  <title>CP-1456</title>
  <subtitle>Add Operator Identification Sequence to Series Modules</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Status</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>June 2015 Voting Packet</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Date of Last Update</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2015/02/23</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2014/09/14</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number CP-1456</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Add Operator Identification Sequence to Series Modules</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.3</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para>
          <para>CP 247 add Operator Identification Sequence to General Series but not all other Series Modules, such as RT Series, and some IODs do
            not include General Series. Add it where missing.</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.3 as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>

    <section label="C.8.8.1" status="4" xml:id="sect_C.8.8.1">
      <title>RT Series Module</title>
      <para/>
      <table frame="box" label="C.8-37" rules="all" xml:id="table_C.8-37">
        <caption>RT Series Module Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>Operators' Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,1070)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Name(s) of the operator(s) supporting the Series.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Operator Identification
                    Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                  >(0008,1072)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Identification of the
                    operator(s) supporting the Series. One or more items may be present in this
                    sequence. If more than one Item, the number and order shall correspond to the
                    value of Operators' Name (0008,1070), if present.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
          
    <section label="C.8.8.21" status="4" xml:id="sect_C.8.8.21">
      <title>RT Beams Session Record Module</title>
      <table frame="box" label="C.8-57" rules="all" xml:id="table_C.8-57">
        <caption>RT Beams Session Record Module Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;Operators' Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,1070)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Name of operator who authorized override.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;&gt;Operator Identification
                Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                >(0008,1072)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Identification of the
                operator who authorized override. Only a single Item is permitted in this sequence.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>

    <section label="C.8.8.22" status="4" xml:id="sect_C.8.8.22">
      <title>RT Brachy Session Record Module</title>
      <table frame="box" label="C.8-58" rules="all" xml:id="table_C.8-58">
        <caption>RT Brachy Session Record Module Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;Brachy Control Point Delivered Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(3008,0160)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Sequence of machine configurations describing this Channel.</para>
              <para>Two or more items shall be included in this sequence.</para>
              <para>See <xref linkend="sect_A.20" xrefstyle="select: title"/> and <xref linkend="sect_C.8.8.22.1" xrefstyle="select: label"/> for description of Brachy Control Point Delivered Sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;Override Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(3008,0060)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Sequence of parameters that were overridden during the administration of the treatment immediately prior to the current control point.</para>
              <para>One or more Items are permitted in this sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;&gt;Override Parameter Pointer</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(3008,0062)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Contains the Data Element Tag of the attribute that was overridden.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;&gt;Operators' Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,1070)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Name of operator who authorized override.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;&gt;&gt;Operator Identification
                Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                >(0008,1072)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Identification of the
                operator who authorized override. Only a single Item is permitted in this sequence.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;Brachy Pulse Control Point Delivered Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(3008,0173)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>List of control points for this pulse.</para>
              <para>See <xref linkend="sect_C.8.8.22.1" xrefstyle="select: label"/>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;&gt;Override Sequence</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(3008,0060)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Parameters which were overridden during the administration of the treatment immediately prior to the current control point.</para>
              <para>One or more Items are permitted in this sequence.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;&gt;&gt;Override Parameter Pointer</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(3008,0062)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Data Element Tag of the attribute that was overridden.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;&gt;&gt;Operators' Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,1070)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Name of operator who authorized override.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;&gt;&gt;&gt;Operator Identification
                Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                >(0008,1072)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Identification of the
                operator who authorized override. Only a single Item is permitted in this sequence.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section label="C.8.8.26" status="4" xml:id="sect_C.8.8.26">
      <title>RT Ion Beams Session Record Module</title>
      <para/>
      <table frame="box" label="C.8.8.26-1" rules="all" xml:id="table_C.8.8.26-1">
        <caption>RT Ion Beams Session Record Module Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;&gt;&gt;Operators' Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,1070)</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Name of operator who authorized override.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">&gt;&gt;&gt;Operator Identification
                Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                >(0008,1072)</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Identification of the
                operator who authorized override. Only a single Item is permitted in this sequence.</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section label="C.31.1" status="3" xml:id="sect_C.31.1">
      <title>RT General Machine Verification Module</title>
      <table frame="box" label="C.31-1" rules="all" xml:id="table_C.31-1">
        <caption>RT General Machine Verification Module Attributes</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Attribute Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Tag</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>&gt;Operators' Name</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>(0008,1070)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Name of operator who authorized override of overridden parameter.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">&gt;Operator Identification
                Sequence</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"
                >(0008,1072)</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Identification of the
                operator who authorized override. Only a single Item is permitted in this sequence.</emphasis></emphasis></para>