    <title>CP-1578</title>
    <subtitle>Correct Pulse Sequence Name in MR Image Library</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/09/15</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>Jörg Riesmeier</para><para><link
                xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:mail@jriesmeier.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2015/08/21</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1578</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Correct Pulse Sequence Name in MR Image Library</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>The code used for Pulse Sequence Name in TID 1606 is already used for something else.</para>
          <para>The definition in PS3.3 for the corresponding attribute is used.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="TID 1606" status="3" xml:id="sect_TID_1606">
        <title>Image Library Entry Descriptors for MR</title>
        <table frame="box" label="TID 1606" rules="all" xml:id="table_TID__1606">
          <caption>Image Library Entry Descriptors for MR</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b86f2bcb-89f5-4143-8c92-9885eaab6f56"/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a5cc9616-c9f2-4ff0-8085-6b90d77bfea4">NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_458dcde2-3c48-4f04-b9df-b89a17814a9f">Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4c667b0e-0bed-4f39-84f8-e1e47a5ccc6d">VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9c063452-5de0-44f1-a9ed-124e7fb6c437">Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_be2dbd6e-8090-4811-94b0-8a4bef3c97c3">VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5cfe7daf-e03d-4897-8e5e-38e42a25f703">Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_1b46b12c-b7b2-4c30-8bc4-8936f8ad6ecc">Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a30d2c5e-0434-422f-ad16-1b0bff397b3b">Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d11c9e48-ac1e-46d1-b71b-7d794bf18e25">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5069bd55-3dec-427d-99df-93e7a753b20b"/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_284793dd-4e24-47a3-a869-ea6b16c5f6b2">HAS ACQ CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a56b4fd2-620a-4055-bd71-34d71afb633a">TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20727b85-1744-47ea-a6fe-d9ea113469b6">EV <link linkend="DCM_ddd001" xrefstyle="select: nopage">(<emphasis role="bold"><emphasis role="strikethrough">110909</emphasis><emphasis role="underline">ddd001</emphasis></emphasis>, DCM, "Pulse Sequence Name")</link>
              </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f303f13f-318c-4e04-b868-2f2d36041c35">1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_784932be-21c9-4bff-bc42-02030baeb3c0">U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e59234f-c752-40dc-897f-cad9c9d114e3"/>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </table>
        <para xml:id="para_6b4b30b3-00f6-44dd-9a43-b63209d88da8">
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable frame="box" rules="all">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a156b016-7000-4adb-95f8-070f37688132">Pulse Sequence Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9892402-5025-4703-917c-6e21424165c3">The value of Pulse Sequence Name (0018,9005) or Sequence Name (0018,0024) in
                  the Image IOD. See PS3.3 Section C.8.13.4 "MR Pulse Sequence Module"<!--<olink targetdoc="PS3.3" targetptr="sect_C.8.13.4" xrefstyle="select: docname label quotedtitle"/>-->.</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      </section>
      
      
      <section label="D" status="1" xml:id="chapter_D">
        <title>DICOM Controlled Terminology Definitions (Normative)</title>
        <para xml:id="para_3cce97e2-c3f2-4dcd-bd55-9bf6bc7f7f2a">This Annex specifies the meanings of codes defined in DICOM, either explicitly or by reference to another part of DICOM or an external reference document or standard.</para>
        <para xml:id="para_89a5e18f-1781-4007-be8c-654d60de917b">DICOM Code Definitions (Coding Scheme Designator "DCM" Coding Scheme Version "01")</para>
        <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
          <caption>DICOM Controlled Terminology Definitions</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">
                  <emphasis role="bold">Definition</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">
                  <emphasis role="bold">Notes</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
        <td align="center" colspan="1" rowspan="1">
          <para xml:id="DCM_110909">110909</para>
        </td>
        <td align="left" colspan="1" rowspan="1">
          <para xml:id="para_a674d783-9654-449d-be3f-7888cfb9043b">Image Orientation (Patient) Column Z</para>
        </td>
        <td align="left" colspan="1" rowspan="1">
          <para xml:id="para_c4bd6902-908b-4e96-bb28-6ad4a9693536">The z value of the first column direction cosine with respect to the patient, with respect to the patient-based coordinate system.</para>
        </td>
        <td align="left" colspan="1" rowspan="1">
          <para xml:id="para_6d2d4355-4d79-42d9-9379-2ded14d2dc86"/>
        </td>
      </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_ddd001"><emphasis role="bold"><emphasis role="underline">ddd001</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Pulse Sequence Name</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Name of an MR pulse sequence for annotation purposes. Potentially vendor-specific name.</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            </tr>
            
          </tbody>
        </table>
      </section>
      
      
    </chapter>
</book>