  <title>CP-1612</title>
  <subtitle>Clarify incorrect multiplier for male James SUV lean body mass formula</subtitle>
    <chapter>
      <title></title>
        <informaltable rules="all" frame="box">
          <tr>
            <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
            <td rowspan="1" colspan="1" valign="top"><para>Sep 2016 Voting Packet</para></td>
          </tr>
          <tr>
            <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
            <td rowspan="1" colspan="1" valign="top"><para>2016/05/25</para></td>
          </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>QIBA FDG-PET group</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2015/10/01</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1612</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Clarify incorrect multiplier for male James SUV lean body mass formula</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16 (2016a+CP1535)</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                    <para>For the James formula (male), the value 128 was misquoted in an article as 120.</para>
                    <para>So, the SUVlbm(James) males should  = 1.1 x Weight - 128 x Weight squared / Height squared</para>
                    <para>However, a survey of different vendors reveals that some use 120 and some use 128.</para>
                    <para>It is proposed that the existing definitions in DICOM be relaxed to allow for both possibilities,
                        given that vendors may have implemented it either way, and that a new code be added for the
                        "correct" (128) value.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td><para><emphasis role="italic">Amend PS 3.3: </emphasis></para></td>
        </tr>
      </informaltable>
    </section>

<section label="C.8.9.1" status="4" xml:id="sect_C.8.9.1">
          <title>PET Series Module</title>
            <para>..</para>
          <para xml:id="para_4d39efd9-51ff-4582-b980-e2136d02962c">.</para>
          <table frame="box" label="C.8-60" rules="all" xml:id="table_C.8-60">
            <caption>PET Series Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7508cef2-d373-4cb4-ab01-16f5a978f8a5">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7b71e149-2d25-4ca6-a50d-c2ffef1e9704">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e4dcd9fd-3650-45d8-a5cd-3dfca89e99e6">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8ce02659-d8cf-4c44-9ea0-b121b53d9bc7">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0474fd1e-b1f3-4b6b-95b4-ca32b0a63e82">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6f095005-7b76-41f8-b194-3f4121107c93">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f209fc29-74c3-4892-b815-9dcd1f31703b">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1fb17223-c8e0-4f62-83dd-1d91c64d7b21">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d7e96cbb-906b-4f3c-a711-9b7414f4ba65">SUV Type</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e96eebfc-a7ea-4c2d-87c7-d5aa08ac6db2">(0054,1006)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9f173004-d315-4903-8f0e-70adffed0a06">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c1d928f9-845e-4c0d-b091-2cee8e17ab56">Type of Standardized Uptake Value (SUV).</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>BSA</term>
                      <listitem>
                        <para xml:id="para_1f09a3e1-fb52-4bb3-8725-22f2fc8a9d2b">body surface area</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>BW</term>
                      <listitem>
                        <para xml:id="para_590e6719-78d0-46fb-8df7-b0470b715f57">body weight</para>
                      </listitem>
                    </varlistentry>
                      <varlistentry>
                          <term>LBM</term>
                          <listitem>
                              <para xml:id="para_56343d9d-0b3b-41d7-b5a4-30d9c20ef0b5">lean body mass by James method<emphasis role="bold"> (see Note)</emphasis></para>
                          </listitem>
                      </varlistentry>
                      <varlistentry>
                          <term><emphasis role="bold"><emphasis role="underline">LBMJAMES128</emphasis></emphasis></term>
                          <listitem>
                              <para><emphasis role="bold"><emphasis role="underline">lean body mass by James method using a multiplier of 128 for males</emphasis></emphasis></para>
                          </listitem>
                      </varlistentry>
                      <varlistentry>
                      <term>LBMJANMA</term>
                      <listitem>
                        <para>lean body mass by Janmahasatian method</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>IBW</term>
                      <listitem>
                        <para>ideal body weight</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <para xml:id="para_c1200938-84b8-4a3a-b0f4-fe43359e9b7e">If absent, and the Units (0054,1001) are GML, then the type of SUV shall be assumed to be BW.</para>
                  <note>
                    <para xml:id="para_522db716-4502-4fd7-875c-607fc7291ca6">The type of SUV cannot reliably be deduced from the units alone, i.e., SUVbw, SUVibw and SUVlbm (James or Janmahasatian) all have units of GML.</para>
                  </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_07430a57-9140-4d67-989e-75fbb700b002">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d80a185e-75f7-4b03-b7b0-98851062bfd7">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6fa587e0-7130-46b4-b420-55c21c8ebc5b">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9b214656-c8f5-405f-b021-0d793413e608">..</para>
                </td>
              </tr>
            </tbody>
          </table>
          <note>
            <para xml:id="para_6dc98ecd-8792-4a35-91c7-e817b2988a3e">The formulas for the determination of SUVbw, SUVbsa, SUVlbm(James) and SUVibw are defined in Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction. Radiology, 1999</emphasis> at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>.</para>
                <para>
                    <emphasis role="bold"><emphasis role="underline">Unfortunately, Sugawara used a parameter of 120 rather than 128 for males, propagating an error in Morgan DJ, Bray KM. Lean Body Mass as a Predictor of Drug Dosage: Implications for Drug Therapy. Clinical Pharmacokinetics. 1994;26(4):292–307, which misquoted the original LBM definition that used 128 in James WPT, Waterlow JC. Research on Obesity: A Report of the DHSS/MRC Group. London: Her Majesty’s Stationery Office; 1976. Implementations differ in whether they have used 120 or 128 when the DICOM Defined Term is LBM. See Kelly M. SUV: Advancing Comparability and Accuracy. Siemens; 2009. Available from: http://www.mpcphysics.com/documents/SUV_Whitepaper_Final_11.17.09_59807428_2.pdf.</emphasis></emphasis></para>
                <para>The Janmahasatian LBM formula is defined in <emphasis role="italic">Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051–65.</emphasis> at <link xl:href="http://dx.doi.org/10.2165/00003088-200544100-00004"/> and its role in SUVlbm(Janma) calculation is discussed in <emphasis role="italic">Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging. Journal of Nuclear Medicine. 2014 Sep 1;55(9):1481–4. </emphasis> at <link xl:href="http://jnm.snmjournals.org/content/55/9/1481"/>.</para>
                <para>See also ... .</para>
          </note>
        </section>

    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td><para><emphasis role="italic">Amend DICOM PS 3.16:</emphasis></para></td>
        </tr>
      </informaltable>
    </section>

    
    <section label="CID 85" status="3" xml:id="sect_CID_85">
      <title>SUV Units</title>
      <variablelist spacing="compact">
        <varlistentry>
          <term>
            <emphasis role="bold">Type:</emphasis>
          </term>
          <listitem>
            <para xml:id="para_1233c115-159e-4923-8177-2fdfb7e5fd11">
              <emphasis role="bold">Extensible</emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term>
            <emphasis role="bold">Version:</emphasis>
          </term>
          <listitem>
            <para xml:id="para_798e4c11-1fcc-41c5-8d20-a5ad53ead4d8">
                            <emphasis role="bold"><emphasis role="strikethrough">20160314</emphasis><emphasis role="underline">yyyymmdd</emphasis></emphasis>
                        </para>
          </listitem>
        </varlistentry>
      </variablelist>
      <table frame="box" label="CID 85" rules="all" xml:id="table_CID_85">
        <caption>SUV Units</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_64d64317-1a6a-41a2-9031-58566be0d9e2">
                <emphasis role="bold">Coding Scheme Designator</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8bc22f66-ffc5-4b2b-a04a-88a43bcd6e00">
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2d06eea7-6f2e-4033-82d0-190fe081f18e">
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6d949794-0fb2-4cd3-95e8-0ead381dd70c">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_4742fa17-0d2f-49b1-9a90-59d067bfff04">{SUVlbm}g/ml{SUVlbm}</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7fd7d41-fca4-4957-b0c5-0bcfb58cfe86">Standardized Uptake Value lean body mass (James)</para>
            </td>
          </tr>
            <tr valign="top">
                <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">UCUM</emphasis></emphasis></para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">g/ml{SUVlbm(James128)}</emphasis></emphasis></para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                    <para><emphasis role="bold"><emphasis role="underline">Standardized Uptake Value lean body mass (James 128 multiplier)</emphasis></emphasis></para>
                </td>
            </tr>
        </tbody>
      </table>
      <note>
        <para xml:id="para_100b164d-1936-487d-97a8-c6cb77f8feaa">The formulas for the determination of SUVbw, SUVbsa, SUVlbm(James) and SUVibw are defined in Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction. Radiology, 1999</emphasis> at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>.</para>
                <para><emphasis role="bold"><emphasis role="underline">Unfortunately, Sugawara used a parameter of 120 rather than 128 for males, propagating an error in Morgan DJ, Bray KM. Lean Body Mass as a Predictor of Drug Dosage: Implications for Drug Therapy. Clinical Pharmacokinetics. 1994;26(4):292–307, which misquoted the original LBM definition that used 128 in James WPT, Waterlow JC. Research on Obesity: A Report of the DHSS/MRC Group. London: Her Majesty’s Stationery Office; 1976. Implementations differ in whether they have used 120 or 128 for ({SUVlbm}g/ml{SUVlbm}, UCUM, "Standardized Uptake Value lean body mass (James)"). See Kelly M. SUV: Advancing Comparability and Accuracy. Siemens; 2009. Available from: http://www.mpcphysics.com/documents/SUV_Whitepaper_Final_11.17.09_59807428_2.pdf.</emphasis></emphasis></para>
                <para>The Janmahasatian LBM formula is defined in <emphasis role="italic">Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051–65.</emphasis> at <link xl:href="http://dx.doi.org/10.2165/00003088-200544100-00004"/> and its role in SUVlbm(Janma) calculation is discussed in <emphasis role="italic">Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging. Journal of Nuclear Medicine. 2014 Sep 1;55(9):1481–4. </emphasis> at <link xl:href="http://jnm.snmjournals.org/content/55/9/1481"/>.</para>
        <para>The patient size correction factors are summarized here, where weight is in kg and height is in cm:</para>
        <para xml:id="para_211f1841-764d-481b-9165-4ecdd3a55a15">SUVbw: males &amp; females: weight</para>
        <para xml:id="para_f990e628-8377-4a92-956a-651e70f6467f">SUVlbm(James): males: 1.10 * weight - <emphasis role="bold"><emphasis role="underline">(</emphasis></emphasis>120<emphasis role="bold"><emphasis role="underline"> or 128)</emphasis></emphasis> * (weight/height) ^2</para>
        <para xml:id="para_3571f8d2-e7b4-432b-9d54-bcb2d7d04342">females: 1.07 * weight - 148 * (weight/height) ^2</para>
        <para>SUVlbm(Janma): males: 9.27E3 * weight / (6.68E3 + 216 * weight / (height^2))</para>
        <para>females: 9.27E3 * weight / (8.78E3 + 244 * weight / (height^2))</para>
        <para xml:id="para_7454a1df-dcc9-4492-b521-ddd007870a34">SUVbsa: males &amp; females: weight^ 0.425 * height^0.725 * 0.007184</para>
        <para xml:id="para_0418a618-881c-4ce1-8f86-1673ade9ebb7">SUVibw: males: 48.0 + 1.06 * (height - 152)</para>
        <para xml:id="para_33526356-6ca9-424a-a9bc-9095a7367e65">females: 45.5 + 0.91 * (height - 152)</para>
      </note>
    </section>
    
      <section label="CID 7180" status="3" xml:id="sect_CID_7180">
         <title>Abstract Multi-dimensional Image Model Component Semantics</title>
         <variablelist spacing="compact">
            <varlistentry>
               <term>
                  <emphasis role="bold">Type:</emphasis>
               </term>
               <listitem>
                  <para xml:id="para_ff3b7038-d405-4c7f-ae17-7191d2826e30">
                     <emphasis role="bold">Extensible</emphasis>
                  </para>
               </listitem>
            </varlistentry>
            <varlistentry>
               <term>
                  <emphasis role="bold">Version:</emphasis>
               </term>
               <listitem>
                  <para xml:id="para_28f1b78a-d2a9-4dc6-aa5e-6bfbdcff75c0">
                            <emphasis role="bold"><emphasis role="strikethrough">20160314</emphasis><emphasis role="underline">yyyymmdd</emphasis></emphasis>
                        </para>
               </listitem>
            </varlistentry>
         </variablelist>
         <table frame="box" label="CID 7180" rules="all" xml:id="table_CID_7180">
            <caption>Abstract Multi-dimensional Image Model Component Semantics</caption>
            <thead>
               <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_2eab0d49-4604-4a49-bf7d-f264812fd5ff">
                        <emphasis role="bold">Coding Scheme Designator</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_45981ceb-ccd2-49db-b35c-dad487b30b46">
                        <emphasis role="bold">Code Value</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_e809725f-b073-41d7-b621-163707cc949a">
                        <emphasis role="bold">Code Meaning</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_8e3ec141-b4cd-416e-bffb-4572bfe13597">
                        <emphasis role="bold">SNOMED-CT Concept ID</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_cc1d9649-7dae-4808-9d4c-725a13405592">
                        <emphasis role="bold">UMLS Concept Unique ID</emphasis>
                     </para>
                  </th>
               </tr>
            </thead>
            <tbody>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_3c5a2a58-501e-451a-bd09-035ead7b0abd">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_034f7adc-9723-4779-9199-aa5139c36aa5"> ...</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_d9b6370a-64e9-4869-93f6-93fef94a52cd">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_9516a1c2-346c-4476-889d-82fc5ce7ced2"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_28a48501-9cb2-4d23-a8ec-f16cb160260c"/>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_435ebcf3-1a36-497d-a80e-db65a4513377">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_3d1d0cd3-6d16-4065-ad3f-550b366bf1b9">
                        <link linkend="DCM_126402" xrefstyle="select: nopage">126402</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6364c99b-6145-4061-b28f-90ffd14fdba5">SUVlbm</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_138384e7-53a3-4090-bd76-60a6da82b6c1"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_1adece85-9dcf-4604-b5c1-b71934b5ef73"/>
                  </td>
               </tr>
                    <tr>
                        <td align="center" colspan="1"><para>DCM</para></td>
                        <td align="center" colspan="1"><para><link linkend="DCM_ddd001" xrefstyle="select: nopage">ddd001</link></para></td>
                        <td align="left" colspan="1"><para>SUVlbm(James128)</para></td>
                        <td align="center" colspan="1"><para></para></td>
                        <td align="center" colspan="1"><para></para></td>
                    </tr>
            </tbody>
         </table>
      </section>

    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td><para><emphasis role="italic">Amend DICOM PS3.16 concept definitions:</emphasis></para></td>
        </tr>
      </informaltable>
    </section>
    
    <section>
      <title/>
      <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
      <caption>DICOM Controlled Terminology Definitions</caption>
      <thead>
        <tr valign="top">
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">
              <emphasis role="bold">Code Value</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">
              <emphasis role="bold">Code Meaning</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">
              <emphasis role="bold">Definition</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">
              <emphasis role="bold">Notes</emphasis>
            </para>
          </th>
        </tr>
      </thead>
      <tbody>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126402">126402</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c2e3e95-49e9-48a1-a260-e335bb835e86">SUVlbm</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d02b3c10-d16f-443f-9a53-fbe81c4be75d">Standardized Uptake Value calculated using lean body mass by James method. The patient size correction factor for males is 1.10 * weight - <emphasis role="bold"><emphasis role="underline">(</emphasis></emphasis>120<emphasis role="bold"><emphasis role="underline"> or 128)</emphasis></emphasis> * (weight/height) ^2, and for females is 1.07 * weight - 148 * (weight/height) ^2.</para>
            <para xml:id="para_9df819b6-d333-4b0a-91eb-cbe77579617d">Defined in Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999 at <link xl:href="http://radiology.rsna.org/content/213/2/521"/><emphasis role="bold"><emphasis role="underline">, except that either 120 or 128 may be used as the multiplier parameter for males).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">Unfortunately, Sugawara used a parameter of 120 rather than 128, propagating an error in Morgan DJ, Bray KM. Lean Body Mass as a Predictor of Drug Dosage: Implications for Drug Therapy. Clinical Pharmacokinetics. 1994;26(4):292–307, which misquoted the original LBM definition that used 128 in James WPT, Waterlow JC. Research on Obesity: A Report of the DHSS/MRC Group. London: Her Majesty’s Stationery Office; 1976. Implementations differ in whether they have used 120 or 128 when using this code. See Kelly M. SUV: Advancing Comparability and Accuracy. Siemens; 2009. Available from: http://www.mpcphysics.com/documents/SUV_Whitepaper_Final_11.17.09_59807428_2.pdf.</emphasis></emphasis></para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cea7ee94-c1bb-4c81-9588-3b9ff43dbbce"/>
          </td>
        </tr>
                    <tr>
                        <td align="center" colspan="1"><para xml:id="DCM_ddd001"><emphasis role="bold"><emphasis role="underline">ddd001</emphasis></emphasis></para></td>
                        <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">SUVlbm(James128)</emphasis></emphasis></para></td>
                        <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">Standardized Uptake Value calculated using lean body mass by James method, using the originally published 128 multiplier for males. The patient size correction factor for males is 1.10 * weight - 128) * (weight/height) ^2, and for females is 1.07 * weight - 148 * (weight/height) ^2.</emphasis></emphasis></para></td>
                        <td align="left" colspan="1"><para></para></td>
                    </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126411">126411</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_115b009c-9a00-4d1b-af00-2724e1778217">SUV lean body mass calculation method</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f4eaf243-c419-4a5c-a16f-d12f274b9fcc">James method of calculating Standardized Uptake Value using lean body mass. The patient size correction factor for males is 1.10 * weight - <emphasis role="bold"><emphasis role="underline">(</emphasis></emphasis>120<emphasis role="bold"><emphasis role="underline"> or 128)</emphasis></emphasis> * (weight/height) ^2, and for females is 1.07 * weight - 148 * (weight/height) ^2.</para>
            <para xml:id="para_68853096-7935-4e69-af20-506c5b778d33">Defined in Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999 at <link xl:href="http://radiology.rsna.org/content/213/2/521"/><emphasis role="bold"><emphasis role="underline">, except that either 120 or 128 may be used as the multiplier parameter for males).</emphasis></emphasis></para>
                            <para><emphasis role="bold"><emphasis role="underline">Unfortunately, Sugawara used a parameter of 120 rather than 128, propagating an error in Morgan DJ, Bray KM. Lean Body Mass as a Predictor of Drug Dosage: Implications for Drug Therapy. Clinical Pharmacokinetics. 1994;26(4):292–307, which misquoted the original LBM definition that used 128 in James WPT, Waterlow JC. Research on Obesity: A Report of the DHSS/MRC Group. London: Her Majesty’s Stationery Office; 1976. Implementations differ in whether they have used 120 or 128 when using this code. See Kelly M. SUV: Advancing Comparability and Accuracy. Siemens; 2009. Available from: http://www.mpcphysics.com/documents/SUV_Whitepaper_Final_11.17.09_59807428_2.pdf.</emphasis></emphasis></para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a6df8ef-87c8-4c1b-aa6e-5edb88eb8636"/>
          </td>
        </tr>
          <tr>
              <td align="center" colspan="1"><para xml:id="DCM_ddd002"><emphasis role="bold"><emphasis role="underline">ddd002</emphasis></emphasis></para></td>
              <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">SUV lean body mass calculation method using 128 multiplier</emphasis></emphasis></para></td>
              <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">James method of calculating Standardized Uptake Value using lean body mass with the originally published 128 multiplier for males. The patient size correction factor for males is 1.10 * weight - 128) * (weight/height) ^2, and for females is 1.07 * weight - 148 * (weight/height) ^2.</emphasis></emphasis></para></td>
              <td align="left" colspan="1"><para></para></td>
          </tr>
          
      </tbody>
    </table>
    </section>
    
        
    </chapter>
</book>