    <title>CP-1549</title>
    <subtitle>Remove description of retired Big Endian Transfer syntax</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/03/18</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                  xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>2015/10/02</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1549</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Remove description of retired Big Endian Transfer syntax</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.5</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                   <para>When Supplement 98 flagged the Big Endian Explicit VR Transfer Syntax as retired in PS3.6, it did not
                       remove the description of it in PS3.5.</para>
                    <para>The illustration of the representation inside big and little endian "machines" is not removed.</para>
                    <para>Also clarify that the  OL VR is not used for Pixel Data or Overlay Data.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title></title>

        <section>
            <title></title>
            <informaltable rules="all" frame="box">
                <tr><td><para><emphasis role="italic">Amend DICOM PS3.5 as follows:</emphasis></para></td></tr>
            </informaltable>
        </section>
        
        <section label="6.2" status="2" xml:id="sect_6.2">
      <title>Value Representation (VR)</title>
      <table frame="box" label="6.2-1" rules="all" xml:id="table_6.2-1">
        <caption>DICOM Value Representations</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_62ffdcad-27ab-482e-8581-ab71ea117436">
                <emphasis role="bold">VR Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ae0cc4bd-efd7-4aa8-ba18-7df14fce5fbe">
                <emphasis role="bold">Definition</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3473682f-25a1-47a4-8fc2-d99762e43296">
                <emphasis role="bold">Character Repertoire</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_482e9707-5e08-45b5-b56a-11321f495679">
                <emphasis role="bold">Length of Value</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9a67e98c-cd28-4023-8358-9aab59aaea07">AT</para>
              <para xml:id="para_8d360848-86d9-4e16-bd8e-a4ecd3a0fa89">Attribute Tag</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53012354-d943-4d36-8103-f7640b7a15bc">Ordered pair of 16-bit unsigned integers that is the value of a Data Element Tag.</para>
              <para xml:id="para_4461a6a0-0a7e-4546-ade2-aa304e3d6212">Example: A Data Element Tag of (0018,00FF) would be encoded as a series of 4 bytes in a Little-Endian Transfer Syntax as 18H,00H,FFH,00H<emphasis role="bold"><emphasis role="strikehtrough"> and in a Big-Endian Transfer Syntax as 00H,18H,00H,FFH</emphasis></emphasis>.</para>
              <note>
                <para xml:id="para_ecc00586-542e-403c-844f-95526ddcf942">The encoding of an AT value is exactly the same as the encoding of a Data Element Tag as defined in Section 7<!--<xref linkend="chapter_7" xrefstyle="template:Section %n"/>-->.</para>
              </note>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_879b1519-9274-4d2c-b9c0-9615a28ccf8c">not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0694845c-375f-4489-9a36-9fbc00cdd285">4 bytes fixed</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_973d1f90-6cb4-4966-86e4-68f32946b6e9">OB</para>
              <para xml:id="para_bd2d566b-6196-4dfa-b40b-2ed6214162f6">Other Byte</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63ef9d65-10d9-445d-b7b3-93cb6d31e218">A string of bytes where the encoding of the contents is specified by the negotiated Transfer Syntax. OB is a VR that is insensitive to <emphasis role="bold"><emphasis role="strikethrough">Little/Big Endian </emphasis></emphasis>byte ordering (see <xref linkend="sect_7.3" xrefstyle="select: label"/>). The string of bytes shall be padded with a single trailing NULL byte value (00H) when necessary to achieve even length.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_401bb3cf-9510-477f-b992-96cd3d0aa2c9">not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_689bdc65-7e1b-4777-a62c-699775eae1d2">see Transfer Syntax definition</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f21f3196-96be-4c5a-a5a9-b014a18c3c13">OD</para>
              <para xml:id="para_3ff0e461-16ef-4035-afbe-d2be9b209fdd">Other Double</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c00ad319-fe87-41ad-83b5-ac20a1a70afe">A string of 64-bit IEEE 754:1985 floating point words. OD is a VR that requires byte swapping within each 64-bit word when changing <emphasis role="bold"><emphasis role="strikethrough">between Little Endian and Big Endian </emphasis></emphasis>byte ordering (see <xref linkend="sect_7.3" xrefstyle="select: label"/>).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98ea7d6a-97b6-4f95-b818-830c8eb86784">not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a427e77-4636-42ea-b20b-ecdc516b2ead">2<superscript>32</superscript>-8 bytes maximum</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_51c4e4f2-2811-4828-9a06-db1c5711c874">OF</para>
              <para xml:id="para_62319e8c-b6d8-4e69-99ef-22b57db8b33e">Other Float</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_411d45f9-2aae-46c4-af81-60d4205f25c2">A string of 32-bit IEEE 754:1985 floating point words. OF is a VR that requires byte swapping within each 32-bit word when changing <emphasis role="bold"><emphasis role="strikethrough">between Little Endian and Big Endian</emphasis></emphasis> byte ordering (see <xref linkend="sect_7.3" xrefstyle="select: label"/>).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e1db68a-7c7b-4565-90ae-db7bdacf2f71">not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b2e68b4-97aa-4d3d-88a0-fae5f37cd6c4">2<superscript>32</superscript>-4 bytes maximum</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>OL</para>
              <para>Other Long</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para>A string of 32-bit words where the encoding of the contents is specified by the negotiated Transfer Syntax. OL is a VR that requires byte swapping within each word when changing <emphasis role="bold"><emphasis role="strikethrough">between Little Endian and Big Endian </emphasis></emphasis>byte ordering (see <xref linkend="sect_7.3" xrefstyle="select: label"/>).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>see Transfer Syntax definition</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b548f29c-2871-4ad6-91ce-5f0ba6bd294f">OW</para>
              <para xml:id="para_b8371695-2729-4b85-8c99-12c64e4d3aaf">Other Word</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_29d49676-964d-48ea-81bc-8551eddc026e">A string of 16-bit words where the encoding of the contents is specified by the negotiated Transfer Syntax. OW is a VR that requires byte swapping within each word when changing <emphasis role="bold"><emphasis role="strikethrough">between Little Endian and Big Endian </emphasis></emphasis>byte ordering (see <xref linkend="sect_7.3" xrefstyle="select: label"/>).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd98f5fb-75f7-4f25-aebe-94715a7d7dbc">not applicable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2e3d2e2-b5a3-4bd4-aaf0-d86abb3bea98">see Transfer Syntax definition</para>
            </td>
          </tr>
        </tbody>
      </table>
      <section label="6.2.2" status="3" xml:id="sect_6.2.2">
        <title>Unknown (UN) Value Representation</title>
          <para xml:id="para_768f8d76-226d-44f6-a9ee-44cf0ef649af">The Unknown (UN) VR shall only be used for Private Attribute Data Elements and Standard Data Elements previously encoded as some DICOM VR other than UN using the DICOM Default Transfer Syntax (Implicit VR Little Endian), and whose Value Representation is currently unknown. As long as the VR is unknown the Value Field is insensitive to <emphasis role="bold"><emphasis role="strikethrough">Little/Big Endian </emphasis></emphasis>byte ordering and shall not be 'byte-swapped' (see <xref linkend="sect_7.3" xrefstyle="select: label"/>). In the case of undefined length sequences, the value shall remain in implicit VR form. See Section 7.8<!--<xref linkend="sect_7.8" xrefstyle="select: label"/>--> for a description of Private Data Attribute Elements and section 10 and Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>--> for a discussion of Transfer Syntaxes.</para>
        <note>
          <orderedlist numeration="arabic">
            <listitem/>
            <listitem>
              <para xml:id="para_c03ba4d0-2094-43e3-9ad1-fb039671675f">If at some point an application knows the actual VR for an Attribute of VR UN (e.g., has its own applicable data dictionary), it can assume that the Value Field of the Attribute is encoded in Little Endian byte ordering with implicit VR encoding, irrespective of the current Transfer Syntax.</para>
            </listitem>
            <listitem/>
            <listitem/>
            <listitem/>
          </orderedlist>
        </note>
      </section>
    </section>
        
        <section label="7.3" status="2" xml:id="sect_7.3">
            <title><emphasis role="strikethrough">Big Endian Versus </emphasis>Little Endian Byte Ordering</title>
            <para><emphasis role="bold"><emphasis role="underline">All unretired Transfer Syntaxes in DICOM require the use of Little Endian Byte Ordering.</emphasis></emphasis></para>
            <para xml:id="para_fe41f7af-a4ee-4a9f-8e74-e115224086d3"><emphasis role="bold"><emphasis role="strikethrough">Another component of the encoding of a Data Set that shall be agreed upon by communicating Application Entities is the Byte Ordering.</emphasis></emphasis></para>
            <para xml:id="para_a38977c0-90cf-409a-8aad-1f9573e4e862">Little Endian byte ordering is defined as follows:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_dedeaa1c-42c1-4d37-90fd-7ae400cc29e2">In a binary number consisting of multiple bytes (e.g., a 32-bit unsigned integer value, the Group Number, the Element Number, etc.), the least significant byte shall be encoded first; with the remaining bytes encoded in increasing order of significance.</para>
                </listitem>
                <listitem>
                    <para xml:id="para_34c940f5-337d-4566-878d-0177daefd5ef">In a character string consisting of multiple 8-bit single byte codes, the characters will be encoded in the order of occurrence in the string (left to right).</para>
                </listitem>
            </itemizedlist>
            <para xml:id="para_a05846b2-0d73-45d9-86b5-7d06a82de928">Big Endian byte ordering <emphasis role="bold"><emphasis role="strikethrough">is defined as follows:</emphasis><emphasis role="underline">was previously described but has been retired, See PS3.5 2015x.</emphasis></emphasis></para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_77a1442b-91bf-490a-931d-c7c492a83b8c"><emphasis role="bold"><emphasis role="strikethrough">In a binary number consisting of multiple bytes, the most significant byte shall be encoded first; with the remaining bytes encoded in decreasing order of significance.</emphasis></emphasis></para>
                </listitem>
                <listitem>
                    <para xml:id="para_66f09c72-4b3a-4f9f-9ead-ab2c3b4bd961"><emphasis role="bold"><emphasis role="strikethrough">In a character string consisting of multiple 8-bit single byte codes, the characters will be encoded in the order of occurrence in the string (left to right).</emphasis></emphasis></para>
                </listitem>
            </itemizedlist>
            <note>
                <para xml:id="para_b20f2fbd-617f-4233-8f28-ceacdcab0ffd">The packing of bits within values of OB<emphasis role="bold"><emphasis role="strikethrough">, OL</emphasis></emphasis> or OW Value <emphasis role="bold"><emphasis role="underline">R</emphasis><emphasis role="strikethrough">r</emphasis></emphasis>epresentation for Pixel Data and Overlay Data is described in Section 8<!--<xref linkend="chapter_8" xrefstyle="template:Section %n"/>-->.<emphasis role="bold"><emphasis role="underline"> The OL Value Representation is not used for Pixel Data or Overlay Data.</emphasis></emphasis></para>
            </note>
            <para xml:id="para_01d35885-9117-4583-a818-eb9468e2e9c3">Byte ordering is a component of an agreed upon Transfer Syntax (see Section 10<!--<xref linkend="chapter_10" xrefstyle="template:Section %n"/>-->). The default DICOM Transfer Syntax, which shall be supported by all AEs, uses Little Endian encoding and is specified in Section A.1<!--<xref linkend="sect_A.1" xrefstyle="select: label"/>-->. Alternate <emphasis role="bold"><emphasis role="underline">Little Endian</emphasis></emphasis>Transfer Syntaxes<emphasis role="bold"><emphasis role="strikethrough">, some of which use Big Endian encoding</emphasis></emphasis>, are also specified in Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->.</para>
            <note>
                <para xml:id="para_e28249f6-e346-45cb-bf58-62c561fcf88d">The Command Set structure as specified in PS3.7<!--<olink targetdoc="PS3.7" targetptr="PS3.7" xrefstyle="select: labelnumber"/>--> is encoded using the Little Endian Implicit VR Transfer Syntax.</para>
            </note>
            <para xml:id="para_f19daaf4-7894-4138-8db1-20e87d232a62">In the <emphasis role="bold"><emphasis role="strikethrough">default</emphasis></emphasis> case of Little Endian encoding, Big Endian Machines interpreting Data Sets shall do 'byte swapping' before interpreting or operating on certain Data Elements. The Data Elements affected are all those having VRs that are multiple byte Values and that are not a character string of 8-bit single byte codes. VRs constructed of a string of characters of 8-bit single byte codes are really constructed of a string of individual bytes, and are therefore not affected by byte ordering. The VRs that are not a string of characters and consist of multiple bytes are:</para>
            <itemizedlist>
                <listitem>
                    <para xml:id="para_8350f735-1f76-4828-93c4-18bb424b95b6">2-byte US, SS, OW and each component of AT</para>
                </listitem>
                <listitem>
                    <para xml:id="para_613a62a3-6098-4d7a-9a68-7e621eeaabd1">4-byte OF, OL, UL, SL, and FL</para>
                </listitem>
                <listitem>
                    <para xml:id="para_c9ceb71a-081c-4e72-b3a1-cf1831cb963f">8 byte OD, FD</para>
                </listitem>
            </itemizedlist>
            <note>
                <para xml:id="para_b0cf516e-2194-469d-855e-fd4a79b26a3d">For the above VRs, the multiple bytes are presented in increasing order of significance when in Little Endian format. For example, an 8-byte Data Element with VR of FD, might be written in hexadecimal as 68AF4B2CH, but encoded in Little Endian would be 2C4BAF68H.</para>
            </note>
        </section>
        
    <section label="8.1" status="2" xml:id="sect_8.1">
      <title>Pixel and Overlay Data, and Related Data Elements</title>
      <section label="8.1.2" status="3" xml:id="sect_8.1.2">
        <title>Overlay Data Encoding of Related Data Elements</title>
        <para xml:id="para_58b4f1a2-07ee-4e09-abf5-d99388f13d38">...</para>
        <note>
          <para xml:id="para_b5a3fd56-1954-49ab-a334-41ce059eae4d">The DICOM default Transfer Syntax (Implicit VR Little Endian) does not explicitly convey Value Representation and therefore the VR of OB may not be used for Pixel Data when using the default Transfer Syntax.</para>
        </note>
          <para xml:id="para_e5bcf268-0b7d-4b21-b495-84b330046490">Overlay Data is encoded as the direct concatenation of the bits of a single Overlay Plane, where the first bit of an Overlay Plane is encoded in the least significant bit, immediately followed by the next bit of the Overlay Plane in the next most significant bit. When the Overlay Data crosses a word boundary in the OW case, or a byte boundary in the OB case, it shall continue to be encoded, least significant bit to most significant bit, in the next word, or byte, respectively (see Annex D<!--<xref linkend="chapter_D" xrefstyle="template:Annex %n"/>-->). For Overlay Data encoded with the Value Representation OW, the byte ordering of the resulting 2-byte words is defined by the Little Endian <emphasis role="bold"><emphasis role="strikethrough">or Big Endian </emphasis></emphasis>Transfer Syntax<emphasis role="bold"><emphasis role="strikethrough">es</emphasis></emphasis> negotiated at the Association Establishment (see Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->).</para>
        <note>
            <para xml:id="para_a7ef1ac3-5a2d-4db2-a26c-885be2db7d91">For Overlay Data encoded with the Value Representation OB, the Overlay Data encoding is unaffected by <emphasis role="bold"><emphasis role="strikethrough">Little Endian or Big Endian </emphasis></emphasis>byte ordering.</para>
        </note>
      </section>
    </section>
    <section label="8.2" status="2" xml:id="sect_8.2">
      <title>Native or Encapsulated Format Encoding</title>
      <para xml:id="para_f9b54410-31ad-4044-9be0-651dd86c84b3">...</para>
      <para xml:id="para_d6734089-67bc-433e-87f7-40f00a859a9d">Native format Pixel Cells are encoded as the direct concatenation of the bits of each Pixel Cell, the least significant bit of each Pixel Cell is encoded in the least significant bit of the encoded word or byte, immediately followed by the next most significant bit of each Pixel Cell in the next most significant bit of the encoded word or byte, successively until all bits of the Pixel Cell have been encoded, then immediately followed by the least significant bit of the next Pixel Cell in the next most significant bit of the encoded word or byte. The number of bits of each Pixel Cell is defined by the Bits Allocated (0028,0100) Data Element Value. When a Pixel Cell crosses a word boundary in the OW case, or a byte boundary in the OB case, it shall continue to be encoded, least significant bit to most significant bit, in the next word, or byte, respectively (see Annex D<!--<xref linkend="chapter_D" xrefstyle="template:Annex %n"/>-->). For
          Pixel Data (7FE0,0010) encoded with the Value Representation OW, the byte ordering of the resulting 2-byte words is defined by the Little Endian <emphasis role="bold"><emphasis role="strikethrough">or Big Endian </emphasis></emphasis>Transfer Syntax<emphasis role="bold"><emphasis role="strikethrough">es</emphasis></emphasis> negotiated at the Association Establishment (see Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->).</para>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
              <para xml:id="para_1e6dec20-1376-4b1e-b909-255ba6abcf59">For Pixel Data (7FE0,0010) encoded with the Value Representation OB, the Pixel Data (7FE0,0010) encoding is unaffected by <emphasis role="bold"><emphasis role="strikethrough">Little Endian or Big Endian </emphasis></emphasis>byte ordering.</para>
          </listitem>
          <listitem>
            <para xml:id="para_754c31cf-0a9a-484e-bfba-2130fc81252e">If encoding Pixel Data (7FE0,0010) with a Value for Bits Allocated (0028,0100) not equal to 16 be sure to read and understand Annex D<!--<xref linkend="chapter_D" xrefstyle="template:Annex %n"/>-->.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_6287384f-06a5-47f6-ab74-625074661b31">...</para>
    </section>
    <section label="8.3" status="2" xml:id="sect_8.3">
      <title>Waveform Data and Related Data Elements</title>
      <para xml:id="para_9a8e1cfc-73c8-4db2-a88a-040b4251f39d">The DICOM protocol provides for the exchange of encoded time-based signals, or waveforms, encoded in the Waveform Data Element (5400,1010).</para>
      <note>
        <para xml:id="para_d143f2ad-4d5f-4107-abff-0f8e36db57ec">Per Section 7.6<!--<xref linkend="sect_7.6" xrefstyle="select: label"/>-->, an IOD supporting multiple sets of Waveform Data will encapsulate Data Element (5400,1010) within a Sequence.</para>
      </note>
      <para xml:id="para_d9e0d1cb-8342-422f-a1fa-e5bc837581f4">Encoded Waveform Data of various bit depths is accommodated through the Waveform Bits Allocated (5400,1004) Data Element. This element defines the size of each waveform data sample within the Waveform Data (5400,1010). Allowed values are 8 and 16 bits.</para>
      <para xml:id="para_4e05e669-69cc-48b4-b632-33d56cefab1c">The Value Representation of the Waveform Data (5400,1010) shall be OW; OB shall be used in cases where Waveform Bits Allocated has a value of 8, but only with Transfer Syntaxes where the Value Representation is explicitly conveyed.</para>
      <note>
        <orderedlist numeration="arabic">
          <listitem>
            <para xml:id="para_0a64c73d-232d-4053-b35b-9679358d93a8">Under the Default Transfer Syntax, OB and OW VRs have the identical byte transfer order.</para>
          </listitem>
          <listitem>
            <para xml:id="para_18b37e8d-9639-4837-924a-443ea870fae7">Conversion of a SOP Instance from the Default Transfer Syntax to an Explicit VR Transfer Syntax (uncompressed) requires the interpretation of the Waveform Bits Allocated (5400,1004) Data Element, to determine the proper VR of the Waveform Data.</para>
          </listitem>
        </orderedlist>
      </note>
      <para xml:id="para_9acd47c0-d9e7-4cde-a7c2-701a2c07b049">The following data elements related to Waveform Data shall be encoded with the same VR as Waveform Data: Channel Minimum Value (5400,0110), Channel Maximum Value (5400,0112) and Waveform Padding Value (5400,100A).</para>
    </section>
       
        <section label="A.3" status="2" xml:id="sect_A.3">
            <title>DICOM Big Endian Transfer Syntax (Explicit VR)</title>
            <para><emphasis role="bold"><emphasis role="underline">This Transfer Syntax was retired in 2006. For the most recent description of it, see PS3.5 2015X.</emphasis></emphasis></para>
            <para xml:id="para_e7d83f64-9ac7-438e-96fd-f8fdd8ff564b"><emphasis role="bold"><emphasis role="strikethrough">This Transfer Syntax applies to the encoding of the entire DICOM Data Set. This implies that when a DICOM Data Set is being encoded with the DICOM Big Endian Transfer Syntax the following requirements shall be met:</emphasis></emphasis></para>
            <orderedlist numeration="loweralpha">
                <listitem>
                    <para xml:id="para_8e2beefb-1d41-49d3-895d-c10b90c617f4"><emphasis role="bold"><emphasis role="strikethrough">The Data Elements contained in the Data Set structure shall be encoded with Explicit VR (with a VR Field) as specified in Section 7.1.2<!--<xref linkend="sect_7.1.2" xrefstyle="select: label"/>-->.</emphasis></emphasis></para>
                </listitem>
                <listitem>
                    <para xml:id="para_5f02b258-bc73-4edc-9d9b-775d3ea9ad86"><emphasis role="bold"><emphasis role="strikethrough">The encoding of the overall Data Set structure (Data Element Tags, Value Length, and Value) shall be in Big Endian as specified in <xref linkend="sect_7.3" xrefstyle="select: label"/>.</emphasis></emphasis></para>
                </listitem>
                <listitem>
                    <emphasis role="bold"><emphasis role="strikethrough">
                            <para xml:id="para_697e0a63-3e2f-4f9b-b518-0493fe0bbdfa">The encoding of the Data Elements of the Data Set shall be as follows according to their Value Representations:</para>
                        </emphasis></emphasis>
                    <itemizedlist>
                        <listitem>
                            <para xml:id="para_4fe7cb96-897a-4e59-bfdd-0ed4526c72ca"><emphasis role="bold"><emphasis role="strikethrough">For all Value Representations defined in this part, except for the Value Representations OB and OW, the encoding shall be in Big Endian as specified in <xref linkend="sect_7.3" xrefstyle="select: label"/>.</emphasis></emphasis></para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_6572b746-02d8-43f9-8193-ab875b2eab4a"><emphasis role="bold"><emphasis role="strikethrough">For the Value Representations OB, OL and OW, the encoding shall meet the following specification depending on the Data Element Tag:</emphasis></emphasis></para>
                            <itemizedlist>
                                <listitem>
                                    <para xml:id="para_bf5c33ac-dd30-41ea-9b23-7643205cad3a"><emphasis role="bold"><emphasis role="strikethrough">Data Element (7FE0,0010) Pixel Data</emphasis></emphasis></para>
                                    <itemizedlist>
                                        <listitem>
                                            <para xml:id="para_106f6342-4c10-400f-a67b-8934203c6f57"><emphasis role="bold"><emphasis role="strikethrough">where Bits Allocated (0028,0100) has a value greater than 8 shall have Value Representation OW and shall be encoded in Big Endian;</emphasis></emphasis></para>
                                        </listitem>
                                        <listitem>
                                            <para xml:id="para_4d1cadc6-627f-4b03-a541-4dd8ad6070d0"><emphasis role="bold"><emphasis role="strikethrough">where Bits Allocated (0028,0100) has a value less than or equal to 8 shall have the Value Representation OB or OW and shall be encoded in Big Endian.</emphasis></emphasis></para>
                                        </listitem>
                                    </itemizedlist>
                                    <note>
                                        <para><emphasis role="bold"><emphasis role="strikethrough">The OL Value Representation is not used for Pixel Data, even if it has a Bits Allocated (0028,0100) of 32, since OL was added to the standard after the encoding of Pixel Data had been established</emphasis></emphasis></para>
                                    </note>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_b9594f10-3fa3-41fa-a6ae-a76e799e7506"><emphasis role="bold"><emphasis role="strikethrough">Data Element (60xx,3000) Overlay Data</emphasis></emphasis></para>
                                    <itemizedlist>
                                        <listitem>
                                            <para xml:id="para_1ef77a76-e96d-42d0-8492-99d00f0e9684"><emphasis role="bold"><emphasis role="strikethrough">shall have the Value Representation OB or OW and shall be encoded in Big Endian.</emphasis></emphasis></para>
                                            <note>
                                                <para xml:id="para_13fc8f6e-fba8-4272-9a0b-aa3e4ac0dfde"><emphasis role="bold"><emphasis role="strikethrough">Previous versions of the Standard specified that the choice of OB or OW VR was based on whether or not Overlay Bits Allocated (60xx,0100) was greater than, or less than or equal to, 8. However, since only one bit plane can be encoded in each Overlay Data Element (60xx,3000), no value of Overlay Bits Allocated other than 1 makes sense. Such a restriction is now present in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->.</emphasis></emphasis></para>
                                            </note>
                                        </listitem>
                                    </itemizedlist>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_dc941892-c8bd-496c-8a03-f140fca3becc"><emphasis role="bold"><emphasis role="strikethrough">Data Element (5400,1010) Waveform Data has the Value Representation specified in its Explicit VR Field. The component points shall be encoded in Big Endian.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_dfcf801f-13e2-4c85-b8b7-67dd9896354e"><emphasis role="bold"><emphasis role="strikethrough">Data Elements (0028,1201), (0028,1202), (0028,1203), (0028,1204) Red, Green, Blue, Alpha Palette Lookup Table Data have the Value Representation OW and shall be encoded in Big Endian.</emphasis></emphasis></para>
                                    <note>
                                        <para xml:id="para_cfc28db1-0df8-4ad9-8c70-9c91a89df759"><emphasis role="bold"><emphasis role="strikethrough">Previous versions of the Standard either did not specify the encoding of Data Elements (0028,1201), (0028,1202), (0028,1203) in this Part, but specified a VR of US or SS in PS3.6-1993, or specified OW in this Part but a VR of US, SS or OW in PS3.6-1996. The actual encoding of the values and their byte order would be identical in each case, though the explicitly encoded VR field would be different. However, an explicit VR of US or SS cannot be used to encode a table of 2<superscript>16</superscript> elements, since the Value Length is restricted to 16 bits.</emphasis></emphasis></para>
                                    </note>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_9e119a97-50cc-4aaa-af6a-96d2e952862c"><emphasis role="bold"><emphasis role="strikethrough">Data Elements (0028,1101), (0028,1102),(0028,1103) Red, Green, Blue Palette Lookup Table Descriptor have the Value Representation SS or US (depending on rules specified in the IOD in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->), and shall be encoded in Big Endian. The first and third values are always interpreted as unsigned, regardless of the Value Representation.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_e9f6bbb8-e8ae-40ce-aa09-0d8b9cef0337"><emphasis role="bold"><emphasis role="strikethrough">Data Elements (0028,1221), (0028,1222), (0028,1223) Segmented Red, Green, Blue Palette Color Lookup Table Data have the Value Representation OW and shall be encoded in Big Endian.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_05a8c43f-69c3-4a06-bcfb-952f4ba8dcf0"><emphasis role="bold"><emphasis role="strikethrough">Data Element (0028,3006) LUT Data has the Value Representation US or OW and shall be encoded in Big Endian.</emphasis></emphasis></para>
                                    <note>
                                        <para xml:id="para_1d82fa01-b144-45c7-bd1e-581ff8f5dc8e"><emphasis role="bold"><emphasis role="strikethrough">Previous versions of the Standard did not specify the encoding of these Data Elements in this Part, but specified a VR of US or SS in PS3.6-1998. However, an explicit VR of US or SS cannot be used to encode a table of 2<superscript>16</superscript> elements, since the Value Length is restricted to 16 bits. Hence a VR of OW has been added. Moreover this element is always unsigned, therefore the VR of SS has been removed. The actual encoding of the values and their byte order would be identical in each case, though the explicitly encoded VR field would be different.</emphasis></emphasis></para>
                                    </note>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_6f1d82b3-e15d-4678-bb2d-1e886295a909"><emphasis role="bold"><emphasis role="strikethrough">Data Element (0028,3002) LUT Descriptor has the Value Representation SS or US (depending on rules specified in the IOD in PS3.3<!--<olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>-->), and shall be encoded in Big Endian. The first and third values are always interpreted as unsigned, regardless of the Value Representation.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_487d95eb-7b12-4728-9a69-c9f2e8692388"><emphasis role="bold"><emphasis role="strikethrough">Data Element (0028,1408) Blending Lookup Table Data has the Value Representation OW and shall be encoded in Big Endian.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_d7343588-4b14-453e-9b3a-a72463dabe50"><emphasis role="bold"><emphasis role="strikethrough">Data Elements (0066,0025) Vertex Point Index List, (0066,0024) Edge Point Index List, (0066,0023) Triangle Point Index List and (0066,0029) Primitive Point Index List have the Value Representation OW and shall be encoded in Big Endian and are always interpreted as unsigned.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para><emphasis role="bold"><emphasis role="strikethrough">Data Element (0066,0129) Track Point Index List has the Value Representation OL and shall be encoded in Big Endian and is always interpreted as unsigned.</emphasis></emphasis></para>
                                </listitem>
                            </itemizedlist>
                        </listitem>
                    </itemizedlist>
                </listitem>
            </orderedlist>
            <note>
                <orderedlist numeration="arabic">
                    <listitem>
                        <para xml:id="para_ca332a54-e4b4-441b-8cce-4d11198be63b"><emphasis role="bold"><emphasis role="strikethrough">For Data encoded with the Value Representation OB, the Data encoding is unaffected by Little Endian or Big Endian byte ordering.</emphasis></emphasis></para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_1e9f75ce-7ed6-4353-83a0-0dc95728b534"><emphasis role="bold"><emphasis role="strikethrough">Encoding of Curve Data and Audio Sample Data was previously defined but has been retired. See PS3.5-2004.</emphasis></emphasis></para>
                    </listitem>
                </orderedlist>
            </note>
            <para xml:id="para_46060c6d-66ba-4363-b747-79c7840e83c9"><emphasis role="bold"><emphasis role="strikethrough">This DICOM Explicit VR Big Endian Transfer Syntax shall be identified by a UID of Value "1.2.840.10008.1.2.2".</emphasis></emphasis></para>
        </section>
        
            <section label="D.1" status="2" xml:id="sect_D.1">
                <title>Detailed Example of Pixel Data Encoding</title>
                <para xml:id="para_797183b5-abb6-4308-a7e1-a1958dabf82d">All other (non-default) DICOM Transfer Syntaxes make use of explicit VR encoding. For these Transfer Syntaxes, all Pixel Data where Bits Allocated is less than or equal to 8 may be encoded with an explicit VR of OB (see Annex A<!--<xref linkend="chapter_A" xrefstyle="template:Annex %n"/>-->). As in the OW case, Pixel Cells are packed together, but in this case the Pixel Data is broken up into a stream of physical 8-bit words.</para>
                <note>
                    <para xml:id="para_e83c50bf-f06f-4c8e-87cb-f15f9ad9d31f">For Pixel Data encoded with an explicit VR of OB, the encoding of the Pixel Data is unaffected by <emphasis role="bold"><emphasis role="strikethrough">Little Endian or Big Endian </emphasis></emphasis>byte ordering.</para>
                </note>
            <para>...</para>
                <para xml:id="para_e472579f-fe77-42ba-aac6-98bb721fc5b0">Byte ordering becomes a consideration when we represent the Pixel Data physically, in memory, a file, or on a network.</para>
                <para xml:id="para_e7daa4d4-f668-4ff9-a024-de8fa98e1554">In the memory of a byte-addressable Big Endian machine, the highest order byte (bits 8 - 15) in each 16-bit word has a binary address of x...x0. While in a byte-addressable Little Endian machine, the lowest order byte (bits 0 - 7) in each 16-bit word has a binary address of x...x0. Figure D-5 pictures our example Pixel Data streams as they would be addressed in the memory of both a Big Endian and a Little Endian machine.</para>
                <para xml:id="para_388c8227-f0e2-413b-81b4-2856db00cebf">
                    <figure label="D-5" pgwide="1" xml:id="figure_D-5">
                        <title>Example Pixel Cells Byte Ordered in Memory (VR = OW)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-5.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                <para xml:id="para_bd0d30fe-e491-4c19-9fc4-dce65e96100e">Byte ordering is also specified as part of the negotiated Transfer Syntax used in the exchange of a DICOM message. Sixteen bit words are transmitted across the network (a byte at a time) least significant byte first in the case of a Little Endian Transfer Syntax <emphasis role="bold"><emphasis role="strikethrough">and most significant byte first when using a Big Endian Transfer Syntax </emphasis></emphasis>(see Figure <xref linkend="figure_D-6" xrefstyle="select: labelnumber"/>).</para>

                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">Replace the existing figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para>
                    <figure label="D-6" pgwide="1" xml:id="figure_D-6old">
                        <title>Sample Pixel Data Byte Streams (VR = OW)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-6_old.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">with the new figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para xml:id="para_c5c3919d-9058-42b3-a74f-967950864a4f">
                    <figure label="D-6" pgwide="1" xml:id="figure_D-6">
                        <title>Sample Pixel Data Byte Streams (VR = OW)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-6_new.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                
                <para xml:id="para_885ab505-2e41-4ee0-b7c4-f8ed5414afe6">As a last pair of examples, for Pixel Data having the Value Representation OW and the following attributes: 8 bits allocated, 8 bits stored, and a high bit of 7; the resulting byte streams pictured in <xref linkend="figure_D-7" xrefstyle="select: label"/> are as they would be transmitted across a network and/or stored on media. For Pixel Data having the same attributes, but having the explicit Value Representation OB; the resulting byte streams are unaffected by byte ordering and are pictured in <xref linkend="figure_D-8" xrefstyle="select: label"/>.</para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">Replace the existing figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para>
                    <figure label="D-7" pgwide="1" xml:id="figure_D-7old">
                        <title>Sample Pixel Data Byte Streams for 8-bits Allocated and 8-bits Stored (VR = OW)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-7_old.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">with the new figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para xml:id="para_4a8056ae-6ea0-49c6-a6aa-7b18e78aef7c">
                    <figure label="D-7" pgwide="1" xml:id="figure_D-7">
                        <title>Sample Pixel Data Byte Streams for 8-bits Allocated and 8-bits Stored (VR = OW)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-7_new.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>

                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">Replace the existing figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para>
                    <figure label="D-8" pgwide="1" xml:id="figure_D-8old">
                        <title>Sample Pixel Data Byte Streams for 8-bits Allocated and 8-bits Stored (Explicit VR = OB)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-8_old.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">with the new figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para xml:id="para_ce47df00-0694-4d94-a3c7-eb7f66c146fd">
                    <figure label="D-8" pgwide="1" xml:id="figure_D-8">
                        <title>Sample Pixel Data Byte Streams for 8-bits Allocated and 8-bits Stored (Explicit VR = OB)</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D-8_new.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
            </section>
            <section label="D.3" status="2" xml:id="sect_D.3">
                <title>Examples of Float and Double Float Pixel Data</title>
                
                <para xml:id="para_34948a5a-269f-4291-9837-c569e451e375">Float Pixel Data having the Value Representation OF always has 32 bits allocated; the resulting byte streams pictured in <xref linkend="figure_D.3-1" xrefstyle="select: label"/> are as they would be transmitted across a network and/or stored on media.</para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">Replace the existing figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para>
                    <figure label="D.3-1" pgwide="1" xml:id="figure_D.3-1old">
                        <title>Sample Float Pixel Data Byte Streams for VR = OF</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D.3-1_old.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">with the new figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para xml:id="para_163e3bac-a3ae-444d-b105-3c315ccf40e4">
                    <figure label="D.3-1" pgwide="1" xml:id="figure_D.3-1">
                        <title>Sample Float Pixel Data Byte Streams for VR = OF</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D.3-1_new.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                
                
                <para xml:id="para_505a9fc2-827e-495e-9b19-1be585bffa7e">Double Float Pixel Data having the Value Representation OD always has 64 bits allocated; the resulting byte streams pictured in <xref linkend="figure_D.3-2" xrefstyle="select: label"/> are as they would be transmitted across a network and/or stored on media.</para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">Replace the existing figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para>
                    <figure label="D.3-2" pgwide="1" xml:id="figure_D.3-2old">
                        <title>Sample Float Pixel Data Byte Streams for VR = OD</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D.3-2_old.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
                <section>
                    <title></title>
                    <informaltable rules="all" frame="box">
                        <tr><td><para><emphasis role="italic">with the new figure:</emphasis></para></td></tr>
                    </informaltable>
                </section>
                <para xml:id="para_4cbb376e-e99b-4a06-bc8b-1505d2297c20">
                    <figure label="D.3-2" pgwide="1" xml:id="figure_D.3-2">
                        <title>Sample Float Pixel Data Byte Streams for VR = OD</title>
                        <mediaobject>
                            <imageobject>
                                <imagedata fileref="cp1549/PS3.5_D.3-2_new.svg"/>
                            </imageobject>
                        </mediaobject>
                    </figure>
                </para>
            </section>