  <title>CP-1388</title>
  <subtitle>Add Real World Value Map Reference to Measurements</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>STATUS</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Assigned</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Date of Last Update</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2014/05/28</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Andriy Fedorov</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:fedorov@bwh.harvard.edu"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2014/02/21</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number CP-1388</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Add Real World Value Map Reference to Measurements</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.16 2013</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Rationale for Correction:</para>
          <para>Measurements on regions of interest may be made using separate Real World Value Maps applied to
            stored pixel values in images, and it is useful to record a reference to which instance was used.</para>
          <para>These references may either be shared by an entire measurement group, or associated
            with an individual measurement.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>

    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 - Content Mapping Resource - Structured
                Reporting Templates to update the following templates to include an optional
                reference to RWV Map instances:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
           <section label="TID 300" status="3" xml:id="sect_TID_300">
        <title>Measurement</title>
        <para>...</para>
        <table rules="all" frame="box" label="TID 300" xml:id="table_TID_300">
          <caption>Measurement</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NUM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UNITS = $Units</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>17</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 4108 "Tracking Identifier"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >18</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/>&gt;</emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">INFERRED FROM</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">COMPOSITE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (<link linkend="DCM_dd3001" xrefstyle="select: nopage">dd3001</link>, DCM, "Real World Value Map used for measurement")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">SOP Class UID shall be Real World Value Mapping Storage ("1.2.840.10008.5.1.4.1.1.67")</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </table>
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Row 18</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">The reference to an RWV in
                      Row 18 allows measurements to be made in units that differ from the stored
                      pixel values in the images referenced in Row 13. E.g., for a PET SUVbw
                      measurement, the mapping from activity/concentration units in the referenced
                      image that was used (and which may be reused for measurements in the future)
                      may be encoded in a referenced RWV instance. This reference overrides any
                      reference in an including template (such as for a Measurement
                      Group).</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      </section>

      <section label="TID 1410" status="3" xml:id="sect_TID_1410">
        <title>Planar ROI Measurements</title>
        <para>...</para>
         <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Order:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Significant</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="TID 1410" xml:id="table_TID_1410">
          <caption>Planar ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_125007" xrefstyle="select: nopage">(125007, DCM, "Measurement Group")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IMAGE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121200" xrefstyle="select: nopage">(121200, DCM, "Illustration of ROI")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">9</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/>&gt;</emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">COMPOSITE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (<link linkend="DCM_dd3001" xrefstyle="select: nopage">dd3001</link>, DCM, "Real World Value Map used for measurement")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">SOP Class UID shall be Real World Value Mapping Storage ("1.2.840.10008.5.1.4.1.1.67")</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">9</emphasis><emphasis role="underline">10</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = $Measurement</para>
                <para>$Units = $Units</para>
                <para>$ModType = $ModType</para>
                <para>$ModValue = $ModValue</para>
                <para>$Method = $Method</para>
                <para>$Derivation = $Derivation</para>
                <para>$TargetSite = $TargetSite</para>
                <para>$TargetSiteMod = $TargetSiteMod</para>
                <para>$Equation = $Equation</para>
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
                <para>$DerivationParameter = $DerivationParameter</para>
                <para>$DerivationParameterUnits = $DerivationParameterUnits</para>
              </td>
            </tr>
          </tbody>
        </table>
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Row
                  9</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">The reference to an RWV in
                      Row 9 allows measurements to be made in units that differ from the stored
                      pixel values in the images referenced elsewhere in the template. E.g., for a
                      PET SUVbw measurement, the mapping from activity/concentration units in the
                      referenced image that was used (and which may be reused for measurements in
                      the future) may be encoded in a referenced RWV instance. This reference
                      applies to any measurements in included templates, unless
                      overridden).</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      </section>


      <section label="TID 1411" status="3" xml:id="sect_TID_1411">
        <title>Volumetric ROI Measurements</title>
        <para>...</para>
        <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Order:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Significant</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="TID 1411" xml:id="table_TID_1411">
          <caption>Volumetric ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_125007" xrefstyle="select: nopage">(125007, DCM, "Measurement Group")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IMAGE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV <link linkend="DCM_121200" xrefstyle="select: nopage">(121200, DCM, "Illustration of ROI")</link></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >13</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/>&gt;</emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">COMPOSITE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (<link linkend="DCM_dd3001" xrefstyle="select: nopage">dd3001</link>, DCM, "Real World Value Map used for measurement")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">SOP Class UID shall be Real World Value Mapping Storage ("1.2.840.10008.5.1.4.1.1.67")</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="strikethrough">13</emphasis><emphasis role="underline">14</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = $Measurement</para>
                <para>$Units = $Units</para>
                <para>$ModType = $ModType</para>
                <para>$ModValue = $ModValue</para>
                <para>$Method = $Method</para>
                <para>$Derivation = $Derivation</para>
                <para>$TargetSite = $TargetSite</para>
                <para>$TargetSiteMod = $TargetSiteMod</para>
                <para>$Equation = $Equation</para>
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
                <para>$DerivationParameter = $DerivationParameter</para>
                <para>$DerivationParameterUnits = $DerivationParameterUnits</para>
              </td>
            </tr>
          </tbody>
        </table>
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Row
                  13</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">The reference to an RWV in
                      Row 13 allows measurements to be made in units that differ from the stored
                      pixel values in the images referenced elsewhere in the template. E.g., for a
                      PET SUVbw measurement, the mapping from activity/concentration units in the
                      referenced image that was used (and which may be reused for measurements in
                      the future) may be encoded in a referenced RWV instance. This reference
                      applies to any measurements in included templates, unless
                      overridden).</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      </section>

      <section label="TID 1419" status="3" xml:id="sect_TID_1419">
        <title>ROI Measurements</title>
        <para>...</para>
        <table rules="all" frame="box" label="TID 1419" xml:id="table_TID_1419">
          <caption>ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1<emphasis role="bold"><emphasis role="underline"
                  >e</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NUM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UNITS = $Units</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>14</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  (121050, DCM, "Equivalent Meaning of Concept Name")
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">15</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/>&gt;</emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">INFERRED
                  FROM</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >COMPOSITE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (<link linkend="DCM_dd3001" xrefstyle="select: nopage">dd3001</link>, DCM, "Real World
                  Value Map used for measurement")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"/></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">SOP Class UID shall be Real
                  World Value Mapping Storage
                  ("1.2.840.10008.5.1.4.1.1.67")</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </table>
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
             <informaltable rules="all" frame="box">
               <tbody>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para>...</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para>...</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para><emphasis role="bold"><emphasis role="underline">Row 15</emphasis></emphasis></para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para><emphasis role="bold"><emphasis role="underline">The reference to an RWV
                      in Row 18 allows measurements to be made in units that differ from the stored
                      pixel values in the images referenced in the parent template. E.g., for a PET
                      SUVbw measurement, the mapping from activity/concentration units in the
                      referenced image that was used (and which may be reused for measurements in
                      the future) may be encoded in a referenced RWV instance. This reference
                      overrides any reference in an including template (such as for a Measurement
                      Group).</emphasis></emphasis></para>
                   </td>
                 </tr>
               </tbody>
             </informaltable>
           </section>


    </section>
    
        <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS3.16 - Content Mapping Resource - Controlled
                Terminology Definitions to add the following new concepts:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table rules="all" frame="box" label="D-1">
        <caption>DICOM Controlled Terminology Definitions</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Definition</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Notes</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd3001">dd3001</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Real World Value Map used for measurement</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A reference to the Real World Value Map applied to the stored image pixel values before their use for a measurement</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>           
        </tbody>
      </table>
    </section>

  </chapter>