<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-1535</title>
  <subtitle>Additional SUVlbm formulas</subtitle>
  <chapter>
    <title></title>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>January 2016 Voting Packet</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>2015/11/11</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
          xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>QIBA FDG-PET group</para><para><link
          xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:mail@jriesmeier.com"/></para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>2015/09/09</para></td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1535</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Additional SUVlbm formulas</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.3, PS3.16</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
          <para>Larger patients are not accurately described by the conventional (James) LBM formula, and an alternative formula has been defined with more robust characteristics across a larger range of patient sizes.</para>
          <para>Add appropriate codes for the additional formula both for the SUV Type attribute used in images and the corresponding concepts used in real World Value Maps and Structured Reports, and link the enumerated values for SUV Type to the PS3.16 concept definitions.</para>
          <para>Also correct the order of curly braces unit annotations (annotations need to follow, not precede, the units if not unity).</para>
          <para>Also change the reference to CID 84 to CID 85,. since SUV units have been factored out of PET units into their own context group.</para>
          <para>The existing Enumerated Value for LBM by the James method is not renamed due to the potential impact on the installed base.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td><para><emphasis role="italic">Amend PS 3.3: </emphasis></para></td>
        </tr>
      </informaltable>
    </section>

<section label="C.8.9.1" status="4" xml:id="sect_C.8.9.1">
          <title>PET Series Module</title>
          <para xml:id="para_4d39efd9-51ff-4582-b980-e2136d02962c">
<xref linkend="table_C.8-60" xrefstyle="select: label"/> contains IOD Attributes that describe a PET Series.</para>
          <table frame="box" label="C.8-60" rules="all" xml:id="table_C.8-60">
            <caption>PET Series Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7508cef2-d373-4cb4-ab01-16f5a978f8a5">Attribute Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7b71e149-2d25-4ca6-a50d-c2ffef1e9704">Tag</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e4dcd9fd-3650-45d8-a5cd-3dfca89e99e6">Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8ce02659-d8cf-4c44-9ea0-b121b53d9bc7">Attribute Description</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0474fd1e-b1f3-4b6b-95b4-ca32b0a63e82">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6f095005-7b76-41f8-b194-3f4121107c93">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f209fc29-74c3-4892-b815-9dcd1f31703b">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1fb17223-c8e0-4f62-83dd-1d91c64d7b21">...</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d7e96cbb-906b-4f3c-a711-9b7414f4ba65">SUV Type</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e96eebfc-a7ea-4c2d-87c7-d5aa08ac6db2">(0054,1006)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9f173004-d315-4903-8f0e-70adffed0a06">3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c1d928f9-845e-4c0d-b091-2cee8e17ab56">Type of Standardized Uptake Value (SUV).</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>BSA</term>
                      <listitem>
                        <para xml:id="para_1f09a3e1-fb52-4bb3-8725-22f2fc8a9d2b">body surface area</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>BW</term>
                      <listitem>
                        <para xml:id="para_590e6719-78d0-46fb-8df7-b0470b715f57">body weight</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>LBM</term>
                      <listitem>
                        <para xml:id="para_56343d9d-0b3b-41d7-b5a4-30d9c20ef0b5">lean body mass <emphasis role="bold"><emphasis role="underline">by James method</emphasis></emphasis></para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term><emphasis role="bold"><emphasis role="underline">LBMJANMA</emphasis></emphasis></term>
                      <listitem>
                        <para><emphasis role="bold"><emphasis role="underline">lean body mass by Janmahasatian method</emphasis></emphasis></para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>IBW</term>
                      <listitem>
                        <para>ideal body weight</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <para xml:id="para_c1200938-84b8-4a3a-b0f4-fe43359e9b7e">If absent, and the Units (0054,1001) are GML, then the type of SUV shall be assumed to be BW.</para>
                  <note>
                    <para xml:id="para_522db716-4502-4fd7-875c-607fc7291ca6">The type of SUV cannot reliably be deduced from the units alone, i.e., SUVbw, SUVibw and SUVlbm <emphasis role="bold"><emphasis role="underline">(James or Janmahasatian)</emphasis></emphasis>all have units of GML.</para>
                  </note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_07430a57-9140-4d67-989e-75fbb700b002">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d80a185e-75f7-4b03-b7b0-98851062bfd7">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6fa587e0-7130-46b4-b420-55c21c8ebc5b">...</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9b214656-c8f5-405f-b021-0d793413e608">..</para>
                </td>
              </tr>
            </tbody>
          </table>
          <note>
            <para xml:id="para_6dc98ecd-8792-4a35-91c7-e817b2988a3e">The formulas for the determination of SUVbw, SUVbsa, SUVlbm<emphasis role="bold"><emphasis role="underline">(James)</emphasis></emphasis> and SUVibw are defined in Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction. Radiology, 1999</emphasis> at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>.
              <emphasis role="bold"><emphasis role="underline">The Janmahasatian LBM formula is defined in <emphasis role="italic">Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051–65.</emphasis> at <link xl:href="http://dx.doi.org/10.2165/00003088-200544100-00004"/>
                and its role in SUVlbm(Janma) calculation is discussed in <emphasis role="italic">Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging. Journal of Nuclear Medicine. 2014 Sep 1;55(9):1481–4. </emphasis> at <link xl:href="http://jnm.snmjournals.org/content/55/9/1481"/>.</emphasis></emphasis>
              See also
              <emphasis role="bold"><emphasis role="strikethrough"><xref linkend="sect_CID_84" xrefstyle="select: labelnumber quotedtitle"/><!--<olink targetdoc="PS3.16" targetptr="sect_CID_84" xrefstyle="select: docname labelnumber quotedtitle"/>--></emphasis></emphasis>
              <emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_85" xrefstyle="select: labelnumber quotedtitle"/><!--<olink targetdoc="PS3.16" targetptr="sect_CID_85" xrefstyle="select: docname labelnumber quotedtitle"/>--></emphasis></emphasis>
              .</para>
          </note>
        </section>

    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td><para><emphasis role="italic">Amend DICOM PS 3.16:</emphasis></para></td>
        </tr>
      </informaltable>
    </section>

    <section label="CID 84" status="3" xml:id="sect_CID_84">
      <title>PET Units</title>
      <variablelist spacing="compact">
        <varlistentry>
          <term>
            <emphasis role="bold">Type:</emphasis>
          </term>
          <listitem>
            <para xml:id="para_d2513e41-d059-43a3-bf65-46d14ecb6f78">
              <emphasis role="bold">Extensible</emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term>
            <emphasis role="bold">Version:</emphasis>
          </term>
          <listitem>
            <para xml:id="para_74089d91-8980-49d8-88c2-7733d3063c89">
              <emphasis role="bold">20141110</emphasis>
            </para>
          </listitem>
        </varlistentry>
      </variablelist>
      <table frame="box" label="CID 84" rules="all" xml:id="table_CID_84">
        <caption>PET Units</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_97268f5e-8292-4bfb-82d9-a1b1706fbaac">
                <emphasis role="bold">Coding Scheme Designator</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0b987147-1a61-460a-ae8c-8ae6344de823">
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1657b646-4e16-44bc-a40a-6ba15d283775">
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="3" rowspan="1">
              <para xml:id="para_b22ef7d3-e1f7-454d-ab82-dd08aefb5e7f">
                <emphasis role="italic">Include <xref linkend="sect_CID_85" xrefstyle="select: labelnumber quotedtitle nopage"/>
                </emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_0b06214c-36c6-41c1-b11b-4c7ee7de0002">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_217e5d63-53e8-42c4-b275-1a9acac890c5">{counts}</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9ea1163-5a61-4a93-b36f-9ff78834f652">Counts</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_0be0367a-971b-4818-82d4-a56d4c627cef">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c99c5a25-ef51-4800-b315-50320672dce2">{counts}/s</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c6d0a85-7aef-44e7-8b84-99ab379fa907">Counts per second</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6643bb83-0b2c-4768-b000-73dffb5aaa9b">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_eebff1cf-81fe-46e1-95a8-b96f21d52cea">{propcounts}</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1105f6d-042a-4315-aa0a-ba0f6d81cce7">Proportional to counts</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_d0256b98-7443-4045-bd71-6efe144a09a1">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_cf3df64f-cbfa-4ce6-b1d3-3043664da53b">{propcounts}/s</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54a5796d-4691-4f90-9157-ce545f520700">Proportional to counts per second</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c483c81f-1ddf-44e7-9d43-0a6c5811c2d1">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_8b0b0c61-c6e3-46fa-a89b-86b5dfa62847">cm2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ff23dc6-d4b2-4141-8384-ecf95325f486">Centimeter**2</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a72c4cf4-5be4-469e-bf5f-34ddee025008">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6cdab932-83ef-4c12-adf7-54a3e117ddcf">%</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3bfe2fdb-c986-4920-85de-3f3bec4da4f4">Percent</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_530a306d-3a9c-461b-99b0-79f988aed2ae">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_583c7120-2619-4b0a-8e52-22c3f2b1a5d0">Bq/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_051a1b6f-2f42-48f5-8227-51f906c7ca48">Becquerels/milliliter</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_93f2a2cf-65c6-465c-9477-453faf6d2488">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_e3687263-62a5-4d33-bbd1-4f8fba052752">mg/min/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d576d8e-eddc-40f0-a657-311cd0f757a4">Milligrams/minute/milliliter</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_a439a427-1b64-4ab4-b50a-64a25f9dbfa9">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_9fbc37b1-2079-47be-8596-3cc7d4644a12">umol/min/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3fd5a00e-d606-44cf-9e3f-dca1feb5e714">Micromole/minute/milliliter</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_bf53b92d-b899-4078-a8a4-597b7e8e3113">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_06f51eb4-4880-4e4a-bdeb-c3a6aa20338e">ml/min/g</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc9aec0e-ab40-4091-bb34-24021a4b177c">Milliliter/minute/gram</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_2e6b5c36-c592-49aa-9da6-a4e91f41d23e">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_b220766b-a09b-4df8-9256-784f13b2c386">ml/g</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62a29e61-e158-468d-9851-92624c076cea">Milliliter/gram</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_34dd648f-be65-43b5-a50d-94705ad6948c">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_3f91d833-bc38-42d4-9efe-c247ab6c4a49">/cm</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd83501a-8e91-4bc5-9307-253fed2ed385">/Centimeter</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c6724ff0-456c-425a-86d1-4cb8cad82a1c">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_0a7167d0-5e5f-42b3-a6b3-ccf15d77e404">umol/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3705ca3e-fab3-4408-a7c2-402ebacfe269">Micromole/milliliter</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section label="CID 85" status="3" xml:id="sect_CID_85">
      <title>SUV Units</title>
      <variablelist spacing="compact">
        <varlistentry>
          <term>
            <emphasis role="bold">Type:</emphasis>
          </term>
          <listitem>
            <para xml:id="para_1233c115-159e-4923-8177-2fdfb7e5fd11">
              <emphasis role="bold">Extensible</emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term>
            <emphasis role="bold">Version:</emphasis>
          </term>
          <listitem>
            <para xml:id="para_798e4c11-1fcc-41c5-8d20-a5ad53ead4d8">
              <emphasis role="bold">20141110</emphasis>
            </para>
          </listitem>
        </varlistentry>
      </variablelist>
      <table frame="box" label="CID 85" rules="all" xml:id="table_CID_85">
        <caption>SUV Units</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_64d64317-1a6a-41a2-9031-58566be0d9e2">
                <emphasis role="bold">Coding Scheme Designator</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8bc22f66-ffc5-4b2b-a04a-88a43bcd6e00">
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2d06eea7-6f2e-4033-82d0-190fe081f18e">
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_7461633a-9d61-4a97-9940-69e4b4ead6a3">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_0dfd9e64-7a63-436d-a0a8-66d978db7743"><emphasis role="bold"><emphasis role="strikethrough">{SUVbw}</emphasis></emphasis>g/ml<emphasis role="bold"><emphasis role="underline">{SUVbw}</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf9c6c61-7a52-4c20-b7a6-306e42f5a88d">Standardized Uptake Value body weight</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6d949794-0fb2-4cd3-95e8-0ead381dd70c">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_4742fa17-0d2f-49b1-9a90-59d067bfff04"><emphasis role="bold"><emphasis role="strikethrough">{SUVlbm}</emphasis></emphasis>g/ml<emphasis role="bold"><emphasis role="underline">{SUVlbm}</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7fd7d41-fca4-4957-b0c5-0bcfb58cfe86">Standardized Uptake Value lean body mass<emphasis role="bold"><emphasis role="underline"> (James)</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">UCUM</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">g/ml{SUVlbm(Janma)}</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Standardized Uptake Value lean body mass (Janma)</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c5ae4cd9-1735-42fd-89ef-6256ed332edb">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_59c9cf00-fe3f-4742-b1dc-397a1bc84e6f"><emphasis role="bold"><emphasis role="strikethrough">{SUVbsa}</emphasis></emphasis>cm2/ml<emphasis role="bold"><emphasis role="underline">{SUVbsa}</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf2e79f6-1993-44e9-8cd4-f3d3ec5e7cef">Standardized Uptake Value body surface area</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_f15fc726-0de8-430d-bafa-a885ff3a6e84">UCUM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_df84794c-3cba-4529-a64f-30383dea2229"><emphasis role="bold"><emphasis role="strikethrough">{SUVibw}</emphasis></emphasis>g/ml<emphasis role="bold"><emphasis role="underline">{SUVibw}</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_056ac97a-80b9-4481-b27c-f1b7ff614161">Standardized Uptake Value ideal body weight</para>
            </td>
          </tr>
        </tbody>
      </table>
      <note>
        <para xml:id="para_100b164d-1936-487d-97a8-c6cb77f8feaa">The formulas for the determination of SUVbw, SUVbsa, SUVlbm<emphasis role="bold"><emphasis role="underline">(James)</emphasis></emphasis> and SUVibw are defined in Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction. Radiology, 1999</emphasis> at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>. <emphasis role="bold"><emphasis role="underline">The Janmahasatian LBM formula is defined in <emphasis role="italic">Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051–65.</emphasis> at <link xl:href="http://dx.doi.org/10.2165/00003088-200544100-00004"/> and its role in SUVlbm(Janma) calculation is discussed in <emphasis role="italic">Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging. Journal of Nuclear Medicine. 2014 Sep
                1;55(9):1481–4. </emphasis> at <link xl:href="http://jnm.snmjournals.org/content/55/9/1481"/>.</emphasis></emphasis></para>
        <para>The patient size correction factors are summarized here, where weight is in kg and height is in cm:</para>
        <para xml:id="para_211f1841-764d-481b-9165-4ecdd3a55a15">SUVbw: males &amp; females: weight</para>
        <para xml:id="para_f990e628-8377-4a92-956a-651e70f6467f">SUVlbm<emphasis role="bold"><emphasis role="underline">(James)</emphasis></emphasis>: males: 1.10 * weight - 120 * (weight/height) ^2</para>
        <para xml:id="para_3571f8d2-e7b4-432b-9d54-bcb2d7d04342">females: 1.07 * weight - 148 * (weight/height) ^2</para>
        <para><emphasis role="bold"><emphasis role="underline">SUVlbm(Janma): males: 9.27E3 * weight / (6.68E3 + 216 * weight / (height^2))</emphasis></emphasis></para>
        <para><emphasis role="bold"><emphasis role="underline">females: 9.27E3 * weight / (8.78E3 + 244 * weight / (height^2))</emphasis></emphasis></para>
        <para xml:id="para_7454a1df-dcc9-4492-b521-ddd007870a34">SUVbsa: males &amp; females: weight^ 0.425 * height^0.725 * 0.007184</para>
        <para xml:id="para_0418a618-881c-4ce1-8f86-1673ade9ebb7">SUVibw: males: 48.0 + 1.06 * (height - 152)</para>
        <para xml:id="para_33526356-6ca9-424a-a9bc-9095a7367e65">females: 45.5 + 0.91 * (height - 152)</para>
      </note>
    </section>
    
      <section label="CID 7180" status="3" xml:id="sect_CID_7180">
         <title>Abstract Multi-dimensional Image Model Component Semantics</title>
         <variablelist spacing="compact">
            <varlistentry>
               <term>
                  <emphasis role="bold">Type:</emphasis>
               </term>
               <listitem>
                  <para xml:id="para_ff3b7038-d405-4c7f-ae17-7191d2826e30">
                     <emphasis role="bold">Extensible</emphasis>
                  </para>
               </listitem>
            </varlistentry>
            <varlistentry>
               <term>
                  <emphasis role="bold">Version:</emphasis>
               </term>
               <listitem>
                  <para xml:id="para_28f1b78a-d2a9-4dc6-aa5e-6bfbdcff75c0">
                     <emphasis role="bold">20141110</emphasis>
                  </para>
               </listitem>
            </varlistentry>
         </variablelist>
         <table frame="box" label="CID 7180" rules="all" xml:id="table_CID_7180">
            <caption>Abstract Multi-dimensional Image Model Component Semantics</caption>
            <thead>
               <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_2eab0d49-4604-4a49-bf7d-f264812fd5ff">
                        <emphasis role="bold">Coding Scheme Designator</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_45981ceb-ccd2-49db-b35c-dad487b30b46">
                        <emphasis role="bold">Code Value</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_e809725f-b073-41d7-b621-163707cc949a">
                        <emphasis role="bold">Code Meaning</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_8e3ec141-b4cd-416e-bffb-4572bfe13597">
                        <emphasis role="bold">SNOMED-CT Concept ID</emphasis>
                     </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                     <para xml:id="para_cc1d9649-7dae-4808-9d4c-725a13405592">
                        <emphasis role="bold">UMLS Concept Unique ID</emphasis>
                     </para>
                  </th>
               </tr>
            </thead>
            <tbody>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_3c5a2a58-501e-451a-bd09-035ead7b0abd">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_034f7adc-9723-4779-9199-aa5139c36aa5"> ...</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_d9b6370a-64e9-4869-93f6-93fef94a52cd">...</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_9516a1c2-346c-4476-889d-82fc5ce7ced2"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_28a48501-9cb2-4d23-a8ec-f16cb160260c"/>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_b8dea7f1-0449-4ab7-94e5-66f3b9f93711">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_74c43ea3-9782-4302-8991-631654c550d4">
                        <link linkend="DCM_126400" xrefstyle="select: nopage">126400</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a02ed2bf-f7e6-4d9d-aff8-147d55a253ba">Standardized Uptake Value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_7bd9ca2c-5ec2-4e88-af5c-6ee6d046cfdb"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_76c6c235-bbd0-4544-88ef-3b283f97d568"/>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_e0df5a00-f41b-4ea5-ac8e-c7c9aeb90fdf">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_fb3e139e-3e47-4e01-83c1-86001d27e471">
                        <link linkend="DCM_126401" xrefstyle="select: nopage">126401</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_003b8be2-d677-49cf-8154-c6c942d2dfa9">SUVbw</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_92f9b047-c173-4c6f-a87b-727be40cdab1"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_e60a6ca7-ab71-48e2-8ad5-c7d0a73692a5"/>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_435ebcf3-1a36-497d-a80e-db65a4513377">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_3d1d0cd3-6d16-4065-ad3f-550b366bf1b9">
                        <link linkend="DCM_126402" xrefstyle="select: nopage">126402</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6364c99b-6145-4061-b28f-90ffd14fdba5">SUVlbm</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_138384e7-53a3-4090-bd76-60a6da82b6c1"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_1adece85-9dcf-4604-b5c1-b71934b5ef73"/>
                  </td>
               </tr>
          <tr>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">
                    <link linkend="DCM_126405" xrefstyle="select: nopage">126405</link>
                  </emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline">SUVlbm(Janma)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
            </td>
            <td align="center" colspan="1">
              <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
            </td>
          </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_4bd42f16-aa32-4718-980e-3b2f27813b2f">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_34d397be-8ce8-4cef-85ab-fba866445cde">
                        <link linkend="DCM_126403" xrefstyle="select: nopage">126403</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_f0cda444-15ae-4c68-8c5e-8f85026a38e1">SUVbsa</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_ae8ac368-04a5-415d-af19-3ea4359fefbf"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_61a60f7b-f4c0-42e2-ba43-e9bc817f1b35"/>
                  </td>
               </tr>
               <tr valign="top">
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_2da213d2-386c-495b-9073-b6800ee7ab6a">DCM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_35cd9883-6f23-470a-96d7-82c174fa6482">
                        <link linkend="DCM_126404" xrefstyle="select: nopage">126404</link>
                     </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_fbcc5948-cbf1-45fe-86c7-d883ac332824">SUVibw</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_69a96770-a341-4c46-acd1-5f4b89b83a0c"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_2ed14e3d-3752-4afb-8f8f-f6b53fd883dd"/>
                  </td>
               </tr>
            </tbody>
         </table>
      </section>

    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td><para><emphasis role="italic">Amend DICOM PS3.16 concept definitions:</emphasis></para></td>
        </tr>
      </informaltable>
    </section>
    
    <section>
      <title/>
      <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
      <caption>DICOM Controlled Terminology Definitions</caption>
      <thead>
        <tr valign="top">
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">
              <emphasis role="bold">Code Value</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">
              <emphasis role="bold">Code Meaning</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">
              <emphasis role="bold">Definition</emphasis>
            </para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">
              <emphasis role="bold">Notes</emphasis>
            </para>
          </th>
        </tr>
      </thead>
      <tbody>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126400">126400</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8e52c585-0861-4885-a636-9ef11a825850">Standardized Uptake Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_06a6e6f0-bfdb-41d2-a82d-5b3d271dd2d1">A ratio of locally measured radioactivity concentration versus the injected radioactivity distibuted evenly throughout the whole body.</para>
            <para xml:id="para_44f438df-abfd-471b-a927-a8eeae406e39">This general concept encompasses all specific methods of calculating the whole body volume of distribution, such as using body weight, lean body mass, body surface area, etc.</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_812dcbe6-fdf4-4938-b593-0ab9a23a68b1"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126401">126401</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dad17923-e87c-492a-9f82-1669be5de388">SUVbw</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d3757af-bda2-48ab-a0a1-cc5d5d343285">Standardized Uptake Value calculated using body weight. The patient size
              correction factor for males and females is body weight.</para>
            <para xml:id="para_311680c9-aaf7-4973-b246-38c93d717702">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e07cddb1-0f04-49ac-8ce7-28e5663c89b7"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126402">126402</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c2e3e95-49e9-48a1-a260-e335bb835e86">SUVlbm</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d02b3c10-d16f-443f-9a53-fbe81c4be75d">Standardized Uptake Value calculated using lean body mass<emphasis role="bold"><emphasis role="underline"> by James method</emphasis></emphasis>. The patient size correction factor for males is 1.10 * weight - 120 * (weight/height) ^2, and for females is 1.07 * weight - 148 * (weight/height) ^2.</para>
            <para xml:id="para_9df819b6-d333-4b0a-91eb-cbe77579617d">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cea7ee94-c1bb-4c81-9588-3b9ff43dbbce"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126403">126403</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_23e408bd-0a55-4006-9387-fca4f7035c99">SUVbsa</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e810ad9a-6794-4bb7-941a-5a4cf5ef1583">Standardized Uptake Value calculated using body surface area. The patient size
              correction factor for males and females is weight^ 0.425 * height^0.725 *
              0.007184.</para>
            <para xml:id="para_c786d4ad-0027-4061-8e9b-50171f01ff2f">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f2ce7037-08f0-454e-bec6-2ac2298e12f8"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126404">126404</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e26244f7-a388-4a4b-8b5a-6a82174f3c41">SUVibw</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c5b0bb1-81d1-4ca2-82ba-c43d61f8dc6a">Standardized Uptake Value calculated using ideal body weight. The patient size
              correction factor for males is 48.0 + 1.06 * (height - 152) and for females is 45.5
              + 0.91 * (height - 152).</para>
            <para xml:id="para_7309f0ac-b024-4482-aeaf-3b13f67fb9ac">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_43a7263a-388c-43e6-a345-11d10e2da8fb"/>
          </td>
        </tr>
          <tr>
            <td align="center" colspan="1"><para xml:id="DCM_126405"><emphasis role="bold"><emphasis role="underline">126405</emphasis></emphasis></para></td>
            <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">SUVlbm(Janma)</emphasis></emphasis></para></td>
            <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">Standardized Uptake Value calculated using lean body mass by Janmahasatian method. The patient size correction factor for males is 9.27E3 * weight / (6.68E3 + 216 * weight / (height^2)) and for females is 9.27E3 * weight / (8.78E3 + 244 * weight / (height^2)).</emphasis></emphasis></para>
              <para><emphasis role="bold"><emphasis role="underline">Defined in <emphasis role="italic">Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051–65.</emphasis> at <link xl:href="http://dx.doi.org/10.2165/00003088-200544100-00004"/> and its role in SUVlbm(Janma) calculation is discussed in <emphasis role="italic">Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging. Journal of Nuclear Medicine. 2014 Sep 1;55(9):1481–4. </emphasis> at <link xl:href="http://jnm.snmjournals.org/content/55/9/1481"/>.</emphasis></emphasis></para></td>
            <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline"/></emphasis></para></td>
          </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126410">126410</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d597971f-217e-4b91-988a-bcf1c23c18bf">SUV body weight calculation method</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_270653ce-766a-476a-b683-74d2b20245a9">Method of calculating Standardized Uptake Value using body weight. The patient
              size correction factor for males and females is body weight.</para>
            <para xml:id="para_5a00de2c-9c86-4b9b-aa0a-19ff0d09d0b0">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_afb39c28-e029-4fee-afd9-0ea50d34dad5"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126411">126411</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_115b009c-9a00-4d1b-af00-2724e1778217">SUV lean body mass calculation method</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f4eaf243-c419-4a5c-a16f-d12f274b9fcc"><emphasis role="bold"><emphasis role="underline">James m</emphasis><emphasis role="strikethrough">M</emphasis></emphasis>ethod of calculating Standardized Uptake Value using lean body mass. The patient size correction factor for males is 1.10 * weight - 120 * (weight/height) ^2, and for females is 1.07 * weight - 148 * (weight/height) ^2.</para>
            <para xml:id="para_68853096-7935-4e69-af20-506c5b778d33">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a6df8ef-87c8-4c1b-aa6e-5edb88eb8636"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126412">126412</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5fb9194d-4319-4447-97fa-7a7b84911da2">SUV body surface area calculation method</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d1ddd69b-d42e-4373-a12c-c43807f70301">Method of calculating Standardized Uptake Value using body surface area. The
              patient size correction factor for males and females is weight^ 0.425 * height^0.725
              * 0.007184.</para>
            <para xml:id="para_711f75ec-1979-451a-9d5c-5215a0356aca">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1b8d6106-aff0-4e1f-8457-60186a3e5059"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="center" colspan="1" rowspan="1">
            <para xml:id="DCM_126413">126413</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d22fee15-5bad-4f11-89f8-fba9dd32c905">SUV ideal body weight calculation method</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ad2311fa-6009-4b47-896f-9afd76c02bde">Method of calculating Standardized Uptake Value using ideal body weight. The
              patient size correction factor for males is 48.0 + 1.06 * (height - 152) and for
              females is 45.5 + 0.91 * (height - 152).</para>
            <para xml:id="para_1b747178-1be8-4bfa-ba20-04938d76dcbf">Defined in
              Sugawara et al. <emphasis role="italic">Reevaluation of the Standardized Uptake Value for
                FDG: Variations with Body Weight and Methods for Correction.</emphasis>Radiology, 1999
              at <link xl:href="http://radiology.rsna.org/content/213/2/521"/>
            </para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b3a3f75e-158f-4f83-a064-495c6ed180c5"/>
          </td>
        </tr>
        <tr>
          <td align="center" colspan="1"><para xml:id="DCM_126414"><emphasis role="bold"><emphasis role="underline">126414</emphasis></emphasis></para></td>
          <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">SUV lean body mass calculation Janmahasatian method</emphasis></emphasis></para></td>
          <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">Janmahasatian method of calculating Standardized Uptake Value using lean body mass. The patient size correction factor for males is 9.27E3 * weight / (6.68E3 + 216 * weight / (height^2)) and for females is 9.27E3 * weight / (8.78E3 + 244 * weight / (height^2)).</emphasis></emphasis></para>
            <para><emphasis role="bold"><emphasis role="underline">Defined in <emphasis role="italic">Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051–65.</emphasis> at <link xl:href="http://dx.doi.org/10.2165/00003088-200544100-00004"/> and its role in SUVlbm(Janma) calculation is discussed in <emphasis role="italic">Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging. Journal of Nuclear Medicine. 2014 Sep 1;55(9):1481–4. </emphasis> at <link xl:href="http://jnm.snmjournals.org/content/55/9/1481"/>.</emphasis></emphasis></para></td>
          <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline"/></emphasis></para></td>
        </tr>
        
      </tbody>
    </table>
    </section>
    
  </chapter>
</book>