  <title>CP-1386</title>
  <subtitle>Addition of Measurement Report Root Template for Planar and Volumetric ROIs</subtitle>
  <chapter>
    <title/>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>STATUS</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>Letter Ballot</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Date of Last Update</para></td>
        <td rowspan="1" colspan="1" valign="top"><para>2014/09/08</para></td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Person Assigned</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>David Clunie</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:dclunie@dclunie.com"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submitter Name</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>Andriy Fedorov</para>
          <para><link xmlns:xlink="http://www.w3.org/1999/xlink"
              xlink:href="mailto:fedorov@bwh.harvard.edu"/></para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Submission Date</para>
        </td>
        <td rowspan="1" colspan="1" valign="top">
          <para>2014/02/21</para>
        </td>
      </tr>
    </informaltable>
    <informaltable rules="all" frame="box">
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Correction Number: CP-1386</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Log Summary: Addition of Measurement Report Root Template for Planar and Volumetric
            ROIs</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Name of Standard</para>
          <para>PS3.6, PS3.16, PS3.17 2014b</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top"><para>Rationale for Correction:</para><para>CP 1112
            added Planar and Volumetric ROIs with the intention that future standard or private root
            templates would use them as necessary.</para>
          <para>There is a need for a root template
            that simply provides a list of measurements made, together with an image library with
            optional content to describe the images referenced.</para>
          <para>The principle of a set of measurements is generic, so the template is defined in a
            general way, and specialized for a particular application by a document title.</para>
          <para>It is desirable to make these heading concepts specifically image-derived
            measurements (as opposed to any kind of measurement). Note a recent survey that confirms
            users preference for separating "measurements" from other findings in reports (<link
              xl:href="http://www.academicradiology.org/article/S1076-6332%2814%2900087-7/abstract"
              >Travis et al. Preferences for Structured Reporting of Measurement
            Data</link>).</para>
          <para>Simple measurements (like distance) are also needed in addition to planar and
            volumetric ROIs, so TID 300 Measurement has also been included (in preference to more
            specific templates like TID 1400 Linear Measurement, which lack the general mechanism
            for including modifiers to define site, method, etc.).</para><para>Since a single ROI often has multiple measurements associated with it, the ROI measurement
            template (TID 1419) is refactored to allow specific modifiers to be present at a common
            as well as per-measurement level, and in particular the finding site and measurement
            method. In allowing this, it is apparent that the "1-n" cardinality should not be permit
            both on the inclusion of TID 1419 as well as the (formerly first) NUM content item of
            TID 1419, and so the former is restricted to 1.</para>
          <para>A mechanism has been added for including simple coded or text qualitative
            evaluations either at the top (subject) level or with each measurement group, since it
            is common to gather such information whilst making measurements.</para>
          <para>Time point information is optionally included at the measurement group level,
            allowing for recording the time point without constraining the scope of a single report
            instance to being one time point or multiple.</para>
          <para>An informative section is added to PS 3.17 to illustrate use of
            the templates.</para>
          <para>In future, the parameters passed to the subordinate templates may define specific
            regions and types of measurements, and context groups can be defined for those if
            necessary (e.g., a set of SUV and related PET measurements on FDG or FLT PET, for
            instance, or DCE-MRI measurements). The intent of this generic root template is not to
            preclude the future definition of more specific templates that specialize it for very
            specific purposes, which could retain the same structure but constrain the sets of
            values, and add other content (such as a description of the procedure reported, +/-
            references, e.g., to specific types of images (like T1 maps and dynamic T1 images for
            DCE-MRI, for instance).</para>
          <para>Also, the list of quantitative procedures reported is expected to be expanded in
            future. LOINC is lacking a lot of specific variants of procedures (e.g., FLT versus FDG
            whole body PET, DCE of prostate, etc.). SNOMED has slightly better coverage of PET, and
            has very patchy coverage of dynamic MR (pelvis and knee only). RadLex doesn't have
            DCE-MR (or any "dynamic" concepts, though there is "perfusion" for CT and NM only); it
            does have PET (modeled in a really weird way that makes it hard to find, as an NM
            modality with a technique of "PET" or "CT-PET"). Furture research applications probably
            need specific codes, and non-oncology apps like neuro and MSK will probably need newly
            procedures like Amyloid PET as well as structural MR (e.g., for hippocampal
            volume).</para>
          <para>There is a balance needed (and potentially some redundancy) between the codes used
            for the Document Title (top level container) and the Procedure Reported, but these are
            kept distinct in order to handle a measurement-specific report created for  standard of
            care images (e.g., oncology measurements on ordinary chest/abdomen/pelvis CT). There is
            a subtle difference between the imaging acquisition procedure reported, and the act of
            making the measurement, which is often performed separately; for arguments sake, this
            proposal assumes that the Procedure Reported is the acquisition procedure, and it turns
            out that the definition is lacking in DICOM, so one is added here.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">
          <para>Editor's Notes:</para>
          <para>Rows will be appropriately renumbered at final text or incorporation into the
            standard.</para>
        </td>
      </tr>
      <tr>
        <td rowspan="1" colspan="1" valign="top">Correction Wording:</td>
      </tr>
    </informaltable>
  </chapter>
  <chapter>
    <title/>
    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource -
                Structured Reporting Templates to add the following:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      <section label="TID t1tt1" status="3" xml:id="sect_TID_t1tt1">
        <title>Measurement Report</title>
        <para>This root template that encodes a list of Measurement Groups each containing lists of
          measurements, together with any derived measurements.</para>
        <para>Each Measurement Group is identified by Tracking ID and UIDs.</para>
        <para>An image library is available to describe characteristics of the images referenced by
          the measurements, if any.</para>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Order:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Non-Significant</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Root:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Yes</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="TID t1tt1" xml:id="table_TID_t1tt1">
          <caption>Measurement Report</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DCID cc1c1 "Measurement Report Document Titles"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Root node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 1204 "Language of Content Item and Descendants"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 1001 "Observation Context"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (121058, DCM, "Procedure reported")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>B<xref linkend="sect_CID_cc1c2" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID CP-1389 t4tt1 "Image Library"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_dd1d91" xrefstyle="select: nopage">dd1d91</link>, DCM, "Measurements")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IF row 10 and 12 are absent</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 1410 "Planar ROI Measurements"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = B<xref linkend="sect_CID_cc1c7" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
                <para>$Units = BCID 7181 "Abstract Multi-dimensional Image Model Component
                  Units"</para>
                <para>$Derivation = BCID 7464 "General Region of Interest Measurement
                  Modifiers"</para>
                <para>$Method = B<xref linkend="sect_CID_cc1c8" xrefstyle="select: labelnumber quotedtitle nopage"/> </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 1411 "Volumetric ROI Measurements"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = B<xref linkend="sect_CID_cc1c7" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
                <para>$Units = BCID 7181 "Abstract Multi-dimensional Image Model Component
                  Units"</para>
                <para>$Derivation = BCID 7464 "General Region of Interest Measurement
                  Modifiers"</para>
                <para>$Method = B<xref linkend="sect_CID_cc1c8" xrefstyle="select: labelnumber quotedtitle nopage"/> </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">9</td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_t1tt2" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = B<xref linkend="sect_CID_cc1c7" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
                <para>$Units = BCID 7181 "Abstract Multi-dimensional Image Model Component
                  Units"</para>
                <para>$Derivation = BCID 7464 "General Region of Interest Measurement
                  Modifiers"</para>
                <para>$Method = B<xref linkend="sect_CID_cc1c8" xrefstyle="select: labelnumber quotedtitle nopage"/> </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_dd1d92" xrefstyle="select: nopage">dd1d92</link>, DCM, "Derived Imaging Measurements")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IF row 6 and 12 are absent</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td>11</td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 1420 "Measurements Derived From Multiple ROI Measurements"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (C0034375, UMLS, "Qualitative Evaluations")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>IF row 6 and 10 are absent</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>13</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>14</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>
        
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 6, 10, 12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The conditions require that all except one of the "heading" containers is
                  present, though each may be empty.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 7, 8, 9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The baseline context groups defined allow for generic intensity and size
                  measurements, regardless of the geometry of the ROI (e.g., linear distance can be
                  measured on volumes, or volume can be estimated from a linear distance), and being
                  baseline, do not constrain the invoker from using other appropriate concepts
                  specific to the application.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>A Measurement Group is used to contain one or more individual measurements that are invocations of TID 300, consistent with TIDs 1410 and 1411,
                  which both already have Measurement Group containers as their roots.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 12, 13, 14</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>These content items allow encoding a flat list of name-value pairs that are
                  coded questions with coded or text answers, for example, to record categorical
                  observations related to the entire subject of the report rather than specific
                  measurement groups.</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        
        <para>...</para>
      </section>

      <section label="TID t1tt2" status="3" xml:id="sect_TID_t1tt2">
        <title>Measurement Group</title>
        <para>This template groups Measurements into a Measurement Group.</para>
        <para>Each Measurement Group is identified by Tracking ID and UIDs, and may be described as
          having being made at a particular time point.</para>
        <para>Measurement groups may contain various common measurement modifiers that are shared by
          all measurements in the group, such as method and finding site.</para>
        <table rules="all" frame="box" label="TID t1tt2" xml:id="table_TID_t1tt2_Parameters">
          <caption>Parameters</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>Parameter Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Parameter Usage</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Coded term or Context Group for Concept Name of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Units</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Units of Measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$ModType</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Modifier Name for Concept Name of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$ModValue</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Modifier Value for Concept Name of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Method</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Value for Measurement Method</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Derivation</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Value for Measurement Derivation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSite</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Value for Anatomic Location of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSiteMod</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Modifier Value for Anatomic Location of measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$Equation</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Coded term or Context Group for the equation or table from which the measurement was derived or computed</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$ImagePurpose</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Purpose of Reference for an image used as a source of the measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$WavePurpose</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Purpose of Reference for a waveform used as a source of the measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$RefAuthority</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Bibliographic reference or authority for statistical properties of a reference population</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$RangeAuthority</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Bibliographic reference or authority for the normal range of the measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameter</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Coded term or Context Group for Concept Name of a derivation parameter</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameterUnits</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Units of derivation parameter</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>$QualitativeEvaluations</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Evaluations encoded with code or text responses</para>
              </td>
            </tr>
          </tbody>
        </table>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Order:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Non-Significant</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Root:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">No</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="TID t1tt2" xml:id="table_TID_t1tt2">
          <caption>Measurement Group</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">1</td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINER</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (125007, DCM, "Measurement Group")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DT (112039, DCM, "Tracking Identifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UIDREF</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (112040, DCM, "Tracking Unique Identifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_t1tt3" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (G-C036, SRT, "Measurement Method")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Method</para>
              </td>
            </tr>
            
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (G-C0E3, SRT, "Finding Site")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSite</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (G-C171, SRT, "Laterality")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DCID 244 "Laterality"</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DT (G-A1F8, SRT, "Topographical modifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSiteMod</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>COMPOSITE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (CP-1388 dd3001, DCM, "Real World Value Map used for measurement")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>SOP Class UID shall be Real World Value Mapping Storage ("1.2.840.10008.5.1.4.1.1.67")</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 300 "Imaging Measurement"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = $Measurement</para>
                <para>$Units = $Units</para>
                <para>$ModType = $ModType</para>
                <para>$ModValue = $ModValue</para>
                <para>$Method = $Method</para>
                <para>$Derivation = $Derivation</para>
                <para>$TargetSite = $TargetSite</para>
                <para>$TargetSiteMod = $TargetSiteMod</para>
                <para>$Equation = $Equation</para>
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
                <para>$DerivationParameter = $DerivationParameter</para>
                <para>$DerivationParameterUnits = $DerivationParameterUnits</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>11</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$QualitativeEvaluations</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$QualitativeEvaluations</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>
        
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 2, 3, 10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The included TID 300 already contains an optional inclusion of TID 1408
                  Tracking Identifier, which in turns allows for either or both Tracking Identifier
                  and Tracking Unique Identifier content items; the intent of Rows 2 and 3 is not to
                  send these content items twice, but rather to specialize their use such that their
                  presence is mandatory, consistent with TIDs 1410 and 1411.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The reference to an RWV in
                      Row 9 allows measurements to be made in units that differ from the stored
                      pixel values in the images referenced elsewhere in the template. E.g., for a
                      PET SUVbw measurement, the mapping from activity/concentration units in the
                      referenced image that was used (and which may be reused for measurements in
                      the future) may be encoded in a referenced RWV instance. This reference
                      applies to any measurements in included templates, unless
                      overridden).</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Rows 11, 12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Allows encoding a flat list of name-value pairs that are coded questions with
                  coded or text answers, for example, to record categorical observations related to
                  the subject of the measurement group.</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        
        <para>...</para>
      </section>



      <section label="TID t1tt3" status="3" xml:id="sect_TID_t1tt3">
        <title>Time Point Context</title>
        <para>This template describes information about the time point, for example, at which a measurement was obtained.</para>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Order:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Non-Significant</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Root:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">No</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="TID t1tt3" xml:id="table_TID_t1tt3">
          <caption>Time Point Context</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_dd1800" xrefstyle="select: nopage">dd1800</link>, DCM,
                  "Subject Time Point Identifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_dd1801" xrefstyle="select: nopage">dd1801</link>, DCM,
                  "Protocol Time Point Identifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (C2348792, UMLS, "Time Point")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_dd1803" xrefstyle="select: nopage">dd1803</link>, DCM, "Time Point Type")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>B<xref linkend="sect_CID_cc1c6" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS OBS CONTEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NUM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (<link linkend="DCM_dd1804" xrefstyle="select: nopage">dd1804</link>, DCM, "Time Point Order")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UNITS = EV (1, UCUM, "no units")</para>
              </td>
            </tr>
          </tbody>
        </table>
        
        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Usually the same value as the Clinical Trial Time Point ID (0012,0050)
                  attribute in the Clinical Trial Study Module, though not confined to clinical
                  trial use. May or may not be human readable, and not required to be a DICOM
                  UID.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>All of the subjects within a treatment protocol that are examined at a
                  particular scheduled time point (e.g., "baseline", "pre-treatment", "first
                  post-treatment") will have the same Protocol Time Point Identifier, but different
                  Subject Time Point Identifiers. However, in different protocols, the Protocol Time
                  Point Identifiers for the same conceptual “time point” will be different. E.g.,
                  the “baseline” time point will have different Protocol Time Point Identifiers in
                  different protocols. May or may not be human readable, and not required to be a
                  DICOM UID.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Typically a short pre-defined label that has the same scope as Protocol Time
                  Point Identifier (i.e., same conceptual time point within a treatment protocol)
                  but is human-readable and understandable, e.g., "BASELINE" or "TP0", "TP1", etc.
                  Usually the same value as Clinical Trial Time Point Description (0012,0051)
                  attribute in the Clinical Trial Study Module, though not confined to clinical
                  trial use. The Concept Name is selected as (C2348792, UMLS, "Time Point") (which
                  is (C68568, NCIt, "Time Point"), defined as "a specific point in the time
                  continuum, including those established relative to an event") in order to be
                  compatible with external terminologies.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>More than one type is permitted, e.g., a time point may be "posttreatment" as well as "unscheduled" or "nadir", etc.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>Row 5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>The order is expected to be monotonically increasing within a particular scope
                  of usage, but is not required to start at 0 or 1, nor required to increase in
                  increments of 1 or even the same increment (e.g., to allow for retrospective
                  insertion of unscheduled time points). In clinical usage, the Time Point Order
                  would be expected to be temporally increasing, but in a clinical trial may be a
                  randomized reading order rather than a temporal order.</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        
        <para>...</para>
      </section>



    </section>

    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource -
                Structured Reporting Templates as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
            <section label="TID 1410" status="3" xml:id="sect_TID_1410">
        <title>Planar ROI Measurements</title>
              <para>...</para>
              <table rules="all" frame="box" label="TID 1410" xml:id="table_TID_1410_Parameters">
                <caption>Parameters</caption>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para>...</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para>...</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline"
                      >$QualitativeEvaluations</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">Evaluations encoded
                      with code or text responses</emphasis></emphasis></para>
                    </td>
                  </tr>
                </tbody>
              </table>
              
              <table rules="all" frame="box" label="TID 1410" xml:id="table_TID_1410">
          <caption>Planar ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UIDREF</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (112040, DCM, "Tracking Unique Identifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">3b</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">INCLUDE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">D<xref linkend="sect_TID_t1tt3" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1<emphasis role="bold"><emphasis role="strikethrough">-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = $Measurement</para>
                <para>$Units = $Units</para>
                <para>$ModType = $ModType</para>
                <para>$ModValue = $ModValue</para>
                <para>$Method = $Method</para>
                <para>$Derivation = $Derivation</para>
                <para>$TargetSite = $TargetSite</para>
                <para>$TargetSiteMod = $TargetSiteMod</para>
                <para>$Equation = $Equation</para>
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
                <para>$DerivationParameter = $DerivationParameter</para>
                <para>$DerivationParameterUnits = $DerivationParameterUnits</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">10</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >$QualitativeEvaluations</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">11</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">TEXT</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >$QualitativeEvaluations</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>
              
              <para>
                <emphasis role="bold">Content Item Descriptions</emphasis>
              </para>
              <informaltable rules="all" frame="box">
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para>...</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para>...</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">Rows 10,
                    11</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para><emphasis role="bold"><emphasis role="underline">Allows encoding a flat
                      list of name-value pairs that are coded questions with coded or text answers,
                      for example, to record categorical observations related to the subject of the
                      measurement group.</emphasis></emphasis></para>
                    </td>
                  </tr>
                </tbody>
              </informaltable>
              
              <para>...</para>
            </section>
      <section label="TID 1411" status="3" xml:id="sect_TID_1411">
        <title>Volumetric ROI Measurements</title>
        <para>...</para>

        <table rules="all" frame="box" label="TID 1411" xml:id="table_TID_1411_Parameters">
          <caption>Parameters</caption>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >$QualitativeEvaluations</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Evaluations encoded with code
                      or text responses</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </table>

        <table rules="all" frame="box" label="TID 1411" xml:id="table_TID_1411">
          <caption>Volumetric ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UIDREF</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (112040, DCM, "Tracking Unique Identifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">3b</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">INCLUDE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">D<xref linkend="sect_TID_t1tt3" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>13</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CONTAINS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>D<xref linkend="sect_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1<emphasis role="bold"><emphasis role="strikethrough">-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement = $Measurement</para>
                <para>$Units = $Units</para>
                <para>$ModType = $ModType</para>
                <para>$ModValue = $ModValue</para>
                <para>$Method = $Method</para>
                <para>$Derivation = $Derivation</para>
                <para>$TargetSite = $TargetSite</para>
                <para>$TargetSiteMod = $TargetSiteMod</para>
                <para>$Equation = $Equation</para>
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
                <para>$DerivationParameter = $DerivationParameter</para>
                <para>$DerivationParameterUnits = $DerivationParameterUnits</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">14</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >$QualitativeEvaluations</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">15</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CONTAINS</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">TEXT</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                      >$QualitativeEvaluations</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1-n</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>

        <para>
          <emphasis role="bold">Content Item Descriptions</emphasis>
        </para>
        <informaltable rules="all" frame="box">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Rows 14,
                    15</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Allows encoding a flat
                  list of name-value pairs that are coded questions with coded or text answers,
                  for example, to record categorical observations related to the subject of the
                  measurement group.</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </informaltable>

      </section>

      
      <section label="TID 1419" status="3" xml:id="sect_TID_1419">
        <title>ROI Measurements</title>
        <para>...</para>
        <table rules="all" frame="box" label="TID 1419" xml:id="table_TID_1419">
          <caption>ROI Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para/>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>NL</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Rel with Parent</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VT</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Concept Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>VM</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Req Type</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Condition</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>Value Set Constraint</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1a</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">HAS CONCEPT
                    MOD</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (G-C036, SRT, "Measurement Method")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">$Method</emphasis></emphasis></para>
              </td>
            </tr>
            
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1b</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"></emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">HAS CONCEPT
                    MOD</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (G-C0E3, SRT, "Finding Site")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">$TargetSite</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1c</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">HAS CONCEPT
                    MOD</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">EV (G-C171, SRT, "Laterality")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">DCID 244 "Laterality"</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >1d</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">&gt;</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">HAS CONCEPT
                    MOD</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">CODE</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">DT (G-A1F8, SRT, "Topographical modifier")</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">1</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">U</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">$TargetSiteMod</emphasis></emphasis></para>
              </td>
            </tr>
            


            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>1<emphasis role="bold"><emphasis role="underline"
                  >e</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NUM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Measurement</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>UNITS = $Units</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$ModType</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$ModValue</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (G-C036, SRT, "Measurement Method")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Method</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (121401, DCM, "Derivation")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Derivation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (G-C0E3, SRT, "Finding Site")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSite</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>EV (G-C171, SRT, "Laterality")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DCID 244 "Laterality"</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CODE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DT (G-A1F8, SRT, "Topographical modifier")</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$TargetSiteMod</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS PROPERTIES</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 310 "Measurement Properties"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$RefAuthority = $RefAuthority</para>
                <para>$RangeAuthority = $RangeAuthority</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INFERRED FROM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NUM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameter</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>UC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Row 10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameterUnits</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>10</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>R-INFERRED FROM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NUM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameter</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1-n</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>UC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Row 9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$DerivationParameterUnits</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>11</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INFERRED FROM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 315 "Equation or Table"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>UC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Row 12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>$Equation = $Equation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>12</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INFERRED FROM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DCID 228 "Equation or Table"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>UC</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>XOR Row 11</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>13</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>INCLUDE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>DTID 1000 "Quotation"</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para>14</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>HAS CONCEPT MOD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>TEXT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>
                  (121050, DCM, "Equivalent Meaning of Concept Name")
                </para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>1</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para/>
              </td>
            </tr>
          </tbody>
        </table>
      </section>


    </section>
    
    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource -
              Context Groups to amend the following context groups to add new values:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <section label="CID 3488" status="3" xml:id="sect_CID_3488">
        <title>Min/Max/Mean</title>
        <para>This context group contains modifiers that indicate whether the measurement is a minimum, maximum or averaged value.</para>
        <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">20040614</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID 3488" xml:id="table_CID_3488">
          <caption>Min/Max/Mean</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>G-A437</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Maximum</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>R-404FB</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Minimum</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>R-00317</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Mean</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      <section label="CID 6143" status="3" xml:id="sect_CID_6143">
        <title><emphasis role="bold"><emphasis role="underline">Lesion </emphasis></emphasis>Response<emphasis role="bold"><emphasis role="strikethrough"> Criteria</emphasis></emphasis></title>
        <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">20030108</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID 6143" xml:id="table_CID_6143">
          <caption><emphasis role="bold"><emphasis role="underline">Lesion </emphasis></emphasis>Response<emphasis role="bold"><emphasis role="strikethrough"> Criteria</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include <xref linkend="sect_CID_6144" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis>
                </para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      <section label="CID 6144" status="3" xml:id="sect_CID_6144">
        <title>RECIST<emphasis role="bold"><emphasis role="underline">-Defined Lesion</emphasis></emphasis> Response<emphasis role="bold"><emphasis role="strikethrough"> Criteria</emphasis></emphasis></title>
        <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">20030108</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID 6144" xml:id="table_CID_6144">
          <caption>RECIST<emphasis role="bold"><emphasis role="underline">-Defined Lesion</emphasis></emphasis> Response<emphasis role="bold"><emphasis role="strikethrough"> Criteria</emphasis></emphasis></caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>112041</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Target Lesion Complete Response</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>...</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>...</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      
      
      <section label="CID 7464" status="3" xml:id="sect_CID_7464">
        <title>General Region of Interest Measurement<emphasis role="bold"><emphasis role="underline"> Modifier</emphasis></emphasis>s</title>
        <para>This context group contains <emphasis role="bold"><emphasis role="underline">modifiers of </emphasis></emphasis>measurements of ROIs.</para>
        <variablelist spacing="compact">
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold"><emphasis role="strikethrough">20121101</emphasis><emphasis role="underline">yyyymmdd</emphasis></emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID 7464" xml:id="table_CID_7464">
          <caption>General Region of Interest Measurement<emphasis role="bold"><emphasis role="underline"> Modifier</emphasis></emphasis>s</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include <xref linkend="sect_CID_3488" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>R-10047</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Standard Deviation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>R-40507</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Total</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>R-00319</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Median</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>R-0032E</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Mode</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"><link linkend="DCM_dd1501" xrefstyle="select: nopage">dd1501</link></emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Peak Value Within ROI</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">UMLS</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">C0681921</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Coefficient of Variance</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >DCM</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <emphasis role="bold"><emphasis role="underline"><link linkend="DCM_dd1710" xrefstyle="select: nopage">dd1710</link></emphasis></emphasis>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Skewness</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline"
                  >DCM</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <emphasis role="bold"><emphasis role="underline"><link linkend="DCM_dd1711" xrefstyle="select: nopage">dd1711</link></emphasis></emphasis>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Kurtosis</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">UMLS</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">C1711260</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Variance</emphasis></emphasis></para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">UMLS</emphasis></emphasis></para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">C2347976</emphasis></emphasis></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para><emphasis role="bold"><emphasis role="underline">Root Mean Square</emphasis></emphasis></para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
    <section label="CID 7474" status="3" xml:id="sect_CID_7474">
      <title>General Volume Calculation Methods</title>
      <variablelist>
        <varlistentry>
          <term>
            <emphasis role="bold">Type:</emphasis>
          </term>
          <listitem>
            <para>
              <emphasis role="bold">Extensible</emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term>
            <emphasis role="bold">Version:</emphasis>
          </term>
          <listitem>
            <para>
              <emphasis role="bold"><emphasis role="strikethrough">20070827</emphasis><emphasis role="underline">yyyymmdd</emphasis></emphasis>
            </para>
          </listitem>
        </varlistentry>
      </variablelist>
      <table rules="all" frame="box" label="CID 7474" xml:id="table_CID_7474">
        <caption>General Volume Methods</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Coding Scheme Designator</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                122503
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Integration of sum of closed areas on contiguous slices</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold"><emphasis role="underline"><link linkend="DCM_dd1301" xrefstyle="select: nopage">dd1301</link></emphasis></emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Sum of segmented voxel volumes</emphasis></emphasis></para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    </section>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource -
              Context Groups to add the following:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      <section label="CID cc1c1" status="3" xml:id="sect_CID_cc1c1">
      <title>Measurement Report Document Titles</title>
      <variablelist>
        <varlistentry>
          <term>
            <emphasis role="bold">Type:</emphasis>
          </term>
          <listitem>
            <para>
              <emphasis role="bold">Extensible</emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term>
            <emphasis role="bold">Version:</emphasis>
          </term>
          <listitem>
            <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
          </listitem>
        </varlistentry>
      </variablelist>
        <table rules="all" frame="box" label="CID cc1c1" xml:id="table_CID_cc1c1">
          <caption>Measurement Report Document Titles</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Coding Scheme Designator</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><link linkend="DCM_dd1dd1" xrefstyle="select: nopage">dd1dd1</link></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Imaging Measurement Report</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><link linkend="DCM_dd1dd2" xrefstyle="select: nopage">dd1dd2</link></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Oncology Measurement Report</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><link linkend="DCM_dd1dd3" xrefstyle="select: nopage">dd1dd3</link></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Dynamic Contrast MR Measurement Report</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><link linkend="DCM_dd1dd4" xrefstyle="select: nopage">dd1dd4</link></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>PET Measurement Report</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
      <section label="CID cc1c2" status="3" xml:id="sect_CID_cc1c2">
        <title>Quantitative Diagnostic Imaging Procedures</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID cc1c2" xml:id="table_CID_cc1c2">
          <caption>Quantitative Diagnostic Imaging Procedures</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>P5-09051</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Magnetic resonance imaging guidance</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><link linkend="DCM_dd1401" xrefstyle="select: nopage">dd1401</link></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Multiparametric (DWI, DCE and T2 Mapping) MRI</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><link linkend="DCM_dd1402" xrefstyle="select: nopage">dd1402</link></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Multiparametric (DWI, DCE and T2 Mapping) MRI of prostate</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para><link linkend="DCM_dd1403" xrefstyle="select: nopage">dd1403</link></para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Multiparametric (DWI, DCE and T2 Mapping) MRI of whole body</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>P5-0907F</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Dynamic magnetic resonance imaging of knee</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>P5-70694</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Dynamic magnetic resonance imaging of pelvis</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>LN</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>44139-4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>PET whole body</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>P5-080FF</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>PET/CT FDG imaging of whole body</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>SRT</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>P5-08118</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>PET/CT MET imaging of whole body</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>RADLEX</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>RPID96</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>CT HEAD PERFUSION WITH IV CONTRAST</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>RADLEX</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>RPID5258</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NM HEAD PERFUSION BRAIN SPECT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>RADLEX</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>RPID5427</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>NM HEAD PERFUSION BRAIN PET-CT AV-45</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>

      <section label="CID cc1c3" status="3" xml:id="sect_CID_cc1c3">
        <title>PET Region of Interest Measurements</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID cc1c3" xml:id="table_CID_cc1c3">
          <caption>PET Region of Interest Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1502" xrefstyle="select: nopage">dd1502</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Metabolic Volume</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1520" xrefstyle="select: nopage">dd1520</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Total Lesion Glycolysis</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1521" xrefstyle="select: nopage">dd1521</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Glycolysis</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1522" xrefstyle="select: nopage">dd1522</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Total Lesion Proliferation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1523" xrefstyle="select: nopage">dd1523</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Proliferative Activity</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1530" xrefstyle="select: nopage">dd1530</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Standardized Added Metabolic Activity</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1531" xrefstyle="select: nopage">dd1531</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Standardized Added Metabolic Activity Background</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1630" xrefstyle="select: nopage">dd1630</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Lesion to Background SUV Ratio</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>
                  <link linkend="DCM_dd1631" xrefstyle="select: nopage">dd1631</link>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Background for Lesion to Background SUV Ratio</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
     
      <section label="CID cc1c4" status="3" xml:id="sect_CID_cc1c4">
        <title>Grey Level Co-occurrence Matrix Measurements</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID cc1c4" xml:id="table_CID_cc1c4">
          <caption>Grey Level Co-occurrence Matrix Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1712" xrefstyle="select: nopage">dd1712</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Entropy of GLCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1713" xrefstyle="select: nopage">dd1713</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Energy of GLCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1714" xrefstyle="select: nopage">dd1714</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Homogeneity of GLCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1715" xrefstyle="select: nopage">dd1715</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Contrast of GLCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1716" xrefstyle="select: nopage">dd1716</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Dissimilarity of GLCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1717" xrefstyle="select: nopage">dd1717</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>ASM of GLCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="center" colspan="1" rowspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1718" xrefstyle="select: nopage">dd1718</link>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para>Correlation of GLCM</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      
      <section label="CID cc1c5" status="3" xml:id="sect_CID_cc1c5">
        <title>Texture Measurements</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID cc1c5" xml:id="table_CID_cc1c5">
          <caption>Texture Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include <xref linkend="sect_CID_cc1c4" xrefstyle="select: labelnumber quotedtitle nopage"/></emphasis>
                </para>
              </td>
            </tr>
            <tr>
              <td align="center" colspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1701" xrefstyle="select: nopage">dd1701</link>
              </td>
              <td align="left" colspan="1">
                <para>Fractal Dimension</para>
              </td>
            </tr>
          </tbody>
        </table>
      </section>
      
      <section label="CID cc1c6" status="3" xml:id="sect_CID_cc1c6">
        <title>Time Point Types</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID cc1c6" xml:id="table_CID_cc1c6">
          <caption>Time Point Types</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr>
              <td align="center" colspan="1">
                <para>UMLS</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C1442488</para>
              </td>
              <td align="left" colspan="1">
                <para>Baseline</para>
              </td>
            </tr>
            <tr>
              <td align="center" colspan="1">
                <para>UMLS</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C3539075</para>
              </td>
              <td align="left" colspan="1">
                <para>Pretreatment</para>
              </td>
            </tr>
            <tr>
              <td align="center" colspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1901" xrefstyle="select: nopage">dd1901</link>
              </td>
              <td align="left" colspan="1">
                <para>Posttreatment</para>
              </td>
            </tr>
            <tr>
              <td align="center" colspan="1">
                <para>DCM</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <link linkend="DCM_dd1902" xrefstyle="select: nopage">dd1902</link>
              </td>
              <td align="left" colspan="1">
                <para>Eligibility</para>
              </td>
            </tr>
            <tr>
              <td align="center" colspan="1">
                <para>UMLS</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C1699701</para>
              </td>
              <td align="left" colspan="1">
                <para>Unscheduled</para>
              </td>
            </tr>
            <tr>
              <td align="center" colspan="1">
                <para>UMLS</para>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para>C1708760</para>
              </td>
              <td align="left" colspan="1">
                <para>Nadir</para>
              </td>
            </tr>
          </tbody>
        </table>
        <note>
          <orderedlist>
            <listitem>
              <para>(C1442488, UMLS, "Baseline") is (C25213, NCIt, "Baseline"). The undefined (121079, DCM, "Baseline") that is used in CID 7003 Diagnostic Imaging Report Purposes of Reference is not used in this context.</para>
            </listitem>
            <listitem>
              <para>(C3539075, UMLS, "Pretreatment") is (C103341, NCIt, "Pretreatment").</para>
            </listitem>
            <listitem>
              <para>(C1708760, UMLS, "Nadir") is (C43517, NCIt, "Nadir"), and is a synonymn for "lowest", though "nadir" is more commnly used in the context of therapeutic response criteria.</para>
            </listitem>
          </orderedlist>
        </note>
      </section>


      <section label="CID cc1c7" status="3" xml:id="sect_CID_cc1c7">
        <title>Generic Intensity and Size Measurements</title>
        <variablelist>
          <varlistentry>
            <term>
              <emphasis role="bold">Type:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">Extensible</emphasis>
              </para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>
              <emphasis role="bold">Version:</emphasis>
            </term>
            <listitem>
              <para>
                <emphasis role="bold">yyyymmdd</emphasis>
              </para>
            </listitem>
          </varlistentry>
        </variablelist>
        <table rules="all" frame="box" label="CID cc1c7" xml:id="table_CID_cc1c7">
          <caption>Generic Intensity and Size Measurements</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Coding Scheme Designator</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para>
                  <emphasis role="bold">Code Meaning</emphasis>
                </para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include CID 7180 "Abstract Multi-Dimensional Image Model Component Semantics"</emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include CID 7470 "Linear Measurements"</emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include CID 7471 "Area Measurements"</emphasis>
                </para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="3" rowspan="1">
                <para>
                  <emphasis role="italic">Include CID 7472 "Volume Measurements"</emphasis>
                </para>
              </td>
            </tr>
            
          </tbody>
        </table>
      </section>
    </section>
    
    <section label="CID cc1c8" status="3" xml:id="sect_CID_cc1c8">
      <title>Response Criteria</title>
      <variablelist>
        <varlistentry>
          <term>
            <emphasis role="bold">Type:</emphasis>
          </term>
          <listitem>
            <para>
              <emphasis role="bold">Extensible</emphasis>
            </para>
          </listitem>
        </varlistentry>
        <varlistentry>
          <term>
            <emphasis role="bold">Version:</emphasis>
          </term>
          <listitem>
            <para>
              <emphasis role="bold">yyyymmdd</emphasis>
            </para>
          </listitem>
        </varlistentry>
      </variablelist>
      <table rules="all" frame="box" label="CID cc1c8" xml:id="table_CID_cc1c8">
        <caption>Response Criteria</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Coding Scheme Designator</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr>
            <td align="center" colspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>112029</para>
            </td>
            <td align="left" colspan="1">
              <para>WHO</para>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><link linkend="DCM_dd1910" xrefstyle="select: nopage">dd1910</link></para>
            </td>
            <td align="left" colspan="1">
              <para>RECIST 1.0</para>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1">
              <para>DCM</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para><link linkend="DCM_dd1910" xrefstyle="select: nopage">dd1911</link></para>
            </td>
            <td align="left" colspan="1">
              <para>RECIST 1.1</para>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1">
              <para>NCIt</para>
            </td>
            <td align="center" colspan="1" rowspan="1">
              <para>C114879</para>
            </td>
            <td align="left" colspan="1">
              <para>RANO</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource -
                Controlled Terminology Definitions to add the definition:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table rules="all" frame="box" label="D-1">
        <caption>DICOM Controlled Terminology Definitions</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Definition</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Notes</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_121058">121058</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Procedure Reported</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">The diagnostic or interventional procedure that is the subject of the report.</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
        </tbody>
      </table>
    </section>
    

    
    
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.16 - Content Mapping Resource -
                Controlled Terminology Definitions to add the following new
              concepts:</emphasis></para>
          </td>
        </tr>
      </informaltable>
      
      <table rules="all" frame="box" label="D-1">
        <caption>DICOM Controlled Terminology Definitions</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Value</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Code Meaning</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Definition</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Notes</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1dd1">dd1dd1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Imaging Measurement Report</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A structured report containing the quantitative results of human or machine
                analysis of images.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1dd2">dd1dd2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Oncology Measurement Report</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A structured report containing the quantitative results of human or machine
                analysis of images for oncology evaluation.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1dd3">dd1dd3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Dynamic Contrast MR Measurement Report</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A structured report containing the quantitative results of human or machine
                analysis of DCE-MR.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1dd4">dd1dd4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>PET Measurement Report</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A structured report containing the quantitative results of human or machine
                analysis of PET images.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1d91">dd1d91</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Imaging Measurements</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurements made on images</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1d92">dd1d92</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derived Imaging Measurements</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurements derived from measurements made on images.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1401">dd1401</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Multiparametric (DWI, DCE and T2 Mapping) MRI</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>An MRI procedure in which multiple parameters including diffusion, dynamic
                contrast and T2 are measured.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1402">dd1402</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Multiparametric (DWI, DCE and T2 Mapping) MRI of prostate</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>An MRI procedure of the prostate in which multiple parameters including
                diffusion, dynamic contrast and T2 are measured.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1403">dd1403</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Multiparametric (DWI, DCE and T2 Mapping) MRI of whole body</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>An MRI procedure of the whole body in which multiple parameters including
                diffusion, dynamic contrast and T2 are measured.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1301">dd1301</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Sum of segmented voxel volumes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The volume derived by summing the volumes of all the voxels (and partial voxels if the segment contains partially occupied voxels) included in the segment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1501">dd1501</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Peak Value Within ROI</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Maximum average gray value that is calculated from a 1 cubic centimeter sphere placed within the region of interest.</para>
              <para>Reference Wahl et al PERCIST article.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>

          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1502">dd1502</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Metabolic Volume</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The volume of a lesion (e.g., a tumor) ascertained through information about its metabolic activity (e.g., SUV on PET).</para>
              <para>Abbreviated "MV". Synonymous with Metabolic Tumor Volume (MTV).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1520">dd1520</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Total Lesion Glycolysis</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The total activity of a lesion obtained as the product of its volume and its glycolytic activity (on FDG-PET).</para>
              <para>The volume may be defined on the same modality (e.g., the MV on FDG-PET by some thresholding or other technique) or on another spatially registered modality (e.g., the lesion outline segmented on CT or MR).</para>
              <para>Does not apply to other radiopharmaceuticals than those involved in glucose metabolism.</para>
              <para>Abbreviated TLG. Synonymnous with "Tumor Lesion Glycolysis".</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1521">dd1521</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Glycolysis</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The amount glycolytic activity summed across all voxels in a defined region or within a defined range of SUV (on FDG-PET).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1522">dd1522</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Total Lesion Proliferation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The total activity of a lesion obtained as the product of its volume and its proliferative activity (on FLT-PET).</para>
              <para>The volume may be defined on the same modality (e.g., the MV on FDG-PET by some thresholding or other technique) or on another spatially registered modality (e.g., the lesion outline segmented on CT or MR).</para>
              <para>Does not apply to other radiopharmaceuticals than those involved in cellular proliferation.</para>
              <para>Abbreviated TLP. Synonymnous with "Tumor Lesion Proliferation".</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1523">dd1523</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Proliferative Activity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The amount proliferative activity summed across all voxels in a defined region or within a defined range of SUV (on FLT-PET).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1530">dd1530</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Standardized Added Metabolic Activity (SAM)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A background-corrected, partial volume independent version of TLG.</para>
              <para>SAM is calculated
                by drawing a volume of interest (VOI1) around the tumour
                and a larger VOI (VOI2) around VOI1. Subtracting the
                background activity in VOI2-VOI1 from VOI1 yields SAM.</para>
              <para>See Mertens et al. "Standardized added metabolic activity (SAM): a partial
                volume independent marker of total lesion glycolysis
                in liver metastases". Eur J Nucl Med Mol Imaging (2012) 39:1441–1448.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1531">dd1531</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Standardized Added Metabolic Activity (SAM) Background</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The background value (VOI2-VOI1) used to calculate Standardized Added Metabolic Activity (SAM).</para>
              <para>SAM is calculated
                by drawing a volume of interest (VOI1) around the tumour
                and a larger VOI (VOI2) around VOI1. Subtracting the
                background activity in VOI2-VOI1 from VOI1 yields SAM.</para>
              <para>See Mertens et al. "Standardized added metabolic activity (SAM): a partial
                volume independent marker of total lesion glycolysis
                in liver metastases". Eur J Nucl Med Mol Imaging (2012) 39:1441–1448.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1630">dd1630</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Lesion to Background SUV Ratio</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The ratio of the SUV within a tumor to the SUV of a pre-defined background region.</para>
              <para>Need reference(s).</para>
              <para>A more general concept than Tumor to Background Ratio (TBR).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1631">dd1631</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Background for Lesion to Background SUV Ratio</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>The SUV of a pre-defined background region used to compute Lesion to Background SUV Ratio.</para>
              <para>Need reference(s).</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1701">dd1701</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Fractal Dimension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>A statistical index of complexity comparing how detail in a fractal pattern changes with the scale at which it is measured; a ratio of the change in detail to the change in scale.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1710">dd1710</para>
            </td>
            <td align="left" colspan="1">
              <para>Skewness</para>
            </td>
            <td align="left" colspan="1">
              <para>Measure of the asymmetry of the probability distribution of a real-valued random variable about its mean.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1711">dd1711</para>
            </td>
            <td align="left" colspan="1">
              <para>Kurtosis</para>
            </td>
            <td align="left" colspan="1">
              <para>Measure of the peakedness of the probability distribution of a real-valued random variable.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1712">dd1712</para>
            </td>
            <td align="left" colspan="1">
              <para>Entropy of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>The zero order entropy of a Gray Level Co-occurrence Matrix (GLCM). A measure of disorder.</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1713">dd1713</para>
            </td>
            <td align="left" colspan="1">
              <para>Energy of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>The energy (uniformity) (square root of the Angular Second Moment (ASM)) of a Gray Level Co-occurrence Matrix (GLCM). A measure of orderliness.</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1714">dd1714</para>
            </td>
            <td align="left" colspan="1">
              <para>Homogeneity of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>The Inverse Difference Moment of a Gray Level Co-occurrence Matrix (GLCM).</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1715">dd1715</para>
            </td>
            <td align="left" colspan="1">
              <para>Contrast of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>The sum of squares variance of a Gray Level Co-occurrence Matrix (GLCM).</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1716">dd1716</para>
            </td>
            <td align="left" colspan="1">
              <para>Dissimilarity of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>The dissimilarity of a Gray Level Co-occurrence Matrix (GLCM).</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1717">dd1717</para>
            </td>
            <td align="left" colspan="1">
              <para>ASM of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>The Angular Second Moment of a Gray Level Co-occurrence Matrix (GLCM).</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1718">dd1718</para>
            </td>
            <td align="left" colspan="1">
              <para>Correlation of GLCM</para>
            </td>
            <td align="left" colspan="1">
              <para>A measure of the linear dependency of grey levels on those of neighbouring pixels of a Gray Level Co-occurrence Matrix (GLCM).</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/equations.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1719">dd1719</para>
            </td>
            <td align="left" colspan="1">
              <para>Gray Level Co-occurrence Matrix (GLCM)</para>
            </td>
            <td align="left" colspan="1">
              <para>A tabulation of how often different combinations of pixel values (grey levels) occur in an image.</para>
              <para>See <link xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://www.fp.ucalgary.ca/mhallbey/the_glcm.htm"/>.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1800">dd1800</para>
            </td>
            <td align="left" colspan="1">
              <para>Subject Time Point Identifier</para>
            </td>
            <td align="left" colspan="1">
              <para>An identifier of a specific time point in a continuum, which is unique within an
                appropriate local context (such as an entire organization, system or treatment
                protocol), which identifies the time point for a specific patient.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1801">dd1801</para>
            </td>
            <td align="left" colspan="1">
              <para>Protocol Time Point Identifier</para>
            </td>
            <td align="left" colspan="1">
              <para>An identifier of a specific time point in a continuum, which is unique within an
                appropriate local context (such as an entire organization, system or treatment
                protocol), which identifies the time point “slot” within a treatment protocol using
                the same value for all patients in the protocol.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1803">dd1803</para>
            </td>
            <td align="left" colspan="1">
              <para>Time Point Type</para>
            </td>
            <td align="left" colspan="1">
              <para>A pre-defined type of a specific time point in a continuum.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1804">dd1804</para>
            </td>
            <td align="left" colspan="1">
              <para>Time Point Order</para>
            </td>
            <td align="left" colspan="1">
              <para>A number indicating the order of a time point relative to other time points in
                the same continuum.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1901">dd1901</para>
            </td>
            <td align="left" colspan="1">
              <para>Posttreatment</para>
            </td>
            <td align="left" colspan="1">
              <para>The time after the treatment of interest.</para>
            </td>
            <td align="left" colspan="1"/>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1902">dd1902</para>
            </td>
            <td align="left" colspan="1">
              <para>Eligibility</para>
            </td>
            <td align="left" colspan="1">
              <para>For the purpose of determining eligibility for a protocol.</para>
            </td>
            <td align="left" colspan="1">
              <para>Similar but not identical to (21954-3, LN, "Protocol eligibility status
                Cancer"), since not constrained to cancer, etc.</para>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1910">dd1910</para>
            </td>
            <td align="left" colspan="1">
              <para>RECIST 1.0</para>
            </td>
            <td align="left" colspan="1">
              <para>Response Evaluation Criteria in Solid Tumors version 1.0. See [RECIST] in Normative References.</para>
            </td>
            <td align="left" colspan="1">
              <para>More specific than (112022, DCM, "RECIST") or (C1709926, UMLS, "RECIST") or (C49164, NCIt, "RECIST") in that a specific version is specified.</para>
            </td>
          </tr>
          <tr>
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="DCM_dd1911">dd1911</para>
            </td>
            <td align="left" colspan="1">
              <para>RECIST 1.1</para>
            </td>
            <td align="left" colspan="1">
              <para>Response Evaluation Criteria in Solid Tumors Version 1.1. See Eisenhauer et al. "New Response Evaluation Criteria in Solid Tumours: Revised RECIST Guideline (version 1.1)." European Journal of Cancer 45, no. 2 (n.d.): 228–47. doi:10.1016/j.ejca.2008.10.026..</para>
            </td>
            <td align="left" colspan="1">
              <para>More specific than (112022, DCM, "RECIST") or (C1709926, UMLS, "RECIST") or (C49164, NCIt, "RECIST") in that a specific version is specified.</para>
            </td>
          </tr>
        </tbody>
      </table>
    </section>


    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.6 - Data Dictionary as follows:</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>
    
    <section>
      <title/>
      <table rules="all" frame="box" xml:id="table_A-3" label="A-3">
        <caption>Context Group UID Values</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Context UID</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Context Identifier</emphasis>
              </para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">Context Group Name</emphasis>
              </para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.840.10008.6.​1.​463</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>6143</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">Lesion </emphasis></emphasis>Response<emphasis role="bold"><emphasis role="strikethrough"> Criteria</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.840.10008.6.​1.​464</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>6144</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>RECIST<emphasis role="bold"><emphasis role="underline">-Defined Lesion</emphasis></emphasis> Response<emphasis role="bold"><emphasis role="strikethrough"> Criteria</emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.840.10008.6.​1.​951</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>7464</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>General Region of Interest Measurement<emphasis role="bold"><emphasis role="underline"> Modifier</emphasis></emphasis>s</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn1</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c1" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c1" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn2</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c2" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c2" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn3</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c3" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c3" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn4</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c4" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c4" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn5</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c5" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c5" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn6</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c6" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c6" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn7</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c7" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c7" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.6.​1.​nn8</emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c8" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_CID_cc1c8" xrefstyle="select: title"/></emphasis></emphasis></para>
            </td>
          </tr>
          
          
        </tbody>
      </table>
    </section>
    
  </chapter>
  
  
  <chapter label="XXX" status="1" xml:id="chapter_XXX">
    <title>Measurement Report SR Document for Planar and Volumetric ROI (Informative)</title>
    <section>
      <title/>
      <informaltable rules="all" frame="box">
        <tr>
          <td>
            <para><emphasis role="italic">Amend DICOM PS 3.17 - Explanatory Information to add new Annex XXX:</emphasis></para>
          </td>
        </tr>
      </informaltable>
    </section>
    <section>
      <title></title>
      <para>This Annex contains examples of the use of ROI templates within Measurement Report SR Documents.</para>
    </section>
    <section label="XXX.1" status="2" xml:id="sect_XXX.1">
      <title>Measurement Report SR Document Volumetric ROI on CT Example</title>
      <para>This CT example describes the minimum content necessary to encode a single measurement
        (volume) made from a single volumetric ROI encoded as a single segment that spans two source
        CT images.</para>
      <note>
        <orderedlist>
          <listitem>
            <para>References to Segmentation Image or Surface Segmentation objects are encoded as IMAGE references, with a single value specified in Referenced Segment Number.</para>
          </listitem>
          <listitem>
            <para>The method of volume calculation is not described in this example.</para>
          </listitem>
        </orderedlist>
      </note>
      <table frame="box" label="XXX.1-1" rules="all" xml:id="table_XXX.1-1">
        <caption>Volumetric ROI on CT Example</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Node</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning of Concept Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning or Example Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>TID</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">Oncology Measurement Report</td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Language of Content Item and Descendants</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>English</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1204</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">1.2</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Observation Context</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1001</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Person Observer Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Doe^Jane</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1003</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Procedure Reported</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Chest+Abd CT W+WO contr IV</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurements</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Group</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Object1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Unique Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.276.0.7230010...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Referenced Segment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - Segmentation, Segment #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.6</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Volume</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>3267.46 mm3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
        </tbody>
      </table>
      
    </section>
    
    <section label="XXX.2" status="2" xml:id="sect_XXX.2">
      <title>Measurement Report SR Document Volumetric ROI on CT Example</title>
      <para>This CT example describes a set of measurements (volume. long axis and mean attenuation
        coefficient) made from a single volumetric ROI encoded as a single segment that spans two
        source CT images, and includes a description of the measurement methods and the finding
        site, as well as an image library to describe characteristics of the images used, and
        categorical observations at the measurement group and entire subject level.</para>
      <note>
        <orderedlist>
          <listitem>
            <para>For a different modality than CT, the choice of measurement for the mean intensity
              would not be (122713, DCM, "Attenuation Coefficient").</para>
          </listitem>
          <listitem>
            <para>For MR one might use (110852, DCM, "MR signal intensity"), or (110804, DCM, "T1
              Weighted MR Signal Intensity"), etc. See also CID 7180 "Abstract Multi-dimensional
              Image Model Component Semantics" for various appropriate signal intensity types for MR
              and other modalities.</para>
          </listitem>
          <listitem>
            <para>For PET one might use (110821, DCM, "Nuclear Medicine Tomographic Activity"), in
              which case the specific type of signal would be apparent from the units, e.g.
              ({SUVbw}g/ml, UCUM, "Standardized Uptake Value body weight") or for
              activity-concentration, (Bq/ml, UCUM, "Becquerels/milliliter"). See also CID 84 PET
              Units For Real World Value Mapping.</para>
          </listitem>
          <listitem>
            <para>Care should be taken when selecting modifiers such as (G-C036, SRT, "Measurement
              Method") versus (121401, DCM, "Derivation").</para>
          </listitem>
          <listitem>
            <para>The finding site and laterality within the measurement template (TID 1419) are
              factored out and shared by both measurements.</para>
          </listitem>
          <listitem>
            <para>The pattern used for the image library uses TID 4020, though commonality may be
              refactored (in a separate CP).</para>
          </listitem>
        </orderedlist>
      </note>
      <table frame="box" label="XXX.2-1" rules="all" xml:id="table_XXX.2-1">
        <caption>Volumetric ROI on CT Example</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Node</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning of Concept Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning or Example Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>TID</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">Oncology Measurement Report</td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Language of Content Item and Descendants</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>English</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1204</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">1.2</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Observation Context</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1001</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Person Observer Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Doe^Jane</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1003</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Procedure Reported</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Chest+Abd CT W+WO contr IV</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Image Library</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Study Date</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>20030417</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Study Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>104607</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Horizontal Pixel Spacing</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>0.810547 mm</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.j</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.n</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Pixel Data Columns</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>512 pixels</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.2.j</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 4020</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurements</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Group</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Object1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Unique Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.276.0.7230010...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Referenced Segment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - Segmentation, Segment #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.6</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Finding Site</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Adrenal Gland</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.6.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Laterality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Right</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.7</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Volume</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>3267.46 mm3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.7.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Sum of segmented voxel volumes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7474</para>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.8</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Long Axis</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>9.21 mm</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7470</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.8.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>RECIST 1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID cc1c8</para>
            </td>
          </tr>
          
          
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.9</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Attenuation Coefficient</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>70.978 Hounsfield unit</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.9.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derivation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Mean</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7464</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.10</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Necrosis</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Present</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.5.1.11</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Hemorrhage</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Absent</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.6</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Qualitative Evaluations</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.6.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Renal Vein Involvement</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Absent</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
        </tbody>
      </table>
      
    </section>

    <section label="XXX.3" status="2" xml:id="sect_XXX.3">
      <title>Measurement Report SR Document Planar ROI on DCE-MR Tracer Kinetic Model Example</title>
      <para>This DCE-MR example illustrates encoding measurements of mean and standard deviation
        Ktrans values in a planar ROI.</para>
      <note>
        <orderedlist>
          <listitem>
            <para>The measurement method and finding site and laterality within the measurement
              template (TID 1419) are factored out and shared by both measurements.</para>
          </listitem>
        </orderedlist>
      </note>
      <table frame="box" label="XXX.3-1" rules="all" xml:id="table_XXX.3-1">
        <caption>Planar ROI on DCE-MR Example</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Node</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning of Concept Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning or Example Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>TID</para>
              <para>CID</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">Oncology Measurement Report</td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Language of Content Item and Descendants</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>English</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1204</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>
                <emphasis role="bold">1.2</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Observation Context</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1001</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Person Observer Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Doe^Jane</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1003</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Procedure Reported</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Breast - bilateral MRI dynamic W contrast IV</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurements</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Group</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Object1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Unique Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.276.0.7230010...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Referenced Segment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - Segmentation, Segment #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - MR image #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Extended Tofts Model</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID cc2c2b in CP-1391</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.6</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Finding Site</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Breast</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.6.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Laterality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Right</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.7</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Ktrans</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>0.0185 /min</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.7.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derivation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Mean</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7464</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.8</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Ktrans</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>0.0102 /min</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.8.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derivation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Standard Deviation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7464</para>
            </td>
          </tr>
        </tbody>
      </table>
      
    </section>
    
    
    <section label="XXX.4" status="2" xml:id="sect_XXX.4">
      <title>Measurement Report SR Document Volumetric and SUV ROI on FDG PET Example</title>
      <para>This FDG PET example illustrates encoding measurements of various SUVbw related
        measurements.</para>
      <note>
        <orderedlist>
          <listitem>
            <para>The real world value map reference (for intensity, not size measurements) and
              finding site within the measurement template (TID 1419) are factored out and shared by
              measurements.</para>
          </listitem>
          <listitem>
            <para>The time point is described in this case only with a simple label.</para>
          </listitem>
        </orderedlist>
      </note>
      <table frame="box" label="XXX.4-1" rules="all" xml:id="table_XXX.4-1">
        <caption>SUV ROI on FDG PET Example</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para>Node</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning of Concept Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>Code Meaning or Example Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para>TID</para>
              <para>CID</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">Oncology Measurement Report</td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Language of Content Item and Descendants</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>English</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1204</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Observation Context</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1001</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Person Observer Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Doe^Jane</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1003</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Procedure Reported</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>PET/CT FDG imaging of whole body</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurements</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para/>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Group</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para></para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Liver</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Tracking Unique Identifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>1.2.276.0.7230010...</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Time Point</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TP0</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID t1tt3</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Referenced Segment</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - Segmentation, Segment #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.5</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - PET image #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.6</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Source image for segmentation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>IMAGE - CT image #1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1411</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.7</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Finding Site</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Liver</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.8</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Real World Value Map used for measurement</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>RWVM - UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.9</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>SUVbw</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>3.90557 {SUVbw}g/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.9.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derivation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Max</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7464</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.10</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>SUVbw</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>3.25653 {SUVbw}g/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.10.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derivation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Peak Value Within ROI</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7464</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.11</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>SUVbw</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>2.34467 {SUVbw}g/ml</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.11.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Derivation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Root Mean Square</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7464</para>
            </td>
          </tr>
          
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.12</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Standardized Added Metabolic Activity (SAM)</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>20400.3 g</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID cc1c3</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.12.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>SUV body weight calculation method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.13</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Volume</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>395512 mm3</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para>1.4.1.13.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Measurement Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>Sum of segmented voxel volumes</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para>TID 1419</para>
              <para>CID 7474</para>
            </td>
          </tr>
        </tbody>
      </table>
      
    </section>
    
  </chapter>