  <title>CP-1475</title>
  <subtitle>Add patient defined term for Ultrasound Acquisition Geometry</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Status</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Letter Ballot</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Date of Last Update</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/06/19</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/02/27</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number CP-1475</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Add patient defined term for Ultrasound Acquisition Geometry</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.3</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
          <para>The Enhanced US Volume object added an Ultrasound Frame Of Reference Module with a descriptive term describing the characteristic of the ultrasound acquisition geometry.</para>
          <para>It is useful to be able to identify when the geometry is patient relative, to signal that the geometry is the same as the patient-relative coordinate system.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>


    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>

        <section label="C.8.24.2" status="4" xml:id="sect_C.8.24.2">
          <title>Ultrasound Frame of Reference Module</title>
          <para><xref linkend="table_C.8.24.2-1" xrefstyle="select: label"/> specifies the attributes of the Ultrasound Frame Of Reference Module. See <xref linkend="sect_C.8.24.2.1" xrefstyle="select: label"/>.</para>
          <table frame="box" label="C.8.24.2-1" rules="all" xml:id="table_C.8.24.2-1">
            <caption>Ultrasound Frame of Reference Module Attributes</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para>
                    <emphasis role="bold">Attribute Name</emphasis>
                  </para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>
                    <emphasis role="bold">Tag</emphasis>
                  </para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>
                    <emphasis role="bold">Type</emphasis>
                  </para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para>
                    <emphasis role="bold">Attribute Description</emphasis>
                  </para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Volume Frame of Reference UID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,9312)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Uniquely identifies this Volume Frame of Reference.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Ultrasound Acquisition Geometry</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,9307)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Characteristic of the ultrasound acquisition geometry.</para>
                  <variablelist spacing="compact">
                    <title>Defined Terms:</title>
                    <varlistentry>
                      <term>APEX</term>
                      <listitem>
                        <para>there exists an apex of the scan lines from which the volume data was acquired.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term><emphasis role="bold"><emphasis role="underline">PATIENT</emphasis></emphasis></term>
                      <listitem><emphasis role="bold"><emphasis role="underline">the ultrasound acquisition geometry is patient relative</emphasis></emphasis></listitem>
                    </varlistentry>
                  </variablelist>
                  <note><para><emphasis role="bold"><emphasis role="underline">When the value is PATIENT, the Plane Position (Patient) and Plane Orientation (Patient) functional group macros will be present, and will contain attributes with identical values to the corresponding attributes in the Plane Position (Volume) and Plane Orientation (Volume) functional group macros.</emphasis></emphasis></para></note>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Apex Position</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,9308)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Position of the apex (or phase center) of the acquisition geometry, encoded as x<subscript>A</subscript>, y<subscript>A</subscript>, and z<subscript>A</subscript> in mm units in the Volume Frame of Reference. The apex (x<subscript>A</subscript>, y<subscript>A</subscript>, z<subscript>A</subscript>) may be located in the volume or exterior to it.</para>
                  <para>Required if value of Ultrasound Acquisition Geometry (0020,9307) is APEX.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Volume to Transducer Relationship</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,930B)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Relationship between the transducer and the acquired volume.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>FIXED</term>
                      <listitem>
                        <para>The transducer position and orientation relative to the volume is constant and specified by Volume to Transducer Mapping Matrix (0020,9309).</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>POSITION_VAR</term>
                      <listitem>
                        <para>The transducer position relative to the volume varies during acquisition and the position specified by Volume to Transducer Mapping Matrix (0020,9309) is a nominal value. The transducer orientation relative to the volume is constant and specified by Volume to Transducer Mapping Matrix (0020,9309).</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>ORIENTATION_VAR</term>
                      <listitem>
                        <para>The transducer orientation relative to the volume varies during acquisition and the orientation specified by Volume to Transducer Mapping Matrix (0020,9309) is a nominal value. The transducer position relative to the volume is constant and specified by Volume to Transducer Mapping Matrix (0020,9309).</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>VARIABLE</term>
                      <listitem>
                        <para>The transducer position and orientation relative to the volume varies during acquisition and the position and orientation specified by Volume to Transducer Mapping Matrix (0020,9309) are nominal.</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <para>Required if the transducer position and/or orientation relative to the volume is not constant. May be present otherwise.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Volume to Transducer Mapping Matrix</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,9309)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>A 4x4 rigid transformation matrix that maps the Volume Frame of Reference homogeneous coordinate system (X<subscript>V</subscript>, Y<subscript>V</subscript>, Z<subscript>V</subscript>) to the Transducer Frame of Reference homogeneous coordinate system (X<subscript>X</subscript>,Y<subscript>X</subscript>, Z<subscript>X</subscript>). Matrix elements shall be listed in row-major order. See <xref linkend="sect_C.8.24.2.1" xrefstyle="select: label"/> for details.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Patient Frame of Reference Source</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,930C)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Indicates how the supplied Image Position (Patient) (0020,0032) and Image Orientation (Patient) (0020,0037) values are obtained.</para>
                  <variablelist spacing="compact">
                    <title>Enumerated Values:</title>
                    <varlistentry>
                      <term>TABLE</term>
                      <listitem>
                        <para>A positioning device, such as a gantry, was used to generate these values.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>ESTIMATED</term>
                      <listitem>
                        <para>Estimated patient position / orientation (e.g., estimated by the user), or if reliable information is not available.</para>
                      </listitem>
                    </varlistentry>
                    <varlistentry>
                      <term>REGISTRATION</term>
                      <listitem>
                        <para>Acquisition has been spatially registered to a prior image set.</para>
                      </listitem>
                    </varlistentry>
                  </variablelist>
                  <para>Required if either Image Position (Patient) (0020,0032) or Image Orientation (Patient) (0020,0037) is present.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Table Frame of Reference UID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,9313)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>Uniquely identifies this Table Frame of Reference.</para>
                  <para>Required if Patient Frame of Reference Source (0020,930C) is TABLE.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para>Volume to Table Mapping Matrix</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>(0020,930A)</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para>1C</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para>A 4x4 rigid transformation matrix that maps the Volume Frame of Reference homogeneous coordinate system (X<subscript>V</subscript>,Y<subscript>V</subscript>, Z<subscript>V</subscript>) to the Table Frame of Reference homogeneous coordinate system (X<subscript>T</subscript>,Y<subscript>T</subscript>, Z<subscript>T</subscript>). Matrix elements shall be listed in row-major order. See <xref linkend="sect_C.8.24.2.2" xrefstyle="select: label"/> for details.</para>
                  <para>Required if Patient Frame of Reference Source (0020,930C) is TABLE.</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
    
  </chapter>