<book xmlns="http://docbook.org/ns/docbook" xmlns:xl="http://www.w3.org/1999/xlink" version="5.0">
  <title>CP-1523</title>
  <subtitle>Requirements for items in Quantity Definition Sequence</subtitle>
  <chapter>
    <title></title>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Status</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>Assigned</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>2015/09/17</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Person Assigned</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>David Clunie</para><para><link
          xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:dclunie@dclunie.com"/></para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Submitter Name</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>Harry Solomon</para><para><link
          xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="mailto:Harry.Solomon@med.ge.com"/></para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Submission Date</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>2015/08/04</para></td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Correction Number CP-1523</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Log Summary: Requirements for items in Quantity Definition Sequence</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.3</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Rationale for Correction:</para>
          <para>RWV examples in CP 1448 (and in PS3.17 Annex QQQ from Sup 172) highlight the issue of the meaning of multiple Content Items in the Quantity Definition Sequence (0040,9220). Additional text should be added to PS3.3 Section C.7.6.16.2.11 Table C.7.6.16-12.</para>
          <para>One of the items should be required to specify the quantified characteristic, and whether or not the order of the items is signicant should be specified (e.g., whether the quantified characteristic is required to be first or not).</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title></title>
    <section>
      <title></title>
      <informaltable rules="all" frame="box">
        <tr><td><para><emphasis role="italic">Amend DICOM PS3.3 as follows:</emphasis></para></td></tr>
      </informaltable>
    </section>
    
                <section label="C.7.6.16.2.11" status="6" xml:id="sect_C.7.6.16.2.11">
              <title>Real World Value Mapping Macro</title>
              <para>...</para>
              <table frame="box" label="C.7.6.16-12b" rules="all" xml:id="table_C.7.6.16-12b">
                <caption>Real World Value Mapping Item Macro Attributes</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_6e8e4f6c-2271-4a9a-974d-8b42545b1b34">Attribute Name</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_1f2022d8-22bb-4217-a733-78b3e68577a9">Tag</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_72a0e165-d196-4752-a3cf-57f98ff1b0c5">Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_052368dc-5b23-4e0c-a131-21746d308b90">Attribute Description</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_dbbe3f18-5452-4826-aebc-5604798969c8">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_77e79797-3365-41ba-b338-616da45eb62d">...</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_362c5559-002e-4f9a-b8cb-82760593e394">...</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_68de3918-f0d3-414d-b8f9-d94c91a041d1">...</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_175d0c33-fd9b-4a9b-8db1-fec20d966ce6">Measurement Units Code Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_fc99a81b-b4cd-4a18-bbbb-e1875d6d498b">(0040,08EA)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_73e90002-ca29-4154-baff-9de725f62dad">1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_87fc085c-854d-4f5e-af27-af9e002e0346">Units of measurement.</para>
                      <para xml:id="para_9bfc2e91-11b3-41b8-a4ae-3941d4971013">Only a single Item shall be included in this Sequence.</para>
                      <para xml:id="para_4db097a1-9162-4e58-ba30-bb1d621feeaa">See <xref linkend="sect_C.7.6.16.2.11.1" xrefstyle="select: label"/> for further explanation.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_7ae5239b-caa8-4674-94c2-b31c5731d80a">
                <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3c76703c-7902-4902-8d53-40d52c8d6806">Defined CID 7181 "Abstract Multi-dimensional Image Model Component Units"<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7181" xrefstyle="select: labelnumber quotedtitle"/>-->, or as specified in the macro invocation.</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ac548701-56ea-4ab5-868c-547401a077d6">Quantity Definition Sequence</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_d3f5a438-038a-4ccc-90dc-4701389019f8">(0040,9220)</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_87752427-78d9-4443-ba06-5dbb79714895">3</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_c724a52c-9812-4c33-a323-accfcea2c487">A list of name-value pairs that describe the characteristics of the quantity represented by the Real World Value.</para>
                      <para xml:id="para_6f00067e-dcc1-42d0-af97-6d0ec63d5f22">One or more Items are permitted in this Sequence.</para>
                      <para><emphasis role="bold"><emphasis role="underline"> One of the items shall have a concept name that specifies the quantified characteristic, though it is not required that (G-C1C6, SRT, "Quantity") be used if there is a reason to use a similar concept from a different coding scheme. Other items may be concept modifiers, such as (G-C036, SRT, "Measurement Method"). The order of the items is not significant.</emphasis></emphasis></para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="3" rowspan="1">
                      <para xml:id="para_b0033d1d-91a9-49fd-b58c-5160074d61d1">
                <emphasis role="italic">&gt;Include Table 10-2 "Content Item Macro Attributes Description"<!--<xref linkend="table_10-2" xrefstyle="select: label quotedtitle"/>-->
                </emphasis>
              </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e52e0aa4-3f62-42c1-9883-dae8f8a9bdea">Baseline CID for Concept Name Code Sequence is CID 9000 "Physical Quantity Descriptors"<!--<olink targetdoc="PS3.16" targetptr="sect_CID_9000" xrefstyle="select: labelnumber quotedtitle"/>-->.</para>
                      <para xml:id="para_96e462b9-f313-4e6e-91c5-350294f557c5">Baseline CID for Concept Code Sequence for Concept Name of (G-C1C6, SRT, "Quantity") is CID 7180 "Abstract Multi-dimensional Image Model Component Semantics"<!--<olink targetdoc="PS3.16" targetptr="sect_CID_7180" xrefstyle="select: labelnumber quotedtitle"/>-->.</para>
                    </td>
                  </tr>
                </tbody>
              </table>
              <section label="C.7.6.16.2.11.1" status="7" xml:id="sect_C.7.6.16.2.11.1">
                <title>Real World Value Representation</title>
                <para>...</para>
                <section label="C.7.6.16.2.11.1.2" status="8" xml:id="sect_C.7.6.16.2.11.1.2">
                  <title>Real World Values Mapping Sequence Attributes</title>
          <para>...</para>
                  <para xml:id="para_fa83a720-b820-4f54-9680-9b2817f949da">The physical units for the real world values obtained from the sequence item are given by the Measurement Units Code Sequence (0040,08EA).</para>
                  <para xml:id="para_992160a0-1ae9-47d2-a38f-1bb52ea3061b">The quantity that the real world values represent may be described by the Quantity Definition Sequence (0040,9220), which consists of a list of name-value pairs, in which the coded concept name specifies what aspect of the physical quantity is being described.</para>
                  <note>
                    <para xml:id="para_d9140333-5a96-46df-95e8-652cef8c6bed">For example, Cerebral Blood Flow (CBF) may be described by units and quantity as follows:
                      <itemizedlist>
                        <listitem>
                          <para xml:id="para_571e5363-504e-4071-8642-da8a42142bee">Measurement Units Code Sequence (0040,08EA) = (ml/[100]g/min, UCUM, "milliliter per 100 gram per minute")</para>
                        </listitem>
                        <listitem>
                          <para xml:id="para_7ed77289-982e-42e3-8c89-5b15c7c3536f">Quantity Definition Sequence (0040,9220):</para>
                          <itemizedlist>
                            <listitem>
<para xml:id="para_05f15416-61a1-4242-a25b-bdbde812503a">(G-C1C6, SRT, "Quantity") = <olink targetdoc="PS3.16" targetptr="DCM_113055">(113055, DCM, "Regional Cerebral Blood Flow")</olink>
</para>
</listitem>
                          </itemizedlist>
                        </listitem>
                      </itemizedlist>
                    </para>
                  </note>
                  <para xml:id="para_3da55e74-08a1-484e-b9e7-bae0816f7863">The Quantity Definition Sequence (0040,9220) describes only the stored pixel values that are mapped using the Real
                    World Values Mapping, and does not describe derived values from multiple pixels to
                    which the Real World Values Mapping applies.</para>
                  <note>
                    <para xml:id="para_97e02896-96d3-4c6f-ae54-254ebec43aa0">I.e., the mapping is a "point" operation, and
                    as a consequence various modifiers that might be applied to a group of pixels,
                    such as in an ROI, should not be used. E.g., an ROI encoded in a Structured Report
                    using TID 1419 "ROI Measurements" might be the mean or maximum value (e.g., SUVbw
                    mean or SUVbw max), and be encoded with <olink targetdoc="PS3.16" targetptr="DCM_121401">(121401, DCM, "Derivation")</olink> = (R-00317,
                    SRT, "Mean") or (G-A437, SRT, "Maximum"), respectively. These would not be
                    appropriate to use within Quantity Definition Sequence (0040,9220), unless the
                    individual pixel values were themselves derived in such a manner, e.g., when
                    multiple images are averaged together. Thus the content items used in an SR to
                    describe an ROI might be a superset of the name-value pairs used in Quantity
                    Definition Sequence (0040,9220).</para>
                  </note>
                </section>
              </section>
            </section>

  </chapter>
</book>