listitem>
            <para xml:id="para_6398d205-00c7-47fd-9f4a-4dcf987f6263">1.2.840.10008.1.2.4.108 corresponding to the HEVC/H.265 Main 10 Profile / Level 5.1 of the <xref linkend="biblio_ISOIEC23008-2"/> HEVC Video standard.</para>
          </listitem>
        </itemizedlist>
      </section>
    </section>
  </section>
      
      <section>
        <title/>
        <informaltable rules="all" frame="box">
          <tr>
            <td><para><emphasis role="italic">Amend DICOM PS 3.11:</emphasis></para></td>
          </tr>
        </informaltable>
      </section>
        
        <section label="C" status="1" xml:id="chapter_C">
    <title>Ultrasound Application Profile (Normative)</title>
    <section label="C.3" status="2" xml:id="sect_C.3">
      <title>General Class Profile</title>
      <section label="C.3.1" status="3" xml:id="sect_C.3.1">
        <title>Abstract and Transfer Syntaxes</title>
        <section label="C.3.1.1" status="4" xml:id="sect_C.3.1.1">
          <title>Ultrasound Single and Multi-frame Pixel Formats Supported</title>
          <para xml:id="para_598408ae-1b4d-402c-998a-13c07d260fec">The STD-US application profile requires that all ultrasound image objects only be stored using the values described in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>--> US Image Module and the specializations used for the Ultrasound Single and Multi-Frame IODs.</para>
          <para xml:id="para_816c905b-ecef-46b7-9e4c-d9d4b635b0f3">In the role of FS-Updater or FS-Creator the application can choose any of the supported Photometric Interpretations described in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>--> US Image Module to create an IOD. In the role of FS-Reader, an application shall support all Photometric Interpretations described in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>--> US Image Module.</para>
          <para xml:id="para_3ba1db63-5c64-4c50-a7e2-745423353760">
<xref linkend="table_C.3-2" xrefstyle="select: label"/> describes restrictions on the use of various Transfer Syntaxes with the supported Photometric Interpretations for both single and multi-frame images.</para>
          <table frame="box" label="C.3-2" rules="all" xml:id="table_C.3-2">
            <caption>Defined Photometric Interpretation and Transfer Syntax Pairs</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f21a76af-6ead-48c0-a14e-787e25a01f49">
                    <emphasis role="bold">Photometric Interpretation Value</emphasis>
                  </para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_148958d5-632d-4c8a-8e67-1177d05f9037">
                    <emphasis role="bold">Transfer Syntax</emphasis>
                  </para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3b6ca0a9-363c-4276-a85b-a04adb2b258f">
                    <emphasis role="bold">Transfer Syntax UID</emphasis>
                  </para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ab5533f2-1ca6-4a80-8f74-84f5fe06ef38">MONOCHROME2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cb7cd602-5ee1-465a-bda0-7f1d073e90e6">Uncompressed</para>
<para xml:id="para_f8e87ae4-75f1-45c4-9e0b-5b699272ab25">RLE Lossless Image Compression</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_54e9a5fe-f87e-41f6-a533-a84f31b9959f">1.2.840.10008.1.2.1</para>
<para xml:id="para_b0c77b38-ab38-4cd1-b74f-5ca5b622bf3e">1.2.840.10008.1.2.5</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_df20ab76-869c-49f9-9cbe-f005f3661bdc">RGB</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7869ca6e-aa21-4a0b-8b82-0b02d26109bc">Uncompressed</para>
<para xml:id="para_1ffe6de5-d7cb-4463-8448-b71c2f6458d3">RLE Lossless Image Compression</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5558ab0c-5813-42ef-a866-e2a7751174bb">1.2.840.10008.1.2.1</para>
<para xml:id="para_17a281c2-2552-44d4-ba5e-b8dcc401df7b">1.2.840.10008.1.2.5</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0f5f5f2d-378a-4f1b-8d84-397dc66fe779">PALETTE COLOR</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ee5c3414-0cac-4a59-b368-16d1a07edb68">Uncompressed</para>
<para xml:id="para_87109394-4f34-4a1a-9f73-81a7fa5462af">RLE Lossless Image Compression</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f9d3564e-03d0-4ab4-bc13-43ac32890918">1.2.840.10008.1.2.1</para>
<para xml:id="para_f44352b4-fb90-4859-a700-2276df45d809">1.2.840.10008.1.2.5</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c6c9b008-5c4e-4f4f-8a1a-7f0238125e2d">YBR_FULL</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8100adcc-fc89-430b-9967-38870d6b25ef">RLE Lossless Image Compression</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d1b508ed-2343-4ce7-b7bd-659166e1bddf">1.2.840.10008.1.2.5</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9ddd6911-e5ab-4954-9eed-51e2b3fb8fc4">YBR_FULL_422</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9f3703cc-e159-49e6-9897-657b1d1a22a3">Uncompressed</para>
<para xml:id="para_c6d0c3e0-fb1b-4d90-bb11-3aed6795aa2b">JPEG Lossy</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a87b5c44-8d0f-4cc2-b731-426c05bc5bdc">1.2.840.10008.1.2.1</para>
<para xml:id="para_af9c3eb3-8747-4711-9b3b-764f7ed00019">1.2.840.10008.1.2.4.50</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8d458109-b4af-4527-8643-9676d45f4d95"><emphasis role="bold"><emphasis role="strikethrough">YBR_PARTIAL_422</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d21ae9d7-3b04-4e9f-b1a5-9381b18ce91b"><emphasis role="bold"><emphasis role="strikethrough">Uncompressed</emphasis></emphasis></para>
                  <para xml:id="para_e947205f-66a2-4d7d-9e97-abe06b6573d3"><emphasis role="bold"><emphasis role="strikethrough">JPEG Lossy</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_71cb8ba4-eaae-4eda-a2e1-2e26ced8d097"><emphasis role="bold"><emphasis role="strikethrough">1.2.840.10008.1.2.1</emphasis></emphasis></para>
                  <para xml:id="para_45d2bcac-dda1-4a67-a4f8-89cc82b936a6"><emphasis role="bold"><emphasis role="strikethrough">1.2.840.10008.1.2.4.50</emphasis></emphasis></para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      </section>
    </section>
  </section>
    </chapter>
</book>