| DICOM PS3.18 2024e - Web Services |
|---|
Table J.2-1 specifies the Attributes of the Storage Commitment Response Module.
Table J.2-1. Storage Commitment Response Module
|
The SOP Instances for which storage has been committed. Required if the request payload contained the Referenced SOP Sequence (0008,1199), and there is at least one SOP Instance in that supplied sequence for which storage has been committed. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” in PS3.3 |
|||
|
The SOP Instances for which storage has not been committed. Required if the request payload contained the Referenced SOP Sequence (0008,1199), and there is at least one SOP Instance in that supplied sequence for which storage has not been committed. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” in PS3.3 |
|||
|
The reason that storage has not been committed for this SOP Instance. See Section C.14.1.1 in PS3.3 for possible values. |
|||
|
The Studies containing Instances for which storage has been committed. Required if the request payload contained the Referenced Study Sequence (0008,1110), and there is at least one SOP Instance for which storage has been committed |
|||
|
The Series containing Instances for which storage has been committed. |
|||
|
The SOP Instances for which storage has been committed grouped by SOP Class. |
|||
|
The Studies containing Instances for which storage has not been committed. Required if the request payload contained the Referenced Study Sequence (0008,1110), and there is at least one SOP Instance in that supplied sequence for which storage has not been committed. |
|||
|
The Series containing Instances for which storage has not been committed. |
|||
|
The SOP Instances for which storage has not been provided grouped by SOP Class. |
|||
|
The reason that storage has not been committed for this SOP Instance. |
|||
| DICOM PS3.18 2024e - Web Services |
|---|