<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-1543</title>
    <subtitle>PET image library corrections</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
               <td rowspan="1" colspan="1" valign="top"><para>March 2016 Voting Packet</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2016/01/19</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xl:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Jörg Riesmeier</para><para><link
                    xl:href="mailto:mail@jriesmeier.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2015/08/24</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
               <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1543</para></td>
            </tr>
            <tr>
               <td rowspan="1" colspan="1" valign="top"><para>Log Summary: PET image library corrections</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Units should be specified as EV for the half-life.</para>
               <para>The Glucose Measurement Date and Time are only needed if the Glucose Measurement is present and it does not have an Observation DateTime.</para>
               <para>Fix incorrectly numbered rows.</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title></title>
        <section>
            <title></title>
            <informaltable rules="all" frame="box">
                <tr><td><para><emphasis role="italic">Amend DICOM PS3.16 as follows:</emphasis></para></td></tr>
            </informaltable>
        </section>
        
         <section label="TID 1607" status="3" xml:id="sect_TID_1607">
            <title>Image Library Entry Descriptors for PET</title>
            <para xml:id="para_8757771d-3325-4465-af5f-e365cb0dcc00">...</para>
            <table frame="box" label="TID 1607" rules="all" xml:id="table_TID__1607">
               <caption>Image Library Entry Descriptors for PET</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2f405261-e510-4283-8b74-1140b71e08c7"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ab60e888-00d7-4040-9e83-d7bc144d2709">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0da1adad-8b79-4b82-9f20-d188d906b02d">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f3474f32-4538-45ca-aa85-accc3153476c">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_176218c3-494a-4de5-801c-15eeecc70fd5">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_acab390c-bd81-432c-841a-05fde71ab819">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bb565f11-1948-4f95-adb8-0f6e0614f2b2">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6fe70009-05b6-4f5f-af09-d35d51a682f3">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bed078a4-7dc5-4a0e-8764-117e3700c651">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_42521206-a5d1-4332-a452-b8e95aa909ba">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e79e5b76-5ab7-434a-a9f0-8767e485787b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_318f9116-4cf4-4cbc-a13c-4348ff91c5ae">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_38ef416f-6052-430c-9660-de5b527c65f9">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_974a8e25-ae4e-48f3-b8ec-f9b20b05eca2">EV (C-10072, SRT, "Radionuclide")</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cc09263d-d44e-4215-b2c5-7e67386e0887">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a07dff21-b3b3-444c-8928-38784ad6a95c">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_42d92018-c359-43b4-97e6-8e945bb5424e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_525ae12a-5e0d-47d7-8fb3-9bf028a26785">D<xref linkend="sect_CID_4020" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66bc03d4-edd1-4089-a130-c48840b955ef">2</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_98d65662-738a-4203-adc1-7db9d443db6b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e437c6b6-0db2-4f46-a7a7-c3a703729e27">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fcab7301-b5d6-44b3-9292-6604b7f403c2">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a76b3eb1-62b6-484a-ab75-b5d1b2a9b354">EV (F-61FDB, SRT, "Radiopharmaceutical agent")</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_868f8ca6-8bbf-47dd-83e1-4371608424ae">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8309ddde-3c16-4bd5-91bf-255914e1109a">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_43dfefc0-5701-416b-87d7-b8309cc470d0"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_69fffba2-10e0-46b4-9ede-839324c8a30f">D<xref linkend="sect_CID_4021" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9e0dedde-ec48-42c3-948c-7c514b7caaca">3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f4173197-feb8-4a3e-be29-485316b82da2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_97e7b8c0-8087-4fe1-acf5-8a56fc73d8dc">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3dae6c66-256a-489b-952d-de8e2859c3ec">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1755fb95-afda-4ceb-a4f7-f3a7c5b9b0fc">EV (R-42806, SRT, "Half-life of radiopharmaceutical")</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_03e89c90-2fb0-4268-b956-1ffc57de4e45">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0b3f8549-07d2-4d40-9d4d-9238cab11e08">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a413e27d-c93c-44b8-a561-527a77a83438"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_81e5287c-b495-4afb-a0b5-6d0e827e1a4d">UNITS = <emphasis role="bold"><emphasis role="underline">EV </emphasis></emphasis>(s, UCUM, "s")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10a19c6f-c3f9-4248-8026-c23967973534">4</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d46a6b48-90d1-4ac6-8ecf-23d510235fed"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ce7af05c-aeec-4cbb-b8cb-613e3c5ccf11">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d8e00f4e-b172-4910-89d3-954ae978cd15">DATETIME</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4a1c688-1bc4-4838-9f30-9c66bc3b5770">EV <link linkend="DCM_123003" xrefstyle="select: nopage">(123003, DCM, "Radiopharmaceutical Start Date Time")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_eeb5b3e7-38ed-4f62-89a4-c74f6f776e67">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a297a7a0-eb10-4218-b0f0-465c41aa8d10">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4f159b28-f98e-4570-a4e1-aa11af595f7b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1"/>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4dc7b393-aa38-4f29-9b50-e367442fcd26"><emphasis role="bold"><emphasis role="strikethrough">4</emphasis><emphasis role="underline">5</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ad6acad4-1e52-4608-9ca8-baf11378fb76"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5bb35671-7df4-4c67-b925-24caa924e2a7">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b1036d55-6a08-4a61-81ee-545bcbdf4a22">DATETIME</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_67882427-b1b2-4737-98d9-96343632abdc">EV <link linkend="DCM_123004" xrefstyle="select: nopage">(123004, DCM, "Radiopharmaceutical Stop Time")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1b99c516-367e-46ac-ae67-0d78d45d655f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2200db31-6957-4df0-af3c-bdfb336c2023">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d19c572e-3361-4c80-a6c5-58722722e9c8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8c0109a4-1753-4a23-8be5-6b6ff3c5216f"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_178bf07b-889a-447d-ab8a-1ad627cd0925"><emphasis role="bold"><emphasis role="strikethrough">5</emphasis><emphasis role="underline">6</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_504a8e90-d930-4852-a790-29f5b5d8ce1a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2b3b1bf7-5b98-4867-98cb-3001bcf48e9d">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8799adb6-9ead-4c36-bd4e-ab274cde59d3">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eef276a8-bcaa-4079-a372-54ae410f43ef">EV <link linkend="DCM_123005" xrefstyle="select: nopage">(123005, DCM, "Radiopharmaceutical Volume")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_063a51d3-3515-4970-8233-a9c1dd7d84a4">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_83a692e9-5339-4695-8ff1-7e903472e9ba">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a4df6817-0183-467d-948b-e55f0183e120"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2073ff02-a2d1-4a5e-9d8f-8980195ea464">UNITS = DT (cm3, UCUM, "cm3")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_91230338-4fbc-44be-871b-a40b7e20b054"><emphasis role="bold"><emphasis role="strikethrough">6</emphasis><emphasis role="underline">7</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ae43bacd-649b-4128-831c-252ef28e7d9c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_43fbb90a-b340-4b14-a544-93055551dcb1">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ed1f4b5-dd15-4ca9-99c3-b9589a9da26d">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_11365024-167f-46e6-9a01-2f379d003775">EV <link linkend="DCM_123006" xrefstyle="select: nopage">(123006, DCM, "Radionuclide Total Dose")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_85c62f7a-0629-4488-9677-f48a8a00633a">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fb764399-9530-4050-b105-b68dd2266e06">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ec2f09bf-ad32-4639-a460-84ba15bd6632"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_095ae827-d3a9-4750-aede-9fd37e1bbfd5">UNITS = DT (Bq, UCUM, "Bq")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e82f319a-bdd4-4f7a-bde8-a4f7035f8dd7"><emphasis role="bold"><emphasis role="strikethrough">7</emphasis><emphasis role="underline">8</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7cb7f757-54a1-47a3-9040-75797ea23652"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_51973ff6-5855-4de5-bd02-7375915f5a8f">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_233cb432-21fc-4034-9d4a-c938c1732433">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c85990c1-c9fa-49d9-9764-04cd7f6186f1">EV <link linkend="DCM_123007" xrefstyle="select: nopage">(123007, DCM, "Radiopharmaceutical Specific Activity")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5f06e3b3-536f-4938-8f0d-0dc67ed3259e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ca09f0eb-38d9-4d9c-9dd0-b97dd48569a4">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_72a09b87-1903-4de1-b8ca-6b4cb5077220"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10de3cfa-550f-46b2-a45f-e12d283ec405">UNITS = DT (Bq/mol, UCUM, "Bq/mol")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_945e170a-11e2-46ab-aba1-7f405251463e"><emphasis role="bold"><emphasis role="strikethrough">8</emphasis><emphasis role="underline">9</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c21403c8-592f-4f47-8088-a32f049a9d84"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2a2fbcc8-d5be-4b41-8ba7-0c52d5031065">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5b93ec38-e0c4-4798-a246-e5ee82aea344">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6f91f1c8-ab2d-4a7c-85ce-55fddae16935">EV (G-C340, SRT, "Route of Administration")</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fb1a1c3a-e7e7-49c5-8703-b8ffe37247d9">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d714ab82-c99f-4ef1-88ad-7177b7a1121b">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4f0f6466-d357-4fba-a7df-0088a37c6309"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_34f8403a-8a36-445b-ab06-66609688e5e8">BCID 11<!--<xref linkend="sect_CID_11" xrefstyle="select: labelnumber quotedtitle nopage"/>-->
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_15bb16fb-4a0b-48ad-825a-ed6efcd76f53"><emphasis role="bold"><emphasis role="strikethrough">9</emphasis><emphasis role="underline">10</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1e76beba-98f1-49bb-b453-26b1fb20e563"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_27824a34-e2b6-421a-b3a1-b58bdfa1f3c1">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_04e0a575-4c80-4736-af6b-a102642cd51a">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d05bcfad-bc87-408b-a124-b70d433004ef">EV <link linkend="DCM_123009" xrefstyle="select: nopage">(123009, DCM, "Radionuclide Syringe Counts")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d8970f02-48a4-431c-887a-930937014622">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_334094e1-fda8-4350-8259-c717597d3006">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5838c892-7415-40e5-9527-9e72e4abcf9e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_948b61d7-f837-4349-814d-82afee09b254">UNITS = DT ({counts}/s, UCUM "counts/s")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a4e179d8-34ae-480b-869d-b5795cde5c03"><emphasis role="bold"><emphasis role="strikethrough">10</emphasis><emphasis role="underline">11</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd28c70d-b2f8-4e7b-a570-5add3e98375a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5e000bcb-1f19-4bc2-b495-ef13a01c713d">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_82640e9f-a2c3-48fe-9e10-311417507a1a">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c5fb8076-0438-4c7b-a891-c3845e438b1a">EV <link linkend="DCM_123010" xrefstyle="select: nopage">(123010, DCM, "Radionuclide Residual Syringe Counts")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_66e77bb6-5bfa-4ae0-828b-fd2fd535d5cc">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f0c2b460-ddbc-47b8-98f4-61cb58be76db">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_85503189-5820-4fad-8021-1943a6c8d5bd"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9c5dc25e-7f62-48a4-ae30-1f61312aa3a6">UNITS = DT ({counts}/s, UCUM "counts/s")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">10</emphasis><emphasis role="underline">12</emphasis></emphasis></td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ae1b9db5-d681-4c75-917a-e93baa5e209c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_93c45947-7926-41ad-8bf4-066c92a6a162">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_76378be5-8d85-4115-8092-c6a0479a2ae6">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_01ad7f78-088f-458f-a2f8-67962d0bbf40">EV <link linkend="DCM_126203" xrefstyle="select: nopage">(126203, DCM, "PET Radionuclide Incubation Time")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a1e0a84e-6fa6-41e5-9ee0-26d666e44581">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_980b6964-15d1-4918-914c-2abc5c5acf74">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_313007b5-9cca-4be0-b596-72366baf63ea"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f042955c-33dc-440a-8605-d83726d5ded8">UNITS = EV (min, UCUM, "min")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c9a20a40-ec20-4e6b-bf9b-5c6176c5514f"><emphasis role="bold"><emphasis role="strikethrough">12</emphasis><emphasis role="underline">13</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3b404b2c-d801-4f69-b455-16f0f9dbd54a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e804d2db-9d1e-4309-9216-de81fcb31a24">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5af5e3fb-0c95-4cee-b1a8-16a7c756caf7">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66312693-4379-4070-9ff9-3c04daeb5a12">EV <link xl:href="http://s.details.loinc.org/LOINC/14749-6.html" xl:show="new">(14749-6, LN, "Glucose")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bcbd1765-80ac-4278-978f-2cfe687230ae">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_949ef4c7-2eb2-4a19-aa8f-fbb0f66f7e77">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d298ac23-9eff-4721-b83c-63fed328befa"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dc8e3db1-b1cf-455e-8a72-fcc4e76fe918">UNITS = EV (mmol/l, UCUM, "mmol/l")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1"><emphasis role="bold"><emphasis role="strikethrough">13</emphasis><emphasis role="underline">14</emphasis></emphasis></td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bef5afb8-0924-4ded-baf5-e180269a6e29"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4449c649-a81c-4827-a474-1cca3502f9c3">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_02900285-f0fe-4322-9c29-6176dfc655a3">DATE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_928bb41c-26ab-4eea-b286-b06067786a98">EV <link linkend="DCM_109081" xrefstyle="select: nopage">(109081, DCM, "Glucose Measurement Date")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b409354c-e4a7-4fbf-b28e-36f18a46fabb">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6d1b344d-823a-4968-a7b5-dc6b8e405212">M<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5cad91c-f163-42f3-813e-1ae7748955bf"><emphasis role="bold"><emphasis role="underline">IF Row 13 Glucose is present and does not contain Observation DateTime (0040,A032).</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd7fa0e0-b227-4ade-912f-71151c345a2a"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9a91fe48-e3ab-4eb7-9afc-f65cf770ea79"><emphasis role="bold"><emphasis role="strikethrough">14</emphasis><emphasis role="underline">15</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_88afb275-1f21-4954-b689-fc7d7502f591"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_df26e9f1-c588-4fe8-9f2c-49e933676ac4">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e1f91711-42f1-405a-a0c1-16e296baa980">TIME</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_82e31834-6572-4f50-ab1e-32ec162fa094">EV <link linkend="DCM_109082" xrefstyle="select: nopage">(109082, DCM, "Glucose Measurement Time")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ada74815-033d-4587-8400-d7a13a11ca57">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bf7445c6-1349-49dd-b1c6-92759118cbb2">M<emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_430f77da-3c96-4494-a09d-fd80ff0df524"><emphasis role="bold"><emphasis role="underline">IF Row 13 Glucose is present and does not contain Observation DateTime (0040,A032).</emphasis></emphasis></para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_940f6c75-0581-4fd0-bdea-a7324c5f6731"/>
                     </td>
                  </tr>
               </tbody>
            </table>
            <para xml:id="para_25176600-ff6d-49a1-b369-6467fe348caf">
               <emphasis role="bold">Content Item Descriptions</emphasis>
            </para>
            <informaltable frame="box" rules="all">
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_731e9af9-8f6c-4c30-b3a8-99f889c447b8">Half-life of radiopharmaceutical</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2a8e97a0-ce2f-4e3b-8459-e46d748e92bd">The units for half life are chosen to be seconds, to match the units used for Radionuclide Half Life (0018,1075). See PS3.3 Section C.8.9.2<!--<olink targetdoc="PS3.3" targetptr="sect_C.8.9.2" xrefstyle="select: docname label quotedtitle"/>-->.</para>
                     </td>
                  </tr>
               </tbody>
            </informaltable>
         </section>


    </chapter>
</book>