  <title>CP-1469</title>
  <subtitle>Remove duplicate rows referencing source image or series for segmentation in volumetric ROI</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Status</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Sep 2015 Votinng Packet</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Date of Last Update</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/06/19</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/02/11</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number CP-1469</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Remove duplicate rows referencing source image or series for segmentation in volumetric ROI</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.16</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>CP 1112 introduced volumetric ROIs that are defined by reference to either images with 2D
            coordinates, rasterized segmentations or surfaces defined by 3D coordinates. In early
            drafts, references to source images or series for segmentation were nested within parent
            content items for the different types, but when these were raised up one level of
            nesting to prevent forbidden content item relationships, it was not noticed that they
            were duplicated.</para>
          <para>The duplicated references can be removed, since the different types of ROI are
            mutually exclusive.</para>
          </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>


    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>
    
          <section label="TID 1411" status="3" xml:id="sect_TID_1411">
        <title>Volumetric ROI Measurements</title>
        <para>This Template provides a general structure to report one or more measurements for some metric, e.g., density, flow, or concentration, over a volumetric region of interest in a set of images or a Frame of Reference. The volumetric ROI may be specified by a set of SCOORDs on an image set representing a volume, by a volumetric Segmentation Image, by a volume defined in a Surface Segmentation, or by a SCOORD3D.</para>
        <table rules="all" frame="box" label="TID 1411" xml:id="table_TID_1411_Parameters">
          <caption>Parameters</caption>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Coded term or Context Group for Concept Name of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Units</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Units for the measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$ModType</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Modifier Name for Concept Name of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$ModValue</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Modifier Value for Concept Name of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Method</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Value for Measurement Method</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Derivation</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Value for Measurement Derivation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSite</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Value for Anatomic Location of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSiteMod</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Modifier Value for Anatomic Location of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Equation</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Coded term or Context Group for the equation or table from which the measurement was derived or computed</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$RefAuthority</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Bibliographic reference or authority for statistical properties of a reference population</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$RangeAuthority</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Bibliographic reference or authority for the normal range of the measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameter</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Coded term or Context Group for Concept Name of a derivation parameter</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameterUnits</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Units of derivation parameter</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$QualitativeEvaluations</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Evaluations encoded with code or text responses</para>
              </td>
            </tr>
          </tbody>
        </table>
        <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Order:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Significant</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="TID 1411" xml:id="table_TID_1411">
          <caption>Volumetric ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_125007" xrefstyle="select: nopage">(125007, DCM, "Measurement Group")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DT <link linkend="DCM_112039" xrefstyle="select: nopage">(112039, DCM, "Tracking Identifier")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UIDREF</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_112040" xrefstyle="select: nopage">(112040, DCM, "Tracking Unique Identifier")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_1502" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>SCOORD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_111030" xrefstyle="select: nopage">(111030, DCM, "Image Region")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>MC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Rows 7, 10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>GRAPHIC TYPE = not {MULTIPOINT}</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>SELECTED FROM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IMAGE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IMAGE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121191" xrefstyle="select: nopage">(121191, DCM, "Referenced Segment")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>MC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Rows 5, 10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Reference shall be to a Segmentation Image or Surface Segmentation object, with a single value specified in Referenced Segment Number</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">8</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">IMAGE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">EV <link linkend="DCM_121233" xrefstyle="select: nopage">(121233, DCM, "Source image for segmentation")</link></emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">1-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">MC</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">XOR Row 9 and IFF Row 7</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">9</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">UIDREF</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">EV <link linkend="DCM_121232" xrefstyle="select: nopage">(121232, DCM, "Source series for image segmentation")</link></emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">MC</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">XOR Row 8 and IFF Row 7</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>SCOORD3D</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121231" xrefstyle="select: nopage">(121231, DCM, "Volume Surface")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>MC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Rows 5, 7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>GRAPHIC TYPE = {ELLIPSOID}</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>11</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IMAGE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121233" xrefstyle="select: nopage">(121233, DCM, "Source image for segmentation")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>MC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Row 12 and IFF <emphasis role="bold"><emphasis role="underline"
                    >(</emphasis></emphasis>Row 10<emphasis role="bold"><emphasis role="underline">
                    or Row7)</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UIDREF</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121232" xrefstyle="select: nopage">(121232, DCM, "Source series for segmentation")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>MC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Row 11 and IFF <emphasis role="bold"><emphasis role="underline"
                    >(</emphasis></emphasis>Row 10<emphasis role="bold"><emphasis role="underline">
                    or Row7)</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>13</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IMAGE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121200" xrefstyle="select: nopage">(121200, DCM, "Illustration of ROI")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>14</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>COMPOSITE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_126100" xrefstyle="select: nopage">126100</link>, DCM, "Real World Value Map used for measurement")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>SOP Class UID shall be Real World Value Mapping Storage ("1.2.840.10008.5.1.4.1.1.67")</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>15</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = $Measurement</para>
                <para>$Units = $Units</para>
                <para>$ModType = $ModType</para>
                <para>$ModValue = $ModValue</para>
                <para>$Method = $Method</para>
                <para>$Derivation = $Derivation</para>
                <para>$TargetSite = $TargetSite</para>
                <para>$TargetSiteMod = $TargetSiteMod</para>
                <para>$Equation = $Equation</para>
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
                <para>$DerivationParameter = $DerivationParameter</para>
                <para>$DerivationParameterUnits = $DerivationParameterUnits</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>16</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$QualitativeEvaluations</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>17</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$QualitativeEvaluations</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 2, 3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The Tracking Identifier and Tracking Unique Identifier are defined as a text label or unique identifier (respectively) used for tracking a finding or feature, potentially across multiple reporting objects, over time. As such, they are distinct from the Observation UID (0040,A171), which is unique identifier of the specific Content Item and its subsidiary Content Items that constitute an individual observation, and would be different for different observations on different occasions of the same finding or feature.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 6, 7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Referenced Segment Number (0062,000B) is an attribute of the IMAGE Content Item, and shall be present with a single value.</para>
                <para>If the Referenced SOP Instance is a Segmentation Image, it shall have a defined Frame of Reference. If it has Segmentation Type (0062,0001) value BINARY, it identifies the volume of defined (measured) region of interest by voxel values in the referenced segment with value 1. If it has Segmentation Type value FRACTIONAL, the volume is defined by an implementation dependent method.</para>
                <para>If the referenced SOP Instance is a Surface Segmentation, the referenced segment shall constitute a finite volume. It identifies the volume of the defined (measured) region of interest by the interior of the finite volume.</para>
                <para>Segment number shall be specified even if the Segmentation SOP Instance has only a single segment.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row<emphasis role="bold"><emphasis role="strikethrough">s 8,</emphasis></emphasis>
                11</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Identifies the source images that were segmented to identify the ROI, when, for example a subset of images in a series was used.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row<emphasis role="bold"><emphasis role="strikethrough">s 9,</emphasis></emphasis>
                12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Identifies the source series of images that were segmented to identify the ROI, when, for example an entire set of images in a series was used.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 13</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>These referenced images may contain "screen shot" illustrating rendered versions of the ROI.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 14</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The reference to an RWV in
                  Row 14 allows measurements to be made in units that differ from the stored
                  pixel values in the images referenced elsewhere in the template. E.g., for a
                  PET SUVbw measurement, the mapping from activity/concentration units in the
                  referenced image that was used (and which may be reused for measurements in
                  the future) may be encoded in a referenced RWV instance. This reference
                  applies to any measurements in included templates, unless
                  overridden).</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 16, 17</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Allows encoding a flat
                  list of name-value pairs that are coded questions with coded or text answers,
                  for example, to record categorical observations related to the subject of the
                  measurement group.</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      </section>


  </chapter>
</book>