    <title>CP-1646</title>
    <subtitle>Add repeat flag and reason to RDSR to help with outlier analysis</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2017/01/24</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>2016/06/10</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1646</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Add repeat flag and reason to RDSR to help with outlier analysis</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>Repetition of part of a CT study for reasons like patient motion or suboptimal contrast timing have been identified
                        as reasons for radiation dose exposure being higher than usual for a procedure
                        [Chen YA, MacGregor K, Li I, Concepcion L, Deva DP, Dowdell T, et al. Tracking and Resolving CT Dose Metric Outliers Using Root-Cause Analysis. Journal of the American College of Radiology. 2016 Jun;13(6):680–7. Available from <link xl:href="http://dx.doi.org/10.1016/j.jacr.2016.01.017"/>.].</para>
                    <para>Knowledge of this information captured at the time of scanning would potentially reduce the number of outliers that needed to be investigated.</para>
                    <para>It is proposed to add a per-event flag and reason to record this.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
        
        <section label="TID 10013" status="3" xml:id="sect_TID_10013">
            <title>CT Irradiation Event Data</title>
            <para xml:id="para_c376d4ec-e7f4-4e4e-a47e-b518930c6ba4">...</para>
            <table frame="box" label="TID 10013" rules="all" xml:id="table_TID_10013">
               <caption>CT Irradiation Event Data</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a7752bbc-ea79-405e-84fb-68cf449ca41f"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_77545633-84cb-4c30-877d-b8b1609a8c15">
                           <emphasis role="bold">NL</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e540d0c1-1f91-453b-b341-d86ce89fec6f">
                           <emphasis role="bold">Rel with Parent</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1b33952f-26eb-4779-a3cc-1f4ebffe6d43">
                           <emphasis role="bold">VT</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ab0c21a6-5fd7-4cdb-800a-f1232806b301">
                           <emphasis role="bold">Concept Name</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9b16203e-dc48-416d-a9da-3aa88eb23fbb">
                           <emphasis role="bold">VM</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3882c12d-d555-4398-9e32-9969038eb012">
                           <emphasis role="bold">Req Type</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a796eeb0-27ec-4010-9f92-04caa56775de">
                           <emphasis role="bold">Condition</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_530b8609-1d9b-448d-b2e9-dd01312d2b58">
                           <emphasis role="bold">Value Set Constraint</emphasis>
                        </para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_37d66cae-ac15-4b07-a564-aad3fca78280">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b19e8fe0-71d4-4740-8c8e-c14ff589b100"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_079ae4f7-6d50-4e78-b390-5bd5a7b100bf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e379aba8-6ac2-4ee5-a85d-4682ce928889">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_071b66b4-c54d-4939-85fd-44eb0b451d27">EV <link linkend="DCM_113819" xrefstyle="select: nopage">(113819, DCM, "CT Acquisition")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b193acb3-32b2-49af-a822-cd59698f4c0d">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a69920e4-2d8b-4f27-96f6-5be3ce35fa5a">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bc338618-9934-4bf4-871a-a93dfc063042"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_296cda33-568d-4db6-91a8-8c55dd97b27a"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c1323f26-ba53-4ac0-9a87-3fe78210d743">2</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7e246dd6-cfdd-44b8-95fd-fc690b2b2cc3">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_00ca9f4d-f4d3-454d-92ea-db03fb2a1523">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_06d64d28-9bb3-4ec4-9029-a48a80114dca">TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c9fc37dc-8a18-4f8e-9637-6bc99a6513b5">EV <link linkend="DCM_125203" xrefstyle="select: nopage">(125203, DCM, "Acquisition Protocol")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_727f5404-f1c3-45fc-bcdb-bb9f54244e90">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cb6273b7-bd72-4fe6-b4f4-eeebab26a03f">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dbed400e-5a71-47e1-87d5-ebbacb0fa874"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_da3dacb3-7a45-4cb4-a93e-e52a88868d38"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ce7a1720-2531-424f-bd50-6c096ab697b7">3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ea27272c-0b56-4846-8c0e-920a1cc5e6d2">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2908dc1f-5ccb-4768-8d97-122f290b0a70">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d44b21d0-76bd-47f0-9c41-484930da930f">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_484b8875-11a3-41b4-8a82-cc05d2661ec4">EV <link linkend="DCM_123014" xrefstyle="select: nopage">(123014, DCM, "Target Region")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_60e7aaaf-d45d-4650-bb11-5edb9397db22">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5804f4d0-6822-4b3b-8dc5-e043e6ba272e">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8b999d2a-4c84-4528-9961-d3d8d6c5c179"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ef1d1a75-62db-4f04-bff4-a8eeadfc1672">D<xref linkend="sect_CID_4030" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cb5fb4e8-6f21-4ba3-9754-ab6c61c2979c">4</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_616eb3a5-66ff-46e0-95af-a61d1a9557fd">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7690f1af-d5e8-43a5-a0f4-4c11f50ab8b2">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_45a89d5e-e05f-4e7e-891a-1cd658fc9596">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_df664e53-7ee7-4053-b3fd-ac58d4013558">EV <link linkend="DCM_113820" xrefstyle="select: nopage">(113820, DCM, "CT Acquisition Type")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_03efc72e-baa2-4050-92ab-0ee8388d9532">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c0d70231-7c37-4bbf-a475-08c8839fd8cc">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8d373077-114a-4ba6-833e-25bf70885e88"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_95d44271-3b73-444d-9131-6286fdff8deb">D<xref linkend="sect_CID_10013" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8403fdfe-e87d-4d76-9ba1-a12b0db5c630">4b</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_23d985b2-6f5d-469a-9694-fb391079f430">&gt;&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c0f4f6ba-cbde-4ea5-bd2b-9f36c3e3478b">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5c940492-5a70-4ff3-8e92-941cc418f799">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_46295279-8a10-46a5-ab2f-150c98ae302b">EV <link linkend="DCM_113961" xrefstyle="select: nopage">(113961, DCM, "Reconstruction Algorithm")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_31740d4a-14c0-48cf-94ee-2b2e4487441d">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_098e3e38-5494-4059-a77d-36305f0a1725">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f94722eb-1c93-419c-be56-c5099a4db159"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a26a1525-2945-41fc-b4ea-d6beb0e1b2ed">D<xref linkend="sect_CID_10033" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9e7a8409-a28b-4ec4-9c45-397c233da575">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f956eeb5-d4eb-493b-809f-f20b249a629b">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4efdd4d7-7bf3-4b33-a95c-0022aa7bc8af">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1de301ee-9bd1-4b93-998c-b8c677f0aa73">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_21743ba8-4820-4a91-9afa-eba1df6ce279">EV (G-C32C, SRT, "Procedure Context")</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1cb88b58-2364-4a75-a58c-aaee02b414e2">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1a3ed39e-85aa-46d1-9e5f-ae0b5f36db24">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7bb27281-88c8-4a5f-b9b5-a67296eb7962"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cf711534-0f0a-4431-98d1-2626802a0400">D<xref linkend="sect_CID_10014" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d769162e-d348-4129-84ff-d634ce7c200b">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ff0e6c1e-505d-4270-ba06-efccc300146e">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_18fcd71b-442b-47f9-aa9b-cbbca4c22b80">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9d72adbe-d5d2-4bc4-b541-e7e96de73f2e">UIDREF</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5b195f70-00c0-4f04-85ac-04df2efdbb94">EV <link linkend="DCM_113769" xrefstyle="select: nopage">(113769, DCM, "Irradiation Event UID")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_65ef3611-da36-413b-b824-c6e9dbaf80cc">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e9b9238a-333d-4f0c-9a21-53bb8f6f1258">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7598ff1f-a7cf-4859-8261-15cf16f6bee1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_33a9beb0-0769-4049-9376-8cd120a263e4"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_31826003-0360-4098-be05-555a4eb8cd45">6b</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_94f2fe4e-49ef-441c-bef9-dfbb83b70fd8">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9a782f11-82f8-43cb-ab4a-7efa14d24320">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8aa8b4c8-1f6e-450c-af30-35bebae4058f">TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_872b987d-a055-47a2-9c7a-46f087c7be5c">EV <link linkend="DCM_113605" xrefstyle="select: nopage">(113605, DCM, "Irradiation Event Label")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ce0bac29-e1f9-49c8-933b-e81f37cf1fe8">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_adc1abb4-7253-4225-8276-826fa2069f37">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4dbbac75-e2d7-4541-a312-ab5c33c4a027"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6b73da9a-b87d-41e7-b1ad-305a68f53482"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b9b716e9-225b-4897-bc78-4f3c5a508d07">6c</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d6c1845f-23bf-4adc-b666-9433b8873bb1">&gt;&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b3dd93a1-403a-4080-8326-30f4aab0abcf">HAS CONCEPT MOD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_32dfcbd1-8a6a-419d-a1cd-541f206101db">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d236e92f-78c6-4fff-b12f-982d7d7c3e15">EV <link linkend="DCM_113606" xrefstyle="select: nopage">(113606, DCM, "Label Type")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d6031bdf-489a-45f5-abf3-0c91b3f0a951">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0166f2a6-6f36-42c7-814f-ae8b04a24d95">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e3798a3d-3394-4685-b4ef-0ac5d1d9f36f">IF the value of Row 6b is the value of an Attribute in the images.</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c2b8e10f-c696-4b7f-84c6-29f76c2e508b">D<xref linkend="sect_CID_10022" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">6d</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">TEXT</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd001" xrefstyle="select: nopage">(ddd001, DCM, "Is Repeated Acquisition")</link>
                                    </emphasis></emphasis></para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                                </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
                                </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para>D<xref linkend="sect_CID_230" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
                       </td>
                   </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="underline">
                                    <para>6e</para>
                                </emphasis></emphasis></td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">HAS CONCEPT MOD</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">EV <link linkend="DCM_ddd002" xrefstyle="select: nopage">(ddd002, DCM, "Reason for Repeating Acquisition")</link>
                                    </emphasis></emphasis></para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
                                </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">M</emphasis></emphasis></para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para><emphasis role="bold"><emphasis role="underline">D<xref linkend="sect_CID_ccc01" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis>
                            </para>
                       </td>
                   </tr>
                   <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1b2974de-ab03-40da-abec-bed89494c908">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_610ab462-7075-4461-803b-ce45ead3c4ef">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_68268976-c1a4-4263-bc33-59691bfb458d">CONTAINS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eb76118e-d763-46ee-95d0-066d29620474">CONTAINER</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_83eb8d0e-a270-4c5a-8c9c-5f35b3e6c898">EV <link linkend="DCM_113822" xrefstyle="select: nopage">(113822, DCM, "CT Acquisition Parameters")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f8d38d87-2df0-45ac-9a0a-d46f0e28bbe3">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_65c007c9-4fa7-4cf6-a73b-ab81f7f60de3">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9fc0ca24-c99f-4dd8-937b-f5f70132d550"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_16271ae8-c83d-49dd-bd28-a15afc519d1f"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc0f2716-9042-4d24-bd4f-8ccfd2af37db">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_38b99da7-cdbe-484c-b2d2-040383b5c28b">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e2ca966d-bc1d-433a-a066-923cf443f3aa">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ba53b1fb-2f69-4ae7-ae45-d0e921f4c771">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6b45850d-ab8b-4e72-a8a7-01814502674e">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_346f202f-a5be-4efb-b826-4ef0ec7c6802"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_67ce948a-85b8-4ab7-b9e1-bbda7fedb74e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eab14c09-7401-4e4b-ad74-0db6cc97136c">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c3b8ae4d-cede-4a1f-bd8d-0008be3f7880"/>
                     </td>
                  </tr>
               </tbody>
            </table>
            <para xml:id="para_2cd2d4de-81b2-44af-abd1-19de63403d5b">
               <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_af32ae80-5455-41b0-b63e-d6737b07f956"><emphasis role="bold"><emphasis role="underline">Row 6d, 6e</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_9325a21a-9bf8-4122-830e-2a3e32bb0b2f"><emphasis role="bold"><emphasis role="underline">If an acquisition is a repeat because an earlier acquisition was unsatisfactory, this may be recorded along with a coded reason. This is intended to help with subsequent analysis by providing a priori information about why the study might be flagged as an outlier with higher dose exposure values than usual for the type of study.</emphasis></emphasis></para>
                     </td>
                  </tr>
               </tbody>
            </informaltable>
         </section>
        
        <section label="CID ccc01" status="3" xml:id="sect_CID_ccc01">
            <title>Reason for Repeating Acquisition</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_b016a9fe-85e1-40d6-9881-e952542253d2">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para>
                            <emphasis role="bold">yyyymmdd</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID ccc01" rules="all" xml:id="table_CID_ccc01">
                <caption>Reason for Repeating Acquisition</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Coding Scheme Designator</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_ddd003" xrefstyle="select: nopage">ddd003</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Patient motion</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para>DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <link linkend="DCM_ddd004" xrefstyle="select: nopage">ddd004</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Suboptimal contrast timing</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="CID 6135" status="3" xml:id="sect_CID_6135">
            <title>Image Quality Finding</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_5f0de022-86d6-4074-a5f2-f61c75b0c58e">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_39155402-0518-4645-a934-ce54c9e9387f">
                            <emphasis role="bold">20090402</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 6135" rules="all" xml:id="table_CID_6135">
                <caption>Image Quality Finding</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d7655813-b8ea-4d70-a0d1-168becbc4efe">
                                <emphasis role="bold">Coding Scheme Designator</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e848a6fb-3358-45a4-8ce1-d67c625ab6e5">
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_17e48fe5-c4d9-4493-9edb-ffd336488893">
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_22aa2727-8c4e-405d-be16-700ef120d255">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d0a0ebbd-a81e-4cbd-85d9-864d4e93cc8c">
                                <link linkend="DCM_111208" xrefstyle="select: nopage">111208</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b9209125-cc90-4c89-88f6-970a0b3213b0">Grid artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6b4e483a-5d74-4f86-be8d-56cf75bee17d">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_faaebc97-df6b-493d-a9c7-f8d4e34ca659">
                                <link linkend="DCM_111209" xrefstyle="select: nopage">111209</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3262c677-722e-4ecc-988c-7eef6fac8382">Positioning</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e3359afd-4d97-4381-a3ce-a46270164372">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c17c459c-dbe5-4209-b830-1ca1fb67184c">
                                <link linkend="DCM_111210" xrefstyle="select: nopage">111210</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_62e97b15-a353-4704-8710-14afbc7198c9">Motion blur</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6b8a0dd9-870a-464b-ae52-50b0ba9732ca">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_366c8e4c-5b3b-4eef-a520-3c6aa8863844">
                                <link linkend="DCM_111211" xrefstyle="select: nopage">111211</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d5d946e5-5e6b-4da2-a980-8b151fc1b739">Under exposed</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3d0fe38e-2b45-45dc-b404-63e38a608045">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b1f8eb39-855c-4248-bdca-487eac3f15ac">
                                <link linkend="DCM_111212" xrefstyle="select: nopage">111212</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3fd6f01f-244c-4578-bd20-59fa6b9ffa09">Over exposed</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_32385fe0-4037-475c-bc80-7aa223462371">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d3af1996-91d3-4d34-80de-5de274f50a30">
                                <link linkend="DCM_111213" xrefstyle="select: nopage">111213</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_cff91847-15cc-4c10-9e1b-a8acb6961031">No image</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0a51778-1d41-4f9f-bac6-d83c872ddd69">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ea95b880-2bfe-4e89-b8fa-4c2f7c629b3c">
                                <link linkend="DCM_111214" xrefstyle="select: nopage">111214</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b0e4d21e-818c-4231-a01f-b74d815107ad">Detector artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f144022d-2c7e-441a-8fab-88ee817950b5">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fd8f8e7b-6f35-4b9d-be13-3c57a3c86bde">
                                <link linkend="DCM_111215" xrefstyle="select: nopage">111215</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7d2dd93c-5180-4d49-8f42-d27e910a532f">Artifact(s) other than grid or detector artifact</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ae19cacd-2780-4008-b7c2-a268c418463f">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_14cb2a94-595d-49fb-a30e-a104898b5f08">
                                <link linkend="DCM_111216" xrefstyle="select: nopage">111216</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a6136197-c885-4e45-bd17-b8e31be771b3">Mechanical failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3297cacd-fbb2-4a17-af48-7d6fa794fbbf">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_11c88669-c7bc-4fb5-9143-5f6249e15148">
                                <link linkend="DCM_111217" xrefstyle="select: nopage">111217</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bf8beaf6-32d8-4745-a256-12860f20777b">Electrical failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9adc3b55-6199-4392-b7f8-b1f20ee64f7f">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_57660e02-3a5f-4df9-8fc4-5de5c4bed576">
                                <link linkend="DCM_111218" xrefstyle="select: nopage">111218</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d0f36b4c-07c3-4c6b-9f6b-4039d5c5ac35">Software failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_801c8ccd-99f8-4a8a-8d4d-df305b7ef500">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7adfbe54-8503-467b-8773-a0e4b67121a9">
                                <link linkend="DCM_111219" xrefstyle="select: nopage">111219</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_43e59197-6982-4024-bbb1-f939f05696d2">Inappropriate image processing</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7dc76dc3-66d7-4c43-808d-9f8bb398c301">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_40835559-4685-48e6-8c33-abee86ba36b3">
                                <link linkend="DCM_111220" xrefstyle="select: nopage">111220</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f3d9f8b5-353e-48dd-a5f2-4903d306c484">Other failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_273d1d03-28d0-4f5f-86fb-8d08b60d991b">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2b2ca92f-8335-492c-b5f7-f05c0dc4f8f3">
                                <link linkend="DCM_111221" xrefstyle="select: nopage">111221</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_26ac4a1a-f353-4c36-ab46-25dbc903abb1">Unknown failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_ba85fefb-7550-48e8-b9a8-4b024a3d86cc">RADLEX</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_20dcc602-ba76-4b10-9706-f745e6c75f63">
                                <link xl:href="http://www.radlex.org/RID/RID11327" xl:show="new">RID11327</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2ceba264-b9fd-43de-a049-c95dca1b2f39">Beam-hardening artifact</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="CID 7011" status="3" xml:id="sect_CID_7011">
            <title>Rejected for Quality Reasons</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_b5b6855f-e5b5-49ad-99ec-dc958d4b1e50">
                            <emphasis role="bold">Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_e2e26106-c7d7-4768-aed3-7f1c4dde2042">
                            <emphasis role="bold">20020904</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 7011" rules="all" xml:id="table_CID_7011">
                <caption>Rejected for Quality Reasons</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_a06a3871-7fc0-4d31-9957-1ff71ab4b9ad">
                                <emphasis role="bold">Coding Scheme Designator</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_94324dc7-ad2a-4f70-9360-060c6e77a63b">
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b4fb24e6-8651-4f87-a9ee-e8191765d1e9">
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f0ebb9a7-59ac-4794-8145-16da9869331f">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6ed99ae5-1597-413d-8c37-c73f9a37e4c1">
                                <link linkend="DCM_111207" xrefstyle="select: nopage">111207</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_feebd2a3-a2af-4966-910c-cd43dfa3b14d">Image artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e4a9bf18-a020-4fc3-b34e-675b608573d4">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_792472eb-4b4e-4cb4-ab40-6d6a373c22c5">
                                <link linkend="DCM_111208" xrefstyle="select: nopage">111208</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_50a8dba8-9c78-49c6-950d-0e4179b3a6d7">Grid artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_9af9b902-d6f9-46db-bc8f-d5faa9b9ab86">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2ab7bdcb-c506-43de-a0a0-b862cddb9faf">
                                <link linkend="DCM_111209" xrefstyle="select: nopage">111209</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_10566e2f-71fd-48f5-8764-2aaac48d2c84">Positioning</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_39774aad-6ebf-41dc-bd07-0515390150bb">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42699deb-511e-4b9f-accd-0468564f6f50">
                                <link linkend="DCM_111210" xrefstyle="select: nopage">111210</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_354837cd-6925-476e-bbb4-7212f10063fa">Motion blur</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2086b79c-262f-4cdc-8676-64511220debd">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_84853ccd-053d-4aa7-8b83-be678c2285d3">
                                <link linkend="DCM_111211" xrefstyle="select: nopage">111211</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_01ce9134-24bb-4c61-918d-91028199d121">Under exposed</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_efd9008b-d56d-40b3-a7d7-53cff4ad43fc">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b0ecf752-750f-4b58-8e68-77e8bb85c379">
                                <link linkend="DCM_111212" xrefstyle="select: nopage">111212</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3b8ba995-9203-4d64-8424-455bab7860f7">Over exposed</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_b9e22841-f0ab-4d9d-9862-a8f1c72b3ca3">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3b5d66d3-1bf1-40ef-82c8-98607899cb79">
                                <link linkend="DCM_111213" xrefstyle="select: nopage">111213</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a30b2051-8f27-4dea-9e1c-317d5213afa3">No image</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_26b8e179-d7a9-4b81-95b9-47fe548156ed">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4e9102cb-54a4-4060-880f-d455f2d76768">
                                <link linkend="DCM_111214" xrefstyle="select: nopage">111214</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e919003e-4800-48be-b669-de22dd5cf415">Detector artifact(s)</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_71a690dd-378e-4c3e-a0a6-ec8587521077">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7fe70faf-76c1-4344-adde-e5c2db2c936a">
                                <link linkend="DCM_111215" xrefstyle="select: nopage">111215</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8918233a-3da4-42fe-83cb-1e4716ff9ef3">Artifact(s) other than grid or detector artifact</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_6cee293b-0a37-41cd-a0f2-aa951074e168">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_acd2078b-0daf-4f61-8b81-f4c3125791f9">
                                <link linkend="DCM_111216" xrefstyle="select: nopage">111216</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d425028c-db9a-4dec-a759-3c825b469a64">Mechanical failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e7cca2cd-67fc-40dc-a42a-53b9a15c90b3">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_30245332-9564-4e72-b341-f9acf3ab2252">
                                <link linkend="DCM_111217" xrefstyle="select: nopage">111217</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8f246218-97bd-49b0-bb2a-a35093106f55">Electrical failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f596c8e3-93a6-415a-8d8d-7f50e8ac55ae">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_c6219a10-292d-47da-9134-142caf9a5bf3">
                                <link linkend="DCM_111218" xrefstyle="select: nopage">111218</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_206ea16c-3d8d-4f94-a6ac-1fa02e822917">Software failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4cc1684f-c6f7-4e8b-bb9e-fb10704a09e8">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_fe6376a8-ec07-41a8-bd1d-daa9b3fb99da">
                                <link linkend="DCM_111219" xrefstyle="select: nopage">111219</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_417bd8ac-27bc-44c4-8623-a2991e65c40f">Inappropriate image processing</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_55a77f6a-7c6d-478e-9122-6f92a5100afa">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_890b77de-6d3f-47a2-ab42-22f8918af10d">
                                <link linkend="DCM_111220" xrefstyle="select: nopage">111220</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_87bf0ebe-430c-4d5e-8d0a-181c273729f8">Other failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_1ba94aa3-b5ef-4af4-a20c-5796ecf13286">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e5b77798-88aa-45aa-a187-726d01ecc0ce">
                                <link linkend="DCM_111221" xrefstyle="select: nopage">111221</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_794bafc7-546c-421c-b8aa-7ad37a96291a">Unknown failure</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_3018a27a-8806-4447-b712-6f602ef625c8">DCM</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_0aeb6408-f8a0-472c-9c24-52d7f494bddd">
                                <link linkend="DCM_113026" xrefstyle="select: nopage">113026</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8692f0b7-89c3-46a4-bc23-26f534343bef">Double exposure</para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="CID 230" status="3" xml:id="sect_CID_230">
            <title>Yes-No</title>
            <variablelist spacing="compact">
                <varlistentry>
                    <term>
                        <emphasis role="bold">Type:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_28b89f28-64bd-40b3-834c-37f8191b8e37">
                            <emphasis role="bold">Non-Extensible</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
                <varlistentry>
                    <term>
                        <emphasis role="bold">Version:</emphasis>
                    </term>
                    <listitem>
                        <para xml:id="para_bad3e5d2-ca9d-4f01-ad3e-01cdb7efd5a3">
                            <emphasis role="bold">20060613</emphasis>
                        </para>
                    </listitem>
                </varlistentry>
            </variablelist>
            <table frame="box" label="CID 230" rules="all" xml:id="table_CID_230">
                <caption>Yes-No</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2704a225-f25e-4937-b98c-45703e0bdd5c">
                                <emphasis role="bold">Coding Scheme Designator</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_2fe32e59-d548-4d34-a528-99e5d154daa3">
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f17aaf8d-6575-47b1-a502-67315023c09b">
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_45057e1b-4582-4b7a-b003-b5a52e3baa86">
                                <emphasis role="bold">SNOMED-CT Concept ID</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5b856fcb-bbf6-4f7c-94b8-9e2b21e33073">
                                <emphasis role="bold">UMLS Concept Unique ID</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_33750814-dd2c-4d66-a10a-585dd61dc3c4">SRT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_64909569-cdc9-4774-a3bb-741f3e1ccb80">
                                <link xl:href="http://browser.ihtsdotools.org/?perspective=full&amp;conceptId1=373066001" xl:show="new">R-0038D</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_c8cf8d7c-0dce-42b3-945d-8cb64373413e">Yes</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_db42016d-3df8-49ef-a43f-7ae0c8283d71">
                                <link xl:href="http://browser.ihtsdotools.org/?perspective=full&amp;conceptId1=373066001" xl:show="new">373066001</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_28c0426b-7cc3-4569-ac0a-01202a792442">
                                <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C1298907" xl:show="new">C1298907</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_41faa94d-5b97-4d13-a9b7-0dd1cb48b8a9">SRT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e8959f71-e65b-4dd6-b706-936bf8c259fe">
                                <link xl:href="http://browser.ihtsdotools.org/?perspective=full&amp;conceptId1=373067005" xl:show="new">R-00339</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f68a2528-68dc-476c-aba2-fabb651a25b3">No</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_355bac7e-5932-4291-8cb5-61ad42a63051">
                                <link xl:href="http://browser.ihtsdotools.org/?perspective=full&amp;conceptId1=373067005" xl:show="new">373067005</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_43250785-95b9-4ff6-b376-966db998c6a5">
                                <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C1298908" xl:show="new">C1298908</link>
                            </para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_dd4c6422-8cf2-4e78-8cad-7190aabdaf6c">SRT</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7867a4ce-86bf-4daf-a2f0-b16bd25e03d4">
                                <link xl:href="http://browser.ihtsdotools.org/?perspective=full&amp;conceptId1=373068000" xl:show="new">R-0038A</link>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_a2118c9d-ed51-4711-834d-f0ebfcce36c5">Undetermined</para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_36234273-8a88-4f84-8268-5c58bb09ee9d">
                                <link xl:href="http://browser.ihtsdotools.org/?perspective=full&amp;conceptId1=373068000" xl:show="new">373068000</link>
                            </para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5a598aee-cf0a-431b-8b77-cdda5acd6a4d">
                                <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C3536725" xl:show="new">C3536725</link>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
        
        <section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            
            
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Code Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Code Meaning</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Definition</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold">Notes</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_111210">111210</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_088a7ece-cd2d-43a1-b317-1d41164c572e">Motion blur</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_60528387-dcf3-4f2f-8180-0713a3c039c6">Unacceptable image blur resulting from motion of the anatomy of interest during exposure or the inadequately compensated motion of X-Ray source with respect to the image detector during exposure.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_69610eff-c152-451f-ad0f-5d98f129f4ee"/>
                        </td>
                    </tr>
                    
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_ddd001">ddd001</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Is Repeated Acquisition</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>This acquisition of data (e.g., for constructing an image) is a repeat of an earlier acquisition that was for some reason unsatisfactory,</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_ddd002">ddd002</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Reason for Repeating Acquisition</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>The reason that data (e.g., for constructing an image) was acquired again.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_ddd003">ddd003</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Patient motion</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>The acquired data is unsatisfactory because the patient moved.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="center" colspan="1" rowspan="1">
                            <para xml:id="DCM_ddd004">ddd004</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>Suboptimal contrast timing</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>The acquired data is unsatisfactory because the contrast timing was not adequate.</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para/>
                        </td>
                    </tr>
                </tbody>
            </table>