  <title>CP-1463</title>
  <subtitle>Clarify Enhanced US Volume Image and Frame Type Values 3 and 4</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Status</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Assigned</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Date of Last Update</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/03/18</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Michael Onken</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
            xlink:href="mailto:onken@open-connections.de"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/02/06</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number CP-1463</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Clarify Enhanced US Volume Image and Frame Type Values 3 and 4</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.3</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
          <para>Beginning with Sup 49 (enhanced MR), the enhanced family of IODs all use the concept of
            "image flavor" for Value 3 and "derivation" for Value 4 of both Image Type and Frame Type.
            Further, the VM of Frame Type is defined in PS3.6 as fixed at 4.
          The Enhanced US Volume (Sup 43) IOD is the one exception, and uses the terminology "are not used"
          for these values, which raises doubt as to whether empty values should be sent, or no values
          (which have different representations, e.g. "ORIGINAL\PRIMARY\\" versus "ORIGINAL\PRIMARY").</para>
          <para>In practice, since Enhanced US Volume images may be acquired in different
            circumstances that can indeed be described with the common "flavor" defined terms, and
            since they may also be derived, it is proposed to permit the reuse of the same common
            terms as all the rest of the enhanced family of IODs.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>


    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows (changes to existing text
              are bold and <emphasis role="bold"><emphasis role="underline"
                >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                  ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
              removals):</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>

    <section label="C.8.24.3.2" status="5" xml:id="sect_C.8.24.3.2">
      <title>Image Type</title>
      <para><emphasis role="bold"><emphasis role="strikethrough">Image Type (0008,0008) is specified to be Type 1 with the following constraints:</emphasis></emphasis></para>
      <variablelist spacing="compact">
        <title><emphasis role="strikethrough">Enumerated Values for Value 1:</emphasis></title>
        <varlistentry>
          <term><emphasis role="bold"><emphasis role="strikethrough">ORIGINAL</emphasis></emphasis></term>
          <listitem>
            <para/>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term><emphasis role="bold"><emphasis role="strikethrough">DERIVED</emphasis></emphasis></term>
          <listitem>
            <para/>
          </listitem>
        </varlistentry>
      </variablelist>
      <variablelist spacing="compact">
        <title><emphasis role="strikethrough">Enumerated Values for Value 2:</emphasis></title>
        <varlistentry>
          <term><emphasis role="bold"><emphasis role="strikethrough">PRIMARY</emphasis></emphasis></term>
          <listitem>
            <para/>
          </listitem>
        </varlistentry>
      </variablelist>
      <para><emphasis role="bold"><emphasis role="strikethrough">Value 3 (Image Flavor) and Value 4 are not used</emphasis></emphasis></para>
      <section label="C.8.24.3.2.1" status="6" xml:id="sect_C.8.24.3.2.1">
        <title><emphasis role="underline">Pixel Data Characteristics</emphasis></title>
        <para><emphasis role="bold"><emphasis role="underline">Value 1 of Image Type (0008,0008) and
              Frame Type (0008,9007) is discussed in <xref linkend="sect_C.8.16.1.1"
                xrefstyle="select: label"/>. No additional requirements or Enumerated
              Values.</emphasis></emphasis></para>
      </section>
      <section label="C.8.24.3.2.2" status="6" xml:id="sect_C.8.24.3.2.2">
        <title><emphasis role="underline">Patient Examination Characteristics</emphasis></title>
        <para><emphasis role="bold"><emphasis role="underline">Value 2 of Image Type (0008,0008) and
              Frame Type (0008,9007) is discussed in <xref linkend="sect_C.8.16.1.2"
                xrefstyle="select: label"/>. No additional requirements or Enumerated
              Values.</emphasis></emphasis></para>
      </section>
      <section label="C.8.24.3.2.3" status="6" xml:id="sect_C.8.24.3.2.3">
        <title><emphasis role="underline">Image Flavor</emphasis></title>
        <para><emphasis role="bold"><emphasis role="underline">Value 3 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in <xref linkend="sect_C.8.16.1.3" xrefstyle="select: label"/>. No additional requirements or Defined Terms.</emphasis></emphasis></para>
      </section>
      <section label="C.8.24.3.2.4" status="6" xml:id="sect_C.8.24.3.2.4">
        <title><emphasis role="underline">Derived Pixel Contrast</emphasis></title>
        <para><emphasis role="bold"><emphasis role="underline">Value 4 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in <xref linkend="sect_C.8.16.1.4" xrefstyle="select: label"/>. No additional requirements or Defined Terms.</emphasis></emphasis></para>
      </section>
    </section>
    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Unchanged DICOM PS3.3 referenced sections:</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>

    <section label="C.8.24.6.1" status="5" xml:id="sect_C.8.24.6.1">
      <title>US Image Description Macro</title>
      <para><xref linkend="table_C.8.24.6.1-1" xrefstyle="select: label"/> specifies the attributes of the US Image Description Functional Group Macro.</para>
      <table frame="box" label="C.8.24.6.1-1" rules="all" xml:id="table_C.8.24.6.1-1">
        <caption>US Image Description Macro Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>US Image Description Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0018,9806)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A sequence that describes a general description of this image or frame.</para>
              <para>Only a single Item shall be included in this sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;Frame Type</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,9007)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Type of Frame. A multi-valued attribute analogous to Image Type (0008,0008).</para>
              <para>Enumerated Values and Defined Terms are the same as those for the four values of Image Type (0008,0008), except that the value MIXED is not allowed. See <xref linkend="sect_C.8.24.3.2" xrefstyle="select: label"/>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;Volumetric Properties</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,9206)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Indication if geometric manipulations are possible with frames in the SOP Instance.</para>
              <para>See <xref linkend="sect_C.8.16.2.1.2" xrefstyle="select: label"/> for a description and Enumerated Values.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;Volume Based Calculation Technique</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,9207)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Method used for volume calculations with frames in the SOP Instance.</para>
              <para>See <xref linkend="sect_C.8.16.2.1.3" xrefstyle="select: label"/> for a description and Defined Terms.</para>
            </td>
          </tr>
        </tbody>
      </table>
      <para>If Volumetric Properties (0008,9206) is VOLUME and Volume Based Calculation Technique (0008,9207) is NONE, all frames in the frame set shall be spaced the same Z-distance from adjacent frames (i.e., spacing between slices is constant).</para>
    </section>
    
    <section label="C.8.16.1.1" status="5" xml:id="sect_C.8.16.1.1">
      <title>Pixel Data Characteristics</title>
      <para>Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) shall use one of the following Enumerated Values from <xref linkend="table_C.8-127" xrefstyle="select: label"/>.</para>
      <para>Value 1 of Image Type (0008,0008) and Value 1 of Frame Type (0008,9007) shall not be zero length.</para>
      <table frame="box" label="C.8-127" rules="all" xml:id="table_C.8-127">
        <caption>Image Type and Frame Type Value 1</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Enumerated Value Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Enumerated Value Description</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>ORIGINAL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>An image or frame is original if its pixel data was directly reconstructed from the original data that is obtained from the sensors of the imaging equipment, Image Type (0008,0008) Value 4 is NONE, and Volume Based Calculation Technique (0008,9207) is NONE.</para>
              <note>
                <orderedlist numeration="arabic">
                  <listitem>
                    <para>For MR, original data is data directly reconstructed from k-space data.</para>
                  </listitem>
                  <listitem>
                    <para>For CT, original frames are those directly reconstructed from projection data.</para>
                  </listitem>
                </orderedlist>
              </note>
              
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>DERIVED</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>An image or frame is derived if its pixel data was calculated from original or other derived pixel data (i.e., it is not original).</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>MIXED</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Used only as a value in Image Type (0008,0008) if frames within the SOP Instance contain different values for Value 1 in their Frame Type (0008,9007).</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    <section label="C.8.16.1.2" status="5" xml:id="sect_C.8.16.1.2">
      <title>Patient Examination Characteristics</title>
      <para>Value 2 for Image Type (0008,0008) and Frame Type (0008,9007) follows the standard definition and shall have the following Enumerated Value from <xref linkend="table_C.8-128" xrefstyle="select: label"/>.</para>
      <para>Value 2 of Image Type (0008,0008) and Value 2 of Frame Type (0008,9007) shall not be zero length.</para>
      <table frame="box" label="C.8-128" rules="all" xml:id="table_C.8-128">
        <caption>Image Type and Frame Type Value 2</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Enumerated Value Name</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Enumerated Value Description</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>PRIMARY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>See <xref linkend="sect_C.7.6.1.1.2" xrefstyle="select: label"/></para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section label="C.8.16.1.3" status="5" xml:id="sect_C.8.16.1.3">
            <title>Image Flavor</title>
            <para>Value 3 is an overall representation of the image type. This value may be a summary of several other attributes or a duplication of one of the other attributes to indicate the most important aspect of this image. Value 3 Image Flavor is to be used with Value 4 Derived Pixel Contrast to indicate the nature of the image set.</para>
            <note>
              <para>For example Value 3 = DIFFUSION together with Value 4 = NONE indicates that the image set was originally collected for DIFFUSION.</para>
            </note>
            <para>If Value 3 = DIFFUSION together with Value 4 = DIFFUSION this indicates that the object contains DIFFUSION weighted post processed images.</para>
            <para>Value 3 of Image Type (0008,0008) shall not be zero length.</para>
            <para>Value 3 of Frame Type (0008,9007) may have the same value as found in Value 3 of Image Type (0008,0008), or may have a different value or may be of zero length.</para>
            <para>The attribute value may not be MIXED as this value needs to be a summary of the primary purpose of the images, whether the frames have the same value or not.</para>
            <para><xref linkend="table_C.8-129" xrefstyle="select: label"/> specifies the Defined Terms for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007) that are common to CT and MR. Additional Defined Terms are defined in the modality-specific Module and Macro definitions.</para>
            <table frame="box" label="C.8-129" rules="all" xml:id="table_C.8-129">
              <caption>Image Type and Frame Type Value 3 Common</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Defined Term Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Defined Term Description</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>ANGIO</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Collected for the purpose of angiography</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>CARDIAC</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Images of the heart</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>CARDIAC_GATED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Cardiac gated images, other than of the heart</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>CARDRESP_GATED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Cardiac and respiratory gated images</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>DYNAMIC</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>An image in which the same anatomical volume is imaged at multiple times in order to capture images of a non-cyclic, time varying event. For example, imaging of the uptake of a tracer or contrast in a specific organ over time.</para>
                    <note>
                      <para>This is different from gating techniques, in which the same anatomical volume is imaged during some portion of a cyclic event, e.g., inspiration or R-R Interval.</para>
                    </note>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>FLUOROSCOPY</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Real-time collection of single slices (e.g., CT or MR Fluoroscopy)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>LOCALIZER</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Collected for the purpose of planning other images.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MOTION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Collected for looking at body motion</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>PERFUSION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Collected for the purposes of perfusion calculations.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>PRE_CONTRAST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Collected before contrast was administered</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>POST_CONTRAST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Collected during or after contrast was administered</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>RESP_GATED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Respiratory gated images</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>REST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Cardiac rest image set</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>STATIC</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>A group of frames at varying spatial locations acquired at the same time.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>STRESS</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Cardiac stress image set</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>VOLUME</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Set of frames that define a regularly sampled volume</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>NON_PARALLEL</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Set of frames that are not parallel</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>PARALLEL</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Set of frames that are parallel but do not constitute a regularly sampled volume</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>WHOLE_BODY</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>A group of frames of the whole body; the frames may be acquired at various times (as distinct from STATIC).</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          <section label="C.8.16.1.4" status="5" xml:id="sect_C.8.16.1.4">
            <title>Derived Pixel Contrast</title>
            <para>Value 4 shall be used to indicate derived pixel contrast - generally, contrast created by combining or processing images with the same geometry. Value 4 shall have a value of NONE when Value 1 is ORIGINAL.</para>
            <note>
              <para>If more than one of the following derived types is applicable, then it is up to the generating application to specify the value that best characterizes the derived image.</para>
            </note>
            <para>Value 4 of Image Type (0008,0008) and Value 4 of Frame Type (0008,9007) shall not be zero length unless the SOP Class UID is "1.2.840.10008.5.1.4.1.1.2.2" or "1.2.840.10008.5.1.4.1.1.4.4" or "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted).</para>
            <para><xref linkend="table_C.8-130" xrefstyle="select: label"/> specifies the Defined Terms for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007) that are common to CT and MR. Additional Defined Terms are defined in the modality-specific Module and Macro definitions.</para>
            <table frame="box" label="C.8-130" rules="all" xml:id="table_C.8-130">
              <caption>Image Type and Frame Type Value 4 Common</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Defined Term Name</emphasis>
                    </para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para>
                      <emphasis role="bold">Defined Term Description</emphasis>
                    </para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>ADDITION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by pixel addition operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>DIVISION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by pixel division operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MASKED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by pixel masking operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MAXIMUM</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by Pixel Maximum operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MEAN</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by pixel mean operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MINIMUM</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by Pixel Minimum operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MTT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Mean Transit Time</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MULTIPLICATION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by pixel multiplication operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>RCBF</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Regional Cerebral Blood Flow (rCBF)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>RCBV</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Regional Cerebral Blood Volume (rCBV)</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>RESAMPLED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Pixels have been spatially re-sampled, e.g., MPR</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>STD_DEVIATION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Standard Deviation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>SUBTRACTION</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Created through Pixel by pixel subtraction operation</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>T_TEST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Student's T-Test</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>TTP</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Time To Peak map</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>Z_SCORE</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Z-Score Map</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>NONE</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Not a calculated image</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para>MIXED</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para>Used only as value in Image Type (0008,0008) if frames within the image SOP Instance contain different values for value 4 in their Frame Type (0008,9007) attribute.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
    
    <section label="C.7.6.1.1.2" status="6" xml:id="sect_C.7.6.1.1.2">
      <title>Image Type</title>
      <para>Image Type (0008,0008) identifies important image identification characteristics. These characteristics are:</para>
      <orderedlist numeration="loweralpha">
        <listitem>
          <para>Pixel Data Characteristics</para>
          <orderedlist numeration="arabic">
            <listitem>
              <para>is the image an ORIGINAL Image; an image whose pixel values are based on original or source data</para>
            </listitem>
            <listitem>
              <para>is the image a DERIVED Image; an image whose pixel values have been derived in some manner from the pixel value of one or more other images</para>
            </listitem>
          </orderedlist>
        </listitem>
        <listitem>
          <para>Patient Examination Characteristics</para>
          <orderedlist numeration="arabic">
            <listitem>
              <para>is the image a PRIMARY Image; an image created as a direct result of the Patient examination</para>
            </listitem>
            <listitem>
              <para>is the image a SECONDARY Image; an image created after the initial Patient examination</para>
            </listitem>
          </orderedlist>
        </listitem>
        <listitem>
          <para>Modality Specific Characteristics</para>
        </listitem>
        <listitem>
          <para>Implementation specific identifiers; other implementation specific identifiers shall be documented in an implementation's conformance statement.</para>
        </listitem>
      </orderedlist>
      <para>The Image Type attribute is multi-valued and shall be provided in the following manner:</para>
      <orderedlist numeration="loweralpha">
        <listitem>
          <para>Value 1 shall identify the Pixel Data Characteristics</para>
          <variablelist spacing="compact">
            <title>Enumerated Values:</title>
            <varlistentry>
              <term>ORIGINAL</term>
              <listitem>
                <para>identifies an Original Image</para>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>DERIVED</term>
              <listitem>
                <para>identifies a Derived Image</para>
              </listitem>
            </varlistentry>
          </variablelist>
        </listitem>
        <listitem>
          <para>Value 2 shall identify the Patient Examination Characteristics</para>
          <variablelist spacing="compact">
            <title>Enumerated Values:</title>
            <varlistentry>
              <term>PRIMARY</term>
              <listitem>
                <para>identifies a Primary Image</para>
              </listitem>
            </varlistentry>
            <varlistentry>
              <term>SECONDARY</term>
              <listitem>
                <para>identifies a Secondary Image</para>
              </listitem>
            </varlistentry>
          </variablelist>
        </listitem>
        <listitem>
          <para>Value 3 shall identify any Image IOD specific specialization (optional)</para>
        </listitem>
        <listitem>
          <para>Other Values that are implementation specific (optional)</para>
        </listitem>
      </orderedlist>
      <para>Any of the optional values (value 3 and beyond) may be sent either with a value or zero-length, independent of other optional values, unless otherwise specified by a specialization of this attribute in an IOD.</para>
      <para>If the pixel data of the derived Image is different from the pixel data of the source images and this difference is expected to affect professional interpretation of the image, the Derived Image shall have a UID different than all the source images.</para>
    </section>
    
    