<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
    <title>CP-1542</title>
    <subtitle>Specify units for positioner angles in image library</subtitle>
    <chapter>
        <title></title>
        <informaltable rules="all" frame="box">
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2015/11/11</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>Jörg Riesmeier</para><para><link
                    xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:mail@jriesmeier.com"/></para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
                <td rowspan="1" colspan="1" valign="top"><para>2015/08/24</para></td>
            </tr>
        </informaltable>
        <informaltable rules="all" frame="box">
            <tr>
               <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1542</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Specify units for positioner angles in image library</para></td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Name of Standard</para>
                    <para>PS3.16</para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">
                    <para>Rationale for Correction:</para>
                    <para>Units should be specified.</para>
                    <para><emphasis role="italic">[Ed.Note: There is inconsistent usage throughout PS3.16 of a code meaning of "degrees" versus "deg"; should these be harmonized?]</emphasis></para>
                </td>
            </tr>
            <tr>
                <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
            </tr>
        </informaltable>
    </chapter>
    <chapter>
        <title></title>
        <section>
            <title></title>
            <informaltable rules="all" frame="box">
                <tr><td><para><emphasis role="italic">Amend DICOM PS3.16 as follows:</emphasis></para></td></tr>
            </informaltable>
        </section>
        
                 <section label="TID 1603" status="3" xml:id="sect_TID_1603">
            <title>Image Library Entry Descriptors for Projection Radiography</title>
            <para xml:id="para_5df46815-9377-4578-bbda-b6224d64fe9c">...</para>
            <table frame="box" label="TID 1603" rules="all" xml:id="table_TID_1603">
               <caption>Image Library Entry Descriptors for Projection Radiography</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3b86817d-6979-4682-8b78-096a8a6b3afa"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bcc4da96-f5a0-4e2c-8da3-24854d06de9a">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1bcce8f5-150f-41ae-9a2d-23eb9e746470">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f5c77ee3-3353-4052-a2e4-520f953582ad">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_05dfc679-67e0-4944-bf6b-76c2213e0f97">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_db5d1c0d-04fe-4d75-af76-ba258a93edd3">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7d7a1024-4b31-484d-97b3-47a76fbce065">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_be5280d2-1409-49c8-8267-e7adc5b898a8">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_46aafe31-c818-4258-b4d8-4e27978ecb01">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_98f18042-d5fe-4e6e-8919-34c2e082a244">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_93ff1b87-8240-4c9b-84e5-93e082ecbdc3">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_092f6bd3-b600-4c00-a83f-18ff4c7184e3">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c42aab89-98d6-4b5d-aa87-2c7ebc2539b9">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_17a48035-e68f-4055-a70e-ca7f598befab">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6bf9884b-4bc3-4d4c-b04b-76ca990aafc1">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4af6573c-84ca-4197-ac8b-bf795f2233be">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9b883813-dbdd-4620-b44a-e785942bd610">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2b8eec03-f424-4c29-962b-9ae11b2efed9">...</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_27be1d08-febd-49f6-b8b4-9c4896747741">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_93362fda-1073-41b3-a930-f24e06ecc6ce"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8a464664-c75f-41f4-acd0-b5037c559f8d">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e580f06d-1db5-4849-9c7a-0b13792d7b87">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d0894c98-b957-4b4b-81ba-e8f2b913e344">EV <link linkend="DCM_111026" xrefstyle="select: nopage">(111026, DCM, "Horizontal Pixel Spacing")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_012dc81b-a76f-4dc1-971f-1583f1064b23">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bf8f1aa0-d3ff-41a9-b108-806bfdca1464">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5c19ac87-026a-4d8c-8e6a-edd8bfb230a2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1"><para>UNITS = EV (mm, UCUM, "millimeter")</para></td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_232a24b6-d399-4218-a960-a0102529287a">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_364361dd-fb63-426d-9b98-25900af09bf5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_36df9117-6de2-4df7-8968-5383f884aa5d">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ac464c73-87fc-432d-b886-176caae412a1">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e99e5c2e-a78c-4444-89d3-27ccd651ef8e">EV <link linkend="DCM_111066" xrefstyle="select: nopage">(111066, DCM, "Vertical Pixel Spacing")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4517d388-e4a8-49d5-b0bc-56770d048351">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2db120ed-d561-4f68-94c8-ab3e721f0480">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4154c864-4cc3-40b8-bf34-8722d09038b4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ee574f48-a453-483d-993e-2a3b4e326581">UNITS = EV (mm, UCUM, "millimeter")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1820f53d-e1d3-4097-bd89-8c920bf57382">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5d7c6ec5-9e1e-4523-aa1c-c158227d0202"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c8f15e32-97b2-4e0e-9e4a-bcca98d3ca86">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5798fd47-6ee0-41bd-8f39-4564d2754a96">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8c2003a1-a6cd-4775-b216-b50e954d074c">EV <link linkend="DCM_112011" xrefstyle="select: nopage">(112011, DCM, "Positioner Primary Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_41e516a9-e073-4867-a07f-718ef5647cfb">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_108bbc63-dc1c-43a2-9c27-13fa073c949e">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_679514a2-7bb6-4712-acc7-820906b92944"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_6a5a8493-8264-47c0-997e-5c9e74ee4ff2"><emphasis role="bold"><emphasis role="underline">UNITS = EV (deg, UCUM, "deg")</emphasis></emphasis></para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fa7f4c3d-800d-4035-95bd-032050b0fb77">8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_04cc0c4e-9fb7-472d-9e95-ac7a96c8ffef"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fad642e8-e362-4c9f-ace8-9f422e629bbf">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_31332c3a-18d0-4358-b59e-ec765aaf86a4">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_921a3c3a-dc4f-42c6-896b-dcce65d2fec6">EV <link linkend="DCM_112012" xrefstyle="select: nopage">(112012, DCM, "Positioner Secondary Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1f670ae5-430c-4501-b352-817ab406c01c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bca90686-bcd9-492a-961d-7fb9f93c7709">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cd0e8035-fffe-4af8-a238-f95f418a3c1b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_4ac71f89-12a8-462d-83d1-e5cd42a29cbc"><emphasis role="bold"><emphasis role="underline">UNITS = EV (deg, UCUM, "deg")</emphasis></emphasis></para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
        
        
                 <section label="TID 3101" status="3" xml:id="sect_TID_3101">
            <title>Image Acquisition</title>
            <para xml:id="para_c32b3845-27c8-467a-bc62-c3159f1cd074">...</para>
            <table frame="box" label="TID 3101" rules="all" xml:id="table_TID_3101">
               <caption>Image Acquisition</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_10cb6a10-ed16-4deb-9b9b-7a1cb070a7d7"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7af8ef32-e7cc-4f79-b633-f23e9d03dd96">NL</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_987b49f4-935e-45d7-af95-c0078b313f70">Rel with Parent</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3158d279-0fdd-4d9a-a01d-4cd805ec1dbd">VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_eaf104c5-8a08-46f6-ac2c-45321e62cf14">Concept Name</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_63bcbab5-d232-4fb0-b478-83333e384fb5">VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2227775d-fc72-4e23-8759-68cca47c8f10">Req Type</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ccb381be-c983-4ca2-b89e-2e6d4f0258af">Condition</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_631cc9d0-f13d-4091-824c-385e1d10dbf5">Value Set Constraint</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e44c60ca-ca3e-4509-96bf-f8448a5af3a5">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9d889441-1b18-41f7-9ec3-f091913a7d7c">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_177d40e6-0b33-482c-b360-4b00f27761e2">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_57adf028-737c-47c7-b248-94c9a23ee6ec">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_30dc5606-259a-4c53-88b1-8b177fac7375">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_086de62a-79f5-4c4e-9b60-818231ac3875">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7d219abc-8f21-4d6e-9cf7-7f553d8fa04a">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_58034dce-6f89-42f2-84d5-782567310ae6">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_819a3675-3b5b-40a4-948e-874cbe1187d3">...</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c843559c-9d5e-4e83-81bf-e283a50d4001">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_73e00536-cd63-4282-8b4f-1396d872cccb">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4bedc660-101a-4159-a114-ff9d771e4b0e">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f2f9cdd5-e356-45c7-bf67-81ab3d1ac2c0">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_056c4c14-70fa-4252-93be-7f20f3d464b8">EV <link linkend="DCM_112011" xrefstyle="select: nopage">(112011, DCM, "Positioner Primary Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_493867a8-5962-4eea-bbcb-d57dd2514563">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cbc2fa84-604e-4213-8b4c-6eb6812f8ed5">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a1deac85-5cb9-468a-a18b-9260b7b4d081"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_59e9849f-417e-42d5-ae94-f39a9dc121d4">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_04095599-992f-42c9-833c-93f4d94038d1">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fdc9bc44-476b-40f2-9770-671d4a0d59ff">&gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_51f282f3-177d-40ea-8bb6-ce4901eff132">HAS ACQ CONTEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_216ce616-8aff-446e-8243-fc53c6050c35">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4d9e1ee-8167-4c2f-8d7c-4310a2307417">EV <link linkend="DCM_112012" xrefstyle="select: nopage">(112012, DCM, "Positioner Secondary Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_507cc233-0770-42a9-a007-84d6733f9c9f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1087d465-b8f0-4319-bc54-c97ffc450e98">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_12f8b55b-04b7-4a23-a36b-3637063930c6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4173502a-3795-4d86-990a-862cce7bc4e6">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_251d3de0-b95e-4e8c-aab0-200b8f548982">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_63bc96e7-1d85-4f9f-a581-eabb6ab558f9">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_064955fb-1c5c-46f3-8153-10e14675f7b3">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6aba1818-7fdd-416b-9ba4-b4922eac6880">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ead80ce-369f-4fda-b9a9-4dc8d77a1870">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1f43e96d-0bc1-442c-afa8-20c2b4153821">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4888d1b2-fce9-4c0a-a908-df0c91826b1b">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8eed7121-a58a-48a3-aedb-f2337b145fa1">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_06d5db24-86b1-4925-a4a1-c2e41d7b567b">...</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>

    
         <section label="TID 10003C" status="3" xml:id="sect_TID_10003C">
            <title>Irradiation Event X-Ray Mechanical Data</title>
            <para xml:id="para_dc68623c-1ced-4414-9bae-b893a11c1598">...</para>
            <table frame="box" label="TID 10003C" rules="all" xml:id="table_TID_10003C">
               <caption>Irradiation Event X-Ray Mechanical Data</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2e9d8d13-d218-426d-9244-b39516fdeefe"/>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4c771417-8382-40fb-9aac-5463f3cefcd1">
                           <emphasis role="bold">NL</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4ab59d78-eb23-4d77-91d0-13ea59e516e5">
                           <emphasis role="bold">Rel with Parent</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_af3085f8-a3aa-4361-bd3c-17524f3c7160">
                           <emphasis role="bold">VT</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6760cc50-840e-439f-a16b-8cb616a42d9d">
                           <emphasis role="bold">Concept Name</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8ad27284-77ff-4a2f-a0ba-0dafdf1a0844">
                           <emphasis role="bold">VM</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9027dbc4-f522-4386-b739-318b8cd8c77f">
                           <emphasis role="bold">Req Type</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b6e92417-bd4b-466e-b2a4-14f481cbf989">
                           <emphasis role="bold">Condition</emphasis>
                        </para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2ae106ff-b403-4dbb-a1b6-3c87e79b51e0">
                           <emphasis role="bold">Value Set Constraint</emphasis>
                        </para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2bde75c7-5e14-42b0-9e21-f3d7d22170bf">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_94bc1dcd-96ff-48f5-8e1a-fd41de3ca3ef"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_272e3c17-82c4-4505-9d73-62156a3e61a2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a8724b8b-7f70-49b3-8622-a1c234a6f3c2">CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2ad8297b-ffa2-4c57-a32e-44db1895a8b6">EV <link linkend="DCM_113956" xrefstyle="select: nopage">(113956, DCM, "CR/DR Mechanical Configuration")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8fe6e9f1-4bf6-43dc-88db-9d4aa40d328f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a00711f8-98f6-43eb-8af0-9f9141564d4a">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dfaf89e4-3d0c-4349-b15b-b1c118f79e35"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8f8a5baa-6da4-4ce8-969e-2b456bef6677">D<xref linkend="sect_CID_10031" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_03ad2cf4-6669-4723-a88f-d361e35473eb">2</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9e55cc1b-4c45-486a-a739-52e0cc082c8e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_01234826-0787-4b90-a46a-5eb0220fd096"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3f2b9aad-5be0-420b-8d70-fb993a1596f1">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6498d3ec-2cc7-492a-9371-6d52d900e196">EV <link linkend="DCM_112011" xrefstyle="select: nopage">(112011, DCM, "Positioner Primary Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_471f6c0f-228c-4720-87e0-666382b4ec45">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_13c62647-1af4-4ce9-9f1b-3a8f4a4aa0e3">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2ff1a1ce-3b33-4fdd-9b29-3b630190d87c">XOR Row 6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7fefd881-5518-4b5b-8e8c-9a121f494e49">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9520dbf0-b0ae-4f09-8b2d-6331e7158866">3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0bfdaf0b-c7d6-4630-b60b-804c8131b931"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_88c39cbf-2e22-466e-a96d-f1306dee784a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_074068b2-4dd9-4dd2-857f-cd3397969204">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66edbea8-80fa-4bf9-9abd-ce8d08819b47">EV <link linkend="DCM_112012" xrefstyle="select: nopage">(112012, DCM, "Positioner Secondary Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ce05b182-a7d5-4f29-b3d1-96e47ad0274c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bd03ca51-9d7c-427b-952f-8c01346d9f1d">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8b504b76-ad11-43eb-9f71-f84a1c281209">XOR Row 6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d6fd0972-d1c8-4210-b197-4ed18827f6ab">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5c320ff3-46bb-44dc-9cbd-7db429443d75">4</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_07321c34-cf05-4b21-aade-d3ae61b3013a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c002d754-cda9-4a08-9382-b8552b72fba8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc66edf4-06d7-4894-b0d6-6523719728eb">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_064263e5-541f-469a-8569-829a353b2ef1">EV <link linkend="DCM_113739" xrefstyle="select: nopage">(113739, DCM, "Positioner Primary End Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_288c49ee-5bde-478d-80c7-b7f27240ad50">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d909dec5-6e66-4392-9ea6-952e278b32a8">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_89936d7e-cd00-4903-a00d-0927273fee9f">IFF TID (10003) Row 7 value = <link linkend="DCM_113613" xrefstyle="select: nopage">(113613, DCM, "Rotational Acquisition")</link>
                        </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd7e0a59-0ec4-4e37-b7cd-1bd37c5909d4">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a0bb852f-6e55-4d8f-ac50-af98df2f7707">5</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5e10c2b6-5361-478c-933c-ba488d261770"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_08820abc-5ae1-419e-bfed-e2814bd9918b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_63643434-db3e-47ae-8aa3-3d1c3c288743">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ca9feac1-eba0-414a-985a-66301eec26c9">EV <link linkend="DCM_113740" xrefstyle="select: nopage">(113740, DCM, "Positioner Secondary End Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f091e135-b52b-4da1-907b-1cc4f2d5582e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2b93dce8-ed84-430a-890c-485394231793">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_894c8279-076c-47b4-8ac2-f27186e7a9c0">IFF TID (10003) Row 7 value = <link linkend="DCM_113613" xrefstyle="select: nopage">(113613, DCM, "Rotational Acquisition")</link>
                        </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1c2af4e8-fc6b-4a11-a205-f68d8b0776a9">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5ab1374a-f441-437b-8f87-6635b8f9c6f5">6</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e0ed1466-d851-482a-a80f-218bc7ec4342"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ccc2b740-dcc8-4f4f-bdf1-6b5cf7ba9263"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e4af3e95-0fb7-43f3-9ec7-fa3a8586a533">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0695c976-ce42-4258-a2a3-743b171ef5f6">EV <link linkend="DCM_113770" xrefstyle="select: nopage">(113770, DCM, "Column Angulation")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8e5e803d-7f31-425f-a0f6-6fec56355bd7">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fda8c010-ba23-4ce9-bf2c-b257eecd9ccb">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_634db4db-d850-4f6e-b000-649d9e3fc292">XOR Rows 2, 3</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a49e0859-6327-4893-809b-9576c68c0ce2">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8b4eb671-07b8-4207-b12e-31d8282cd693">7</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ad4b6c39-978d-488b-81e2-09e9489b1a6e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7b125385-7832-4935-9b9b-339970223092"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_81b1cf39-f9d7-48db-8951-74fbae89eab6">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_60c2b3c9-9ed3-4459-acaf-3e85f07a2ec8">EV <link linkend="DCM_113754" xrefstyle="select: nopage">(113754, DCM, "Table Head Tilt Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6f8849b9-933a-45d2-84da-3b171bc43478">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8d8b06e6-6520-45dd-a3d8-21d696e4f1b8">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6fc4a039-6e2c-4883-b550-55b4284c6d31"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_812c25d5-8131-4dbb-a765-da98e036ed85">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_81e9e2e1-6f8c-4a0b-8b53-c2dfc14bc165">8</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bb95cb0d-a2f4-4bca-94c7-5335b1a7c819"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_611b3e09-f85d-4212-ae62-5f74381c447c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6211c88d-716a-4941-b600-7fadd511b104">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2fa4a6b3-0e3a-4824-bb45-201603e6b37c">EV <link linkend="DCM_113755" xrefstyle="select: nopage">(113755, DCM, "Table Horizontal Rotation Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d732dc16-181d-41b1-b403-b2318c287e85">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_73430765-d22f-4df5-a17c-71371522c0bc">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b514797a-96fd-44a5-8c36-8f4df5380da2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c924de4c-daae-48fc-8245-cbc86216e3eb">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_28c3c37b-4920-49af-b8ef-c41ab403897e">9</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_62fbb0ff-c589-4892-b3a3-273823277abb"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_88e3d2eb-d38b-42e1-98b9-d24b5e491fe6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c2574847-df27-4c9a-88ef-8cfbab793d2d">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b294faba-1fb3-4137-a716-3777275a662d">EV <link linkend="DCM_113756" xrefstyle="select: nopage">(113756, DCM, "Table Cradle Tilt Angle")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a30fa836-5da9-46fa-a2eb-9332e613428b">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_35221189-5e28-4921-9ed8-45b9a4e7d5c5">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10f4b641-ca2b-4139-8e78-c7069dd0bf48"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d4af1cf6-8ad1-49f4-8afb-caf0f1a7ccbd">UNITS = EV (deg, UCUM, "deg")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_16255d2c-ef89-4675-be5d-406a720f6578">10</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_78571c59-c5b5-4a4d-a8f0-3d420cdb5551"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9b795e7f-5ff1-4c09-82c2-b6a4381c9432"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0b0f46b9-0d52-48a2-89db-c4f36632f9ed">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0a23c328-6622-4d71-b7aa-7a672c8df5bf">EV <link linkend="DCM_111633" xrefstyle="select: nopage">(111633, DCM, "Compression Thickness")</link>
                        </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_50105ba3-c20e-48be-8f9e-b64d92620b57">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4a5f365a-0ad4-4b36-a078-ee0c2faa374c">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a10afe7a-1d03-421a-836d-ca5b04fe2271"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a6d156f7-1f7a-4237-981d-4daef9c24e58">UNITS = EV (mm, UCUM, "mm")</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f7b36b41-692b-4c13-8d49-0dda30062283">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_603960dc-6f46-44b8-964b-e6d8f50e399a">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7fe962d8-6637-4288-a73c-59dfe354bbfa">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0acc52ce-a121-4d89-aeec-2756673c7be1">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6fe56613-0659-43d7-9e15-9dce82e991ff">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4c1ef3e4-e9a5-4645-8bc9-0a2216a77c79">...</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_da7ad4b5-d447-4f9d-9aaa-b99f509bb533">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8cb7e0d6-d5d7-4607-afa0-e5787f69a050">...</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0ac4829e-5272-4339-8daf-38ad787eee63">...</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>


    </chapter>
</book>