<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-1569</title>
    <subtitle>Define CT Reconstruction Diameter more precisely and correct Enhanced CT illustration</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
              <td rowspan="1" colspan="1" valign="top"><para>May 2016 Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/03/18</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/01/07</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1569</para></td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Define CT Reconstruction Diameter more precisely and correct Enhanced CT illustration</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.3</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
                  <para>The single-valued Attribute Reconstruction Diameter (0018,1100) has been present in the Standard since ACR-NEMA 1985, with the rather vague definition (never changed) of "Diameter in mm of the region from within which data were used in creating the reconstruction of the image. Data can exist outside this region and portions of the patient can exist outside this region".</para>
          <para>Traditionally, for square CT images it has been populated with values such that for Pixel Spacing (0028,0030):</para>
          <para>Pixel Spacing value 1 = row spacing = Reconstruction Diameter (0018,1100) / Rows (0028,0010)</para>
          <para>Pixel Spacing value 2 = col spacing = Reconstruction Diameter (0018,1100) / Columns (0028,0011).</para>
          <para>In otherwords, the circlular region defined by Reconstruction Diameter (0018,1100) has been populated as defining the largest circle fitting within the square region encoded as the Pixel Data.</para>
          <para>When Supplement 58 added the Enhanced CT Image IOD, additional attributes to describe the reconstructed area were added, as well as a figure; that figure incorrectly shows the Reconstruction Diameter defining a circular region that entirely contains the square region encoded as the Pixel Data, rather than the other way around.</para>
          <para>By contrast, Supplement 117 added the Enhanced PET Image IOD and illustrated the correct use of Reconstruction Diameter.</para>
          <para>Change the figure to reflect the traditionally usage, and add a more precise definition of Reconstruction Diameter (0018,1100) wherever it is already used in the legacy image IODs, consistent with traditional usage.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title/>

      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text are bold and <emphasis role="bold"><emphasis role="underline">underlined</emphasis></emphasis> for additions and <emphasis role="bold"><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="C.8.2.1" status="4" xml:id="sect_C.8.2.1">
          <title>CT Image Module</title>
          <para xml:id="para_aaba9d3f-d3ab-42ff-bc46-14d6dbce2893">...</para>
          <table frame="box" label="C.8-3" rules="all" xml:id="table_C.8-3">
            <caption>CT Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_fd13c408-cbc2-4251-8cef-0245ae467852">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_dcede1c9-6274-4e32-93f3-0e1137aa6253">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_06d59a79-b42e-4a60-9e6a-7f66d0436d5b">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0a9d4323-9134-4dc6-aa6c-411f65190bbd">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_40e01d26-559d-42f7-ad13-74db445f4de8">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7770bd3c-9270-4491-a7ae-dc59b94714c9">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f0b584b5-8826-452d-a2f4-8afa72f94995">Data Collection Diameter</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6c69cc19-44e3-4da2-90e9-131a003b53a7">(0018,0090)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5e565f9c-946d-42ac-9dfb-4b2a73e7dc3a">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_795f1f35-daab-4a3f-aa05-6f8edcdba01f">The diameter in mm of the region over which data were collected</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a5aeb08b-e323-4b12-a651-2ba4261601c5">Data Collection Center (Patient)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7bedfacc-e27b-4c44-9580-251c97fd1334">(0018,9313)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_09fda8e5-5982-4570-83b5-897e3dcabfa9">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7af6b366-9fe1-4bf6-b619-1832c7282fbd">The x, y, and z coordinates (in the patient coordinate system) in mm of the center of the region in which data were collected. See <xref linkend="sect_C.8.15.3.6.1" xrefstyle="select: label"/>.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6851c6f6-45b9-484b-b2ec-e0e1bc9f6e1d">Reconstruction Diameter</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e1b7dec2-2f52-43eb-bbed-59c8014e129d">(0018,1100)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d4710eb1-8f9c-47a4-9e57-fc1987c25acf">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_aa2a5b02-cefa-44cb-ba43-3357bf49f6d1">Diameter in mm of the region from within which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region.</para>
              <para><emphasis role="bold"><emphasis role="underline"> The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction.</emphasis></emphasis></para>
              <note>
                <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for square images with square pixels, both values of Pixel Spacing (0028,0030) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
              </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_13c86aaa-4be4-4550-afed-8d17c376eca1">Reconstruction Target Center (Patient)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2c147868-9d80-452c-8783-d865aca1b3b5">(0018,9318)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8faecc2d-fd9c-4377-9ff9-b934dd6db497">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_650cbccb-d9ec-4c0b-8de5-f4966a2ce62f">The x, y, and z coordinates (in the patient coordinate system) of the reconstruction center target point as used for reconstruction in mm. See <xref linkend="sect_C.8.15.3.6.1" xrefstyle="select: label"/>.</para>
                  <note>
                    <para xml:id="para_06a14d9d-f32a-42e2-a68d-fa1372a8dbd8">If the reconstructed image is not magnified or panned the value corresponds with the Data Collection Center (Patient) (0018,9313) attribute.</para>
                  </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_895bab77-85a5-4e69-b5c8-cfb82a0a7cf3">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6514dffc-6aeb-4bfa-82d0-6af0cf65d085">...</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      
      <section label="C.8.3.1" status="4" xml:id="sect_C.8.3.1">
          <title>MR Image Module</title>
      <para>...</para>
          <table frame="box" label="C.8-4" rules="all" xml:id="table_C.8-4">
            <caption>MR Image Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_cfd5fcac-5575-405c-902f-95f6dc74eb1e">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3ada24bb-7e9e-45a9-b89d-ddf2261064cf">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b3ec4e11-3722-4bf5-88d0-3b1e9e1a62a1">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5025cbac-d236-4a46-b194-1892f2890a7f">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_91dd4ef3-b782-4872-9faa-b44081e7f874">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_78a4f65b-a8e3-4145-943e-58506e28cea7">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_57752905-bfbc-465f-9184-7f21c236f738">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1c1145e0-9bef-451a-aa36-3158f741ef32">Reconstruction Diameter</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7fd590ae-d11a-4743-8434-2a227f5acdaf">(0018,1100)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_13e7acec-2de8-44aa-9ff2-9c8bcb169dd3">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_40d31644-8e9c-4ffe-99ed-33fbe9e413e3">Diameter in mm. of the region from within which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region.</para>
                  <para><emphasis role="bold"><emphasis role="underline"> The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction.</emphasis></emphasis></para>
                  <note>
                    <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for square images with square pixels, both values of Pixel Spacing (0028,0030) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                  </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9357e90a-b421-403e-8945-3124982de34a">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">...</td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6bdc5ac9-3911-4cc0-8346-fd8b6c01cf49">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bbcd8a62-a55d-4400-8434-ce184059d878">...</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      
      <section label="C.8.4.15" status="4" xml:id="sect_C.8.4.15">
        <title>NM Reconstruction Module</title>
        <para xml:id="para_d3bd4078-a645-4f74-9784-e280de97d74a">...</para>
        <table frame="box" label="C.8-15" rules="all" xml:id="table_C.8-15">
          <caption>NM Reconstruction Module Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d79e8a45-0c59-4f6c-b969-c34a2b830d98">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_fae9b55e-5e28-4bd2-a715-94d675e4dbf7">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_fea362d9-7a2d-4da0-a7ed-84bb466f4606">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0e6c1d7e-fb3d-437e-b10b-5647f58dfc26">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">...</td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_7a9681d7-4306-4eda-b5f0-e1b471139f72">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_83d5018e-8260-4650-ab02-0dc46070905a">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65f29505-0175-432d-9af2-00e7d81c6b49">....</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccc10989-cc80-49fe-bc19-da4aba4faf0e">Reconstruction Diameter</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f027a53f-1d16-4796-9195-807bc2903f93">(0018,1100)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f78153f1-2afc-4082-a96e-db701c9d4d0e">3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1682e264-1a3d-4d74-82f1-dab6ff147b60">Diameter, in mm, of the region from within which the data was used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region.</para>
                <para><emphasis role="bold"><emphasis role="underline"> The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction.</emphasis></emphasis></para>
                <note>
                  <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for square images with square pixels, both values of Pixel Spacing (0028,0030) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                </note>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5ee6730a-eefd-4f7a-9ced-76472cda0dd1">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_6358b31a-f12c-49aa-a2de-08c070bcc2dd">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_4359ab8d-0c35-440b-bc84-b88761a7faaf">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_53f52b56-53bf-4e34-864d-627f7a484383">...</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      <section label="C.8.9.1" status="4" xml:id="sect_C.8.9.1">
          <title>PET Series Module</title>
          <para xml:id="para_4d39efd9-51ff-4582-b980-e2136d02962c"> ...</para>
          <table frame="box" label="C.8-60" rules="all" xml:id="table_C.8-60">
            <caption>PET Series Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7508cef2-d373-4cb4-ab01-16f5a978f8a5">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7b71e149-2d25-4ca6-a50d-c2ffef1e9704">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e4dcd9fd-3650-45d8-a5cd-3dfca89e99e6">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8ce02659-d8cf-4c44-9ea0-b121b53d9bc7">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fdc05d9d-fb18-4f52-aad6-212b769052c0">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8d3125a2-c462-4bf6-8749-71eb7c210dc1">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_52eaef3f-0e96-48a7-8e9f-94a4a7f21e08">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0b8e6180-9404-4dbe-b64f-ef16c929de06">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b005f9f8-1fcd-4ea8-83c2-e41a2e981d82">Reconstruction Diameter</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d0ba61f2-25a6-4681-999c-c0a299f45d5a">(0018,1100)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e2665de0-f559-4828-b07f-2c394b8fda9b">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cd7d42d4-d0c6-4c1c-b1a7-18066f9ccf61">Diameter, in mm, of the region within which the data was used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region.</para>
                  <para><emphasis role="bold"><emphasis role="underline"> The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction.</emphasis></emphasis></para>
                  <note>
                    <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for square images with square pixels, both values of Pixel Spacing (0028,0030) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                  </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_493a0031-b184-4f04-b677-8550a2e11b0d">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d38c1ee4-75c0-40e7-857e-4422835b5e27">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0bed3a60-7dd7-4357-8c65-1a42d0411cbb">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2e988929-fdf8-4989-ac21-14aac22d2616">...</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      
      
      
      <section label="C.8.15.3.6" status="5" xml:id="sect_C.8.15.3.6">
        <title>CT Geometry Macro</title>
        <para xml:id="para_c49146b9-b8c4-4016-93ac-5bd0f60b5b70">...</para>
        <table frame="box" label="C.8-122" rules="all" xml:id="table_C.8-122">
          <caption>CT Geometry Macro Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_1e352a5e-ab3d-43cb-a33f-9ca7ff3150b5">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5deb33be-cedd-4395-b1bb-3917d667b252">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_467c6b23-e649-45f7-a4de-0b57c4f45601">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_df090a80-2378-45bd-aed3-5a6a4c962dc0">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa46fb4e-a040-41ac-9623-ac351157c1a4">CT Geometry Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_6473faa1-b700-471d-aace-26630e683e2b">(0018,9312)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_cb1cf9f7-cb28-41c8-b6e0-c00a11716e6f">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0bd0c5c1-07bc-41ba-81fc-f2b38b25bc8b">Contains the attributes defining the CT geometry.</para>
                <para xml:id="para_ee4c8761-a44d-4dfa-b7c8-285efada7665">Only a single Item shall be included in this Sequence.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc857d34-34c8-4187-bfd2-a470ac5ce262">&gt;Distance Source to Detector</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_ea04123d-6c74-4fde-b1aa-a7440e82f84d">(0018,1110)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_19733ec4-db4f-4bc2-8f69-696959008732">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9ceabf0d-4ff2-43f8-bd1c-26bba087b14d">Distance in mm from source to detector center. See <xref linkend="sect_C.8.15.3.6.1" xrefstyle="select: label"/>.</para>
                <note>
                  <para xml:id="para_cf268f31-fa79-41c3-84be-7f51f2169f14">This value is traditionally referred to as Source Image Receptor Distance (SID).</para>
                </note>
                <para xml:id="para_ca63ebb2-f5eb-49e8-b888-573ae6281aa9">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6d98b3c-add6-4cc4-8b2c-ebbd46488f68">&gt;Distance Source to Data Collection Center</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_fb122ab4-d357-4b72-894c-e032aff99417">(0018,9335)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_076a8d54-78f4-4be5-bc12-1adba6adfe5d">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b969997d-cfcb-427c-80a5-8b227fcdebeb">Distance in mm from source to data collection center. See <xref linkend="sect_C.8.15.3.6.1" xrefstyle="select: label"/>.</para>
                <para xml:id="para_ecb88bce-b992-4708-81b7-d2471e4a511e">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.</para>
              </td>
            </tr>
          </tbody>
        </table>

        <section label="C.8.15.3.6.1" status="6" xml:id="sect_C.8.15.3.6.1">
          <title>Relationships Between CT Geometric Attributes (Informative)</title>
          <para xml:id="para_43bd82e7-85b2-4f47-9de2-089529dd340a">In <xref linkend="figure_C.8-19" xrefstyle="select: label"/> the relationship of the Geometric Attributes within the CT Geometry and CT Reconstruction functional groups is shown.<emphasis role="bold"><emphasis role="underline"> In this example, the Pixel Data is shown encoding the entire reconstructed region with no outside padding or clipping, and the reconstructed region is illustrated as being square, i.e., the value of Reconstruction Diameter would be the same as both values of Reconstruction Field of View, whichever Attribute is present.</emphasis></emphasis>
          <xref linkend="figure_C.8-19" xrefstyle="select: label"/>, viewed from the front of the gantry (where the table enters the gantry), is informative only and is not meant to represent a standardization of an equipment-based frame of reference.</para>
          <informaltable rules="all" frame="box">
            <tr>
              <td>
                <para><emphasis role="italic">Replace incorrect figure:</emphasis></para>
              </td>
            </tr>
          </informaltable>
          <para xml:id="para_192e5a1b-07dd-470c-87e2-5af1f2eb7a20">
            <figure label="C.8-19" pgwide="1" xml:id="figure_C.8-19">
              <title>Geometry of CT Acquisition System</title>
              <mediaobject>
                <imageobject>
                  <imagedata fileref="cp1569/PS3.3_C.8-19_old.svg"/>
                </imageobject>
              </mediaobject>
            </figure>
          </para>
          <informaltable rules="all" frame="box">
            <tr>
              <td>
                <para><emphasis role="italic">with corrected figure:</emphasis></para>
              </td>
            </tr>
          </informaltable>
          <para xml:id="para_192e5a1b-07dd-470c-87e2-5af1f2eb7a20new">
            <figure label="C.8-19" pgwide="1" xml:id="figure_C.8-19new">
              <title>Geometry of CT Acquisition System</title>
              <mediaobject>
                <imageobject>
                  <imagedata fileref="cp1569/PS3.3_C.8-19_new.svg"/>
                </imageobject>
              </mediaobject>
            </figure>
          </para>
        </section>
      </section>
      
      <section label="C.8.15.3.7" status="5" xml:id="sect_C.8.15.3.7">
        <title>CT Reconstruction Macro</title>
        <para xml:id="para_a95c6dcd-5d27-4b7a-b95b-de374b439b6f"> ...</para>
        <table frame="box" label="C.8-123" rules="all" xml:id="table_C.8-123">
          <caption>CT Reconstruction Macro Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_255b38ac-cdd3-43cc-867d-5740eea1ece2">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_2a01a937-ab5f-4b35-9081-5c992338d446">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_553d22d0-8719-4f13-a1a1-340d814102be">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5ea4a872-9819-49a7-88a7-3597b753fac4">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bb6e266e-6cd7-485c-a7ee-3ae70e097258">CT Reconstruction Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_515a99b2-0d4f-4ef5-8291-bc19f97bfa2f">(0018,9314)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_038d1d59-33f4-421f-b39f-b15eff24fc40">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_86cacfc4-9cab-45f9-9ef2-1c0b50eb8b44">Contains the attributes holding information about the reconstruction techniques used.</para>
                <para xml:id="para_d7d9bdca-a492-4832-ac45-bba6fc3164f6">Only a single Item shall be included in this Sequence.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25fb8ab1-f287-4ac4-953e-693507f35d6a">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_8c5bb866-3d4a-4b24-a09f-e7c31175c2e6">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_83547fd1-3d27-498b-a98d-12af41ecc792">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">...</td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b25b038-cdc2-4d4e-90f1-78551ab06d97">&gt;Reconstruction Diameter</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_45e10372-7f98-4e84-af58-2e1125d3cb72">(0018,1100)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_1662b6d0-ac15-4958-a66a-924310b9c151">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ff19b38-6fd2-446c-8768-1e2daccab697">The diameter in mm of the region from which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region. See <xref linkend="sect_C.8.15.3.6.1" xrefstyle="select: label"/>.</para>
                <para><emphasis role="bold"><emphasis role="underline"> The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped or padded after reconstruction.</emphasis></emphasis></para>
                <note>
                  <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for square images with square pixels, both values of Pixel Spacing (0028,0030) and Reconstruction Pixel Spacing (0018,9322) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                </note>
                <para xml:id="para_83102abb-d64e-4c4d-a09b-837b5ebd3f0c">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and Reconstruction Field of View (0018,9317) is not present.</para>
                <para xml:id="para_c4ecfba6-4bc5-466e-bb8a-df92af325f93">Otherwise may be present if Frame Type (0008,9007) Value 1 of this frame is DERIVED and Reconstruction Field of View (0018,9317) is not present.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_10ea6e12-aec2-406b-a45a-eb093ca25ef4">&gt;Reconstruction Field of View</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_a90ca03f-7b10-4569-80b9-6d0588a1eb18">(0018,9317)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_d19cc34a-79b7-4d2d-8b9b-2523a059149b">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56c6aa4c-1abd-4898-9ff4-cc7561a2d867">The field of view width (x-dimension) followed by height (y-dimension) as used for reconstruction in mm.</para>
                <note>
                  <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for images with square pixels, both the corresponding values of Pixel Spacing (0028,0030) and Reconstruction Pixel Spacing (0018,9322) will be equal.</emphasis></emphasis></para>
                  <para><emphasis role="bold"><emphasis role="underline">The first value (adjacent row spacing) will be equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010).</emphasis></emphasis></para>
                  <para><emphasis role="bold"><emphasis role="underline">The second value (adjacent column spacing) will be equal to Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                </note>
                <para xml:id="para_723c6a83-2cbb-4c17-96e5-0e4da6b11df8">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and Reconstruction Diameter (0018,1100) is not present.</para>
                <para xml:id="para_20329faf-ac93-4ebd-ad8a-5fdb13171aa6">Otherwise may be present if Frame Type (0008,9007) Value 1 of this frame is DERIVED and Reconstruction Diameter (0018,1100) is not present.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_151ae12d-f782-457d-a88c-e1dc36a227a2">&gt;Reconstruction Pixel Spacing</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_6a66c9eb-a63e-4737-a32e-489a95976c68">(0018,9322)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_6817652e-5991-4cc1-ae15-de304ede4bea">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_07c3d0ec-64f8-4234-9221-09797e5cd631">Physical distance in the patient between the center of each reconstructed pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing in mm. See Section 10.7.1.3 for further explanation of the value order.</para>
                <para xml:id="para_da43546d-9742-423b-b10d-cf9c386bdb48">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1625cad8-d715-4ef8-9c60-a949d2b73fb9">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_83c9120a-c02f-4ea2-abab-40cf19b46524">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_0aedc48d-bca9-47c9-86ef-79e83b9000b8">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f7e06414-7484-4817-a8d5-da04af42e4d6">...</para>
            </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      
      <section label="C.8.22.5.4" status="5" xml:id="sect_C.8.22.5.4">
        <title>PET Position Macro</title>
        <para xml:id="para_c7e1c2aa-a4de-4729-a2d4-e2fbbedb34b5"> ...</para>
        <table frame="box" label="C.8.22-14" rules="all" xml:id="table_C.8.22-14">
          <caption>PET Position Macro Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4659a3cb-d323-45eb-82c0-522c2ece6ce8">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9d2edb0c-e199-49c2-8540-4d4ed8db5774">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d29a4841-e598-487b-920a-635f476c3fec">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_df4aa8a5-90ef-419a-a06f-1055b9123eb5">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70751851-62ce-4150-90c0-cede0ae4084b">PET Position Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_22f2edf1-9571-4c36-a121-aadeb8ea7c3a">(0018,9735)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_bd69899a-4523-47de-b741-f36582904b93">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_892350a6-96de-46e3-aca4-15898c8fea24">Contains the attributes defining the PET geometry.</para>
                <para xml:id="para_c56dff7a-3022-4ee7-a263-7caa6292f6d1">Only a single Item shall be included in this Sequence.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_afcb307e-1baa-4e9b-aeee-f5b5db67ca62">&gt;Table Position</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_12f68922-8c30-4c2e-921b-63f1852018d2">(0018,9327)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_a51579b5-4db3-4358-a8ad-bbe82a603d78">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d65f3803-1278-4c27-a382-7a5bc7e60278">Relative longitudinal position of acquisition location of this frame in mm from an implementation specific reference point. Shall be relative to the same reference point for all frames in this SOP Instance, but may be different from the reference point in other SOP Instances. Positions as the table moves into the gantry viewed from the front are more negative.</para>
                <note>
                  <orderedlist numeration="arabic">
                    <listitem>
                      <para xml:id="para_99e8b360-574f-4b2b-9a24-c8a925385584">For contiguous slices reconstructed from multiple detectors one would expect different values for adjacent slices.</para>
                    </listitem>
                    <listitem>
                      <para xml:id="para_6fd34df0-e58a-4c32-8e91-1f62654310d0">Lateral positioning or tilting or swiveling are not described.</para>
                    </listitem>
                  </orderedlist>
                </note>
                <para xml:id="para_568daaf7-4332-4fcb-a2c7-f5297433e079">Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a995d94-d106-4b04-8326-e0f120ca08e8">&gt;Data Collection Center (Patient)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_cc7f6d06-ab9f-4da6-afb3-6f7e4698d247">(0018,9313)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_64be44ca-f523-4402-b5f1-94d02877b837">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a6c0231-4e62-4b57-845b-2a1160c17b04">The x, y, and z coordinates (in the patient coordinate system) in mm of the center of the region in which data were collected. See <xref linkend="sect_C.8.22.5.4.1" xrefstyle="select: label"/>.</para>
                <para xml:id="para_e2921655-dcc0-4dcc-ae99-ed628ba342f6">Required if Frame Type (0008,9007) Value 1 equals ORIGINAL. May be present otherwise.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_53f3a20b-b666-4da8-8868-a90110f95f42">&gt;Reconstruction Target Center (Patient)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_170e4c6b-68aa-4543-92e1-ec648be81edb">(0018,9318)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_b87c35d1-368b-4aff-8f65-652b602f9447">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_326a199e-e841-4cf5-b692-d55b249df8ce">The x, y, and z coordinates (in the patient coordinate system) of the reconstruction center target point as used for reconstruction in mm. See <xref linkend="sect_C.8.22.5.4.1" xrefstyle="select: label"/>.</para>
                <note>
                  <para xml:id="para_7c50f985-3466-459e-be63-a07410c1672e">If the reconstructed image is not magnified or panned the value corresponds with the Data Collection Center (0018,9313) attribute.</para>
                </note>
                <para xml:id="para_f8be9b8e-6d8a-4834-92e6-2f80810e3a9f">Required if Frame Type (0008,9007) Value 1 equals ORIGINAL. May be present otherwise.</para>
              </td>
            </tr>
          </tbody>
        </table>
        <section label="C.8.22.5.4.1" status="6" xml:id="sect_C.8.22.5.4.1">
          <title>Relationships Between PET Geometric Attributes (Informative)</title>
          <para xml:id="para_17cfd725-9a65-4216-b658-412484997be3">In <xref linkend="figure_C.8.22-1" xrefstyle="select: label"/> the relationship of the Geometric Attributes within the PET Geometry and PET Reconstruction functional groups is shown.<emphasis role="bold"><emphasis role="underline"> In this example, the Pixel Data is shown encoding the entire reconstructed region with no outside padding or clipping, and the reconstructed region is illustrated as being square, i.e., the value of Reconstruction Diameter would be the same as both values of Reconstruction Field of View, whichever Attribute is present.</emphasis></emphasis>
          <xref linkend="figure_C.8.22-1" xrefstyle="select: label"/>, viewed from the front of the gantry (where the table enters the gantry), is informative only and is not meant to represent a standardization of an equipment-based frame of reference.</para>
          <para xml:id="para_c8f137e0-591c-4846-8f89-5356e60836c5">
            <figure label="C.8.22-1" pgwide="1" xml:id="figure_C.8.22-1">
              <title>Geometry of PET Acquisition System</title>
              <mediaobject>
                <imageobject>
                  <imagedata fileref="cp1569/PS3.3_C.8.22-1.svg"/>
                  <!--<imagedata fileref="part03_fromword_files/C.8.22-1.png"/>-->
                </imageobject>
              </mediaobject>
            </figure>
          </para>
        </section>
      </section>
      
      <section label="C.8.22.5.6" status="5" xml:id="sect_C.8.22.5.6">
        <title>PET Reconstruction Macro</title>
        <para xml:id="para_50c56f34-a548-456d-ad2c-4917492b5d5a">
          <xref linkend="table_C.8.22-17" xrefstyle="select: label"/> specifies the attributes of the PET Reconstruction Functional Group Macro, which describe the method used to reconstruct this image.</para>
        <table frame="box" label="C.8.22-17" rules="all" xml:id="table_C.8.22-17">
          <caption>PET Reconstruction Macro Attributes</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_f6168e76-8c27-4984-9089-0a06825051fe">Attribute Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_35ab6c21-6158-4f14-a075-ebbe437d810f">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_f0cd9f97-5b36-437b-9994-df46853ba434">Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b2077560-2d44-4e90-b7ad-628f2f6b7a78">Attribute Description</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b5d2536-37e9-44d6-8e38-db0157eeb07e">PET Reconstruction Sequence</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_231ea80b-48e8-42b9-b019-9e2ead46c1d1">(0018,9749)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_553d2c88-8cb3-423a-adbd-4b94478ed658">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a6772aa6-0a5f-4db5-863e-976b27a740ac">Contains the attributes describing the reconstruction process for this frame.</para>
                <para xml:id="para_c8dcaa43-0ee9-4a6e-a549-31689891a303">Only a single Item shall be included in this Sequence.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d64276e-89e3-41f5-9650-b677f0a67e35">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_9f318afe-701e-4f0a-bc8b-5e17429fc1d6">...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_cf90ade1-8d5c-4334-864b-c88161a39fcb">...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">...</td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75657195-1c00-4b03-8d0e-71caef1c301c">&gt;Reconstruction Diameter</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_0f71519a-c35a-4030-a03a-2bb405c4bfd1">(0018,1100)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_eabc9ab0-3833-4bdd-a464-9ab9c49daa2c">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3be45b6c-e6f0-4107-8ab3-32cb95971b68">The diameter in mm of the region from which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region. See <xref linkend="sect_C.8.22.5.4.1" xrefstyle="select: label"/>.</para>
                <para><emphasis role="bold"><emphasis role="underline"> The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction.</emphasis></emphasis></para>
                <note>
                  <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for square images with square pixels, both values of Pixel Spacing (0028,0030) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                </note>
                <para xml:id="para_9ab6f141-55cf-463a-b862-61ccc6a3445e">Required if Frame Type (0008,9007) Value 1 equals ORIGINAL and Reconstruction Field of View (0018,9317) is not present.</para>
                <para xml:id="para_5ea98a7f-3041-4f84-903d-5e87383c1884">Otherwise may be present if Frame Type (0008,9007) Value 1 equals DERIVED and Reconstruction Field of View (0018,9317) is not present.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab5d70f2-3571-4e6c-9e47-1d5dc17e3489">&gt;Reconstruction Field of View</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_b028ce65-78e1-4c69-9dd3-1be5b939b1f1">(0018,9317)</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_be690edd-8578-4cb5-9b2b-f78ee7c91057">1C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b78bd3d7-0fb1-4209-bf46-a80c17954548">The field of view width (x-dimension) followed by height (y-dimension) as used for reconstruction in mm.</para>
                <note>
                  <para><emphasis role="bold"><emphasis role="underline">If not cropped or padded, for images with square pixels, both the corresponding values of Pixel Spacing (0028,0030) and Reconstruction Pixel Spacing (0018,9322) will be equal.</emphasis></emphasis></para>
                  <para><emphasis role="bold"><emphasis role="underline">The first value (adjacent row spacing) will be equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010).</emphasis></emphasis></para>
                  <para><emphasis role="bold"><emphasis role="underline">The second value (adjacent column spacing) will be equal to Reconstruction Diameter (0018,1100) / Columns (0028,0011).</emphasis></emphasis></para>
                </note>
                <para xml:id="para_863d7eab-5a49-471d-b398-ebfd7aa7a7b5">Required if Image Type (0008,9007) Value 1 equals ORIGINAL and Reconstruction Diameter (0018,1100) is not present.</para>
                <para xml:id="para_751f634f-64af-4f2c-afff-b8c7b5c61ade">Otherwise may be present if Frame Type (0008,9007) Value 1 equals DERIVED and Reconstruction Diameter (0018,1100) is not present.</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      
      
    </chapter>
</book>
